Erythrocyte sedimentation rate (ESR), also called a sedimentation rate or Biemacki Reaction, is the rate at which red blood cells sediment, typically measured over a period of one (1) hour. It is a common hematology test and is a non-specific measure of inflammation. To perform the test using a traditional technique, anti-coagulated blood is placed in an upright tube, known as a Westergren-Katz tube, and the rate at which the red blood cells sediment is measured and reported in mm/hour. Specifically, the Westergren method requires collecting 2 ml of venous blood into a tube containing 0.5 ml of sodium citrate. The sample should be stored no longer than 2 hours at room temperature or 6 hours at 4° C. The blood is drawn into the Westergren-Katz tube to the 200 mm mark. The tube is placed in a rack in a strictly vertical position for one hour at room temperature, at which time the distance from the lowest point of the surface meniscus to the interface between red-cell free plasma and the portion of the sample occupied by red-cells measured. The distance moved by the erythrocyte interface, expressed as millimeters in 1 hour (mm/h) is the ESR.
The ESR is governed by the balance between pro-sedimentation factors, mainly fibrinogen (but possibly also the levels of serum C-reactive protein (CRP), immunoglobulins A and G, alpha(1)-acid-glycoprotein and alpha(1)-antitrypsin), and sedimentation resisting factors, mainly the negative charge of the erythrocytes (zeta potential). In one example of the effects of inflammation, high concentrations of fibrinogen in blood plasma causes red blood cells to adhere to each other. The red blood cells adhere to form stacks called ‘rouleaux,’ which settle faster than individual red cells. Rouleaux formation can also occur in association with some lymphoproliferative disorders in which one or more immunoglobulins are found in high concentrations. Rouleaux formation can, however, be a normal physiological finding in horses, cats, and pigs.
ESR is increased by any cause or focus of inflammation. ESR is increased in pregnancy and rheumatoid arthritis, and decreased in polycythemia, sickle cell anemia, hereditary spherocytosis, and congestive heart failure. The basal ESR is slightly higher in females.
The standard predicate method for measuring ESR is the Westergren test, and the test uses a large volume of blood, typically several ml. It typically requires one hour incubation since many samples have ESRs as low as 10 mm/hour. Inflammatory factors which increase ESR include fibrinogen, C-Reactive Protein (CRP) and some immunoglobulins, which can increase ESR to as high as 100 mm/hour.
Traditional techniques of performing sedimentation tests have various limitations. For instance as discussed, Westergren sedimentation tests require a substantially high volume of blood to be withdrawn. Additionally, traditional sedimentation test techniques take a substantial period of time and may result in time lags in obtaining test results that could lead to delays in diagnoses and treatments which can have a deleterious effect on a patient's health.
All publications, patents, and patent applications mentioned in this specification are herein incorporated by reference to the same extent as if each individual publication, patent, or patent application was specifically and individually indicated to be incorporated by reference.
It may be desirable to have sedimentation rate test that can be completed in a very short time, such as but not limited to being on the order of seconds to a few minutes. For distributed test settings, it may also be desirable to have sedimentation rate measurements that use only small blood volumes, such as can be obtained by alternate site, non-venous blood draws or minimal venous draws. It may be further desirable to make the sedimentation measurement in an automated fashion (no human observation required) and to create an objective record of the measurement Additionally, further information useful in optimizing management of patients may be obtained by performing and/or maximizing the speed of multiplexed measurement of other analytical parameters in parallel with sedimentation rate measurement.
In one embodiment described herein, the sedimentation rate measurement method may use (1) centrifugal techniques for separating red blood cells from plasma and (2) video and/or still imaging capability. Both may be used alone or in combination to accelerate erythrocyte sedimentation and to measure its rate. Of course, techniques other than centrifugation for accelerating sedimentation may be used in place of or in combination with centrifugation to separate blood components.
In one non-limiting example, the method may advantageously enable (1) rapid measurement of ESR (seconds) with small blood sample volumes such as about 20-25 microliters (“uL” or “μL”) or less, (2) use of automated image analysis to determine both red blood cell sedimentation rate and hematocrit, and/or (3) automated techniques to compensate for effects of hematocrit on uncorrected ESR so as to provide a value corresponding to the traditional Westergren method. Of course, alternative embodiments using large volumes of blood are not excluded. Because of the ability to correct for hematocrit, some embodiments of sedimentation measurement techniques described herein is more robust than traditional Westergren techniques and can be used on sample with fibrinogen and/or hematocrit levels outside the narrow range required by Westergren testing.
Using an embodiment herein, corrected ESR can be acquired in a matter of seconds using a small blood volume and which compensates for effects of hematocrit ESR. The results acquired in a matter of seconds during initial centrifugation can accelerate deliver of a diagnosis to the patient.
Moreover, in the context of multiplexed assay procedures, a common pre-processing step already involves separating red and white cells from plasma or serum prior to measurements of cellular markers and of analytes present in plasma/serum. Thus, it is convenient to incorporate an ESR measurement along with such a pre-processing procedure that will already be performed during the course of assay preparation. The ESR measurement will not create significant burden in terms of additional processing time or use of limited quantities of blood available from non-venous collection methods. By way of non-limiting example, it should be understood that assay processing, including pre-processing step(s), may occur in a single instrumented system. Optionally, some embodiments may perform one or more steps in one instrument and another one or more steps in another instrument.
It should also be understood that embodiments described herein may be adapted to have one or more of the features described below. In one non-limiting example, a typical protocol may take 20 uL of blood in a centrifuge vessel and spin in a swing-out centrifuge rotor at 4000 rpm (580*g) for about 10 s. During this time, the interface between the portion of the sample containing the red blood cells and that cleared of red blood cells is observed by video imaging. Although other time periods are not excluded, it can be advantageous to obtain the ESR measurement in this short period of time. Optionally, some embodiments may correct these “raw” ESR values for the effects of hematocrit. Hematocrit may be measured in the same operation as that used for measurement of raw ESR. In one non-limiting example, following a relatively low speed spin during centrifugation to measure ESR, the spin speed is increased to pack the red blood cells. Hematocrit is determined by image analysis of the packed red blood cells and the supernatant plasma volumes. Optionally, other techniques for measuring hematocrit may also be used to correct “raw” ESR values.
At least some of the embodiments herein may have ESR corrected without using calculations of the slope of an essentially linear transform of the non-linear (exponential) portion of the sedimentation curve.
At least some of the embodiments herein may have ESR corrected without calculating a mathematical function for a plurality of the erythrocyte/plasma interface positions occurring in a non-linear portion of the sedimentation curve.
At least some of the embodiments herein may have ESR corrected without selecting a segment of the sedimentation curve which lies in said non-linear portion of the sedimentation curve.
At least some of the embodiments herein may have ESR corrected based only on measurements of linear portion(s) of the sedimentation curve.
At least some of the embodiments herein may have ESR corrected based on measurements which consists essentially of linear portion(s) of the sedimentation curve. By “consists essentially of”, we mean at least 90% or more of the measurement is based on the linear portion(s).
At least some of the embodiments herein may have ESR corrected without determining a mathematical function for a non-linear segment of the sedimentation curve representative of the magnitude of intercellular erythrocyte repulsion in the blood sample.
At least some of the embodiments herein may have ESR corrected without negating the time period during the centrifugation of the sample during which a linear portion of the sedimentation curve is formed.
At least some of the embodiments herein may have ESR corrected for hematocrit using hematocrit measurements not derived from centrifugal techniques, such as for example, lysis of red cells with detergent and mixing with ferricyanide and cyanide followed by measurement of the absorbance of the cyan-met-hemoglobin formed.
At least some of the embodiments herein may have the blood sample adjusted so that it is at a known hematocrit level for the sedimentation measurement.
In at least one embodiment described herein, a method is provided comprising: using an accelerated blood component separation technique on a blood sample for a period of time to separate formed blood components from plasma; determining a sedimentation rate of the formed blood component based on at least the following: a time-related compaction curve and a hematocrit correction factor, wherein the time-related compaction curve for at least one formed blood component in said blood sample is determined after accelerated blood component separation has begun, said compaction curve having an initial approximately linear portion and a non-linear portion after the linear portion.
In at least one embodiment described herein, a method is provided comprising: centrifuging a blood sample in a vessel for a period of time; establishing a time-related compaction curve for at least one formed blood component in said blood sample after centrifuging has begun, said compaction curve having an initial approximately linear portion; correcting for hematocrit effect on sedimentation rate of the formed blood component by using a hematocrit correction factor on the approximately linear portion of said compaction curve.
It should be understood that embodiments in this disclosure may be adapted to have one or more of the features described below. In one non-limiting example, the method comprises calibrating sedimentation rates from centrifuge based technique with sedimentation rates from a reference technique. Optionally, the reference technique is the Westergren technique. Optionally, the sample is about 25 uL or less. Optionally, centrifuging occurs at a first speed for a first period of time and then at a second, faster speed for a second period of time. Optionally, centrifuging comprises using a centrifuge configured to allow the blood sample to be visually observed during centrifugation to establish interface positions of one or more formed blood components in the blood sample. Optionally, centrifuging comprises using a centrifuge having a window thereon to enable visual observation of the blood sample to establish erythrocyte/plasma interface positions over time. Optionally, centrifuging comprises using a centrifuge, a light source, and an image capture device to enable visual observation of the blood sample to establish formed blood component/plasma interface positions over time. Optionally, compaction curve data is collected by capturing a plurality of images of interface positions of one or more formed blood components in a centrifuge vessel over the time period. Optionally, pixel positions in the plurality of images are used to accurately determine interface position. Optionally, compaction curve data is collected by capturing a single image of an interface position of one or more formed blood components in the centrifuge vessel after a time period, wherein sedimentation rate is calculated based on position of the meniscus of the supernatant liquid and the interface position. Optionally, compaction curve data is collected while the sample is being centrifuged. Optionally, centrifugation is used to obtain hematocrit measurement and to correct for hematocrit impact on sedimentation rate measurement. Optionally, correcting for hematocrit comprises calculating a mathematical function for a plurality of formed blood component interface positions occurring in said curve, said function being operative to correct for sedimentation rate variations due to hematocrit. Optionally, hematocrit measurement in the sample is derived from a technique separate from centrifugation Optionally, image transformation is used for conversion of a curved interface to a flat interface. Optionally, image transformation parameters are selected, video of formed blood component interface position is put through image transformation, and then a region of interest is chosen that covers both the whole range of positions for both air/plasma interface and erythrocyte interface. Optionally, each timepoint in the video, pixel intensity values for each row across a sample vessel containing the sample, within the region of interest are averaged to produce a single column representing the intensity radially down the sample vessel. Optionally, a linear region of a sedimentation profile is used to extract a sedimentation rate. Optionally, the formed blood component is white blood cells. Optionally, the formed blood component is platelets.
It should be understood that embodiments in this disclosure may be adapted to have one or more of the features described below. In one non-limiting example, the method comprises performing image transformation on said images to transform images with curved interfaces into corrected images with straight line interfaces; establishing a time-related compaction curve based on interface positions in said corrected images, for at least one formed blood component in said blood sample after centrifuging has begun. Optionally, the method comprises: using a programmable processor-controlled system to transfer at least a portion of a blood sample from a blood sample location into a centrifugation vessel; using a sample handling system under programmable processor control to transfer said vessel from a first addressable position to a centrifuge with a second addressable position; centrifuging the blood sample in the vessel for a period of time; collecting at least one image of formed blood component and plasma interface position after centrifuging; establishing a time-related compaction curve based on interface position(s) in the image, for at least one formed blood component in said blood sample after centrifuging has begun. Optionally, the vessel is removed from the centrifuge to obtain said image. Optionally, the vessel is returned to the centrifuge after said image is obtained. Optionally, the method comprises varying centrifuging speed to establishing a linear compaction curve of at least one formed blood component over the period of time until compacting has completed; monitoring centrifuging speed profile for at least a portion of the time period; and determining blood component sedimentation rate based on the centrifuging speed profile. Optionally, the method comprises collecting at least a first single image of formed blood component and plasma interface positions at an initial time; collecting at least a second single image of formed blood component and plasma interface positions at a second time while rate of sedimentation is still linear; calculating sedimentation rate for at least one formed blood component in said blood sample based on linear sedimentation rate calculated and a hematocrit correction factor.
In yet another embodiment described herein, a method is provided comprising: centrifuging a blood sample in a vessel for a period of time; using imaging of the vessel in a single state condition to establish sedimentation rate; and correcting for hematocrit effect on sedimentation rate of the formed blood component by using a hematocrit correction factor. As used herein, using imaging can comprise using a single image of the vessel to determine sedimentation rate. By of non-limiting example, when using a single image of the vessel, a meniscus of supernatant liquid in the vessel shows an initial level and an interface position of formed components with the supernatant liquid shows current position, from which sedimentation rate is calculated. Optionally, some embodiment may use a plurality of images for sedimentation rate calculation, but all of the images are of the vessel while the vessel is in a single state condition. Optionally, all of the images are of the vessel at a single point in time. Optionally, all of the images are of the vessel while the formed component interface position is not changing in the vessel.
This Summary is provided to introduce a selection of concepts in a simplified form that are further described below in the Detailed Description. This Summary is not intended to identify key features or essential features of the claimed subject matter, nor is it intended to be used to limit the scope of the claimed subject matter.
It is to be understood that both the foregoing general description and the following detailed description are exemplary and explanatory only and are not restrictive of the invention, as claimed. It may be noted that, as used in the specification and the appended claims, the singular forms “a”, “an” and “the” include plural referents unless the context clearly dictates otherwise. Thus, for example, reference to “a material” may include mixtures of materials, reference to “a compound” may include multiple compounds, and the like. References cited herein are hereby incorporated by reference in their entirety, except to the extent that they conflict with teachings explicitly set forth in this specification.
In this specification and in the claims which follow, reference will be made to a number of terms which shall be defined to have the following meanings:
“Optional” or “optionally” means that the subsequently described circumstance may or may not occur, so that the description includes instances where the circumstance occurs and instances where it does not. For example, if a device optionally contains a feature for a sample collection well, this means that the sample collection well may or may not be present, and, thus, the description includes both structures wherein a device possesses the sample collection well and structures wherein sample collection well is not present.
Referring now to
Sedimentation Curve Measurement
A variety of techniques may be used to establish a sedimentation rate curve for one or more formed blood components. Although the present application is described mostly in the context of measuring erythrocyte sedimentation rate, systems and methods herein can also be adapted for use in measuring sedimentation rates for other formed blood components such as but not limited to white blood cells, platelets, or the like.
In one nonlimiting example, one technique described herein comprises taking images at several time points during sedimentation by placing the sample vessel in a centrifuge, spinning for a few seconds, stopping the spin, removing the vessel, placing it in a viewer, taking an image, and repeating the above to obtain multiple images over time. From a device simplicity standpoint, it is helpful in that it simplifies hardware implementation for obtaining such images. The ability to measure sedimentation is discussed elsewhere herein where the slope from the initial (linear) part of the sedimentation curve is used to calculate the ESR.
It should be understood of course, that some embodiments may obtain such images/data regarding interface position while the container is in-situ in the centrifuge and without having to stop the centrifuge to remove the sample vessel for imaging. The in-situ images may be taken while the centrifuge rotor is in motion or at rest. It should also be understood that although discrete images may be taken, videos, continuous imaging, and multi-frame per second imaging may also be used.
Referring now to
Other suitable centrifuge designs and features including dimensions for the centrifuge vessels, construction of the centrifuge rotor, and centrifuge size are disclosed in co-pending U.S. patent application Ser. Nos. 13/355,458 and 13/244,947, all fully incorporated herein by reference for all purposes. Other components of the present system including suitable imaging devices and fluid handling systems are also described in the applications incorporated by reference. For example, the ability of digital cameras such as those described in those applications may be used to measure very small distances and rates of change of distances to measure ESR. Image analysis can be used to measure the movement of the interface between red cells and plasma.
By way of nonlimiting example, in some embodiments, only two measurements taken at early times (seconds) after centrifugation has begun is sufficient to define the sedimentation rate with high precision. In one embodiment, some may take the first image after an initial minimum centrifuge speed is reached and then a second image may be taken about 10 seconds later. Of course, other time periods for the images are not excluded so long as they are in the linear portion of the sedimentation curve.
Viewing
As seen in
In one embodiment, a method is provided which uses a transparent window in the centrifuge housing so that a video record of sedimentation can be made during the low-speed centrifugation. Moreover, the centrifugal field causes the meniscus to become straighter (at right angles to the centrifugal force vector) making measurement of small settling distances easier. This may be particularly true when images are captured while the centrifuge rotor is spinning. By spinning a small volume (20-25 uL) of blood at intermediate speeds (typically 4000 rpm, although 2000 to 6000 rpm may also suitable), almost complete sedimentation of red blood cells is achieved in this embodiment within about three minutes. In practice, one method may take sedimentation measurements for a few seconds at relatively low speed (4000 rpm) then the speed would be increased to about 10,000 rpm for about three minutes to pack the red blood cells and determine the hematocrit. This multi-stage spinning at different centrifugation speeds allows for imaging for sedimentation and then rapid spin down to achieve compaction of blood components and separation from blood plasma.
Referring now to
Any descriptions of cameras, or other detection devices described elsewhere herein may apply. In one example, the image capture device 110 may be a digital camera. Image capture devices may also include charge coupled devices (CCDs) or photomultipliers and phototubes, or photodetector or other detection device such as a scanning microscope, whether back-lit or forward-lit. In some instances, cameras may use CCDs, CMOS, may be lensless (computational) cameras (e.g., Frankencamera), open-source cameras, or may use any other visual detection technology known or later developed in the art. Cameras may include one or more feature that may focus the camera during use, or may capture images that can be later focused. In some embodiments, imaging devices may employ 2-d imaging, 3-d imaging, and/or 4-d imaging (incorporating changes over time). Imaging devices may capture static images. The static images may be captured at one or more point in time. The imaging devices may also capture video and/or dynamic images. The video images may be captured continuously over one or more periods of time. Any other description of imaging devices and/or detection units may also be applied, preferably so long as they are able to detect changes in interface position.
In one non-limiting example, a light source 112 may be a light-emitting diode (LED) (e.g., gallium arsenide (GaAs) LED, aluminum gallium arsenide (AlGaAs) LED, gallium arsenide phosphide (GaAsP) LED, aluminum gallium indium phosphide (AlGaInP) LED, gallium(III) phosphide (GaP) LED, indium gallium nitride (InGaN)/gallium(III) nitride (GaN) LED, or aluminum gallium phosphide (AlGaP) LED). In another example, a light source can be a laser, for example a vertical cavity surface emitting laser (VCSEL) or other suitable light emitter such as an Indium-Gallium-Aluminum-Phosphide (InGaAIP) laser, a Gallium-Arsenic Phosphide/Gallium Phosphide (GaAsP/GaP) laser, or a Gallium-Aluminum-Arsenide/Gallium-Aluminum-Arsenide (GaAIAs/GaAs) laser. Other examples of light sources may include but are not limited to electron stimulated light sources (e.g., Cathodoluminescence, Electron Stimulated Luminescence (ESL light bulbs), Cathode ray tube (CRT monitor), Nixie tube), incandescent light sources (e.g., Carbon button lamp, Conventional incandescent light bulbs, Halogen lamps, Globar, Nernst lamp), electroluminescent (EL) light sources (e.g., Light-emitting diodes—Organic light-emitting diodes, Polymer light-emitting diodes, Solid-state lighting, LED lamp, Electroluminescent sheets Electroluminescent wires), gas discharge light sources (e.g., Fluorescent lamps, Inductive lighting, Hollow cathode lamp, Neon and argon lamps, Plasma lamps, Xenon flash lamps), or high-intensity discharge light sources (e.g., Carbon arc lamps, Ceramic discharge metal halide lamps, Hydrargyrum medium-arc iodide lamps, Mercury-vapor lamps, Metal halide lamps, Sodium vapor lamps, Xenon arc lamps). Alternatively, a light source may be a bioluminescent, chemiluminescent, phosphorescent, or fluorescent light source.
As seen in
In the present embodiment of
In one nonlimiting example, the positions of the top and/or bottom of the centrifuge vessel are obtained by imaging as reference points, and later these are used to calibrate the liquid and interface levels.
Referring now to
It should be understood that strobe illumination or capture frames synchronized to the rotor position are not excluded, but in the present embodiment, are not required for image capture. In this nonlimiting example for the image of
As seen in
In one embodiment, images are captured for three minutes at five frames per second, with long exposure (˜200 ms), then processed to extract the sedimentation curve. Optionally, the rate of imaging includes but is not limited to 1, 2, 4, 8, 16, 32, 64, or 128 images per second. Optionally, exposure time includes but is not limited to 10, 20, 40, 80, 160, 320, or 640 ms. Temperature during measurement may also be varied. Although many embodiments herein had measurements performed at room temperature, but other temperatures e.g. 37 C are not excluded. Effect of temperature would be taken into account in the calibration, such as for determining empirical parameters of the correction factor. Also, time to centrifuge spin up was typically about 3 seconds, but faster or slower spin up times are not excluded.
By way of nonlimiting example, the sedimentation rate of the desired formed blood component being measured may be defined by:
Although others settings are not excluded, it should be understood that times of sedimentation are usually defined as starting following the rotor 116 reaching its target speed when the buckets holding the centrifuge vessels 114 are oriented radially in the spin plane and so are in optimal position for image capture and processing.
Image Transformation
Referring now to
As seen in
This image transformation is desirable to remove the effects of the vertical lines seen in
Referring now to
For each dataset using a script that produces a series of images with different arc and rotation angle correction, superimposing a series of horizontal lines 160 on the images allows for judgment of when the interfaces are flat (horizontal in the images of
Once these parameters are selected, the acquired image information, which may be a video, is put through the transformation. A region of interest may be chosen that covers both the whole range of positions for both the air/plasma interface 130 and the erythrocyte interface 132. Optionally, some embodiments may choose a region of interest covering only one of the interfaces 130 or 132. Optionally, some embodiments may be configured to target one or more other areas of interest in the sample.
Sedimentation Curve Extraction
Referring now to
In one nonlimiting example, the positions of the two local maxima of the image in
As seen in
Calculating ESR from Sedimentation Curve
Once the sedimentation curve of
One such example is shown in
Referring now to
Although these simple techniques of using standard nonlinear least squares fitting may yield some information related to ESR, when comparing such measurements with traditional Westergren ESR measurements, the correlation based on the nonlinear least squares (NLS) fitting leaves room for improvement as NLS by itself does not take into account certain correction factors.
Plasma Protein Impact on ESR
To extract ESR parameters that more closely correlate with traditional Westergren ESR measurements, it is helpful to understand some factors which may impact ESR measurements. The parameter of interest (ESR) responds to the concentration of certain plasma proteins and can be directly affected/manipulated by adding one of these proteins, (e.g., fibrinogen) to the blood sample.
In the present example, as a technique to provide samples with a wide range of ESR values, exogenous fibrinogen was used to create blood samples with ESR values spanning the whole range of interest (0-120 mm/h in the Westergren method).
As seen in
Hematocrit Impact on ESR
It should be understood that, in addition to fibrinogen, hematocrit is another factor that affects Westergren and other ESR measurements. In fact, Westergren erythrocyte sedimentation is strongly affected by hematocrit. In the Westergren method, many laboratories either do not report results for samples with hematocrits greater than about 45% or adjust the sample hematocrit to a fixed level (usually 45%) before measuring ESR. The present embodiment of the method is actually better than the Westergren technique, in that Westergren saturates (i.e. does not respond to fibrinogen <10 mg/ml), whereas the present embodiment of the method does not saturate out to 15 mg/ml.
Centrifuge-based ESR sedimentation is even more strongly affected by hematocrit levels than measurements under gravity. For at least some embodiments here, the increased dependency on hematocrit is also because of the lower volumes—and consequently smaller vessel dimensions. Increasing hematocrit typically means the erythrocytes start closer together, increasing the viscosity of the blood by presenting physical barriers to free movement, and decreasing the maximum distance the interface can move before the cells become packed, all of which decrease the ESR, independent of fibrinogen from inflammation.
To illustrate the dramatic confounding effect of hematocrit, centrifuge-based ESR measurements, performed by taking the same sample of blood and adjusting the hematocrit before measuring the ESR, show that a person with a typical hematocrit of 45% and a normal ESR of 22 mm/h would register as 5 mm/h (very low) if the hematocrit were 60% and 93 mm/h (very high) if the hematocrit were 35%, even though there are no changes in the plasma protein levels which are clinically important. In other words, variations of ESR due to hematocrit can dominate variations of ESR due to plasma proteins the clinician is interested in.
There are several traditional approaches to compensating for this confounding effect of hematocrit. One approach is using the hematocrit compensation curves, e.g. from Dintenfass (1974). Rather than correcting for hematocrit using a chart, a more accurate (if more labor-intensive) way to remove the confounding effect is simply to alter the hematocrit to a standard value before the test. Some ESR techniques, e.g. the ‘hematocrit corrected ESR’ include an initial such step that fixes the hematocrit to a set value e.g. 45%, so that the measured ESR really reflects the protein content of the plasma (clinically relevant), rather than the hematocrit (Borawski and Myśliwiec 2001).
As seen in
Referring now to
In
As a further example,
To obtain a better correlation between the centrifuge-based and Westergren sedimentation rates, the centrifuge-based sedimentation rates were corrected for effect of hematocrit. The correction used can be represented by,
where Uuncorr and Ucorr are the uncorrected (raw) and corrected sedimentation rates respectively, φ is the volume fraction of cells (hematocrit), and φmax and γ are empirical parameters obtained by curve fitting. The correction factor represents a simple mathematical form to account for the increased drag exerted by red blood cells. It should be understood that this functional form was found to be able to correct for hematocrit, but other functions would work too.
By way of non-limiting example, one way of calculating φmax and γ is by way of a calibration technique such as but not limited to the following: for a diverse set of samples (different hematocrits, ESR values, etc. . . . ), the ESR value is determined using a reference method, and by the centrifuge-based method. The φmax and γ parameters are determined as a calibration for each centrifuge setup and may change based at least in part on vessel geometry and volume of sample. Thus, if at least one of those factors is changed, it may be desirable to re-calculate the parameters. For one centrifuge setup as described herein, optimal values of these parameters were obtained as: φmax=1.67 and γ=3.85. It should be understood that these parameters are for fit optimization and do not relate directly to physical parameters.
Hematocrit Measurement Techniques
For purposes of calculating the hematocrit correction factor, it should be understood that the value for hematocrit may be known prior to the start of the centrifuge-based sedimentation test and in such situations, corrected ESR results can be obtained quickly based on the initial linear portion of the sedimentation and the known hematocrit level, without having to wait until the erythrocytes have been fully compacted by centrifugation. Optionally, some embodiments may determine hematocrit levels during or after centrifugation.
Hematocrit measurement by non-centrifugal before, during, or after centrifugation includes at least the following. One technique involves measurement of hemoglobin concentration. For example, in roughly 99% of the population, there is a 1:1 correlation between hemoglobin measurements and hematocrit levels. Thus, if hemoglobin test data is available, the hematocrit level is generally already known before start of the centrifuge-based sedimentation test.
Referring now to
Using an assay protocol for hemoglobin-based measurement, correlation of results with hematocrit measurement will now be discussed. Human blood samples were processed by recombining plasma and red cells (collected by centrifugation) to provide a wide range of hematocrit values. These samples were assayed as above and by a standard centrifugal capillary tube hematocrit assay and results correlated as shown below. As seen in
Another technique for hematocrit measurement involves microscopic imaging. Hematocrit can also be measured in the devices using a cuvette with a fixed depth and a blood sample diluted to a known extent. Description of a system with such a cuvette can be found in U.S. patent application Ser. No. 13/244,947 fully incorporated herein by reference for all purposes. Hematocrit can be determined by microscopic measurement of the (1) the red cell count per field of view and (2) the average red cell volume. Favored methods are: (1) Dark field microscopy and (2) (1)+Fluorescence microscopy using fluorescently-labeled anti-human CD-35 (red cell surface antigen). Image analysis techniques are then applied.
Specifically, one method of measuring hematocrit may involve measuring optical density of the sample. See for example Lipowsky et al. “Hematocrit determination in small bore tubes from optical density measurements under white light illumination” Microvascular Research, Volume 20, Issue 1, July 1980, Pages 51-70; http://dx.doi.org/10.1016/0026-2862(80)90019-9, fully incorporated herein by reference for all purposes. Lipowsky discusses the relationship between the hematocrit of blood flowing in small bore glass tubes and its optical density (OD) under white light (tungsten) illumination has been examined for various tube luminal diameters. In at least some embodiments herein, all this data is available since a small-bore tube of blood is being illuminated.
In another embodiment, hematocrit level can be determined by testing a portion of the blood sample under microscopy or other magnified observation, such as but not limited to measuring the number and mean size of red blood cells in a defined observation area which may have a known size. In this manner, the hematocrit may be determined based on such visual characterization of the red blood cells.
In yet another example, hematocrit level can be measured based on a completed centrifugation of the blood sample which compacts the red blood cells. This compacted level can be used to determine hematocrit. In this example, only linear portions of the centrifuge based sedimentation test are used to determine a corrected ESR. By way of non-limiting example, the first initial portion of interface position measurement which is linear, along with the final end portion which is also linear, are two portions of the sedimentation which may be used to calculate an ESR corrected for hematocrit. As seen in
The above is a non-exhaustive listing of hematocrit calculation techniques and other methods of measuring hematocrit levels are not excluded from use with the sedimentation measurement techniques described herein.
In one non-limiting example of how this works together for an ESR measurement, a vessel containing the sample may be centrifuged under controlled conditions for a first period of time, such as but not limited to time period associated with the linear portion of the sedimentation curve. In one non-limiting example, the centrifugation is controlled to a specific rpm range for the rotor of the centrifuge such that there is a substantially consistent force applied to the accelerated sedimentation process. The sedimentation profile during the initial period, as previously noted, is generally linear, and it may be desirable to capture the sedimentation image of formed components in the material after centrifugation during the initial linear period. By way of non-limiting example, the image can be taken in several scenarios such as but not limited to: a) while the vessel is in the centrifuge, b) when the vessel is in the centrifuge but the centrifuge is stopped, or c) by removing the vessel from the centrifuged and imaging it. In some embodiments, the sedimentation can be measured with a single picture. It should be understood that the level at the start time t0 is the meniscus level of the supernate or the remaining solution above the settled formed components. The level of sedimentation is the level of the settled formed components in the image after the initial period of accelerated sedimentation.
Hematocrit measurement, used for the ESR calculation, can be performed using at least one method such as but not limited to those described herein. It can be performed in the same system as the one with the centrifuge, or optionally, it can be performed using a physically separate instrument. In one non-limiting example, after the image of the sedimentation is obtained, the sample may be centrifuged to complete the sedimentation and pack the formed components into a pellet. The effective gravitational force on a vessel can completely cause the precipitate (“pellet”) to gather on the bottom of the tube. The supernatant liquid is then either be decanted from the tube without disturbing the precipitate, or withdrawn with by a pipette.
Graphs of Hematocrit Corrected ESR
With hematocrit-adjusted clinical samples, good correlations with ESR within each hematocrit level, and, as expected, significant effects of hematocrit were also found. The centrifugation method can also be used to obtain accurate values for the hematocrit, and the impact of hematocrit can be corrected.
Referring now to
Estimated Westergren ESR=10{circumflex over ( )}(((LOG(HCT corrected ESR)−LOG(644.11))/0.1367)), where the relationship and parameters used are derived from the analysis of FIG. 18B.
Experimental Methods
The data obtained for the various charts were obtained using the following techniques. These are provided as examples and are meant to be non-limiting.
Samples:
Fresh EDTA-anticoagulated blood samples were used. EDTA is used as this is the standard for the “Modified Westergren” method. Samples were kept at room temperature and re-suspended prior to measurement.
Hematocrit Adjustment:
Samples were spun down for hematocrit packing (e.g. 5000 Relative centrifugal force (RCF) for 20 minutes), and plasma is separated from the cells. Red blood cells were slurried with plasma from the same sample and more plasma added to give a desired hematocrit level.
Westergren ESR Measurements:
A 1 mL sample is required to perform the Westergren ESR measurement (‘Sedigren’ brand tubes used, following protocol enclosed therein). Red blood cell sedimentation was observed and measured by video recording.
Adjustment of RBC Zeta Potential (and ESR) with Fibrinogen:
For the examples shown in
Measurement of Centrifuge Sedimentation Curve:
A 25 uL of whole blood sample was added to a centrifuge vessel. A swinging bucket centrifuge as described in co-pending U.S. patent application Ser. Nos. 13/355,458 and 13/244,947 were modified with a slot cut to allow light to pass through the bucket when spinning in a horizontal fashion (axis of rotation vertical). In this non-limiting example, the light source may be a 1 W green LED, such as available from Thorlabs of Newton N.J., that is brightness adjusted (typically ˜10%) so that light reaching the detector did not saturate it. A webcam or other imaging device such as available from Logitech was positioned at 10 mm above the plane of rotation as shown in FIG. 3. Integration time was 200 ms. Images were taken at 5 frames/second (fps) using a loss-less compression codec (“huffyuv”) over known times up to three minutes.
Image Transformation:
images obtained visual observations of the centrifuge vessels during centrifugation were processed in the manner as described herein for
Sedimentation Curve Extraction:
the position over time of the red blood cell/plasma and other interfaces in the images were then plotted in the manner as described herein for
Curve Fitting with Hematocrit Correction Factor:
The sedimentation curve is then further processed by way of curve fitting using various techniques described herein for
Measurement of Non-Red Blood Cell Blood Components
Although the present description is written primarily in the context of measuring erythrocyte sedimentation rate, it should be understood that the techniques herein can be adapted for use to measure sedimentation rates of other formed blood components that are not erythrocytes. Some embodiments may measure platelet sedimentation. Some embodiments may measure white blood cell sedimentation. Optionally, sedimentation of other formed components can also be measured.
By way of nonlimiting example in
Thus as seen in
By way of nonlimiting example, it should be understood that the use of change in refractive index or possibly light scattering can be used as a measure of the blood component interface position, rather than change in absorbance. Optionally, some embodiments may use both.
Assay Processing in an Integrated. Automated System
Referring now to
In one non-limiting example as shown in
As seen in
All of the foregoing may be integrated within a single housing 520 and configured for bench top or small footprint floor mounting. In one example, a small footprint floor mounted system may occupy a floor area of about 4 m2 or less. In one example, a small footprint floor mounted system may occupy a floor area of about 3 m2 or less. In one example, a small footprint floor mounted system may occupy a floor area of about 2 m2 or less. In one example, a small footprint floor mounted system may occupy a floor area of about 1 m2 or less. In some embodiments, the instrument footprint may be less than or equal to about 4 n2, 3 n2, 2.5 n2, 2 m2, 1.5 m2, 1 m2, 0.75 m2, 0.5 m2, 0.3 m2, 0.2 m2, 0.1 m2, 0.08 m2, 0.05 m2, 0.03 m2, 100 cm2, 80 cm2, 70 cm2, 60 cm2, 50 cm2, 40 cm2, 30 cm2, 20 cm2, 15 cm2, or 10 cm2. Some suitable systems in a point-of-service setting are described in U.S. patent application Ser. Nos. 13/355,458 and 13/244,947, both fully incorporated herein by reference for all purposes. The present embodiments may be configured for use with any of the modules or systems described in those patent applications.
While the invention has been described and illustrated with reference to certain particular embodiments thereof, those skilled in the art will appreciate that various adaptations, changes, modifications, substitutions, deletions, or additions of procedures and protocols may be made without departing from the spirit and scope of the invention. For example, with any of the above embodiments, it should be understood that other techniques for plasma separation may also be used with or in place of centrifugation. For example, one embodiment may centrifuge the sample for an initial period, and then the sample may be located into a filter that then removes the formed blood components to complete separation. Although the present embodiments are described in the context of centrifugation, other accelerated separation techniques may also be adapted for use with sedimentation rate measurement methods described herein. Some embodiments may optionally combine the hematocrit correction techniques described herein with measurement techniques as described in U.S. Pat. No. 6,204,066 fully incorporated herein by reference for all purposes. Some embodiments herein may pre-process the blood sample to pre-set the hematocrit value in the blood sample to a pre-determined value so that the variable due to hematocrit is removed. Some embodiments may also use traditional techniques for adjusting for hematocrit levels. It should also be understood that although the present embodiments are described in the context of blood samples, the techniques herein may also be configured to be applied to other samples (biological or otherwise).
Optionally, at least one embodiment may use a variable speed centrifuge. With feedback, such as but not limited to imaging of the position of interface(s) in the sample, the speed of the centrifuge could be varied to keep the compaction curve linear with time (until fully compacted), and the ESR data extracted from the speed profile of the centrifuge rather than the sedimentation rate curve. In such a system, one or more processors can be used to feedback control the centrifuge to have a linear compaction curve while speed profile of the centrifuge is also recorded. Depending on which interface is being tracked, the sedimentation rate data is calculated based centrifuge speed. In one non-limiting example, a higher centrifuge speed is used to keep a linear curve as the compaction nears completion.
Furthermore, those of skill in the art will recognize that any of the embodiments of the present invention can be applied to collection of sample fluid from humans, animals, or other subjects. Optionally, the volume of blood used for sedimentation testing may be 1 mL or less, 500 μL or less, 300 μL or less, 250 μL or less, 200 μL or less, 170 μL or less, 150 μL or less, 125 μL or less, 100 μL or less, 75 μL or less, 50 μL or less, 25 μL or less, 20 μL or less, 15 μL or less, 10 μL or less, 5 μL or less, 3 μL or less, 1 μL or less, 500 nL or less, 250 nL or less, 100 nL or less, 50 nL or less, 20 nL or less, 10 nL or less, 5 nL or less, or 1 nL or less.
Additionally, concentrations, amounts, and other numerical data may be presented herein in a range format. It is to be understood that such range format is used merely for convenience and brevity and should be interpreted flexibly to include not only the numerical values explicitly recited as the limits of the range, but also to include all the individual numerical values or sub-ranges encompassed within that range as if each numerical value and sub-range is explicitly recited. For example, a size range of about 1 nm to about 200 nm should be interpreted to include not only the explicitly recited limits of about 1 nm and about 200 nm, but also to include individual sizes such as 2 nm, 3 nm, 4 nm, and sub-ranges such as 10 nm to 50 nm, 20 nm to 100 nm, etc . . . .
The publications discussed or cited herein are provided solely for their disclosure prior to the filing date of the present application. Nothing herein is to be construed as an admission that the present invention is not entitled to antedate such publication by virtue of prior invention. Further, the dates of publication provided may be different from the actual publication dates which may need to be independently confirmed. All publications mentioned herein are incorporated herein by reference to disclose and describe the structures and/or methods in connection with which the publications are cited. The following are fully incorporated herein by reference for all purposes: U.S. patent application Ser. Nos. 13/355,458 and 13/244,947 and U.S. Provisional Application Ser. No. 61/673,037 entitled “Rapid Measurement of Formed Blood Component Sedimentation Rate from Small Sample Volumes” filed Jul. 18, 2012; U.S. Pat. Nos. 8,380,541, 8,088,593; U.S. Patent Publication No. 2012/0309636; U.S. Pat. App. Ser. No. 61/676,178, filed Jul. 26, 2012; PCT/US2012/57155, filed Sep. 25, 2012; U.S. application Ser. No. 13/244,946, filed Sep. 26, 2011; U.S. patent application Ser. No. 13/244,949, filed Sep. 26, 2011; and U.S. Application Ser. No. 61/673,245, filed Sep. 26, 2011.
Various aspects of at least some embodiments described herein are enumerated in the following paragraphs:
Aspect 1. A method comprising: using an accelerated blood component separation technique on a blood sample for a period of time to separate formed blood components from plasma; establishing a time-related compaction curve for at least one formed blood component in said blood sample after accelerated blood component separation has begun, said compaction curve having an initial approximately linear portion; determining sedimentation rate of the formed blood component based on at least the following: the compaction curve and a hematocrit correction factor.
Aspect 2. A method comprising: centrifuging a blood sample in a vessel for a period of time; establishing a time-related compaction curve for at least one formed blood component in said blood sample after centrifuging has begun, said compaction curve having an initial approximately linear portion; correcting for hematocrit effect on sedimentation rate of the formed blood component by using a hematocrit correction factor on the approximately linear portion of said compaction curve.
Aspect 3. A method comprising: centrifuging a blood sample in a vessel for a period of time; establishing a time-related compaction curve for at least one formed blood component in said blood sample after centrifuging has begun; correcting for impact of hematocrit on sedimentation rate of the formed blood component using a hematocrit correction factor based on the formula:
where Uuncorr and Ucorr are the uncorrected (raw) and corrected sedimentation rates, φ is the volume fraction of cells (hematocrit), and φmax and γ are empirical parameters obtained by curve fitting
Aspect 4. The method of any one of the foregoing aspects, wherein curve fitting for the hematocrit correction factor comprises calibrating sedimentation rates from centrifuge based technique with sedimentation rates from a reference technique.
Aspect 5. The method of any one of the foregoing aspects, wherein the reference technique is the Westergren technique.
Aspect 6. The method of any one of the foregoing aspects, wherein fibrinogen levels as high as 15 mg/ml does not impact sedimentation rate measurement.
Aspect 7. The method of any one of the foregoing aspects, wherein said blood sample is about 100 uL or less.
Aspect 8. The method of any one of the foregoing aspects, wherein said blood sample is about 50 uL or less.
Aspect 9. The method of any one of the foregoing aspects, wherein said blood sample is about 25 uL or less.
Aspect 10. The method of any one of the foregoing aspects, wherein centrifugation occurs at a first speed for a first period of time and then at a second, faster speed for a second period of time.
Aspect 11. The method of any one of the foregoing aspects, wherein centrifugation uses a centrifuge configured to allow the blood sample to be visually observed during centrifugation to establish interface positions of one or more formed blood components in the blood sample.
Aspect 12. The method of any one of the foregoing aspects, wherein centrifugation uses a centrifuge having a window thereon to enable visual observation of the blood sample to establish erythrocyte/plasma interface positions over time.
Aspect 13. The method of any one of the foregoing aspects, wherein centrifugation uses a centrifuge, a light source, and an image capture device to enable visual observation of the blood sample to establish formed blood component/plasma interface positions over time.
Aspect 14. The method of any one of the foregoing aspects, wherein compaction curve data is collected by capturing a plurality of images of interface positions of one or more formed blood components in the centrifuge vessel over the time period.
Aspect 15. The method of aspect 14, wherein pixel positions in the plurality of images are used to accurately determine interface position.
Aspect 16. The method of aspect 14, wherein capturing of images begins once the centrifuge has reached a minimum operating speed.
Aspect 17. The method of aspect 14, wherein capturing of images begins when the centrifuge has begins rotation.
Aspect 18. The method of any one of the foregoing aspects, wherein compaction curve data is collected while the sample is being centrifuged.
Aspect 19. The method of any one of the foregoing aspects, wherein centrifugation is used to obtain accurate values for the hematocrit and to correct for hematocrit impact on sedimentation rate measurement.
Aspect 20. The method of any one of the foregoing aspects, wherein correcting for hematocrit comprises calculating a mathematical function for a plurality of formed blood component interface positions occurring in said curve, said function being operative to correct for sedimentation rate variations due to hematocrit.
Aspect 21. The method of any one of the foregoing aspects, wherein hematocrit correction factor is determined without using data from a non-linear portion of the compaction curve.
Aspect 22. The method of any one of the foregoing aspects, wherein hematocrit level in the sample is derived from a technique separate from centrifugation.
Aspect 23. The method of any one of the foregoing aspects, wherein φmax and γ are for fit optimization and do not relate directly to physical parameters.
Aspect 24. The method of any one of the foregoing aspects, further comprising image transformation for conversion of a curved interface to a flat interface.
Aspect 25. The method of any one of the foregoing aspects, wherein hematocrit correction is capable of essentially eliminating the effects of hematocrit on formed blood component sedimentation rate.
Aspect 26. The method of any one of the foregoing aspects, wherein image transformation parameters are selected, video of formed blood component interface position is put through image transformation, and then a region of interest is chosen that covers both the whole range of positions for both air/plasma interface and erythrocyte interface.
Aspect 27. The method of any one of the foregoing aspects, wherein for each timepoint in the video, pixel intensity values for each row across the sample vessel within the region of interest are averaged to produce a single column representing the intensity radially down the sample vessel.
Aspect 28. The method of any one of the foregoing aspects, wherein columns for each timepoint are then assembled into a kymograph.
Aspect 29. The method of aspect 28 wherein positions of the two local maxima of the image, one representing the air/plasma interface and other the plasma/erythrocyte interface are determined.
Aspect 30. The method of aspect 28 comprising converting pixel positions into volume occupied by the whole sample and volume occupied by red blood cells, wherein the y-position of the top and bottom of the centrifuge vessel are used as reference locations together with knowledge of the shape of the centrifuge vessel.
Aspect 31. The method of any one of the foregoing aspects comprising converting plasma/erythrocyte interface position to the volume fraction occupied by red blood cells and plotted against time as a centrifuge sedimentation curve.
Aspect 32. The method of any one of the foregoing aspects, wherein a linear region of a sedimentation profile is used to extract a sedimentation rate.
Aspect 33. The method of any one of the foregoing aspects, further comprising deriving an estimate of the sedimentation rate linearly related to the Westergren ESR, the centrifuge-derived, hematocrit corrected data further corrected using the formula: Estimated Westergren ESR=10{circumflex over ( )}(((LOG(HCT corrected ESR)−LOG(a))/b)).
Aspect 34. The method of any one of the foregoing aspects, further comprising hematocrit-correcting and linearly-transforming Log(ESR) values to establish a linear graph of sedimentation rate.
Aspect 35. The method of any one of the foregoing aspects wherein the blood sample is whole blood.
Aspect 36. The method of any one of the foregoing aspects wherein the blood sample is an anti-coagulated sample.
Aspect 37. The method of any one of the foregoing aspects wherein the formed blood component is white blood cells.
Aspect 38. The method of any one of the foregoing aspects wherein the formed blood component is platelets.
Aspect 39. The method of any one of the foregoing aspects, further comprising determining white cell sedimentation rate after centrifugation has begun, wherein measuring white cell sedimentation rate characterizes at least one of the following regarding the white blood cells: cell density, shape, and aggregation state.
Aspect 40. A method comprising: collecting a plurality of images of formed blood component and plasma interface positions over time from an accelerated blood sample compaction process; performing image transformation on said plurality of images to transform images with curved interfaces into corrected images with straight line interfaces; establishing a time-related compaction curve based on interface positions in said corrected images, for at least one formed blood component in said blood sample.
Aspect 41. A method comprising: centrifuging a blood sample in a vessel for a period of time; collecting a plurality of images of formed blood component and plasma interface positions over time; performing image transformation on said images to transform images with curved interfaces into corrected images with straight line interfaces; establishing a time-related compaction curve based on interface positions in said corrected images, for at least one formed blood component in said blood sample after centrifuging has begun.
Aspect 42. A method comprising: using a programmable processor-controlled system to transfer at least a portion of a blood sample from a blood sample location into a centrifugation vessel; using a sample handling system under programmable processor control to transfer said vessel from a first addressable position to a centrifuge with a second addressable position; centrifuging the blood sample in the vessel for a period of time; collecting a plurality of images of formed blood component and plasma interface positions over time;
establishing a time-related compaction curve based on interface positions in said corrected images, for at least one formed blood component in said blood sample after centrifuging has begun.
Aspect 43. The method of any one of the foregoing aspects, wherein the centrifuge has a rotor with a diameter of about 15 cm or less.
Aspect 44. The method of any one of the foregoing aspects, wherein the centrifuge has a rotor with a diameter of about 10 cm or less.
Aspect 45. The method of any one of the foregoing aspects, wherein the centrifuge has a rotor when in motion circumscribes an area with a longest dimension of about 15 cm or less.
Aspect 46. The method of any one of the foregoing aspects, wherein the centrifuge has a rotor when in motion circumscribes an area with a longest dimension of about 10 cm or less.
Aspect 47. A method comprising: centrifuging a blood sample in a vessel for a period of time; varying centrifuging speed to establishing a linear compaction curve of at least one formed blood component over the period of time until compacting has completed; monitoring centrifuging speed profile for at least a portion of the time period; and determining blood component sedimentation rate based on the centrifuging speed profile.
Aspect 48. A method comprising: centrifuging a blood sample in a vessel for a period of time; collecting at least a first single image of formed blood component and plasma interface positions at an initial time; collecting at least a second single image of formed blood component and plasma interface positions at a second time while rate of sedimentation is still linear; calculating sedimentation rate for at least one formed blood component in said blood sample based on linear sedimentation rate calculated and a hematocrit correction factor.
Aspect 49. A device for use with a sample, the device comprising:
a centrifuge having a centrifuge vessel holder configured to allow for detection of blood component interface position in the vessel holder during centrifugation.
Aspect 50. The device of aspect 49, wherein the centrifuge has window to allow for visual observation of the centrifuge vessel holder during centrifugation.
Aspect 51. The device of aspect 49, wherein the centrifuge an illumination source to allow for detection of blood component interface position in the sample.
Aspect 52. A system comprising: a centrifuge having a centrifuge vessel holder configured to allow for detection of blood component interface position in the vessel holder in the vessel holder during centrifugation; a sample handling system for transporting a blood sample from a first location to a location on the centrifuge; and a processor programmed to record interface position during a least a portion of centrifugation.
This document contains material subject to copyright protection. The copyright owner (Applicant herein) has no objection to facsimile reproduction of the patent documents and disclosures, as they appear in the US Patent and Trademark Office patent file or records, but otherwise reserves all copyright rights whatsoever. The following notice shall apply: Copyright 2013 Theranos, Inc.
While the above is a complete description of the various embodiments of the present invention, it is possible to use various alternatives, modifications and equivalents. Therefore, the scope of the present invention should be determined not with reference to the above description but should, instead, be determined with reference to the appended claims, along with their full scope of equivalents. Any feature, whether preferred or not, may be combined with any other feature, whether preferred or not. The appended claims are not to be interpreted as including means-plus-function limitations, unless such a limitation is explicitly recited in a given claim using the phrase “means for.” It should be understood that as used in the description herein and throughout the claims that follow, the meaning of “a,” “an,” and “the” includes plural reference unless the context clearly dictates otherwise. Also, as used in the description herein and throughout the claims that follow, the meaning of “in” includes “in” and “on” unless the context clearly dictates otherwise. Finally, as used in the description herein and throughout the claims that follow, the meanings of “and” and “or” include both the conjunctive and disjunctive and may be used interchangeably unless the context expressly dictates otherwise. Thus, in contexts where the terms “and” or “or” are used, usage of such conjunctions do not exclude an “and/or” meaning unless the context expressly dictates otherwise.
Number | Name | Date | Kind |
---|---|---|---|
2982170 | Wyss | May 1961 | A |
3288019 | Blumenfeld | Nov 1966 | A |
3848796 | Bull | Nov 1974 | A |
4010893 | Smith et al. | Mar 1977 | A |
4197735 | Munzer et al. | Apr 1980 | A |
4558947 | Wardlaw | Dec 1985 | A |
4745279 | Karkar et al. | May 1988 | A |
4848900 | Kuo et al. | Jul 1989 | A |
5003488 | Hardy | Mar 1991 | A |
5086784 | Levine et al. | Feb 1992 | A |
5578269 | Yaremko et al. | Nov 1996 | A |
5731513 | Bull | Mar 1998 | A |
5844128 | Bull | Dec 1998 | A |
5914272 | Dufresne et al. | Jun 1999 | A |
6098451 | Bull | Aug 2000 | A |
6204066 | Wardlaw | Mar 2001 | B1 |
6336358 | Kishimori et al. | Jan 2002 | B1 |
6403328 | Clampitt | Jun 2002 | B1 |
6517475 | Brown et al. | Feb 2003 | B1 |
8862448 | Holmes et al. | Oct 2014 | B2 |
9464981 | Gibbons et al. | Oct 2016 | B2 |
20010044584 | Kensey | Nov 2001 | A1 |
20020012117 | Wardlaw | Jan 2002 | A1 |
20020032149 | Kensey | Mar 2002 | A1 |
20020061835 | Kensey | May 2002 | A1 |
20030078517 | Kensey | Apr 2003 | A1 |
20030113930 | Winkelman et al. | Jun 2003 | A1 |
20030205538 | Dorian et al. | Nov 2003 | A1 |
20040065143 | Husher | Apr 2004 | A1 |
20080213743 | Ellsworth et al. | Sep 2008 | A1 |
20110226045 | McQuillan | Sep 2011 | A1 |
20110293686 | Hyde | Dec 2011 | A1 |
20120015794 | Holmes et al. | Jan 2012 | A1 |
20130078733 | Holmes et al. | Mar 2013 | A1 |
20130203107 | Lalpuria et al. | Aug 2013 | A1 |
20140020457 | Dayel et al. | Jan 2014 | A1 |
20140315240 | Dayel et al. | Oct 2014 | A1 |
20150198515 | Dayel et al. | Jul 2015 | A1 |
20150233893 | Dayel et al. | Aug 2015 | A1 |
20160341650 | Dayel et al. | Nov 2016 | A1 |
20190170630 | Dayel et al. | Jun 2019 | A1 |
20210140866 | Dayel | May 2021 | A1 |
Number | Date | Country |
---|---|---|
2230222 | Sep 1998 | CA |
2259629 | Aug 1997 | CN |
1584591 | Feb 2005 | CN |
0864854 | Sep 1998 | EP |
1063513 | Dec 2000 | EP |
1063515 | Aug 2006 | EP |
2-075959 | Mar 1990 | JP |
1996039618 | Dec 1996 | WO |
2008021626 | Feb 2008 | WO |
2014015177 | Jan 2014 | WO |
Entry |
---|
Borawski J. et al. The hematocrit-corrected erythrocyte sedimentation rate can be useful in diagnosing inflammation in hemodialysis patients. Nephron, 2001, 89(4), p. 381-383 (abstract) [online]. Retrieved from PubMed, PMID:11721153. |
Bull et al. The Zeta Sedimentation Ratio, Blood, vol. 40, No. 4 (Oct. 1972). |
Bull. Is a Standard ESR Possible? Laboratory Medicine. vol. 6, No. 11, Nov. 1975. |
Harper College. Measuring the Volume of Liquids, 2004. |
Hematek, ESR Stat Plus Laser Kinetic Micro Sed Rate Analyzer, 2008. |
Lipowsky, et al. Hematocrit determination in small bore tubes from optical density measurements under while illumination, Microvascular Research 20, 51-70 (1980). |
Notice of Allowance dated Jan. 20, 2016 for U.S. Appl. No. 13/945,147. |
Notice of Allowance dated Jan. 25, 2018 for U.S. Appl. No. 14/598,561. |
Notice of Allowance dated Nov. 3, 2017 for U.S. Appl. No. 14/627,307. |
Notice of Allowance dated Dec. 29, 2014 for U.S. Appl. No. 14/319,644. |
Office Action dated Nov. 25, 2016 for U.S. Appl. No. 15/132,906. |
Office Action dated Mar. 13, 2017 for U.S. Appl. No. 14/627,307. |
Office Action dated May 31, 2017 for U.S. Appl. No. 14/598,561. |
Office Action dated Jun. 23, 2015 for U.S. Appl. No. 13/945,147. |
Office Action dated Aug. 25, 2017 for U.S. Appl. No. 15/132,906. |
Pawlotsky, et al. An Erythrocyte Sedimentation Rate Adjusted for the Hematocrit and Hemoglobin Concentalion, American Journal of Clinical Pathology, vol. 122, No. 5, Nov. 1, 2004, pp. 802-810. |
QBC Europe, 2008 distributor meeting technical support, 2008. |
Qin Renjia, Study of Erythrocyte Aggregation With Sedimentation Rate, (abstract) Chinese Journal of Hemorheology, vol. 6, No. 3 19961231. |
Saleem et al. Comparison of Zeta Sedimentation Ratio with Westergren Sedimentation Rate, Annals of Clinical and Laboratory Science, vol. 7, No. 4, Institute for Clinical Science. |
Schembri, et al. Centrifugation and capillarity integrated into a multiple analyte whole blood analyser, Journal of Automatic Chemistry, vol. 17, No. 3 (May-Jun. 1995), pp. 99-104. |
Shelat et al. Difference in Erythrocyte Sedimentation Rates Using the Westergren Method and a Centrifugation Method, American Journal of Clinical Pathology, vol. 103, No. 1, Jul. 1, 2008, pp. 127-130. |
The international search report and the written opinion dated Dec. 19, 2013 for Application No. PCT/US2013/051143. |
The international search report and the written opinion dated Jun. 4, 2015 for PCT/US2015/012537. |
Office Action dated Jan. 16, 2020 for U.S. Appl. No. 16/001,760. |
Number | Date | Country | |
---|---|---|---|
20190226967 A1 | Jul 2019 | US |
Number | Date | Country | |
---|---|---|---|
61673037 | Jul 2012 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 14598561 | Jan 2015 | US |
Child | 16259511 | US | |
Parent | PCT/US2013/051143 | Jul 2013 | US |
Child | 14598561 | US |