The present invention relates to radio frequency (RF) detector coils. It has application, for example, in magnetic resonance (MR) imaging and monitoring systems.
Small resonant RF detectors have many applications for in-vivo internal magnetic resonance imaging (MRI). Although small coils generally have low Q-factors, this disadvantage is mitigated by the overall increase in signal-to-noise ratio obtained from close coupling to the source. Suitable coil arrangements include single- or multi-turn loops, parallel conductor transmission lines, opposed solenoids and meanders. More compact alternatives for intravascular imaging include the so-called “loopless catheter antenna”, which measures electric rather than magnetic fields. A similar range of coils has been used for the alternative application of catheter tracking.
In each case, the need for matching and tuning has limited widespread clinical application. This problem can be illustrated as follows. Generally, the coil (which has inductance L and resistance R) must be matched to a load RL at the angular frequency ω0 at which the coil is to resonate.
Despite advances in modelling methods that allow for 3D coils, skin effects and material losses, it is unfortunately difficult to predict either the resistance R or the inductance L accurately. The coil resistance is inherently frequency-dependent, joints provide further resistive losses, self-capacitance strongly affects resonant frequency, and hand winding introduces variability. Consequently, both CM and CT must generally be determined experimentally, using values that are successively updated to improve the degree of matching and the resonant frequency. The restricted set of readily available capacitance values generally forces the use of multiple components for both CM and CT. As a result, the final assembly is often bulky, and may have been soldered and re-soldered many times. For large coils, the end product may be acceptable, but the general approach cannot achieve the low cost, small form factor and reproducibility needed for mass deployment of catheter-based probes, especially disposable ones. One solution is to locate the matching and tuning components remotely, using a λ/2 length of cable. This approach allows a suitable form factor. Matching and tuning can also be carried out using automated varactor-based systems. However, both solutions are generally too complex for low-cost, mass-produced coils.
Micro-fabrication can improve the situation, since it allows repeatable values of R and L to be obtained. Electroplated spiral coils have been formed on rigid substrates such as GaAs, Si and glass. Micro-fabricated Helmholtz coils and gradient coils have been constructed, solenoids have been fabricated on capillaries, planar coils have been integrated with micro-fluidics, and pre-amplifiers incorporated. More recently, attention has turned to flexible plastics such as polyimide and polyether-ether-ketone [Coutrot A.-L., Dufour-Gergam E., Quemper J.-M., Martincic E., Gilles J.-P., Grandchamp J. P., Matlosz M., Sanchez A., Darasse L., Ginefri J.-C. “Copper micromolding process for NMR microinductors realisation” Sensors and Actuators A99, 49-54 (2002)] and polytetrafluoroethylene [Eroglu S., Gimi B., Roman B., Friedman G., Magin R. L. “NMR spiral surface microcoils: design, fabrication and imaging” Conc. Mag. Res. B 17, 1-10 (2003)], which are much more suitable for in-vivo use, and flexible 3D coils have been constructed [Woytasik M., Grandchamp J.-P., Dufour-Gergam E., Martincic E., Gilles J.-P., Megherbi S., Lavally V., Mathet V. “Fabrication of planar and three-dimensional microcoils on flexible substrates” Microsyst. Tech. 12, 973-978 (2006)]. However, matching and tuning have largely been ignored. Some attempts have been made to integrate capacitors, using coplanar conductors [Ellersiek D., Harms S., Casanova F., Blümich B., Mokwa W., Schnakenberg U. “Flexible RF microcoils with integrated capacitor for NMR applications” Proc. MME'05, Göteborg, Sweden, September 4-6, pp 256-259 (2005)], double-layer windings [Woytaskik M., Ginefri J.-C., Raynaud J.-S., Poirier-Wuinot M., Dufour-Gergam E., Grandchamp J.-P., Girard O., Robert P., Gilles J.-P., Martincic E., Darasse L. “Characterisation of flexible RF microcoils dedicated to local MRI” Microsyst. Tech. 13, 1575-1580 (2007)] or optically-variable MOS structures [Uelzen Th., Fandrey S., Müller J. “Mechanical and electrical properties of electroplated copper for MR-imaging coils” Microsyst. Tech. 12, 343-351 (2006)]. Previously [Ahmad M. M., Syms R. R. A., Young I. R., Mathew B., Casperz W., Taylor-Robinson S. D., Wadsworth C. A., Gedroyc W. M. W. “Catheter-based flexible microcoil RF detectors for internal magnetic resonance imaging” J. Micromech. Microeng., submitted], it has been demonstrated that high-resolution MRI using catheter mounted microfabricated coils with discrete capacitors, based on the multi-turn rectangular spiral inductor can be produced However, no convincing solution has yet been found to the problem of optimising component values without expensive iteration of planar processing.
The present invention provides a resonant radiofrequency (RF) detector comprising a substrate, an inductor coil formed on a front surface of the substrate, and two capacitors, each capacitor having a front plate formed on the front surface of the substrate and a rear plate formed on the rear surface of the substrate. The two front surface capacitor plates are each connected electrically to a different end of the coil, and the two rear capacitor plates are connected electrically to each other, so that the whole circuit represents a resonant electrical loop containing one inductor and two capacitors.
The resonant circuit may then provide the function of detecting RF signals. One capacitor CM may then provide the function of matching the electrical impedance seen across its plates at a target resonant frequency to a target value while the other capacitor CT may provide the function of tuning the resonant frequency of the circuit to a target value.
The coil may comprise one or more full turns, or it may comprise one or more half turns, or part turns, or it may be of any other suitable shape for detecting a RF signal. The front plate of one of the capacitors may be formed within the coil. The front plate of one of the capacitors may be formed outside the coil. The two rear capacitor plates may be formed from a common layer of conductive material. Connections between the two capacitors may also be formed in the same common layer of material. A similar approach may be use to add additional coils and capacitors in series.
The coil may have two sections and each of the front plates may be connected to an end of a respective one of the sections. In some cases the sections each have a respective winding sense, the winding senses being opposite to each other. For example the winding may be arranged in a figure-of-eight configuration. The sections may each have the same number of turns, or may be otherwise arranged so as to have the same, or substantially the same, inductance. Each of the coil sections may form a respective loop and each of the front plates may be inside a respective one of the loops.
The substrate may be thin, to allow capacitors of a given size to be formed using a small surface area. Using a thin substrate, the inductor and capacitors may be flexible. The whole assembly may therefore be flexible so that it can be wrapped around a catheter.
The present invention further provides a method of producing an RF detector assembly comprising:
The coil and the front capacitor plates may be formed simultaneously, or they may be formed separately in separate steps. The coil and the front capacitor plates may be formed as a common layer of conductive material. The two rear capacitor plates may be formed simultaneously, and may also be formed as a common layer of conductive material. The connection between the two rear capacitor plates may also be formed in this layer. In some embodiments this can allow the whole assembly to be made using just two steps of patterning a surface conductive layer on a substrate, one layer being provided on the front surface of the substrate and the other layer on the rear surface.
The present invention further provides a method of producing a resonant RF detector comprising:
In one embodiment the target application is a catheter-based probe for magnetic resonance imaging of the bile duct, but similar approaches would be appropriate for vascular imaging, or for other forms of internal magnetic resonance imaging such as oral, rectal or vaginal imaging that require a small flexible probe.
The overall aim of some embodiments of the invention is a resonant detector in the form of a flexible sheet that may be wrapped around a catheter and connected to receive electronics, for example via a subminiature co-ax cable.
In some embodiments a three-stage approach is used. In the first, a RF resonator is formed from a micro-fabricated coil and discrete capacitors. Conventional matching and tuning of this structure allows the values of CM and CT to be found. In the second, a fully integrated device is constructed. The same design of coil is used, together with micro-fabricated capacitors whose areas are estimated from the previous experimental values of CM and CT. Mechanical trimming of the capacitors following a systematic procedure then allows exact matching and tuning. In the third, a fully integrated device is constructed using an identical micro-fabrication process, but with the now-known capacitor areas. The result is a flexible monolith requiring only connection to a co-axial output, and the method is easily applicable to other coil arrangements.
Preferred embodiments of the present invention will now be described by way of example only with reference to the accompanying drawings.
a is a circuit diagram of a detector coil;
b is a diagram of an equivalent circuit to that of
c is a diagram of a further equivalent circuit to that of
a is a diagram of a known detector coil assembly;
b is a diagram of a detector coil assembly according to an embodiment of the invention;
a is a front view of a set of detector coil assemblies according to the invention;
b shows one of the detector coil assemblies of
c is an enlarged view of the part of the detector coil assembly of
Referring to
The admittance of the combination of RL and CM is Y=1/RL+jωCM. Hence, the corresponding impedance Z=1/Y can be found as:
Z=(1/jωCM){1/(1+1/jωCMRL)} (1)
If ωCMRL>>1, Equation 1 may be approximated as:
Z≈1/jωCM+1/(ω2CM2RL) (2)
This approximation assumes that RL>>1/ωCM, i.e. that the load is much greater than the modulus of the impedance of CM. Equation 2 implies that Z is equivalent to a capacitor CM in series with a frequency-dependent load RL′=1/(ω2CM2RL) as shown in
C
M=1/{ω0√(RRL)} (3)
This result implies that 1/ω0CM=√(RRL), i.e. that the modulus of the impedance of CM should equal the geometric mean of the two resistors that require matching. CM must clearly reduce as the frequency rises or as resistive losses increase. We may then rewrite the condition for the earlier approximation as RL>>R. Since RL will typically be 50Ω, and R a few Ohms, Equation 3 will be almost universally valid, independent of the value of L and the operating frequency.
The circuit of
C
T
=CC
M/(CM−C) (4)
Clearly, CT must be positive and finite to achieve a meaningful solution. This condition requires CM>C, or ω0L>√(RRL), so that modulus of the impedance of the inductor must exceed the geometric mean of the two resistors. This condition can normally be satisfied, provided the coil resistance R is low enough. The impedance matching problem may be illustrated graphically as shown in
The operating frequency f0 for 1H magnetic resonance (MR) scales linearly with flux at 42.57 MHz/T. In a 1.5 T system (for example), f0=63.5 MHz.
In a first stage of a method of producing detector coils according to an embodiment of the invention, hybrid integrated RF detectors are first produced. Referring to
Referring to
The devices of
Coils were fabricated on 25 μm thick polyimide film (Kapton® HN, DuPont, Circleville, Ohio). This material is mechanically and thermally stable, flexible, pinhole free, resistant to dielectric breakdown, and commercially available in a range of thicknesses [Data Sheet HK-15345: DuPont Kapton® HN polyimide film” DuPont High Performance Films, Circleville, Ohio, http://www.dupont.com]. To provide a rigid surface for processing, the film was first stretched over a 100 mm diameter silicon wafer and anchored using Kapton® tape. Seed layers of Ti (30 nm) and Cu (200 nm) metal were then deposited by RF sputtering. A layer of AZ 9260 thick positive photoresist (Microchemicals GmbH, Ulm, Germany) was then deposited by spin coating, and patterned using UV contact lithography to form a mould. 20 μm thick Cu conductor tracks were then formed by electroplating inside this mould, using Technic FB Bright Acid copper plating solution (Lektrachem Ltd., Nuneaton, UK). The mould was stripped, and exposed seed layer was removed by etching. A 2.5 μm thick layer of SU-8 2000 negative epoxy photoresist (Microchem Corp., Newton, Mass.) was then deposited and patterned to act as an interlayer. Finally, the air bridge was formed, by repeating the steps of seed layer deposition, mould formation, plating, mould removal and seed layer etching. The Kapton® sheet was detached from its carrier, and individual devices were separated using a scalpel. The devices were highly flexible, and could be distorted considerably without conductor detachment.
Electrical performance was measured using an Agilent E5061A network analyser. Matching to RL=50Ω and tuning were carried out with the coil loosely attached with heat shrink tubing to an 8 Fr dual lumen catheter, using discrete 0805 series non-magnetic capacitors (SRT Micro Céramique, France). The additional components were located just beyond the end of the catheter, and a length of 0.8 mm diameter non-magnetic 50Ω Bluetooth co-axial cable (Axon Cable, Dunfermline, UK) was connected across CM and passed down one of the catheter lumens. The heat shrink was then tightened to protect the coil, capacitors and cable joint. Matching and tuning were achieved by minimising the value of the scattering parameter S11 at 63.8 MHz using component values of CM=139 pF, CT=19.5 pF. Ignoring self-capacitance, these values imply a total capacitance of C=17.1 pF, a coil inductance of L=1/(ω02C)=0.36 μH and a resistance of R=1/(ω02CM2RL)=6.4Ω (and hence a quality factor of Q=ω0L/R=23). Confirmation of the Q-factor was provided by measurement of the frequency variation of sensitivity for completed resonators.
Referring to
The coil assembly of
Prototype devices were fabricated using conductor dimensions similar to those above, but a number of coil assemblies in a range of coil lengths were formed on a single substrate as shown in
Similar patterning and electroplating processes were used to form the conductors on each side. The rear side conductors were formed first, using a smaller Cu thickness (5 μm) and then protected with a layer of photoresist while the front side conductors were formed with the standard Cu thickness (20 μm). The overall thickness was therefore approximately 40 μm.
Matching and tuning of fully integrated devices was then carried out by mechanically trimming each of the matching and tuning capacitors CM and CT, using a systematic process.
Completed devices were connected electrically using subminiature co-ax cable, which was soldered across CM. The initial condition of each device was then assessed, by measuring the frequency variation of the scattering parameter S11 with the devices held flat.
The devices were then mounted on a short section of 8 Fr catheter, and the frequency variation of S11 was re-measured. This process would be expected to reduce L without changing R, and hence simply increase the resonant frequency. In each case, the matching degraded. Since the effective load is RL=1/(ω02CM2RL), and CM and RL are both fixed, the only possible conclusion is that CM is too large, in line with earlier estimates. The shortest (40 mm) device re-tuned immediately to 63.8 MHz frequency. However, since the longest (60 mm) device only re-tuned to 52 MHz, this device required adjustment.
Matching and tuning of the 60 mm device was carried out as shown in
The final areas of CM and CT were 52 mm2 and 9.3 mm2, respectively, corresponding to estimated capacitances of 127 pF and 23 pF (close to the target values). These results show that a method for identifying the initial state of an unknown device exists, and (provided the two capacitances are both too large), there is a convergent algorithm to match and tune by reduction of overlap areas. However, if necessary adjustment can still be carried out when capacitors are too small, simply by extending overlap areas with silver-loaded epoxy.
It can be a problem that mechanical trimming can short one of the capacitors, due to small slivers of metal tracking between the closely spaced conductors. This problem can be eliminated, by replacing mechanical trimming of the entire structure with laser trimming of just one, or both, of the conductors, so that an insulating dielectric layer remains in place throughout. In addition, excess solder flow during cable attachment can spoil the flexibility of the metal layers. This problem can be addressed using additional patterned surface layers to limit the solderable area.
Once the final area is known, a mask redesign may be carried out to fabricate further devices whose initial state is even closer to the desired final state.
1H magnetic resonance imaging experiments were carried out on phantoms, using resonators as described above with reference to Figures 6a to c to demonstrate the imaging capability of these integrated resonators.
Imaging was performed using a 1.5 T GE HD Signa Excite scanner. The system body coil was used for transmission and a 40 mm long catheter mounted micro-coil as shown in
High-resolution imaging was then demonstrated using a phantom consisting of an M4 nylon nut and cheese-head bolt (tooth pitch 0.7 mm), which was placed in solution in a small glass cuvette as shown in
In addition to providing a means of realising spiral coils with integrated tuning and matching capacitors, as described in the embodiments above, other embodiments of the invention make use of its ability to allow more complex coil arrangements, which can have advantages in certain applications.
Referring to
The resulting winding on the front face of the substrate has two loops each having two full turns, the winding being broken in the outer turn at one end of the first loop, to allow connection to two plates 103 and 104 and also broken at its centre where the resulting gap, between the two capacitor plates 105 and 106 which form ends of the coil on the front face, is bridged by the interconnecting bridge 108.
The arrangement of
ω0=1/(LC)1/2 (5)
Where L is the total inductance and C is the total capacitance, and L and C are given by:
L=L
1
+L
2 (6)
1/C=1/CM+1/CT1+1/CT2+1/CT3 (7)
It will be appreciated that the circuit can operate as a resonant detector for RF signals, and that matching and tuning may be carried out as described above for the embodiment of
It will be appreciated that if the inductors L1 and L2 are identical, a uniform time-varying external magnetic flux B1 acting perpendicular to the coil will induce equal and opposite emf in each half winding, which will therefore cancel to yield zero net emf and zero current. Consequently, the coil will have low sensitivity to a spatially uniform RF magnetic field, such as the field generated by the body coil of a MRI scanner during excitation. As a result, directly induced voltages and local modification of the excitation pattern may be minimised. However, the coil can still have sensitivity to the locally generated RF fields that arise during signal reception.
This feature provides an inherent passive de-coupling between the transmitter and receiver of the MRI system, which can avoid the need for other methods of de-coupling such as diode-switched de-tuning. Consequently, the arrangement can provide a de-coupled coil in thin-film form, that can be fabricated entirely by patterning of conductor layers without the need for additional semiconductor components.
Finally, it will also be appreciated that the use of paired capacitors to allow a figure-of-eight coil winding without the use of an airbridge can be extended to provide windings that are further subdivided into additional sections, providing a multi-section coil whose winding alternates in sense between adjacent sections or loops. If there are an even number of sections, then the induced emfs in the sections can be balanced by making all of the loops the same size and with the same number of turns. In other cases the sections can be of different sizes, and there may be different numbers of sections with the two winding senses, but with the correct selection of coil size and shape, the emfs can still be balanced.
Number | Date | Country | Kind |
---|---|---|---|
0910039.7 | Jun 2009 | GB | national |
1004721.5 | Mar 2010 | GB | national |
Filing Document | Filing Date | Country | Kind | 371c Date |
---|---|---|---|---|
PCT/GB2010/050982 | 6/11/2010 | WO | 00 | 2/27/2012 |