1. Field of Endeavor
The present invention relates to micromachined cantilevers and more particularly to physics-based signal processing algorithm for micromachined cantilever arrays.
2. State of Technology
U.S. Pat. No. 5,908,981 issued Jun. 1, 1999 to Abdullah Atalar et al and assigned to the Board of Trustees of the Leland Stanford, Jr. University provides the following state of technology information, “a microcantilever includes a pattern of interdigitated fingers that together form a phase grating. The phase grating is used to sense deflection of the microcantilever. In the pattern, movable fingers alternate with reference fingers. The movable fingers are physically connected to the tip of the microcantilever and move with the cantilever as it deflects; the reference fingers are physically connected to the fixed end of the cantilever and do not move as the cantilever deflects. Each reference finger is bounded on either side by movable fingers, and each movable finger is bounded on either side by a reference finger (ignoring the fingers at the ends of the pattern).”
U.S. Patent Application No. 2004/0120577 by Igor Touzov published Jun. 24, 2004 provides the following state of technology information, “Development of diverse set of applications that employs micro and nano scale properties of matter created equally wide range of equipment that is able to operate at such small scales. One of primary advantages of such technologies is the ability to efficiently and cheaply employ parallel processing for large number of entities. These parallel technologies have been developed for processing of thousands and even millions of chemicals on a single microfluidic/microarray device. Microoptical devices accounts several millions of parallel processing channels suitable for diverse tasks such as maskless lithography, printing, network switching, etc. Micromechanical and micro-electro mechanicals systems are capable of simultaneous execution of thousands and sometimes millions of simultaneous mechanical operations required for microfluidics, microoptics and micromachining.”
The article “Micromechanical Cantilever-based Biosensors” by Roberto Raiteri, Massimo Grattarola, Hans-Jürgen Butt, and Petr Skládal in Sensors and Actuators B:Chemical, Volume 79, Issues 2-3, Oct. 15, 2001, pages 115-126 provides the following state of technology information, “Microcantilevers can transduce a chemical signal into a mechanical motion with high sensitivity.” Generally, biosensing is a more demanding task than physical or chemical sensing because of the complexity of the biochemical processes involved and the nature of the operation environment. Biosensors have attracted considerable interest in the last few years since the monitoring of a specific substance is central in many applications ranging from clinical analysis to environmental control and for monitoring many industrial processes. A biosensor, as any other sensing device, can be divided into three main components: a detector which recognizes the signal of interest, a transducer which converts the signal into a more useful output, typically an electronic signal, and a read-out system which filters, amplifies, displays, records, or transmits the transduced signal. A biosensor employs a biological or biochemical detector, which can range from single proteins and enzymes up to whole cells and microorganisms. In biosensing applications, detection is usually carried out in a liquid (aqueous) environment. Flow and mixing of the solution cause turbulence which directly affects cantilever deflection. Additional drifts in deflection have been observed. They can be due to both slow electrochemical processes on either side of the cantilever and to rearrangements of the sensing surface, which is usually composed by multilayers of complex molecules like proteins.
Features and advantages of the present invention will become apparent from the following description. Applicants are providing this description, which includes drawings and examples of specific embodiments, to give a broad representation of the invention. Various changes and modifications within the spirit and scope of the invention will become apparent to those skilled in the art from this description and by practice of the invention. The scope of the invention is not intended to be limited to the particular forms disclosed and the invention covers all modifications, equivalents, and alternatives falling within the spirit and scope of the invention as defined by the claims.
Micromachined cantilevers can sense chemicals and biological molecules of interest for biodefense and pollution-control applications. Microcantilevers function in liquid and in air environments, and they are inexpensive because they are batch fabricated in large numbers using techniques similar to those used to make microelectronic circuits. It is often difficult to extract meaningful signals from the cantilevers; however, due to the sensitivity of these devices to a wide variety of stimuli.
The present invention provides methods of detecting a parameters utilizing deflection of a micromachined cantilever. The methods utilize deflection of a micromachined cantilever that represents the particular parameter being detected. Embodiments of the present invention provide methods of detecting chemical and biological agents utilizing micromachined cantilevers. One embodiment of the method comprises the steps of modeling the deflection of the micromachined cantilever producing a deflection model, sensing the deflection of the micromachined cantilever and producing a signal representing the deflection, and comparing the signal representing the deflection with the deflection model. In another embodiment, the present invention includes a physics-based signal extraction algorithm that makes used of thermodynamic, fluidic and mechanical properties of a micromachined cantilever array and its environment in order to maximize signal to noise ratios. For micromachined cantilever-based sensors, large, nonspecific background signals prevent deflections caused by target chemicals and biological molecules from being identified.
The present invention enables detection of chemicals and biological molecules of interest. The present invention can be used for environmental control, biodefense, and nonproliferation. The present invention can also be used for drug discovery, detection of pollutants, and chemical monitoring in fabrication processes. The present invention can also be used for can any system (e.g., making smart sensors that use the underlying physics).
The invention is susceptible to modifications and alternative forms. Specific embodiments are shown by way of example. It is to be understood that the invention is not limited to the particular forms disclosed. The invention covers all modifications, equivalents, and alternatives falling within the spirit and scope of the invention as defined by the claims.
The accompanying drawings, which are incorporated into and constitute a part of the specification, illustrate specific embodiments of the invention and, together with the general description of the invention given above, and the detailed description of the specific embodiments, serve to explain the principles of the invention.
Referring to the drawings, to the following detailed description, and to incorporated materials, detailed information about the invention is provided including the description of specific embodiments. The detailed description serves to explain the principles of the invention. The invention is susceptible to modifications and alternative forms. The invention is not limited to the particular forms disclosed. The invention covers all modifications, equivalents, and alternatives falling within the spirit and scope of the invention as defined by the claims.
Micromachined cantilevers are powerful transducers for sensing inorganic, organic and biological molecules, since they readily deflect in the presence of a very small number of target molecules (nanomolar to femtomolar concentrations). The number of potential target chemicals is large, ranging from DNA to explosives, suggesting that cantilevers may be useful in defense, medicine, drug discovery, and environmental monitoring. Microcantilevers have already been demonstrated to be capable of recognizing antibodies and nerve agent products such as hydrofluoric acid in solution, for example. Other cantilever-based sensors that have been developed perform other functions.
Referring now to the drawings and in particular to
A limitation on cantilever sensors in liquid, however, is that their signal-to-noise ratio (SNR) is low, often 5:1 or smaller. SNR is expected to be significantly lower in many operational environments of interest. Further, the reliability of fielded cantilever sensors ultimately depends on their being incorporated into arrays, which increase system complexity and can make response interpretation difficult. Fortunately, these difficulties are exactly the type that can be overcome with effective signal extraction techniques such as the present invention model-based approach. The present invention takes advantage of multiple redundant signals available to the signal processing algorithm. This approach improves the SNR and helps provide a physical basis for interpreting the deflection signal.
The micromachined cantilever system 10 can function as a detection device when one side is fabricated to be chemically distinct from the other, as illustrated in
The micromachined cantilever system 10 can be used for detecting chemical and biological agents. Examples of the use of the micromachined cantilever system 10 range from clinical analysis to environmental control and for monitoring many industrial processes. The micromachined cantilever system 10 can also be used for detecting chemical and biological agents of defense interest. The article “Micromechanical Cantilever-based Biosensors” by Roberto Raiteri, Massimo Grattarola, Hans-Jürgen Butt, and Petr Skládal in Sensors and Actuators B:Chemical, Volume 79, Issues 2-3, Oct. 15, 2001, pages 115-126 describes uses of the micromachined cantilever system 10. The article “Micromechanical Cantilever-based Biosensors” by Roberto Raiteri, Massimo Grattarola, Hans-Jürgen Butt, and Petr Skládal in Sensors and Actuators B:Chemical, Volume 79, Issues 2-3, Oct. 15, 2001, pages 115-126 is incorporated herein by reference. Other uses of the micromachined cantilever system 10 are described in the article, “Biosensors for Chemical and Biological Agents of Defence Interest,” by Brian M. Paddle in Biosensors & Bioelectronics, Vol. 11, No. 11, pp. 1079-1113, Feb. 13, 1996. The article “Biosensors for Chemical and Biological Agents of Defence Interest,” by Brian M. Paddle in Biosensors & Bioelectronics, Vol. 11, No. 11, pp. 1079-1113, Feb. 13, 1996 is incorporated herein by reference.
The total free energy change of the cantilever surface, ΔG, can be decomposed into four contributions: ΔGCANT, the mechanical energy change associated with bending the cantilever, ΔGPOLY, free energy change resulting from macromolecular conformational entropy and non-electrostatic interactions, ΔGOSM, free energy change from osmotic pressure of counter-ions near the surface of the cantilever, and ΔGELEC, the electrostatic free energy change.
ΔG=ΔGCANT+ΔGPOLY+ΔGOSM+ΔGELEC (Equation 1)
The free energy change is related to Δσ, the surface stress difference between top and bottom side of the cantilever by:
Δσ(t)=ΔG(t)Γ(t)/MA (Equation 2)
where ΔG has units of J/mole and is the change in the sum of all of the contributions to the free energy of the surface of the cantilever, Γ(t) is the surface concentration of the species of interest (typically in molecules per cm2) on the surface of the cantilever and MA is Avogadro's number. Sulfur-terminated molecules bind preferentially to the Au-coated side of the cantilever.
The surface concentration of the interacting molecules, Γ(t), is estimated using Langmuir kinetics. The equation describing first-order Langmuir kinetics has the following form:
Here c(t) is the bulk concentration of the target molecule in solution in moles per liter, or [M], ka is the adsorption rate constant in [M]−1s−1, kd′ (=kd/Γmax) is the desorption rate constant in cm2 molecule−1 s−1. Γmax is the maximum possible surface concentration of the species of interest in molecules cm−2, which is approached asymptotically at equilibrium as the solution concentration becomes very large. Finally, differential surface stress, from (2), in the cantilever induces a deflection, Δz(t), using a variant of Stoney's equation:
where E is the Young's modulus, vis the Poisson's ratio, and l and δ are the cantilever length and thickness, respectively.
The terms multiplying the differential surface stress are system-specific constants and can be combined and expressed as a single coefficient,
Δz(t)=βΔσ(t) (Equation 5)
Here, β=3l2(1−ν)/(Eδ2). (Equation 6)
The proportionality constant that relates cantilever deflection, Δz(t), to induced surface stress due to surface coverage, Δσ(t), can be written equivalently in terms of the spring constant of the rectangular cantilever, krect,=Eδ3W/(4l3):
where W is the width of the cantilever and the superscript c on Δz indicates deflection from cantilever coverage by chemicals, as opposed to deflection from thermal effects. Correspondingly, the coefficient to the surface stress, β, given above can be rewritten as,
The equations describing the cantilever deflection (4-8) assume that the strain is small, an assumption that is valid for all of the deflections encountered in Applicants' experiments.
Applicants will now describe the prediction of changes in surface stress as a function of surface loading. The analysis also begins with Stoney's equation (4), which states that the deflection of the cantilever is directly proportional to the difference in surface stress on the cantilever surface. This stress differential constitutes the signal. Applicants follow the approach as described by Lavrik et al. to relate the surface stress difference to the surface coverage and the free energy of adsorption (2). What remains is to develop physical models for Γ(t) and ΔG.
When incorporating the rate of change of the dimensionless surface concentration, Γ(t)/Γmax, for the MBP, Applicants employ a modified form of (3) by re-defining a rate of desorption, kd, where kd=k′dΓmax and has units of s−1. With this new definition (3) becomes:
All terms are as described above in (3) with the exception of kd. This form, (9), was chosen to enable a one-to-one comparison between rate constants predicted by the MBP presented here.
Once the input signal, a bulk concentration c0 of target molecules, is turned off, Applicants fit the desorption process with an nth order Langmuir desorption model:
The fit to the actual desorption data revealed a third-order dependence on the surface concentration, or n=3. Desorption processes are typically modeled with a first or second order model to extract kd, and the preference for a third order model is somewhat surprising. The higher order behavior could indicate a cooperative desorption process, but it is more likely due to a distribution of activation energies associated with the desorption process.
Also included in the model-based processor is the cantilever response to thermal effects, deflections due to differences in the coefficient of thermal expansion of Au and Si. Thermally-induced deflections are added directly to chemically-induced deflections to obtain the total cantilever deflection. From,
where ΔZT is the normal (z-direction) deflection of the free end of the cantilever, ΔT is the temperature difference, Δα is the difference in thermal expansion coefficients (αSi−αAu), l is the cantilever length, and δ is the thickness of the Si or Au. The superscript T on Δz in Equation (11) indicates deflection from thermal effects, as opposed to chemical surface coverage which was previously indicated in Equation (7) by superscript c. The coefficient k1 is given by
Here E is Young's modulus for Si or Au, and ν is Poisson's ratio for Si or Au. Note the Poisson effect has been included in k1, and that the gold layer is relatively thin so the final term dominates the expression for k1.
The models describing the physics of the cantilever bending and the surface chemistry were built into the model-based signal processor, and a parameter estimator was developed to fit these coupled equations predicting the optimal values for ka, kd and Γmax. With these parameters specified, Applicants calculate the free energy for adsorption, adsorption
For any target species, Applicants can then fit to find the appropriate adsorption parameters.
A prototype cantilever detection system from Veeco Instruments, Inc., was used for Applicants' experiments. In this system, a single Si chip (fabricated by IBM) with eight identical cantilevers is loaded into a 50 microliter flow cell, and cantilever deflections are measured optically using a focused light beam from a super-luminescent diode reflected into a position-sensitive detector by each cantilever. Levers are interrogated in series, with an overall measurement frequency of ˜2 Hz for all eight levers. Each Si cantilever is 500 microns long, 100 microns wide and 1 micron thick; levers are parallel to one another on the chip, spaced by 100 microns. The top side of the levers (as depicted in
During the experiment, levers were exposed to a thermal pulse, then a chemical stimulus in the form of a c0=0.014 M concentration of 2-mercaptoethanol in pure water, and finally a second heat pulse. Except for brief pauses to exchange syringes in the syringe pump, the entire experiment was performed at a constant flow rate of 20 microliter/min. Results of this experiment are shown in
Design for cantilever arrays—Applicants will describe the development of the model-based approach, first for the generic model sets and then for the specific embellishment discussed previously. Applicants start with the development of an approximate Gauss-Markov model, which can be used to capture the general cantilever signal enhancement problem. Applicants then apply it to Applicants specific microcantilever array sensor system. The scope of this section on model development involves specification of the process and measurement models for the cantilever system, construction of the model-based processor algorithm, estimation of the model parameters, verification of model performance on simulated data and validation of the model with real data. The testing with real experiments involves both single channel and multi-channel data.
If nonlinear dynamics (differential equations) describe the system under investigation, then an approximate representation of the deterministic process and associated measurement is easily captured in state-space form. State-space is simply converting an nth-order set of coupled differential equations into an equivalent set of n first-order differential equations. With this accomplished, Applicants obtain the general nonlinear vector functional relations defined by the process and measurement models,
{dot over (x)}(t)=a[x,u;θ] [process (state)]
y(t)=c[x,u;θ] [measurement] (Equation 13)
where x is the Nx-dimensional state vector; y is the Ny-dimensional measurement. u is a known input, and θ is a generic model parameter. If these processes are contaminated by additive zero-mean, Gaussian noise, then the approximate Gauss-Markov model evolves as
x(t)=a[x,u;θ]+w(t) [process (state)]
y(t)=c[x,u;θ]+v(t) [measurement] (Equation 14)
Here a[.], c[.] are the Nx-dimensional vector process function and Ny-dimensional measurement functions, respectively, for the process noise given by w˜N(0,Rww) and the corresponding measurement noise, ν˜N(0,Rvv)·N(·,·) is the Gaussian distribution specified by mean and covariance. With this representation in mind, Applicants can now define the generic cantilever signal enhancement problem as:
Given a set of noisy displacement measurements, {y(t)}, with known inputs, {u(t)}, and parameters, {θ}, specified by the approximate Gauss-Markov model of (14), FIND the best (minimum error variance) estimate of the displacement, ŷ(t), and surface concentrations, î(t).
The solution to this problem can be derived in a wide variety of approaches. Here, Applicants will use the common Bayesian approach. Applicants summarize the algorithm as:
{dot over ({circumflex over (x)})}(t|t−1)=a[{circumflex over (x)},u;θ] [State Prediction]
ŷ(t|t−1)=c[{circumflex over (x)},u;θ] [Measurement Prediction]
ε(t)=y(t)−ŷ(t|t−1) [Innovation or Residual]
{circumflex over (x)}(t|t)={circumflex over (x)}(t|t−1)+K(t)ε(t) [Correction] (Equation 15)
where K(t) is the gain of the processor, which must be calculated from the underlying process statistics. Here Applicants have discretized time and measured it in units of the time step, so the time at the step before t is t-1.
For the cantilever array problem, Applicants must convert the physical relations, (2-12), into the state-space form above. Applicants chose to solve the differential equation and incorporate the resulting relations into the measurement (cantilever sensor) model. For the state, Applicants modeled the free energy as a piecewise constant function, converted it to discrete-time using the first difference approximation and excited it with zero-mean, white Gaussian (process) noise creating a random walk model for this parameter. Therefore, Applicants start with defining the state vector as x:=ΔG and the deflection measurement y:=Δz; then Applicants obtain the following relations. From (14), the surface concentration relation becomes simply
x(t)=Δx(t−1) [Process Model] (Equation 16)
The measurement model is more complicated. Applicants must first solve for the physical variables to obtain the generic form of (11); therefore, Applicants have from (2-12) that
y
i(t)=Δzic(t)+ΔzT(t) [Measurement Model] (Equation 17)
where Δzic(t) is the chemically-induced deflection, different for different cantilevers. ΔzT(t) thermal deflection, assumed to be the same for all cantilevers. In addition to accounting for the adsorption-desorption kinetics, Applicants also developed an approximation, based on the stirred tank reactor, to estimate the target concentration as a function of time under continuous flow conditions. In the experiments presented in this communication the applied chemical signal was a step function, i.e., a constant concentration at was turned on at time, tON, and off at time tOFF. The dynamic surface concentration, ν(t), has the following form:
As described above, the differential surface stress on the cantilever is a function of the surface concentration and free energy
Δσ(t)=Γ(t)ΔG(t)=Γ(t)x(t) (Equation 19)
The deflection of the ith-lever is weighted by the Stoney equation with a specific value of β, from (5):
Δzic(t)=βiΔσ(t), for βi:=3l2(1−ν)/(Eiδi2) (Equation 20)
where βi is the ith-Stoney coefficient with cantilever modulus, Ei, with lever length and thickness, l and δi, respectively, and ν is Poisson's ratio for silicon. Here Applicants allow different values of β for the different cantilevers to account for the well-known variations in cantilever properties, something that has been investigated extensively in the context of AFM cantilever spring constants. In principle, variations in either E or δ could cause appreciable variations in β; in practice, the variation of the Young's modulus of silicon is much less than the variation of δ2. Fortunately, it is the same combination (E δ2) that enters both β and the leading term of k1, so a single fitting parameter suffices to account for variations in both. In practice Applicants have taken this parameter to be Ei. Thus, Applicants obtain the measurement equation at the ith-cantilever as
yi(t)=βiΓ(t)x(t)+ΔzT(t) (Equation 21)
Finally, assuming that both noise sources are Gaussian random processes (as before), then the result is a time-varying Gauss-Markov (not approximate due to linearity) multi-channel cantilever model defined by
x(t)=x(t−1)+w(t−1)
yi(t)=βiΓ(t)x(t)+ΔzT(t)+vi(t) (Equation 22)
for w˜N(0, Rww); v˜N(0, Rw). As before, Applicants can develop the model-based processor based on this Gauss-Markov model. First, Applicants define the signal enhancement problem in terms of the cantilever models as:
Given a set of noisy Ny-vector displacement measurements, {y(t)} with known Nu-vector inputs, {u(t)} and parameters, {θk} specified by the Gauss-Markov model of (14), Find the best (minimum error variance) estimate of the displacement and Nx-vector surface concentrations, ŷ(t|t−1), {circumflex over (x)}(t|t), respectively.
The model-based algorithm to solve this problem using the specified models is:
{circumflex over (x)}(t|t−1)={circumflex over (x)}(t−1|t−1) [Surface Concentration Prediction]
ŷi(t|t−1)=βiΓ(t){circumflex over (x)}(t|t−1)+{circumflex over (τ)}(t) [Displacement Prediction]
εi(t)=yi(t)−ŷi(t|t−1) [Innovation or Residual]
{circumflex over (x)}(t|t−1)={circumflex over (x)}(t−1|t−1)+k(t)ε(t) [Surface Concentration Correction] (Equation 23)
This completes the development of the MBP algorithm for cantilever sensor arrays. Note that once this framework is developed, it is straightforward to define other problems of high interest (e.g., detection problems).
Model-based processor performance evaluation—Applicants will discuss the performance of the model-based processor (MBP) for signal enhancement of an L-element cantilever sensor array. The basic approach Applicants take for MBP performance evaluation is illustrated in
Parameter Estimation. The basic approach Applicants use is to first “parameterize” the cantilever array model by performing parameter estimation (nonlinear least squares method) on the raw deflection measurements to extract the critical absorption, desorption and maximum concentration, that is, θi={ka(i), kd(i), Γmax(i)}; i=1, . . . , L. The parameter estimator Applicants employed was a nonlinear least-squares criterion using the Nelder-Meade polytope search algorithm. This algorithm is based on minimizing
where the estimated or filtered cantilever measurement at the ith-lever is given by
ŷi(t;θ)=Δzic(t;θ)+ΔzT(t) (Equation 25)
Once these parameters are extracted from the data, they are averaged to give
Applicants tested the parameter estimator with raw experimental deflection data (
Gauss-Markov Model Simulation. For Applicants problem Applicants chose to use ΔG(t) as an unknown but constant parameter (dΔG(t)/dt=0) and the nonlinear deflection and known temperature measurement; that is, defining
ΔG(t)=ΔG(t−1)+w(t−1) (Equation 26)
with cantilever array measurement
yi(t)=βiΓ(t)ΔG(t)+ΔzT(t)+vi(t) (Equation 27)
where ΔG(t) is the free energy at the surface, and w, vi are the additive, zero-mean, Gaussian noise processes with covariances, Rww and RvvεRL×L with diagonals, σv2(i); i=1, . . . , L. Applicants assume that the measurement uncertainty is uncorrelated producing the diagonal matrix. Each of the cantilevers has a different value of βi creating a set of Stoney coefficients, β→βi, one representing each of the individual lever properties. A typical set of cantilever simulation data is shown in
where Δztrue is the “true” deflection available from the Gauss-Markov simulation and σ2Δz,true is its variance. Once the noisy deflection measurements are synthesized, then the processors are applied to extract the “true” deflections. Applicants chose to evaluate two methods: smoothing processor and model-based processor. The smoothing processor is simply a running window average that is equivalent to a low-pass filtering operation. This smoothing is an example of a typical approach taken by scientists in this field. The MBP is the “optimal” (approximately) solution to this problem. Applicants used SSPACK_PC, a commercial model-based signal processing package in MATLAB, to perform these calculations. Table 1 gives the values of parameters for the cantilever used in the calculations.
MBP application to single channel cantilever data—The MBP was developed using the cantilever measurement model of (29) with the average parameter estimates of Table II. Concerning the values given in Table II, it is helpful to review results obtained in related chemical systems. For example, ka is 1.3×10−2 S−1 for CH3 (CH2)15 S/Au (10−3 M) in ethanol, relatively constant between concentrations of 10−3 and 10−5 M; Applicants average value of 4.6×10−3 s−1 for ka for 2-mercaptoethanol, C2H6OS, (1.4×10−2 M) in water is smaller, possibly because of different kinetic mechanisms that are active for this smaller molecule at higher concentrations, or because of solvent differences. Unlike the longer CH3(CH2)15 S, for example, 1,2-mercaptoethanol does not form well-ordered monolayers on Au. For 20-base pair thiolated DNA on Au in buffer, kd is 4.7×10−3 s−1, and Γmax=1.3×1013 molecule/cm2. Applicants value of 4.8×10−4 s−1 for kd is smaller, again possibly because of molecular size or solvent differences. Γmax for C2H6OS is larger (1.3×1015 molecule/cm2), likely because the much smaller C2H6OS is better able to concetrate on the Au surface.
Applicants will discuss the application of the MBP to simulated cantilever array measurement data at a variety of signal-to-noise ratios. Applicants first show the results of the MBP design for the −20 dB SNR case, then summarize the results at a variety of signal-to-noise ratios to evaluate its overall performance. The MBP design is based not only on the average parameter estimates (see Table II), but also using a smoothed temperature data estimate, Δ{circumflex over (z)}T(t), to be more realistic in performance evaluation. Applicants expect this processor to provide an outstanding performance, once tuned. In fact, the results of applying it to the −20 dB data indicate an approximate optimal performance, since the underlying prediction errors or innovations associated with each lever are statistically zero-mean (mean less than bound) and white (less than 5% of the points outside the bound) as shown in
Next Applicants investigate the overall performance of both the standard smoother/averager and the MBP on synthesized data sets. In order to quantitatively evaluate the performance of the processors individually, Applicants calculate the residual deflection errors defined by
Δ{tilde over (z)}ci(t):=Δzctrue,i(t)−Δ{circumflex over (z)}ci(t) (Equation 29)
where Δzctrue,i(t) is the true (noise free) deflection at the ith-lever and Δ{circumflex over (z)}ci(t) is the filtered or estimated deflections as shown for the −20 dB case in
This ratio represents the enhancement provided by each processor. The smaller the residual error variance (σ2Δ{tilde over (z)}), the higher the SNRout, providing a reasonable metric. Averaging these statistics over the cantilever array gives a measure of overall processing gain.
Applicants performed a set of simulations at 0, −20, −40 dB SNRin with the typical run outputs shown in the previous figures for the −20 dB case. The results for each run are summarized in Table IV. Here Applicants see that output SNR produced by both processors are respectable with a significant gain in enhancement; however, it is clear that the MBP consistently demonstrates superior performance with an overall average enhancement of 80 dB and enhancement gain over the Smoother of 38, 43 and 60 dB, respectively. The MBP is insensitive at these SNRs to the measurement noise variance changes yielding identical performance at each level. By contrast, the Smoother performance clearly deteriorates as the input SNR decreases. The MBP performance will also deteriorate with decreases in input SNR, but not at realistic experimental levels. This demonstrates outstanding performance for multi-channel cantilever arrays on these simulated data sets. Next Applicants apply the process to the measured array data.
Model-based processor application to multichannel cantilever data—Applicants developed MBP for two cases: (1) average deflection data and an averaged cantilever model; and (2) multi-channel deflection data. First Applicants averaged the six cantilever deflection data to obtain the complete deflection response over the entire array. Applicants next designed the MBP: Applicants first obtained the parameter estimates, then applied those parameters to the processor. Finally, Applicants developed the multi-channel processor using the same approach: estimating the individual cantilever parameters, taking their average and calculating the appropriate parameters for the multi-channel model. In both cases Applicants performed simulations first to develop the optimal MBP.
The average cantilever model was developed by performing the model-based parameter estimation obtaining the adsorption, desorption and free energy, then developing the corresponding Gauss-Markov simulation model and corresponding MBP. The raw and simulated data (0 dB) are shown in
Next Applicants applied the MBP to the actual deflection and temperature profile data shown in
Next Applicants developed the MBP for the multi-channel case using the same model-based approach: simulation and application. Applicants used the average model parameters developed over the entire cantilever array data set with the nonlinear least-squares model-based parameter estimator, then applied it to the raw cantilever data to investigate its performance. Applicants again used the MBP with the free energy as Applicants piecewise constant parameter (state) and the nonlinear cantilever array model with six elements. Applicants used the smoothed estimate of the temperature profile in Applicants estimator as in the synthesized data case. Applicants also used the estimated the Stoney coefficients and the average parameter estimates as before.
The measured cantilever data, MBP estimates and the corresponding errors or innovations are shown in
Applicants have developed a successful model-based approach to the microcantilever array signal enhancement problem. Experimental data were obtained from a multi-cantilever detection system and quantitatively analyzed with mathematical tools from physical chemistry and solid state physics. Applicants incorporated the results of this analysis directly into the generic signal processing approach. A proof-of-concept solution was created to parameterize Applicants theoretical model, enabling us to test an average model. This model was then used to develop the MBP for enhancing noisy cantilever measurements. Applicants investigated data averaged over the array and the multi-channel cases.
Through simulations with additive Gaussian noise at SNR of 0, −20, and −40 dB, Applicants demonstrated the ability of the processor to extract the cantilever deflection response with a large improvement in signal gain (˜80 dB). Applicants compared the performance of the processor to that of a “smoother” (averager) at several different signal-to-noise ratios, and the MBP demonstrated superior performance with an overall average processing gain of ˜40-60 dB over the averager. Finally, Applicants applied the MBP to noisy, smoothed (averaged) cantilever data and demonstrated that the processor could perform quite well except for a bias error, which is easily corrected.
To demonstrate the full utility of the MBP for chemical sensing of low levels of signature chemicals, necessary next steps are (1) verify the physical models used in this study for a variety of solvents and target molecules (2) make use of control levers, and (3) extend the experimental library to include low concentrations of chemical targets of practical interest.
The present invention provides model based signal processing based on developing (signal processing type models) models of the process under investigation (chemistry), measurement or sensors (cantilever array) and noise/uncertainty (background noise, chemical noise, cantilever parameter uncertainty, etc.) and incorporating them into a signal processing scheme. The models developed are usually simplified or lumped versions of the process/measurement. Applicants developed simplified models of the system and process and incorporated them into a processor. When the actual sensor measurement (experiment) and gather data is made, since the MBP has this knowledge in the form of mathematical models, it is able to outperform other processors that do not. Advantages of the present invention include incorporating the process chemistry and flow across the cantilevers, using simple structural models to capture their bending and extract the parameter of most interest the deflection to obtain the information (species) being sought. The present invention provides a method of detecting chemical or biological agents utilizing deflection of a micromachined cantilever that represents the chemical or biological agent being detected employing a model-based signal processing scheme, comprising the steps of using process system model development including the thermodynamics, chemistry fluidics and statistical uncertainties associated with parameters and noise; using measurement system model development modeling the deflection of the micromachined cantilever producing a deflection model, along with its parametric uncertainties and noise; using noise models representing the inherent noise (instrumentation, background, etc.) of cantilever array measurement system; and incorporating said models into an optimal signal processing system able to extract both parameters of high interest as well of enhanced the noisy measurements, that is, significantly increasing the final signal-to-noise ratio.
The present invention has use with cantilever measuring and sensing systems. Cantilever measuring and sensing systems are used widely. Examples of cantilever measuring and sensing systems are described in U.S. Patent Application No. 2006/0075836 by Anis Zribi, Luana Emiliana Iorio, and Daniel Joseph Lewis for a pressure sensor and method of operation thereof; U.S. Patent Application No. 2006/0016270 by Roberto Cavenago and Massimo Gherlinzoni for a load sensor; and the article Measuring Contact Stress Inside Weapons Systems in the April 2006 issue of Science and Technology Review. U.S. Patent Application No. 2006/0075836 by Anis Zribi, Luana Emiliana Iorio, and Daniel Joseph Lewis for a pressure sensor and method of operation thereof; U.S. Patent Application No. 2006/0016270 by Roberto Cavenago and Massimo Gherlinzoni for a load sensor; and the article Measuring Contact Stress Inside Weapons Systems in the April 2006 issue of Science and Technology Review are incorporated herein by reference.
While the invention may be susceptible to various modifications and alternative forms, specific embodiments have been shown by way of example in the drawings and have been described in detail herein. However, it should be understood that the invention is not intended to be limited to the particular forms disclosed. Rather, the invention is to cover all modifications, equivalents, and alternatives falling within the spirit and scope of the invention as defined by the following appended claims.
This application claims the benefit of U.S. Provisional Patent Application No. 60/683,121 filed May 17, 2005 and titled “Physics-Based Signal Processing Algorithm for Micromachined Cantilever Arrays.” U.S. Provisional Patent Application No. 60/683,121 filed May 17, 2005 and titled “Physics-Based Signal Processing Algorithm for Micromachined Cantilever Arrays” is incorporated herein by this reference.
The United States Government has rights in this invention pursuant to Contract No. W-7405-ENG-48 between the United States Department of Energy and the University of California for the operation of Lawrence Livermore National Laboratory.
Number | Date | Country | |
---|---|---|---|
60683121 | May 2005 | US |