The present invention is directed to a microfluidic device and to methods of using it.
Molecular biology based technologies, such as the polymerase chain reaction (PCR), for detection of pathogenic microorganisms are slowly replacing culture based detection methods (Kow et al., Journal of Medical Entomology 378(4):475-479 (2001); Laue et al., Journal of Clinical Microbiology 37(8):2543-2547 (1999); and Illen et al., Journal of Virol. Methods 41(2):135-146 (1993)). While molecular methods tend to be more sensitive, specific, and faster than culture based methods, they are also limited by expensive equipment requirements (Baeumner et al., Analytical Chemistry 74:1442-1448 (2002)). Scientists are overcoming this limitation by miniaturizing molecular assays into a microfluidic format (Mondesire et al., IVD Magazine 9-14 (2000); Yu et al., Micro Total Analysis Systems Conference, Enschede, Netherlands 545-548 (2000); Kopp et al., Science 280:1046-1048 (1998); and Manz et al., Journal of Chromatography 593:253-258 (1992)). Microfluidics is the enabling technology base for the development of miniature devices that move, mix, control, and react fluid volumes in the micron range. Microfluidics offer obvious advantages in the reduced consumption of reagents; faster and more sensitive reactions due to enhanced effects of processes such as diffusion and mass transport; increased throughput through parallel processing; and reduced expenses in terms of power and reagent consumption. Most importantly, fabrication of microfluidic devices is inexpensive and allows the integration of several modules to automate analytical processes (Duffy et al., Analytical Chemistry 70:4974-4984 (1998); Jingdong et al., Analytical Chemistry 72:1930-1933 (2000); and Martynova et al., Analytical Chemistry 69(23):4783-4789 (1997)).
A common feature of all nucleic acid detection methods in microarray chips and microchannels is the use of labels coupled to target specific probes. Typically, these labels are molecules that fluoresce, change, or produce color to indicate target hybridization to a probe (Ramsay, G., Nature Biotech 16:40-44 (1998)). Nanoparticles such as magnetic beads (Edelstein et al., Biosensors & Bioelectronics 14:805 (2000)), liposomes (Esch et al., Analytical Chemistry 73:2952-2958 (2001)) and gold particles (Taton et al., Science 289:1756-1760 (2002) and Cao et al., Science 297:1536-1540 (2002)) have also been used as labels. In most cases, these particle-labelled assays have proven to be more sensitive as they offer a means for further signal amplification that is not possible with conventional labels. Taton et al., for instance, use silver reduction to enhance visualization of gold particles in their assay (Taton et al., Science 289:1756-1760 (2002)). The least expensive and perhaps the simplest signal amplification scheme has been achieved with liposomes. Liposomes are phospholipid vesicles that entrap hundreds of thousands of marker molecules to provide a large signal amplification and enhanced sensitivity, 3 orders of magnitude greater than single fluorophore detection (Lee et al., Analytica Chimica Acta 354:23-28 (1997)).
Microfluidic mixers are an integral component of microscale total analysis systems (μTAS), which contain various modular units in a compact system (Manz et al., “Miniaturized total chemical analysis systems. A novel concept for chemical sensing,” Transducers '89: Proceedings of the 5th International Conference on Solid-State Sensors and Actuators and Eurosensors III. Part 1, Montreux, Switzerland (Jun. 25-30, 1989); van den Berg et al., Proceedings of the International Symposium on Micromechantronics and Human Science, pages 181-184 (1994); and Dhawan et al., Analytical and Bioanlytical Chemistry, 373:421-426 (2002)). Turbulence, the primary mechanism for macro-scale mixing, is effectively absent under normal conditions in most microfluidic systems due to low Reynolds numbers. Thus, alternative strategies for mixing in microfluidic systems must be employed. Several different strategies have been suggested in recent years based on a variety of different principles. Passive mixers use only the geometry of the channel to achieve mixing. Examples of passive mixers include those that generate transverse flows using a rigid arrangement of herring-bone structures to increase the interfacial area between liquids to be mixed (Stroock et al., Science 295:647-651 (2002)); that use a serpentine channel to simulate a partial packed bed of a chromatography column (He et al., “A Picoliter Volume Mixer for Microfluidic Analytical Systems,” Analytical Chemistry 73:1942 (2001)); and that use a T-junction mixer with deep well structures (Johnson et al., “Rapid Microfluidic Mixing,” Analytical Chemistry 74:45 (2002)). A thorough review of passive micromixers gives an overview of the basic physics involved in microscale mixing systems, and a discussion of the various geometries currently being employed in micromixers (Nguyen et al., “Micromixers—A review,” J. Micromech. Microeng. 15:R1-R16 (2005)). Active mixers generally use physical motion to induce mixing. An example of such a device is one that is based upon the movement of a stir bar under the influence of a magnetic field (Barbic et al., “Electromagnetic micromotor for microfluidics applications,” Applied Physics Letters 79:1399 (2001)). Another reported device includes a microfluidic device capable of recirculating nanoliter volumes within closed microfluidic channels, using counterbalancing hydrodynamic pressure against an electro-osmotically generated flow in a dead-end chamber (Lammertink et al., Anal. Chem. 76:3018-3022 (2004)). However, there is a need for a microfluidic mixer that does not have the deficiencies (noted above).
The present invention is directed to overcoming the above deficiencies in the art.
The present invention relates to a microfluidic test device for detecting or quantifying an analyte in a test sample. The device includes a non-absorbent substrate having at least one inlet and outlet extending therethrough. The inlet and outlet are connected by at least one microchannel imbedded in the substrate. The at least one microchannel includes an inlet portion and an analysis portion. The device also includes a non-specific capture device located at or upstream of the analysis portion. The device further includes one or more stationary mixing structures extending into the at least one microchannel.
The present invention also relates to a method for detecting or quantifying an analyte in a test sample. This involves providing at least one test mixture where the test mixture includes a test sample, potentially containing analyte, a capture conjugate, and a marker conjugate. The capture conjugate includes a capture support and a first binding material, where the first binding material is selected to bind with a portion of the analyte. The marker conjugate includes a particle, a marker, and a second binding material, where the second binding material is selected to bind with a portion of the analyte other than the portion of the analyte for which the first binding material is selected. The method also involves providing a microfluidic test device for detecting or quantifying an analyte in a test sample. The test device includes a non-absorbent substrate having at least one inlet and outlet extending therethrough, where the inlet and outlet are connected by at least one microchannel imbedded in the substrate, and where the at least one microchannel includes an inlet portion and an analysis portion. The test device also includes a non-specific capture device located at or upstream of the analysis portion. The test device further includes one or more stationary mixing structures extending into the at least one microchannel. Reaction is permitted to occur, within the microfluidic test device, in the test mixture between analyte present in the test sample and the first and second binding materials, thereby forming a product complex that includes analyte present in the test sample, the capture conjugate, and the marker conjugate. The reacted test mixture is contacted to the non-specific capture device (e.g., a device having non-specific affinity for the capture support), whereby product complex present in the reacted test mixture is immobilized from the reacted test mixture. The presence or amount of the marker from the immobilized product complex is detected at the analysis portion of the microfluidic test device and correlated with the presence or amount, respectively, of the analyte in the test sample.
Another aspect of the present invention relates to a method for detecting or quantifying an analyte in a test sample, as follows: This method includes providing at least one test mixture which includes a test sample potentially containing an analyte, a capture support complex including a capture support and a first member of a first coupling group, a first binding material selected to bind with a portion of the analyte and including a second member of the first coupling group, a marker complex which includes a particle, a marker, and a first member of a second coupling group, and a second binding material selected to bind with a portion of the analyte other than the portion of the analyte for which the first binding material is selected and including a second member of the second coupling group. The method also involves providing a microfluidic test device for detecting or quantifying an analyte in a test sample. The test device includes a non-absorbent substrate having at least one inlet and outlet extending therethrough, where the inlet and outlet are connected by at least one microchannel imbedded in the substrate, and where the at least one microchannel includes an inlet portion and an analysis portion. The test device also includes a non-specific capture device located at or upstream of the analysis portion. The test device further includes one or more stationary mixing structures extending into the at least one microchannel. Reaction is permitted to occur, within the microfluidic test device, in the at least one test mixture between the first and second members of the first coupling group, between the first and second members of the second coupling group, and between analyte present in the test sample and the first and second binding materials. As a result, a product complex including analyte present in the test sample, the capture support complex, the first binding material, the marker conjugate, and the second binding material is formed. The reacted test mixture is contacted to a non-specific capture device (e.g., a device having non-specific affinity for the capture support) so that product complex present in the reacted test mixture is immobilized from the reacted test mixture. The presence or amount of the marker from the immobilized product complex is detected at the analysis portion of the microfluidic test device and correlated with the presence or amount, respectively, of the analyte in the test sample.
Another aspect of the present invention relates to a method for detecting or quantifying an analyte in a test sample, as follows: This involves providing at least one test mixture including a test sample potentially containing an analyte, a capture conjugate (including a capture support and a first binding material), where the first binding material is selected to bind with a portion of the analyte, and a marker conjugate (including a particle, a marker, and an analyte analog). The method also involves providing a microfluidic test device for detecting or quantifying an analyte in a test sample. The test device includes a non-absorbent substrate having at least one inlet and outlet extending therethrough, where the inlet and outlet are connected by at least one microchannel imbedded in the substrate, and where the at least one microchannel includes an inlet portion and an analysis portion. The test device also includes a non-specific capture device located at or upstream of the analysis portion. The test device further includes one or more stationary mixing structures extending into the at least one microchannel. Competition is permitted to occur, within the microfluidic test device, in the at least one test mixture between analyte present in the test sample and the analyte analog for the first binding material. As a result, a product complex, including the capture conjugate and the marker conjugate, is formed. The reacted test mixture is contacted to a non-specific capture device (e.g., a device having non-specific affinity for the capture support) so that product complex present in the reacted test mixture is immobilized from the reacted test mixture. The immobilized product complex is detected at the analysis portion. The presence or amount of the marker from the immobilized product complex is correlated with the presence or amount, respectively, of the analyte in the test sample.
The present invention also relates to a microfluidic device (also referred to herein as a recirculating microfluidic device, a microfluidic mixing device, or the like). This device includes a non-absorbent substrate having at least one inlet and outlet extending therethrough and one or more stationary mixing structures. The at least one inlet and outlet are connected by at least one microchannel imbedded in the substrate. The one or more stationary mixing structures extend into the at least one microchannel.
Microfluidics combined with a liposome signal amplification scheme, in accordance with the present invention, promises an inexpensive solution to the heightened need for technology that can rapidly and accurately detect pathogenic organisms in environmental, clinical, and food samples in the wake of recent threats of bioterrorism. Liposome technology has been used in analogous membrane detection systems with great success (Baeumner et al., Analytical Chemistry 74:1442-1448 (2002); Esch et al., Analytical Chemistry 73:3162-3167 (2001); and Rule et al., Clinical Chemistry 42:206-1209 (1996), which are hereby included by reference in their entirety). It has been reported that gains in sensitivity can be achieved by converting a liposome-based membrane detection assay for Cryptosporidium parvum to a microfluidic format (Esch et al., Analytical Chemistry 73:3162-3167 (2001); and Rule et al., Clinical Chemistry 42:206-1209 (1996); and Taton et al., Science 289:1756-1760 (2002), which are hereby incorporated by reference in its entirety) (see also Goral et al., “Electrochemical microfluidic biosensor for the detection of nucleic acid sequences,’ Lab on a Chip 6(6):414-421 (2006); Zaytseva et al., “Microfluidic biosensor for the serotype-specific detection of dengue virus RNA,” Analytical Chemistry 77(23):7520-7527 (2005); and Zaytseva et al., “Development of a microfluidic biosensor module for pathogen detection,” Lab on a Chip 5(8):805-811 (2005), which are hereby incorporated by reference in their entirety).
The passive microfluidic mixer of the present invention is capable of establishing a recirculating flow inside mobile and open volumes from the nanoliter to the microliter range. Mixing in the device occurs not by generating transverse flows perpendicular to the length of the channel (see Stroock et al., Science 295:647-651 (2002), which is hereby incorporated by reference in its entirety), but instead by generating transverse flows parallel to the length of the channel, such that streamline segments at different lengths of the channel can be brought into contact with each other. It takes advantage of a fluid-exchange principle (described in U.S. Pat. No. 6,331,073 to Chung et al., which is hereby incorporated by reference in its entirety): the device provides order-changing functions to a microfluid, i.e., allowing sections of fluid separated by a length of the channel to interact directly. The device effectively “folds” the solution to permit streamlines that are normally linearly separated to come into contact. In one embodiment, the microfluidic device is a microfluidic mixer that is pressure driven in an open-end chamber using an attached syringe controlled by an external motor.
The present invention relating to the recirculating microfluidic mixer can find application in a variety of bioanalytical and chemical micro/nano systems, such as (but not limited to) microfluidic sensors, micro-Total Analysis Systems. For example, it can be used for the effective and rapid mixing of several solutions, it can be used to decrease the time needed for a nucleic acid sequence-based amplification (NASBA) reaction, or any catalytically derived reaction, any hybridization reaction, any binding reaction (e.g., RNA-DNA hybridization reactions using liposome and magnetic beads with immobilized DNA oligonucleotides).
The present invention relates to a microfluidic test device for detecting or quantifying an analyte in a test sample. The device includes a non-absorbent substrate having at least one inlet and outlet extending therethrough. The inlet and outlet are connected by at least one microchannel imbedded in the substrate. The at least one microchannel includes an inlet portion and an analysis portion. The device also includes a non-specific capture device located at or upstream of the analysis portion. The device further includes one or more stationary mixing structures extending into the at least one microchannel.
In one embodiment, there are a plurality of stationary mixing structures extending into the at least one microchannel. These stationary mixing structures can extend different lengths into the at least one microchannel. In another embodiment, each microchannel has opposite sides with at least some of the stationary mixing structures extending into the microchannel from the opposite sides in directions generally toward one another. In a further embodiment, the microfluidic test device can include one or more stationary mixing structures that extend into the one or more microchannels at an inclined angle. There can be a plurality of the stationary mixing structures with at least some extending into the one or more microchannels at different angles. In still another embodiment, there can be a plurality of inlets to each microchannel.
As used herein, the term stationary mixing structure can also be referred to as a sawtooth. The sawteeth can be of variable length (in order to catch different streamlines). The sawteeth can have variable angles. The sawtooth arrangement can included reversing sawteeth (i.e., having one set of sawteeth as shown plus a set mirrored to is in a second part of the channel), or putting sawteeth on either or both sides of the channel walls. Examples are shown in
The channel length is not critical and will simply provide more or less volume. Thus, devices have been fabricated that can have 5 nL total volume and those that can have 15 nL volume (as an example). The mixer can be made of other materials than PDMS, including (but not limited to) Si, SiO2, SU8, quartz, acrylic, etc. (see
In one embodiment, the microfluidic device is capable of recirculating microliter volumes is described. The device consists of molded polydimethyl siloxane (PDMS) channels with pressure inlet and outlet holes sealed by a glass lid (see
The recirculating microfluidic mixer of the present invention can be used in a variety of bioanalytical and chemical micro/nano systems, such a (but not limited to) microfluidic sensors, micro-Total Analysis Systems. For example, it can be used for the effective and rapid mixing of several solutions, it can be used to decrease the time needed for a nucleic acid sequence-based amplification (NASBA) reaction, or any catalytically derived reaction, any hybridization reaction, any binding reaction.
The microfluidic device can also be used in an NASBA reaction, and subsequently also for RNA-DNA hybridization reactions using liposome and magnetic beads with immobilized DNA oligonucleotides.
The non-absorbent substrate is formed from a material like quartz, glass, polymethylacrylate, polydimethyl siloxane, or polymeric materials.
The microfluidic test device can additionally include an incubation portion upstream of the analysis portion.
When the capture device and the analysis portion are at the same location, the complex containing analyte, capture conjugate, and marker conjugate can be detected at the capture device. When the analysis portion is downstream of the capture device, marker is released from the complex immobilized to the capture device and detected as it moves with fluid in the direction from inlet 102 to outlet 110. In a third embodiment, the analysis portion is located upstream of the capture device so that when marker is released from the immobilized complex, it is carried to the analysis portion by then reversing flow of fluid in the direction from outlet 110 to inlet 102.
The electrochemical detection assembly comprises a microcontroller-based analysis system. An example of such a system is described in the following paragraphs.
The current instrumentation is at once a potentiostat for electrochemical detection, a data acquisition/storage system, and a controller for the active components (such as the pump actuator and electromagnet) of the microfluidic biosensor. Requirements for portability, low power consumption, and a small form factor are achieved with an electronic design that uses as few components as possible.
The heart of the system is the low power, highly integrated MSP430FG439 microcontroller (“MCU”) from Texas Instruments. Texas Instruments produces a large range of devices that differ only in terms of the number of I/O pins, integrated peripherals, memory, and price. The underlying architecture of all the MCUs are the same. Thus, code written for one MCU will work on all MCUs with a few changes to the initialization setup. The flexibility offered by MCU choice allows the manufacture of inexpensive basic analysis systems as well as deluxe systems using the same code-base. Furthermore, the system can be easily upgraded with an advanced MCU.
The MSP430 has 4 main sections —CPU, memory, clock, and peripherals. See
The MSP430FG439 has 60 KB of program memory and 2K SRAM. Program memory is flash and self-programmable. This feature allows about 100 data files to be stored for 1-minute measurements taken in 1 sec intervals. The storage capacity can be increased with extra non-volatile memory modules.
The clock system is very flexible and allows the device to operate in a very low power mode at 32 KHz for unattended periodic measurements for instance, and up to a fast 8 MHz for data acquisition, analysis, transmission, and display in real time.
Most of the system's functionality is provided by the MCU's peripherals. The MSP430FG439 has a built in liquid crystal display (“LCD”) controller, 1 universal synchronous asynchronous receiver transceiver (“USART”), an 8 channel 12-bit analogue-to-digital-conversion (“ADC”) port, 2 channel 12 bit digital-to-analogue conversion (“DAC”) port, 3 operational amplifiers, a built in supply voltage supervisor, 6 general input/output (I/O) ports and 4 timers. In this application, the basic timer is used to maintain a real time clock and time stamps for logged data. It also supplies the LCD frame frequency rate. Timer A is used to generate alarm and distinctive status beeps on a buzzer. Timer B is used to generate PWM outputs used to control peripherals external to the MCU. Any of the timers can be set to keep track of the interval and duration of measurement. The watchdog timer can also reset the device when errors occur during operation.
The firmware for the MCU is written mainly in C and compiled with the open-source MSPGCC compiler for the MSP430 line of microcontrollers. The microcontroller is in-circuit programmable via a JTAG interface. The current design calls for the JTAG headers to be left in the circuit so that the firmware can be upgraded and easily debugged. However, the interface can also be removed to prevent tampering. Writing the code in C offers another distinctive advantage to this system, with the addition of a hardware configuration file for the parts and peripherals, any capable microcontroller can be substituted for the MSP430 line of microcontrollers.
The other major components of the system are the analogue chain couplings to the ADC and DAC channels of the MCU. Each ADC channel is coupled to a programmable gain current-to-voltage amplifier. The amplifier converts the current induced in an IDUA sensor to a voltage and amplifies it. The signal is then captured and logged by the analogue-to-digital converter of the MCU. The potential signal is converted back to current in software before display.
The built-in DAC peripheral supplies a bias potential of up to 2.5V for the IDUA. The potential is adjusted by a user via the user interface described in more detail below.
The ADC and DAC analogue chains form a potentiostat for the electrochemical detection scheme of the biosensor. As mentioned earlier, this circuit was derived from a standalone analogue version that was thoroughly tested.
The operation of the microcontroller-based device is interrupt driven. For the most part, the MCU stays in low power mode. In this mode, the real time clock is on but most of the peripherals are turned off. The device enters active mode only in response to interrupts generated by pressing one of the push buttons; communication received on the USART; a power-on reset; low battery alarm; or any of the timers. The operations are summarized in the
The device is battery powered. When connected, a power-on reset starts up the MCU. It goes through an initialization sequence and preps its peripherals and timers. The MCU then enters and stays in the main loop of low power mode with bursts of activity generated by other interrupts.
Every interrupt received is processed in order of priority. Each interrupt wakes the MCU up and puts it in active mode to perform whatever activity is required. Once all the instructions have been processed in active mode, the MCU goes back into low power mode to wait for the next event.
Four push buttons generate interrupts that turn the LCD display on or off, initiate measurements, initiate a change in the parameters, and puts the device in monitor mode where the device wakes up periodically to take a measurement at predetermined intervals. The functions are not fixed and can be re-programmed as needed.
The device also wakes up when it receives an input on its USART port. This input may be a request to retrieve logged data for instance. A low battery interrupt disables most of the MCU activity and generates alarms that may include—beeping and/or flashing the low battery sign on the LCD. The watchdog timer generates an interrupt if there is a problem with the execution n of an instruction. This interrupt will cause the device to re-initialize itself with default parameters and notify the user accordingly.
The user interface currently includes an LCD, a serial connection to a computer, 4 buttons as well as connections to a keypad. The interface also includes a cross-platform graphical user interface (“GUI”) with access to the underlying platform's internet capabilities that a client may use to change measurement or control parameters, upload/download data, and visualize sensor output. The modular design of the system allows other communication schemes such as ethernet, infrared, and wireless to be easily integrated as needed.
The GUI provides an easy to use menu-driven interface for adjusting sensor potential, full scale measurement range, measurement interval, communication settings and setting the correct time. Currently, sensor potential may range from 0 to 1500 mV. Full scale range (+/−) may be 10 nA to 1 mA. Measurement interval is a minimum of 0.5 seconds at the moment.
There is no restriction on measurement duration if data storage is not required of the MCU. The capacity of the MCU at the moment is restricted to 6000 data points. Duration thus depends on capacity. Thus, for 1 sec intervals, measurement duration should not exceed 100 mins. Capacity can be increased to 30000 data points in the MCU's flash memory. Also, as mentioned earlier, capacity can be increased with external dataflash.
The GUI also allows the user to watch sensor signals change in real time on a graph or graph data downloaded from the MCU. The data can also be saved as comma delimited files for viewing and analysis in third party applications.
The present invention also relates to a method for detecting or quantifying an analyte in a test sample. This involves providing at least one test mixture where the test mixture includes a test sample, potentially containing analyte, a capture conjugate, and a marker conjugate. The capture conjugate includes a capture support and a first binding material, where the first binding material is selected to bind with a portion of the analyte. The marker conjugate includes a particle, a marker, and a second binding material, where the second binding material is selected to bind with a portion of the analyte other than the portion of the analyte for which the first binding material is selected. The method also involves providing a microfluidic test device for detecting or quantifying an analyte in a test sample. The test device includes a non-absorbent substrate having at least one inlet and outlet extending therethrough, where the inlet and outlet are connected by at least one microchannel imbedded in the substrate, and where the at least one microchannel includes an inlet portion and an analysis portion. The test device also includes a non-specific capture device located at or upstream of the analysis portion. The test device further includes one or more stationary mixing structures extending into the at least one microchannel. Reaction is permitted to occur, within the microfluidic test device, in the test mixture between analyte present in the test sample and the first and second binding materials, thereby forming a product complex that includes analyte present in the test sample, the capture conjugate, and the marker conjugate. The reacted test mixture is contacted to the non-specific capture device (e.g., a device having non-specific affinity for the capture support), whereby product complex present in the reacted test mixture is immobilized from the reacted test mixture. The presence or amount of the marker from the immobilized product complex is detected at the analysis portion of the microfluidic test device and correlated with the presence or amount, respectively, of the analyte in the test sample. In one embodiment, the permitting reaction to occur and the contacting steps are carried out by cycling the test mixture in opposite directions in the at least one microchannel.
The term “analyte” is meant to include the compound or composition to be measured or detected. It is capable of binding to the first and second binding materials. Suitable analytes include, but are not limited to, antigens (e.g., protein antigens), haptens, cells, and target nucleic acid molecules. A preferred analyte is a target nucleic acid molecule. The present invention is applicable to procedures and products for determining a wide variety of analytes. As representative examples of types of analytes, there may be mentioned: environmental and food contaminants, including pesticides and toxic industrial chemicals; drugs, including therapeutic drugs and drugs of abuse; hormones, vitamins, proteins, including enzymes, receptors, and antibodies of all classes; prions; peptides; steroids; bacteria; fungi; viruses; parasites; components or products of bacteria, fungi, viruses, or parasites; aptamers; allergens of all types; products or components of normal or malignant cells; etc. As particular examples, there may be mentioned T4; T3; digoxin; hCG; insulin; theophylline; leutinizing hormones; and organisms causing or associated with various disease states, such as Streptococcus pyrogenes (group A), Herpes Simplex I and II, cytomegalovirus, chlamydiae, etc. The invention may also be used to determine relative antibody affinities, and for relative nucleic acid hybridization experiments, restriction enzyme assay with nucleic acids, binding of proteins or other material to nucleic acids, and detection of any nucleic acid sequence in any organism, i.e., prokaryotes and eukaryotes. A more preferred analyte is a target nucleic acid molecule found in an organism selected from the group consisting of bacteria, fungi, yeast, viruses, protozoa, parasites, animals (e.g., humans), and plants. Suitable organisms include, but are not limited to, Cryptosporidium parvum, Escherichia coli, Bacillus anthracis, Dengue virus, and Human immunodeficiency virus (HIV-1).
The term “binding material” is meant to include a bioreceptor molecule such as an immunoglobulin or derivative or fragment thereof having an area on the surface or in a cavity which specifically binds to and is thereby defined as complementary with a particular spatial and polar organization of another molecule—in this case, the analyte. Suitable binding materials include antibodies, antigens, nucleic acid molecules, aptamers, cell receptors, biotin, streptavidin, and other suitable ligands. When the analyte is a target nucleic acid molecule, the first binding material can be a nucleic acid molecule (e.g., reporter probe, selected to hybridize with a portion of the target nucleic acid molecule) and the second binding material can be a nucleic acid molecule (e.g., capture probe, selected to hybridize with a separate portion of the target nucleic acid molecule), or other moiety, such as an antibody or other agent capable of binding to and interacting with the analyte.
Antibody binding materials can be monoclonal, polyclonal, or genetically engineered (e.g., single-chain antibodies, catalytic antibodies) and can be prepared by techniques that are well known in the art, such as immunization of a host and collection of sera, hybrid cell line technology, or by genetic engineering. The binding material may also be any naturally occurring or synthetic compound that specifically binds the analyte of interest.
The first and second binding materials are selected to bind specifically to separate portions of the analyte. For example, when the analyte is a nucleic acid sequence, it is necessary to choose probes for separate portions of the target nucleic acid sequence. Techniques for designing such probes are well-known. Probes suitable for the practice of the present invention must be complementary to the target analyte sequence, i.e., capable of hybridizing to the target, and should be highly specific for the target analyte. The probes are preferably between 17 and 25 nucleotides long, to provide the requisite specificity, while avoiding unduly long hybridization times and minimizing the potential for formation of secondary structures under the assay conditions. Thus, in this embodiment, the first binding material is reporter probe, which is selected to, and does, hybridize with a portion of target nucleic acid sequence. The second binding material, referred to herein as a capture probe for the nucleic acid detection/measurement embodiment, is selected to, and does, hybridize with a portion of target nucleic acid sequence other than that portion of the target with which reporter probe hybridizes. The capture probe may be immobilized in a capture portion of the microchannel or on a magnetic bead. In addition, the first and second binding materials (reporter and capture probes) should be capable of no or limited interaction with one another. Techniques for identifying probes and reaction conditions suitable for the practice of the invention are described in Sambrook et al., Molecular Cloning: A Laboratory Manual, Cold Spring Harbor Laboratory Press (1989), which is hereby incorporated by reference in its entirety. A software program known as “Lasergene”, available from DNASTAR, or similar products may optionally be used.
The method of the invention employs marker complexes which include a particle, a marker, and one member of a coupling group. Suitable particles include liposomes (the marker may be encapsulated within the liposome, or incorporated in the bilayer), latex beads, gold particles, silica particles, dendrimers, quantum dots, magnetic beads (e.g., antibody-tagged magnetic beads and nucleic acid probe-tagged magnetic beads), or any other particle suitable for derivatization. Where multiple marker complexes are used, the marker in each complex may be the same or different.
The use of liposomes as described in the present application provides several advantages over traditional signal production systems employing, for example, enzymes. These advantages include increased signal intensity, shelf stability, and instantaneous release of signal-producing markers, as described in Siebert et al., Analytica Chimica Acta 282:297-305 (1993); Yap et al., Analytical Chemistry 63:2007 (1991); Plant et al., Analytical Biochemistry 176:420-426 (1989); Locascio-Brown et al., Analytical Chemistry 62:2587-2593 (1990); and Durst et al., Eds., Flow Injection Analysis Based on Enzymes or Antibodies, vol. 14, VCH, Weinheim (1990), each of which is hereby incorporated by reference in its entirety.
Liposomes can be prepared from a wide variety of lipids, including phospholipids, glycolipids, steroids, relatively long chain alkyl esters; e.g., alkyl phosphates, fatty acid esters; e.g. lecithin, fatty amines, and the like. A mixture of fatty materials may be employed, such as a combination of neutral steroid, a charge amphiphile and a phospholipid. Illustrative examples of phospholipids include lecithin, sphingomyelin, and dipalmitoylphosphatidylcholine, etc. Representative steroids include cholesterol, chlorestanol, lanosterol, and the like. Representative charge amphiphilic compounds generally contain from 12 to 30 carbon atoms. Mono- or dialkyl phosphate esters, or alkylamines; e.g. dicetyl phosphate, stearyl amine, hexadecyl amine, dilaurylphosphate, and the like are representative.
The liposome vesicles are prepared in aqueous solution containing the marker, whereby the vesicles will include the marker in their interiors. The liposome vesicles may be prepared by vigorous agitation in the solution, followed by removal of the unencapsulated marker. Alternatively, reverse phase evaporation plus sonication can be used. Further details with respect to the preparation of liposomes are set forth in U.S. Pat. No. 4,342,826 and PCT International Publication No. WO 80/01515, both of which are hereby incorporated by reference in their entirety.
The concentration of electrolytes in the medium will usually be adjusted to achieve isotonicity or equi-osmolality (or up to about 50 to about 100 mmol/kg hypertonic) with the solution in the interior of liposomes to prevent their crenation or swelling.
With some increased complexity of the excitation waveform applied by the electroanalyzer, electrochemical measurement in accordance with the invention may also be carried out using stripping voltammetry, employing, for example, liposome encapsulated metal ions for detection and measurement.
Moderate, and desirably substantially constant, temperatures are normally employed for carrying out the assay. The temperatures for the assay and production of a detectable signal will generally be in the range of about 4-65° C., more usually in the range of about 20-38° C., and frequently, will be about 15-45° C.
The solvent for the test mixture will normally be an aqueous medium, which may be up to about 60 weight percent of other polar solvents, particularly solvents having from 1 to 6, more usually of from 1 to 4, carbon atoms, including alcohols, formamide, dimethylformamide and dimethylsulfoxide, dioxane, and the like. Usually, the cosolvents will be present in less than about 30-40 weight percent. Under some circumstances, depending on the nature of the sample, some or all of the aqueous medium could be provided by the sample itself.
The pH for the medium will usually be in the range of 2-11, usually 5-9, and preferably in the range of about 6-8. The pH is chosen to maintain a significant level of binding affinity of the binding members and optimal generation of signal by the signal producing system. Various buffers may be used to achieve the desired pH and maintain the pH during the assay. Illustrative buffers include borate, phosphate, carbonate, tris, barbital, and the like. The particular buffer employed is usually not critical, but in individual assays, one buffer may be preferred over another. For nucleic acid analytes, it is necessary to choose suitable buffers. Such buffers include SSC, sodium chloride, sodium citrate buffer, and SSPE (sodium chloride, sodium phosphate, EDTA).
This method can be carried out with the bioanalytical microsystem which includes a sample preparation module and a biosensor module. This entire system is preferably produced in a microfluidic platform.
The principle of the biosensor of the present invention is based on DNA/RNA hybridization system and liposome signal amplification (
The microfluidic device of the present invention can be designed to carry out fluorescent or electrochemical methods for signal detection. The approach for construction of the microfluidic device was based on providing precise sample handling in terms of volume and flow-rates, zero-dead volume at inlet and outlet points (no sample losses during the analysis and 100% waste disposal), ability to disassemble the device for replacing the microfluidic channel or transducer parts. Liposome, Dengue virus RNA, reporter and capture probes, and hybridization and washing buffers were used as optimized in experiments previously carried out in the development of membrane strip-based biosensors for Dengue virus detection (Baeumner et al., “A Biosensor for Dengue Virus Detection: Sensitive, Rapid and Serotype Specific,” Analytical Chemistry, 74(6):1442-1448 (2002) and Zaytseva et al., “Multi-Analyte Single-Membrane Biosensor for Serotype-Specific Detection of Dengue Virus,” Anal. Bioanal. Chem. 380:46-53 (2004), which are hereby incorporated by reference in their entirety).
Microfluidic channels can be fabricated as raised structures on 4 inch silicone wafers using standard photolithography processes. A 1 mL of freshly prepared 7:1 by volume mixture of silicone elastomer and silicone elastomer curing agent (Sylgard, 184 Silicone elastomer kit) was poured onto the silicone template and covered with another flat silicone wafer. Covering the resulting polydimethylsiloxane (PDMS) layer with a silicone wafer allowed thickness and thickness uniformity of the layer to be controlled. The obtained sandwich structure (
Microfluidic channels employed in the microfluidic device of the present invention should fulfill the following requirements: 1) The geometry and dimensions of a channel should be suitable to avoid large pressure drops in the liquid flow upon entering it. Experimentally, it was found that 100 micrometer wide, 50 micrometer deep channel satisfied this requirement; 2) The channel should have a small region with 5 times slower linear fluid flow compared to the rest of it. In this region, magnetic beads that are utilized during the analysis are captured. The signal transducer is placed downstream of the captured beads area; and 3) The PDMS layer with embedded channels structure should have inlet and outlet holes that go vertically through the entire width of the PDMS and that have a diameter not bigger than the width of the adjacent channel. These appropriate inlet dimensions are especially needed in order to avoid volumes of stagnated flow during the analysis.
Typical channel geometries with dimensions in micrometers are presented in
The formation of the PDMS 22 with embedded channels 24 is shown in
For the microfluidic device with fluorescent detection, glass plate 26 is clear. In the case of electrochemical detection, glass plate 26 is provided with patterned interdigitated ultramicroelectrode arrays (IDUAs) which are in fluid communication with channels 24 so that material passing through channel 24 can be in contact with and analyzed by the IDUAs.
Leak tight sealing was achieved by applying pressure from above PDMS layer 22 and beneath glass plate 26. For this purposes two plates 28 and 4-8 screws 30 were used as shown in
It should be noted that the upper plate has tubing 32 and 34 glued into the locations that line up with inlet 102 and outlet 110 of the PDMS device. The pressure applied onto the PDMS-glass plate device also provides a seal for the PDMS-Plexiglas interface. Initially, metal tubing was used in the Plexiglas inlet and outlet holes. However, due to high background signal during the electrochemical detection, these were replaced with plastic tubing.
In order to accommodate a magnet required to capture magnetic beads during the analysis in the capture zone, a groove can be made in the upper Plexiglas plate. The distance between the magnet and the upper wall of the PDMS channel can be precisely controlled by the depth of the groove and the thickness of the PDMS layer. These parameters and the strength of the magnetic field have a great influence on the ability to quantitatively capture beads during the analysis under varying flow rates. The closer the magnet is positioned with respect to the upper wall of the microchannel, the higher flow rates can be used during the analysis. In the microfluidic device of the present invention, the magnet (35DNE1304-NI, Magnet Applications, Inc.) is placed at a distance of 270 μm from the upper wall of the channel. This allows all the beads (1 μm diameter) to be captured at a linear flow rate of 0.2 m/min or 5 μL/min.
The capture device can be any device which achieves non-specific binding (i.e. does not involve use of any of the above-described binding materials). A magnetic field generating device or a filter with a binding material are particularly preferred capture devices. Any suitable solid support can be utilized as the capture support to which the capture device has an affinity. It is particularly preferred to use magnetic beads as the capture support, while the capture device comprises a magnetic field generating device. In this embodiment, as shown in
When a filter is used as the capture device, any porous material having a pore size of from about 0.1 μm to about 100 μm, preferably from about 1 μm to about 30 μm, which allows an aqueous medium to flow therethrough can be used. The pore size has an important impact on the performance of the device. The pore size has to be larger than the mean diameter of the marker. Also, the pores should not be too large so that a good volume to surface ratio can be obtained and to hold back the magnetic, polymer, or silica beads coupled to capture probes. Additionally, the filter could function as a conventional filter and retain large particles. Thus, liposomes bound to silica or other particles (the silica or other particles being too large to fit through the filter) would be retained but all other liposomes would pass through the filter. As a result, the amount of target present could be measured by the amount of liposomes bound via target to silica or other particles retained on the filter.
Suitable filter membranes for the device and methods of the present invention include nitrocellulose membranes, nitrocellulose mixed esters, mylar membranes, polysulfonyl based membranes, plain filter paper, glass fiber membranes, and membranes of any plastic material with defined pore size, such as polycarbonate filters, porous gold, and porous magnetic material. It can also be fabricated using microfabrication tools directly inside the microchannel using photoresist materials, such as SU-8 or also PDMS. The filter membranes can be of a variety of shapes, including rectangular, circular, oval, trigonal, or the like.
When the optical detection embodiment of the present invention is utilized, an optical marker is immobilized in the liposome. Suitable optical markers include a fluorescent dye, visible dyes, bio- or chemi-luminescent materials, quantum dots, and enzymatic markers. A qualitative or semi-quantitative measurement of the presence or amount of an analyte of interest may be made with the unaided eye when visible dyes are used as the marker. The intensity of the color may be visually compared with a series of reference standards, such as in a color chart, for a semi-quantitative measurement. Alternatively, when greater precision is desired, or when the marker used necessitates instrumental analysis, the intensity of the marker may be measured directly on the membrane using a quantitative instrument such as a reflectometer, fluorimeter, spectrophotometer, electroanalyzer, etc.
When using liposomes as the particle, the amount of marker material present can be measured without lysis of the liposomes. However, lysis can be used to enhance such visualization. This may be accomplished by applying a liposome lysing agent. Suitable liposome lysing materials include surfactants such as octylglucopyranoside, sodium dioxycholate, sodium dodecylsulfate, saponin, polyoxyethylenesorbitan monolaurate sold by Sigma under the trademark Tween-20, and a non-ionic surfactant sold by Sigma under the trademark Triton X-100, which is t-octylphenoxypolyethoxyethanol. Octylglucopyranoside is a preferred lysing agent for many assays, because it lyses liposomes rapidly and does not appear to interfere with signal measurement. Alternatively, complement lysis of liposomes may be employed, or the liposomes can be ruptured with electrical, optical, thermal, or other physical means.
A suitable arrangement for the embodiment of the present invention using optical detection is shown in
Interdigitated ultramicroelectrode arrays (“IDUA”) can be fabricated on glass wafers using standard photolithographic and lift-off techniques. A typical IDUA was produced by evaporation deposition of 70 nm Ti followed by 500 nm Au on patterned Pyrex glass wafers (7740, Corning, N.Y.). IDUAs with different dimensions were studied as signal transducers in oxidation-reduction reaction of the potassium ferro/ferrihexacyanide, Fe2+/Fe3+ (CN)6, pair. It has been shown that both, the background noise and the specific signal depend on the microelectrode's finger/gap ratio as well as on the total amount of fingers (Min et al., “Characterization and Optimization of Interdigitated Ultramicroelectrode Arrays as Electrochemical Biosensor Transducers,” Electroanalysis, 16(9):724-729 (2004), which is hereby incorporated by reference in its entirety). The IDUA designed with 3.8 μm wide fingers and 2.5 μm wide gaps with a total of 1000 electrode fingers demonstrated the best characteristics in terms of sensitivity and signal to noise ratio. Typical microphotographs of IDUAs are present in
The general principles described above have been used for the device assembling.
IDUAs fabricated on glass plates are used as signal transducers for the electrochemical signal detection scheme. During assembly, the PDMS channel is positioned on the glass in such a way that the IDUA detection zone is located downstream of the capture zone. In addition, the PDMS channel should be on top of the active microelectrode fingers (
When the electrochemical detection embodiment of the present invention is utilized, an electroactive species, such as potassiumhexaferrocyanide and potassium hexaferricyanide, is encapsulated in the marker, e.g., liposomes. The microchannel is placed above reusable electrodes, such as an interdigitated electrode array, as described above. After lysis of the liposomes, the quantity of the electroactive species is determined.
Suitable electrochemical markers, as well as methods for selecting them and using them are disclosed, for example, in U.S. Pat. No. 5,789,154 to Durst et al., U.S. Pat. No. 5,756,362 to Durst et al., U.S. Pat. No. 5,753,519 to Durst et al., U.S. Pat. No. 5,958,791 to Roberts et al., U.S. Pat. No. 6,086,748 to Durst et al., U.S. Pat. No. 6,248,956 to Durst et al., U.S. Pat. No. 6,159,745 to Roberts et al., U.S. Pat. No. 6,358,752 to Roberts et al., and co-pending U.S. patent application Ser. No. 10/264,159, filed Oct. 2, 2002, which are hereby incorporated by reference in their entirety. Briefly, the test device may be designed for amperometric detection or quantification of an electroactive marker. In this embodiment, the test device includes a working electrode portion(s), a reference electrode portion(s), and a counter electrode portion(s) in the microfluidic device. The working electrode portion(s), reference electrode portion(s), and counter electrode portion(s) are each adapted for electrical connection to one another via connections to a potentiostat. The test device can instead include a working electrode portion and a counter electrode portion. Alternatively, the microfluidic device may be designed for potentiometric detection or quantification of an electroactive marker. In this embodiment, the device includes an indicator electrode portion(s) and a reference electrode portion(s). The indicator electrode portions and reference electrode portions are adapted for electrical connection to potentiometers. In another embodiment, the test device may include an interdigitated electrode array positioned to induce redox cycling of an electroactive marker released from liposomes upon lysis of the liposomes.
Suitable electroactive markers are those which are electrochemically active but will not degrade the particles (e.g., liposomes) or otherwise leach out of the particles. They include metal ions, organic compounds such as quinones, phenols, and NADH, and organometallic compounds such as derivatized ferrocenes. In one embodiment, the electrochemical marker is a reversible redox couple. A reversible redox couple consists of chemical species for which the heterogeneous electron transfer rate is rapid and the redox reaction exhibits minimal overpotential. Suitable examples of a reversible redox couple include, but are not limited to, ferrocene derivatives, ferrocinium derivatives, mixtures of ferrocene derivatives and ferrocinium derivatives, cupric chloride, cuprous chloride, mixtures of cupric chloride and cuprous chloride, ruthenium-tris-bipyridine, potassium ferrohexacyanide, potassium ferrihexacyanide, and mixtures of potassium ferrohexacyanide and potassium ferrihexacyanide. Preferably, the electrochemical marker is encapsulated within a liposome, in the bilayer, or attached to a liposome membrane surface.
A suitable arrangement for the embodiment of the present invention using electrical detection is shown in
As hereinabove indicated, the assay may be qualitative (presence or absence of certain level of target) or quantitative or semi-quantitative. The preparation of suitable standards and/or standard curves (the term “standard curve” is used in a generic sense to include a color chart) is deemed to be within the scope of those skilled in the art from the teachings herein.
In one embodiment, the test device includes multiple capture portions, each of which is modified to bind a distinctive second binding material specific for one of several analytes. Thus, each analyte may be determined by assignment of each conjugate/analyte to its own measurement portion for concentration and measurement. Alternatively, the conjugate of each of the analytes to be determined in this embodiment of the present invention, may include a marker which is distinctly detectable from the other markers. With different encapsulated dyes (e.g., fluorescent dyes) or quantum dots, the results of the assay can be “color coded”. In particular, a multi-wavelength detector can be used in a capture portion.
As a matter of convenience, the present device can be provided in a kit in packaged combination with predetermined amounts of reagents for use in assaying for an analyte or a plurality of analytes. Included within the kit are stabilizers, buffers, and the like. The relative amounts of the various reagents may be varied widely, to provide for concentration in solution of the reagents which substantially optimizes the sensitivity of the assay. The reagents can be provided as dry powders, usually lyophilized, including excipients, which on dissolution will provide for a reagent solution having the appropriate concentrations for performing the assay. The kit or package may include other components such as standards of the analyte or analytes (analyte samples having known concentrations of the analyte).
As described above, the method and device of the present invention can be used in a variety of assays, such as competitive binding assays and sandwich assays, as described in U.S. Pat. No. 5,789,154 to Durst et al., U.S. Pat. No. 5,756,362 to Durst et al., U.S. Pat. No. 5,753,519 to Durst et al., U.S. Pat. No. 5,958,791 to Roberts et al., U.S. Pat. No. 6,086,748 to Durst et al., U.S. Pat. No. 6,248,956 to Durst et al., U.S. Pat. No. 6,159,745 to Roberts et al., U.S. Pat. No. 6,358,752 to Roberts et al., co-pending U.S. patent application Ser. No. 09/698,564, filed Oct. 27, 2000, and co-pending U.S. patent application Ser. No. 10/264,159, filed Oct. 2, 2002, which are hereby incorporated by reference in their entirety.
Another aspect of the present invention relates to a method for detecting or quantifying an analyte in a test sample, as follows: This method includes providing at least one test mixture which includes a test sample potentially containing an analyte, a capture support complex including a capture support and a first member of a first coupling group, a first binding material selected to bind with a portion of the analyte and including a second member of the first coupling group, a marker complex which includes a particle, a marker, and a first member of a second coupling group, and a second binding material selected to bind with a portion of the analyte other than the portion of the analyte for which the first binding material is selected and including a second member of the second coupling group. The method also involves providing a microfluidic test device for detecting or quantifying an analyte in a test sample. The test device includes a non-absorbent substrate having at least one inlet and outlet extending therethrough, where the inlet and outlet are connected by at least one microchannel imbedded in the substrate, and where the at least one microchannel includes an inlet portion and an analysis portion. The test device also includes a non-specific capture device located at or upstream of the analysis portion. The test device further includes one or more stationary mixing structures extending into the at least one microchannel. Reaction is permitted to occur, within the microfluidic test device, in the at least one test mixture between the first and second members of the first coupling group, between the first and second members of the second coupling group, and between analyte present in the test sample and the first and second binding materials. As a result, a product complex including analyte present in the test sample, the capture support complex, the first binding material, the marker conjugate, and the second binding material is formed. The reacted test mixture is contacted to a non-specific capture device (e.g., a device having non-specific affinity for the capture support) so that product complex present in the reacted test mixture is immobilized from the reacted test mixture. The presence or amount of the marker from the immobilized product complex is detected at the analysis portion of the microfluidic test device and correlated with the presence or amount, respectively, of the analyte in the test sample. In a preferred embodiment, the marker is released from the immobilized product before the contacting and after the detection steps.
The components and steps used to carry out this aspect of the present invention are substantially the same as those described above.
Another aspect of the present invention relates to a method for detecting or quantifying an analyte in a test sample, as follows: This involves providing at least one test mixture including a test sample potentially containing an analyte, a capture conjugate (including a capture support and a first binding material), where the first binding material is selected to bind with a portion of the analyte, and a marker conjugate (including a particle, a marker, and an analyte analog). The method also involves providing a microfluidic test device for detecting or quantifying an analyte in a test sample. The test device includes a non-absorbent substrate having at least one inlet and outlet extending therethrough, where the inlet and outlet are connected by at least one microchannel imbedded in the substrate, and where the at least one microchannel includes an inlet portion and an analysis portion. The test device also includes a non-specific capture device located at or upstream of the analysis portion. The test device further includes one or more stationary mixing structures extending into the at least one microchannel. Competition is permitted to occur, within the microfluidic test device, in the at least one test mixture between analyte present in the test sample and the analyte analog for the first binding material. As a result, a product complex, including the capture conjugate and the marker conjugate, is formed. The reacted test mixture is contacted to a non-specific capture device (e.g., a device having non-specific affinity for the capture support) so that product complex present in the reacted test mixture is immobilized from the reacted test mixture. The immobilized product complex is detected at the analysis portion. The presence or amount of the marker from the immobilized product complex is correlated with the presence or amount, respectively, of the analyte in the test sample. In a preferred embodiment, the marker is released from the immobilized product before the contacting and after the detection steps.
In this embodiment of the present invention, an analyte analog is used, because this embodiment involves a competitive binding assay format. Thus, the term “analyte analog” is meant to include an analog which binds to the capture conjugate. When an analog is employed, however, it is necessary that the particular characteristics of the analyte necessary for recognition by the first binding material in the competition reaction be present in the analyte analog conjugated with the marker complex.
In all other respects, the components and steps used to carry out this aspect of the present invention are substantially the same as those described above.
The present invention also relates to a microfluidic device (also referred to herein as a recirculating microfluidic device, a microfluidic mixing device, or the like). This device includes a non-absorbent substrate having at least one inlet and outlet extending therethrough and one or more stationary mixing structures. The at least one inlet and outlet are connected by at least one microchannel imbedded in the substrate. The one or more stationary mixing structures extend into the at least one microchannel. In all other respects, the components and steps used to carry out this aspect of the present invention are substantially the same as those described above.
In one particular embodiment, the microfluidic device is capable of recirculating microliter volumes. This embodiment of the device can includes molded polydimethyl siloxane (PDMS) channels with pressure inlet and outlet holes sealed by a glass lid. Recirculation is accomplished by repeatedly changing the direction of flow over an iterated sawtooth structure. The sawtooth structure serves to change the fluid velocity of individual streamlines differently dependent on whether the fluid is flowing backwards or forwards over the structure. In this manner, individual streamlines can be accelerated or decelerated relative to the other streamlines to allow sections of the fluid to interact that would normally be linearly separated. Low Reynolds numbers imply that the process is reversible, neglecting diffusion. Fluorescent indicators can be employed to verify numerical simulations. It has been found that mixing of a Carboxyfluorescein labeled DMSO plug with an unlabeled DMSO plug across an immiscible hydrocarbon plug reached steady state in the channels with the sawtooth structures after 7.1 min, versus 34.8 min in the channels without sawtooth structures, which verified what would be expected based on numerical simulations.
Through inlet 102 of the microfluidic device shown in
A series of experiments was performed in order to optimize the detection of RNA in the microfluidic channels. These experiments were done without any liposome lysis and were monitored using the fluorescence microscope. The amount of liposomes (1.61 OD value for 1/100 dilution in PBS+ sucrose buffer, pH 7.0, osmolality 630 nmol/kg) with immobilized reporter probe (
To test the IDUA response in microfluidic system, as shown in
These results demonstrated that indeed the system based on the IDUA is capable of a fast response to the electrochemical composition changes inside the channel. The delay time between injection and the maximum signal reached was about 5-7 sec. In all the experiments the IDUA itself demonstrates a good reproducibility and the ability to function for prolonged periods of time without mechanical cleaning.
The typical results of RNA analysis by means of electrochemical detection is present in
Hybridization mixture was injected into inlet 102 at 3 μl/min. After all the beads were captured on the magnet and washed with 15 μl buffer, 25 mM solution of OG was injected into inlet 108 at 0.8 μl/min to lyse liposomes. Electrochemical responses of the IDUA in the presence and in the absence of RNA in the hybridization mixture are present in
A microfluidic biosensor for the highly specific and sensitive detection of pathogens via their nucleic acid sequence has been developed. The biosensor module employs the two alternative methods of detection, fluorescent or electrochemical. A microfabrication approach allows one to use microliter amounts of reagents to perform a single analysis. The microfluidic system was tested and optimized with a model Dengue virus target sequence. It has been shown that as low as 0.5 fmol of the synthetic target can be detected using a microfluidic platform, fluorescence detection method, and nonlysed liposomes.
Experimental data relating to a microfluidic device capable of recirculating nano to microliter volumes in order to efficiently mix solutions is described in this Example 4 and in the below Examples 5-9. The device consists of molded polydimethyl siloxane (PDMS) channels with pressure inlet and outlet holes sealed by a glass lid. Recirculation is accomplished by a repeatedly reciprocated flow over an iterated sawtooth structure. The sawtooth structure serves to change the fluid velocity of individual streamlines differently dependent on whether the fluid is flowing backwards or forward over the structure. Thus, individual streamlines can be accelerated or decelerated relative to the other streamlines to allow sections of the fluid to interact that would normally be linearly separated. Low Reynolds numbers imply that the process is reversible, neglecting diffusion. Computer simulations were carried out using FLUENT (Fluent, Inc.). Subsequently, fluorescent indicators were employed to experimentally verify these numerical simulations of the recirculation principal. Finally, mixing of a carboxyfluorescein labeled DMSO plug with an unlabeled DMSO plug across an immiscible hydrocarbon plug was investigated. At cycling rates of 1 Hz across five sawtooth units, the time was recorded to reach steady state in the channels, i.e., until both DMSO plugs had the same fluorescence intensity. In the case of the sawtooth structures, steady state was reached five times faster than in channels without sawtooth structures, which verified what would be expected based on numerical simulations. The microfluidic mixer is unique due to its versatility with respect to scaling, its potential to also mix solutions containing small particles such as beads and cells, and its ease of fabrication and use.
Microfluidic structures were designed using L-Edit (Tanner Research, Inc.) CAD software and fabricated using a silicon master mold and PDMS elastomer (Dow Corning, Corning, N.Y.). The mold was formed by DRIE into a photoresist patterned 100 mm Si wafer at the Cornell NanoScale Science and Technology Facility. After cleaning, Teflon AF (601S1-100-6) was poured, spun, and cured at 170° C. in an oven for 30 minutes. The channels were formed using 7 parts of PDMS elastomer and 1 part curing agent poured over the leveled silicon mold to a thickness of 1 mm and baked at 60° C. for 55 minutes in a vacuum oven at 0.5 bar. After curing, 0.75 mm holes were punched into the PDMS using a cork borer. The PDMS was then sliced into individual channels, oxidized using a Tesla coil, and placed in contact with a cleaned glass lid, where it was left for at least 30 minutes to seal permanently. An acrylic base and lid were used to secure the channels and align them accurately with a set of pressure inlets and outlets (
Carboxyfluorescein was obtained from Sigma-Aldrich Co. DMSO was obtained from Fisher Scientific. Mineral oil was obtained locally. A plug of mineral oil was injected between streams of pure DMSO and 1 mM Carboxyfluorescein labeled DMSO and streams with Carboxyfluorescein labeled DMSO on both sides of the plug, and was visualized using a Leica type DM LB microscope and Coolsnap camera and software package with an exposure time of 1 s using a 300 W UV arc lamp and subsequently color enhanced in Photoshop 7.0 (Adobe Systems, Inc.) using the auto-contrast and auto-level functions only.
Simulations were carried out using FLUENT software (Fluent, Inc.). Meshes were constructed using GAMBIT (Fluent, Inc.). Two-dimensional channels with stationary walls and pressure inlet-outlet ports were simulated. Thirty stream lines across 140 μm of channel length and 50 μm of channel width (at the inlets and outlets, 25 μm at the tip of the sawtooth) were tracked using simulated particle injections. It should be noted that the simulated structure was only one sawtooth unit (for reasons of computational practicality), whereas the experimental device consisted of 200 sawtooth units, each 150 μm long, connected together over a 3 cm channel, with a total volume of approximately 0.5 nL. The same device was fabricated with greater length in order to accommodate about 15 μL of solution.
The sawtooth structure of the micromixer was designed to cause mixing of solution in the microchannel based on recirculation. Thus, by repeatedly reciprocating the flow of a solution, some parts of the solution will be relocated with respect to their neighboring volume elements. Mixing occurs by generating transverse flows parallel to the length of the channel, such that streamline segments at different lengths of the channel can be brought into contact with each other. Computational simulations with GAMBIT and FLUENT were used to understand the effects of the sawtooth unit on the flow profiles.
The velocities of the individual streamlines and their profiles are shown in
Three streamlines were chosen for analyzing the separation efficiency of the streamlines by the sawtooth unit. These three streamlines represent distinct locations in the channel, i.e., in the top quarter (y=0.75*50 μm at x=0 μm), the middle of the channel (y=0.5*50 μm at x=0 μm) and the bottom quarter (y=0.25*50 μm at x=0 μm). The “middle” streamline has the highest velocity due to the parabolic flow profile of pressure driven microfluidic systems. In the left to right flow, a significant decrease of 10% per sawtooth unit in the “bottom” and “middle” streamline velocities was observed compared to right to left flow. It is this difference in rightward and leftward flow profiles that allows for the unique recirculating mixing based on transverse flows parallel to the length of the channel.
Based on the findings of the flow modeling studies, an optimal saw tooth mixer was designed by varying the unit lengths and sawtooth angles. Sawtooth angles were varied in increments of 5° from 20° to 70° and lengths were varied in increments of 2.5 μm from 10 μm to 40 μm. An optimal angle of 45° and length of 25 μm was chosen. The device was subsequently fabricated using standard photolithography and soft-lithography processes. A microfluidic system was assembled consisting of molded PDMS bonded to a glass lid, and connected to a KD Scientific Model 210 Syringe Pump using a machined acrylic assembly as shown in
A single pump was utilized for the micromixer in order to simplify the requirements of the ultimate design. Thus, positive and negative pressure for the fluid flow were used. By applying a pressure gradient across the entire microchannel, a parabolic flow profile is developed. It was found that devices with unoxidized PDMS leaked under positive pressure flow. Therefore, the surface of the PDMS was modified by oxidation using a Tesla coil to allow for permanent bonding to the glass lid, which created a sufficiently strong bond to allow for both positive and negative pressure to be applied from the same port.
Theoretically, the differences in backpressure between rightward (forward) and leftward (backward) flows that lead to the altered streamline velocity profiles shown in
In the second experiment, photographic investigations of the recirculation process were performed in order to prove the presence of recirculation generated by the sawtooth structures by mixing two plugs of DMSO, one fluorescently labeled and one unlabeled each occupying one half of the channel. The fluid was then rapidly moved back and forth (at a set flow rate of 10 μL/min) at a frequency of 1 Hz across the sawtooth structures. During this process photographs were taken of a 200 □m segment (
Finally, the mixing efficiency of the micromixer was compared to a straight channel (a channel without sawtooth units) of the same dimensions with respect to channel length, height and width. A hydrocarbon plug was injected between two streams of DMSO, but this time only one DMSO stream was fluorescently labeled. Thus, appearance of fluorescence in the second DMSO plug was again an indication of the recirculation mixing principle based on mixing between different streamlines. The time required for reaching homogeneous fluorescent DMSO plugs of the same fluorescence intensity was an indication of the mixing efficiency. It was determined that diffusion across the solution plug was insignificant by utilizing a plug consisting of unlabeled DMSO/Hydrocarbon/Carboxyfluorescein labeled DMSO left unmoved in the PDMS channel for 48 hours. At the conclusion of the 48 hours, the unlabeled DMSO still showed no fluorescence. In Table 3, typical experimental data obtained are summarized by providing the time needed to reach steady state, which is defined as the amount of time necessary, at continuous mixing of approximately 1 Hz across five sawtooth units (750 μm in the straight channel), for both DMSO plugs to show equal fluorescence.
Recirculation during mixing is necessary for many microfluidic applications such as enzyme catalyzed reactions, hybridization and binding reactions. It can be readily accomplished using the sawtooth structure described herein. Recirculation was obtained by the structures since these introduce an asymmetry in backward and forward flow that serves to introduce a separation between previously adjacent elements in neighboring streamlines. The fact that recirculation of the solution was obtained was demonstrated via fluid modeling and with three separate experiments. The design can easily be scaled up in length to house microliter volumes, and can bear broader straight channels at either end of the mixer segment to allow all of the solution to pass the entire length of the sawteeth structure. More radical variations on sawtooth placement, such as placing sawteeth on both sides of the channel, using more than one sawtooth length, width and angle in a single channel, etc., can be used. The microfluidic mixer described herein is unique due to its versatility with respect to scaling, its potential to also mix solutions containing small particles such as beads and cells, and its ease of fabrication and use.
Examples 10-15 relate to experiments regarding the electrochemical microfluidic biosensor and the recirculating microfluidic mixer of the present invention. An electrochemical biosensor for the detection of nucleic acid sequences was developed (Goral et al., “Electrochemical microfluidic biosensor for the detection of nucleic acid sequences,’ Lab on a Chip 6(6):414-421 (2006), which is hereby incorporated by reference in its entirety). To summarize, the target molecule was first hybridized with a capture probe which was immobilized on a paramagnetic bead, and a reporter probe which was conjugated to a liposome. The liposomes encapsulated ferrihexacyanide and ferrohexacyanide, Fe2+/3+(CN)6. The hybridization solution was then pumped through a 100 μm channel. A magnet was placed on the channel to capture the magnetic beads. Unhybridized liposomes would flow past the magnet and out an outlet port. A solution containing the surfactant octyl glucoside (OG) was then pumped toward the capture zone. Once the OG is in the capture zone, liposomes bound to the magnetic beads via the nucleic acid hybridizations lysed, releasing the redox solution into the channel.
Directly downstream of the capture zone, an interdigitated ultramicroelectrode array (IDUA) was able to measure the current which was proportional to the captured liposome concentration. A dose response curve estimated a limit of detection of 1 fmol of synthetic DNA target. Advantages of this electrochemical detection system include ease of use, cost effectiveness, and portability. Equivalent fluorescent systems require the use of a complicated detection device such as a photomultiplier tube or CCD camera in addition to an excitation source and filters.
Experiments have been conducted on the microfluidic mixer, which can be included in all three modules of the microfluidic device in order to enhance reaction and binding kinetics and avoid diffusion-based limitations. For example, its mixing characteristics, fabrication in larger dimensions (so that NASBA and liposome-RNA binding reactions can be carried out effectively), and fabrication using hot embossing rather than soft lithography were investigated.
The biosensor can be designed to perform three distinct steps: mRNA isolation, RNA amplification, and RNA detection. A single flow channel pattern was designed to accommodate characteristics needed for all three steps (see
Preliminary work has been done using the original reciprocating mixer using sawtoothed channels with a width of 50 μm using a polydimethyl siloxane (PDMS) body bonded to a glass microscope slide. This channel had two separate inlets and a single outlet (
One inlet was loaded with DI water while the other inlet was loaded with 50 mM fluorescein. The fluorscein was in a concentration high enough to experience self quenching. Therefore, any dilution of the fluorescein with the DI water would result in an increase in fluorescence during excitation. The flow was controlled using a syringe pump. The flow rate of both inlets was 1 μL/min. The fluorescence across the channel was observed using a microscope mounted with a CCD camera. The pixel intensity was later quantified using Image-Pro Express (MediaCybernetics, Silver Spring, Md.).
The pixel intensity was measured at the midpoint between the sawteeth for the first two centimeters (
The standard deviation of the pixel intensity across the channel at various distances could also be used as a measure of uniformity of fluorescein concentration (
The device fabrication was performed in part at the Cornell NanoScale Facility (Ithaca, N.Y.). A negative print of a channel was etched on to a silicon wafer using lift off photolithography. Initially, a blank wafer was coated with a layer of primer followed by a layer of S1813 photoresist. After a baking step, the mask was exposed to UV light using a contact aligner (HTG system III-HR, Hybrid Technology Group) for 10 seconds. The overlayed mask allowed exposure only of the areas between the channel structures. The UV light exposed regions become soluble to the photoresist developer. Following a post exposure baking of 90° C. for one minute, the wafer was developed in 1165 developer using an automatic MIF300 to remove the exposed regions of the photoresist resulting in the underlying silicon being exposed.
The wafer was then placed in a Unaxis SLR 770 to etch the channels. Silicon exposed to the inductively coupled plasma/reactive ion environment inside the chamber is etched at a rate of approximately 2 μm per minute. The etching process was allowed to run long enough to obtain a 100 μm deep etch, resulting in a channel height of that depth.
Following etching, the wafer was cleaned of any residual photoresist with acetone. The channel height and width were confirmed using a Tencor P10 profilometer.
The channel patterns were hot embossed into a polymethyl methacrylate (PMMA) substrate using an EV520HE semi-automated hot embossing system. This system allows controlled temperature, a high compression force and a high vacuum. The PMMA is sandwiched between two wafers, the structured wafer on top and a blank wafer below. The sandwich is placed between two temperature controlled plates. The top plate is hydraulically controlled to provide a desired compression force. After the chamber is set to a high vacuum, the top and bottom plates heat to 115° C. before applying 4000 N of force. The high temperature softens the PMMA while the applied force imprints the channel structures into the top of the PMMA. The high vacuum environment ensures that no air bubbles are trapped in the softened PMMA thereby causing channel distortion. The compression is held for 15 minutes before the temperature of both top and bottom compression plates are brought to 100° C. Below the softening temperature, the pressure can be released without disturbing the newly embossed channel structures. After the chamber pressure is normalized the PMMA is removed and the inlet and outlet holes are drilled.
There have been several methods used to bond two pieces of PMMA together while preserving a channel. The most common method is thermal bonding (Yahng et al., “Fabrication of microfluidic devices by using a femtosecond laser micromachining technique and mu-PIV studies on its fluid dynamics,” Journal of the Korean Physical Society 47(6):977-981 (2005); Li et al., “Low-temperature thermal bonding of PMMA microfluidic chips,” Analytical Letters 38(7):1127-1136 (2005); Chen et al., “Vacuum-assisted thermal bonding of plastic capillary electrophoresis microchip imprinted with stainless steel template,” Journal of Chromatography A 1038(1-2):239-245 (2004); and Keynton et al., “Design and development of microfabricated capillary electrophoresis devices with electrochemical detection,” Analytica Chimica Acta 507(1):95-105 (2004), which are hereby incorporated by reference in their entirety), which is hereby incorporated by its entirety), which is hereby incorporated by its entirety), which is hereby incorporated by its entirety). This process uses equipment similar to that used in the hot embossing. Two heated plates heat sandwiched pieces of PMMA until they soften. Under pressure, the softening allows the interface of the two plastics to fuse. This is a fast and simple method of bonding two PMMA pieces. The drawback is the deformation of the channels observed during the process. Another technique used for thermal bonding takes advantage of the ideal bonding temperature being close to 100° C. The two pieces of PMMA are tightly clamped together and immersed in boiling water for one hour (Kelly et al., “Thermal bonding of polymeric capillary electrophoresis microdevices in water,” Analytical Chemistry 75(8):1941-1945 (2003), which is hereby incorporated by reference in its entirety). The advantage is a good heat control. The disadvantage is that commercially available PMMA have different thermal properties depending on the manufacturer. This technique was tested with Optix® PMMA (Plaskolite, Inc., Columbus, Ohio) and found to cause a collapse in the channel due to too high a temperature. The conditions of these bonding techniques would cause the sawtooth structures of the micromixers to distort and lose the intended design. Therefore, a lower temperature technique is needed.
Solvent-assisted thermal bonding involves the application of a very thin layer of solvent on the surface of the unstructured PMMA piece (Klank et al., “CO2-laser micromachining and back-end processing for rapid production of PMMA-based microfluidic systems,” Lab on a Chip 2(4):242-246 (2002), which is hereby incorporated by reference in its entirety). When pressed into the structured piece, the solvent fuses the two pieces together. If this is performed in an 80° C. environment, which is below the thermal distortion range of PMMA not plasticized with solvent, a very uniform seal is made. A spincoater at 4500 rpm for 3 seconds was used in order to have a very thin layer of solvent on the unstructured PMMA. Because the surface of the PMMA is fairly hydrophobic, and most of the solvents are polar, the surface is first treated with O2 plasma in order to oxidize the surface. This has been shown to increase the hydrophilicity of a PMMA surface.
A lab designed O2 plasma unit was built using a tesla coil (Model BD-10A Electro-Technic Products Inc., Chicago, Ill.) which is typically used for activation of PDMS by corona discharge as the power source. The point of the tesla coil was placed through a drilled rubber stopper. The stopper was then placed on a PVC cylinder and the pressure in the cylinder was dropped to below 100 mbar. A ten minute treatment was found to be enough to alter the water contact angle from approximately 60 degrees to 42 degrees using a Tantec CAM-Plus (Schaumburg, Ill.) water contact angle meter.
Acetone was not used as a solvent because it was found to be too volatile and was almost completely volatilized after spinning. A higher molecular weight solvent, 2,4-pentadione was found to have ideal volatility properties (Wang et al., “Towards disposable lab-on-a-chip: Poly(methylmethacrylate) microchip electrophoresis device with electrochemical detection,” Electrophoresis 23(4):596-601 (2002), which is hereby incorporated by reference in its entirety). Following plasma treatment, the unstructured PMMA was placed on the spin coater. Enough solvent was placed on the PMMA to completely cover the surface. After 15 seconds, the PMMA was spun at 1,250 rpm for 6 seconds including ramping time. The PMMA was then removed and clamped together with the structures PMMA piece. The clamped pieces were then placed in an 80° C. oven for one hour.
Once the device has cured, inlet and outlet ports are inserted into the predrilled holes. The tubes were constructed of stainless steel and are held in place by epoxy. The device was again placed in the 80° C. oven for 24 hours to ensure the volatilization of any remaining solvent.
The IDUA will be manufactured using a gold deposition procedure (Goral et al., “Electrochemical microfluidic biosensor for the detection of nucleic acid sequences,’ Lab on a Chip 6(6):414-421 (2006), which is hereby incorporated by reference in its entirety) on a 0.5 mm glass wafer (
Although preferred embodiments have been depicted and described in detail herein, it will be apparent to those skilled in the relevant art that various modifications, additions, substitutions, and the like can be made without departing from the spirit of the invention and these are therefore considered to be within the scope of the invention as defined in the claims which follow.
This application claims the benefit of U.S. Provisional Patent Application Ser. No. 60/689,720, filed Jun. 10, 2005, which is hereby incorporated by reference in its entirety.
The subject matter of this application was made with support from the United States Government under CSRESS Contract No. NYC-123-404 and National Institutes of Health Grant No. 1 R01 HD37109-01A1. The U.S. Government may have certain rights.
Filing Document | Filing Date | Country | Kind | 371c Date |
---|---|---|---|---|
PCT/US06/22638 | 6/9/2006 | WO | 00 | 10/9/2008 |
Number | Date | Country | |
---|---|---|---|
60689720 | Jun 2005 | US |