Variations in tissue perfusion have critically important consequences throughout medicine. This can be evident when not enough perfusion is available to keep tissues alive, when perfusion is restored to tissue after an acute event interrupting flow to that tissue, and when an additional source of blood flow, such as a bypass graft, is created to increase perfusion to the tissue supplied by a diseased vessel.
There are two general classifications of tissue perfusion variation, revascularization and devascularization.
Revascularization occurs when an intervention is performed to increase or restore blood flow to tissue, either by pharmacologic, catheter-based, or surgical interventions. The physiological benefit of successful revascularization is not only angiographic vessel patency, but in addition a demonstrable increase in tissue perfusion in the tissue supplied by flow within that vessel. In both circumstances, angiographic patency (vessel or graft) is one traditional marker of success. A more recently emerging consideration in the literature is the functional or physiologic success of revascularization, which is an index of the increase in perfusion to the tissue supplied by the vessel that was revascularized.
Devascularization is when tissue is deprived, either artificially or through a disease process, of enough blood flow and perfusion to compromise tissue viability. This can occur in a wide variety of surgical procedures, such as when tissue reconstruction flaps are created, or when a bowel tumor is removed and an anastomosis is performed. In these cases, maintenance of a normal threshold of perfusion to all parts of the tissue is critical to overall clinical procedural success, and to the avoidance of complications.
An example of revascularization that illustrates this principle is the setting of coronary artery bypass grafs (CABG). Here, where a stenotic area of narrowing in the vessel is bypassed, the increase in tissue perfusion results from a combination of flow down the bypass graft and the native vessel.
An example of devascularization that illustrates this principle is breast reconstruction after mastectomy, where removal of all or part of the breast is performed because of cancer. The remaining skin and underlying tissue needs to be stretched (“expanded”) to create a new breast; these skin and tissue edges can be devascularized in this process, resulting in wound breakdown and scar tissue formation.
In both these examples, the ability to directly assess perfusion at the time of surgery creates the opportunity to generate new, important information for decision-making. Examples include 1) measurement of the physiologic benefit of revascularization in CABG in a way quite distinctive and supplemental to angiographic graft patency alone; and 2) measurement leading to the avoidance of areas of tissue devascularization, which would decrease the incidence of complications from this surgical procedure.
Accordingly, there is a need for an analysis platform to intra-operatively visualize, display, analyze, and quantify angiography, perfusion, and the change in angiography and perfusion in real-time in tissues imaged by indocyanine green (ICG) near-infrared (NIR) fluorescence angiography technology (ICG-NIR-FA).
Some embodiments of the present invention provide for the derivation of unique analyzed data from ICG-NIR-FA that describes simple and complex angiography and perfusion, and their combination, across multiple clinical applications of the imaging technology.
In all embodiments, we define the term Full Phase Angiography (FPA) as consisting of three phases: 1) an arterial phase, 2) a micro-vascular phase, and 3) a venous phase. More specifically, the arterial phase is an arteriographic inflow phase, 2) the micro-vascular phase is a tissue perfusion phase in between phases 1 and 3, and 3) the venous phase is the venous outflow phase.
In some embodiments, Full Phase Angiography can be derived from any ICG-NIR-FA video, if properly captured. A properly captured video in this context would be one captured according to a protocol standardized with respect to time, dosage and image parameters.
In further embodiments, it has been determined that these three phases can be captured and elucidated in essentially all applications of the ICG-NIR-FA studied clinically thus far, and should be present in all applications of the technology assessing tissue perfusion with angiography. The characteristics of the real-time video generated by the NIR-FA system will vary according to the clinical application, in terms of length and image capture characteristics, but included in each image video are data for these three phases in all application areas. Importantly, the image capture characteristics need to be optimized in order to capture data from all three phases for the subsequent analysis platform to be accurate in its application. Therefore, the specific image capture characteristics are linked to the subsequent analysis. This approach substantially reduces the need for surgeons to make subjective judgments regarding perfusion and patency.
In still other embodiments, using our discovery of these full phase angiographic characteristics in fluorescent angiography, we have developed a core analytic platform for combined angiography and perfusion analysis, using these and other embodiments described herein. The core analytic platform is the basis for all assessments of perfusion across surgical specialties.
In still other embodiments, the core platform has been and can be extended to be applicable across Clinical Application Areas studied to date, and has been designed to be extended to new Clinical Application Areas where angiography and perfusion are important for intraoperative and experimental decision-making. Examples of Clinical Application Areas, not intended to be limiting in any way, are plastic and reconstructive surgery, wound care, vascular surgery and GI surgery.
In still other embodiments, this core analytic platform and its Clinical Application Area-specific component secondary applications are based on the following principles:
1) In some embodiments, by analyzing the arterial phase, angiographic inflow can be assessed (similar to conventional angiography). However, unlike some conventional angiography studies, the real-time characteristics of this inflow under true physiologic conditions can be readily imaged, assessed and evaluated. An example of this type of analysis is the real-time, intra-operative imaging of competitive flow in the context of CABG.
2) In some embodiments, by analyzing both the arterial phase and the microvascular phase, tissue perfusion can be imaged, assessed and quantified. An example in this context is the imaging of limb perfusion in vascular surgery.
3) In some embodiments, by analyzing the venous phase, venous congestion and outflow from tissue problems can be imaged, assessed and quantified. An example in this context is the assessment of possible venous congestion in breast reconstruction surgery.
4) In some embodiments, by capturing all three phases, with the appropriate image acquisition protocol, a complete description of the combination of angiography and perfusion as applied to that clinical application setting can be acquired and analyzed in real-time. This type of analysis might be performed in the context of esophageal or GI surgery.
5) In some embodiments, by capturing all three phases, with the appropriate image acquisition protocol, this complete description of the combination of angiography and perfusion can be evaluated against important, physiologic changes in hemodynamics and/or other conditions that would affect these angiography and perfusion comparison results.
6) In some embodiments, because this NIR imaging technology allows for capture of real-time physiology and changes over time, a dynamic analysis platform is necessary to fully describe these changes over time and accurately reflect physiology. A static, single “snapshot” analytical approach can't and won't accurately describe these physiologic changes, and is not representative of the physiologic changes that are captured by this full phase angiography analysis.
In still other embodiments, each Clinical Application Area and procedure within that Clinical Application Area relies on a certain combination of phase information derived from the FPA; this combination may be relatively specific for that procedure. All Clinical Application Areas and procedures, however, rely at a minimum on information from at least two phases, emphasizing the requirement for a dynamic analytical approach.
In further embodiments, because the anatomy and physiology varies across these Clinical Application Areas, a core analytic platform has been developed with characteristics that are applicable across all applications; additions to this core analytic platform make up the specific analytical toolkits used in each of the Clinical Application Areas.
In further embodiments, because this fluorescence technology captures information in the near-infrared (NIR) spectrum, the standard display is in 255 grey scale black and white. With the development of the analysis platform, new color schemes based on the full phase angiography components have been developed to highlight the arterial, microvascular (perfusion) and venous phases differently, based on the same NIR image. An accurate depiction of the underlying physiology requires more than just the NIR black and white image display.
In still further embodiments, because in some Clinical Application Areas there is a need to evaluate perfusion to multiple anatomic areas at the same operative setting, capturing the metadata imbedded in each of the individual analyses and combining these data into 2-D and 3-D representations is an important component and attribute of the analytic platform. These representations, in turn, are best presented as dynamic displays. Solely by way of example, in the cardiac surgery context, NIR fluorescence imaging can be performed on multiple coronary artery grafts and the data can be aggregated together to produce a dynamic 3D image of the heart showing all of the grafts and the resulting changes in perfusion of the heart muscle.
It is noted that aspects of the invention described with respect to some embodiments, may be incorporated in different embodiments although not specifically described relative thereto. That is, all embodiments and/or features of any embodiment can be combined in any way and/or combination. These and other objects and/or aspects of the present invention are explained in detail in the specification set forth below. Further features, advantages and details of the present invention will be appreciated by those of ordinary skill in the art from a reading of the figures and the detailed description of the embodiments that follow, such description being merely illustrative of the present invention.
The patent or application file contains at least one drawing executed in color. Copies of this patent or patent application publication with color drawing(s) will be provided by the Office upon request and payment of the necessary fee.
Embodiments of the present invention will now be described more fully hereinafter with reference to the accompanying figures, in which preferred embodiments of the invention are shown. The invention may, however, be embodied in many different forms and should not be construed as limited to the embodiments set forth herein.
The terminology used herein is for the purpose of describing particular embodiments only and is not intended to be limiting of the invention. As used herein, the singular forms “a”, “an” and “the” are intended to include the plural forms as well, unless the context clearly indicates otherwise. It will be further understood that the terms “comprises” and/or “comprising,” when used in this specification, specify the presence of stated features, integers, steps, operations, elements, and/or components, but do not preclude the presence or addition of one or more other features, integers, steps, operations, elements, components, and/or groups thereof. As used herein, the term “and/or” includes any and all combinations of one or more of the associated listed items. As used herein, phrases such as “between X and Y” and “between about X and Y” should be interpreted to include X and Y. As used herein, phrases such as “between about X and Y” mean “between about X and about Y.” As used herein, phrases such as “from about X to Y” mean “from about X to about Y.”
Unless otherwise defined, all terms (including technical and scientific terms) used herein have the same meaning as commonly understood by one of ordinary skill in the art to which this invention belongs. It will be further understood that terms, such as those defined in commonly used dictionaries, should be interpreted as having a meaning that is consistent with their meaning in the context of the specification and relevant art and should not be interpreted in an idealized or overly formal sense unless expressly so defined herein. Well-known functions or constructions may not be described in detail for brevity and/or clarity.
It will be understood that when an element is referred to as being “on”, “attached” to, “connected” to, “coupled” with, “contacting”, etc., another element, it can be directly on, attached to, connected to, coupled with or contacting the other element or intervening elements may also be present. In contrast, when an element is referred to as being, for example, “directly on”, “directly attached” to, “directly connected” to, “directly coupled” with or “directly contacting” another element, there are no intervening elements present. It will also be appreciated by those of skill in the art that references to a structure or feature that is disposed “adjacent” another feature may have portions that overlap or underlie the adjacent feature.
It will be understood that, although the terms first, second, etc. may be used herein to describe various elements, components, regions, layers and/or sections, these elements, components, regions, layers and/or sections should not be limited by these terms. These terms are only used to distinguish one element, component, region, layer or section from another element, component, region, layer or section. Thus, a first element, component, region, layer or section discussed below could be termed a second element, component, region, layer or section without departing from the teachings of the invention. The sequence of operations (or steps) is not limited to the order presented in the claims or figures unless specifically indicated otherwise.
As will be appreciated by one of skill in the art, embodiments of the present invention may be embodied as a method, system, data processing system, or computer program product. Accordingly, the present invention may take the form of an embodiment combining software and hardware aspects. Furthermore, the present invention may take the form of a computer program product on a non-transitory computer usable storage medium having computer usable program code embodied in the medium. Any suitable computer readable medium may be utilized including hard disks, CD ROMs, optical storage devices, or other electronic storage devices.
Computer program code for carrying out operations of the present invention may be written in an object oriented programming language such as Matlab, Mathematica, Java, Smalltalk, C or C++. However, the computer program code for carrying out operations of the present invention may also be written in conventional procedural programming languages, such as the “C” programming language or in a visually oriented programming environment, such as Visual Basic.
Certain of the program code may execute entirely on one or more of a user's computer, partly on the user's computer, as a standalone software package, partly on the user's computer and partly on a remote computer or entirely on the remote computer. In the latter scenario, the remote computer may be connected to the user's computer through a local area network (LAN) or a wide area network (WAN), or the connection may be made to an external computer (for example, through the Internet using an Internet Service Provider).
The invention is described in part below with reference to flowchart illustrations and/or block diagrams of methods, devices, systems, computer program products and data and/or system architecture structures according to embodiments of the invention. It will be understood that each block of the illustrations, and/or combinations of blocks, can be implemented by computer program instructions. These computer program instructions may be provided to a processor of a general-purpose computer, special purpose computer, or other programmable data processing apparatus to produce a machine, such that the instructions, which execute via the processor of the computer or other programmable data processing apparatus, create means for implementing the functions/acts specified in the block or blocks.
These computer program instructions may also be stored in a computer readable memory or storage that can direct a computer or other programmable data processing apparatus to function in a particular manner, such that the instructions stored in the computer-readable memory or storage produce an article of manufacture including instruction means which implement the function/act specified in the block or blocks.
The computer program instructions may also be loaded onto a computer or other programmable data processing apparatus to cause a series of operational steps to be performed on the computer or other programmable apparatus to produce a computer implemented process such that the instructions which execute on the computer or other programmable apparatus provide steps for implementing the functions/acts specified in the block or blocks.
The IDSs produced are DICOM or AVI video loops of variable duration, depending upon the Clinical Application of the imaging technology. The invention is applicable to IDSs generated from ICG-NIR-FA clinical and non-clinical research applications of the imaging technology where arteriography and/or perfusion assessment is important.
The modeling of a generic FPA, and the modifications for its application to a specific CAA, is as follows. The definition of FPA using an average intensity over time curve is detailed in
Let B1=the average baseline intensity before the arterial phase, let P=the peak intensity, and let B2=the average baseline intensity after the venous phase. The Arterial phase starting time is defined as when the average intensity first increases to
B1+(P−B1)×k1 Equation 1
And the Arterial phase ending time is defined as when the average intensity first increases to
B1+(P−B1)×k2 Equation 2
Where k1 is a percentage defining the beginning of arterial phase (e.g. 5%), and k2 is a percentage defining the ending of the arterial phase (e.g. 95%).
The Venous phase starting time is defined as when the average intensity first decreases to
B2+(P−B2)×k3 Equation 3
And the Venous phase ending time is defined as when the average intensity first decreases to
B2+(P−B2)×k4 Equation 4
Where k3 is a percentage defining the beginning of arterial phase (e.g. 95%), and k4 is a percentage defining the ending of the arterial phase (e.g. 5%).
The Micro-vascular phase is defined as when the average intensity ranges between
B1+(P−B1)×k2 Equation 5
and
B2+(P−B2)×k3 Equation 6
nearby the peak.
The percentages will be somewhat different across different CAAs. The collection and analysis of clinical data is used to validate these percentages and to increase the specificity of these percentage values for each CAA utilization of the FPA ‘filter.’
Importantly, the IDS synchronization step occurs before the CAW/CAWT step, to avoid comparing data that are inadequate for analysis.
The Results of the analysis are reported in a format that is most applicable to the specific CAA, to assist the surgeon with new, real-time information in the operating room with which to make better decisions and decrease the incidence of complications.
In
Fluorescence angiography relies on low-energy, NIR laser excitation of ICG in blood vessels and perfused tissues, with capture of the intensity of fluorescence based upon the ICG infrared absorption and emission spectra. Importantly and in addition, imaging and its interpretation are influenced by a number of physiologic and/or pathophysiologic circumstances. The imaging data are captured as standard AVI and/or DICOM video loops at 30 fps, which can be directed imported into the core analytical platform. These standard image formats make the analytical platform widely applicable from a technical perspective. The frame rate was accounted for in the development of the CAPA core platform, as it limits the fidelity of the image analysis. An example of this is shown in
The known behavior of ICG dye in the blood has established that on the first pass through the heart, the fluorescence intensity is proportional to the concentration of ICG, which in turn is directly related to the injected dose. This allows for tailoring of the ICG dosage/injection for specific Clinical Application Areas and procedures within those areas. Importantly, this behavior also creates the possibility of fluorescence saturation, where the quantification of the intensity exceeds the 0-255 scale. This creates a problem of being unable to quantify how much greater than 255 the actual fluorescence intensity actually is; this is particularly a problem in other ICG-NIR-FA analysis approaches. As demonstrated, the CAPA analysis accounts for saturation correction when it does occur.
The known behavior of ICG dye as a bolus injection, with or without a saline flush, allows for specific detailing of how the ICG injection should be administered in order to optimize image quality. This understanding has specific importance in those CAAs where the angiography analysis is of particular relevance. The ICG bolus stays relatively undispersed as it passes through the central cardiac circulation, and ultimately out to the peripheral tissue microvasculature. Even at this anatomic location extremely distant physiologically from the heart, the FPA and its phase components can be readily identified in the ICG-NIR-FA IDS sequences. This documented discovery creates the opportunity to establish the CAPA core platform as an independent claim applicable across all ICG-NIR-FA applications involving angiography and perfusion. Now and in the future, supplemental analytical components that are specific to the existing and new CAAs can and will be developed as dependent claims.
The known behavior of ICG dye in blood and in circulation is fundamental to this imaging technology and analysis. ICG binds to the circulating proteins in serum, and to endothelial proteins attached to the inner surface of arterial and venous blood vessels. The half-life of ICG in humans is about 3 minutes, and the dye is metabolized by the liver and excreted in the kidney. Because the surface area on the venous side of the circulation is so much greater than the arterial side, there is more endothelial binding on the venous side, creating residual fluorescence, which typically is ‘washed out’ in 4-5 minutes after an injection. As demonstrated, our discovery and analysis of FPA, however, led to the understanding of how to deal with residual, background fluorescence in a physiologically-accurate manner that meets the time frame for this imaging technology to be adopted and used clinically by surgeons during complex operative procedures.
As with any imaging technology, image data acquisition is key to sustained, successful analysis across multiple providers in multiple settings. The standardization of these image acquisition parameters for each Clinical Application Area is critical for the analysis claim of the invention to be used appropriately and for the results to be used accurately in the clinical setting. As related to the invention, it is critically important that the image acquisition process for each CAA enables the complete capture of the FPA information, which is, as demonstrated, a key component for the CAPA platform analysis of angiography and perfusion in that CAA, and that surgical procedure.
We have defined the term Image Data Sequence (IDS) as the captured video loop with all the imbedded metadata. This IDS may be of variable duration, depending upon the application. As shown in
We have defined the term Image Data Acquisition Protocol (IDAP) as the specific, step-by-step process of coordinated capture of the IDS. This includes: 1) machine setup and positioning of the field of view, specific to the application and procedure; 2) the dosage, administration route and timing of administration of the ICG fluorescent dye coupled with management of the data capture software on the ICG Fluorescence machine; and 3) any specific technical, clinical or hemodynamic management processes necessary for optimization of the IDAP.
In addition, there are specific subset applications of the IDAP, depending upon the relative predominance of the arterial, microvascular and venous phases in that particular CAA and surgical procedure application. In these cases, the IDAP needs to be designed and executed so as to assure the time frame of data capture encompasses the necessary FPA spectrum. For example, in a CAA that is dependent upon the arterial phase, starting the video capture without a stable baseline makes a comparative analysis unfeasible. Similarly, truncating the video capture, or moving the machine, or shining the surgeon's headlight into the field, before the necessary venous phase information is captured creates an analytical problem. The specific IDAP must reflect a very real understanding of the FPA, its principles, and the CAPA platform.
In certain CAAs, specific IDAPs are developed for imaging purposes specific to either angiography or perfusion. For example, in the cardiac application, at the end of the revascularization procedure, with the heart in the anatomic position in the mediastinum, the ‘aortic root shot” is obtained, to illustrate flow and subjective rate of flow down the graft conduits, and to assess the anastomoses constructed to the ascending aorta, and to identify subtle technical issues (air bubble, low flow rate vs. other grafts) (
As is demonstrated in
Also in certain CAAs, intraoperative techniques have been developed to specifically facilitate IDS capture in a framework that enables subsequent analysis. For example, in the cardiac application, we have determined that the most reliable approach to consistent angiography and perfusion analysis is the following Coronary Bypass Graft Image Protocol (CBGIP) (
As shown in
We define the image area to which the FPA ‘filter’ intensity vs. time curve is applied as the Clinical Application Window (CAW), and/or to a sub-set of this window, termed the Clinical Application Window Target (CAWT).
This CAW is the area of clinical interest for imaging, and will be variable from application to application, but as shown in
The CAWT can be individual image pixels in a CAW, a certain selection and/or identified grouping of pixels, or an anatomic subset of the CAW as defined by the clinical application. The target can be manually selected, or automatically computer generated. The physiology of arterial flow and perfusion predicts that different CAWTs will, at any point in time, have different intensity vs. time curve characteristics.
Because the opportunity inherent in FPA and the CAPA is a dynamic analysis that reflects physiology, an important observational finding present in all CAAs studied thus far and critical for the analytical platform is that the predominant blood supply source engages the tissue being imaged be identified. This allows identification of a proximal (nearest to the blood supply origin) and a distal end (farthest away from the proximal end). The perfusion analysis must account for the entirety of the arterial and micro vascular phases in real time rather than just a single static frame from the image sequence. As mentioned, if the CAWT is defined as a certain selection and/or identified grouping of pixels in a CAW, during a single ICG injection that selection/grouping of pixels image arterial, micro-vascular and venous phases of full phase angiography. For that pixel CAWT and for the CAW as a whole, the image characteristics are very different from phase to phase. Since adjacent CAWT will have different characteristics, these differences in intensity and time can be used to derive comparative and contrasting data throughout the CAW.
Due to the limitations of 8-bit cameras, the intensity of fluorescence measurement in any IDS is limited to 255. At times, based on physiological or pathophysiologic circumstances, the same dose and concentration of ICG dye could in theory create saturation (intensity >255) in the IDS for part of the sequence. This saturation effect has been observed, especially with multiple injections, and this might jeopardize the accuracy of the perfusion comparison. To address this, we created an algorithm to estimate “real” intensity of the saturated pixels from the image histogram and approximate their distribution above intensity 255 by estimating the distribution of the pixels with intensity smaller than 255. Their geographical locations can be also estimated using non-saturated frames previous to the saturated frame.
In
The same imaged segment of large bowel is analyzed to emphasize this point. The bowel segment takes 12 seconds to perfuse the left-sided CAWT reference point (red box) to the CAWT point on the far right. The blue curve is the intensity vs time curve for the left-sided CAWT, and the red curve is the right-sided CATW. In the top panel, if a static reference point is chosen (black line at 46 sec), then the red CATW is higher than the blue CATW, reflected by the normalized percentage of 156% for this point. However, on the bottom panel, if the reference point is chosen at the 32 second point, a completely different normalize result occurs, despite the fact that the same blue CATW reference was used in both analyses. The visual appearance of the dynamic image sequence is dependent upon these physiologic arteriography and perfusion characteristics, depending on which part of the tissue the fluorescent wave front will reach first.
Only by synchronizing these CAWT curves by some parameter (time, distance) can the perfusion of different part of tissue can be quantified and validly compared in a dynamic manner.
Also as shown here in
Therefore, for a valid perfusion comparison, the corresponding phases have to be accurately aligned by a common parameter, whether the comparison is between different IDSs with the same CAW, or between different CAWCs within the CAW, derived from a single IDS (
An illustration of a method for synchronization is shown in
The effect of curve synchronization impacts on both analysis and display components of CAPA. Using average intensity vs. time curves, a correlation coefficient is calculated at each alignment time position and the largest correlation coefficient yields the optimal synchronization result. The extra segments in the beginning and/or end of the IDSs will be truncated. Therefore a fundamental principle of this present invention(s) is that the intensity vs time curve is the basis for synchronization of the phases of FPA.
This venous residual creates the need to account for residual fluorescence in any type of comparative analysis. In this core analytical platform, we define the baseline as described in
During multiple ICG-NIR-FA dye injections, the residue of dye accumulates and images acquired later tend to be brighter than the previous ones, mostly due to binding in the venules.
To investigate how residue of fluorescent dye from the previous injection affects intensity of the current IDS, we performed multiple sequential, paired IDSs without any change to the tissue or position of the camera. Since these two IDSs are recorded under same physiologic and CAW conditions, by studying their average intensity vs. time curves the optimal baseline management strategy was developed.
In
Importantly, from
Where BD is the baseline difference between pre and post IDSs with x, y as pixel coordinates and t as time; C(x, y) is the constant background difference between pre and post images estimated from the first few seconds of the IDSs; AICpost(t) is the average intensity curve of the post image acquisition and
is used to adjust the baseline difference across time. From
Examples of two important novel paradigms are documented herein. These are 1) the ability to recognize and document arterial-phase competitive flow between native and grafted sources of blood flow under physiologic conditions, and 2) the ability to recognize microvascular-phase collateral flow in adjacent and/or related areas of perfused tissues.
In
Competitive flow is currently most appropriately understood in the context of the arterial phase of FPA, although extension into the microvascular phase is being examined.
In,
Collateral flow is currently most appropriately understood in the context of the microvascular phase of FPA. Again the cardiac application is used as an example, in part because the heart is typically able to develop collaterals with non-acute, regional occlusions of the blood supply to a territory of the heart.
The CAPA perfusion quantification is a relative measurement based on a comparison, as illustrated in
k=mean(Imax)+m×std(Imax) Equation 8
Where ′max is the still frame that has the maximum average intensity in one IDS; mean is the average function; std is the standard deviation function; m is a constant parameter between 0-1 to adjust this Equation 8. The threshold k is used in one or several IDSs depending on the application and only pixels with intensity above the value are used in the perfusion calculation.
The arterial phase of IDS records perfusion as a process of blood being delivered by arteries to the tissue. Correspondingly, this process starts from the beginning (baseline part) to the peak (maximum) of the average intensity vs. time curve. Visually, this process includes arterial and part of micro vascular phases in the IDS. We are assuming not only the Verfusion strength” (corresponds to the average intensity above the threshold) but also the Verfusion area” (corresponds to the number of the pixels with intensity above the threshold) should be included in estimation of the perfusion level. Equation 9 is applied in all the still frames of the IDSs till the maximum of the average intensity curve is reached.
Al(t)=Num(I(x,y,t)>k)x mean(I(x,y,t)>k) Equation 9
Where Al is a number representing combination of perfusion strength and area at time t. 1(x, y, t) is a still frame at one time location of an IDS; Num is the function to calculate the number of pixels; mean is the average function.
Then we estimate the accumulation effect of the Al (t) from the beginning (baseline part) to the peak (maximum) of the average intensity vs. time curve as
Where T is any time at the peak (maximum); Al (0) is the residue from baseline. In the cardiac application we calculate this area-intensity value in sequential IDSs of the same CAW tissue area. In other CAAs identified thus far, we calculate this area-intensity value relatively across two or more CAWTs identified in one CAW identified in one IDS.
Notice that this is a relative value in both cases, and it does not reflect the estimation of perfusion directly. In the cardiac application, to estimate the perfusion change, we normalized the post area-intensity value by the pre one by
In the other CAAs identified thus far, to estimate the perfusion change, we normalize the current CAWT by the reference CAWT
The opportunity inherent in FPA and CAPA extends to image and image analysis display. The NIR part of the spectrum is outside the visible color spectrum, and therefore is inherently a black and white, 255-level grey scale image. This is actually quite sufficient for imaging the arterial phase of full phase imaging, but is not optimal or optimized for microvascular or venous phase imaging. We have developed different color schemes to optimize the display for combined (arteriography and perfusion) display using a modified RGB format, and for the microvascular (perfusion) image display and analysis. This in turn means that in many CAAs combination of displays of the same NIR image data is optimal for understanding the context and content of the image(s) and analyses for decision-making.
As illustrated in
We also designed an Overview Display as a unique way to visualize the IDS+FPA data. In
However, as previously articulated, to visually capture the inference of the FPA and CAPA construct requires that two points can be accurately compared. As depicted in
In
An additional opportunity inherent in the FPA and CAPA invention is to analyze angiography and perfusion as a dynamic process, rather than assuming that a selected static image accurately represents these physiologic processes. In some CAAs, multiple CAWS (for example, bypass grafts to the anterior, lateral and inferior territories of the heart) can be captured and analyzed individually; following this, the CAPA analysis metadata can be combined into 2-D and 3-D reconstructions to more accurately display the physiologic effects of perfusion increases or decreases, reperfusion, and/or devascularization.
The importance of this component of the present invention is in the ability to modify the CAPA core analysis display capabilities to specifically represent the critical information display that is necessary to optimize real-time decision-making by the surgeons in the operating room. The display results must be entirely accurate, intuitively presented, and simple enough to be grasped and understood in a visual display format from across the operating room.
As an example of this display capability, we can use the cardiac application of the 3-D model for revascularization-induced change in myocardial perfusion (
As an image analysis platform, it is necessary to be able to assess the quality of the IDSs for subsequent analysis. This is part of the analytical platform, and consists of the IDS Image Quality Test (
Baseline test
Timing test
Brightness test
IDS overall quality test
The foregoing is illustrative of the present invention and is not to be construed as limiting thereof. Although a few exemplary embodiments of the present invention have been described, those skilled in the art will readily appreciate that many modifications are possible in the exemplary embodiments without materially departing from the novel teachings and advantages of this invention. Accordingly, all such modifications are intended to be included within the scope of the invention as defined in the claims. In the claims, means-plus-function clauses, where used, are intended to cover the structures described herein as performing the recited function and not only structural equivalents but also equivalent structures. Therefore, it is to be understood that the foregoing is illustrative of the present invention and is not to be construed as limited to the specific embodiments disclosed, and that modifications to the disclosed embodiments, as well as other embodiments, are intended to be included within the scope of the appended claims. The invention is defined by the following claims, with equivalents of the claims to be included therein.
This application is a continuation of U.S. patent application Ser. No. 16/356,766, filed Mar. 18, 2019, which is a continuation of U.S. patent application Ser. No. 13/922,996, filed Jun. 20, 2013, now U.S. Pat. No. 10,278,585, which claims the benefit of U.S. Provisional Application No. 61/662,885, filed Jun. 21, 2012, the entire contents of each of which are incorporated herein by reference.
Number | Date | Country | |
---|---|---|---|
61662885 | Jun 2012 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 16356766 | Mar 2019 | US |
Child | 17656872 | US | |
Parent | 13922996 | Jun 2013 | US |
Child | 16356766 | US |