The present disclosure relates to the field of clinical analysis devices.
COVID-19 has had a profound effect on global health, and there is clinical evidence that COVID-19 is a contagious inflammatory disorder that can affect any organ system. The pathophysiology of this novel viral disease is complex and not fully understood. Why some patients infected with the virus remain asymptomatic while others succumb to organ failure remains to be explained.
At this point there remains to be found a specific treatment for COVID-19, but a variety of existing treatments are being applied with varying levels of success. This is especially evident in the ICU where the sickest victims are empirically treated with a cocktail of therapies, including anti-inflammatory and anti-thrombotic agents.
This therapeutic approach is justified by the belief that the intensity of the inflammatory response to COVID-19 infection may be the major factor responsible for organ failure, thus use of corticosteroids. Also, the viral infection is accompanied by pervasive vascular complications—including myocardial infarction, stroke and renal failure. Without being bound by any particular theory, it has been proposed that the cardiovascular and cerebrovascular systems can be compromised by a pervasive vasculitis that disposes the blood to intravascular thrombosis, thus use of anticoagulants.
It is not usual in medicine to prescribe a treatment for a serious physiological disorder without objectively monitoring the physiological consequences of the treatment. If insulin is administered for management of diabetes, we carefully follow the blood glucose levels, and adjust treatment accordingly. This level of care for diabetes is not a problem because our diagnostic technology is easy to use and provides immediate results. The same level of care should apply to the use of powerful anti-inflammatory and anticoagulant medications, and especially when administered to our most vulnerable COVID-19 patients in the ICU.
Unfortunately, testing for and titrating the concentration of inflammatory (also termed “reactive”) and coagulopathic proteins in the blood are expensive, inconvenient, and time-consuming clinical laboratory procedures. For this reason, standard laboratory practice may not be adequate to monitor the effects of powerful drugs on an unconscious patient whose life is being supported by a respirator. ICU clinicians need more timely information to guide their use of these potentially life-saving empirical treatments.
Thus, there is a need in the field for an ICU “bedside” or other point-of-care (POC) test, in particular one that can take a small sample of blood (e.g., from an existing IV port) to measure the effects of these empirical treatments. It is useful to know how the drugs modify reactive proteins and clotting factors that reflect the body's inflammatory reaction to the infection and response to treatment; it is also useful to analyze the effects of the infection on blood flow, blood clotting and vascular integrity. In addition, there is also a long-felt need for method and systems that utilize whole blood viscosity (WBV) as a diagnostic indicator.
Rapid Profile Viscometer
C-Reactive Protein (CRP), erythrocyte sedimentation rate (ESR), and plasma viscosity (PV) tests generate similar findings in inflammatory disorders. But CRP and ESR are expensive and time consuming and are best performed in the clinical lab. Also, CRP is very sensitive and can remain elevated after clinical remission, while ESR is an indirect measure that depends on RBC clumping to disclose elevated reactive proteins. CRP and ESR are not conducive to on-the-spot testing.
The disclosed rapid profile viscometer (which can be termed an “RPV”) can provide clinically important POC (bedside) data about the actual sources of hyperviscosity and reduced clotting time in a faster and more convenient test than either CRP or ESR. In less than a minute the RPV can provide a full profile of the whole blood viscosities that exist throughout the body, ranging from the aorta to the deep veins of the leg, from either a fresh or anticoagulated blood sample. The test gives a picture of blood flow conditions existing in various parts of the vasculature.
WBV and PV can be quickly performed bedside as many times as needed using fresh or anticoagulated blood samples. Equipment for such use can include, e.g., Vacutainer™ tubes, a small centrifuge (for PV), and the disclosed RPV and any other disposables. The test can use a 1-2 ml sample of plasma or whole blood and takes only seconds to perform. WBV can be measured to assess the flow state and likelihood of abnormal clotting in vital organs by determining viscosity and clotting time with the RPV Viscometer and Coagulometer functions.
The test can be repeated using the same disposable with the same (or additional) anticoagulated samples to confirm the findings. Also, the test can be repeated as frequently as needed to monitor the patient's response to treatment, and to immediately act on that data. Alternative methods for performing blood viscometry are costly and time-consuming, and require specially trained personnel in the clinical lab setting.
The disclosed RPV provides a quick and easy point of care (POC) or even bedside test that can help in titrating anti-inflammatory, anticoagulant, and intravascular plasma replacement therapies in the ICU hospitalized patient, and especially in those patients with COVID-19. Blood samples can be withdrawn, e.g., from a vein or an IV port using a syringe or Vacutainer system under applicable ICU protocol. Here is provided is a simple, safe technology that does not require expensive personnel to utilize.
In meeting the long-felt needs in the field, the present disclosure provides methods, comprising: encouraging one or more of (1) anticoagulated whole blood from a subject patient, (2) a plasma from the subject, and/or (3) fresh whole blood undergoing coagulation through a conduit into and/or out of an enclosed volume; monitoring one or more pressures related to the encouraging as a function of time so as to generate one or more sets of pressure vs. time data; and based on the one or more sets of pressure vs. time data, determining a viscosity (or other measure of resistance to flow) of the anticoagulated whole blood, the plasma, and/or the fresh whole blood during coagulation.
In one exemplary embodiment, the method can include first collecting blood or plasma sample in a Vacutainer or equivalent receptacle; drawing sample into an enclosed (sealed) space; and discharging sample back into the Vacutainer. A time-pressure curve is recorded for both the drawing up and discharging, and the data can be averaged so as to cancel out hydrostatic, cohesive/adhesive and diabatic errors (as well as creating a replication of the experiment to improve and assure accuracy/precision).
Also provided are apparatuses, comprising: a first conduit configured to communicate a fluid therein; a first enclosed volume, the first enclosed volume being in fluid communication with the first conduit; a transducer, the transducer configured to measure a pressure within the first enclosed volume related to fluid entry and/or fluid egress from the first enclosed volume; and a source of positive and/or negative pressure, the source being capable of fluid communication with the first enclosed volume, the source configured to encourage fluid from the first conduit into the first enclosed volume and/or to encourage fluid out of the first enclosed volume.
Also provided are methods, comprising: encouraging a fluid sample through a conduit into and/or out of a reservoir, the fluid sample optionally comprising one or more of (1) whole blood, (2) a blood plasma, or (3) whole blood during coagulation; monitoring a pressure within the reservoir related to the encouraging as a function of time so as to generate at least a first set of pressure vs. time data; and based at least in part on the first set of pressure vs. time data, determining one or more of a viscosity, a viscosity as a function of shear rate, or a flow resistance of the fluid sample.
Further provided is an apparatus, comprising: a conduit configured to communicate a fluid therein; a reservoir, the reservoir being in fluid communication with the conduit, the reservoir optionally being expandable; a transducer, the transducer configured to measure a pressure within the first reservoir related to fluid entry and/or fluid egress from the expandable reservoir; and a memory in communication with the transducer, the memory configured to record at least a first set of pressure vs. time data related to fluid entry and/or fluid egress from the expandable reservoir.
Also disclosed are methods, comprising effecting operation of an apparatus according to the present disclosure, e.g., according to any one of Aspects 14-20.
Also provided is a method, comprising: generating n records from a subject, n being a positive integer, and an n-th record comprising at least an n-th whole blood viscosity (WBV) that is collected at an n-th time and at a shear rate of from about 1/s to about 1000/s; and determining, by a trained algorithm that receives the n records, at least one of (1) a relation between at least two of the n records and (2) a clinical significance of one or more of the n records.
Further disclosed is a method, comprising: receiving, at a trained algorithm, first data including a first WBV of a subject collected at a first time and at a shear rate of from about 1/s to about 1000/s; and (1) determining, by the trained algorithm, a correlation between the first data and a first physiological state, the correlation being indicative of a physiological state of the subject, or (2) receiving, at the trained algorithm, second data including a first WBV of a subject collected at a second time and at a shear rate of from about 1/s to about 1000/s, and determining, by the trained algorithm, a correlation between the first data and the second data, the correlation being indicative of a current or future physiological state of the subject.
Additionally provided is a method, comprising: receiving, at a trained algorithm, at least (i) a first data including to a first WBV of a subject collected at a first time and at a shear rate of from about 1/s to about 1000/s and (ii) a second data including a second WBV of a subject collected at a second time and at a shear rate of from about 1/s to about 1000/s; determining, by the trained algorithm, a correlation between the first data and the second data, wherein the correlation is indicative of a current or future physiological state of the subject.
Further provided is a system, comprising a viscometer stage, the viscometer stage configured to receive a whole blood sample from a subject and determine a WBV of the sample at one or more shear rates of from 1/s to 1000/s; a processing stage, the processing stage configured to implement a trained algorithm that receives at least a first data from the subject, the first data including a WBV collected at a first time and at a shear rate of from about 1/s to about 1000/s, the trained algorithm configured to determine a correlation between the first data and a first physiological state, the correlation being indicative of a current or future physiological state of the subject, or (2) the processing stage configured to implement a trained algorithm that receives (i) a first data from a subject, the first data including a WBV collected at a first time and at a shear rate of from about 1/s to about 1000/s and (ii) a second data from the subject, the second data including a WBV collected at a second time and at a shear rate of from about 1/s to about 1000/s, the trained algorithm configured to determine a correlation between the first data and the second data, wherein the correlation is indicative of a current or future physiological state of the subject.
Also provided is a method of training an algorithm for detecting a current or future physiological state of a subject, the method comprising: inputting a plurality of datasets for each of a plurality of subjects into a model function, wherein the plurality of datasets include: a first dataset including a first WBV of a subject collected at a first time and at a shear rate of from about 1/s to about 1000/s; a second dataset including a second WBV of a subject collected at a second time and at a shear rate of from about 1/s to about 1000/s; and a third dataset including at least one or more physiological states of the subject; and calculating weighted values for each data type of the first dataset and the second dataset via the model function; and generating a trained model by assigning the weighted values to the data type of the first dataset and the second dataset.
In the drawings, which are not necessarily drawn to scale, like numerals can describe similar components in different views. Like numerals having different letter suffixes can represent different instances of similar components. The drawings illustrate generally, by way of example, but not by way of limitation, various aspects discussed in the present document. In the drawings:
The present disclosure may be understood more readily by reference to the following detailed description of desired embodiments and the examples included therein.
Unless otherwise defined, all technical and scientific terms used herein have the same meaning as commonly understood by one of ordinary skill in the art. In case of conflict, the present document, including definitions, will control. Preferred methods and materials are described below, although methods and materials similar or equivalent to those described herein can be used in practice or testing. All publications, patent applications, patents and other references mentioned herein are incorporated by reference in their entirety. The materials, methods, and examples disclosed herein are illustrative only and not intended to be limiting.
The singular forms “a,” “an,” and “the” include plural referents unless the context clearly dictates otherwise.
As used in the specification and in the claims, the term “comprising” can include the embodiments “consisting of” and “consisting essentially of” The terms “comprise(s),” “include(s),” “having,” “has,” “can,” “contain(s),” and variants thereof, as used herein, are intended to be open-ended transitional phrases, terms, or words that require the presence of the named ingredients/steps and permit the presence of other ingredients/steps. However, such description should be construed as also describing compositions or processes as “consisting of” and “consisting essentially of” the enumerated ingredients/steps, which allows the presence of only the named ingredients/steps, along with any impurities that might result therefrom, and excludes other ingredients/steps.
As used herein, the terms “about” and “at or about” mean that the amount or value in question can be the value designated by some other value approximately or about the same. It is generally understood, as used herein, that it is the nominal value indicated ±10% variation unless otherwise indicated or inferred. The term is intended to convey that similar values promote equivalent results or effects recited in the claims. That is, it is understood that amounts, sizes, formulations, parameters, and other quantities and characteristics are not and need not be exact, but can be approximate and/or larger or smaller, as desired, reflecting tolerances, conversion factors, rounding off, measurement error and the like, and other factors known to those of skill in the art. In general, an amount, size, formulation, parameter or other quantity or characteristic is “about” or “approximate” whether or not expressly stated to be such. It is understood that where “about” is used before a quantitative value, the parameter also includes the specific quantitative value itself, unless specifically stated otherwise.
Unless indicated to the contrary, the numerical values should be understood to include numerical values which are the same when reduced to the same number of significant figures and numerical values which differ from the stated value by less than the experimental error of conventional measurement technique of the type described in the present application to determine the value.
All ranges disclosed herein are inclusive of the recited endpoint and independently of the endpoints (e.g., “between 2 grams and 10 grams, and all the intermediate values includes 2 grams, 10 grams, and all intermediate values”). The endpoints of the ranges and any values disclosed herein are not limited to the precise range or value; they are sufficiently imprecise to include values approximating these ranges and/or values. All ranges are combinable.
As used herein, approximating language can be applied to modify any quantitative representation that can vary without resulting in a change in the basic function to which it is related. Accordingly, a value modified by a term or terms, such as “about” and “substantially,” may not be limited to the precise value specified, in some cases. In at least some instances, the approximating language can correspond to the precision of an instrument for measuring the value. The modifier “about” should also be considered as disclosing the range defined by the absolute values of the two endpoints. For example, the expression “from about 2 to about 4” also discloses the range “from 2 to 4.” The term “about” can refer to plus or minus 10% of the indicated number. For example, “about 10%” can indicate a range of 9% to 11%, and “about 1” can mean from 0.9-1.1. Other meanings of “about” can be apparent from the context, such as rounding off, so, for example “about 1” can also mean from 0.5 to 1.4. Further, the term “comprising” should be understood as having its open-ended meaning of “including,” but the term also includes the closed meaning of the term “consisting.” For example, a composition that comprises components A and B can be a composition that includes A, B, and other components, but can also be a composition made of A and B only. Any documents cited herein are incorporated by reference in their entireties for any and all purposes.
Provided here is, inter alia, a Rapid Profile Viscometer (RPV) that can measure the viscosity of whole blood and/or blood plasma at the bedside in a matter of seconds. The disclosed technology also generates a viscosity profile of non-Newtonian fluids such as whole blood. The viscosity of whole blood increases as flow rate decreases. Elevated blood viscosity predisposes the slow flow vessels (such as the deep veins of the leg) to intravascular clotting and embolism. This is the bane of the COVID-19 ICU patient.
The disclosed technology measures the blood viscosity through a range of flow rates found in the cardiovascular system and can do so over the full range of physiological flow rates and/or pressures. The disclosed technology can be performed on fresh or anticoagulated whole blood to predict the flow properties (e.g., viscosity) anywhere in the body from the aorta to the deep veins of the leg. The disclosed technology can provide a flow-rate dependent blood viscosity curve (viscosity profile) that helps the clinician predict and manage the patient's vulnerability to thrombosis and embolism.
First Order Decay is the physical principle that can be used to measure and/or estimate (1) Whole Blood Viscosity (WBV), (2) Plasma Viscosity (PV) and (3) blood coagulation with a Rapid Profile Viscometer (RPV). First Order Decay describes the relationship between pressure and time in the RPV, and as a consequence of First Order Decay, the slope of pressure plotted on a log scale vs. time plotted on a linear scale gives the relative viscosity of Newtonian fluids such as blood plasma (PV), while the instantaneous plot of the log of pressure vs. time provides the shear rate dependent viscosity of non-Newtonian fluids such as blood. Each point on the log pressure curve reflects the relative viscosity of the sample (e.g., whole blood vs the control) at that shear rate. One can measure the volumetric flow within the conduit; one can know the total air volume in a system (which is enclosed), which can include the volume of the reservoir, tubing, and other components. By measuring the change in pressure when the volume of the system is changed (e.g., via pulling up on a syringe of air, by adding 1 mL of air, and the like.).
An example relationship between pressure and time in an RPV is shown in
An example of first order decay is provided by
Plasma Viscosity
The disclosed technology can be applied to plasma obtained from spinning down a whole blood sample in a centrifuge. Plasma viscosity reflects the amount of reactive proteins in the blood. Thus, plasma viscosity is a measure of the inflammatory response, helpful in detecting and managing the excessive immune response that underlies much COVID-19 pathology. Plasma viscometry can reveal the presence of acute phase reactants; whole blood viscometry can reveal the effects of the acute phase reactants on intravascular flow, especially in slow flow vessels such as the deep veins of the leg.
Coagulometer
In addition, the disclosed technology can be used to objectively measure the physical manifestations of the blood clotting cascade and determine the clotting time by continually measuring the viscosity of a fresh blood sample during coagulation.
The test is performed by repeatedly measuring the viscosity of the same blood sample until coagulation prevents the passage of fluid between the sample container and reservoir. A blood plasma viscosity can use ˜1 mL of blood plasma supernatant created by spinning down a small sample of anticoagulated whole blood. The plasma viscosity test can take only 30 seconds to perform, and all components that contact the sample are disposable. Each coagulation test involves a series of viscosity measures (e.g., every 30 seconds) performed on a sample of fresh whole blood to record changes in viscosity during coagulation. The resulting viscosity curve plots the physical cascade of coagulation. In this way, a clinician can determine the coagulation condition of a patient and administer and/or adjust the patient's treatment accordingly.
The Rapid Profile Viscometer has been used to study efficacy of hemodilution for treatment of experimental stroke in dogs, and to study blood viscosity reducing medications for managing human peripheral arterial occlusive disorders such as intermittent claudication. A Point-Of-Care (bedside) version of the Rapid Profile Viscometer is a companion test in the critical care management of anti-inflammatory and anticoagulant treatments for hospitalized COVID-19 patients.
Additional Disclosure—COVID-19 Application
As explained elsewhere herein, infection with COVID-19 (or other pathogens, including other viruses) can result in severe inflammation and intravascular clotting (coagulopathy). In the case of inflammation, the virus can induce extreme reactivity of the immune system leading to production of large quantities of acute reactive proteins including fibrinogen (an important clotting factor), which enter the bloodstream and cause a highly elevated blood viscosity. Unlike blood plasma, whole blood viscosity is shear rate dependent (i.e., the slower the flow, the higher the viscosity) so the highest blood viscosity is usually in the deep veins of the legs where the blood flow is slow. Increased blood viscosity associated with diminished blood flow sets up ideal conditions for intravascular clotting.
As to the viral infection itself, the blood vessel wall has specific receptors for the corona virus that permit the virus to enter and injure the vessel lining (endotheliopathy). These injuries permit intravascular blood clots to form, especially in the slow flow blood vessels. The end result of these interacting inflammatory and coagulopathic pathways can be Venous Thromboembolism (VTE) and/or Disseminated Intravascular Coagulation (DIC), which can lead to widespread embolism. Thus, an elevated shear rate-dependent blood viscosity can generate a vicious cycle, in which the viscosity continually increases as the flow rate decreases, culminating in stasis and intravascular clotting. For these (and other reasons), it is clinically important to monitor blood viscosity, as a tendency toward a thickening of blood (i.e., increased viscosity with decreased flow rate) can result in thrombosis and/or embolism. VTE and DIC are the bane of the COVID-19 patient in the ICU. As a consequence, elevated blood viscosity must be detected before VTE and DIC take place, and appropriate preventive treatments/measures administered to mitigate the risk of embolization and irreversible injury to vital organs.
Pathophysiology
Existing tests for the markers of inflammation (e.g., C-reactive protein, CRP, and the erythrocyte sedimentation rate, ESR) are complex and time-consuming, and are also best performed by trained technicians in a laboratory setting. The disclosed technology provides the novel, non-obvious insight that blood viscosity can measure (or act as a proxy for) the amount of these reactive proteins, and blood viscosity is easily and quickly measured at the bedside using the RPV.
A major cause of Morbidity and Mortality in the ICU is the increase in DIC, venous thromboembolism (VTE) and microvascular thrombi from uncontrolled intravascular clotting and hyperviscosity of blood. At present, neither Whole Blood Viscosity (WBV) nor Plasma Viscosity (PV) are generally utilized in the diagnosis and treatment of COVID-19. For this reason, the disclosed point-of-care (“bedside”) WBV and PV test can help mitigate mortality from intravascular blood clotting by enabling timely detection and early management of coagulopathy in ICU patients before damage to vital organs.
It is thought that the morbidity and mortality resulting from COVID-19 is as much a consequence of the body's over-response to the virus as it is due to the tissue injury directly caused by the virus. The immune over-response to SARS-CoV-2 can cause acute respiratory distress syndrome (ARDS), Disseminated Intravascular Coagulation (DIC), “cytokine storm”, immuno-thrombosis, multiple organ failure, and death. Therefore, current empirical treatment for COVID-19 is directed to diminishing this excessive immune response. A convenient companion test is needed to monitor the efficacy of these therapeutic interventions on blood viscosity. The disclosed RPV devices and methods provide an objective measure.
Increased thrombogenicity also plays an important role in the pathophysiology of severe COVID-19 disease. Empirical treatment for COVID-19 depends on anticoagulants to decrease susceptibility to intravascular clotting. This approach may not be effective. In any event, the thrombogenicity is accentuated by elevated blood viscosity (due to excessive acute phase reactants produced as part of the immune over-response to the virus) and further compounded by direct viral injury to the vascular intima. A convenient companion test is needed to monitor the efficacy of the therapeutic interventions for coagulopathy. The RPV can also serve as a bedside Coagulometer.
COVID-19 Treatment
At present, venous thromboembolism (VTE) prevention and treatment with unfractionated heparin (UH) and low molecular weight heparin (LMWH) provides mixed results in treating thrombotic complications of COVID-19. Multiple trials are currently underway to determine the effectiveness of UH and LMWH in COVID-19 treatment. Alternative novel anticoagulants are also being studied.
Measures of whole blood viscosity (WBV) and plasma viscosity (PV) reflect elevation in acute phase reactants produced during COVID-19 disease. However, C-reactive protein (CRP) and fibrinogen levels are not significantly improved by the use of heparin. Also, elevated whole blood viscosity (WBV) predisposes the patient to VTE and microthrombosis despite heparin use. Further, elevated plasma viscosity (PV) and measures of disease severity, such as Sequential Organ Failure Assessment (SOFA) scores, are highly correlated. A number of studies have demonstrated that the amount of acute phase reactants in the blood correlates highly with morbidity and mortality in COVID-19.
SARS-CoV-2 infection of vascular endothelial cells leads to their destruction. This endotheliopathy leads, in turn, to excessive platelet activation predisposing the patient to the coagulative DIC—a serious complication of COVID-19 disease. Heparins do not address this excessive platelet activation.
Role of Viscometry in COVID-19
A study of 15 critically ill COVID-19 patients found that all had blood plasma viscosity (PV) levels 95% or more above the normal range (1.4-1.8 centipoise), and of the 4 patients with a PV of 3.5 centipoise or greater, all had venous thromboembolism (VTE). D-dimer levels are used to monitor the severity of the activation of the coagulation cascade, and elevated D-dimer levels in COVID-19 are associated with poor prognosis. The Rapid Profile Viscometer (RPV) can measure both PV and WBV. It can record the viscosity profile of a fresh or anticoagulated whole blood sample at bedside in less than 60 seconds and also serve as a bedside test to measure coagulation. Whole blood viscosity is superior to PV in that the viscosity profile better reflects the actual shear rate dependent blood flow conditions existing in vivo, especially the increased resistance to flow in the low flow vasculature (such as deep veins of the leg) where VTE is most prevalent.
Heparin may not be adequate to prevent the VTE and micro thrombotic complications of COVID-19. “Timing is everything”: the elevated D-dimer is from clot lysis (“clot breaking up”) which occurs later in the disease process than the increase in whole blood viscosity (WBV) and plasma viscosity (PV) which take place early on in the disease. Increased viscosity is caused by the elevated fibrinogen and other acute phase reactants prior to clot formation. If the elevated viscosity is detected early enough, it will be possible to employ preventive interventions.
Early and repeated measures of WBV and PV will be helpful in guiding treatment needed to suppress the coagulation cascade—use of alternative anticoagulants—apixaban, fondaparinux and GP IIb/IIIa inhibitors, or the use of plasma exchange (a classic modality for the treatment of hyperviscosity). These promising treatments are under investigation in ongoing clinical trials.
Point-of-care (POC) fresh whole blood viscometry (WBV) provides real-time information regarding the state of blood flow and the potential for in vivo intravascular clotting. These situations will benefit from immediate timely information to guide preventive measures and medical management of abnormal blood clotting.
POC blood plasma viscometry (PV) provides real time information regarding the level of acute phase reactants. These situations will also benefit from immediate timely information to guide medical management of severe inflammatory disorders.
POC measures of coagulation provide real-time information about the physical manifestation of the coagulation cascade in situations that benefit from immediate timely information to guide medical management of abnormal blood clotting.
Summary of Recent Findings
There is growing evidence that the flow properties of blood in the macro-circulation play a key role in the pathogenesis of COVID-19. In particular, the high volume of acute phase reactants (such as fibrinogen) entering the bloodstream because of the intense viral inflammation significantly increases blood viscosity, especially in the slow flow vasculature. Elevated low shear rate viscosity plus viral vasculitis can lead to hemostasis and intravascular clotting in the deep veins of the leg. Decreased blood flow combined with widespread viral injury to the vascular intima create conditions conducive to vascular microthrombosis, the bane of the COVID-19 ICU patient. A blood viscosity test can be the key to early detection of impending viral coagulopathy and create opportunity for preventive intervention and convenient monitoring of response to treatment.
While COVID-19, caused by the SARS-CoV-2 virus, usually presents with symptoms of viral upper respiratory infection such as fever, headache, dry cough, and muscle aches, much of the morbidity and mortality resulting from the illness is the consequence of a severe generalized inflammation and coagulopathy. There are many biological markers that are found in severe COVID-19 disease including leukocytosis, lymphopenia, and elevated C-reactive protein (CRP), ferritin, IL-6 (suggestive of severe inflammation,) D-dimer (suggestive of clot lysis) and LDH.
One study found that the incidence of symptomatic venous thromboembolism is approximately 6% in hospitalized COVID-19 patients but is approximately 24% in those requiring higher levels of care. Another study of patients with mild-to-moderate COVID-19 (using CT pulmonary angiography and CT venography) found asymptomatic venous thromboembolism in approximately 83% of those screened for clots. A study of over 300 post-mortem COVID-19 patients conducted by pathologists observed a 91% incidence of micro-thrombosis at autopsy.
Based on studies reported to date, it can be helpful to remove the excess volume of acute phase reactants and the excess coagulopathic proteins that contribute to increased blood viscosity and intravascular clotting. Therapeutic plasma exchange (TPE) is an accepted treatment for hyperviscosity syndromes. TPE is currently being studied as a method of treatment for hyperviscosity associated with severe COVID-19. As yet, they have not collected enough data from clinical trials to determine whether TPE can have outcome related benefits in severe COVID-19. However, early results show TPE reduces plasma viscosity and decreases IL-6, CRP, fibrinogen, D-dimer, and ferritin levels (biochemical factors associated with more severe COVID-19) and normalizes PaO2 and FiO2 levels (demonstrating an improvement in lung function). These results suggest that replacing COVID-19 affected plasma with healthy fresh frozen plasma can cause reductions in the markers of inflammation and coagulopathy that are associated with the morbidity and mortality seen in severe COVID-19. It is also possible that earlier intervention with TPE may prevent the severe manifestations. While CRP and D-dimer are biochemical markers that can show presence of severe COVID-19 disease, an elevation in whole blood or plasma viscosity is an early biophysical marker that can help to predict the severity of COVID-19 as well as be an early manifestation of the intense immune response (cytokine storm) and the coagulopathy (disseminated thromboembolism) that foretell a poor outcome in severe COVID-19.
As explained elsewhere herein, the disclosed RPV can help to identify patients at risk and measure treatment response of those patients who are likely to develop severe COVID-19. The RPV can distinguish patients who may benefit most from treatments such as early therapeutic anticoagulation, a variety of anti-inflammatory treatments, and/or therapeutic plasma exchange. Because much remains unanswered regarding if and at what point to implement these treatments, applying blood viscometry in COVID-19 can also play a role in developing the guidelines for these therapies. Possibly, the most important contribution of viscometry in COVID-19 will be the ability to predict the likely severity of inflammation and coagulopathy prior to irreversible organ damage. In contrast, usual markers such as elevated D-dimer levels inform the intensivist that the irreversible damage caused by coagulopathy (DIC) may already be present.
Recent studies suggest that intervention (such as prophylactic therapeutic anticoagulation or the monoclonal antibody drug directed at GM-CSF, lenzilumab) during moderate early disease, before cytokine storm, ARDS, and coagulopathy have set in, has been more effective at producing improved outcomes such as decreased overall mortality and survivability without ventilation suggesting the possibility that earlier preventive interventions may be of greater value as opposed to waiting to intervene once the disease has become severe. The latest research also indicates that (1) Whole blood viscosity on admission in COVID-19 is higher in fatal disease than in non-fatal disease; (2) microthrombosis and venous thromboembolism (VTE) are different types of clots and clinically distinct phenomena explaining the difference in the incidence of microthrombosis and VTE in other studies, and (3) a scoping review has found a high probability that therapeutic plasma exchange would be beneficial in treating the cytokine storm in severe disease. In a study investigating the effect of TPE on outcomes in severe COVID-19, the results of 43 ICU patients randomized to TPE plus standard treatment vs. 44 ICU patients with standard treatment only found that duration of mechanical ventilation, duration of ICU stay, and SOFA scores were significantly reduced in the TPE group. While the difference in day mortality was deemed non-significant, the results of 20.9% for TPE vs. 34.1% for no TPE were noteworthy and suggest the need for further investigation. Finally, another retrospective propensity-score matched study of 90 patients assigned to two equal groups found that 28 day survival was significantly higher in the TPE group (91.1%) vs. controls (61.5%), p<0.001 and that in the TPE group where TPE was performed before day 12 of hospitalization there was a 0% mortality. Thus, a POC test that can detect elevated volumes of acute phase reactants and indicators of coagulopathy earlier in the disease (such as whole blood viscosity) can help guide therapy (possibly TPE) that would decrease morbidity and improve survivability in COVID-19 disease.
In the management of COVID-19, time is of the essence. Early detection of coagulopathy with timely intervention can save lives and avoid post COVID invalidism. Current biochemical markers of inflammation and coagulopathy have not proven as helpful as required to defeat this agile and deceptive virus, and a test is needed that can serve as a “leading indicator” that can predict the likelihood of organ damage. For example, the D-dimer usually reveals the presence of coagulopathic injury after the fact, when disseminated clots and emboli are already undergoing lysis. The intensivist needs markers that predict the future, early indicators that provide time for preventive interventions that protect vital organs from injury.
In meeting these needs, the disclosed RPV is a practical tool for the intensivist to track the rheological and clotting characteristics of blood in the macro-vasculature, and to monitor the biophysical response to treatments for the hyperviscosity of blood associated with severe inflammation, and the intravascular clotting of blood associated with hypercoagulability. The RPV is a convenient lab or bedside whole blood viscometer, plasma viscometer, and/or coagulometer that can detect and track the biophysical changes that serve as a proxy for the early biochemical changes associated with severe inflammation and coagulopathy caused by COVID-19, among other pathogens.
Summary
RPV quickly measures the magnitude of the immune response—any increase in acute phase reactants increases both whole blood and plasma viscosity, but blood viscosity (unlike plasma viscosity) is flow rate dependent. Therefore, the Rapid Profile Viscometer provides information about the varying degrees of resistance to flow that blood experiences in various parts of the vasculature, as for example, highly elevated viscosity in slow flow vessels such as the deep veins of the leg where a combination of injury to the vascular endothelium from the virus plus near stasis flow due to elevated acute phase reactants can result in VTE and DIC. Other blood tests for inflammation are complex and time consuming. Therefore, C-Reactive Protein (CRP) and Erythrocyte Sedimentation Rate (ESR) are not convenient or adequately informative as POC tests. Whole Blood viscosity (WBV) and Plasma viscosity (PV) are POC tests that can measure the physical effects of reactive proteins on blood flow—and whole blood viscosity is especially easy to perform at the bedside (i.e., no centrifuge is required) to quickly (i.e., takes less than a minute) determine the viscosity profile to detect elevated low shear rate viscosity and monitor response to treatment. The disclosed technology can serve as, inter alia, a whole blood viscometer, a blood plasma viscometer, and a blood coagulometer.
As shown and explained herein, the disclosed technology can comprise a device that uses disposables that contain a blood sample which is loaded into a reservoir area within a reservoir. A reservoir can be a syringe that also serves as the pump (air pulse generator) that draws up a sample to generate a negative pressure curve and then expels the sample to generate a positive pressure curve. The positive and negative curves can be averaged to cancel out the hydrostatic, diabatic, cohesive/adhesive and atmospheric pressure errors that afflict most conventional capillary viscometers. It should be understood, however, that averaging is not a requirement, as a single pressure vs. time curve can suffice. Serial curves performed (e.g., every 30 seconds) with fresh whole blood samples can be used to measure coagulopathy (coagulometer function), and/or anticoagulated blood samples can be spun down and plasma viscosity measured. Thus, the disclosed technology provides multifunction devices.
The circulatory system is a vital component of good health. It serves as the means of transport for blood which allows for the absorption of nutrients and removal of waste. For this reason, the flow of blood has been a focus of research for centuries and its malfunction has been determined to be the root cause of many medical conditions. The properties of this bodily fluid are distinct due to its constituents and therefore one must fully understand the elements of blood in order to predict its rheological behavior.
The basis of our experiment revolves around the fundamentals of fluid dynamics. A fundamental concept that is used to characterize liquids in rheology is the resistance of a fluid to flow, which is known as viscosity. A French scientist in the early 19th century by the name of Jean Leonard Marie Poiseuille derived an equation which incorporated viscosity to describe the flow of a fluid in his attempt to understand blood flow. Poiseuille's law states:
Where Δp is the pressure difference between the ends of the conduit, μ is the dynamic viscosity, L is the length of the conduit, Q is the volumetric flow rate, and R is the conduit radius.
Therefore, by knowing the pressure difference along a given tube with known dimensions one can determine the viscosity of any given fluid flowing through that tube. When comparing viscosity versus shear rate, which is the rate at which fluid layers move past each other, one can notice that viscosity is not constant during blood flow.
The disclosed technology thus presents significant value to a number of stakeholders. To the medical community, the disclosed technology presents a device to diagnose elevated inflammation and clotting in the blood vessels of severely ill COVID-19 patients, as well as an opportunity for early detection and more effective management to save lives. To patients, the technology provides a non-painful test to aid doctors in diagnosis and treatment of what can be a life-or-death condition, which can in turn improve chances of survival in the ICU.
Figures
Method for eliminating hydrostatic, diabatic, atmospheric, adhesive, and cohesive errors. Generate a negative pulse. Generate a positive pulse. Take both time-pressure curves, slice equivalent T-P segments, flip one T-P segment, average resulting T-P segments to eliminate errors related to direction of flow. Permits development of a quick, simple, safe, portable, low cost and disposable precision viscometer suitable for bedside use (POC) as well as in the clinical laboratory.
For example, one can analyze (e.g., via selection) a portion of a curve that corresponds to pressures of 120 mm Hg to 0 mm Hg, from 110 mm Hg to 0 mm Hg, from 100 mm Hg to 0 mm Hg, from 90 mm Hg to 0 mm Hg, from 80 mm Hg to 0 mm Hg, from 70 mm Hg, from 60 mm Hg to 0 mm Hg, from 50 mm Hg to 0 mm Hg, from 40 mm Hg to 30 mm Hg, from 30 mm Hg to 0 mm Hg, from 20 mm Hg to 0 mm Hg, from 10 mm Hg to mm Hg, or even from 5 mm Hg to 0 mm Hg. All intermediate ranges can also be used, e.g., from 50 mm Hg to 10 mm Hg, from 40 mm Hg to 20 mm Hg, or even from 35 mm Hg to 30 mm Hg.
Arterial blood flow velocities can be, e.g., in the range of from, e.g., about 4.9 to about 19 cm/sec; venous blood flow can be in the range of from about 1.5 to 7.1 cm/sec. Taking into consideration the diameters of the blood vessel diameters, one can determine blood flow rates of 3.0-26 ml/min in arteries and 1.2-4.8 ml/min. Klarhofer M, et al., High-resolution blood flow velocity measurements in the human finger. Magn Reson Med. 2001 April; 45(4):716-9. doi: 10.1002/mrm.1096. PMID: 11284002. The foregoing values are illustrative only, and flow pressures and flow rate ranges for the body's blood vessels will be known to the person of ordinary skill in the art.
The viscosity profile curve represents a summary of expected resistance to flow throughout the subject's macrocirculation (blood vessels larger than capillaries, including veins, arteries, venules, arterioles, and the like. Certain illnesses are related to viscosity in the high flow vasculature such as the aorta, while other disorders are related to viscosity in the low flow vasculature such as the deep veins in the leg. In COVID-19 the major interest is the effect of virus induced damage to the lining of the blood vessels (vasculitis and endotheliopathy) that creates a nidus for clotting, and the large amounts of acute reactive proteins created by the intense inflammatory response to the viral infection that causes slowing (and possible interruptions of blood flow). The effect is near stasis conditions due to the highly elevated blood viscosity, especially in the slow flow vasculature. These two mutually facilitating factors (elevated viscosity and damage to the vessel lining) create a perfect physiological storm leading to a potentially lethal coagulopathy—disseminated intravascular coagulation (DIC) with embolization (clots breaking loose within the blood vessels). The resulting barrage of clots causes severe damage to the lungs (PE—pulmonary embolism) and other vital organs if the patient has a patent foramen ovale, a common birth defect. In the case of COVID-19, the low shear rate viscosity of whole blood reflects the amount of acute phase reactants (inflammatory proteins) present, and can predict the potential for intravascular clotting early enough to allow prophylactic intervention before development of disseminated intravascular coagulation (DIC). As shown, the GUI can then output viscosity profiles and relevant statistics.
Fluid 102 can be encouraged to travel through capillary 104 and into reservoir 106. This can be accomplished by, e.g., expanding the volume within reservoir 106. As but one example, if reservoir 106 is a syringe or other container with a moveable barrier, the volume within reservoir 106 can be increased so as to give rise to a relatively low pressure within reservoir 106, which pressure in turn encourages fluid 102 through capillary 104 and into reservoir 106.
Reservoir 106 can be in fluid communication with pressure transducer 110, e.g., via line 108 that places reservoir 106 into fluid communication with the pressure transducer. Reservoir 106 can also be open to the exterior environment, e.g., via a valve in fluid communication with line 108 or otherwise in fluid communication with reservoir 106. Priming volume 112 can be fluid communication with line 108 and also in fluid communication with reservoir 106. In this way, a pressure (e.g., a reduced pressure) within priming volume 112 can also motivate fluid into or out of reservoir 106. A valve (e.g., a two- or three-way valve) can be used at the intersection of priming volume 112 and line 108 to control the fluid communication between priming volume 112 and reservoir 106.
As shown in
As shown in
As an example (with reference to
Thus, the disclosed technology allows for rapid measurements of blood viscosity. The output of the disclosed technology can be pressure vs. time, flow rate vs. time, flow velocity vs. time, among other outputs. A user can compare data obtained with the present technology to baseline values, as well as comparing such data to expected or literature values, thereby allowing the user to assess a given patient's condition and to prescribe (or adjust) the patient's treatment as appropriate. A user can also evaluate a sample based on the overall time necessary for a sample to return to a baseline pressure after pressure is applied (or removed). Although monitoring the time to return to baseline pressure does not necessarily provide a quantitative viscosity measurement, the time to return can have clinical significance, as the time to return can provide a qualitative indication of viscosity (which can be correlated to disease and/or coagulation state), which qualitative indication can in turn be used to make treatment decisions for the patient.
The system 2900 can include one or more machine learning (ML) models 64. Some embodiments of models 64 can comprise a convolutional neural network (CNN). A CNN can comprise an input and an output layer, as well as multiple hidden layers. The hidden layers of a CNN typically comprise a series of convolutional layers that convolve with a multiplication or other dot product. The activation function can be a rectifier (RELU) layer, and is subsequently followed by additional convolutions such as pooling layers, fully connected layers and normalization layers, referred to as hidden layers because their inputs and outputs are masked by the activation function and final convolution.
The CNN computes an output value by applying a specific function to the input values coming from the receptive field in the previous layer. The function that is applied to the input values is determined by a vector of weights and a bias (typically real numbers). Learning, in a neural network, progresses by making iterative adjustments to these biases and weights. The vector of weights and the bias are called filters and represent particular features of the input (e.g., a particular shape).
Training component 32 of
For example, training a model can include: inputting a plurality of datasets for each of a plurality of subjects into a model function, wherein the plurality of datasets include: a first dataset including a first WBV of a subject collected at a first time and at a shear rate of from about 1/s to about 1000/s; a second dataset including a second WBV of a subject collected at a second time and at a shear rate of from about 1/s to about 1000/s; and a third dataset including at least one or more physiological states of the subject; and calculating weighted values for each data type of the first dataset and the second dataset via the model function; and generating a trained model by assigning the weighted values to the data type of the first dataset and the second dataset.
A model implementing a neural network can be trained using training data obtained by information component 30 from training data 62 storage/database. The training data can include many attributes of objects or other portions of a content item. For example, this training data obtained from prediction database 60 of
The validation set can be a subset of the training data, which is kept hidden from the model to test accuracy of the model. The test set can be a dataset, which is new to the model to test accuracy of the model. The training dataset used to train prediction models 64 can leverage, via inference component 34, a server or database for data storage and extraction purposes.
In some embodiments, training component 32 can be configured to obtain training data from any suitable source, via electronic storage 22, external resources 24 (e.g., which can include sensors), network 70, and/or user interface (UI) device(s) 18. The training data can comprise WBV data, and/or other discrete instances of clinical or sensed information.
In some embodiments, training component 32 can enable one or more prediction models 64 to be trained. The training of the neural networks can be performed via several iterations. For each training iteration, a classification prediction (e.g., output of a layer) of the neural network(s) can be determined and compared to the corresponding, known classification. As such, the neural networks can be configured to receive at least a portion of the training data as an input feature space. Once trained, the model(s) can be stored in database/storage 64 of prediction database 60, as shown in
Electronic storage 22 of
External resources 24 can include sources of information (e.g., databases, websites, and the like), external entities participating with system 10, one or more servers outside of system 10, a network, electronic storage, equipment related to Wi-Fi technology, equipment related to Bluetooth® technology, data entry devices, a power supply (e.g., battery powered or line-power connected, such as directly to 110 volts AC or indirectly via AC/DC conversion), a transmit/receive element (e.g., an antenna configured to transmit and/or receive wireless signals), a network interface controller (NIC), a display controller, a set of graphics processing units (GPUs), and/or other resources. In some implementations, some or all of the functionality attributed herein to external resources 24 can be provided by other components or resources included in system 10. Processor 20, external resources 24, UI device 18, electronic storage 22, a network, and/or other components of system 10 can be configured to communicate with each other via wired and/or wireless connections, such as a network (e.g., a local area network (LAN), the Internet, a wide area network (WAN), a radio access network (RAN), a public switched telephone network (PSTN), and the like), cellular technology (e.g., GSM, UMTS, LTE, 5G, and the like), Wi-Fi technology, another wireless communications link (e.g., radio frequency (RF), microwave, IR, ultraviolet (UV), visible light, cm wave, mm wave, and the like), a base station, and/or other resources.
UI device(s) 18 of system 10 can be configured to provide an interface between one or more users and system 10. UI devices 18 are configured to provide information to and/or receive information from the one or more users. UI devices 18 include a UI and/or other components. The UI can be and/or include a graphical UI (GUI) configured to present views and/or fields configured to receive entry and/or selection with respect to particular functionality of system 10, and/or provide and/or receive other information. In some embodiments, the UI of UI devices 18 can include a plurality of separate interfaces associated with processors 20 and/or other components of system 10. Examples of interface devices suitable for inclusion in UI device 18 include a touch screen, a keypad, touch sensitive and/or physical buttons, switches, a keyboard, knobs, levers, a display, speakers, a microphone, an indicator light, an audible alarm, a printer, and/or other interface devices. The present disclosure also contemplates that UI devices 18 include a removable storage interface. In this example, information can be loaded into UI devices 18 from removable storage (e.g., a smart card, a flash drive, a removable disk) that enables users to customize the implementation of UI devices 18.
In some embodiments, UI devices 18 are configured to provide a UI, processing capabilities, databases, and/or electronic storage to system 10. As such, UI devices 18 can include processors 20, electronic storage 22, external resources 24, and/or other components of system 10. In some embodiments, UI devices 18 are connected to a network (e.g., the Internet). In some embodiments, UI devices 18 do not include processor 20, electronic storage 22, external resources 24, and/or other components of system 10, but instead communicate with these components via dedicated lines, a bus, a switch, network, or other communication means. The communication can be wireless or wired. In some embodiments, UI devices 18 are laptops, desktop computers, smartphones, tablet computers, and/or other UI devices.
Data and content can be exchanged between the various components of the system 10 through a communication interface and communication paths using any one of a number of communications protocols. In one example, data can be exchanged employing a protocol used for communicating data across a packet-switched internetwork using, for example, the Internet Protocol Suite, also referred to as TCP/IP. The data and content can be delivered using datagrams (or packets) from the source host to the destination host solely based on their addresses. For this purpose the Internet Protocol (IP) defines addressing methods and structures for datagram encapsulation. Of course other protocols also can be used. Examples of an Internet protocol include Internet Protocol Version 4 (IPv4) and Internet Protocol Version 6 (IPv6).
In some embodiments, processor(s) 20 can form part (e.g., in a same or separate housing) of a user device, a consumer electronics device, a mobile phone, a smartphone, a personal data assistant, a digital tablet/pad computer, a wearable device (e.g., watch), augmented reality (AR) googles, virtual reality (VR) googles, a reflective display, a personal computer, a laptop computer, a notebook computer, a work station, a server, a high performance computer (HPC), a vehicle (e.g., embedded computer, such as in a dashboard or in front of a seated occupant of a car or plane), a game or entertainment system, a set-top-box, a monitor, a television (TV), a panel, a space craft, or any other device. In some embodiments, processor 20 is configured to provide information processing capabilities in system 10. Processor 20 can comprise one or more of a digital processor, an analog processor, a digital circuit designed to process information, an analog circuit designed to process information, a state machine, and/or other mechanisms for electronically processing information. Although processor 20 is shown in
As shown in
It should be appreciated that although components 30, 32, and 34 are illustrated in
Aspects
The following Aspects are exemplary only and do not serve to limit the scope of the present disclosure or the appended claims. Any part or parts of any Aspect can be combined with any part or parts of any other Aspect or Aspects.
Aspect 1. A method, comprising: encouraging a fluid sample through a conduit into and/or out of a reservoir, the fluid sample optionally comprising one or more of (1) whole blood, (2) a blood plasma, or (3) whole blood during coagulation; monitoring a pressure within the reservoir related to the encouraging as a function of time so as to generate at least a first set of pressure vs. time data; and based at least in part on the first set of pressure vs. time data, determining one or more of a viscosity, a viscosity as a function of shear rate, or a flow resistance of the fluid sample.
As described, the disclosed technology is useful on whole blood (which can be uncoagulated), plasma, and/or whole blood during coagulation. The pressure within the reservoir can be modulated by, e.g., increasing the volume within the reservoir, e.g., if the reservoir is a syringe by pulling up on the plunger. Alternatively, the volume within the reservoir can be decreased, e.g., if the reservoir is a syringe by pressing down on the plunger. As another (non-limiting) alternative, one can exert a negative and/or positive pressure with an element (e.g., a syringe) that is in fluid communication with the reservoir.
As described herein, the method can include generating at least a first set of pressure vs. time data. Such data can be further processed (as described elsewhere herein) to provide pressure vs. flow data, viscosity data, and the like. The pressure vs. time data (or other data developed at least in part on said pressure vs. time data) can be displayed, recorded, transmitted, or any combination thereof. As explained one can determine one or more of a viscosity, a viscosity as a function of shear rate, or a flow resistance of the fluid sample. Also as explained, one can also determine a time required for a sample to achieve and/or return to a given pressure, which time (which can be termed a “recovery time” in some instances) can be indicative of a viscosity of the sample under study. For example, one could compare the recovery time of a given sample to the recovery time associated with a sample that is indicative of an acute blood clotting state. This can be accomplished by, e.g., encouraging a sample from a sample container into a reservoir (as shown in
Aspect 2. The method of Aspect 1, further comprising relating the one or more of a viscosity, a viscosity as a function of shear rate, or a flow resistance of the fluid sample to a physiologic state of a subject. Such a state can be, e.g., a disease state.
Aspect 3. The method of any one of Aspects 1-2, further comprising administering a treatment or withdrawing a treatment in response to the physiologic state of the subject. Examples of such treatments include, e.g., anti-inflammatories, blood thinners, and the like.
Aspect 4. The method of any one of Aspects 1-3, comprising determining, based at least in part on the first set of pressure vs. time data, a viscosity of the fluid sample. The viscosity can be compared to a standard or a range of standards, the result of said comparison being then used to assess a condition of the patient.
Aspect 5. The method of any one of Aspects 1-4, comprising determining, based at least in part on the first set of pressure vs. time data, a viscosity as a function of shear rate of the fluid sample.
Aspect 6. The method of any one of Aspects 1-5, comprising determining, based at least in part on the first set of pressure vs. time data, a flow resistance of the fluid sample.
Aspect 7. The method of any one of Aspects 1-6, wherein the fluid sample is encouraged into the reservoir by expanding the reservoir. The fluid sample can also, in some embodiments, be encouraged into the reservoir by applying a pressure to a sample container in which the fluid sample is disposed. It should be understood that the reservoir can be placed into fluid communication with the environment exterior to the reservoir (or another volume, which volume can be enclosed) at any point, e.g., to equalize pressures or even to vent excess pressure. Such fluid communication can be done before fluid is encouraged into or out of the reservoir, during fluid encouragement into or out of the reservoir, or even after fluid is encouraged into or out of the reservoir. As an example, a reservoir can be opened and then closed to the environment exterior to the reservoir to as to set a baseline pressure. Sample can be encouraged into the reservoir, with the related development of a pressure vs. time curve related to that sample encouragement. After the pressure within the reservoir stabilizes, the reservoir can be vented to the exterior and then sealed. After re-sealing, fluid sample within the reservoir can then be encouraged out of the reservoir, with a further development of a pressure vs. time curve related to the encouragement.
Aspect 8. The method of any one of Aspects 1-7, wherein the fluid sample is encouraged into the reservoir by applying a negative pressure from a source in fluid communication with the reservoir.
Aspect 9. The method of any one of Aspects 1-8, further comprising reducing a pressure within the reservoir by expanding a priming volume in fluid communication with the reservoir. Such a priming volume can be, e.g., a syringe. The priming volume can be manually controllable, but can also be automated or machine-controlled.
Aspect 10. The method of Aspect 9, further comprising interrupting fluid communication between the priming volume and the reservoir.
Aspect 11. The method of any one of Aspects 1-10, further comprising encouraging the fluid sample out of the reservoir following encouraging the fluid sample into the reservoir and monitoring a pressure within the reservoir related to the encouraging out of the reservoir as a function of time so as to generate a second set of pressure vs. time data.
Aspect 12. The method of Aspect 11, further comprising determining the viscosity, the viscosity as a function of shear rate, or the flow resistance of the fluid sample based at least in part on the second set of pressure vs. time data.
Aspect 13. The method of Aspect 12, wherein the viscosity, the viscosity as a function of shear rate, or the flow resistance of the fluid sample is based at least in part on an average that comprises the first set of pressure vs. time data and the second set of pressure vs. time data. The development of such an average is described elsewhere herein, and is also shown in
Aspect 14. An apparatus, comprising: a conduit configured to communicate a fluid therein; a reservoir, the reservoir being in fluid communication with the conduit, the reservoir optionally being expandable; a transducer, the transducer configured to measure a pressure within the first reservoir related to fluid entry and/or fluid egress from the expandable reservoir; and a memory in communication with the transducer, the memory configured to record at least a first set of pressure vs. time data related to fluid entry and/or fluid egress from the expandable reservoir.
A conduit can be, e.g., a syringe needle, although this is not a requirement. Other conduits include, e.g., capillaries, cannulas, and the like.
Aspect 15. The apparatus of Aspect 14, further comprising a processor in communication with the transducer, the processor being configured to execute instructions to relate at least the first set of pressure vs. time data of the transducer to a viscosity of a fluid communicated into and/or out of the first reservoir to any one or more of a viscosity, a viscosity as a function of shear rate, or a flow resistance of a fluid sample communicated within the conduit. The processor can be embodied within a computer, a mobile computing device, and the like.
Aspect 16. The apparatus of Aspect 15, the processor being configured to execute instructions to relate an average that comprises least the first set of pressure vs. time data of the transducer and a second set of pressure vs. time data of the transducer to a viscosity of a fluid communicated into and/or out of the first reservoir to any one or more of a viscosity, a viscosity as a function of shear rate, or a flow resistance of the fluid. The instructions can also instruct to transform, invert, or otherwise “flip” a data set so that the data set shares a basis with and is thus average-able with another set of data. Such a procedure is described elsewhere herein and also shown in
Aspect 17. The apparatus of any one of Aspects 14-16, further comprising a priming volume, the priming volume in interruptible fluid communication with the reservoir. A priming volume can be, e.g., a syringe or other volume that is expandable or contractable. This is not a requirement, as the priming volume can be of fixed volume and then be pressurized and/or depressurized.
Aspect 18. The apparatus of Aspect 17, wherein the priming volume is operable to draw a fluid through the conduit into the reservoir. One such arrangement is provided in
Aspect 19. The apparatus of Aspect 17, wherein the priming volume is a syringe.
Aspect 20. The apparatus of any one of Aspects 14-19, wherein the reservoir is a syringe. As described elsewhere herein, the reservoir can be of variable volume, but this is not a requirement, as the reservoir can be of fixed volume.
Aspect 21. A method, comprising effecting operation of an apparatus according to any one of Aspects 14-20.
Aspect 22. The method of Aspect 21, wherein the operation is effected on a subject's sample of whole blood, plasma, or coagulating blood.
Aspect 23. The method of Aspect 22, further comprising classifying a disease state of the subject based on the operation of the apparatus.
Aspect 24. The method of Aspect 23, further comprising recommending a treatment for the disease state of the subject. The disclosed methods can also include administering a treatment in accordance with a disease state of the subject identified by analysis of pressure vs. time information or on data related to such a pressure vs. time information.
Aspect 25. The method of any one of Aspects 21-24, wherein the operation is automated.
Aspect 26. A system, comprising: a viscometer, the viscometer configured to communicate a fluid sample to and/or from a reservoir and develop pressure vs. time data of the reservoir related to the communication of said fluid; and a processor, the processor configured to relate the pressure vs. time data to a viscosity of the fluid sample. The pressure vs. time data can also be related to a flow resistance of the fluid sample, which fluid resistance can be developed based on the dimensions of a conduit through which the fluid sample is communicated.
Aspect 27. A method, comprising: encouraging a fluid sample into and/or out of a reservoir, the fluid sample optionally comprising one or more of (1) whole blood, (2) a blood plasma, or (3) whole blood during coagulation; monitoring a pressure within the reservoir related to the encouraging as a function of time so as to generate at least a first set of pressure vs. time data; and based at least in part on the first set of pressure vs. time data, determining one or more of a viscosity, a viscosity as a function of shear rate, or a flow resistance of the fluid sample. The methods can further include any one or more of the features recited in Aspects 2-13.
It should be understood that in some embodiments, the conduit and the reservoir can be one and the same, i.e., the fluid sample is communicated directly into a conduit/reservoir, with the pressure within the conduit/reservoir being monitored. For example, as shown in
As shown in
Fluid 102 can be encouraged to capillary/reservoir 104/106. This can be accomplished by, e.g., expanding the volume within capillary/reservoir 104/106. As but one example, if capillary/reservoir 104/106 is a syringe or other container with a moveable barrier, the volume within capillary/reservoir 104/106 can be increased so as to give rise to a relatively low pressure within capillary/reservoir 104/106, which pressure in turn encourages fluid 102 to capillary/reservoir 104/106.
Capillary/reservoir 104/106 can be in fluid communication with pressure transducer 110, e.g., via line 108 that places reservoir 106 into fluid communication with the pressure transducer. Capillary/reservoir 104/106 can also be open to the exterior environment, e.g., via a valve in fluid communication with line 108 or otherwise in fluid communication with reservoir 106. Priming volume 112 can be fluid communication with line 108 and also in fluid communication with capillary/reservoir 104/106. In this way, a pressure (e.g., a reduced pressure) within priming volume 112 can also motivate fluid into or out of capillary/reservoir 104/106. A valve (e.g., a two- or three-way valve) can be used at the intersection of priming volume 112 and line 108 to control the fluid communication between priming volume 112 and capillary/reservoir 104/106.
As shown in
Whole Blood Viscosity (WBV)
Although the topics of whole blood viscosity (WBV) and plasma viscosity (PV) have been a part of the medical and scientific communities, little has been done to date to demonstrate the importance of knowing a patient's whole blood (or plasma) viscosity. Thus, there has been little clinical interest in measuring or modifying the flow properties of blood, except in the few obvious medical conditions such as polycythemia vera, multiple myeloma (and some analogous conditions where there is a substantial increase in the number of blood cells and/or plasma protein levels), or severe elevations in plasma lipid levels. These disorders can lead to a symptomatic Hyperviscosity Syndrome for which therapeutic reduction of blood viscosity serves as the foundational intervention to reduce symptoms. Thus, one can surmise that maintaining normal whole blood viscosity may be required to assure overall health, akin to the importance of maintaining well-regulated blood pressure to promote cardiovascular health.
For some decades it has been recognized that the health of the lining of the blood vessels (the endothelium or endothelial cells) is essential in order to maintain good cardiovascular health. The endothelial cells are virtually everywhere in the human body, and no living cell of any tissue type resides far from an endothelial cell because the capillaries that channel the blood and deliver oxygen and nutrients and remove carbon dioxide and cellular waste products are primarily composed of endothelial cells. Thus, endothelial cells communicate their state (whether optimally healthy or dysfunctional) to every other cell in the body. Acutely elevated whole blood viscosity (WBV) is a pathogen-agnostic sign of inflammatory illness. WBV reflects the sum of many interacting physiological phenomena, and thus knowledge of WBV values alone may have clinical utility in a pathogen-agnostic manner.
Acute inflammation caused by any pathogen stimulates the production of “acute phase reactants” (APR), which are high molecular weight immune and pro-clotting proteins like fibrinogen, that enter the bloodstream and increase blood viscosity. The increased presence of APR proteins plus the resulting increase in blood viscosity is the hematological hallmark of inflammatory disease. Therefore, the degree of acute elevation in whole blood viscosity can serve as a proxy for both: (1) the volume of the APR proteins, and (2) the relative severity of the inflammation.
It has been determined that a factor needed to maintain healthy endothelial function is the “vessel wall shear stress” produced by the directional flow of blood in the blood vessels. But vessel wall shear stress produced by flowing blood is highly dependent upon the blood's viscosity, and both too-high and too-low shear stress along the blood vessels' lining impair endothelial health.
Optimal vessel wall shear stress from ideally viscous whole blood helps to maintain the endothelium in its normal (i.e., anti-oxidant, anti-inflammatory, anticoagulant, and insulin-sensitive) state. When whole blood viscosity is not ideal, the beneficial effect of the blood flow on the vessel wall is impaired, moving the endothelium in the direction of a pro-oxidant, pro-inflammatory, pro-coagulant, and insulin-resistant state. The majority of illnesses that modern humans experience display some direct or indirect relation to impaired endothelial function, and abnormal whole blood viscosity is an important contributor to “endothelial dysfunction”. Death can result from a progressive hyperinflammatory vascular dysfunction that must be detected and remediated early if treatment is to succeed.
To place this into healthcare perspective using just one disease, one quarter of all money expended on healthcare in the United States each year is spent on care for diabetes. Although endothelial dysfunction is a component of all types of diabetes, one can focus solely on Type 2 Diabetes. Ninety-five percent of diabetics have Type 2 Diabetes, accounting for more than $225 billion annually spent on their healthcare, in addition to the more than $80 billion that results from lost productivity. However, at this point in time, the clinical value of knowing a Type 2 Diabetic patient's whole blood viscosity has yet to be explored.
The current state of viscometers contribute to the lack of knowledge relating to the value of knowing a patient's whole blood viscosity. Clinical scientists need a convenient device, much like a sphygmomanometer, to assess blood viscosity. It must be accurate, inexpensive, quick, and easy to use. Such a clinical tool would facilitate the study of blood viscosity as a potential vital sign, like blood pressure, in evaluating health status.
There are generally two types of viscometers in use in hospital clinical laboratories, neither of which is optimal for conducting a clinical trial to determine the clinical utility of knowing a patient's whole blood viscosity. (a) The first type is simple U-shaped capillary tube viscometers that are comparatively inexpensive devices, but imprecise, slow and labor-intensive tests. The components are non-disposable and require cleaning after each test making them less than optimal if the lab needs to perform numerous whole blood viscosity tests per day. (b) The second type is highly precise but expensive rotational instruments that are also time-consuming with non-disposable components that require preparation and cleaning and are thus not ideal for conducting a clinical trial.
When it comes to demonstrating the value of knowing whole blood viscosity in a very serious and potentially fatal disorder like sepsis, there are very few clinical trials in the literature that have a large enough patient sample size to show a significant effect. There are plenty of sepsis-prone patients, but no truly convenient clinical viscometer.
Studies that consider how whole blood viscosity changes over the course of an acute or chronic illness (for instance during an inpatient hospitalization for incipient sepsis or outpatient management of diabetes) have not been the focus of substantial clinical studies, despite the fact that blood viscosity can and does change over time in both incipient sepsis and diabetes (e.g., increasing WBV may indicate the development of sepsis vs. decreasing WBV may reflect improvement in response to treatment, or vice versa). These changes may reflect how the disease and/or the treatment are affecting viscosity, and how the resulting changes in blood viscosity are in turn affecting the body's homeostasis (e.g., altering endothelial health). In other words, whole blood viscosity may serve both as (1) a proxy for the myriad of disease markers in the blood (i.e., hemorheological indicators of the presence of a disease state) and (2) the contribution of blood viscosity per se as an independent source of disordered vascular function (e.g., directly causing decreased tissue perfusion and endothelial dysfunction). The body has mechanisms to sense and regulate WBV—similar to regulation of blood pressure, though not with the same rapid response.
Application of Machine Learning to WBV
Within the category of unexplored opportunities for whole blood viscosity research is the absence of clinical trials that consider patient genetic and non-genetic individuality, namely a consideration of the meaning of the numerical value of whole blood viscosity in a particular patient.
Every patient is unique, even if they are presenting to the hospital for admission with the same diagnosis, for instance, COVID-19. Of all the many ways in which the value of whole blood viscosity has been studied to establish how knowledge of its value could be of clinical utility, it may not be that a given number in any specific disease is a perfect or even adequate warning sign or red flag to guide more immediate intervention, without additional information. Hence, there is a need for an approach that allows a physician to have a tool that will not merely provide a value of the whole blood viscosity that results from using the disclosed viscometer devices but also to have a machine learning (ML) algorithm that follows not only changes in whole blood viscosity at each blood draw but takes into consideration possibly relevant clinical and lab factors that are unique to this particular patient. Such data points can include (but are be limited to) age, gender, past medical and surgical history, pre-existing conditions, medications currently being taken, scores of social stress (such as a deprivation index measure), daily lab values that are produced for that patient in addition to some lab values which may not be routinely performed but which have shown some association with worsening disease, types of inpatient treatment being received (supplemental oxygen, mechanical ventilation, as examples), and a variety of different outcomes upon discharge from the hospital. The ML algorithm can have a two-fold purpose:
(1) To provide greater context to the real meaning of a given value of whole blood viscosity unique to any individual patient in their current state of health/disease.
(2) To be able to consider both whole blood viscosity, and the total confluence of all studied factors with the goal of turning the RPV from being simply an inexpensive, accurate, and portable capillary viscometer into a “early warning radar” for physicians so that they can be guided to make appropriate interventions earlier and more effectively in the hope of saving lives.
The ML algorithm can be a component of a device. The ML algorithm can also be applied to a clinical trial to make the relevant associations that can predict/detect, for instance, the onset of sepsis at the earliest possible moment in a hospitalized patient being treated for an infection. The ML algorithm can do this in a “pathogen-agnostic manner,” meaning that the RPV with ML can ultimately (after being trained on the data it receives in a clinical trial) serve as a sort of “early warning radar” for a condition (for example, for the development of sepsis) even in the absence of knowing the causal organism implicated as the source of the condition.
As but one example, the ML algorithm can be trained on data that includes WBV information as well as other information of clinical value (for example, age, gender, medical history, surgical history, social history, a pending medical condition, medication, and/or physiological condition). In this way, the ML algorithm can assess a patient's current condition based on WBV information and other clinical information. The ML algorithm can also predict a patient's future condition based on WBV information and other clinical information and/or based on changes in the WBV information and other clinical information.
After the pattern related to the confluence of factors in addition to whole blood viscosity has been determined for one condition, the device can refine the nature of the pattern during its ongoing use for that medical condition. Each time a new medical condition is determined through a clinical trial to demonstrate a pattern of factors in association with the numerical value (or range of values) of whole blood viscosity that predicts onset, severity, or progression, that information can be added to the software of the device so that it may then function with respect to that new medical condition, again with ongoing and continual refinement as it is used more and more for that condition.
Integration of ML with RPV
An RPV integrated with ML is a quick, low-cost, safe, and simple host-based diagnostic test for (a) early detection of acutely elevated WBV, and (b) holistically (i.e., by using ML to augment and amplify RPV findings) determining the relative severity of inflammation and predicting the potential mortal risk to the patient. In daily use with hospitalized patients as a pathogen-agnostic test (i.e. RPV with ML does not identify the specific agent that is causing the inflammation, but instead measures the intensity of the body's reaction to that causal agent), the RPV serves as a host-based diagnostic test that will facilitate early detection of the patient's response to the disease state (e.g. by detecting the acute increase in WBV) in advance of the more severe and potentially deadly manifestations of sepsis and sepsis-like disorders (e.g., COVID-19.)
In these infectious disease disorders the WBV is usually initially acutely increased, but should the infection become overwhelming, the relationship between WBV and disease severity can become inverted as the usual markers of inflammation become dissipated. For example, with severe inflammation, the APR proteins and the hematocrit can decrease due to inflammatory coagulopathy, as the fibrinogen and red blood cells become sequestered as intravascular clots. Under such circumstances, the WBV may evidence an acute paradoxical reduction to “near normal” levels despite severe vascular endothelial injury and disseminated intravascular clotting (DIC) and embolization severely impairing vital organ tissue perfusion. A protocol calling for serial WBV tests each day with concurrent ML can be used to detect and respond quickly to this deadly, but potentially misleading finding.
An RPV integrated with ML can detect and interpret paradoxical changes in WBV based on collateral lab and clinical findings and thus facilitate early intervention before irreversible injury to vital organs proves deadly or creates chronic disability. Incorporating an ML component into the RPV confers an advanced clinical “intelligence” to the device that can recognize when a severely ill patient has crossed the threshold between elevated WBV and paradoxical decreased WBV, by considering a diversity of concurrent laboratory and clinical factors and other data points. As currently designed, the RPV with ML will quickly learn to detect a patient whose illness is progressing rapidly, rather than being fooled by a seemingly “improving” WBV measurement (as would be the case with a clinical viscometer lacking ML).
With particular regard to sepsis, no alternative diagnostic or prognostic tests currently exist that effectively detect the earliest manifestations of sepsis or sepsis-like disorders, and humans do not possess the cognitive capacity for managing the large data sets that may confer the context, meaning, and value of knowing a patient's WBV in the setting of a worsening infectious disease or inflammatory syndrome.
By the time the presence of sepsis is discovered, therapeutic intervention is often of little benefit. Of the more than 1.5 million patients in the US who develop some form of sepsis each year, more than 250,000 will perish. Early detection and intervention are required. The prognosis for unrecognized sepsis is bleak. The mortality rate of those diagnosed with severe sepsis is approximately 50% and survivors experience a high rate of physical & cognitive disability and persistently lowered health status. Thus, a test that can detect patients at high risk for developing sepsis, or who are already in the earliest stages of sepsis, informs clinicians when and in whom to institute preventive measures or early interventions that will save lives.
Additional Aspects
The following Additional Aspects are illustrative only and do not limit the scope of the present disclosure or the appended claims. Any part or parts of any one or more Additional Aspects can be combined with any part or parts of any one or more other Aspects or Additional Aspects.
Aspect 28. A method, comprising: generating n records from a subject, n being a positive integer, and an n-th record comprising at least an n-th whole blood viscosity (WBV) that is collected at an n-th time and at a shear rate of from about 1/s to about 1000/s; and determining, by a trained algorithm that receives the n records, at least one of (1) a relation between at least two of the n records and (2) a clinical significance of one or more of the n records.
Aspect 29. The method of Aspect 29, wherein an n-th record further comprises a clinical information of the subject, the clinical information being information other than WBV.
Aspect 30. The method of Aspect 29, wherein the clinical information comprises any one or more of age, gender, medical history, surgical history, social history, a pending medical condition, or a medication.
Aspect 31. The method of Aspect 29, wherein an n-th record and an (n+1)-th record each include respective clinical information.
Aspect 32. The method of Aspect 31, wherein the clinical information of the n-th record is representative of that clinical information of the subject at the n-th time and wherein the clinical information of the (n+1)th record is representative of that clinical information of the subject at the (n+1)th time.
Aspect 33. The method of any one of Aspects 28-32, wherein the relation is indicative of a current or future physiological state of the subject.
Aspect 34. The method of any one of Aspects 28-33, wherein the determining comprises any one or more of (1) comparing a data of an n-th record to a threshold, and (2) comparing a difference between a data of an n-th record and a corresponding data of an (n+1)-th record to a threshold.
Aspect 35. The method of Aspect 28, wherein an n-th time and an (n+1)-th time are separated by less than about 12 hours, optionally separated by less than about 6 hours.
Aspect 36. A method, comprising: receiving, at a trained algorithm, first data including a first WBV of a subject collected at a first time and at a shear rate of from about 1/s to about 1000/s; and (1) determining, by the trained algorithm, a correlation between the first data and a first physiological state, the correlation being indicative of a physiological state of the subject, or (2) receiving, at the trained algorithm, second data including a first WBV of a subject collected at a second time and at a shear rate of from about 1/s to about 1000/s, and determining, by the trained algorithm, a correlation between the first data and the second data, the correlation being indicative of a current or future physiological state of the subject.
Aspect 37. The method of Aspect 36, wherein the first data further comprises a clinical information of the subject, the clinical information being other than WBV, and the clinical information optionally representing any one or more of age, gender, medical history, surgical history, social history, a pending medical condition, or a medication.
Aspect 38. The method of any one of Aspects 36-37, wherein the physiological state is a septic state.
Aspect 39. The method of any one of Aspects 36-38, wherein the second data further comprises a clinical information of the subject, the clinical information being other than WBV, and the clinical information optionally comprising any one or more of age, gender, medical history, surgical history, social history, a pending medical condition, or a medication.
Aspect 40. The method of any one of Aspects 36-39, further comprising training the algorithm.
Aspect 41. A method, comprising: receiving, at a trained algorithm, at least (i) a first data including to a first WBV of a subject collected at a first time and at a shear rate of from about 1/s to about 1000/s and (ii) a second data including a second WBV of a subject collected at a second time and at a shear rate of from about 1/s to about 1000/s; determining, by the trained algorithm, a correlation between the first data and the second data, wherein the correlation is indicative of a current or future physiological state of the subject.
Aspect 42. The method of Aspect 41, wherein the first data further includes a characteristic of the subject, the characteristic being other than WBV, and the characteristic optionally comprising any one or more of age, gender, medical history, surgical history, social history, a pending medical condition, or a medication.
Aspect 43. The method of any one of Aspects 41-42, wherein the second data further includes a characteristic of the subject, the characteristic being other than WBV, and the characteristic optionally comprising any one or more of age, gender, medical history, surgical history, social history, a pending medical condition, or a medication.
Aspect 44. The method of any one of Aspects 41-43, wherein the physiological state is a septic state.
Aspect 45. A system, comprising a viscometer stage, the viscometer stage configured to receive a whole blood sample from a subject and determine a WBV of the sample at one or more shear rates of from 1/s to 1000/s; a processing stage, the processing stage configured to implement a trained algorithm that receives at least a first data from the subject, the first data including a WBV collected at a first time and at a shear rate of from about 1/s to about 1000/s, the trained algorithm configured to determine a correlation between the first data and a first physiological state, the correlation being indicative of a current or future physiological state of the subject, or (2) the processing stage configured to implement a trained algorithm that receives (i) a first data from a subject, the first data including a WBV collected at a first time and at a shear rate of from about 1/s to about 1000/s and (ii) a second data from the subject, the second data including a WBV collected at a second time and at a shear rate of from about 1/s to about 1000/s, the trained algorithm configured to determine a correlation between the first data and the second data, wherein the correlation is indicative of a current or future physiological state of the subject.
Aspect 46. The system of Aspect 45, wherein the viscometer stage comprises a pressure transducer in fluid communication with a volume configured to receive a whole blood sample of a subject.
Aspect 47. The system of any one of Aspects 45-46, wherein the viscometer train is configured to collect a pressure vs. time curve of a whole blood sample within 60 seconds.
Aspect 48. A method of training an algorithm for detecting a current or future physiological state of a subject, the method comprising: inputting a plurality of datasets for each of a plurality of subjects into a model function, wherein the plurality of datasets include: a first dataset including a first WBV of a subject collected at a first time and at a shear rate of from about 1/s to about 1000/s; a second dataset including a second WBV of a subject collected at a second time and at a shear rate of from about 1/s to about 1000/s; and a third dataset including at least one or more physiological states of the subject; and calculating weighted values for each data type of the first dataset and the second dataset via the model function; and generating a trained model by assigning the weighted values to the data type of the first dataset and the second dataset.
The present application is a continuation in part of U.S. patent application Ser. No. 17/497,292, “Rapid Profile Viscometer Devices and Methods” (filed Oct. 8, 2021) (now allowed); which application claims priority to and the benefit of U.S. patent application No. 63/090,093, “Rapid Profile Viscometer” (filed Oct. 9, 2020). All foregoing applications are incorporated herein by reference in their entireties for any and all purposes.
Number | Date | Country | |
---|---|---|---|
63090093 | Oct 2020 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 17497292 | Oct 2021 | US |
Child | 18459947 | US |