The present disclosure generally relates to a diagnostic monitoring and interpretation of physiological waveforms (e.g., electrocardiograms (ECG), photoplethysmogram (PPG) and capnogram (CO2)). The present disclosure specifically relates a summarizing of a physiological waveform to facilitate a more accurate and reliable diagnostic monitoring and interpretation of the physiological waveform.
Many physiological waveforms (e.g., electrocardiograms (ECG), photoplethysmograms (PPG) and capnograms (CO2)) display a repeated pattern of limited number of shapes. For example, in a resting healthy person, a standard recording of ECG could simply look like one particular shape for one heart beat is repeated over and over. Of course, even in such case there really are minor differences between these similar-looking “beats” due to both physiological and environmental reasons. These changes may be more pronounced in unhealthy subjects and/or when the environmental conditions change. Therefore, monitoring the changes in repeating pattern is widely used to screen or diagnose a disease.
For example, a stress test (sometimes called a treadmill test or exercise test) gradually increases the workload on the subject's heart in order to find out how well the heart handles work. In the presence of obstructive coronary artery disease (CAD), the morphology of ECG beats changes. More precisely, usually the shape of ST segment in the ECG beat changes, through either depression or elevation. Such morphology changes often suggest presence of CAD and warrants further management.
Physicians and the expert reviewers of the physiological waveforms are often given long recordings to annotate and extract the abnormalities and events by looking for the changes in repeating patterns and standard parameters determined by measurements. This procedure is performed either manually or by the aid of computer applications and is usually difficult, time-consuming and error-prone especially in presence of noise and artifact.
ECG, as a typical example of a physiological waveform reviewed by experts, is the main tool in identification of the life-threatening arrhythmias, and detection of the cardiac diseases by measuring the standard parameters from beat fiducial points, such as, for example, ST elevation, QT duration, and QRS width, and inspection of the presence of normal ECG beat components such as P-wave and normal T-wave.
In a long record of a physiological waveform, it would be challenging and time-consuming to detect the gradual and major changes to the repeating signal pattern (i.e. pulses). Furthermore, keeping track of all changes and measurements could be impacted by human error factors. Additionally, the noise and artifact on the signal will make it difficult to make valid and reliable measurements of the pattern.
Moreover, a failure in detection and reliable measurement of each of these parameters in an ECG record reviewed by experts may result in drastic and irreversible errors with potentially detrimental consequences for the patient undergoing the diagnosis.
The present disclosure describes physiological waveform summarizing that is applicable to numerous and various diagnostic monitoring and interpretation applications of clinical physiological waveforms (e.g., ECG waveform, PPG waveform, CO2 waveform, etc.).
The physiological waveform summarizing of the present disclosure identifies one or more major physiological templates for each interval of consecutive intervals (overlapping or non-overlapping) of a clinical physiological waveform to thereby exclude minor physiological templates, which may contain noise and/or artefacts, from the clinical physiological waveform.
One major physiological template is a dominant physiological template constituting a dominating major physiological rhythm of the clinical physiological waveform within an interval of the clinical physiological waveform whereby any deviation in the morphology of the dominating major physiological rhythm may be an indication of a predominant abnormality in the clinical physiological waveform, and further whereby a diagnostic analysis of physiological parameters of the dominating major physiological rhythm will be more accurate and reliable than diagnostic analysis of physiological parameters of the clinical physiological waveform.
A second major physiological template is a secondary physiological template constituting a presence of a secondary major physiological rhythm within an interval of the clinical physiological waveform. A secondary physiological template is usually a sign of an underlying abnormality in the clinical physiological waveform. For example, in an ECG rhythm, a secondary template may be indicative of arrhythmias such as Premature Atrial Complex (PAC), Premature Ventricular Complex (PVC—bigeminy or trigeminy rhythms), Atrial Fibrillation (AFib), Atrial Flutter (AFlutter), and Ventricular Tachycardia (VT).
The present disclosure may be embodied as:
Various embodiments of a physiological waveform summarizing monitor of the present disclosure encompass a physiological waveform summarizing recorder and a physiological waveform summarizing analyzer for summarizing a clinical physiological waveform having a dominating major physiological rhythm and possibly one or more secondary major physiological rhythms.
In operation, the physiological waveform summarizing recorder receives a clinical physiological waveform in real-time or as a recording. The recorder processes the clinical physiological waveform to extract a set of dominant physiological templates from the clinical physiological waveform. The set of dominant physiological templates represent the dominating major physiological rhythm of the clinical physiological waveform temporally spanning over consecutive intervals of the clinical physiological waveform, and each dominant physiological template is derived from a different interval of the consecutive intervals of the clinical physiological waveform
The recorder may also extract one or more secondary physiological template from the clinical physiological waveform. The secondary physiological template(s) represent(s) secondary major physiological rhythm(s) of the clinical physiological waveform.
In operation, the physiological waveform summarizing analyzer provides a major physiological rhythm log of the clinical physiological waveform including a diagnostic plotting of one or more of extracted dominant physiological templates, and optionally one or more of the extracted secondary physiological templates.
Various embodiments of a physiological waveform summarizing systems of the present disclosure encompass a clinical physiological waveform interface in addition to a physiological waveform summarizing monitor of the present disclosure. The clinical physiological waveform interface employs machines, devices, appliances, equipment, tools, accessories, etc. as known in the art of the present disclosure and hereinafter conceived for interfacing with a clinical subject to acquire and communicate the clinical physiological waveform to the monitor. Examples of a clinical physiological waveform interface include, but are not limited to, an ECG pads/ECG lead system based interface for acquiring and communicating an electrocardiogram (ECG) waveform to the monitor, a pulse oximeter based interface for acquiring and communicating a photoplethysmogram (PPG) waveform to the monitor and a CO2 sensor based interface for acquiring and communicating a capnogram (CO2) waveform to the monitor.
Various embodiments of a physiological waveform summarizing method of the present disclosure encompass a summary of a clinical physiological waveform having a dominating major physiological rhythm and possibly one or more secondary major physiological rhythms.
In operation, the physiological waveform summarizing recorder receives a clinical physiological waveform in real-time or as a recording. The recorder processes the clinical physiological waveform to extract dominant physiological templates from the clinical physiological waveform. The dominant physiological templates represent the dominating major physiological rhythm of the clinical physiological waveform temporally spanning over consecutive intervals (overlapping or non-overlapping) of the clinical physiological waveform, and each dominant physiological template is derived from a different interval of the consecutive intervals of the clinical physiological waveform
The recorder may also extract one or more secondary physiological template from the clinical physiological waveform if secondary major physiological rhythm(s) are present in the clinical physiological waveform.
In operation, the physiological waveform summarizing analyzer provides a major physiological rhythm log of the clinical physiological waveform including a diagnostic plotting of one or more extracted dominant physiological templates, and one or more of any extracted secondary physiological template(s).
For purposes of the description and claims of the present disclosure:
The foregoing embodiments and other embodiments of the present disclosure as well as various structures and advantages of the present disclosure will become further apparent from the following detailed description of various embodiments of the present disclosure read in conjunction with the accompanying drawings. The detailed description and drawings are merely illustrative of the present disclosure rather than limiting, the scope of the present disclosure being defined by the appended claims and equivalents thereof.
The present disclosure will present in detail the following description of exemplary embodiments with reference to the following figures wherein:
The present disclosure is applicable to numerous and various diagnostic monitoring and interpretation applications of clinical physiological waveforms (e.g., ECG waveform, PPG waveform, CO2 waveform, etc.).
The present disclosure improves upon the prior art by extracting dominant physiological templates from a clinical physiological waveform whereby the dominant physiological templates represent a dominating major physiological rhythm of the clinical physiological waveform, and further whereby a diagnostic analysis of physiological parameters of the dominant physiological templates will be more accurate and reliable than a diagnostic analysis of the dominating major physiological rhythm of the clinical physiological waveform.
The present disclosure also improves upon the prior art by optionally extracting secondary physiological templates from the clinical physiological waveform whereby the secondary physiological templates represent a secondary major physiological rhythm of the clinical physiological waveform, and further whereby an analysis of the physiological parameters of the secondary physiological templates will be more accurate and reliable than an analysis of the secondary major physiological rhythm of the clinical physiological waveform,
To facilitate an understanding of the present disclosure, the following description of
Referring to
In practice, interface 20 employs machines, devices, appliances, equipment, tools, accessories, etc. as known in the art of the present disclosure and hereinafter conceived for interfacing 21 with a clinical subject 10 to acquire and communicate a clinical physiological waveform CPW to monitor 30.
In one exemplary embodiment of interface 20 as shown in
In a second exemplary embodiment of interface 20 as shown in
In a third exemplary embodiment of interface 20 as shown in
Referring back to
In practice, recorder 40 records clinical physiological waveform CPW as received from clinical physiological waveform interface 20. Concurrently or subsequently to the recording of clinical physiological waveform CPW, recorder 40 extracts dominant physiological templates 41 from clinical physiological waveform CPW.
More particularly, during a stage S62 of flowchart 60, recorder 40 identifies a dominant physiological template 41 for each interval of a plurality consecutive intervals (overlapping or non-overlapping) of clinical physiological waveform CPW. For example, as shown in stage S62, clinical physiological waveform CPW is segmented into six (6) intervals 70a-70f whereby a dominant morphology within each interval 70 of clinical physiological waveform CPW is extracted and designated as a dominant physiological template 41.
In practice, a dominant morphology of an interval of clinical physiological waveform CPW is the morphology best representative of a physiological status of the clinical subject during that interval of clinical physiological waveform CPW for diagnostic monitoring and interpretation purposes.
Examples of a dominant morphology of an interval of clinical physiological waveform CPW include, but are not limited to, a mean morphology or function thereof among the periods/cycles within the interval of clinical physiological waveform CPW, a median morphology or function thereof among the periods/cycles of within the interval of clinical physiological waveform CPW, and a mode morphology or function thereof among the periods/cycles of within the interval of clinical physiological waveform CPW.
At the conclusion of stage S62, recorder 40 temporally aligns the dominant physiological templates 41 as a representation of a dominating major physiological rhythm DMPR of clinical physiological waveform CPW, which may be communicated to analyzer 50 as a dominating major physiological rhythm as shown or as individual dominant physiological templates 41 in sequential order.
In one exemplary embodiment of monitor 30 as shown in
In practice, recorder 40a records clinical electrocardiogram waveform CECG as received from clinical electrocardiogram waveform interface 20a. Concurrently or subsequently to the recording of clinical electrocardiogram waveform CECG, recorder 40a extracts dominant electrocardiogram templates 41a from clinical electrocardiogram waveform CECG.
More particularly, during a stage S62a of flowchart 60a, recorder 40a identifies a dominant electrocardiogram template 41a for each interval of consecutive intervals (overlapping or non-overlapping) of clinical electrocardiogram waveform CECG. For example, as shown in stage S62a, clinical electrocardiogram waveform CECG is segmented into six (6) intervals 71a-71f whereby a dominate morphology within each interval 71 of clinical electrocardiogram waveform CECG is extracted and designated as a dominant electrocardiogram template 41a.
In practice, a dominant morphology of an interval of clinical electrocardiogram waveform CECG is the morphology best representative of an electrocardiogram status of the clinical subject during that interval of clinical electrocardiogram waveform CECG for diagnostic monitoring and interpretation purposes.
Examples of a dominant morphology of an interval of clinical electrocardiogram waveform CECG include, but are not limited to, a mean morphology or function thereof among the cardiac cycles within the interval of clinical electrocardiogram waveform CECG, a median morphology or function thereof among the cardiac cycles of within the interval of clinical electrocardiogram waveform CECG, and a mode morphology or function thereof among the cardiac cycles of within the interval of clinical electrocardiogram waveform CECG.
At the conclusion of stage S62a, recorder 40 temporally aligns the dominant electrocardiogram templates 41a as a representation of a dominant electrocardiogram rhythm DECGR of clinical physiological waveform CPW, which may be communicated to analyzer 50a as dominant electrocardiogram waveform as shown or as individual dominant electrocardiogram templates 41a in sequential order.
In a second exemplary embodiment of monitor 30 as shown in
In practice, recorder 40b records clinical photoplethysmogram waveform CPPG as received from clinical photoplethysmogram waveform interface 20a. Concurrently or subsequently to the recording of clinical photoplethysmogram waveform CPPG, recorder 40b extracts dominant photoplethysmogram templates 41b from clinical photoplethysmogram waveform CPPG.
More particularly, during a stage S62b of flowchart 60b, recorder 40b identifies a dominant photoplethysmogram template 41b for each interval of consecutive intervals (overlapping or non-overlapping) of clinical photoplethysmogram waveform CPPG. For example, as shown in stage S62a, clinical photoplethysmogram waveform CPPG is segmented into six (6) intervals 72a-72f whereby a dominant morphology within each interval 72 of clinical photoplethysmogram waveform CPPG is extracted and designated as a dominant photoplethysmogram template 41b.
In practice, a dominant morphology of an interval of clinical photoplethysmogram waveform CPPG is the morphology best representative of a photoplethysmogram status of the clinical subject during the interval of clinical photoplethysmogram waveform CPPG for diagnostic monitoring and interpretation purposes.
Examples of a dominant morphology of an interval of clinical photoplethysmogram waveform CPPG include, but are not limited to, a mean morphology or function thereof among the cardiac cycles within the interval of clinical photoplethysmogram waveform CPPG, a median morphology or function thereof among the cardiac cycles of within the interval of clinical photoplethysmogram waveform CPPG, and a mode morphology or function thereof among the cardiac cycles of within the interval of clinical photoplethysmogram waveform CPPG.
At the conclusion of stage S62b, recorder 40 temporally aligns the dominant photoplethysmogram templates 41b as a representation of a dominant photoplethysmogram rhythm DPPGR, which may be communicated to analyzer 50b as dominant photoplethysmogram waveform as shown or as individual dominant photoplethysmogram templates 41b in sequential order.
In a third exemplary embodiment of monitor 30 as shown in
In practice, recorder 40c records clinical capnogram waveform CCO2 as received from clinical capnogram waveform interface 20a. Concurrently or subsequently to the recording of clinical capnogram waveform CCO2, recorder 40c extracts dominant capnogram templates 41b from clinical capnogram waveform CCO2.
More particularly, during a stage S62c of flowchart 60c, recorder 40c identifies a dominant capnogram template 41b for each interval of consecutive intervals (overlapping or non-overlapping) of clinical capnogram waveform CCO2. For example, as shown in stage S62a, clinical capnogram waveform CCO2 is segmented into six (6) intervals 73a-73f whereby a dominant morphology within each interval 73 of clinical capnogram waveform CCO2, is extracted and designated as dominant capnogram template 41b.
In practice, a dominant morphology of an interval of clinical capnogram waveform CCO2 is the morphology best representative of a capnogram status of the clinical subject during the interval of clinical capnogram waveform CCO2 for diagnostic monitoring and interpretation purposes.
Examples of a dominant morphology of an interval of clinical capnogram waveform CCO2 include, but are not limited to, a mean morphology or function thereof among the respiratory cycles within the interval of clinical capnogram waveform CCO2, a median morphology or function thereof among the respiratory cycles of within the interval of clinical capnogram waveform CCO2, and a mode morphology or function thereof among the respiratory cycles of within the interval of clinical capnogram waveform CCO2.
At the conclusion of stage S62c, recorder 40 temporally aligns the dominant capnogram templates 41c as a representation of a dominant capnogram rhythm DCO2R, which may be communicated to analyzer 50c as dominant capnogram waveform as shown or as individual dominant capnogram templates 41c in sequential order.
Referring back to
More particularly, during a stage S64 of flowchart 60, for each interval of clinical physiological waveform CPW, recorder 40 determines a presence or an absence of a secondary major physiological rhythm with clinical physiological waveform CPW whereby a detected presence of a secondary major physiological rhythm within an interval is extracted and designated as a secondary physiological template 42.
An exemplary implementation of stage S64, as shown in
At the conclusion of stage S64, recorder 40 temporally aligns the secondary physiological templates 41 to as a representation of the secondary major physiological rhythm SMPR of clinical physiological waveform CPW, which may be communicated to analyzer 50 as secondary physiological waveform as shown or as individual secondary physiological templates 42 in sequential order.
In one exemplary embodiment of stage S64 as shown in
More particularly, during a stage S64a of flowchart 60a, for each interval of clinical electrocardiogram waveform CECG, recorder 40a determines a presence or an absence of secondary major electrocardiogram rhythm with clinical electrocardiogram waveform CECG and embodies any identified secondary major electrocardiogram rhythm within a secondary electrocardiogram template 42a.
An exemplary implementation of stage S64a, as shown in
At the conclusion of stage S64a, recorder 40a temporally aligns the secondary electrocardiogram templates 42a as a representation of the secondary electrocardiogram rhythm SECGR of clinical electrocardiogram waveform CECG, which may be communicated to analyzer 50a (
In a second exemplary embodiment of stage S64 as shown in
More particularly, during a stage S64b of flowchart 60b, for each interval of clinical photoplethysmogram waveform CPPG, recorder 40b determines a presence or an absence of a secondary major photoplethysmogram rhythm with clinical photoplethysmogram waveform CPPG and embodies any identified secondary major photoplethysmogram rhythm within a photoplethysmogram template 42b.
An exemplary implementation of stage S64b, as shown in
At the conclusion of stage S64a, recorder 40b temporally aligns the secondary photoplethysmogram templates 42b as a representation of the secondary photoplethysmogram rhythm SPPGR of clinical photoplethysmogram waveform CPPG, which may be communicated to analyzer 50b (
In a third exemplary embodiment of stage S64 as shown in
More particularly, during a stage S64c of flowchart 60c, for each interval of clinical capnogram waveform CCO2, recorder 40c determines a presence or an absence of a secondary major capnogram rhythm with clinical capnogram waveform CCO2 and embodies a secondary major capnogram rhythm within a capnogram template 42c.
An exemplary implementation of stage S64c, as shown in
At the conclusion of stage S64c, recorder 40c temporally aligns the secondary capnogram templates 42c as a representation of the secondary capnogram rhythm SCO2R of clinical capnogram waveform CCO2, which may be communicated to analyzer 50c (
Referring back to
More particularly, during a stage S66 of flowchart 60, analyzer 50 may display a dominating major physiological rhythm log including an integrated or a segregated display of a subset or the entire set of dominant physiological templates 41 extracted from clinical physiological waveform CPW during stage S62, and may further include an individual, an integrated or a segregated display of secondary physiological template(s) 42 extracted from clinical physiological waveform CPW during stage S64, if implemented.
For example,
The display of templates 41, 42 as shown in
Alternatively or concurrently, the display of templates 41, 42 as shown in
In one exemplary embodiment of stage S66 of flowchart 60 as shown in
For example,
The display of templates 41a, 42a as shown in
In a second exemplary embodiment of stage S66 of flowchart 60 as shown in
For example,
The display of templates 41b, 42b as shown in
In a third exemplary embodiment of stage S66 of flowchart 60 as shown in
For example,
The display of templates 41c, 42c as shown in
To facilitate an understanding of the present disclosure, the following description of
Referring to
A stage S202 of flowchart 200 encompasses a monitor of the present disclosure dividing a clinical physiological waveform into an X number of overlapping or non-overlapping segments, X≥2.
For example, a thirty (30) minute clinical physiological waveform may be divided into thirty (30) one-minute non-overlapping segments.
A stage S204 of flowchart 200 encompasses the monitor of the present disclosure reading of a first segment in a manner to facilitate a cycle detection by the monitor of the present disclosure during a stage S206 of flowchart 200 and a template rhythm matching by the monitor of the present disclosure during a stage S208 of flowchart 200.
Still referring to
In one exemplary embodiment of stage S206 for a clinical electrocardiogram, the cycle detection identifies each cardiac cycle (pulse), if any, within the first segment of the clinical electrocardiogram as known in the art of the present disclosure or hereinafter conceived.
In a second exemplary embodiment of stage S206 for a clinical photoplethysmogram, the cycle detection identifies each cardiac cycle (pulse), if any, within the first segment of the clinical photoplethysmogram as known in the art of the present disclosure or hereinafter conceived.
In a third exemplary embodiment of stage S206 for clinical capnogram, the cycle detection identifies each respiratory cycle (breath), in any, within the first segment of the clinical capnogram as known in the art of the present disclosure or hereinafter conceived.
Still referring to
In practice of stage S208, supervised learning model(s) and/or unsupervised learning model(s) may serve as a basis for functionally mapping each identified cycle to one of a plurality of matching rhythm groups.
In one exemplary embodiment of stage S208 as shown in
Referring to
In practice, a supervised neural network classifier will be trained on designated rhythms of the clinical physiological waveform. For example, a supervised neural network classifier may be trained on designated rhythms of a clinical electrocardiogram including, but not limited to, a normal rhythm, a Premature Atrial Complex (PAC), a Premature Ventricular Complex (PVC—bigeminy or trigeminy rhythms), an Atrial Fibrillation (AFib), an Atrial Flutter (AFlutter), and a Ventricular Tachycardia (VT). Alternatively in practice, an unsupervised neural network classifier will be designed to determine and group related rhythms of the clinical physiological waveform. For example, a unsupervised neural network classifier may be designed to determine and group related rhythms of a clinical electrocardiogram including, but not limited to, a normal rhythm, a Premature Atrial Complex (PAC), a Premature Ventricular Complex (PVC—bigeminy or trigeminy rhythms), an Atrial Fibrillation (AFib), an Atrial Flutter (AFlutter), and a Ventricular Tachycardia (VT).
Still referring to
Referring back to
In practice, if the first segment of the clinical physiological waveform did not include any cycles (e.g., a pulse-less segment of a clinical electrocardiogram or a clinical photoplethysmogram, or a breath-less segment of a capnogram), then neither a dominant physiological template nor a secondary physiological template of the first segment would have been identified during stage S208 and the monitor of the present disclosure would proceed to a stage S216 of the flowchart 200 to ascertain if there are any remaining segments. If there are remaining segments, then the monitor of the present disclosure proceeds to iterate stages S204-S210 for the next segment. Otherwise, flowchart 200 is terminated.
In practice, if the first segment of the clinical physiological waveform included one or more cycles (e.g., a pulsing segment of a clinical electrocardiogram or a clinical photoplethysmogram, or a breathing segment of a capnogram), then a dominant physiological template of the first segment and if present, a secondary physiological template of the first segment would have been identified during stage S208 and the monitor of the present disclosure would proceed to a stage S212 of flowchart 200 to ascertain if the identified template(s) of stage S208 are within the first segment.
If the identified templates of stage S208 are within the first segment, then the monitor of the present disclosure proceeds to a stage S214 of flowchart 200 to plot the identified template(s) within a major physiological rhythm log.
If the identified template(s) of stage S208 are not within the first segment, then the monitor of the present disclosure proceeds to stage S214 to plot an identified dominant physiological template within a major physiological rhythm log if the dominant physiological template is a major deviation from a previously plotted dominant physiological template within the major physiological rhythm log and/or to plot an identified secondary physiological template within the major physiological rhythm log if the secondary physiological template is a major deviation from a previously plotted secondary physiological template within the major physiological rhythm log.
In practice of stage S212, a flowchart 240 of
Referring back to
More particularly,
More particularly,
More particularly,
To facilitate an understanding of the present disclosure, the following description of
Referring to
Each processor 131 may be any hardware device, as known in the art of the present disclosure or hereinafter conceived, capable of executing instructions stored in memory 132 or storage or otherwise processing data. In a non-limiting example, the processor(s) 131 may include a microprocessor, field programmable gate array (FPGA), application-specific integrated circuit (ASIC), or other similar devices.
The memory 132 may include various memories, as known in the art of the present disclosure or hereinafter conceived, including, but not limited to, L1, L2, or L3 cache or system memory. In a non-limiting example, the memory 132 may include static random access memory (SRAM), dynamic RAM (DRAM), flash memory, read only memory (ROM), or other similar memory devices.
The user interface 133 may include one or more devices, as known in the art of the present disclosure or hereinafter conceived, for enabling communication with a user such as an administrator. In a non-limiting example, the user interface may include a command line interface or graphical user interface that may be presented to a remote terminal via the network interface 134.
The network interface 134 may include one or more devices, as known in the art of the present disclosure or hereinafter conceived, for enabling communication with ventilation system 220, pharmacological source(s) 250, physiological sources(s) 260, psychological source(s) 270 and an electronic medical records system 210. In a non-limiting example, the network interface 134 may include a network interface card (NIC) configured to communicate according to the Ethernet protocol. Additionally, the network interface 134 may implement a TCP/IP stack for communication according to the TCP/IP protocols. Various alternative or additional hardware or configurations for the network interface 134 will be apparent.
The storage 136 may include one or more machine-readable storage media, as known in the art of the present disclosure or hereinafter conceived, including, but not limited to, read-only memory (ROM), random-access memory (RAM), magnetic disk storage media, optical storage media, flash-memory devices, or similar storage media. In various non-limiting embodiments, the storage 136 may store instructions for execution by the processor(s) 131 or data upon with the processor(s) 131 may operate. For example, the storage 136 may store a base operating system for controlling various basic operations of the hardware. The storage 136 also stores application modules physiological waveform summarizing recorder 140 and physiological waveform summarizing analyzer 150 in the form of executable software/firmware for implementing the various functions of physiological waveform summarizing recorder 40 (
In practice, physiological waveform summarizing monitor 130 may be a stand-alone monitor including a display, user interface, etc., as known in the art of the present disclosure or hereinafter conceived, or may be incorporated/integrated as a controller into devices and systems for acquiring, monitoring and/or interpreting clinical physiological waveforms as known in the art of the present disclosure or hereinafter conceived. Examples of such devices and systems include, but are not limited to, defibrillators, pacemakers, ECG monitors, PPG monitors, CO2 monitors, Holter monitors and stress test systems.
Referring to
Further, as one having ordinary skill in the art will appreciate in view of the teachings provided herein, structures, elements, components, etc. described in the present disclosure/specification and/or depicted in the Figures may be implemented in various combinations of hardware and software, and provide functions which may be combined in a single element or multiple elements. For example, the functions of the various structures, elements, components, etc. shown/illustrated/depicted in the Figures can be provided through the use of dedicated hardware as well as hardware capable of executing software in association with appropriate software for added functionality. When provided by a processor, the functions can be provided by a single dedicated processor, by a single shared processor, or by a plurality of individual processors, some of which can be shared and/or multiplexed. Moreover, explicit use of the term “processor” or “controller” should not be construed to refer exclusively to hardware capable of executing software, and can implicitly include, without limitation, digital signal processor (“DSP”) hardware, memory (e.g., read only memory (“ROM”) for storing software, random access memory (“RAM”), non-volatile storage, etc.) and virtually any means and/or machine (including hardware, software, firmware, combinations thereof, etc.) which is capable of (and/or configurable) to perform and/or control a process.
Moreover, all statements herein reciting principles, aspects, and embodiments of the invention, as well as specific examples thereof, are intended to encompass both structural and functional equivalents thereof. Additionally, it is intended that such equivalents include both currently known equivalents as well as equivalents developed in the future (e.g., any elements developed that can perform the same or substantially similar function, regardless of structure). Thus, for example, it will be appreciated by one having ordinary skill in the art in view of the teachings provided herein that any block diagrams presented herein can represent conceptual views of illustrative system components and/or circuitry embodying the principles of the invention. Similarly, one having ordinary skill in the art should appreciate in view of the teachings provided herein that any flow charts, flow diagrams and the like can represent various processes which can be substantially represented in computer readable storage media and so executed by a computer, processor or other device with processing capabilities, whether or not such computer or processor is explicitly shown.
Having described preferred and exemplary embodiments of the various and numerous inventions of the present disclosure (which embodiments are intended to be illustrative and not limiting), it is noted that modifications and variations can be made by persons skilled in the art in light of the teachings provided herein, including the Figures. It is therefore to be understood that changes can be made in/to the preferred and exemplary embodiments of the present disclosure which are within the scope of the embodiments disclosed herein.
Moreover, it is contemplated that corresponding and/or related systems incorporating and/or implementing the device/system or such as may be used/implemented in/with a device in accordance with the present disclosure are also contemplated and considered to be within the scope of the present disclosure. Further, corresponding and/or related method for manufacturing and/or using a device and/or system in accordance with the present disclosure are also contemplated and considered to be within the scope of the present disclosure.
Number | Date | Country | Kind |
---|---|---|---|
20152739.7 | Jan 2020 | EP | regional |
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/EP2020/087941 | 12/28/2020 | WO |
Number | Date | Country | |
---|---|---|---|
62955890 | Dec 2019 | US |