This invention relates to hearing aids and, in particular, to a hearing aid powered by a substantially non-replaceable battery.
As used herein, a “primary” battery is one that is not intended for charging even though, in fact, one can safely recharge the battery one or a few times. A “secondary” battery is one that is intended for recharging a plurality of times. In general, primary batteries have a greater capacity (store more energy) than rechargeable batteries. Secondary batteries have a different internal structure from primary batteries, even when the chemistry involved is nominally the same.
Hearing aids powered by a battery have been known for almost a century; see U.S. Pat. No. 1,219,411 (Williams), for example. Modern technology has increased battery life greatly, yet it is annoying to have to replace batteries. Rechargeable batteries are a partial solution but require removal of the hearing aid and placement in a charger. Unless a user has two sets of hearing aids, the charging can be inconvenient.
Hearing aids having rechargeable batteries have been known in the art for a long time; e.g., see U.S. Pat. No. 3,297,933 (McCarthy). The trade-off between rechargeable batteries and non-rechargeable batteries is the inconvenience of having to replace the battery. There is also a trade-off in capacity. A non-rechargeable battery lasts much longer than a rechargeable battery having the same outside dimensions as the non-rechargeable battery. The similarity in shape and dimensions can and does cause confusion among users. The industry has adopted color codes on packaging to distinguish batteries but the problem persists.
Substituting a secondary battery for a primary battery is not as dangerous as substituting a primary battery for a secondary battery. Secondary batteries have specific charging requirements. If the requirements are not met, damage to the battery is likely and catastrophic damage is possible. Carefully charging a primary battery in a laboratory is relatively safe. Accidentally charging a primary battery substituted for a secondary battery is a very different situation and much more serious. Catastrophic failure is likely.
The need to replace batteries means that one must open a hearing aid. Moisture, wax, dirt, oils and so on, can work their way into a hearing aid, causing problems. Hearing aids can be made relatively impervious to ambient conditions. A hearing aid that is relatively impervious is also likely to be relatively difficult to open for a user.
The inconvenience of having to remove the battery from a hearing aid initially applied both to rechargeable batteries and non-rechargeable batteries. The sole advantage of rechargeable batteries was not having to be replaced. Then, chargers were developed that made electrical contact with the hearing aid, obviating the need to remove the rechargeable battery; e.g. see U.S. Pat. No. 3,493,695 (Stork). This simplified matters for those lacking the dexterity to remove and insert a battery. Having exposed electrical contacts is undesirable and inductive chargers solved this problem; e.g. see U.S. Pat. No. 4,379,988 (Mattatall).
Inductive chargers have their own set of difficulties, including adequate coupling between the primary inductor in the charger and the secondary inductor in the hearing aid; e.g. see U.S. Pat. No. 6,658,124 (Meadows). Even with adequate coupling, rechargeable batteries are not a panacea. Many rechargeable batteries, e.g. nickel cadmium, lithium ion, and others, have “memory.” Memory in a battery relates to the amount of stored energy that is available after several discharge-charge cycles. If, for example, half the energy is used and a battery is recharged, then, eventually, only half the energy is available. Also, some rechargeable batteries do not like being overcharged, such as lithium ion batteries. These batteries overheat and rupture, sometimes violently, or catch fire. Currently, nickel-metal-hydride (NiMH) batteries are preferred for hearing aids because they have little memory and are more tolerant of overcharging.
The problems of memory and overcharging are particularly acute for hearing aids because a hearing aid may partially discharge a battery during the day and then be placed on a charger overnight. If more than one hearing aid is used, the batteries may be in different states of charge but are charged simultaneously.
It is known in the art to provide a reduced voltage for operating a hearing aid; e.g. see the Williams patent cited above and U.S. Pat. No. 7,315,626 (Pedersen). The Williams patent also discloses sealing the terminals of the battery in a hearing aid. It is known in the art to provide a regulated supply voltage for the audio processing stages of a hearing aid; e.g. see U.S. Pat. Nos. 5,332,928 (Johnson), 6,741,715 (Andersen), 6,831,988 (Vonlanthen), and published application US2005/0234572 (Heubi et al.). It is also known in the art to store in memory a variable in a software program for control of battery recharging; e.g. see U.S. Pat. No. 6,565,603 (Leysieffer, et al.).
In view of the foregoing, it is therefore an object of the invention to provide a hearing aid that distinguishes a primary battery from a secondary battery and charges only a secondary battery.
Another object of the invention is to provide a hearing aid that recognizes secondary batteries and selects the charging sequence (algorithm) appropriate for the battery.
A further object of the invention is to provide a hearing aid that uses a reduced operating voltage for a reference voltage.
The foregoing objects are achieved by this invention in which a hearing aid includes a secondary battery and means for identifying the kind of battery for charging. After the battery is identified, an appropriate algorithm for charging the battery is selected and used for charging the battery. Primary batteries are not charged. The hearing aid also includes a voltage reduction circuit for supplying a constant voltage to circuitry within the hearing aid. The battery preferably has an energy density greater than 500 Wh/l.
A more complete understanding of the invention can be obtained by considering the following detailed description in conjunction with the accompanying drawings, in which:
In
Battery 16 (
In both of the following examples, a fully discharged battery is presumed at the start of the charging cycle. Charging follows the manufacturer's recommended procedure. Discharge is the maximum rated current.
There are several different batteries that can fit a socket for battery 16.
Although the charging patterns for a silver-zinc battery and NiMH battery are obviously quite different, the discharge patterns are significantly different also. The silver-zinc battery charges for approximately eight hours and can be discharged in approximately eight and one half hours. The NiMH battery charges for a total of about eight hours and discharges in approximately twenty-seven and one half hours. More significant is the discharge rate. The silver-zinc battery charges at approximately ten ma. and discharges at approximately nine ma., after a brief period discharging at eleven ma. The NiMH battery discharges at only two ma. Thus, significantly more current can be drawn, if need be, from a silver-zinc battery and the battery can discharged at substantially the same current as used for charging.
In accordance with one aspect of the invention, the type of battery inserted into hearing aid 10 (
In
As illustrated in
By applying selected currents and loads to battery 16, microprocessor 44 can identify what kind of battery it is. At a minimum, AgZn is distinguished from NiMH. The proper charging algorithm is then selected and battery 16 is charged when necessary. For example, the following procedure is used.
For detecting primary batteries, the decision tree becomes more complicated but is based, as above, on the characteristics of the batteries. For example, a discharged NiMH battery will increase in voltage only slightly after being charged for a minute or two but a discharged AgZn primary battery will increase in voltage very quickly. On the other hand, the voltage will not exceed 1.55 volts for a AgZn primary battery. Thus, one can readily distinguish a primary AgZn battery from a NiMH battery. When a primary battery is detected, the charging procedure terminates.
In one embodiment of the invention, power for charging is derived from signals received by antenna 52, as disclose in co-pending application Ser. No. ______, filed Feb. 25, 2008, and entitled RF Power Supply for Hearing Aids, the entire contents of which are hereby incorporated by reference. In alternative embodiments of the invention, inductive charging or direct electrical connection can be used.
The invention thus enables one to manufacture hearing aids that are sealed but which can be properly charged by identifying the battery within the sealed case. If fully discharged each time, a silver-zinc battery can be recharged approximately two hundred times. If less than fully discharged, a silver zinc battery can be recharged more than four hundred times and battery life starts approaching the useful life of a hearing aid. If, for some reason, there is a failure of a battery, the seal can be broken and the battery replaced. If the wrong type of battery is used, the type is detected and the battery can be charged properly despite the error. If a primary battery is installed, no charging takes place and a suitable indicator informs the user of the error.
Having thus described the invention, it will be apparent to those of skill in the art that various modifications can be made within the scope of the invention. For example, the invention is applicable to hearing aids other than the types illustrated in