Certain embodiments described herein are generally related to digital imaging, and more specifically, to standard and Fourier ptychographic imaging systems and methods with convex relaxation.
Ptychography imaging involves collecting lower resolution images and then reconstructing the image data to form a higher resolution image. Over the past two decades, ptychographic imaging has been used in a variety of regimes to produce high-resolution, wide field-of-view images of microscopic and nanoscopic phenomena. Whether in the X-ray regime at third-generation synchrotron sources, in the electron microscope for atomic scale phenomena, or in the in the optical regime for biological specimens, ptychography has shown an unparalleled ability to acquire hundreds of megapixels of sample information near the diffraction limit. Typically, the underlying operation of ptychography is to sample a series of diffraction patterns from a specimen as it is scanned through a focused beam. These intensity-only measurements are then reconstructed into a complex (i.e. amplitude and phase), high-resolution image with more pixels of sample information than any single recorded diffraction pattern.
Most recently, a Fourier ptychographic microscope (FPM) was introduced that uses a Fourier ptychographic technique that can reconstruct gigapixel optical images from a sequence of lower resolution images collected using a low NA objective lens from a conventional microscope. In one example, Fourier ptychographic microscope activates different LEDs in an LED array to illuminate a sample from different directions while the low-resolution images are captured. As in standard ptychography, Fourier ptychography recovers the sample's phase as it merges together the captured image sequence into a high-resolution output.
Conventional ptychographic imaging systems can avoid the need for a high NA, well-corrected objective lens to image at the diffraction-limit by resolving resolution-limiting factors in their data capture and reconstruction techniques. However, these systems lack stable, robust, and accurate reconstruction methods. For example, conventional ptychographic systems reconstruct the phase of the scattered field from measured intensities using non-convex algorithms. Most of these conventional systems solve the phase retrieval problem by applying known constraints in an iterative manner using an “alternating projection” (AP) strategy. Reconstruction techniques that use AP strategies tend to converge to incorrect local minima and/or to stagnate.
Certain aspects of this disclosure pertain to standard and Fourier ptychographic imaging systems and methods with convex relaxation.
Certain aspects pertain to ptychographic imaging systems with convex relaxation. In some aspects, a ptychographic imaging system with convex relaxation comprises one or more electromagnetic radiation sources, a digital radiation intensity detector, and a processor in communication with the digital radiation detector. In some cases, the one or more electromagnetic radiation sources configured to provide coherent radiation to a specimen from a plurality of incidence angles at a sequence of sample times. For example, the one or more electromagnetic radiation sources may be an LED array. The digital radiation intensity detector configured to receive light transferred from the sample by diffractive optics. In these cases, the digital radiation intensity detector further configured to capture intensity distributions for a sequence of low resolution intensity images associated with the plurality of incidence angles. The processor in communication with the digital radiation detector to receive image data with intensity distributions for the sequence of low resolution intensity images. The processor configured to generate a convex problem based on the sequence of low resolution intensity images and optimize the convex problem to reconstruct a high-resolution image of the specimen. In other cases, the diffractive optics comprises a spatial light modulator configured to provide a pattern at a plurality of locations at a Fourier plane of the specimen. In these cases, the sequence of low resolution images capture by the digital radiation detector is associated with the plurality of locations of the pattern.
Certain aspects pertain to ptychographic imaging methods with convex relaxation. In some aspects, a ptychographic imaging method with convex relaxation comprises collecting a sequence of low resolution images of a specimen, generating a convex problem based on the sequence of low resolution intensity images, and optimizing the convex problem to reconstruct a high-resolution image of the specimen. In some cases, collecting a sequence of low resolution images of a specimen comprises providing coherent radiation to a specimen from a plurality of incidence angles at a sequence of sample times, transferring light from the specimen through diffractive optics to a digital radiation intensity detector, and sampling a sequence of low resolution intensity images associated with the plurality of incidence angles. In some cases, generating the convex problem comprises stacking image data from the low-resolution images into a combined image matrix, constructing measurement matrices, and generating the convex problem using convex relaxation.
These and other features are described in more detail below with reference to the associated drawings.
Embodiments of the present invention will be described below with reference to the accompanying drawings. The features illustrated in the drawings may not be to scale.
Certain aspects are directed to ptychographic (standard or Fourier) imaging systems and methods with convex relaxation. Typically, these ptychographic (standard or Fourier) imaging systems comprise coherent electromagnetic (EM) radiation sources for illuminating the specimen, diffraction optics, and a digital radiation intensity detector for taking intensity measurements. The imaging method with convex relaxation starts by providing coherent EM radiation to illuminate the specimen being imaged. The diffractive optics receives light from the specimen and transfers diffraction patterns to the digital radiation intensity detector, which samples intensity measurements at a sequence of sample times. In the case of a standard ptychography imaging system, the diffraction optics are free-space propagation. In the case of a Fourier ptychography imaging system, the diffraction optics are a combination of free-space propagation and a collection optical element. The digital intensity measurements captured at each sample time provide image data of a low-resolution image. The method reconstructs a high resolution image from the image data of the sequence of low-resolution images. This method manipulates the image matrix to form a convex problem by stacking the reduced resolution images in combined matrix, constructing measurement matrices, and using convex relaxation to create the convex program. The method then reconstructs a high resolution image from the convex image matrix by either (Option A) relaxing the convex image matrix into low-rank formulation and then solving the low-rank image matrix using a low-rank ptychography (LRP) process or (Option B) determining the high resolution image directly with a convex lifted ptychography (CLP) technique.
Certain aspects are directed to ptychographic (standard or Fourier) imaging systems and methods with convex relaxation. In some aspects, ptychographic imaging systems and methods with convex relaxation may provide the desired stability, robustness and/or reliability in reconstructing high resolution images from a collection of lower resolution images. In some aspects, these ptychographic imaging systems and methods use reconstruction techniques with convex relaxation that do not have local minima, incorporate noise compensation techniques, and/or use multiple a priori constraints. In one aspect, a reconstruction technique with convex relaxation uses low-rank factorization, whose runtime and memory usage are near-linear with respect to the size of the output image. In this example, the reconstruction technique may be able to provide 25% lower background variance than conventional ptychographic reconstruction methods.
In certain aspects, ptychographic imaging systems and methods with convex relaxation include an unaided recovery technique that does not use prior sample knowledge or an appropriate heuristic, which may be especially relevant in biological imaging. Moreover, these imaging systems and methods do not have local minima so that a single solution can be found efficiently. In addition, these imaging systems and methods are more noise-tolerant than imaging systems that use AP strategies, which makes results more reproducible. Furthermore, a factorization technique can be implemented to obtain solutions at scale. Thus, these aspects provide noise-tolerant and efficient reconstruction techniques that may provide for more efficiently and accurately processing multi-gigapixel high resolution images than conventional systems.
I. Ptychographic Imaging Systems with Convex Relaxation
As mentioned above, the image collection assembly 12 comprises a diversity element 25 that may be a feature(s) of either the one or more coherent EM radiation sources 20 or the diffractive optics 30. The diversity element 25 refers to one or more feature(s) that implement a change between sample times to cause diversity in the captured image data. In some cases, the diversity element 25 may be provided by configuring the one or more coherent EM radiation sources 20 with multiple light sources (i.e. point emitters) providing illumination from different incidence angles to the specimen during sampling. For example, the coherent EM radiation sources 20 may be in the form of a two-dimensional LED array (n×m dimensions) of LEDs acting as point emitting light sources at different locations at the illumination plane of the two-dimensional LED array. In another example, the diversity element 25 may be provided by mechanically shifting the specimen to different locations at the sample plane using, for example, an X-Y stage. In another example, the diversity element 25 may be provided by configuring the diffractive optics 30 to comprise a spatial light modulator with its display located at the Fourier plane. The diversity can then be generated during sampling by displaying a pattern at different locations on the spatial light modulator display. In yet another example, the diversity element 25 may be provided by shifting another coded mask around the Fourier plane.
In some cases, the one or more coherent EM radiation sources 20 and diffractive optics 30 are configured to operate in a trans-illumination mode directing illumination through the specimen and toward a collection element of the diffractive optics 30. In other cases, the one or more coherent EM radiation sources 20 and diffractive optics 30 are configured in epi-illumination mode directing illumination toward the specimen and away from a collection element of the diffractive optics 30.
In certain aspects, a digital radiation intensity detector comprises a two-dimensional grid of equally spaced discrete elements (e.g., pixels) at a detection plane. At each sample time, each element samples intensity of radiation received. As a group, the grid samples a two-dimensional intensity distribution associated with the location of the elements. The digital radiation intensity detector generates a signal(s) with frames of image data of the intensity distribution measured by the grid of radiation detecting elements at the detection plane at each sample time. If visible light radiation is being used to illuminate the specimen, the digital radiation intensity detector may be in the form of a charge coupled device (CCD), a CMOS imaging sensor, an avalanche photo-diode (APD) array, a photo-diode (PD) array, a photomultiplier tube (PMT) array, or like device. If using THz radiation is used, the digital radiation intensity detector may be, for example, an imaging bolometer. If using microwave radiation, the digital radiation intensity detector may be, for example, an antenna. If X-ray radiation is used, the digital radiation intensity detector may be, for example, an x-ray sensitive CCD. If acoustic radiation is used, the digital radiation intensity detector may be, for example, a piezoelectric transducer array. These examples of digital radiation intensity detectors and others are commercially available. In some aspects, the digital radiation intensity detector may be a color detector e.g., an RGB detector. In other aspects, the digital radiation intensity detector may be a monochromatic detector.
As shown in
The one or more processors 52 may receive instructions stored on the CRM 54 (e.g., memory) and execute those instructions to perform one or more functions of system 10. For example, the processor(s) 52 may execute instructions to perform one or more steps of the imaging with convex relaxation method. For example, the processor(s) 52 may execute instructions stored on the CRM 54 to perform one or more functions of the system 10 such as, for example, 1) interpreting image data, 2) reconstructing a higher resolution image from the image data, and 3) providing display data for displaying one or more images or other output on the display 56. As another example, the processor(s) 52 may provide control instructions for controlling the illumination to the coherent EM radiation source(s) 20. In one case, the processor(s) 52 may provide control instructions to synchronize the illumination by coherent EM radiation source(s) 20 with the sampling times of the digital radiation intensity detector 40. In addition to storing instructions for preforming certain functions of the system 10, the CRM 54 can also store the (lower resolution) intensity and higher resolution image data, and other data produced by the system 10. The display 56 may be a color display or a black and white display. In addition, the display 56 may be a two-dimensional display or a three-dimensional display. In one embodiment, the display 56 may be capable of displaying multiple views.
More specifically,
To collect low-resolution images using the illustrated Fourier ptychographic imaging system with convex relaxation 300, the low NA lens 330 receives illumination altered by the specimen and filters the illumination based on its low NA. The digital radiation intensity detector 340 receives the filtered illumination from the low NA lens 330 and samples a low-resolution image set b comprising a sequence of j low-resolution images at j sample times during illumination by the j illumination sources. Each low-resolution image is captured at a different sample time during illumination from a different incidence angle. As illustrated by the bottom illustration, the system 300 uses an imaging method with convex relaxation that uses a convex lifted ptychographic (CLP) technique to transform the image set b into a high-resolution complex sample image ψ.
To collect low-resolution images using this Fourier ptychographic imaging system with convex relaxation 400, the low NA lens 230 receives illumination altered by the specimen and filters the illumination based on its low NA. The digital radiation intensity detector 240 receives the filtered illumination from the low NA lens 230 and samples a sequence of j low-resolution images at j sample times during illumination by the j illumination sources. Each low-resolution image is captured at a different sample time during illumination from a different incidence angle.
The locations of neighboring apertures have an overlapping area between neighboring apertures such as, for example, the overlapping area 437 between aperture 436(1) and aperture 436(2). When using the reconstruction method with convex relaxation, the overlapping area need only be about 50% or lower of the area of one of the neighboring apertures to converge to a single imaging solution. Conventional FPM reconstruction required a more extensive overlapping area in the range of 80 to 90% in order to converge to an accurate solution. Since more overlap is required, more images and iterations are needed to cover the same area in the conventional systems. Thus, conventional systems required more exposure time and more resources to reconstruct the high resolution image.
II. Ptychographic Imaging Methods with Convex Relaxation
At this juncture in the method, there are two possible options (Option A) go to step 730 or (Option B) go to step 750. Generally speaking, if the convex combined image matrix contains large-scale ptychographic data (i.e. the low resolution images have a high number of pixels and/or the sequence has a high number of low resolution images), then Option A may be the more appropriate option. For example, Option A may be used if the number of pixels is more than 50×50 pixels for each low resolution image and/or the number of images is more than 200. If the convex combined image matrix has smaller scale data, Option B may be the appropriate option.
If Option A is used, the method relaxes the convex combined image matrix into low-rank formulation at step 730. Once in low-rank formulation, the method a minima of the new low-rank formulation to determine a high-resolution image. The method determines the minima based on a low-rank ptychographic (LRP) technique at step 740.
If Option B is used, the method determines a minima of X to determine a high-resolution image at step 750. In this case, the minima is determined using a convex lifted ptychographic (CLP) technique. An example of an appropriate convex solver can be found, for example, in the Templates for First-Order Conic Solvers (TFOCS) of CVS Research, Inc. and California Institute of Technology.
In the subsections that follow, certain steps described with reference to the flowchart in
A. Collection of Low (Reduced) Resolution Images
This section describes example substeps of the collection of low resolution images step 710 of
With respect to the Fourier ptychographic imaging systems in
Based on this point emitter assumption, the jth light source (e.g., LED illuminated in
s(x,j)=ψ(x)eikxp
Where: wavenumber k=2π/λ and
Neglecting scaling factors and a quadratic phase factor for simplicity, a Fourier optics setup gives the field at the imaging system pupil plane, A(x′), as [s(x,j)]={circumflex over (ψ)}(x′−pj). Here,
represents the Fourier transform between conjugate variables x and x′, where {circumflex over (ψ)} is the Fourier transform of ψ, and the Fourier shift property has been applied. The shifted sample spectrum field {circumflex over (ψ)}(x′−pj) is then modulated by the imaging system's aperture function a(x′), which acts as a low-pass filter. In
The spectrum {circumflex over (ψ)} is considered discretized into n pixels with a maximum spatial frequency k. The bandpass cutoff of the aperture function a is denoted as k·m/n, where m is an integer less than n. The modulation of {circumflex over (ψ)} by a results in a field characterized by m discrete samples, which propagates to the imaging plane at g(x) and is sampled by an m-pixel digital radiation intensity detector 240. The m-pixel digital radiation intensity detector 240 samples intensity distribution related to reduced-resolution images that can be combined into a reduced resolution (low resolution) image matrix, g, as:
g(x,j)=|[a(x′){circumflex over (ψ)}(x′−pj)]|2 (Eqn. 2)
Where: g(x,j) is an (m×q) dimensional Fourier ptychographic data matrix.
That is, the jth column contains a low-resolution image of sample intensity while under illumination from the jth optical source.
In the imaging method with convex relaxation, a higher resolution (n-pixel) complex spectrum {circumflex over (ψ)}(x′) is reconstructed from the plurality of low-resolution (m-pixel) intensity measurements contained within the data matrix g. Once {circumflex over (ψ)} is found, an inverse-Fourier transform yields the desired complex sample reconstruction, ψ.
Standard ptychographic systems resolve the inverse problem using alternating projections (AP) strategies: after initializing a complex sample estimate, ψ0, iterative constraints help force ψ0 to obey all known physical conditions. First, its amplitude is forced to obey the measured intensity measurement set from the detector plane (i.e., the values in g). Second, its spectrum {circumflex over (ψ)}0 is forced to lie within a known support in the plane that is Fourier conjugate to the detector. While these AP strategies are known to converge when each constraint set is convex, the intensity constraint applied at the detector plane is not convex, which may sometimes lead to erroneous solutions and/or stagnation in finding a solution.
In one example, the components of the Fourier ptychographic system shown in
B. Generate Convex Combined Image Matrix
In this subsection, an embodiment of the step 720 of
To solve Eqn. 2 as a convex problem, it is expressed in matrix form. First, the unknown sample spectrum {circumflex over (ψ)} is represented as an (n×1) vector where n is the known sample resolution before being reduced by the finite bandpass of the lens aperture. Next, the jth detected lower resolution image becomes an (m×1) vector gj, where m is the number of pixels in each low-resolution image. The ratio n/m defines the ptychographic resolution improvement factor. It is equivalent to the largest angle of incidence from an off-axis optical source, divided by the acceptance angle of the imaging lens. Third, each lens aperture function a(x+pj) is expressed as an (n×1) discrete aperture vector aj, which modulates the unknown sample spectrum {circumflex over (ψ)}.
To rewrite Eqn. 2 as a matrix product {Aj}j=1q is defined as a sequence of (m×n) rectangular matrices that contain a deterministic aperture function aj along a diagonal. For an aberration-free rectangular aperture, each matrix Aj has a diagonal of ones originating at (0, p′j) and terminating at (m, p′j+m−1), where p′j is now a discretized version of the shift variable, pj. Finally, m×m discrete Fourier transform matrix F(m) is introduced to express the transformation of the low-pass filtered sample spectrum through the fixed imaging system for each low-resolution image gj:
gj=|F(m)Aj{circumflex over (ψ)}|2, 1<j<q (Eqn. 3)
The ptychographic system collects a sequence of q lower resolution images, {gj}j=1q such as, for example, at step 710 of
b=|FA{circumflex over (ψ)}|2=|D{circumflex over (ψ)}|2 (Eqn. 4)
This is an example of stacking in substep 910 of
The transpose of the ith row of the D matrix is denoted as di, which is a column vector. The set {di} forms the measurement vectors. The measured intensity in the ith pixel is the square of the inner product between di and the spectrum {circumflex over (ψ)}: bi=|di, {circumflex over (ψ)}
|2.
The method “lifts” the solution {circumflex over (ψ)} out of the quadratic relationship in Eqn. 4 and expresses the solution {circumflex over (ψ)} in the space of (n×n) positive-semidefinite matrices:
bi=Tr({circumflex over (ψ)}*didi*{circumflex over (ψ)})=Tr(didi*{circumflex over (ψ)}{circumflex over (ψ)}*)=Tr(DiX) (Eqn. 6)
where Di=didi* is a rank-1 measurement matrix constructed from the ith measurement vector di, X={circumflex over (ψ)} {circumflex over (ψ)} * is an (n×n) rank-1 outer product, and 1≤i≤q·m. This is an example of constructing measurement matrices at step 920 of
Eqn. 6 states that quadratic image measurements {bi}i=1q·m are linear transforms of {circumflex over (ψ)} in a higher dimensional space. These q·m linear transforms are combined into a single linear operator to summarize the relationship between the stacked image vector b and the matrix X as,
(X)=b.
The phase retrieval problem in ptychography can be posed as the following rank minimization process:
minimize rank(X)
subject to (X)=b,
X>0, where X>0 denotes that X is positive-semidefinite (Eqn. 7)
However, the minimization problem in Eqn. 7 is not convex. In order to transform the minimization program in Eqn. 7 into a convex problem, convex relaxation is performed on Eqn. 7 by replacing the rank of matrix X with its trace, which generates a convex semidefinite program as follows:
minimize Tr(X)
subject to (X)=b,
X>0, (Eqn. 8)
To account for the presence of noise, Eqn. 8 may be reformed such that the measured intensities in b are no longer strictly enforced constraints, but instead appear in the objective function as follows:
minimize αTr(X)+½∥A(X)−b∥
subject to X>0 (Eqn. 9)
Here, α is a scalar regularization variable that directly trades off goodness for complexity of fit. Its optimal value depends upon the assumed noise level. Eqn. 9 forms the final convex combined image matrix problem that can be used to recover a resolution improved complex sample ψ from a set of obliquely illuminated images in b.
Both Eqn. 8 and Eqn. 9 use convex relaxation to generate a convex program with a function to minimize (minimization function) and solution constraints. In addition to generating a convex program, Eqn. 9 also accounts for noise. In certain aspects, the minimization function has no local minima.
If Option B is used in
C. Results from Using Convex Lifted Ptychographic (CLP) Technique
In this Section, some simulated results of using Option B of
In certain aspects, using a CLP process returns a low-rank matrix X, with a rapidly decaying spectrum, as the optimal solution of Eqn. 9. The trace term in the CLP objective function is primarily responsible for enforcing the low-rank structure of X. While this trace term also appears like an alternative method to minimize the unknown signal energy, a fair interpretation should consider its effect in a lifted (n×n) solution space. The final complex image estimate ψ can be obtained by first performing a singular value decomposition of X. Given low-noise imaging conditions and spatially coherent illumination, ψ is set to the Fourier transform of the largest resulting singular vector. Viewed as an autocorrelation matrix, useful statistical measurements may also be found within the remaining smaller singular vectors of X. One may also identify X as the discrete mutual intensity matrix of a partially coherent optical field: X={circumflex over (ψ)}{circumflex over (ψ)}*
, where
denotes an ensemble average. Under this interpretation, Eqn. 9 becomes an alternative solver for the stationary mixed states of a ptychography setup.
Three points distinguish Eqn. 9 from conventional AP-based ptychography strategies. First, the convex CLP process has a larger search space. If AP strategies are used to iteratively update an n-pixel estimate, Eqn. 9 must solve for an n×n positive-semidefinite matrix. Second, this boost in the solution space dimension guarantees that the convex program may find a global optimum with tractable computation. This allows CLP technique to avoid AP's frequent convergence to local minima (i.e., failure to approach the true image). Unlike conventional solvers for the ptychography problem, no local minima exist in the CLP process of embodiments. Finally, Eqn. 9 considers the presence of noise by offering a parameter (a) to tune with an assumed noise level. AP-based solvers lack this parameter and can be easily led into incorrect local minima by even low noise levels as discussed in the following section.
CLP Technique Simulations and Noise Performance
In the example results described in this section, the Fourier ptychographic systems of
In the noiseless case, five (5) iterations of (Non-Convex) AP reconstruction introduced unpredictable artifacts to both the recovered amplitude and phase images as shown in
The AP and CLP reconstructions were repeated again setting α=0.001 in Eqn. 9 while varying two relevant parameters: the number of captured images g, and their signal-to-noise ratio (SNR). In these reconstructions, SNR=10 log10(ψ|2
/
|N2|
), where
|ψ|2
is the mean sample intensity and <|N2|> is the mean intensity of uniform Gaussian noise added to each simulated raw image. To account for the unknown constant phase offset in all phase retrieval reconstructions, the reconstruction mean-squared error defined as MSE=Σx|ψ(x)−ρs(x)|2/Σx|ψ(x)|2, where ρ=Σx ψ(x)s*(x)/Σx|s(x)|2 is a constant phase factor shifting the reconstructed phase to optimally match the known phase of the ground truth sample.
When using the reconstruction method with convex relaxation of certain embodiments, each image spectrum need only overlap with each neighboring image spectrum by about 50% or lower to converge to a single imaging solution. Conventional FPM reconstruction methods require more extensive overlapping, e.g., 80-90%, to converge to an accurate solution. Since more overlap is required in conventional systems, more images and iterations are needed to cover the same Fourier space. Thus, conventional systems require more exposure time and more resources to reconstruct the high resolution image.
D. Factorization for Low-Rank Ptychography (LRP)
Within this subsection, certain details of steps 730 and 740 of
Although generating a convex program based on Eqn. 9 makes reconstruction more efficient, the constraint that X remain positive-semidefinite can be computationally burdensome since each iteration may require a full eigenvalue decomposition of X. In order to process large-scale ptychographic data, certain aspects segment each detected image into tiles (e.g., 103 pixel tile images), and process each tile (segment) separately, and then “tile” the resulting reconstructions back together into a final full resolution image. Tiling parallelization may increase efficiency for processing large-scale ptychographic data.
In certain embodiments, the method of convex relaxation provides another process for processing large-scale ptychographic data that may be used as an alternative or in conjunction with tiling parallelization. In these embodiments, the method may use Option A of
In these cases, instead of solving for an n×n matrix X, the method can use a low-rank ansatz and factorize the matrix X as X=RRT, where the decision variable R is now an n×r rectangular matrix containing complex entries, with r<n. Inserting this factorization into the optimization problem in Eqn. 8 and writing the constraints in terms of the measurement matrix Di=didiT generates the non-convex program,
Minimize Tr(RRT)
subject to Tr(DiRRT)=bi for all i. (Eqn. 10)
Besides removing the positive semidefinite constraint in Eqn. 8, the factored form of Eqn. 10 presents two more key adjustments to the original convex formulation. First, using the relationship Tr(RRT)=∥R∥F2, where F denotes a Frobenius norm, it is direct to rewrite the objective function and each constraint in Eqn. 10 with just one n×r decision matrix, R. Now instead of storing an n×n matrix like CLP, LRP must only store an n×r matrix. Since most practical applications of ptychography require coherent optics, the desired solution rank r will typically be close to 1, thus significantly relaxing storage requirements (i.e., coherent light satisfies X={circumflex over (ψ)}{circumflex over (ψ)}*, so one would expect R as a column vector and RRT a rank-1 outer product). Fixing r at a small value, LRP memory usage now scales linearly instead of quadratically with the number of reconstructed pixels, n. Second, the feasible set of Eqn. 10 is no longer convex and thus, the solution strategy must be shifted away from a simple semidefinite program to use an LRP technique instead.
In one aspect, the method can use the non-convex program of minimizing Tr(RRT)+|R|_1 is used with the same constraints of Eqn. 10. This latter term is the L1 norm of the unknown sample spectrum matrix, R.
At step 740 of
In certain aspects, to solve Eqn. 10, the following augmented Lagrangian function is minimized:
Where RϵCn×r is the unknown decision variable and the two variables yϵRq·m and σϵR+ are parameters to guide the method to its final reconstruction. Each bi is the intensity measured by one image sensor pixel. The LRP technique iteratively minimizes the function L by sequentially updating R and the parameters Y and σ.
The first term in Eqn. 11 is the objective function from Eqn. 10, indirectly encouraging a low-rank factorized product. This tracks the original assumption of a rank-1 solution within a “lifted” solution space. The second term contains the known equality constraints in Eqn. 10 (i.e., the measured intensities), each assigned a weight yi. The third term is a penalized fitting error that is abbreviated with label v. It is weighted by one penalty parameter σ, mimicking the role of a Lagrangian multiplier.
With an appropriate fixed selection of yi's and σ, the minimization of L(R, y, σ) with respect to R identifies the desired optimum of Eqn. 10. Specifically, if a local minimum of L is identified each iteration (which is nearly always the case in practice), then the minimization sequence accumulation point may be a guaranteed solution. As an unconstrained function, the minimum of L can be found quickly using a quasi-Newton approach.
In some aspects, the goal of a low-rank ptychography (LRP) process is to determine a suitable set of (yi, σ) to minimize Eqn. 11 with respect to R, which leads to the desired high resolution image solution. An iterative process is used to sequentially minimize L with respect to Rk at iteration k, and then update a new parameter set (yk+1, σk+1) at iteration k+1. The parameters (yk+1, σk+1) are updated to ensure their associated term's contribution to the summation forming L is relatively small. This suggests Rk+1 is proceeding to a more feasible solution. The relative permissible size of the second and third terms in L are controlled by two important parameters, η<1 and γ>1: if the third term ν sufficiently decreases such that νk+1≤ηνk, then multiplier σ is held fixed and the equality constraint multipliers, yi, is updated. Otherwise, σ is increased by a factor γ such that σk+1=γσk.
In one example, the LRP process is initialized with an estimate of the unknown high-resolution complex sample function ψ0, contained within a low-rank matrix R0. The LRP process terminates if it reaches a sufficient number of iterations or if the minimizer fulfills some convergence criterion. R0 is formed using a spectral method, which can help increase solver accuracy and decrease computation time. Specifically, the r columns of R0 are selected as the leading r eigenvectors of D*diag[b]D, where D is the measurement matrix in Eqn. 4. While this spectral approach works quite well in practice, a random initialization of R0 also often produces an accurate reconstruction.
In certain aspects, the LRP process uses two parameters, γ and η, which help guide its solution process. In most of the included experiments, the LRP process set γ=1.5. The most observable consequence of selecting a different value of γ is its influence on the number of iterations needed for a desired level of performance. A larger value for γ will cause a larger change within the augmented Lagrangian function each iteration, and thus a quicker progression to an optimized reconstruction.
In certain aspects, the LRP process sets η=0.5. A different value of η may also significantly alter the quality of reconstruction output with experimental input images (i.e., in the presence of noise).
LRP Simulations and Noise Performance
Following the same procedure used above to simulate the CLP process, MSE performance of the LRP process was tested as a function of SNR and the results are shown in
These simulations arbitrarily fix η and γ at 0.5 and 1.5, respectively, and set the desired rank of the solution, r, to 1. In some embodiments, these free variables may be changed which offers significant freedom to tune the response of LRP process to noise. For example, similar to the noise parameter α in Eqn. 9, the multiplier σ (controlled via γ) in Eqn. 11 helps trade off complexity for goodness of fit by re-weighting the quadratic fitting error term.
In addition to reducing required memory, the LRP process also improves upon the computational cost of CLP process for large scale data. For an n-pixel sample reconstruction, the per iteration cost of the CLP algorithm is currently O(n3), using big-O notation. The positive-semidefinite constraint in Eqn. 9, which requires a full eigenvalue decomposition, defines this behavior limit. The per-iteration cost of the LRP algorithm, on the other hand, is O(n log n). This large per-iteration cost reduction is the primary source of LRP speedup.
Experimental Results
Experiments were used to verify that the LRP process of certain embodiments can improves the accuracy and noise stability of ptychographic reconstruction by using a Fourier ptychographic (FP) microscope with the configuration of the system in
Quantitative Performance
In these examples, the Fourier ptychographic microscope was comprised of a 15×15 array of surface-mounted LEDs (e.g., model SMD 3528, center wavelength λ=632 nm, 4 mm LED pitch, 150 μm active area diameter), which served as quasi-coherent optical sources. The LED array was placed l=80 mm beneath the sample plane, and each LED has an approximate 20 nm spectral bandwidth.
To quantitatively verify resolution improvement, each of the 15×15 LEDs was turned on beneath a U.S. Air Force (USAF) resolution calibration target. A microscope objective (e.g. 2× e.g., Olympus® objective with apochromatic Plan APO 0.08 NA) transferred each resulting optical field to a CCD detector (e.g., Kodak® KAI-29050 detector with 5.5 μm sized pixels), which sampled 225 low resolution images. Using this 0.08 NA microscope objective (5° collection angle) and a 0.35 illumination NA (θmax=20° illumination angle), the Fourier ptychographic microscope provided a total complex field resolution gain of n/m=25. Each image spectrum overlapped by ol≈70% in area with each neighboring image spectrum. For reconstruction, n=25·m was selected and the same aperture parameters were used with both AP and LRP processes to create the high-resolution images shown in
Both ˜1 megapixel reconstructions achieved their maximum expected resolving power (i.e., resolved Group 9, Element 3: 1.56 μm line pair spacing). This is approximately 5 times sharper than the smallest resolved feature in one raw image (e.g., Group 7, Element 2). The LRP process avoids certain artifacts that are commonly observed during the nonlinear descent of AP process. Both reconstructions slowly fluctuate in background areas that are expected to be uniformly bright or dark. These fluctuations are caused in part by experimental noise, an imperfect aperture function estimate, and possible misalignments in the LED shift values, pj. In a representative background area marked by a 402 pixel blue box in
The LRP process was then demonstrated on a “high-NA” FP microscope configuration comprising a larger 0.5 NA microscope objective lens with a 30° collection angle (e.g., 20× Olympus 0.5 NA UPLFLN). For sample illumination, 28 LEDs are arranged into 3 concentric rings of 8, 8 and 12 evenly spaced light sources (ring radii=16, 32 and 40 mm, respectively). This new light source array was placed 40 mm beneath the sample to create a 0.7 illumination NA with a θmax=45° illumination angle. The synthesized numerical aperture of this FP microscope, computed as the sum of the illumination NA and objective lens NA, is NAs=1.2. With a greater-than-unity synthetic NA, reconstructions can offer oil-immersion quality resolution (˜385 nm smallest resolvable feature spacing), without requiring any immersion medium between the sample and objective lens. A monolayer of polystyrene microspheres (index of refraction nm=1.5 87) immersed in oil (no=1.515, both indexes for λ=632 nm light) was imaged by this high-NA″ FP microscope configuration.
By using a LRP process of certain embodiments, the process was shown to reconstruct quantitatively accurate phase. Using the same data and parameters for AP and LRP input, the high-resolution phase reconstructions were obtained of two adjacent microspheres in
Biological Sample Reconstruction
The third imaging example uses the same high-NA FP configuration (i.e. collection angle=30°, θmax=45°) of the embodiments discussed with respect to
The infected blood sample was prepared to maintain erythrocyte asexual stage cultures of the P. falciparum strain 3D7 in culture medium, then smeared, fixed with methanol, and then a Hema 3 stain was applied. An example sample region containing two infected cells, imaged with a conventional high-NA oil-immersion microscope (NA=1.25) under Kohler illumination. Twenty eight (28) uniquely illuminated images were captured of these two infected cells using the high-NA FP microscope. The top two right images contain an example normally illuminated raw image, which does not clearly resolve the parasite infection. Bottom left six images presents phase retrieval reconstructions using the standard AP algorithm, where m=1202, n=2402, run 6 iterations, and again subtract a constant phase offset. Reconstructions from three data sets were included: images captured with a 1 second exposure (top), a 0.25 second exposure (middle), and 0.1 second exposure (bottom). A shorter exposure time implies increased noise within each raw image. While the 1 sec exposure-based AP reconstruction resolves each parasite, blurred cell boundaries and non-uniform fluctuations in amplitude suggest an inaccurate AP convergence. However, both parasite infections remain visible within the reconstructed phase. The parasites become challenging to resolve within the phase from 0.25 sec exposure data, and are not resolved within the phase from the 0.1 sec exposure data, due to increased image noise. The normalized background variance of each AP amplitude reconstruction, from a representative 402-pixel window (marked blue square), is σ2=0.0020, 0.0027, and 0.0059 for the 1 sec, 0.25 sec, and 0.1 sec exposure reconstructions, respectively.
For comparison, reconstructions using the LRP process are shown (sharpest solutions after 15 iterations). For each reconstructed amplitude, the desired solution rank is set to r=3. The 3 modes of the resulting reconstruction are added in an intensity basis to create the displayed amplitude images. For each reconstructed phase, the desired solution matrix rank is set to r=1 and all other parameters are left unchanged. For all three exposure levels, the amplitude of the cell boundaries remains sharper than in the AP images. Both parasite infections are resolvable in either the reconstructed amplitude or phase, or both, for all three exposure levels. The normalized amplitude variances from the same background window are now σ2=0.0016 (1 sec), 0.0022 (0.25 sec), and 0.0035 (0.1 sec), an average reduction (i.e., improvement) of 26% with respect to the AP results. The AP reconstructions here offer a generally flatter background phase profile than LRP (i.e., less variation at low spatial frequencies). Without additional filtering or post-processing, the AP algorithm here might offer superior quantitative analysis during e.g. tomographic cell reconstruction, where low-order phase variations must remain accurate. However, it is clear within
Through the relaxation in Eqn. 8, the traditionally nonlinear phase retrieval process for ptychography is transformed into a convex program. The convex program can be solved with a CLP process if it is a small-scale image set. If it is a large-scale image set, the convex program can be relaxed into low rank formulation resulting semidefinite program with an appropriate factorization, and then solved with a LRP process. This method of convex relaxation provides a process that is robust to noise.
Besides removing local minima from the recovery process, perhaps the most significant departure from conventional phase retrieval solvers is a tunable solution rank, r. As noted earlier, r connects to statistical features of the ptychographic experiment, typically arising from the partial coherence of the illuminating field. Coherence effects are significant at third-generation X-ray synchrotron sources and within electron microscopes. An appropriately selected r may eventually help LRP process measure the partial coherence of such sources. The solution rank may also help identify setup vibrations, sample auto-fluorescence, or even 3D sample structure. In some cases, the method of convex relaxation can artificially enlarge the solution rank to encourage the transfer of experimental noise into its smaller singular vectors. Other extensions of LRP include simultaneously solving for unknown aberrations (i.e., the shape of the probe in standard ptychography), systematic setup errors, and/or inserting additional sample priors such as sparsity.
III. Subsystems
The various components previously described in the Figures may operate using one or more of the subsystems to facilitate the functions described herein. Any of the components in the Figures may use any suitable number of subsystems to facilitate the functions described herein. Examples of such subsystems and/or components are shown in a
In some embodiments, an output device such as the printer 2430 or display 56 of the Fourier camera system can output various forms of data. For example, the Fourier camera system can output 2D color/monochromatic images (intensity and/or phase), data associated with these images, or other data associated with analyses performed by the Fourier camera system.
Modifications, additions, or omissions may be made to any of the above-described embodiments without departing from the scope of the disclosure. Any of the embodiments described above may include more, fewer, or other features without departing from the scope of the disclosure. Additionally, the steps of the described features may be performed in any suitable order without departing from the scope of the disclosure.
It should be understood that the present invention as described above can be implemented in the form of control logic using computer software in a modular or integrated manner. Based on the disclosure and teachings provided herein, a person of ordinary skill in the art will know and appreciate other ways and/or methods to implement the present invention using hardware and a combination of hardware and software.
Any of the software components or functions described in this application, may be implemented as software code to be executed by a processor using any suitable computer language such as, for example, Java, C++ or Perl using, for example, conventional or object-oriented techniques. The software code may be stored as a series of instructions, or commands on a CRM, such as a random access memory (RAM), a read only memory (ROM), a magnetic medium such as a hard-drive or a floppy disk, or an optical medium such as a CD-ROM. Any such CRM may reside on or within a single computational apparatus, and may be present on or within different computational apparatuses within a system or network.
Although the foregoing disclosed embodiments have been described in some detail to facilitate understanding, the described embodiments are to be considered illustrative and not limiting. It will be apparent to one of ordinary skill in the art that certain changes and modifications can be practiced within the scope of the appended claims.
One or more features from any embodiment may be combined with one or more features of any other embodiment without departing from the scope of the disclosure. Further, modifications, additions, or omissions may be made to any embodiment without departing from the scope of the disclosure. The components of any embodiment may be integrated or separated according to particular needs without departing from the scope of the disclosure.
This application claims priority to U.S. Provisional Patent Application No. 61/992,505 titled “Solving Conventional and Fourier Ptychography with Convex Optimization,” filed on May 13, 2014, which is hereby incorporated by reference in its entirety and for all purposes.
Number | Name | Date | Kind |
---|---|---|---|
5475527 | Hackel et al. | Dec 1995 | A |
6144365 | Young et al. | Nov 2000 | A |
6154196 | Fleck et al. | Nov 2000 | A |
6320648 | Brueck et al. | Nov 2001 | B1 |
6747781 | Trisnadi | Jun 2004 | B2 |
6905838 | Bittner | Jun 2005 | B1 |
7436503 | Chen et al. | Oct 2008 | B1 |
7460248 | Kurtz et al. | Dec 2008 | B2 |
7706419 | Wang et al. | Apr 2010 | B2 |
7787588 | Yun et al. | Aug 2010 | B1 |
8271251 | Schwartz et al. | Sep 2012 | B2 |
8313031 | Vinogradov | Nov 2012 | B2 |
8497934 | Milnes et al. | Jul 2013 | B2 |
8624968 | Hersee et al. | Jan 2014 | B1 |
8942449 | Maiden | Jan 2015 | B2 |
9029745 | Maiden | May 2015 | B2 |
9426455 | Horstmeyer et al. | Aug 2016 | B2 |
9497379 | Ou et al. | Nov 2016 | B2 |
9829695 | Kim et al. | Nov 2017 | B2 |
9864184 | Ou et al. | Jan 2018 | B2 |
9892812 | Zheng et al. | Feb 2018 | B2 |
9983397 | Horstmeyer et al. | May 2018 | B2 |
9993149 | Chung et al. | Jun 2018 | B2 |
9998658 | Ou et al. | Jun 2018 | B2 |
20010055062 | Shioda et al. | Dec 2001 | A1 |
20020141051 | Vogt et al. | Oct 2002 | A1 |
20030116436 | Amirkhanian et al. | Jun 2003 | A1 |
20040146196 | Van Heel | Jul 2004 | A1 |
20040190762 | Dowski, Jr. et al. | Sep 2004 | A1 |
20050211912 | Fox | Sep 2005 | A1 |
20060098293 | Garoutte et al. | May 2006 | A1 |
20060158754 | Tsukagoshi et al. | Jul 2006 | A1 |
20060173313 | Liu et al. | Aug 2006 | A1 |
20060291707 | Kothapalli et al. | Dec 2006 | A1 |
20070057184 | Uto | Mar 2007 | A1 |
20070133113 | Minabe et al. | Jun 2007 | A1 |
20070159639 | Teramura | Jul 2007 | A1 |
20070171430 | Tearney et al. | Jul 2007 | A1 |
20070189436 | Goto et al. | Aug 2007 | A1 |
20080101664 | Perez | May 2008 | A1 |
20090046164 | Shroff et al. | Feb 2009 | A1 |
20090079987 | Ben-Ezra et al. | Mar 2009 | A1 |
20090125242 | Choi et al. | May 2009 | A1 |
20090284831 | Schuster et al. | Nov 2009 | A1 |
20090316141 | Feldkhun | Dec 2009 | A1 |
20100135547 | Lee et al. | Jun 2010 | A1 |
20100271705 | Hung | Oct 2010 | A1 |
20110075928 | Jeong et al. | Mar 2011 | A1 |
20110192976 | Own et al. | Aug 2011 | A1 |
20110235863 | Maiden | Sep 2011 | A1 |
20120069344 | Liu | Mar 2012 | A1 |
20120099803 | Ozcan et al. | Apr 2012 | A1 |
20120105618 | Brueck et al. | May 2012 | A1 |
20120118967 | Gerst | May 2012 | A1 |
20120157160 | Ozcan et al. | Jun 2012 | A1 |
20120218379 | Ozcan et al. | Aug 2012 | A1 |
20120248292 | Ozcan | Oct 2012 | A1 |
20120250032 | Wilde | Oct 2012 | A1 |
20120281929 | Brand | Nov 2012 | A1 |
20130083886 | Carmi et al. | Apr 2013 | A1 |
20130093871 | Nowatzyk | Apr 2013 | A1 |
20130094077 | Brueck | Apr 2013 | A1 |
20130100525 | Chiang et al. | Apr 2013 | A1 |
20130170767 | Choudhury | Jul 2013 | A1 |
20130182096 | Boccara | Jul 2013 | A1 |
20130223685 | Maiden | Aug 2013 | A1 |
20140007307 | Routh, Jr. et al. | Jan 2014 | A1 |
20140029824 | Shi | Jan 2014 | A1 |
20140043616 | Maiden | Feb 2014 | A1 |
20140050382 | Adie et al. | Feb 2014 | A1 |
20140118529 | Zheng et al. | May 2014 | A1 |
20140126691 | Zheng et al. | May 2014 | A1 |
20140152801 | Fine et al. | Jun 2014 | A1 |
20140153692 | Larkin et al. | Jun 2014 | A1 |
20140160236 | Ozcan et al. | Jun 2014 | A1 |
20140160488 | Zhou | Jun 2014 | A1 |
20140267674 | Mertz et al. | Sep 2014 | A1 |
20140347672 | Pavillon | Nov 2014 | A1 |
20140368812 | Humphry | Dec 2014 | A1 |
20150036038 | Horstmeyer et al. | Feb 2015 | A1 |
20150054979 | Ou et al. | Feb 2015 | A1 |
20150160450 | Ou et al. | Jun 2015 | A1 |
20150264250 | Ou et al. | Sep 2015 | A1 |
20160088205 | Horstmeyer et al. | Mar 2016 | A1 |
20160178883 | Horstmeyer et al. | Jun 2016 | A1 |
20160202460 | Zheng | Jul 2016 | A1 |
20160210763 | Horstmeyer et al. | Jul 2016 | A1 |
20160216208 | Kim et al. | Jul 2016 | A1 |
20160216503 | Kim et al. | Jul 2016 | A1 |
20160266366 | Chung et al. | Sep 2016 | A1 |
20160320595 | Horstmeyer et al. | Nov 2016 | A1 |
20160320605 | Ou et al. | Nov 2016 | A1 |
20160341945 | Ou et al. | Nov 2016 | A1 |
20170178317 | Besley et al. | Jun 2017 | A1 |
20170273551 | Chung et al. | Sep 2017 | A1 |
20170299854 | Kim et al. | Oct 2017 | A1 |
20170354329 | Chung et al. | Dec 2017 | A1 |
20170363853 | Besley | Dec 2017 | A1 |
20170371141 | Besley | Dec 2017 | A1 |
20180088309 | Ou et al. | Mar 2018 | A1 |
Number | Date | Country |
---|---|---|
101408623 | Apr 2009 | CN |
101868740 | Oct 2010 | CN |
101872033 | Oct 2010 | CN |
102608597 | Jul 2012 | CN |
103201648 | Jul 2013 | CN |
2007-299604 | Nov 2007 | JP |
2010-012222 | Jan 2010 | JP |
10-1998-0075050 | Nov 1998 | KR |
WO9953469 | Oct 1999 | WO |
WO 2002102128 | Dec 2002 | WO |
WO 2003062744 | Jul 2003 | WO |
WO 2008-116070 | Sep 2008 | WO |
WO 2011-093043 | Aug 2011 | WO |
WO 2012037182 | Mar 2012 | WO |
WO 2014070656 | May 2014 | WO |
WO 2015017730 | Feb 2015 | WO |
WO 2015027188 | Feb 2015 | WO |
WO 2016090331 | Jun 2016 | WO |
WO 2016106379 | Jun 2016 | WO |
WO 2016118761 | Jul 2016 | WO |
WO 2016123156 | Aug 2016 | WO |
WO 2016123157 | Aug 2016 | WO |
WO 2016149120 | Sep 2016 | WO |
WO 2016187591 | Nov 2016 | WO |
WO 2017081539 | May 2017 | WO |
WO 2017081540 | May 2017 | WO |
WO 2017081542 | May 2017 | WO |
Entry |
---|
A. Chai et al., Array imaging using intensity-only measurements, 2011, IOP Publishing, pp. 1-16. |
A. Maiden et al., Superresolution imaging via ptychography, 2011, J.Opt. Soc. Am., vol. 28, No. 4, pp. 604-612. |
U.S. Appl. No. 14/960,252 filed Dec. 4, 2015 entitled “Multiplexed Fourier Ptychography Imaging Systems and Methods”. |
U.S. Appl. No. 14/979,154 filed Dec. 22, 2015 entitled “EPI-Illumination Fourier Ptychographic Imaging for Thick Samples”. |
Office Action dated Oct. 5, 2015 in U.S. Appl. No. 14/065,305. |
Notice of Allowance dated Dec. 4, 2015 in U.S. Appl. No. 14/065,305. |
International Search Report and Written Opinion dated Feb. 21, 2014 in PCT/US2013/067068. |
International Preliminary Report on Patentability dated May 14, 2015 in PCT/US2013/067068. |
International Search Report and Written Opinion dated Dec. 5, 2014 in PCT/US2014/052351. |
International Search Report and Written Opinion dated Nov. 13, 2014 in PCT/US2014/049297. |
“About Molemap,” [Downloaded from internet at http://molemap.net.au/about-us/], 2 pages. |
“Doctor Mole—Skin Cancer App,” [Downloaded from internet at http://www.doctormole.com], 1 page. |
“Immersion Media,” Olympus, Microscopy Resource Center, http://www.olympusmicro.com/primer/anatomy/immersion.html. |
“Lytro,” [Downloaded from internet at https://www.lytro.com/], 6 pages. |
“Melafind,” [Downloaded from internet at http://www.melafind.com/], 4 pages. |
“TFOCS: Templates for First-Order Conic Solvers,” CVX Research, CVX Forum, http://cvxr.com/tfocs/. |
Maiden, A. et al., “A new method of high resolution, quantitative phase scanning microscopy,” in: M.T. Postek, D.E. Newbury, S.F. Platek, D.C. Joy (Eds.), SPIE Proceedings of Scanning Microscopy, 7729, 2010. |
Alexandrov, S. A. et al., “Synthetic Aperture Fourier holographic optical microscopy,” Phys. Rev. Left. 97, 168102 (2006). |
Alexandrov, S. et al., “Spatial information transmission beyond a system's diffraction limit using optical spectral encoding of the spatial frequency,” Journal of Optics A: Pure and Applied Optics 10, 025304 (2008). |
Arimoto, H. et al. “Integral three-dimensional imaging with digital reconstruction,” Opt. Lett. 26, 157-159 (2001). |
Balan, R. et al., “Painless reconstruction from magnitudes of frame coefficients,” J Fourier Anal Appl 15:488-501 (2009). |
Bauschke, HH et al., “Phase retrieval, error reduction algorithm, and Fienup variants: a view from convex optimization,” J Opt Soc Am A 19:1334-1345 (2002). |
Becker, S. et al., “Templates for convex cone problems with applications to sparse signal recovery,” Technical report, Department of Statistics, Stanford University, (2010), 48 Pages. |
Betti, R., et al., “Observational study on the mitotic rate and other prognostic factors in cutaneous primary melanoma arising from naevi and from melanoma de novo,” Journal of the European Academy of Dermatology and Venereology, 2014. |
Bian, L. et al., “Fourier ptychographic reconstruction using Wirtinger flow optimization,” Opt. Express 23:4856-4866 (2015). |
Bian, Z. et al., “Adaptive system correction for robust Fourier ptychographic imaging,” Optics express, 2013. 21(26): p. 32400-32410. |
Blum, A. et al, “Clear differences in hand-held dermoscopes,” JDDG: Journal der Deutschen Dermatologischen Gesellschaft, 2006, 4(12): p. 1054-1057. |
Blum, A., et al., Dermatoskopie von Hauttumoren: Auflichtmikroskopie; Dermoskopie; digitale Bildanalyse; mit 28 Tabellen. 2003: Springer DE, Chapter 4 “Dermatoskopisch sichtbare Strukturen” p. 15-66. |
Brady, D. et al., “Multiscale gigapixel photography,” Nature 486, 386-389 (2012). |
Burer S, Monteiro RDC (2003) A nonlinear programming algorithm for solving semidefinite programs via low-rank factorization. Math Program, Ser B 95:329-357. |
Burer, S. et al., “Local minima and convergence in low-rank semidefmite programming. Math Program,” Ser A 103:427-444 (2005). |
Candes, EJ. Et al., “Phase retrieval via matrix completion,” SIAM J. Imaging Sci. 6:199-225 (2012). |
Candes, EJ. Et al., “PhaseLift: exact and stable signal recovery from magnitude measurements via convex programming.,” Comm Pure Appl Math 66:1241-1274 (2013). |
Candes, EJ. et al., “Soltanolkotabi M Phase retrieval via Wirtinger flow: theory and algorithms,” IEEE Trans. Info. Theory 61:1985-2007 (2015). |
Chen, T. et al., “Polarization and phase shifting for 3D scanning of translucent objects,” Proc. CVPR, (2007). |
Chin, L. et al., “Malignant melanoma: genetics and therapeutics in the genomic era,” Genes & development, 2006, 20(16): p. 2149-2182. |
Colomb, T. et al., “Automatic procedure for aberration compensation in digital holographic microscopy and applications to specimen shape compensation,” Appl. Opt. 45, 851-863 (2006). |
De Sa, C. et al., “Global convergence of stochastic gradient descent for some non convex matrix problems,” Proc. 32nd Int. Conf. Machine Learning (2015). |
Denis, L. et al., “Inline hologram reconstruction with sparsity constraints,” Opt. Left. 34, pp. 3475-3477 (2009). |
Di, J. et al., “High resolution digital holographic microscopy with a wide field of view based on a synthetic aperture technique and use of linear CCD scanning,” Appl. Opt. 47, pp. 5654-5659 (2008). |
Dierolf, M. et al., “Ptychographic coherent diffractive imaging of weakly scattering specimens,” New J. Phys. 12, 035017 (2010). |
Dong, S. et al., “Aperture-scanning Fourier ptychography for 3D refocusing and super-resolution macroscopic imaging,” pp. 13586-13599 (Jun. 2, 2014). |
Eldar, Y.C. et al., “Sparse phase retrieval from short-time Fourier measurements,” IEEE Signal Processing Letters 22, No. 5 (2015): 638-642. |
Emile, O. et al., “Rotating polarization imaging in turbid media,” Optics Letters 21(20), (1996). |
Faulkner, H. et al., “Movable aperture lensless transmission microscopy: a novel phase retrieval algorithm,” Phys. Rev. Left. 93, 023903 (2004). |
Faulkner, H. M. L. et al., “Error tolerance of an iterative phase retrieval algorithm for moveable illumination microscopy,” Ultramicroscopy 103(2), 153-164 (2005). |
Fazel, M. (2002) Matrix rank minimization with applications. PhD thesis (Stanford University, Palo Alto, CA). |
Feng, P. et al., “Long-working-distance synthetic aperture Fresnel off-axis digital holography,” Optics Express 17, pp. 5473-5480 (2009). |
Fienup, J. R., “Invariant error metrics for image reconstruction,” Appl. Opt. 36(32), 8352-8357 (1997). |
Fienup, J. R., “Lensless coherent imaging by phase retrieval with an illumination pattern constraint,” Opt. Express 14, 498-508 (2006). |
Fienup, J. R., “Phase retrieval algorithms: a comparison,” Appl. Opt. 21, 2758-2769 (1982). |
Fienup, J. R., “Reconstruction of a complex-valued object from the modulus of its Fourier transform using a support constraint,” J. Opt. Soc. Am. A 4, 118-123 (1987). |
Fienup, J. R., “Reconstruction of an object from the modulus of its Fourier transform,” Opt. Lett. 3, 27-29 (1978). |
Gan, X. et al., “Image enhancement through turbid media under a microscope by use of polarization gating methods,” JOSA A 16(9), (1999). |
Ghosh, A. et al., “Multiview face capture using polarized spherical gradient illumination,” ACM Transactions on Graphics 30(6) (2011). |
Goodman J. “Introduction to Fourier Optics,” Roberts & Company Publication, Third ⋅ Edition, chapters 1-6, pp. 1-172 (2005). |
Goodson, A.G., et al., “Comparative analysis of total body and dermatoscopic photographic monitoring of nevi in similar patient populations at risk for cutaneous melanoma,” Dermatologic Surgery, 2010. 36(7): p. 1087-1098. |
Granero, L. et al., “Synthetic aperture superresolved microscopy in digital lensless Fourier holography by time and angular multiplexing of the object information,” Appl. Opt. 49, pp. 845-857 (2010). |
Grant, M. et al., “CVX: Matlab software for disciplined convex programming,” version 2.0 beta. http://cvxr.com/cvx, (Sep. 2013), 3 pages. |
Greenbaum, A. et al., “Increased space—bandwidth product in pixel super-resolved lensfree on-chip microscopy,” Sci. Rep. 3, p. 1717 (2013). |
Guizar-Sicairos M. “Phase retrieval with transverse translation diversity: a nonlinear optimization approach,” Opt. Express 16, 7264-7278 (2008). |
Gunturk, B. K. et al., “Image Restoration: Fundamentals and Advances,” vol. 7, Chapter 3, pp. 63-68 (CRC Press, 2012). |
Gustafsson, M. G., “Surpassing the lateral resolution limit by a factor of two using structured illumination microscopy,” J. Microsc. 198, 82-87 (2000). |
Gutzler, T. et al., “Coherent aperture-synthesis, wide-field, high-resolution holographic microscopy of biological tissue,” Opt. Lett. 35, pp. 1136-1138 (2010). |
Hillman, T. R. et al., “High-resolution, wide-field object reconstruction with synthetic aperture Fourier holographic optical microscopy,” Opt. Express 17, pp. 7873-7892 (2009). |
Hong, S-H. et al., “Three-dimensional volumetric object reconstruction using computational integral imaging,” Opt. Express 12, 483-491 (2004). |
Hoppe, W., “Diffraction in inhomogeneous primary wave fields. 1. Principle of phase determination from electron diffraction interference,” Acta Crystallogr. A25, 495-501 1969. |
Horstmeyer, R. et al., “A phase space model of Fourier ptychographic microscopy,” Optics Express, 2014. 22(1): p. 338-358. |
Horstmeyer, R. et al., “Overlapped fourier coding for optical aberration removal,” Manuscript in preparation, 19 pages (2014). |
Hüe, F. et al., “Wave-front phase retrieval in transmission electron microscopy via ptychography,” Phys. Rev. B 82, 121415 (2010). |
Humphry, M. et al., “Ptychographic electron microscopy using high-angle dark-field scattering for sub-nanometre resolution imaging,” Nat. Commun. 3, 730 (2012). |
Rodenburg, J., “Ptychography and related diffractive imaging methods,” Adv. Imaging Electron Phys.150, 87-184 (2008). |
Jaganathan, K. et al., “Phase retrieval with masks using convex optimization,” IEEE International Symposium on Information Theory Proceedings (2015): 1655-1659. |
Jaganathan, K. et al., “Recovery of sparse 1-D signals from the magnitudes of their Fourier transform,” IEEE International Symposium on Information Theory Proceedings (2012): 1473-1477. |
Jaganathan, K. et al., “STFT Phase retrieval: uniqueness guarantees and recovery algorithms,” arXiv preprint arXiv:1508.02820 (2015). |
Sun, J. et al., “Coded multi-angular illumination for Fourier ptychography based on Hadamard codes,” 5 pages (2015). |
Kim, M. et al., “High-speed synthetic aperture microscopy for live cell imaging,” Opt. Lett. 36, pp. 148-150 (2011). |
Kittler, H., et al., Morphologic changes of pigmented skin lesions: a useful extension of the ABCD rule for dermatoscopy. Journal of the American Academy of Dermatology, 1999. 40(4): p. 558-562. |
Levoy, M. et al., “Light field microscopy,” ACM Trans. Graphics 25, (2006). |
Levoy, M. et al., “Recording and controlling the 4D light field in a microscope using microlens arrays,” J. Microsc. 235 (2009). |
Li X. et al., “Sparse signal recovery from quadratic measurements via convex programming,” SIAM Journal on Mathematical Analysis 45, No. 5 (2013): 3019-3033. |
Lohmann, A. W., Dorsch, R. G., Mendlovic, D., Zalevsky, Z. & Ferreira, C., “Space—bandwidth product of optical signals and systems,” J. Opt. Soc. Am. A 13, pp. 470-473 (1996). |
Lue, N. et al., “Live Cell Refractometry Using Hilbert Phase Microscopy and Confocal Reflectance Microscopy,” The Journal of Physical Chemistry A, 113, pp. 13327-13330 (2009). |
Maiden et al., “Superresolution imaging via ptychography,” Journal of the Optical Society of America A. Apr. 2011, vol. 28 No. 4, pp. 604-612. |
Maiden, A. et al., “An improved ptychographical phase retrieval algorithm for diffractive imaging,”Ultramicroscopy 109(10), 1256-1262 (2009). |
Maiden, A. M. et al., “Optical ptychography: a practical implementation with useful resolution,” Opt. Lett. 35, 2585-2587 (2010). |
Marchesini S., “A unified evaluation of iterative projection algorithms for phase retrieval,” Rev Sci Instrum 78:011301 (2007). |
Marchesini S. et al., “Augmented projections for ptychographic imaging,” Inverse Probl 29:115009 (2013). |
Marrison, J. et al., “Ptychography—a label free, high-contrast imaging technique for live cells using quantitative phase information,” Sci. Rep. 3, 2369 (2013). |
Miao et al., “High Resolution 3D X-Ray Diffraction Microscopy,” Physical Review Letters, Aug. 19, 2002, vol. 89, No. 8, pp. 1-4. |
Mico, V. et al., “Synthetic aperture microscopy using off-axis illumination and polarization coding,” Optics Communications, pp. 276, 209-217 (2007). |
Mico, V. et al., “Synthetic aperture superresolution with multiple off-axis holograms,” JOSA A 23, pp. 3162-3170 (2006). |
Mir M. et al., “Blood screening using diffraction phase cytometry,” Journal of Biomedical Optics 15, pp. 027016-027014 (2010). |
Mir, M. et al., “Optical measurement of cycle-dependent cell growth,” Proceedings of the National Academy of Sciences 108, pp. 13124-13129 (2011). |
Nayar, S. K. et al., “Fast separation of direct and global components of a scene using high frequency illumination,” ACM Transactions on Graphics 25(3) (2006). |
Ng, R. et al., “Light field photography with a hand-held plenoptic camera”, Computer Science Technical Report CSTR, 2005. 2(11). |
Nomura, H. et al., “Techniques for measuring aberrations in lenses used in photolithography with printed patterns,” Appl. Opt. 38(13), 2800-2807 (1999). |
Ohlsson, H. et al., “Compressive phase retrieval from squared output measurements via semidefinite programming,” arXiv:1111.6323 (2011). |
Ou, X. et al., “High numerical aperture Fourier ptychography: principle, implementation and characterization,” Opt. Express 23:3472-3491 (2015). |
Ou, X.. et al., “Quantitative phase imaging via Fourier ptychographic microscopy,” Optics Letters, 2013. 38(22): p. 4845-4848. |
Ou. et al., “Embedded pupil function recovery for Fourier ptychographic microscopy,” Optics Express 22 (5), pp. 4960-4972 (2014). |
Balan, R. et al., “On signal reconstruction without phase, Applied and Computational Harmonic Analysis 20,” No. 3 (2006): 345-356. |
Recht, B. et al., “Guaranteed minimum-rank solutions of linear matrix equations via nuclear norm minimization,” SIAM Review 52, No. 3 (2010): 471-501. |
Reinhard, E. et al., “High Dynamic Range Imaging: Acquisition, Display, and Image-based Lighting,” (Morgan Kaufmann, 2010). |
Rodenburg, J. M. et al., “A phase retrieval algorithm for shifting illumination,” Appl. Phys. Left. 85, 4795-4797 (2004). |
Rodenburg, J. M. et al., “Hard-X-ray lensless imaging of extended objects,” Phys. Rev. Left. 98, 034801 (2007). |
Rodenburg, J. M. et al., “The theory of super-resolution electron microscopy via Wigner-distribution deconvolution,” Phil. Trans. R. Soc. Lond. A 339, 521-553 (1992). |
Schnars, U. et al., “Digital recording and numerical reconstruction of holograms,” Measurement Science and Technology, 13, R85 (2002). |
Schwarz, C. J. et al., “Imaging interferometric microscopy,” Optics letters 28, pp. 1424-1426 (2003). |
Shechner, Y.Y.et al., “Polarization-based vision through haze,” Applied Optics 42(3), (2003). |
Shechtman, Y. et al., “Sparsity based sub-wavelength imaging with partially incoherent light via quadratic compressed sensing,” Opt Express 19:14807-14822 (2011). |
Siegel, R. et al. “Cancer statistics 2013,” CA: a cancer journal for clinicians, 2013. 63(1): p. 11-30. |
Stoecker, W.V., R.K. Rader, and A. Halpern, Diagnostic Inaccuracy of Smartphone Applications for Melanoma Detection: Representative Lesion Sets and the Role for Adjunctive Technologies. JAMA Dermatology, 2013. 149(7): p. 884. |
Sun, D. L. et al., “Estimating a signal from a magnitude spectrogram via convex optimization,” arXiv:1209.2076 (2012). |
Thibault, P. et al., “Probe retrieval in ptychographic coherent diffractive imaging,” Ultramicroscopy 109(4), 338-343 (2009). |
Thibault P. et al., “High-resolution scanning X-ray diffraction microscopy,” Science 321, 379-382 (2008). |
Thomas L. et al.. Semiological value of ABCDE criteria in the diagnosis of cutaneous pigmented tumors. Dermatology, 1998. 197(1): p. 11-17. |
Tippie, A.E. et al., “High-resolution synthetic-aperture digital holography with digital phase and pupil correction,” Opt. Express 19, pp. 12027-12038 (2011). |
Turpin, T. et al., “Theory of the synthetic aperture microscope,” pp. 230-240 (1995). |
Tyson, R., “Principles of Adaptive Optics” (CRC Press, 2010). |
Mahajan V. N. “Zernike circle polynomials and optical aberrations of systems with circular pupils,” Appl. Opt. 33(34), 8121-8124 (1994). |
Waldspurger, I. et al., “Phase recovery, maxcut and complex semidefinite programming,” Mathematical Programming 149, No. 1-2 (2015): 47-81. |
Wang. Z. et al., “Tissue refractive index as marker of disease,” Journal of Biomedical Optics 16, 116017-116017 (2011). |
Watanabe, M. et al., “Telecentric optics for focus analysis,” IEEE trans. pattern. anal. mach. intell., 19 1360-1365 (1997). |
Wesner J. et al., “Reconstructing the pupil function of microscope objectives from the intensity PSF,” in Current Developments in Lens Design and Optical Engineering III, R. E. Fischer, W. J. Smith, and R. B. Johnson, eds., Proc. SPIE 4767, 32-43 (2002). |
Wolf, J.A. et al., “Diagnostic Inaccuracy of Smartphone Applications for Melanoma Detection,” JAMA Dermatology, 2013, 149(7): p. 885-885. |
Wu, J. et al., “Focal plane tuning in wide-field-of-view microscope with Talbot pattern illumination,” Opt. Lett. 36, 2179-2181 (2011). |
Wu, J. et al., “Wide field-of-view microscope based on holographic focus grid illumination,” Opt. Lett. 35, 2188-2190 (2010). |
Xu, W. et al., “Digital in-line holography for biological applications,” Proc. Natl Acad. Sci. USA 98, pp. 11301-11305 (2001). |
Yuan, C. et al., “Angular multiplexing in pulsed digital holography for aperture synthesis,” Optics Letters 33, pp. 2356-2358 (2008). |
Zhang Y. et al., “Self-learning based fourier ptychographic microscopy,” Optics Express, 16pgs (2015). |
Zheng, G. et al.. “Characterization of spatially varying aberrations for wide field-of-view microscopy,” Opt. Express 21, 15131-15143 (2013). |
Zheng, G. et al., “Microscopy refocusing and dark-field imaging by using a simple LED array,” Opt. Lett. 36, 3987-3989 (2011). |
Zheng, G. et al., “Sub-pixel resolving optofluidic microscope for on-chip cell imaging,” Lab Chip 10, pp. 3125-3129 (2010). |
Zheng, G. et al., “The ePetri dish, an on-chip cell imaging platform based on subpixel perspective sweeping microscopy (SPSM),” Proc. Natl Acad. Sci. USA 108, pp. 16889-16894 (2011). |
Zheng, G. et at, “Wide-field, high-resolution Fourier ptychographic microscopy,” Nature Photonics (2013). |
Zheng, G.A. et al., “0.5 gigapixel microscopy using a flatbed scanner,” Biomed. Opt. Express 5, 1-8 (2014). |
Tian, L. et al., “Multiplexed Coded Illumination for Fourier Ptychography with an LED Array Microscope,” Optical Society of America, 14 pages (2014). |
Schechner, Y., “Multiplexing for Optimal Lighting,” IEEE Transactions on Pattern Analysis and Machine Intelligence, vol. 29, No. 8, 1339-1354 (2007). |
Ma, W. et al., “Rapid Acquisition of Specular and Diffuse Normal Maps from Polarized Spherical Gradient Illumination,” University of Southern California, Institute for Creative Technologies, 12 pages (2007). |
Rowe, M.P. et al., “Polarization-difference imaging: a biologically inspired technique for observation through scattering media,” Optics Letters, vol. 20, No. 6, 3 pages. (1995). |
Gruev, V. et al., “Dual-tier thin film polymer polarization imaging sensor,” Optics Express, vol. 18, No. 18, 12 pages. (2010). |
Preliminary Amendment dated Mar. 17, 2014 filed in U.S. Appl. No. 14/065,280. |
Preliminary Amendment dated Nov. 28, 2016 filed in U.S. Appl. No. 15/206,859. |
Preliminary Amendment dated Mar. 17, 2014 filed in U.S. Appl. No. 14/065,305. |
Preliminary Amendment dated Nov. 28, 2016 filed in U.S. Appl. No. 15/209,604. |
U.S. Notice of Allowance dated Jan. 14, 2016 in U.S. Appl. No. 14/448,850. |
U.S. Notice of Allowance dated Jan. 22, 2016 in U.S. Appl. No. 14/466,481. |
U.S. Notice of Allowance dated Apr. 13, 2016 in U.S. Appl. No. 14/448,850. |
U.S. Notice of Allowance dated Apr. 22, 2016 in U.S. Appl. No. 14/466,481. |
U.S. Office Action dated Jul. 14, 2016 in U.S. Appl. No. 15/007,196. |
U.S. Notice of Allowance dated Aug. 23, 2016 in U.S. Appl. No. 14/466,481. |
U.S. Office Action dated Aug. 16, 2016 in U.S. Appl. No. 14/065,280. |
U.S. Office Action dated Sep. 16, 2016 I U.S. Appl. No. 14/065,305. |
U.S. Notice of Allowance dated Nov. 2, 2016 in U.S. Appl. No. 14/572,493. |
U.S. Office Action dated Nov. 22, 2016 in U.S. Appl. No. 15/003,559. |
U.S. Supplemental Notice of Allowance dated Dec. 12, 2016 in U.S. Appl. No. 14/572,493. |
U.S. Notice of Allowance dated Jan. 13, 2017 in U.S. Appl. No. 14/065,305. |
U.S. Final Office Action dated Jan. 23, 2017 in U.S. Appl. No. 15/007,196. |
U.S. Office Action dated Feb. 21, 2017 in U.S. Appl. No. 14/960,252. |
U.S. Supplemental Notice of Allowability dated Mar. 2, 2017 in U.S. Appl. No. 14/065,305. |
U.S. Notice of Allowance dated Mar. 8, 2017 in U.S. Appl. No. 14/572,493. |
U.S. Office Action dated Mar. 13, 2017 in U.S. Appl. No. 14/658,019. |
U.S. Notice of Allowance dated Mar. 22, 2017 in U.S. Appl. No. 15/007,196. |
U.S. Notice of Allowance dated Mar. 31, 2017 in U.S. Appl. No. 14/572,493. |
U.S. Final Office Action dated Apr. 3, 2017 in U.S. Appl. No. 14/065,280. |
U.S. Notice of Allowance dated Jun. 9, 2017 in U.S. Appl. No. 14/065,305. |
U.S. Notice of Allowance dated Jun. 9, 2017 in U.S. Appl. No. 15/206,859. |
U.S. Notice of Allowance dated Jun. 9, 2017 in U.S. Appl. No. 15/007,196. |
U.S. Notice of Allowance dated Jun. 20, 2017 in U.S. Appl. No. 14/572,493. |
U.S. Supplemental Notice of Allowance dated Jun. 28, 2017 in U.S. Appl. No. 15/206,859. |
U.S. Final Office Action dated Jul. 27, 2017 in U.S. Appl. No. 15/003,559. |
U.S. Notice of Allowance dated Aug. 16, 2017 in U.S. Appl. No. 15/209,604. |
U.S. Notice of Allowance dated Sep. 1, 2017 in U.S. Appl. No. 15/206,859. |
European Third-Party Observations, dated Jan. 20, 2016 in EP Application No. 13851670.3. |
European Extended Search Report dated Mar. 31, 2016 in EP Application No. 13851670.3. |
International Preliminary Report on Patentability dated Mar. 3, 2016 issued in PCT/US2014/052351. |
International Preliminary Report on Patentability dated Feb. 11, 2016 issued in PCT/US2014/049297. |
International Search Report and Written Opinion dated Feb. 22, 2016 issued in PCT/US2015/064126. |
International Search Report and Written Opinion dated Apr. 19, 2016 issued in PCT/US2015/067498. |
International Search Report and Written Opinion dated May 4, 2016 issued in PCT/US2016/015001. |
International Search Report and Written Opinion dated May 11, 2016 issued in PCT/US2016/015002. |
International Search Report and Written Opinion dated Jun. 27, 2016 issued in PCT/US2016/022116. |
International Search Report and Written Opinion dated Jun. 30, 2016 issued in PCT/US2016/014343. |
International Search Report and Wrtitten Opinion dated Sep. 5, 2016 issued in PCT/US2016/033638. |
Chinese Office Action [Description in English] dated Jul. 11, 2016 issued in Application No. CN 201380068831.6. |
Chinese Office Action [Description in English] dated Dec. 13, 2016 issued in Application No. CN201480057911.6. |
Extended European Search Report dated Feb. 16, 2017 issued in Application No. 14837844.1. |
Extended European Search Report dated Feb. 15, 2017 issued in Applicatoin No. 14832857.8. |
Chinese Second Office Action [Description in English] dated Feb. 17, 2017 issued in Application No. CN201380068831.6. |
International Preliminary Report on Patentability dated Jun. 15, 2017 issued in Application No. PCT/US2015/064126. |
European Office Action dated May 16, 2017 issued in European Patent Application No. 13851670.3. |
International Preliminary Report on Patentability dated Jul. 6, 2017 issued in Application No. PCT/US2015/067498. |
International Preliminary Report on Patentability dated Aug. 3, 2017 issued in Application No. PCT/US2016/014343. |
International Preliminary Report on Patentability dated Aug. 10, 2017 issued in Application No. PCT/US2016/015001. |
International Preliminary Report on Patentability dated Aug. 10, 2017 issued in Application No. PCT/US2016/015002. |
Chinese Third Office Action [Summary in English] dated Jul. 24, 2017 issued in Application No. 201380068831.6. |
Chinese First Office Action [Summary in English] dated Aug. 2, 2017 issued in Application No. CN 201480054301.0. |
Abramowitz, M., et al, “Field Curvature,” Olympus Microscopy Resource Center, 2012 Olympus America Inc., pp. 1-3. [retrieved on Feb. 24, 2016] <URL:http://www.olympusmicro.com/primer/anatomy/fieldcurvature.html>. |
Age-Related Macular Degeneration (AMD)|National Eye Institute. 2010 Table, pp. 1-8. [retrieved Apr. 5, 2016] <URL: https://www.nei.nih.gov/eyedata/amd#top.>. |
Bian, L., et al, “Fourier ptychographic reconstruction using Poisson maximum likelihood and truncated Wirtinger gradient,” Nature Publishing Group; Scientific Reports, vol. 6, No. 27384, Jun. 10, 2016, pp. 1-10. <doi: 10.1038/srep27384>. |
BioTek® Brochure: BioTek's Multi-Mode Microplate Reading Techonologies, BioTek Instruments, Inc. pp. 2. [retrieved on Mar. 14, 2016] <URL: http://www.biotek.com>. |
Bishara, W., et al, “Holographic pixel super-resolution in portable lensless on-chip microscopy using a fiber-optic array,” NIH-PA, Lab Chip, Author manuscript; available in PMC Aug. 8, 2011, pp. 1-9. (Published in final edited form as: Lab Chip. Apr. 7, 2011; 11(7): 1276-1279. <doi:10.1039/c01c00684j>). |
Bishara, W., et al, “Lensfree on-chip microscopy over a wide field-of-view using pixel super-resolution,” Optics Express, vol. 18, No. 11, May 24, 2010, pp. 11181-11191. <doi: 10.1364/OE.18.011181>. |
Born, M., et al, “Principles of Optics: Electromagnetic theory of propagation, interference and diffraction of light,” Seventh (Expanded) Edition, Cambridge University Press, England 1999, pp. 1-31. [ISBN 0 521 642221 hardback]. |
Bunk, O., et al, “Influence of the overlap parameter on the convergence of the ptychographical iterative engine,” Ultramicroscopy, vol. 108, (2008), pp. 481-487. <doi: 10.1016/j.ultramic.2007.08.003>. |
Carroll, J., “Adaptive Optics Retinal Imaging: Applications for Studying Retinal Degeneration,” Archives of Ophthalmology, vol. 126, No. 6, Jun. 9, 2008, pp. 857-858. [retrieved Feb. 24, 2016] <doi:10.1001/archopht.126.6.857>. |
Chao, W. et al, “Soft X-ray microscopy at a spatial resolution better than 15 nm,” Nature|Letters, vol. 435, Jun. 30, 2005, pp. 1210-1213. <doi:10.1038/nature03719>. |
Choi, W., et al, “Tomographic phase microscopy,” NPG: Nature Methods | Advance Online Publication, Aug. 12, 2007, pp. 1-3. <doi:10.1038/NMETH1078>. |
Chung, J., et al, “Counting White Blood Cells from a Blood Smear Using Fourier Ptychographic Microscopy,” PLoS ONE, vol. 10, No. 7, Jul. 17, 2015, pp. 1-10. <doi:10.1371/journal.pone.0133489>. |
Chung, J., et al, “Wide field-of-view fluorescence image deconvolution with aberration-estimation from Fourier ptychography,” Biomedical Optics Express, vol. 7, No. 2, Feb. 1, 2016, pp. 352-368. <doi: 10.1364/BOE.7.000352>. |
Chung, J., et al, pre published manuscript of“Wide-field Fourier ptychographic microscopy using laser illumination source,” ArXiv e-prints (Submitted on Feb. 9, 2016 (v1), last revised Mar. 23, 2016 (this version, v2)). [retrieved on May 20, 2016] <URL:arXiv:1602.02901v2 [physics.optics] Mar. 23, 2016>. |
Debailleul, M., et al, “High-resolution three-dimensional tomographic diffractive microscopy of transparent inorganic and biological samples,” Optics Letters, Optical Society of America, vol. 34, No. 1, Jan. 1, 2009, pp. 79-81. <doi: 10.1364/OL.34.000079>. |
Dierolf, M., et al, “Ptychographic X-ray computed tomography at the nanoscale,” Nature|Letter, vol. 467, Sep. 23, 2010, pp. 436-439. <doi:10.1038/nature09419>. |
Dong, S., et al, “FPscope: a field-portable high-resolution microscope using a cellphone lens,” Biomedical Optics Express, vol. 5, No. 10, Oct. 1, 2014, pp. 3305-3310. <doi:10.1364/BOE.5.003305>. |
Dong, S., et al, “High-resolution fluorescence imaging via pattern-illuminated Fourier ptychography,” Optics Express, vol. 22, No. 17, Aug. 25, 2014, pp. 20856-20870. <doi:10.1364/OE.22.020856>. |
Essen BioScience, “Real-time, quantitative live-cell analysis, IncuCyte® ZOOM System,” IncuCyte Zoom System Brochure 2016, pp. 1-4. [retrieved Feb. 25, 2016] [URL: http:/!www.essenbioscience.com/IncuCyte]. |
Gerke T.D., et al, “Aperiodic volume optics,” Nature Photonics, vol. 4, Feb. 7, 2010, pp. 188-193. <doi:10.1038/nphoton.2009.290>. |
Godara, P., et al, “Adaptive Optics Retinal Imaging: Emerging Clinical Applications,” NIH-PA Author Manuscript; available in PMC Dec. 1, 2011. Published in final edited form as: Optom. Vis. Sci.. Dec. 2010; 87(12): 930-941. <doi: 10.1097/OPX.0b013e3181ff9a8b>. |
Greenbaum, A., et al, “Field-portable wide-field microscopy of dense samples using multi-height pixel super-resolution based lensfree imaging,” Lab Chip, The Royal Society of Chemistry, vol. 12, No. 7, Jan. 31, 2012, pp. 1242-1245. [retrieved on Feb. 27, 2016] <URL:http://dx.doi.org/10.1039/C2LC21072J>. |
Guo, K., et al, “Optimization of sampling pattern and the design of Fourier ptychographic illuminator,” Optics Express, vol. 23, No. 5, Mar. 9, 2015, pp. 6171-6180. <doi: 10.1364/OE.23.006171>. |
Haigh, S. J., et al, “Atomic structure imaging beyond conventional resolution limits in the transmission electron microscope,” Physical Review Letters, vol. 103, Issue 12, Sep. 18, 2009, pp. 126101.1-126101.4. <doi:10.1103/PhysRevLett.103.126101>. |
Han, C., et al, “Wide Field-of-View On-Chip Talbot Fluorescence Microscopy for Longitudinal Cell Culture Monitoring from within the Incubator” Analytical Chemistry, vol. 85, No. 4, Jan. 28, 2013, pp. 2356-2360. <doi:10.1021/ac303356v>. |
Hofer, H., et al, “Dynamics of the eye's wave aberration,” Journal of Optical Society of America A., vol. 18, No. 3, Mar. 2001, pp. 497-506. <doi: 10.1364/JOSAA.18.000497>. |
Hofer, H., et al, “Organization of the Human Trichromatic Cone Mosaic,” The Journal of Neuroscience, vol. 25, No. 42, Oct. 19, 2005, pp. 9669-9679. <doi: 10.1523/JNEUROSCI.2414-05.2005>. |
Hoppe, W., “Diffraction in inhomogeneous primary wave fields. 1. Principle of phase determination from electron diffraction interference.” Acta Crystallographica Section a—Crystal Physics Diffraction Theoretical and General Crystallography, A25, Jan. 1, 1969, pp. 495-501. (English Machine Translation Incl.). |
Horstmeyer, R., et al, “Diffraction tomography with Fourier ptychography,” Optica, Optical Society of America, vol. 3, No. 8, Aug. 2016, pp. 827-835. <doi:10.1364/OPTICA.3.000827>. |
Horstmeyer, R., et al, “Digital pathology with Fourier Ptychography,” Computerized Medical Imaging and Graphics, vol. 42, Jun. 2015, pp. 38-43. <doi:10.1016/j.compmedimag.2014.11.005>. |
Horstmeyer, R., et al, “Solving ptychography with a convex relaxation,” New Journal of Physics, vol. 17, May 27, 2015, pp. 1-14. <doi: 10.1088/1367-2630/17/5/053044> [URL: http://iopscience.iop.org/1367-2630/17/5/053044]. |
Horstmeyer, R., et al, “Standardizing the resolution claims for coherent microscopy,” Nature Photonics | Commentary, vol. 10, No. 2, Feb. 2016, pp. 68-71. <doi:10.1038/nphoton.2015.279> [URL: http://dx.doi.org/10.1038/nphoton.215.279]. |
Joeres, S., et al, “Retinal Imaging With Adaptive Optics Scanning Laser Ophthalmoscopy in Unexplained Central Ring Scotoma,” Arch. Ophthalmol., vol. 126, No. 4, Apr. 2008, pp. 543-547. [retrieved Jun. 10, 2015] [URL: http://archopht.jamanetwork.com/]. |
Jung, J.H., et al, Author Manuscript of“Microfluidic-integrated laser-controlled microactuators with on-chip microscopy imaging functionality,” Published in final edited form as: Lab Chip, Oct. 7, 2014, vol. 14, No. 19, pp. 3781-3789. <doi: 10.1039/c41c00790e>. |
Kawata, S. et al, “Optical microscope tomography. I. Support constraint,” Journal Optical Society America A, vol. 4, No. 1, Jan. 1987, pp. 292-297. <doi:10.1364/JOSAA.4.000292>. |
Kay, D. B., et al, Author Manuscript of “Outer Retinal Structure in Best Vitelliform Macular Dystrophy,” Published in final edited form as: JAMA Ophthalmol., Sep. 2013, vol. 131, No. 9, pp. 1207-1215. <doi: 10.1001/jamaophthalmo1.2013.387>. |
Kim, J., et al, “Incubator embedded cell culture imaging system (EmSight) based on Fourier ptychographic microscopy,” Biomedical Optics Express, vol. 7, No. 8, Aug. 1, 2016, pp. 3097-3110. <doi: 10.1364/BOE.7.003097>. |
Kim, M., et al, “High-speed synthetic aperture microscopy for live cell imaging,” Optics Letters, vol. 36, No. 2, Jan. 15, 2011, pp. 148-150. <doi:10.1364/OL.36.000148>. |
Kirkland, A.I., et al, “Multiple beam tilt microscopy for super resolved imaging,” Journal of Electron Microscopy (Tokyo) Jan. 1, 1997, vol. 46, No. 1, pp. 11-22. [doi: 10.1093/oxfordjournals.jmicro.a023486]. |
Kirkland, A.I., et al, “Super-resolution by aperture synthesis: tilt series reconstruction in CTEM,” Elsevier Science B.V., Ultramicroscopy 57, Mar. 1995, pp. 355-374. <doi: 10.1016/0304-3991(94)00191-O>. |
Kozak, I., “Retinal imaging using adaptive optics technology,” Saudi Journal of Ophthalmology, vol. 28, No. 2, Feb. 25, 2014, pp. 117-122. <doi:10.1016/j.sjopt.2014.02.005>. |
Lauer, V., “New Approach to optical diffraction tomography yielding a vector equation of diffraction tomography and a novel tomographic microscope,” Journal of Microscopy, Feb. 2002, vol. 205, No. 2, pp. 165-176. <doi: 10.1046/j.0022-2720.2001.00980.x>. |
Lee, K., et al, “Synthetic Fourier transform light scattering,” Optics Express, vol. 21, No. 19, Sep. 23, 2013, pp. 22453-22463. <doi:10.1364/OE.21.022453>. |
Lu, H., et al, “Quantitative phase imaging and complex field reconstruction by pupil modulation differential phase contrast,” Optics Express, vol. 24, No. 22, Oct. 31, 2016, pp. 25345-25361. <doi:10.1364/OE.24.025345>. |
LUXEXCEL® Brochure, “LUXEXCEL: 3D Printing Service Description” pp. 1-5. [retrieved on Mar. 7, 2016] <URL: http://www.luxexcel.com. |
Medoff, B.P., et al, “Iterative convolution backprojection algorithms for image reconstruction from limited data,” Journal of the Optical Society of America, vol. 73, No. 11, Nov. 1, 1983, pp. 1493-1500. <doi: 10.1364/JOSA.73.001493>. |
Meyer, R.R., et al, “A new method for the determination of the wave aberration function of high-resolution TEM. 2. Measurement of the antisymmetric aberrations,” Ultramicroscopy, vol. 99, No. 2-3, May 2004, pp. 115-123. <doi: 10.1016/j.ultramic.2003.11.001>. |
Moreno, I., “Creating a desired lighting pattern with an LED array,” Proceedings of SPIE, Eighth International Conference on Solid State Lighting, vol. 705811, Sep. 2, 2008, pp. 9. <doi:10.1117/12.795673>. |
Mrejen, S., et al, “Adaptive Optics Imaging of Cone Mosaic Abnormalities in Acute Macular Neuroretinopathy,” Ophthalmic Surgery, Lasers & Imaging Retina, vol. 45, No. 6, Nov./Dec. 2014, pp. 562-569. <doi: 10.3928/23258160-20141118-12>. |
Ou, X., et al, “Aperture scanning Fourier ptychographic microscopy,” Biomedical Optics Express, vol. 7, No. 8, Aug. 1, 2016, pp. 3140-3150. <doi:10.1364/BOE.7.003140>. |
Ou. X., et al, pre published manuscript of “Embedded pupil function recovery for Fourier ptychographic microscopy,” (submitted on Dec. 26, 2013 (this version, v1); revised Feb. 12, 2014; accepted Feb. 17, 2014; published Feb. 24, 2014) pp. 1-13. <doi: 10.1364/0E.22.004960>. |
Pacheco, S., et al, “Reflective Fourier Ptychography,” Journal of Biomedical Optics, vol. 21, No. 2, Feb. 18, 2016, pp. 026010-1-026010-7. <doi: 10.1117/1.JBO.21.2.026010> [retrieved on Mar. 8, 2016] <URL: http://biomedicaloptics.spiedigitallibrary.org>. |
Phillips, Z., et al, “Multi-Contrast Imaging and Digital Refocusing on a Mobile Microscope with a Domed LED Array,” PLoS One, vol. 10, No. 5, May 13, 2015, pp. 1-13. <doi:10.1371/journal.pone.0124938>. |
Reinhard, E., et al, “High Dynamic Range Imaging: Acquisition, Display, and Image-based Lighting” Second Edition § 5.2 HDR Image Capture: Morgan Kaufmann, May 28, 2010, pp. 148-151. <ISBN: 9780123749147>. |
Rossi, E.A., et al, “In vivo imaging of retinal pigment epithelium cells in age related macular degeneration,” Biomedical Optics Express, vol. 4, No. 11, Nov. 1, 2013, pp. 2527-2539. <doi: 10./1364/BOE.4.0025271. |
Sankaranarayanan, Aswin C., et al, “CS-MUVI: Video Compressive Sensing for Spatial-Multiplexing Cameras,” Proceedings of the IEEE International Conference Computational Photography (ICCP), Apr. 2012, pp. 11. <doi:10.1109/ICCPhot.2012.6215212>. |
Tam, K., et al, “Tomographical imaging with limited-angle input,” Journal of the Optical Society of America, vol. 71, No. 5, May 1981, pp. 582-592. <doi:doi.org/10.1364/JOSA.71.000582>. |
Tian, L., et al, “3D differential phase-contrast microscopy with computational illumination using an LED array,” Optics Letters, vol. 39, No. 5, Mar. 1, 2014, pp. 1326-1329. <doi:10.1364/OL39.001326>. |
Tian, L., et al, “Computional illumination for high-speed in vitro Fourier ptychographic microscropy,” Optica: Research Article, vol. 2, No. 10, Oct. 14, 2015, pp. 904-911. <doi:10.1364/OPTICA.2.000904>. |
Vulovic, M., et al, “When to use the projection assumption and the weak-phase object approximation in phase contrast cryo-EM,” Ultramicroscopy, vol. 136, Jan. 2014, pp. 61-66.<doi: 10.1016/j.ultramic.2013.08.002>. |
Wang, Q., et al, “Adaptive Optics Microperimetry and OCT Images Show Preserved Function and Recovery of Cone Visibility in Macular Telangiectasia Type 2 Retinal Lesions,” Investigative Ophthalmology Visual Science, vol. 56, No. 2, Feb. 2015, pp. 778-786. <doi:10.1167/iovs.14-15576> [retrieved on Apr. 5, 2016] [URL: http://iovs.arvojournals.org]. |
Williams, A., et al, “Fourier ptychographic microscopy for filtration-based circulating tumor cell enumeration and analysis,” Journal of Biomedical Optics, vol. 19, No. 6, Jun. 20, 2014, pp. 066007.1-66007.8. <doi:10.1117/1.JBO.19.6.066007> [retrieved Feb. 10, 2016] <URL:http://biomedicaloptics.spiedigitallibrary.org>. |
Wu, J., et al, “Harmonically matched grating-based full-field quantitative high-resolution phase microscope for observing dynamics of transparent biological samples,” Optics Express, vol. 15, No. 26, Dec. 24, 2007, pp. 18141-18155. <doi:10.1364/OE.15.018141>. |
Wu, J., et al, “Paired-angle-rotation scanning optical coherence tomography forward-imaging probe,” Optics Letters, vol. 31, No. 9, May 1, 2006, pp. 1265-1267. <doi:10.1364/OL.31.001265>. |
Yeh, et al., “Experimental robustness of Fourier ptychography phase retrieval algorithms,” Optics Express, vol. 23, No. 26, Dec. 28, 2015, pp. 33214-33240. <doi: 10.1364/OE.23.033214>. |
Zeiss, C., “Microscopy: Cells Need the Perfect Climate. System Solutions for Live Cell Imaging under Physiological Conditions,” ZEISS Product Brochure, Carl Zeiss Microscopy GmbH Co., Feb. 2008, pp. 42. <URL: http://www.zeiss.de/incubation>. |
Zhang, Y., et al, “Photoreceptor perturbation around subretinal drusenoid deposits as revealed by adaptive optics scanning laser ophthalmoscopy,” HHS Public Access, Am J Ophthalmol. Author Manuscript,Sep. 1, 2015, pp. 22. (Published in final edited form as: Am J Ophthalmol. Sep. 2014; 158(3): 584-96.e1.). |
Zheng, G., “Fourier Ptychographic Imaging: A MATLAB tutorial,” IOP Concise Physics, Morgan & Claypool Publication, San Rafael, CA., May 2016, pp. 96. <ISBN: 978-1-6817-4272-4 (ebook)> <doi: 10.1088/978-1-6817-4273-1>. |
U.S. Appl. No. 15/081,659, filed Mar. 25, 2016, Chung, J. et al. |
U.S. Appl. No. 15/620,674, filed Jun. 12, 2017, Chung, J. et al. |
U.S. Appl. No. 15/636,494, filed Jun. 28, 2017, Kim, J. et al. |
Office Action dated May 19, 2017 in U.S. Appl. No. 15/081,659. |
Office Action dated Aug. 31, 2017 in U.S. Appl. No. 15/636,494. |
Notice of Allowance dated Sep. 20, 2017 in U.S. Appl. No. 15/007,196. |
Notice of Allowance dated Oct. 11, 2017 in U.S. Appl. No. 14/572,493. |
Notice of Allowance dated Oct. 20, 2017 in U.S. Appl. No. 15/081,659. |
Office Action dated Nov. 3, 2017 in U.S. Appl. No. 15/068,389. |
Office Action Interview Summary dated May 3, 2018 in U.S. Appl. No. 15/068,389. |
Final Office Action dated Jun. 6, 2018 issued in U.S. Appl. No. 15/068,389. |
Office Action dated Nov. 30, 2017 in U.S. Appl. No. 15/007,159. |
Notice of Allowance dated Dec. 4, 2017 in U.S. Appl. No. 14/065,305. |
Final Office Action dated Dec. 14, 2017 in U.S. Appl. No. 14/960,252. |
Final Office Action dated Jan. 17, 2018 in U.S. Appl. No. 14/658,019. |
Notice of Allowance dated Jan. 23, 2018 in U.S. Appl. No. 15/206,859. |
Office Action dated Jan. 25, 2018 in U.S. Appl. No. 14/065,280. |
Notice of Allowance dated Jan. 26, 2018 in U.S. Appl. No. 15/209,604. |
Notice of Allowance dated Feb. 9, 2018 in U.S. Appl. No. 15/081,659. |
Office Action dated Apr. 4, 2018 issued in U.S. Appl. No. 15/003,559. |
Office Action dated Apr. 13, 2018 issued in U.S. Appl. No. 15/160,941. |
European Extended Search Report dated Jun. 6, 2018 issued in Application No. 15865492.1. |
Australian Office Action dated Sep. 18, 2017 issued in Application No. AU 2014296034. |
International Preliminary Report on Patentability dated Sep. 28, 2017 issued in Application No. PCT/US2016/022116. |
Japanese Office Action dated Oct. 17, 2017 issued in Application No. 2015-539884. |
Chinese Office Action [Summary in English] dated Oct. 26, 2017 issued in CN 201480057911.6 . |
International Preliminary Report on Patentability dated Nov. 30, 2017 issued in PCT/US2016/033638. |
Australian Examination Report 1/Office Action dated Jan. 18, 2018 issued in AU 2014308673. |
Chinese First Office Action dated Feb. 24, 2018 issued in CN 201680003937.1. |
Abrahamsson, S., et al., “Fast multicolor 3D imaging using aberration-corrected mulitfocus microscopy,” Brief Communications: Nature Methods, vol. 10, No. 1, Jan. 2013, pp. 60-65. <doi:10.1038/nmeth.2277>. |
Holloway, J., et al. “SAVI: Synthetic apertures for long-range, subdiffraction-limited visible imaging using Fourier ptychography,” Science Advances | Research Article, vol. 3, No. 4, Apr. 14, 2017, pp. 1-11. <doi:10.1126/sciadv.1602564> [retrieved on Nov. 28, 2017] <URL:http://advances.sciencemag.org/>. |
Jenson, et al. “Types of imaging, Part 2: An Overview of Fluorescence Microscopy.” The Anatomical Record, vol. 295, No. 10, Oct. 1, 2012, pp. 1621-1627. |
Kner, P., “Phase diversity for three-dimensional imaging,” Journal of the Optical Society of America A, vol. 30, No. 10, Oct. 1, 2013, pp. 1980-1987. <doi:10.1364/JOSAA.30.001980>. |
Wills, S., “Synthetic Apertures for the Optical Domain,” Optics & Photonics News Article [webpage], The Optical Society (OSA), Apr. 18, 2017, pp. 2. <URL:https://www.osa-opn.org/home/newsroom/2017/april/synthetic_apertures_for_the_optical_domain/>. |
Zheng, G., et al, “Wide-field, high-resolution Fourier ptychographic microscopy,” Nature Photonics, vol. 7, Sep. 2013, Published Online Jul. 28, 2013, pp. 739-746. <doi:10.1038/NPHOTON.2013.187>. |
U.S. Appl. No. 15/963,966, filed Apr. 26, 2018, Ou et al. |
U.S. Appl. No. 15/959,050, filed Apr. 20, 2018, Horstmeyer et al. |
Preliminary Amendment dated Jun. 13, 2018 filed in U.S. Appl. No. 15/820,295. |
U.S. Notice of Allowance dated Jun. 27, 2018 in U.S. Appl. No. 15/636,494. |
U.S. Notice of Allowance dated Jul. 16, 2018 in U.S. Appl. No. 15/007,159. |
Extended European Search Report dated Jul. 3, 2018 issued in Application No. EP 15874344.3. |
Jacques, et al., “Imaging Superficial Tissues With Polarized Light,” Lasers in Surgery and Medicine, vol. 26, No. 2, Apr. 25, 2000, pp. 119-129. |
Sarder, et al. “Deconvolution Methods for 3-D Fluorescence Microscopy Images,” IEEE Signal Processing Magazine, vol. 23, No. 3, May 2006, pp. 32-45. |
Preliminary Amendment filed Jul. 11, 2018 in U.S. Appl. No. 15/959,050. |
Japanese First Office Action dated Jul. 31, 2018 issued in Application No. JP 2016-531919. |
Extended European Search Report dated Aug. 8, 2018 issued in Application No. EP 16744002.3. |
Chinese Second Office Action dated Jul. 3, 2018 issued in Application No. CN 201480054301.0. |
Chinese Third Office Action dated Jul. 13, 2018 issued in CN 201480057911.6. |
Preliminary Amendment filed Jul. 23, 2018 in U.S. Appl. No. 15/963,966. |
Chinese Office Action [Description in English] dated May 31, 2016 issued in Application No. CN 201380068831.6. |
Chinese Second Office Action [Description in English] dated Jan. 22, 2017 issued in Application No. CN201380068831.6. |
Notice of Allowance dated Sep. 17, 2018 in U.S. Appl. No. 15/820,295. |
U.S. Office Action dated Oct. 4, 2018 in U.S. Appl. No. 14/658,019. |
U.S. Notice of Allowance dated Oct. 5, 2018 in U.S. Appl. No. 15/636,494. |
U.S. Office Action dated Sep. 7, 2018 in U.S. Appl. No. 14/979,154. |
European Extended Search Report dated Aug. 14, 2018 issued in EP 16744003.1. |
Extended European Search Report dated Sep. 12, 2018 issued in Application No. EP 16740769.1. |
U.S. Appl. No. 16/162,271, filed Oct. 16, 2018, Kim et al. |
Number | Date | Country | |
---|---|---|---|
20150331228 A1 | Nov 2015 | US |
Number | Date | Country | |
---|---|---|---|
61992505 | May 2014 | US |