The present invention relates to medical applications of ultrasound and in particular, to ultrasound for cardiac ablation.
Cardiac arrhythmias are characterized by erratic cardiac contractions. These erratic contractions are often confined to areas of the cardiac muscle that have abnormal electric conduction, refractoriness, or impulse formation. These abnormalities disturb the normal propagation of the electric signals through the muscle resulting in abnormal muscle contraction. A variety of surgical and non-surgical treatments are available for cardiac arrhythmias. The non-surgical treatments are anti-arrhythmetic drugs designed to alter the electrophysiologic properties of the cardiac tissue. Though these drugs decrease the likelihood that an arrhythmia will occur, their efficacy is limited. Moreover, they have potentially fatal side effects. For these reasons, pharmacological approaches for treating cardiac arrhythmia have been widely supplanted by surgical approaches that irreversibly damage or ablate the tissue regions that cause and sustain the arrhythmias.
Over the past two decades open-heart ablative surgery has been replaced by catheter cardiac ablation, a minimally-invasive procedure in which a catheter is inserted transcutaneously into an artery or a vein and guided fluoroscopically to the heart. The catheter delivers energy to the problematic site. This energy heats the arrhythmogenic tissue until it coagulates, thus destroying the tissue. Radio frequency (RF) energy is most commonly used, though a variety of energy sources, including direct currents, microwaves, cryothermic sources, and lasers can be used for ablating tissue.
Although catheter ablation has become a standard form of treatment, it has major limitations in both efficacy and safety. Conventional catheter cardiac ablation techniques are limited in their ability to accurately identify arrhythmogenic tissue. In addition, catheter cardiac ablation has great difficulty producing deep transmural, continuous lesions. The invasive nature of catheter cardiac ablation can lead to significant complications including severe pain, adverse drug reaction from anesthesia, infection, thrombophlebitis, myocardial infarction, perforation, hemopericardium, and cardiac tamponade that can ultimately prove fatal. Furthermore, catheter cardiac ablation is performed under fluoroscopy guidance, a procedure that emits ionizing radiation. If the procedure is prolonged, the patient and the physician are at risk for sustaining a hazardous level of exposure.
In an aspect, the invention features a system for performing ablation of target tissue. The system includes an ultrasound phased array having a plurality of ultrasonic transducers and drive circuitry coupled to the ultrasonic transducers. The drive circuitry is configured to generate signals that cause the ultrasonic transducers to focus ultrasound radiation at the target tissue. The system also includes a diagnostic system configured to percutaneously collect diagnostic data that is indicative of a condition of the target tissue and computational circuitry that is interfaced to the drive circuitry and to the diagnostic system. The computational circuitry is configured to control the drive circuitry based on the diagnostic data.
In some embodiments, the ultrasound phased array is two-dimensional and configured for placement in a patient's esophagus. In some embodiments, the drive circuitry includes multi-channel radio-frequency drivers and the diagnostic system includes an imaging system for producing an image of the target tissue. Examples of the imaging system include a magnetic resonance imaging (MRI) system, an ultrasound imaging system, a computed tomography imaging system, an x-ray imaging system, and a positron-emission tomography imaging system. In some embodiments the diagnostic system includes a temperature monitoring system which may, for example, include an MRI system, and is configured to collect the diagnostic data in real-time.
In another aspect, the invention features methods and computer readable mediums for performing ablation of target tissue. The method includes identifying a target location of the target tissue from an image of the target tissue; focusing ultrasound radiation from an ultrasound phased array at the target location; collecting diagnostic data percutaneously, the diagnostic data being indicative of a condition of the target tissue; and controlling a characteristic of the ultrasound radiation (e.g., phase, frequency, and power) based on the diagnostic data such that the ultrasound radiation ablates the target tissue without damaging surrounding tissue. The computer readable medium includes instructions for performing the method.
In some embodiments, the target location is determined in relation to a periodic triggering event (e.g., a heartbeat) and the ultrasound radiation is focused at the target location for a predefined period of time in response to detecting the triggering event. In other embodiments the target location is determined in real-time. In some embodiments the coordinates of the image are transformed to coordinates of the ultrasound phased array. In some embodiments collecting diagnostic data also includes acquiring temperature data that is indicative of ablation.
The details of one or more embodiments of the invention are set forth in the accompanying drawings and the description below. Other features, objects, and advantages of the invention will be apparent from the description and drawings, and from the claims.
The MRI diagnostic system 26 collects diagnostic data 10 from the target tissue 24 and the surrounding anatomy 8 of the patient. The diagnostic data is collected percutanously, i.e., without puncturing or breaking the skin. In one implementation, the MRI diagnostic system 26 acquires an image on a line-by-line basis using fast line-scan MRI. The fast line-scan MRI technique enables the image data in a line to be analyzed in real-time as the line is scanned without waiting for the whole scan to finish. The MRI diagnostic system 26 processes the diagnostic data 10 into a series of images and temperature measurements that collectively comprise MRI data 28. This MRI data 28, and optimally, ECG data 16, is sent to a computer system 30.
From the MRI data 28 and the ECG data 16, the computer system 30 identifies the location of the target tissue 24 as a function of time. The computer system 30 could also identify the location of the target tissue in real-time. The computer system 30 calculates the radio-frequency (RF) signals 18 required to focus the ultrasonic radiation 22 at the target-tissue location and sends information representative of those RF signals 18 as control signals 32 to the drive circuitry 34. In response to the control signals 32, the drive circuitry 34 generates radio frequency (RF) signals 18 that cause the ultrasound array 20 to focus ultrasonic radiation 22 at the target-tissue location. While the tissue is undergoing sonication, the MRI diagnostic system 26 sends a stream of MRI data 28 to the computer system 30. Using the MRI Data 28, the computer system 30 monitors the movement of the target tissue, tracks the progress of the sonication, and determines when the ablation is complete. In addition to MRI, other imaging techniques such as ultrasound echo imaging could be used to track the target-tissue movement and monitor the sonication.
Referring to
As seen in
The ablation process begins when the coordinate registration function 120 registers the coordinates of the ultrasound array 20 with the coordinates of the MRI data 28. Then, the target-tissue identification function 122 identifies the image space coordinates of the target tissue. These coordinates are translated from image space 264 to array space 260 (see
In
The procedures for performing continuous ablation and periodic ablation are not limited to those illustrated in
The feasibility of transesophageal cardiac thermal ablation using a planar ultrasound two-dimensional phased array was investigated in computer simulation studies.
Summary of Results
The results of the study showed that by varying sonication duration and power, the array can produce controllable tissue coagulation without damage to the overlaying or surrounding tissues. The array modeled in the studies was a two-dimensional planar ultrasound phased array having a 1 MHz output frequency, dimensions of 60×10 mm2, 0.525 mm inter-element spacing, and 114×20 transducer elements. Using electronic beam steering, three groups of foci (total 39 foci) in cardiac muscle were defined at short, medium and long (20, 40 and 60 mm) radial distances from the transducer surface and at different steering angles from the transducer radial axis. A full range of ultrasound pressure distribution in a volume of 60×80×80 mm3, including esophageal wall, was calculated using a multilayer acoustic wave transmission model for each focus. The corresponding thermal effect in both esophageal wall and cardiac tissue due to the acoustic energy absorption was simulated using the bioheat transfer equation. For short, medium and long (1-, 10-, and 20-second sonications that did not produce thermal lesions in the esophageal wall, the acoustic power ranges needed to achieve a 60° C. maximum temperature in cardiac muscle were 105 W to 727 W, 28 W tol 17 W, 21 W to 79 W, respectively. Similarly, for the same sonications, the acoustic power ranges needed to achieve a 70° C. maximum temperature in cardiac muscle were 151 W to 1044 W, 40 W to 167 W, and 30 W tol 14 W, respectively. A thermal dose in equivalent minutes at 43° C. (denoted T43) was applied to the foci for at least 240 min. The resulting tissue lesion lengths at these foci were 1-6, 3-11, 3-13 mm and 3-15, 5-19, 6-23 mm, respectively. The lesion widths were 1-4, 2-7, 3-9 mm and 3-9, 4-13, 4-17 mm, respectively. The following text describes the studies in greater detail.
Simulation Configuration
The tissue lesions produced by various transesophageal ultrasound fields under different sonication powers and durations were simulated on a computer. The two-dimensional planar ultrasound phased array that was modeled in the studies had a length of 60 mm, a width of 10 mm, produced ultrasound at 1 MHz, and included 2280 transducers having a 0.525 mm inter-element center-to-center spacing arranged as a 1 14×20 grid such that the diagonal and maximum transducer array inter-element center-to-center spacing was less than half a wavelength (0.75 mm at 1 MHz). The planar phased array was capable of steering the beam by proper element phasing and amplitude weighting with respect to the distances from the elements to the focus. The full steering functionality enabled the array to aim the beam and track cardiac tissue motion during sonication. The planar phased array was also modeled to be positioned inside an esophagus, facing the heart, with water filling the space between the transducer and the esophageal wall. The simulation assumed a 4.4 mm esophageal wall thickness and defined the inner and outer surfaces of the esophageal wall as being the surfaces that were in contact with water and cardiac muscle. Each of the surfaces of the esophageal wall was interpolated from seven arcs that were evenly distributed along z-axis. Each arc was constructed as either a half circle having a 10 mm radius for the outer surface or a 5.6 mm radius for the inner surface and being centered on the z-axis. Sixty evenly spaced points were randomly selected along the arc such that the distances from the points to the z-axis varied randomly by a distance up to ±2 mm. The variation of the points deformed the half-circle arc to an irregular arc centered at the z-axis, which was constructed by a spline interpolation on the points. The whole outer surface was then linearly interpolated from the seven smooth arcs centered at the z-axis. In this way, up to 2 mm curvature variations were added onto the esophageal wall surfaces to approximate the uneven surfaces.
The simulation evaluated the near field heating in the esophageal wall that was caused by absorbed acoustic intensity. To accomplish this, a full range of acoustic pressure field distributions were calculated in a three-dimensional orthogonal grid of field points for each focus. The pressure field distribution calculated by the simulation spanned from −40 to 40 mm in the x-axis, 0 to 80 mm in the y-axis, and −40 to 40 mm in the z-axis. The same three-dimensional grid was also used in a finite difference thermal simulation.
Transesophageal Acoustic Field Calculation
Continuous-wave sonications were modeled in the computer simulation studies. The transesophageal ultrasound pressure fields in cardiac muscle were calculated with a multilayer acoustic wave transmission model that considered both attenuation in a tissue layer and refraction at a curved tissue layer interface. In this model, a tissue layer interface was partitioned into planar rectangular mesh patches that were small enough (about a quarter wavelength in dimension) to be treated as simple sources. The working variables were the particle normal velocities on these patches. The particle normal velocity at a field point in front of a tissue layer interface was calculated using a Rayleigh-Sommerfeld surface integral over the simple sources on the tissue interface. Each simple source was assumed to be only radiating in its forward half space to model the ultrasound non-illuminating area obstructed by tissue geometry. The refracted particle normal velocities at each tissue layer interface were approximated using Snell's law on the planar patches. To simulate the non-transmitting situation, a total possible reflection was considered by calculating the incident angle of each acoustic beam from any simple source to the current patch. For a multilayer problem, the propagation-refraction processes at the multiple interfaces cascade one layer after another to produce the transmitted acoustic field from the curved tissue layers.
The transducer surface was treated as a simple acoustic source interface that radiated acoustic waves. The waves then propagated through the water layer, the inner esophageal wall, the esophagus layer and the outer esophageal wall and into the cardiac muscle. The tissue interface partitioning mesh size was 0.5×0.5 mm2 in the study. The acoustic properties of the media are listed below in Table 2.
The ratio of the acoustic pressure to the associated normal particle velocity in a medium is the specific acoustic impedance of the medium. Ignoring non-linearity, the acoustic pressure in a small (about a quarter wavelength in dimension) patch can be obtained from the product of the particle normal velocity and the specific acoustic impedance of the tissue. A reported human esophagus speed of sound measurement could not be found in published literature. Therefore, the speed-of-sound value corresponds to a speed-of-sound measurement of a pig esophagus sample using a scanning laser acoustic microscope.
Tissue Coagulation Simulation
The temporal profile of tissue temperature spatial distribution (T(x, y, z, t)) during ultrasound sonication was modeled using the Pennes bioheat transfer equation:
where ρt, Ct and kt are the density, specific heat capacity, and thermal conductivity of the tissue, Cb and W are the specific heat capacity and the perfusion rate of the blood. The variables a, c, p(x, y, z) represent acoustic pressure attenuation, the speed of sound of the tissue, and the acoustic pressure amplitude in the tissue. The variable Ta is the body temperature (37° C.). The first, second, and third term in the right hand side of Equation 1 simulate the heat conduction in tissue, heat loss due to blood perfusion and energy absorption from the acoustic field, respectively. The third term is the specific absorption rate (SAR) as a measure of energy absorption rate from external energy sources. The pressure amplitude p(x, y, z) was calculated in the ultrasound field simulation. The thermal properties of the media are listed in Table 2.
The temperature profile (T(x,y,z,t)) obtained by solving Equation 1 was then mapped to a thermal dose in equivalent minutes at 43° C. (T43 (x, y, z)) using Sapareto and Dewey's thermal dose function expressed as:
Tissue was considered necrotic when T43(x,y,z) exceeded 240 minutes in the simulation volume.
Numerical Implementation
The acoustic pressure amplitude spatial distribution (p(x, y, z)) in a volume of 60×80×80 mm was calculated for each of the 39 foci using the multilayer acoustic wave transmission model. The pressure distribution in the volume was sampled with a three-dimensional rectangular grid of field points. To determine the array element phase necessary for forward transesophageal beam steering, a reverse transesophageal propagation process, in which a point source was radiating at the desired focus, was first simulated using the multilayer transmission model to obtain the reverse complex pressure at each element. The conjugated phase of the reverse complex pressure was fed to each element in the forward transesophageal propagation. This ensured the necessary phase compensation for both beam steering and phase aberration correction caused by esophageal wall. The source intensity of each element was weighted by the distance between the element and the focus so that the propagated wave amplitude at the focus from all elements would be the same.
Because the transducer array had a 60 mm length along the z-axis and a 10 mm width along the x-axis, the beam width along the z-axis was narrower than that along the x-axis. The spatial grid spacing had to be very small to accurately capture the beam pressure amplitude profile along the z-axis. Furthermore, the pressure distribution (p(x, y, z)) was used to calculate the specific absorption rate (SAR=a|p(x, y, z)|2/ρC) in thermal simulation. Therefore, the narrow beam width in z-axis also facilitated the use of a fine grid in the thermal simulation domain to ensure spatial convergence of T(x, y, z, t) using a finite difference scheme. However, a fine spatial sampling of the pressure amplitude would have led to an expensive calculation cost to obtain the p(x, y, z) in the given volume. To reduce the computational cost and yet to maintain a fine resolution of acoustic pressure field for the following thermal simulation, a non-uniform three-dimensional grid was used in calculating the p(x, y, z) in the given volume, with the smallest spacing in the focal region and the largest spacing in the marginal region outside of the beam. Several small spatial grid spacings in the focal region (0.5 mm, 0.25 mm and 0.125 mm) were examined in the thermal simulation to evaluate spatial convergence of the calculated temperature field.
The bioheat equation (Equation 1) was solved using a finite difference scheme in Cartesian coordinates to obtain T(x, y, z, t). Because of the minor differences in their values, the inhomogeneity of thermal conductivity kt in different media was ignored for implementation simplicity. Due to the non-uniform grid, a modified spatial derivative operator using a central differencing scheme was adopted instead of its conventional counterpart for uniform grid. The resulting discrete equation of Equation 1 is
with the discrete operators P(·), Q(·) and R(·) being defined as:
where n is the discrete time step, i, j, k are the indices for the nonuniform grids in the x-, y- and z-axes and take all the integer values between the second and the second to the last indices. The specific-absorption rate (SAR) term in Equation 3 used an average value in the voxel by taking arithmetic mean among the |pi,j,k|2 an d its neighboring values. As a reasonable approximation when dealing with a large volume in which the thermal source was far away from the boundaries, Neumann boundary conditions
were set on the tissue volume surfaces. The temporal derivative was implemented using a forward differencing scheme in Equation 3. Corresponding with the smallest spatial grid spacing aforementioned, the time step size was chosen as 0.05 s for a stable finite difference thermal simulation.
Before each sonication, a pre-cooling phase was used to lower the initial temperature of the esophageal wall and to reduce the risk of thermal damage in esophagus. In the pre-cooling phase, 20° C. degassed water was filled in between the transducer and the esophageal wall as coupling medium. The initial water temperature and the tissue temperature were 20° C. and 37° C., after which the temperature field evolved to its steady state without external sonication (SAR=0 in Equation 1). Approximately 190 seconds later, the water-tissue system reached its steady condition as defined by a maximum temperature change between two consecutive time steps that was less than 0.01° C. At this steady condition, the mean temperatures in the inner and outer surfaces of the esophageal wall were 20° C. and 32.5° C., respectively. The steady condition was then used as the initial condition for a following sonication.
Results
Transesophageal focal beam steering was achieved in a wide range of the field. The near-field squared pressure amplitudes increased when the beam steering angle increased. The average value that the foci shifted away from their intended focal locations was 0.9±0.7 mm for the 39 foci.
Three sonication durations of 1-, 10- and 20-seconds were adopted to simulate the short, medium and long ultrasound exposure times. The peak temperature at each focus was set as 60° C. or at 70° C. at the end of each sonication. The cooling times for the 1-, 10-, and 20-second sonications were adequately set as 24, 40 and 50 seconds to allow for tissue temperature dropping back close to 37° C. The simulated lesions at the 39 foci at 1-, 10-, and 20-second sonications to reach a 60° C. or 70° C. peak temperature were examined, with a total of 234 simulated lesions.
The disadvantage of using an intra-cavity planar probe for thermal ablation is the higher-than-normal transducer power requirement needed to achieve high enough focal intensity for tissue coagulation. Consequently, potential thermal damage to the intervening tissue layer could occur due to the proximity between the probe and the cavity wall. This may spatially compromise the safe thermal ablation zone for the planar phase array. To evaluate the safety of these sonications to patients, the thermal dose accumulation inside esophageal wall was calculated. Sonications that did not cause the thermal dose in equivalent minutes at 43° C. (T43) to be greater than 5 minutes in the esophageal wall are referred to as “safe” sonications. Sonications that produced a T43 greater than or equal to 5 minutes in the esophageal wall are referred to as “unsafe” sonications. The 5-minute T43 threshold imposed a conservative safety criterion for thermal damage estimation.
The 15 foci of the first group, having focus numbers 1 to 15, were in the x=0 plane.
The 15 foci of the second group (focus numbers 7, 8, 9, and 16 to 27) were in the z=0 plane.
The 15 foci of group 3 were in a slanted plane between x=0 and z=0 planes, spanning a slanted slice in the cardiac muscle.
a and 20b show plots that summarize the occurrence of the 10- and 20-second safe and unsafe sonications in the three foci groups at peak temperatures of 60° C. and 70° C., respectively. The occurrences of safe sonications are marked by an “o” and the occurrences of unsafe sonications are marked by an “x”. The safe sonication zone was not symmetric due to the surface curvature variations added in constructing the esophageal wall. The range of the beam steering angle and distance at which safe sonications could be achieved were more limited for the simulations with higher peak temperatures and longer sonication times than for the simulations with the lower peak temperature and shorter sonication times.
Table 4 lists the maximum, minimum and mean peak pressure amplitude at the foci for the 1-, 10-, 20-second safe sonications at 60° C. and 70° C. peak temperatures.
Discussion
At 1 MHz, the simulated planar two-dimensional phased array (60×10 mm2) was able to steer and focus its beam through esophageal wall into cardiac muscle through a wide range of angles. By varying sonication duration and power, the array produced a thermal dose that was high enough to cause tissue necrosis of different sizes. Therefore, on it was feasible to use a two-dimensional planar ultrasound phased array for transesophageal cardiac thermal ablation.
The esophagus offers a convenient ultrasound window to the heart, particularly, the back structures, such as the atria. Such proximity makes the proposed transesophageal ultrasound ablation technique promising since the esophagus tissue layer induces minimal distortion of the wave. The flexible transesophageal three-dimensional beam steering can produce continuous thermal lesions by properly planning ablation locations. Furthermore, such a flexible three-dimensional beam steering capability enables the motion of a beating heart to be tracked during sonication.
Ultrasound pressure greater than a certain threshold may cause acoustic cavitation in biological tissues. The possibility of inertial cavitation under these power levels for the three sonication durations was examined by comparing the peak pressure at the foci with the cavitation pressure threshold in muscle in vivo. The peak pressure values for all the 10- and 20-second sonications (shown in Table 4) were below the cavitation threshold in dog muscle (5.3 MPa at 1 MHz). The peak pressure values for all the 1-second sonications (shown in Table 4) were greater than the cavitation threshold because more acoustic energy was needed to coagulate tissue in a very short time. One should, however, note that there are no cavitation threshold measurements for cardiac tissue. The cavitation phenomena can be utilized to enhance tissue heating and may also be useful for cardiac ablation. The cavitation phenomena, however, was not simulated in this study. To achieve high enough peak temperature for a short sonication time while suppressing cavitation, a higher operating frequency has to be used to raise the cavitation threshold. Or, the sonication duration must be long enough to allow the acoustic pressure, which is lower than cavitation threshold, to slowly produce a thermal lesion.
Power requirement is a practical concern when designing a phased array for thermal ablation. The proposed array size in this study was 60×10 mm2. The acoustic intensity on an array surface is high when thermal lesions are to be produced rapidly. For example, the transducer acoustic power requirements for achieving 70° C. peak temperature with 1-, 10- and 20-second sonication at foci (0, 40, 0) mm were 377, 80, and 58 W and corresponded to acoustic intensities of about 63, 13, 10 W/cm2 on the transducer surface, respectively. The planar transducer array size can be increased to increase the focal pressure gain and add transducer surface area to reduce the power requirement. The human esophagus is about 25 mm in diameter. The proposed transducer array width was only 10 mm in this study; however, the transducer array width could be enlarged to reduce the power requirement of the array. The corresponding acoustic powers for a larger transducer (60×20 mm2) were 261, 60, 45 W (acoustic intensity on the transducer surface about 22, 5, 4 W/cm2), respectively. These acoustic power outputs are within the reach of current transducer array technology. With these arrays, however, the number of transducer elements becomes an issue in RF power amplifier design and channel wiring. There are tradeoffs between the transducer element size, sonication duration and acoustic power.
A number of embodiments of the invention have been described. Nevertheless, it should be understood that various modifications may be made without departing from the spirit and scope of the invention. Other embodiments are within the scope of the following claims.
Under 35 U.S.C. 119(e)(1), this application claims the benefit of provisional application serial number, 60/603,050, filed Aug. 20, 2004.
| Number | Date | Country | |
|---|---|---|---|
| 60603050 | Aug 2004 | US |