The present invention is directed to measurement of an analyte, such as ethanol, and more particularly to non-invasive, in vivo optical measurement of such an analyte.
Blood alcohol content (BAC), also called blood alcohol concentration, blood ethanol concentration, or blood alcohol level, is most commonly used as a metric of alcohol intoxication for legal or medical purposes. Blood alcohol tests have a flaw in that they assume that the person being tested is average in various ways.
For example, on average the ratio of blood alcohol content to breath alcohol content (the partition ratio) is 2100 to 1. In other words, there are 2100 parts of alcohol in the blood for every part in the breath. However, the actual ratio in any given individual can vary from 1300:1 to 3100:1, or even more widely. This ratio varies not only from person to person, but within one person from moment to moment. Thus a person with a true blood alcohol level of 0.08 but a partition ratio of 1700:1 at the time of testing would have a 0.10 reading on a Breathalyzer calibrated for the average 2100:1 ratio.
A similar assumption is made in urinalysis. When urine is analyzed for alcohol, the assumption is that there are 1.3 parts of alcohol in the urine for every 1 part in the blood, even though the actual ratio can vary greatly.
Breath alcohol testing further assumes that the test is post-absorptive—that is, that the absorption of alcohol in the subject's body is complete. If the subject is still actively absorbing alcohol, their body has not reached a state of equilibrium where the concentration of alcohol is uniform throughout the body. Most forensic alcohol experts reject test results during this period as the amounts of alcohol in the breath will not accurately reflect a true concentration in the blood.
U.S. Patent Application Publication No. 2006/0002598 teaches a noninvasive alcohol sensor. An illumination subsystem provides light at discrete wavelengths to a skin site of an individual. A detection subsystem receives light scattered from the skin site. A computational unit is interfaced with the detection system. The computational unit has instructions for deriving a spatially distributed multispectral image from the received light at the discrete wavelengths. The computational unit also has instructions for comparing the derived multispectral image with a database of multispectral images to identify the individual.
The illumination subsystem may comprise a light source that provides the light to the plurality of discrete wavelengths and illumination optics to direct the light to the skin site. In some instances, a scanner mechanism may also be provided to scan the light in a specified pattern. The light source may comprise a plurality of quasi-monochromatic light sources, such as LEDs or laser diodes. Alternatively, the light source may comprise a broadband light source, such as an incandescent bulb or glowbar, and a filter disposed to filter light emitted from the broad band source. The filter may comprise a continuously variable filter in one embodiment. In some cases, the detection system may comprise a light detector, an optically dispersive element, and detection optics. The optically dispersive element is disposed to separate wavelength components of the received light, and the detection optics direct the received light to the light detector. In one embodiment, both the illumination and detection subsystems comprise a polarizer. The polarizers may be circular polarizers, linear polarizers, or a combination. In the case of linear polarizers, the polarizers may be substantially crossed relative to each other.
However, it would be desirable to provide a simpler and more compact way of achieving the same result.
It is therefore an object of the invention to provide such a simpler and more compact way for optical detection of an analyte such as ethanol.
To achieve the above and other objects, the present invention is directed to a technique called temperature-modulated spectrometry (TMS). The development of TMS included an investigation into the relationship between the spectral response of infrared light emitting diodes (IR-LEDs) and temperature. The TMS approach uses the active control of temperature to vary the spectral response of the IR-LED output, effectively sliding a spectral pulse across the ethanol sample, revealing the peaks and valleys of ethanol's spectral response. TMS can be used in any other field of endeavor using spectroscopic analysis.
Throughout the present specification, the term “LED” is meant to include other semiconductor emitters such as laser diodes, vertical cavity surface emitting lasers, and other such devices.
A simulation of the TMS approach was created using COTS IR-LEDs with a spectral response in the region of 2.1 μm-2.50 μm. The TMS approach will yield a low cost, compact system with very few parts (no moving parts). The estimated resulting unit cost is less than $50 per unit in low volume (thousands) and less than $15 in higher volumes.
Initial research was done to do first order approximations of the signal to noise ratio (SNR) of actual components, and to affirm the repeatability of the temperature modulation in the IR-LED. An experimental setup was constructed using IR-LEDS to show the peaks and valleys of the differentiated spectral response of various concentrations of ethanol (95%, 40%, 0.15% and 0.015%) and tap water. The IR-LED was modulated to sweep across 2.3 μm, 2.32 μm, 2.35 μm and 2.38 μm and clearly demonstrated that the measurement of ethanol, even at small levels, is possible with this approach.
This is a sensitive procedure requiring careful environmental control in order to get the precise measurement necessary for an accurate reading. However, even with a rudimentary experimental apparatus, the peaks and valleys in ethanol at 0.015% ethanol or 15 mg ethanol/di water are measurable.
The TMS approach disclosed herein effectively detects ethanol in mixture down to 0.015% (in tap water). This new spectroscopic approach has great promise.
A preferred embodiment of the present invention will be set forth in detail with reference to the drawings, in which:
A preferred embodiment will be set forth in detail with reference to the drawings, in which like reference numerals refer to like elements or steps throughout.
More specifically, in
The key insight behind the proposed solution is that LEDs and solid state lasers generate photons depending on temperature, which shift the spectral range and multitude. This dependence has led to dozens of patents and processes using temperature to stabilize the LED or IR laser's output. As temperature increases, the LED's overall wavelength distribution value shifts to longer wavelengths. Additionally, the overall efficiency of the radiation decreases, as shown in the spectral curves in
This procedure is specific to the experimental apparatus and data drivers used and was built to quickly establish the proof of concept for the proposed solution. Better designs would obviously be used for the final system, which would increase reliability. Multiple iterations of each experiment were conducted. Frequently gross errors in the procedure would occur requiring more measurements, (e.g. misalignment of items or condensation on elements). This report describes the final setup; earlier experiments were conducted, producing inconsistent results that lead to process and/or apparatus changes. Across all designs, more than 300 measurements were taken, leading to the 50 consistent and repeatable measurements shown in the final experimental results.
The 3 main hardware components to the proposed solution are the photodiode, the IR-LED, the peltier cooler and the drivers for these items. Three IR-LEDs, 1 IR photodiode and a peltier cooler were purchased to show feasibility for the proposed solution. LED23a was used for the measurements and has a spectral response centered at approximately 2.3 μm, shown in
As shown in
The temperature of the IR-LED must be controlled and measured for the experimental process. The LED was thermally bonded to the peltier cooler and the driver was used to adjust the temperature. The larger distances and masses involved in this indirect temperature management and measurement increases the potential for error and also varies the speed at which the temperature changes and measurements occur. The temperature and photodiode output voltage measurements were time stamped and aligned in software. As described below, the measured voltage and temperature sequences were then processed and analyzed using Matlab.
In this the final experiments measurements of Colorado Springs tap water, a 95% ethanol mixture, a 40% ethanol/60% water mixture, a 150 mg/dl (an approximately 0.15% ethanol mixture) and a 15 mg/dl (an approximately 0.015% ethanol mixture) were taken. At least 12 iterations for each mixture were processed.
The mixtures were mounted on quartz slides, with a bead of hot glue around the edge of the slide to provide for a reservoir holding the liquid to be measured. Path lengths through the liquid were much less than 1 mm and difficult to control. The experimental process adjusts for total response so variations in path-length are less critical to the results, except that sufficient material is necessary to ensure the system does not saturate the photodiode at the lowest temperature, but does not have too much material so that there is still measurable voltage at the highest temperatures. The initial setup is done so that at the start temperature the voltage, as seen visually on the photodiode controller, is in a specified range.
The core steps to the data acquisition and processing of the procedure are:
Step 1: Record the data
Step 2: Sort the data
Step 3: Extract useful information
Step 4: Calculate Absorbance
Step 5: Calculate Mean Absorbance
Step 6: Calculate Mean Absorbance Derivatives
Step 7: Subtract Water
Step 8: Normalize Data
Step 9: Calculate Derivatives
Step 1: Record the Data
The output voltage of the photodiode driver was measured and recorded using an oscilloscope. Because the IR-LED is being driven as a cyclic pattern, the output voltage of the photodiode measuring the light intensity is itself a wave pattern rather than a constant value. The information from the signal is located in the region of the peak of the photodiode voltage.
Step 2: Sort Data, Smooth Over Time and Temperature
Next the data from the oscilloscope, such as that shown in
Step 3: Extract Useful Information
Several problems were encountered with our simple approach to extracting useful information from the measurement. The peak of the voltage signal from the photodiode shifted between measurements and the oscilloscope, with 500 samples per cycle, measured much more frequently than the temperature sensor. This caused misalignments in the data. To address this, all measurements were time stamped, thus reducing the misalignment. There are 20-30 such single-pulse intensity estimations per temperature measurement. Each estimation of the spectral response of the medium is averaged per temperature measurement. This rate of sampling is sufficient to keep the SNR low enough.
Step 4: Calculate Absorbance
By employing an approximation of Beer-Lambert, absorbance is approximated from the smoothed intensity measurement (I) of ethanol, water or a mixture of water and ethanol, divided by the calibration phase smoothed average intensity of air (I0) at the same temperature. This calculation gives the absorbance of the medium at each temperature recorded as:
A
λ=−log10(I/I0).
Only one photodiode was being used in this experiment. This gave rise to more error because the normalization process uses measurements of air done at a different time and if there are variations in the LED output (e.g., because of current or temperature lag) these differences increase the noise in absorbance computation. This could be overcome by using beam splitters and multiple photodiodes, one for air and one for the medium, allowing the normalization to be on the actual IRLED output.
Step 5: Calculate Mean
The mean and standard deviations are then calculated from the smoothed absorbance calculations in
The ethanol curve shows dear peaks and valleys and the standard deviations are small enough to see that the sizes of the peaks and valleys are well within our measurement tolerance. While water shows some inflections, there are no peaks/valleys in the region of interest.
Step 6: Calculate Derivatives
The derivatives are then calculated using the values obtained from the smooth mean from Step 5 (
Step 7 and Step 8: Dealing with Mixtures: Subtract Water and Normalize
When combining mixtures of material the resulting absorbance will be a combination of the underlying materials. For low ethanol mixtures, there will be no visible peaks and valleys because increasing proportion of water will dominate the overall absorbance. Normalizing, by subtracting out the response of water from the measurements, addresses this issue. (For a final system this would probably involve subtracting out skin response, and it will be important to determine if it must be person specific). This is followed by a normalization process, mapping the resulting data to [0, 1]. The 40/60 ethanol mixture still looks a lot like ethanol and has very clear peaks/valley of ethanol, though somewhat shifted. The 0.015% and 0.15% have similar shapes, but are still moderately different from ethanol, possibly from differences in measurements (or LED output) during the measurement of water (which is subtracted). Rather than just subtracting water, a better approach may be to model the overall data as: y=x(CH3CH2OH), solving for y in the least squares sense over the database. Still these results are promising even in concentrations as small as 15 mg/dl (shown in bottom left) and consistent with the results from simulations: see
Step 9: Calculate Derivatives on Mixtures
By taking the derivative of the normalized absorbance the peaks and valleys become more apparent. The derivatives of the normalized data show how the peaks and valleys of ethanol stand out when the absorbance of water is subtracted from the measurement for the mixtures in 15 mg/dl and 150 mg/dl. The derivatives for the normalized data change the scale as a result of the normalization that occurs. This is why the raw derivative for ethanol is smaller than those for the normalized mixtures.
These results are promising but not without some errors, as there are two added zero crossings and some shifting of the zero with respect to ethanol. For idealized data, with subtraction of water, the shifting should be smaller and there should not be significant added zero crossings. Errors in temperature measurement, relative to the actual LED temperature, would cause localized shifts and produce the increasing slope of water, and decreasing slope of ethanol, could introduce the spurious zeros.
Specific Difficulties in Measurements
Along the path to the above experimental analysis, hundreds of experimental trials were attempted sometimes showing inconsistent measurements. As the experimental process was refined the following issues significantly impacted measurements and experimental repeatability. Some of these were oversights, e.g., the IR absorption of the Petri dishes and beakers, and easily removed once identified. Others were more problematic.
Condensation and Lens Distortion
Condensation and lens distortion were two factors played the most significant roles with regard to the problems encountered with repeatability and incorrect readings. When the LED is made colder than the dew-point condensation, and sometimes ice crystals, would form on the lens. Light would pass through the water or ice crystals and cause erroneous intensity measurements. One possible way to avoid this would be to use a LED with the response in higher temperatures, thus discouraging condensation. Others would include having the TEC inside a sealed dry environment. The experimental setup was improved by adding an AC unit to the darkroom and reducing the ambient temperature and humidity to reduce the potential for condensation.
Alignment and Intensity
Alignment and intensity is critical for successful measurement of ethanol. The photodiode amplifier may over-saturate the signal if the intensity is too high. If oversaturation of the amplifier occurs, the changes in the temperature sweep will still occur but the change in intensity will be lost. In particular air requires lower intensity than water and ethanol, and variations in the output/measurement caused by differential driving increase the experimental error. The initial experimental setup used alignment of the photodiode to adjust the intensity value. During this process a slight movement in the alignment of the photodiode would change the measurement drastically and hinder repeatability. However, once proper alignment and intensity was achieved (as long as the experimental setup was not modified) the results were repeatable. The values displayed on the photodiode driver as the temperature increased correlated to the values as the temperature decreased.
Temperature Control and Temperature Measurement
The temperature measurement and control for this proof of concept is relatively weak and an area where there can be substantial improvement. Because of the high level of dependence on temperature, measuring the temperature change accurately is critical. The most significant issue for our measurements was the fact that the temperature sensor was measuring the temperature on peltier and was on the outside of the LED case, rather than on the LED itself. Thus, the temperature of the LED could be different than the change in the external temperature, and the difference could be temporally varying causing more error.
A secondary advantage of directly cooling the LED is that cooling it directly would reduce the thermal mass that needs to be adjusted, increasing the speed at which the temperature sweeps could be conducted. Another issue is that our current temperature sensor sampled at 1 Hz and frequently dropped packets causing more error in the measurement. The sensor measures temperature at 0.1 C resolution, which may be sufficient to show difference in peaks and valleys for ethanol, but limits alignment in the case of subtractive normalization needed for mixtures.
Experimental Conclusions
The experimental process in research is often a meandering process, full of unexpected obstacles to be overcome. This experimental development had many such side turns, but in the end developed a protocol that shows strong promise for alcohol measurements in the ranges needed. Repeatable measurements that detect peaks/valleys from alcohol at 0.015% were obtained using TMS with a rudimentary apparatus built with plywood frame, quarts slides and plenty of hot glue. Many engineering challenges remain to prove it in actual subjects and design a better device/process to reduce errors, but these experiments show basic feasibility of the concept.
Simulation modeling and results will now be discussed.
There are two parts to the simulation. The first part was completely theoretical, and preceded all experiments, the second part was to relate back the experimental data and allow more what-if based on the theory.
All of the spectral responses are interpolated from manufacture datasheets. In the first part of the simulation GaSb substrates are assumed and it was further assumed that their spectral responses are temperature dependent: As temperature increases their peak value shifts and their efficiency decreases.
This change was modeled as a linear shift and scale. The shift and scale for an emitter with a peak located at 2350 nm is shown in
The simulated method is a voltage sweep from 5 C to 50 C. As the temperature changes the emitter's response becomes more or less efficient depending on a decrease or increase of temperature. This change causes the spectral response to move left or right, sweeping across a portion of the medium's spectral response. The change in the emitter can be divided out, using the Beer-Lambert law, leaving the response in transmittance or absorbance of peaks and valleys in a given medium. This process is simulated with data from two actual emitters: LED 23 and LED 22 and modified data from a hypothetical emitter: LED 22—modified. The targeted peaks and valleys of ethanol in the region of interest do not have a COTS IR-LED available giving the shift needed. Feasibility for this region was shown by gene rating a specification for a desired emitter and then simulated. By using the spectral data on the datasheet a curve was interpolated. This curve is then shifted right and scaled down slightly as the temperature is increased.
The theoretical spectral response of ethanol is shown in
The second part of the simulation combined actual data for pure ethanol and water with simulated mixtures of the two. This provided a quantifiable measure to determine if actual tests are consistent. Once the baseline measurements for 95% ethanol and water were established theoretical measurements of several different mixtures were made. This data showed that water must be subtracted in order to see the effects ethanol was having on the mixture. While water made the most sense to be used in this ease any base-line spectral measurement, such as human skin, could be used. The graphically displayed results are shown below.
Simulation Results: Part 1
The first part of the simulation provided intuition to the region being scanned and helped us determine if the process was feasible and worth doing physical experiments. The initial work used models which were somewhat inaccurate but still showed the potential of the TMS approach, and justified using TMS in the actual experiments.
Simulation Results: Part 2 Relating Back to Real Data
The second state of simulation was feeding back data from the real measurements to help analyze SNR and determine if a viable approach existed. This helped develop the algorithms and address ideas on how to deal with calibration. The simulation graphs are slightly different from the actual data, e.g. the water measured had more slowly increasing absorption, but in general, results were consistent with the actual data.
This part of the simulation now follows Steps 5 and 7 in the actual procedure: Subtract the absorbance response of water from the measured absorbance of a mixture and find the derivative of the differenced value. The simulation suggests the location of the peaks and valleys of the mixtures should be closer to the pure ethanol than our measured value, but overall there is good agreement. In particular, between 15 C and 20 C.
Background on the LED will be provided.
An ideal LED should have a high radiance (light intensity), fast response time and high quantum efficiency. These characteristics are best achieved via double hetero-structure devices. A double heterostructure semiconductor device has junctions between different band-gap materials. It is important that the region in which recombination occurs there is a high carrier concentration. The double heterostructure enables the source radiation to be much better defined, but further, the optical power generated per unit volume is much greater as well. When the structure is connected the Fermi level must remain constant at thermal equilibrium. Because the middle p-layer is smaller in band gap than the other two layers, when the structure is forward biased electrons would flow to the middle p region but would be confined in that region, since there is a potential barrier due to the difference in band gap, limiting them from diffusing further in the adjacent p region. When the electron combines with a hole from the other side of the gap a photon is created. The energy of the photon is a function of the separation energy between the electron and the hole.
By keeping the middle layer extremely small (−0.1 μm) the emitted photon can be confined to a very small area and photons generated in other layers cannot be absorbed since it will have a different energy value than the band gap of the middle layer.
Emitted wavelength depends on band-gap energy. The order of increasing voltages is the order of increasing energy required for emitting light from the LED. The wavelength of light emitted depends on the band gap energy, depending on how strongly the bonding electrons are held in localized, depending on the size of the atom, some small atoms hold their electrons more tightly.
Temperature Dependence
Temperature dependence of the LED is critical to our TMS concept. As the temperature increases the diode gain decreases and more current is required to overcome the losses and produce forward bias. By increasing the temperature, the energy of the electrons and holes increases allowing more to be free outside the active layer (in the n and p layers). More recombination happens outside the active layer with free carriers that would have reached the active layer but recombine instead.
Higher energy band-gap means longer wavelength:
λd(FinalJunctionTemperature)=λd(InitialJunctionTemperature)+ΔT*γ(nm/° C.)
In other words for every change in junction temperature there is a change of W nm in dominant wavelength λd, where γ is a function of materials used and geometry of device. Usually this is considered a problem to be controlled, but in our case it allows us to sweep the wavelengths of the LED output over a range.
Reflection/Refraction is temperature dependent as it changes the effective index of refraction of the material compared to air. The impact is greater at larger angles of incidence and it shifts the angle at which TIR occurs. This impacts both optical spreading of light emitted and spectral spreading/shifting of light emitted. This is a much smaller effect, mostly impacting the wavelength spread, allowing edge-emitting LEDs to maintain a narrower beam with reduced wavelength spreading.
The radiation emerges in the direction perpendicular to the junction plane for SLED's, so SLEDs emit light over a wide area giving a wide far-field angle. SLEDs are more commonly used in communication as they support a more efficient coupling to the optical fiber than edge emitting LEDs. A SLED has an active region (the part of the LED actually emitting photons) from 20 μm to 50 μm.
This is in contrast to the ELED. The primary active region of an ELED is a narrow stripe, which lies below the semiconductor substrate. The substrate is then cut or polished so that the stripe is runs between the front and back of the device. These polished surfaces are called facets. The rear facet is highly reflective and the front facet is antireflection-coated. The rear facet reflects the light so it all travels out the front of the LED. ELEDs emit light in a narrow emission angle resulting in a narrower spectral line width and are typically more sensitive to temperature fluctuations than SLEDs.
Because the TMS approach is sweeping the pulse across the sample, a narrow sample beam would help and so if possible, an ELED would be preferred.
A part of the effort was to assess the measurement feasibility of ethanol (CH3CH2OH) at various skin sites to determine the blood alcohol concentration of a subject (SAC). The spectral response region of ethanol to be investigated is in the 2.2-2.4 μm and possibly 1.6-1.8 μm, both areas are highlighted in
Aside from plasma, which contains about 90% water, red blood cells are the most predominant structure found in blood. The oxygen carrying protein in the red blood cells is called hemoglobin. The chemical formula for hemoglobin is C34H32FeN4O4. The chemical structure of hemoglobin is shown in
Regardless of the site at which ethanol levels are measured, light will have to come in contact with skin. Along with differences in concentration of melanin various skin conditions need to be examined as well as the changes of the spectral absorptivity of skin due to temperature changes. Ideally the ethanol measurement device will function uniformly throughout the range of skin colors and conditions. While many studies exist on the spectral properties of human tissue, few exist which discuss the absorption or transmission of skin at wavelengths longer than 2.4 μm. A study was done on the differing spectral characteristics of skin color.
The next factor to be considered when light enters the skin is the scattering of light. The spatial distribution and intensity of scatter light depends upon the size and shape of the inhomogeneities relative to the wavelength, and upon the difference in refractive index between the medium and the inhomogeneities. By having small variation of tissue present in the measurement sample, less scattering can be expected, producing a more accurate measurement of ethanol. This guides the measurement to an area with relatively little muscle or other interfering tissue, namely the fingers and upper palm. The Kubelka-Munk theory can offer a simple quantitative treatment of the optics of the skin.
Some nonmelanoma skin cancers have shown statistically different scattering at the wavelengths between 1.05 μm to 1.4 μm. Another study on the blanching of skin also found that the spectral response of skin seemed to converge at the longer wavelengths. Some of these results are shown in
Multiple types of sensors can be used, including:
Examining the above list it should be clear that all of these except DIDC, and possibly DICS, require a dispersive element to spread the wavelengths for measurement. Thus looking at approaches for low-cost dispersive elements was the first element we considered. The dispersion or diffraction is only controllable if the light is collimated, that is if all the rays of light are parallel, or practically so. A source, like the sun, which is very far away, provides collimated light. In a practical monochromator however, the light source is close by, and an optical system in the monochromator converts the diverging light of the source to collimated light. There are two major choices for the monochromatic in dispersive elements, ones that use focusing gratings that do not need separate collimator, and the more common (and cheaper) use collimating mirrors. Reflective optics are preferred because they do not introduce dispersive effects of their own. The most common approach is the use of a Czerny-Turner monochromator, which greatly reduce the total size at only a modest increase in cost.
To deliver light to a sensor, we analyzed one of the most widely used compact dispersive elements, the classical asymmetrical Czerny-Turner optical bench. One of the possible configurations is shown in
The emerging use of nanowires for narrow band-pass optical filters became another topic of interest. The basic concept is that a very fine array of nanowires produces a localized surface plasma resonance. The resonance produces a very narrow band-pass filter, often with half-widths around 50 m and peak-to-peak separations of 10 nm in the visible range. Early nanowires were done for polarization techniques, but required Electron Beam Lithography making them expensive. More recently, the process of production of nanowires has been demonstrated at the University of Pittsburgh, using a much lower cost imprinting technique; potentially allow direct application on sensor chips. They have licensed the technology and prototypes for the visible spectrum are now in production by Nanolambda, though yields on the imprinting process are still pretty low. Similar to a Bayer pattern, with this concept placement of a wide range of narrowband filters directly on the CMOS sensor can give a similar effect. The concept extends down to the 2000-3000 nm range, thought the half-widths get broader because of the constraints on the nanowires needed to sustain the surface plasma field. This approach has not been demonstrated physically in the wavelength of interest, but it should apply.
One concern for the nanowire approach is the cost of the imaging array. The imaging array may be why HP never pushed once they looked into it. While CMOS imagers are particularly cheap, the material and lower volume in the spectral range needed for our application will make the nanofilter+area sensing approach more expensive.
A mixed approach using a collection of the nano-filters combined with the TMS approach with a small number of photo-detectors is also a possible area for feasibility. This approach is different from the standard filter usage for broader spectrum measurements which involve spatially separating the measurements of incoming light. This approach may increase the efficacy of TMS by sweeping in temperature and time and so modulating the sensor response or LED output around a few regions of interest then effectively measuring the sum of them. The concern of light levels is still an issue, as these filters will reject most of the light impinging on them, so covering the photodetector with an array of these filters would significantly reduce overall efficiency of the sensor. An unexplored solution may be placing the filters on the cover glass of the LED, which may be more effective as the rejected photons can be reflected down ward back into the reflector surrounding the LED and eventually have a chance of emitting through the small part of the filter that is appropriate for it.
The most significant errors are most likely caused by the variation in responses and need to calibrate to absorption (dividing by the response in air) and dividing by water since both of these are using measurements from the LED at a different point in time. A beam-splitting approach with multiple photodiodes would increase costs but have the potential to simplify these issues.
The second potential major source of error is the limited temperature control and measurement. A diode with a built-in peltier cooler (see above) would help this situation, as would a more controlled test environment. Cooling the diode using gas also seems to have a residual effect on warm-up rate.
The third major source of error is the variations in physical measurements caused by the limited experimental setup. A vertical or sealed testing apparatus would solve that problem, as the path length would be the same between measurements. Reflective measurements would also solve the problem.
There are multiple issues in the current setup that can be corrected for liquid measurements and others for reflective measurements.
In summation,
The spectroscopic analysis subsystem 208 analyzes the detection signal to detect the peaks and valleys corresponding to the known spectroscopic peaks and valleys of the analyte, e.g., ethanol. By detecting and measuring the peaks and valleys, the spectroscopic analysis subsystem can determine both the presence and the concentration of the analyte. In the example of ethanol, the spectroscopic analysis subsystem can determine the presence and concentration of the ethanol and use that information to make an ultimate determination such as blood alcohol content.
While a preferred embodiment has been set forth in detail above, those skilled in the art who have reviewed the present disclosure will readily appreciate that other embodiments can be realized within the scope of the invention. Therefore, the invention should be construed as limited only by the appended claims.
The present application claims the benefit of U.S. Provisional Patent Application No. 61/362,914, filed Jul. 9, 2010. Related information is disclosed in U.S. Provisional Patent Application No. 61/362,922, filed Jul. 9, 2010. The disclosures of the above applications are hereby incorporated by reference in their entireties into the present disclosure.
Filing Document | Filing Date | Country | Kind | 371c Date |
---|---|---|---|---|
PCT/US11/43549 | 7/11/2011 | WO | 00 | 12/9/2013 |
Number | Date | Country | |
---|---|---|---|
61362914 | Jul 2010 | US |