This application is generally related to a radiation-shielding sliding assembly.
Radiation therapy facilities, especially those involving high energy X radiation or neutron radiation, require particularly thick walls, doors, and barriers. Particle accelerators, such as linear particle accelerators, use electromagnetic fields to propel charged particles, such as electrons, protons, or ions, at high speeds along defined beams. Due to radiation from particle accelerators, particle facilities must be designed and constructed to provide adequate shielding.
Known radiation therapy facilities are generally constructed as a room housing the source of radiation, with concrete walls, ceilings, and floors that can reach thicknesses of up to 15 feet. In addition, a maze entry is usually used to provide a wing wall to capture scatter radiation. The entrance to a maze entry or direct entry radiation therapy room can include at least one shielded door to further prevent radiation leakage outside of the room. The shielded door for a radiation therapy room can be constructed as a hinged door having a very thick core, for example 20 inches thick, to provide sufficient shielding. Known shielded doors are also extremely heavy, typically 10,000-20,000 lbs for radiation therapy rooms, and cannot be opened and closed quickly. The time that it takes to open and close a hinged shielded door is especially important in radiation therapy rooms where an operator may need to enter and exit the room repeatedly to make adjustments. For example, in medical applications, several rounds of low energy radiation may be used for diagnostic purposes and patient positioning before treating the patient's tumor with the high energy radiation. After each round of low energy radiation, the operator must either progress down a very long maze corridor leading to the treatment room or alternatively wait for the shielded door to fully open before entering the treatment room to make adjustments to the patient, and then wait for the shielded door to fully close again before starting the next round of low energy radiation testing or high energy radiation treatment. This process can be very time consuming and tiring to the patient.
Bi-parting sliding doors typically permit shorter opening and closing times compared to hinged doors. Because existing bi-parting sliding doors have a relatively linear leading edge at the seam between both doors, they lack the necessary seal required to prevent radiation leakage. One known method to reduce radiation leakage is to equip one of the bi-parting doors with an astragal at its leading edge to cover the seam between the doors.
The increased speed of heavy radiation shielded members introduces additional safety concerns especially when objects obstruct the closing path of the sliding members.
A need exists for a sliding assembly for radiation therapy rooms that provides a sufficient seal to eliminate radiation leakage and improved safety when closing.
A radiation-shielding movable barrier for a radiation therapy facility is disclosed. The movable barrier includes a first member movable between an open position, and a closed position. The first member comprises a front face, a rear face, a trailing face, and a leading face including a sine-wave shaped leading edge. The first member has a core constructed of high-density or high-Z material having radiation-shielding characteristics. Radiation is defined as at least one of high energy X radiation, neutron radiation, photon radiation, proton radiation, X ray radiation, gamma radiation, or high energy radiation. A second member has a sine-wave shaped edge that is complementary to the sine-wave shaped leading edge of the first member. The second member has a core constructed of high-density or high-Z material having radiation-shielding characteristics. When the first member is in the closed position, the sine-wave shaped leading edge of the first member and the sine-weave shaped edge of the second member interlock to form a tight seal between the first member and the second member to prevent leakage of radiation through the seal.
The foregoing summary, as well as the following detailed description of the preferred embodiments of the invention, will be better understood when read in conjunction with the appended drawings. For the purpose of illustrating the invention, there is shown in the drawings embodiments which are presently preferred. It should be understood, however, that the invention is not limited to the precise arrangement shown.
Certain terminology is used in the following description for convenience only and is not limiting. The words “top,” “bottom,” “inner,” and “outer” designate directions in the drawings to which reference is made. The terminology includes the words specifically noted above, derivatives thereof, and words of similar import.
The outer surface of the doors 12, 13 are preferably constructed of carbon steel plate face panels and a minimum ½ inch thick edge banding along the top, bottom, and trailing edge of the door. The outer surface of the doors 12, 13 can be coated and finished with any suitable material including plastic, wood or metal laminates.
The leading edge of each of the bi-parting doors 12, 13 preferably have complementary tortuous paths to prevent radiation leakage when the doors 12, 13 are closed. The tortuous paths extend the length of the doors 12, 13 in a direction perpendicular to the seam 11 formed between the two doors 12, 13 when the doors 12, 13 are closed.
As shown in
In an alternate embodiment shown in
Highly efficient hinged shielded doors used in direct entry radiation therapy rooms take approximately 10-12 seconds to move from an open position to a closed position, and vice-versa. The bi-parting door assembly 10 of the present application can move from an open position to a closed position in approximately 5-6 seconds, which reduces the waiting time for a treatment technician to move in and out of the room.
A drive assembly 16 drives the bi-parting doors 12, 13 or single door 412 between an open and closed configuration. The drive assembly 16 can include any suitable driving mechanism. Preferably, the drive assembly 16 includes magnets to magnetically propel the doors 12, 13 along a track 15 preferable having a linear shape. Because the doors 12, 13 are magnetically propelled, there are fewer mechanical problems related to gears and drive systems. Due to the lack of moving parts in the drive assembly 16, the overall failure rate of the sliding door assembly 10 is reduced. Alternatively, a track support mechanism having guidance rollers can be used to opening and closing the doors 12, 13.
The width of the passageway 60 to the radiation therapy room 2 when the sliding door assembly 10 is open may vary depending on the type of room the sliding door assembly 10 is used in, but should at least be suitable for a person to walk through, for example approximately 36-46 inches wide. In research or medical particle facilities, the passageway 60 may be wider to accommodate equipment to be moved in and out of the room, such as wheel chairs, stretchers, and lab equipment. In addition, the sliding door or doors 12, 13 can be removable in order to create additional space to move equipment in and out of the room.
To prevent the sliding door assembly 10 from closing when a person or object is in the passageway 60, a sensor 18 may be arranged to detect whether an object is in the passageway 60. A sensor 18 may be placed in the floor, ceiling, or in the area adjacent to the sliding door assembly 10 to detect when a person or object is approaching the passageway 60. Preferably, a plurality of sensors are used to enhance accuracy. The sensor 18 may be, for example and without limitation, a pressure sensor arranged in the floor of the sliding door assembly 10, an ultrasonic presence detecting sensor, or an infra-red light sensor. The sensor 18 may be configured to relay signals to a control system 40 which includes a programmable touch screen interface and is electrically connected to the drive assembly 16 to control operation of the sliding door assembly 10. When the sensor 18 detects a person or object in the passageway 60, the control system 40 prevents the drive assembly 16 from moving the door or doors 12, 13.
A secondary sliding door assembly 34 comprised or one or more sliding panels 32 can be positioned exterior to the sliding door assembly 10 as an additional safety precaution against the sliding door assembly 10 closing on a person or object in the passageway 60. The panel or panels 32 are preferably made of a thin, lightweight material, such as plastic or plexiglass. The panel or panels 32 can be operated to close before the sliding door assembly 10. The panel or panels 32 can be driven by either the same drive assembly 16 or a separate drive assembly as the sliding door assembly 10. The panels 32 are prevented from closing if the sensor 18 detects an object or person within the detection area.
The sliding door assembly 10 preferably operates on a 220 volt, three-phase, 30 amp power supply with low voltage wiring to the drive assembly 16, control system 40, sensor 18, and any other electronic components. In the event of a power failure, the magnetic propulsion drive assembly 16 would fail. The sliding door assembly 10 includes a manual operation mode wherein at least one of the doors 12, 13 and the panel 32 can manually open and close under their own power or by a battery back-up system.
While a sliding door assembly has been described herein, one of ordinary skill in the art would also recognize that the sliding door assembly could also be modified for use as a window. As shown in
While various methods, configurations, and features of the present invention have been described above and shown in the drawings, one of ordinary skill in the art will appreciate from this disclosure that any combination of the above features can be used without departing from the scope of the present invention. It is also recognized by those skilled in the art that changes may be made to the above described methods and embodiments without departing from the broad inventive concept thereof.
The following documents are incorporated herein by reference as if fully set forth: U.S. Provisional Application No. 61/787,702, filed Mar. 15, 2013; and U.S. patent application Ser. No. 14/215,797, filed Mar. 17, 2014.
Number | Date | Country | |
---|---|---|---|
61787702 | Mar 2013 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 14215797 | Mar 2014 | US |
Child | 15262838 | US |