The present invention generally relates to medical procedures and devices pertaining to heart valves such as replacement techniques and apparatus. More specifically, the invention relates to the replacement of heart valves having various malformations and dysfunctions.
Complications of the mitral valve, which controls the flow of blood from the left atrium into the left ventricle of the human heart, have been known to cause fatal heart failure. In the developed world, one of the most common forms of valvular heart disease is mitral valve leak, also known as mitral regurgitation, which is characterized by the abnormal leaking of blood from the left ventricle through the mitral valve and back into the left atrium. This occurs most commonly due to ischemic heart disease when the leaflets of the mitral valve no longer meet or close properly after multiple infarctions, idiopathic and hypertensive cardiomyopathies where the left ventricle enlarges, and with leaflet and chordal abnormalities, such as those caused by a degenerative disease.
In addition to mitral regurgitation, mitral narrowing or stenosis is most frequently the result of rheumatic disease. While this has been virtually eliminated in developed countries, it is still common where living standards are not as high.
Similar to complications of the mitral valve are complications of the aortic valve, which controls the flow of blood from the left ventricle into the aorta. For example, many older patients develop aortic valve stenosis. Historically, the traditional treatment had been valve replacement by a large open heart procedure. The procedure takes a considerable amount of time for recovery since it is so highly invasive. Fortunately, in the last decade, great advances have been made in replacing this open heart surgery procedure with a catheter procedure that can be performed quickly without surgical incisions or the need for a heart-lung machine to support the circulation while the heart is stopped. Using catheters, valves are mounted on stents or stent-like structures, which are compressed and delivered through blood vessels to the heart. The stents are then expanded and the valves begin to function. The diseased valve is not removed, but instead it is crushed or deformed by the stent which contains the new valve. The deformed tissue serves to help anchor the new prosthetic valve.
Delivery of the valves can be accomplished from arteries which can be easily accessed in a patient. Most commonly this is done from the groin where the femoral and iliac arteries can be cannulated. The shoulder region is also used, where the subclavian and axillary arteries can also be accessed. Recovery from this procedure is remarkably quick.
Not all patients can be served with a pure catheter procedure. In some cases the arteries are too small to allow passage of catheters to the heart, or the arteries are too diseased or tortuous. In these cases, surgeons can make a small chest incision (thoractomy) and then place these catheter-based devices directly into the heart. Typically, a purse string suture is made in the apex of the left ventricle and the delivery system is placed through the apex of the heart. The valve is then delivered into its final position. These delivery systems can also be used to access the aortic valve from the aorta itself. Some surgeons introduce the aortic valve delivery system directly in the aorta at the time of open surgery. The valves vary considerably. There is a mounting structure that is often a form of stent. Prosthetic leaflets are carried inside the stent on mounting and retention structure. Typically, these leaflets are made from biologic material that is used in traditional surgical valves. The valve can be actual heart valve tissue from an animal or more often the leaflets are made from pericardial tissue from cows, pigs or horses. These leaflets are treated to reduce their immunogenicity and improve their durability. Many tissue processing techniques have been developed for this purpose. In the future, biologically engineered tissue may be used or polymers or other non-biologic materials may be used for valve leaflets. All of these can be incorporated into the inventions described in this disclosure.
There are, in fact, more patients with mitral valve disease than aortic valve disease. In the course of the last decade, many companies have been successful in creating catheter or minimally invasive implantable aortic valves, but implantation of a mitral valve is more difficult and to date there has been no good solution. Patients would be benefited by implanting a device by a surgical procedure employing a small incision or by a catheter implantation such as from the groin. From the patient's point of view, the catheter procedure is very attractive. At this time there is no commercially available way to replace the mitral valve with a catheter procedure. Many patients who require mitral valve replacement are elderly and an open heart procedure is painful, risky and takes time for recovery. Some patients are not even candidates for surgery due to advanced age and frailty. Therefore, there exists a particular need for a remotely placed mitral valve replacement device.
While previously, it was thought that mitral valve replacement rather than valve repair was associated with a more negative long-term prognosis for patients with mitral valve disease, this belief has come into question. It is now believed that the outcome for patients with mitral valve leak or regurgitation is almost equal whether the valve is repaired or replaced. Furthermore, the durability of a mitral valve surgical repair is now under question. Many patients, who have undergone repair, redevelop a leak over several years. As many of these are elderly, a repeat intervention in an older patient is not welcomed by the patient or the physicians.
The most prominent obstacle for catheter mitral valve replacement is retaining the valve in position. The mitral valve is subject to a large cyclic load. The pressure in the left ventricle is close to zero before contraction and then rises to the systolic pressure (or higher if there is aortic stenosis) and this can be very high if the patient has systolic hypertension. Often the load on the valve is 150 mmHg or more. Since the heart is moving as it beats, the movement and the load can combine to dislodge a valve. Also, the movement and rhythmic load can fatigue materials leading to fractures of the materials. Thus, there is a major problem associated with anchoring a valve.
Another problem with creating a catheter delivered mitral valve replacement is size. The implant must have strong retention and leak avoidance features and it must contain a valve. Separate prostheses may contribute to solving this problem, by placing an anchor or dock first and then implanting the valve second. However, in this situation, the patient must remain stable between implantation of the anchor or dock and implantation of the valve. If the patient's native mitral valve is rendered non-functional by the anchor or dock, then the patient may quickly become unstable and the operator may be forced to hastily implant the new valve or possibly stabilize the patient by removing the anchor or dock and abandoning the procedure.
Another problem with mitral replacement is leak around the valve, or paravalvular leak. If a good seal is not established around the valve, blood can leak back into the left atrium. This places extra load on the heart and can damage the blood as it travels in jets through sites of leaks. Hemolysis or breakdown of red blood cells is a frequent complication if this occurs. Paravalvular leak was one of the common problems encountered when the aortic valve was first implanted on a catheter. During surgical replacement, a surgeon has a major advantage when replacing the valve as he or she can see a gap outside the valve suture line and prevent or repair it. With catheter insertion, this is not possible. Furthermore, large leaks may reduce a patient's survival and may cause symptoms that restrict mobility and make the patient uncomfortable (e.g., short of breathe, edematous, fatigued). Therefore, devices, systems, and methods which relate to mitral valve replacement should also incorporate means to prevent and repair leaks around the replacement valve.
A patient's mitral valve annulus can also be quite large. When companies develop surgical replacement valves, this problem is solved by restricting the number of sizes of the actual valve produced and then adding more fabric cuff around the margin of the valve to increase the valve size. For example, a patient may have a 45 mm valve annulus. In this case, the actual prosthetic valve diameter may be 30 mm and the difference is made up by adding a larger band of fabric cuff material around the prosthetic valve. However, in catheter procedures, adding more material to a prosthetic valve is problematic since the material must be condensed and retained by small delivery systems. Often, this method is very difficult and impractical, so alternative solutions are necessary.
Since numerous valves have been developed for the aortic position, it is desirable to avoid repeating valve development and to take advantage of existing valves. These valves have been very expensive to develop and bring to market, so extending their application can save considerable amounts of time and money. It would be useful then to create a mitral anchor or docking station for such a valve. An existing valve developed for the aortic position, perhaps with some modification, could then be implanted in the docking station. Some previously developed valves may fit well with no modification, such as the Edwards Sapien™ valve. Others, such as the Corevalve™ may be implantable but require some modification for an optimal engagement with the anchor and fit inside the heart.
A number of further complications may arise from a poorly retained or poorly positioned mitral valve replacement prosthesis. Namely, a valve can be dislodged into the atrium or ventricle, which could be fatal for a patient. Prior prosthetic anchors have reduced the risk of dislodgement by puncturing tissue to retain the prosthesis. However, this is a risky maneuver since the penetration must be accomplished by a sharp object at a long distance, leading to a risk of perforation of the heart and patient injury.
Orientation of the mitral prosthesis is also important. The valve must allow blood to flow easily from the atrium to the ventricle. A prosthesis that enters at an angle may lead to poor flow, obstruction of the flow by the wall of the heart or a leaflet and a poor hemodynamic result. Repeated contraction against the ventricular wall can also lead to rupture of the back wall of the heart and sudden death of the patient.
With surgical mitral valve repair or replacement, sometimes the anterior leaflet of the mitral valve leaflet is pushed into the area of the left ventricular outflow and this leads to poor left ventricular emptying. This syndrome is known as left ventricular tract outflow obstruction. The replacement valve itself can cause left ventricular outflow tract obstruction if it is situated close to the aortic valve.
Yet another obstacle faced when implanting a replacement mitral valve is the need for the patient's native mitral valve to continue to function regularly during placement of the prosthesis so that the patient can remain stable without the need for a heart-lung machine to support circulation.
In addition, it is desirable to provide devices and methods that can be utilized in a variety of implantation approaches. Depending on a particular patient's anatomy and clinical situation, a medical professional may wish to make a determination regarding the optimal method of implantation, such as inserting a replacement valve directly into the heart in an open procedure (open heart surgery or a minimally invasive surgery) or inserting a replacement valve from veins and via arteries in a closed procedure (such as a catheter-based implantation). It is preferable to allow a medical professional a plurality of implantation options to choose from. For example, a medical professional may wish to insert a replacement valve either from the ventricle or from the atrial side of the mitral valve.
Therefore, the present invention provides devices and methods that address these and other challenges in the art.
In one illustrative embodiment, the invention provides a system for replacing a native heart valve including an expansible helical anchor formed as multiple coils adapted to support a heart valve prosthesis. At least one of the coils is normally at a first diameter, and is expandable to a second, larger diameter upon application of radial outward force from within the helical anchor. A gap is defined between adjacent coils sufficient to prevent engagement by at least one of the adjacent coils with the native heart valve. An expansible heart valve prosthesis is provided and is configured to be delivered into the helical anchor and expanded inside the multiple coils into engagement with the at least one coil. This moves at least that coil from the first diameter to the second diameter while securing the helical anchor and the heart valve prosthesis together. The system further includes a seal on the expansible heart valve prosthesis configured to engage the helical anchor and prevent blood leakage past the heart valve prosthesis after implantation of the heart valve prosthesis in the helical anchor.
The system may include one or more additional aspects. For example, the helical anchor may include another coil that moves from a larger diameter to a smaller diameter as the heart valve prosthesis is expanded inside the multiple coils. The seal may take many alternative forms. For example, the seal can include portions extending between adjacent coils for preventing blood leakage through the helical anchor and past the heart valve prosthesis. The seal may be comprised of many different alternative materials. The seal may further comprise a membrane or panel extending between at least two coils of the helical anchor after implantation of the heart valve prosthesis in the helical anchor. For example, one example is a biologic material. The helical anchor may further comprise a shape memory material. The heart valve prosthesis includes a blood inflow end and a blood outflow end and at least one of the ends may be unflared and generally cylindrical in shape. In an illustrative embodiment, the blood outflow end is flared radially outward and includes a bumper for preventing damage to tissue structure in the heart after implantation. The gap may be formed by a coil portion of the helical anchor that extends non-parallel to adjacent coil portions of the helical anchor.
In another illustrative embodiment, a system is provided as generally described above, except that the seal is alternatively or additionally carried on the helical anchor instead of being carried on the heart valve prosthesis. Any other features as described or incorporated herein may be included.
In another illustrative embodiment, a system for docking a heart valve prosthesis includes a helical anchor formed as multiple coils adapted to support a heart valve prosthesis with coil portions positioned above and/or below the heart valve annulus. An outer, flexible and helical tube carries the coils of the helical anchor to form an assembly. A helical delivery tool carries the assembly and is adapted to be rotated into position through a native heart valve. Additional or optional features may be provided. For example, a heart valve prosthesis may be expanded inside the multiple coils. The outer tube may be formed from a low friction material adapted to slide off of the multiple coils of the helical anchor after rotating into position through the native heart valve. The outer tube may be secured to the helical delivery tool with suture or by any other method. The helical delivery tool may formed with a plurality of coils, and the outer tube may further be secured to the distal end. The distal end may further comprise a bullet or tapered shape to assist with delivery. The distal end can further comprise a resilient element, and the distal ends of the outer tube and the helical delivery tube are secured to the resilient element.
In another illustrative embodiment, a system for replacing a native heart valve includes a helical anchor formed as multiple coils adapted to support a heart valve prosthesis at the native heart valve. An expansible heart valve prosthesis is provided in this system and is capable of being delivered into the helical anchor and expanded inside the multiple coils into engagement with the at least one coil to secure the helical anchor and the heart valve prosthesis together. A guide structure on the expansible heart valve prosthesis is configured to guide the helical anchor into position as the helical anchor is extruded from a helical anchor delivery catheter.
The guide structure may further comprise an opening within a portion of the expansible heart valve prosthesis, such as an opening in a loop, a tube or simply an opening in the stent structure of the expansible heart valve prosthesis, for example. The opening may be configured to receive a helical anchor delivery catheter that carries the helical anchor during the implantation procedure. The opening may be located on an arm of the expansible heart valve prosthesis and the prosthesis may further comprise a plurality of arms configured to engage beneath the native heart valve. The guide structure may further comprise a tubular arm of the expansible heart valve prosthesis.
In another illustrative embodiment, a system for docking a mitral valve prosthesis and replacing a native mitral valve is provided and includes a coil guide catheter and a helical anchor adapted to be received in and delivered from the coil guide catheter. The helical anchor is formed as multiple coils having a coiled configuration after being delivered from the coil guide catheter and adapted to support the mitral valve prosthesis upon being fully delivered from the coil guide catheter and implanted at the native mitral valve. The system further includes a tissue gathering catheter including loop structure configured to be deployed to surround and gather the native chordea tendinae for allowing easier direction of the helical anchor in the left ventricle.
In another illustrative embodiment, an anchor for docking a heart valve prosthesis includes an upper helical coil portion, a lower helical coil portion, and a fastener securing the upper helical coil portion to the lower helical coil portion.
In another illustrative embodiment, a method of implanting a heart valve prosthesis in the heart of a patient includes holding a helical anchor in the form of multiple coils within an outer, flexible tube. The assembly of the outer, flexible tube and the helical anchor is secured to a helical delivery tool. The helical delivery tool is rotated adjacent to a native heart valve of the patient to position the assembly on either or both sides of the native heart valve. The assembly is removed from the helical delivery tool, and the outer tube is removed from the helical anchor. The heart valve prosthesis is then implanted within the helical anchor.
Securing the assembly may further comprise positioning coils of the assembly generally along adjacent coils of the helical delivery tool. Removing the outer tube may further comprise holding the helical anchor with a pusher element, and pulling the outer tube off the helical anchor.
In another illustrative embodiment, a method of implanting an expansible heart valve prosthesis in the heart of a patient includes delivering an expansible helical anchor in the form of multiple coils proximate the native heart valve. The expansible heart valve prosthesis is positioned within the multiple coils of the expansible helical anchor with the expansible heart valve prosthesis and the expansible helical anchor in unexpanded states. The expansible heart valve prosthesis is expanded against the expansible helical anchor thereby expanding the expansible heart valve prosthesis while securing the expansible heart valve prosthesis to the expansible helical anchor. A seal is carried on the helical anchor and/or on the heart valve prosthesis and extends between at least two adjacent coils for preventing blood leakage through the helical anchor and past the heart valve prosthesis.
In another illustrative embodiment, a method of implanting an expansible heart valve prosthesis to replace a native heart valve of a patient includes delivering a helical anchor in the form of multiple coils proximate the native heart valve. The expansible heart valve prosthesis is delivered proximate the native heart valve. The helical anchor is guided generally around a periphery of the expansible heart valve prosthesis using guide structure carried on the expansible heart valve prosthesis. The expansible heart valve prosthesis is expanded against the helical anchor. As discussed above, the guide structure may take many different forms.
In another illustrative embodiment, a method of implanting a helical anchor for docking a mitral heart valve prosthesis in a patient includes gathering the chordea tendinae using a tissue gathering catheter. A helical anchor is then delivered in the form of multiple coils proximate a native heart valve and around the gathered chordae tendinae.
In another illustrative embodiment, a method of implanting a helical anchor for docking a heart valve prosthesis in a patient includes delivering an upper helical anchor portion comprised of upper coils to a position above a native heart valve, and delivering a lower helical anchor portion comprised of lower coils to a position below the native heart valve. The upper and lower helical anchor portions are secured together with a fastener either before or after delivery of each helical anchor portion.
In another illustrative embodiment, a system for replacing a native heart valve is provided and includes an expansible helical anchor formed as multiple coils adapted to support a heart valve prosthesis. At least one of the coils is normally at a first diameter, and is expandable to a second, larger diameter upon application of radial outward force from within the helical anchor. A gap is defined between adjacent coils sufficient to prevent engagement by at least one of the adjacent coils with the native heart valve. An expansible heart valve prosthesis is provided and is capable of being delivered into the helical anchor and expanded inside the multiple coils into engagement with the at least one coil. In this manner, the expansible coil moves from the first diameter to the second diameter while securing the helical anchor and the heart valve prosthesis together. The expansible heart valve prosthesis includes an inflow end and an outflow end. The inflow end is unflared and generally cylindrical, while the outflow end is flared in a radially outward direction.
Various additional advantages, methods, devices, systems and features will become more readily apparent to those of ordinary skill in the art upon review of the following detailed description of the illustrative embodiments taken in conjunction with the accompanying drawings.
It will be appreciated that like reference numerals throughout this description and the drawings refer generally to like elements of structure and function. The differences between embodiments will be apparent from the drawings and/or from the description and/or the use of different reference numerals in different figures. For clarity and conciseness, description of like elements will not be repeated throughout the description.
Referring first to
Now referring to
A procedure will now be described for introducing or implanting a helical anchor 12 in connection with
In another embodiment shown in
Referring to
When the helical anchor 12 is advanced or extruded as is initially shown in
Other mitral valve prosthesis or replacement valves may be used and have a wide range of attachment arms or wings, or stent structure, that wrap around the native mitral valve leaflets 18, 20. The arms or other similar structures in such prostheses could all be fitted with a loop 120, or tube or other similar guidance structure, to perform similar functions as the loop 120 described immediately above. This function generally relates to directing the delivery of the helical anchor 12. Furthermore, it is not necessary that a loop 120 directs the helical anchor delivery. For example, a cell or opening of the replacement valve stent structure 86 could also perform the same function as the loop 120 shown and described in these figures. A hook or a tube may also be used in lieu of the illustrated loop 120. Any structure that can function to direct the helical anchor 12 around the native mitral valve leaflets 18, 20 may be added to the prosthetic or replacement heart valve 60. The structure may be permanently fabricated as part of the replacement valve 60 or may be temporary structure used only during the procedure. For example, a loop of suture (not shown) may be used to guide delivery of a helical anchor 12 including any helical anchor delivery catheter 10 associated therewith. After use of the suture, it may be withdrawn from the patient.
The arms 110 illustrated in these figures are quite narrow or slender. In practice, it may be more useful to have arms that are composed of pairs or triplets of wires that are fused at the ends. The narrow terminal ends of the arms 110 facilitate the arms 110 passing between the chordae tendinae 24 at their margins with the free edge of the native mitral leaflets 18, 20 to allow the arms 110 to wrap around the native leaflets 18, 20. The chordae 24 are closely packed in some areas and slender arms 110 will allow the arms 110 to pass between the chordae tendinae 24. Once the slender portion of the arms 110 pass, thicker portions of the arms 110 may move between the chordae 24 by spreading them apart. Therefore, an arm 110 that is slender or composed of a single wire or fusion of wires at the tip and that is more robust or thicker closer to the main body of the prosthetic or replacement valve 60, may be a desirable arrangement. The wires or arms 110 may also be much shorter than those shown in these illustrative figures. In the illustrated method, delivery of the helical anchor 12 may be started at any desired location and not necessarily at the commissure 14 of the native mitral valve 16. For example, delivery may start in the middle portion of a native mitral leaflet 18 or 20. This would be advantageous for the surgeon who would not have to precisely locate the commissure 14 to begin the procedure, thereby greatly simplifying the procedure.
In clinical practice, there are virtually limitless variations in the size of the native mitral leaflets 18, 20, character of the native mitral leaflets 18, 20, the chordal lengths and the attachment of the chordae 24 as well as the diameter of the mitral annulus 126. The use of a helical anchor 12 or other anchor structure under the native leaflets 18, 20 neutralizes many of these variables since the fixation point of the arms 110 may be brought to the lowest coil 22 of the helical anchor 12. This position may also be determined in advance by selecting the number of coils 22 in the helical anchor 12 as well as the thickness of the coils 22 in the helical anchor 12 to match the turning point of the arms 110 on the lowest portion of the replacement valve 60. Thus, an important feature of the helical anchor 12 delivered under the native mitral annulus 126 is that it can create a common and predefined plane for anchoring the arms 110 of the replacement valve 60. In the situation described above in which some of the chordae 24 are stretched, the attachment in this region of the replacement valve 60 could be to the helical anchor 12. This would create a common plane for the lowest point on the replacement valve 60. To ensure that the valve 60 anchors at a common lowest plane throughout its perimeter, additional coils 22 may be added to the helical anchor 12, or the diameter of the coils 22 may be made larger. Additional options are, for example, waves or undulations may be added to the coils 22 of the helical anchor 12 to expand the overall height of the helical anchor 12. The helical anchor 12 therefore improves stability of the replacement valve 60 by providing an anchoring point or location for the arms of the replacement valve 60 to wrap around while, at the same time, the helical anchor 12 can trap the perimeter of the replacement valve 60 along its length. The combination of these features provides for increased stability to the replacement valve 60 and can also seal the replacement valve 60 against the native mitral valve 16 to prevent perivalvular leakage of blood flow. As mentioned, the native mitral valve and heart structure of patients comes in many varieties and combinations. It is not practical for a manufacturer to make different lengths and depths of anchoring arms 110 and for the user to deliver these products optimally into position for each case. Rather, it is much more practical to adjust for these variations by placing a helical anchor 12 below the native mitral valve 16 and using this to create a lowest plane for the arms 110 to anchor against. The delivery system for the helical anchor 12 may be any delivery or deployment system, for example, described in the above-incorporated PCT applications. It will be appreciated that such deployment methods and apparatus may be used to deliver the helical anchor 12 such that the anchor 12 is positioned only below the native mitral valve 16 as shown herein.
While the present invention has been illustrated by a description of preferred embodiments and while these embodiments have been described in some detail, it is not the intention of the Applicants to restrict or in any way limit the scope of the appended claims to such detail. Additional advantages and modifications will readily appear to those skilled in the art. The various features and concepts of the invention may be used alone or in any combination depending on the needs and preferences of the operator. This has been a description of the present invention, along with the preferred methods of practicing the present invention as currently known. However, the invention itself should only be defined by the appended claims.
This application is a divisional of U.S. patent application Ser. No. 14/912,067, filed Feb. 12, 2016 and issued as U.S. Pat. No. 10,226,330, which is a U.S. national phase application of PCT patent application PCT/US2014/051095, filed Aug. 14, 2014, which claims the priority of U.S. Provisional Application Ser. No. 61/865,657, filed Aug. 14, 2013; U.S. Provisional Application Ser. No. 61/942,300, filed Feb. 20, 2014; and U.S. Provisional Application Ser. No. 61/943,125, filed Feb. 21, 2014, the disclosures of each of the foregoing are hereby incorporated by reference herein.
Number | Name | Date | Kind |
---|---|---|---|
3564617 | Sauvage et al. | Feb 1971 | A |
3755823 | Hancock | Sep 1973 | A |
4035849 | Angell et al. | Jul 1977 | A |
4490859 | Black et al. | Jan 1985 | A |
4512338 | Balko et al. | Apr 1985 | A |
4759758 | Gabbay | Jul 1988 | A |
4777951 | Cribier et al. | Oct 1988 | A |
4787899 | Lazarus | Nov 1988 | A |
4790843 | Carpentier et al. | Dec 1988 | A |
4796629 | Grayzel | Jan 1989 | A |
4856516 | Hillstead | Aug 1989 | A |
4878495 | Grayzel | Nov 1989 | A |
4966604 | Reiss | Oct 1990 | A |
4994077 | Dobben | Feb 1991 | A |
5059177 | Towne et al. | Oct 1991 | A |
5192297 | Hull | Mar 1993 | A |
5282847 | Trescony et al. | Feb 1994 | A |
5370685 | Stevens | Dec 1994 | A |
5403305 | Sauter et al. | Apr 1995 | A |
5411552 | Andersen et al. | May 1995 | A |
5443500 | Sigwart | Aug 1995 | A |
5545214 | Stevens | Aug 1996 | A |
5554185 | Block et al. | Sep 1996 | A |
5558644 | Boyd et al. | Sep 1996 | A |
5584803 | Stevens et al. | Dec 1996 | A |
5591195 | Taheri et al. | Jan 1997 | A |
5607464 | Trescony et al. | Mar 1997 | A |
5665115 | Cragg | Sep 1997 | A |
5769812 | Stevens et al. | Jun 1998 | A |
5800508 | Goicoechea et al. | Sep 1998 | A |
5840081 | Andersen et al. | Nov 1998 | A |
5855597 | Jayaraman | Jan 1999 | A |
5855601 | Bessler et al. | Jan 1999 | A |
5925063 | Khosravi | Jul 1999 | A |
5957949 | Leonhardt et al. | Sep 1999 | A |
6027525 | Suh et al. | Feb 2000 | A |
6092520 | Hasegawa | Jul 2000 | A |
6120534 | Ruiz | Sep 2000 | A |
6168614 | Andersen et al. | Jan 2001 | B1 |
6221091 | Khosravi | Apr 2001 | B1 |
6235042 | Katzman | May 2001 | B1 |
6245102 | Jayaraman | Jun 2001 | B1 |
6302906 | Goicoechea | Oct 2001 | B1 |
6306141 | Jervis | Oct 2001 | B1 |
6406492 | Lytle | Jun 2002 | B1 |
6409758 | Stobie et al. | Jun 2002 | B2 |
6419696 | Ortiz et al. | Jul 2002 | B1 |
6425916 | Garrison et al. | Jul 2002 | B1 |
6432134 | Anson et al. | Aug 2002 | B1 |
6454799 | Schreck | Sep 2002 | B1 |
6458153 | Bailey et al. | Oct 2002 | B1 |
6461382 | Cao | Oct 2002 | B1 |
6482228 | Norred | Nov 2002 | B1 |
6527979 | Constantz | Mar 2003 | B2 |
6582462 | Andersen et al. | Jun 2003 | B1 |
6625578 | Spaur et al. | Sep 2003 | B2 |
6652578 | Bailey et al. | Nov 2003 | B2 |
6663322 | Listle | Dec 2003 | B1 |
6730118 | Spenser et al. | May 2004 | B2 |
6730121 | Ortiz et al. | May 2004 | B2 |
6733525 | Yang et al. | May 2004 | B2 |
6767362 | Schreck | Jul 2004 | B2 |
6797002 | Spence et al. | Sep 2004 | B2 |
6830584 | Seguin | Dec 2004 | B1 |
6878162 | Bales et al. | Apr 2005 | B2 |
6893460 | Spenser et al. | May 2005 | B2 |
6908481 | Cribier | Jun 2005 | B2 |
6971998 | Rosenman et al. | Dec 2005 | B2 |
7018406 | Seguin et al. | Mar 2006 | B2 |
7018408 | Bailey et al. | Mar 2006 | B2 |
7037334 | Hlavka et al. | May 2006 | B1 |
7077861 | Spence | Jul 2006 | B2 |
7101395 | Tremulis et al. | Sep 2006 | B2 |
7125421 | Tremulis et al. | Oct 2006 | B2 |
7166126 | Spence et al. | Jan 2007 | B2 |
7166127 | Spence et al. | Jan 2007 | B2 |
7276078 | Spenser et al. | Oct 2007 | B2 |
7276084 | Yang et al. | Oct 2007 | B2 |
7318278 | Zhang et al. | Jan 2008 | B2 |
7374571 | Pease et al. | May 2008 | B2 |
7393360 | Spenser et al. | Jul 2008 | B2 |
7404824 | Webler et al. | Jul 2008 | B1 |
7431726 | Spence et al. | Oct 2008 | B2 |
7445632 | McGuckin, Jr. et al. | Nov 2008 | B2 |
7462191 | Spenser et al. | Dec 2008 | B2 |
7510575 | Spenser et al. | Mar 2009 | B2 |
7527646 | Randert et al. | May 2009 | B2 |
7585321 | Cribier | Sep 2009 | B2 |
7618446 | Andersen et al. | Nov 2009 | B2 |
7618449 | Tremulis et al. | Nov 2009 | B2 |
7637946 | Solem et al. | Dec 2009 | B2 |
7655034 | Mitchell et al. | Feb 2010 | B2 |
7708775 | Rowe et al. | May 2010 | B2 |
7737060 | Strickler et al. | Jun 2010 | B2 |
7758639 | Mathis | Jul 2010 | B2 |
7780726 | Seguin | Aug 2010 | B2 |
7785366 | Maurer et al. | Aug 2010 | B2 |
7942927 | Kaye et al. | May 2011 | B2 |
7951195 | Antonsson et al. | May 2011 | B2 |
7955385 | Crittenden | Jun 2011 | B2 |
7959672 | Saiahieh et al. | Jun 2011 | B2 |
7967857 | Lane | Jun 2011 | B2 |
7993394 | Hariton et al. | Aug 2011 | B2 |
8016882 | Macoviak et al. | Sep 2011 | B2 |
8029556 | Rowe | Oct 2011 | B2 |
8052750 | Tuval et al. | Nov 2011 | B2 |
8128691 | Keranen | Mar 2012 | B2 |
8142492 | Forster et al. | Mar 2012 | B2 |
8167932 | Bourang et al. | May 2012 | B2 |
8236049 | Rowe et al. | Aug 2012 | B2 |
8323335 | Rowe et al. | Dec 2012 | B2 |
8377115 | Thompson | Feb 2013 | B2 |
8398708 | Meiri et al. | Mar 2013 | B2 |
8449605 | Lichtenstein et al. | May 2013 | B2 |
8449606 | Eliasen et al. | May 2013 | B2 |
8652203 | Quadri et al. | Feb 2014 | B2 |
8657872 | Seguin | Feb 2014 | B2 |
8672998 | Lichtenstein et al. | Mar 2014 | B2 |
8685086 | Navia et al. | Apr 2014 | B2 |
8734507 | Keranen | May 2014 | B2 |
8795352 | O'Beirne et al. | Aug 2014 | B2 |
8801776 | House et al. | Aug 2014 | B2 |
8986373 | Chau et al. | Mar 2015 | B2 |
9078747 | Conklin | Jul 2015 | B2 |
9095434 | Rowe | Aug 2015 | B2 |
9119718 | Keranen | Sep 2015 | B2 |
9180006 | Keranen | Nov 2015 | B2 |
9192471 | Bolling | Nov 2015 | B2 |
9237886 | Seguin et al. | Jan 2016 | B2 |
9314335 | Konno | Apr 2016 | B2 |
9364326 | Yaron | Jun 2016 | B2 |
9463268 | Spence | Oct 2016 | B2 |
9474599 | Keranen | Oct 2016 | B2 |
9597205 | Tuval | Mar 2017 | B2 |
9622863 | Karapetian et al. | Apr 2017 | B2 |
9700409 | Braido et al. | Jul 2017 | B2 |
20020026094 | Roth | Feb 2002 | A1 |
20020032481 | Gabbay | Mar 2002 | A1 |
20020045936 | Moe | Apr 2002 | A1 |
20020055773 | Campbell et al. | May 2002 | A1 |
20020107535 | Wei et al. | Aug 2002 | A1 |
20020138135 | Duerig et al. | Sep 2002 | A1 |
20020151970 | Garrison et al. | Oct 2002 | A1 |
20020173841 | Ortiz et al. | Nov 2002 | A1 |
20030050694 | Yang et al. | Mar 2003 | A1 |
20030100939 | Yodfat et al. | May 2003 | A1 |
20030158597 | Quiachon et al. | Aug 2003 | A1 |
20030167089 | Lane | Sep 2003 | A1 |
20030212454 | Scott et al. | Nov 2003 | A1 |
20030225420 | Wardle | Dec 2003 | A1 |
20040111006 | Alferness et al. | Jun 2004 | A1 |
20040167620 | Ortiz et al. | Aug 2004 | A1 |
20040186563 | Lobbi | Sep 2004 | A1 |
20040186565 | Schreck | Sep 2004 | A1 |
20040260389 | Case et al. | Dec 2004 | A1 |
20050096736 | Osse et al. | May 2005 | A1 |
20050119682 | Nguyen et al. | Jun 2005 | A1 |
20050119735 | Spence et al. | Jun 2005 | A1 |
20050137691 | Salahien et al. | Jun 2005 | A1 |
20050149178 | Spence | Jul 2005 | A1 |
20050182486 | Gabbay | Aug 2005 | A1 |
20050203614 | Forster et al. | Sep 2005 | A1 |
20050203617 | Forster et al. | Sep 2005 | A1 |
20050228496 | Mensah et al. | Oct 2005 | A1 |
20050234546 | Nugent et al. | Oct 2005 | A1 |
20060020327 | Lashinski et al. | Jan 2006 | A1 |
20060025857 | Bergheim et al. | Feb 2006 | A1 |
20060033542 | Lin et al. | Feb 2006 | A1 |
20060052867 | Revuelta et al. | Mar 2006 | A1 |
20060058872 | Salahieh et al. | Mar 2006 | A1 |
20060149350 | Patel et al. | Jul 2006 | A1 |
20060195134 | Crittenden | Aug 2006 | A1 |
20060195185 | Lane et al. | Aug 2006 | A1 |
20060265056 | Nguyen et al. | Nov 2006 | A1 |
20070005131 | Taylor | Jan 2007 | A1 |
20070010677 | Saiahieh et al. | Jan 2007 | A1 |
20070010876 | Saiahieh et al. | Jan 2007 | A1 |
20070112422 | Dehdashtian | May 2007 | A1 |
20070185572 | Solem et al. | Aug 2007 | A1 |
20070203503 | Saiahieh et al. | Aug 2007 | A1 |
20070203575 | Forster et al. | Aug 2007 | A1 |
20070213813 | Von Segesser et al. | Sep 2007 | A1 |
20070232898 | Huynh et al. | Oct 2007 | A1 |
20070244553 | Rafiee et al. | Oct 2007 | A1 |
20070265700 | Ellasen et al. | Nov 2007 | A1 |
20070293808 | Williams et al. | Dec 2007 | A1 |
20080004697 | Lichtenstein et al. | Jan 2008 | A1 |
20080021547 | Davidson | Jan 2008 | A1 |
20080077235 | Kirson | Mar 2008 | A1 |
20080114442 | Mitchell et al. | May 2008 | A1 |
20080125853 | Bailey et al. | May 2008 | A1 |
20080200977 | Paul et al. | Aug 2008 | A1 |
20080200980 | Robin | Aug 2008 | A1 |
20080208330 | Keranen | Aug 2008 | A1 |
20080243245 | Thambar et al. | Oct 2008 | A1 |
20080275503 | Spence et al. | Nov 2008 | A1 |
20090005863 | Goetz et al. | Jan 2009 | A1 |
20090088836 | Bishop et al. | Apr 2009 | A1 |
20090099653 | Suri et al. | Apr 2009 | A1 |
20090157175 | Benichou | Jun 2009 | A1 |
20090177278 | Spence | Jul 2009 | A1 |
20090192601 | Rafiee et al. | Jul 2009 | A1 |
20090234318 | Loulmet et al. | Sep 2009 | A1 |
20090259307 | Gross et al. | Oct 2009 | A1 |
20090276038 | Tremulis et al. | Nov 2009 | A1 |
20090276040 | Rowe et al. | Nov 2009 | A1 |
20090281619 | Le et al. | Nov 2009 | A1 |
20090299471 | Keranen | Dec 2009 | A1 |
20090319037 | Rowe et al. | Dec 2009 | A1 |
20100036484 | Hariton et al. | Feb 2010 | A1 |
20100049313 | Alon et al. | Feb 2010 | A1 |
20100076548 | Konno | Mar 2010 | A1 |
20100076549 | Keidar et al. | Mar 2010 | A1 |
20100145440 | Keranen | Jun 2010 | A1 |
20100152639 | Shandas et al. | Jun 2010 | A1 |
20100161047 | Cabiri | Jun 2010 | A1 |
20100168844 | Toomes | Jul 2010 | A1 |
20100198347 | Zakay et al. | Aug 2010 | A1 |
20100217382 | Chau et al. | Aug 2010 | A1 |
20100217385 | Thompson et al. | Aug 2010 | A1 |
20100312333 | Navia et al. | Dec 2010 | A1 |
20100318183 | Keranen | Dec 2010 | A1 |
20100318184 | Spence | Dec 2010 | A1 |
20100331971 | Keranen | Dec 2010 | A1 |
20100331973 | Keranen | Dec 2010 | A1 |
20110015729 | Jimenez et al. | Jan 2011 | A1 |
20110098802 | Braido et al. | Apr 2011 | A1 |
20110106247 | Miller et al. | May 2011 | A1 |
20110137397 | Chau et al. | Jun 2011 | A1 |
20110208298 | Tuval et al. | Jun 2011 | A1 |
20110178597 | Navia et al. | Jul 2011 | A9 |
20110196480 | Cartledge | Aug 2011 | A1 |
20110208297 | Tuval et al. | Aug 2011 | A1 |
20110218621 | Antonsson et al. | Sep 2011 | A1 |
20110224785 | Hacohen | Sep 2011 | A1 |
20110245911 | Quill et al. | Oct 2011 | A1 |
20110288634 | Tuval et al. | Nov 2011 | A1 |
20110295361 | Claiborne, III et al. | Dec 2011 | A1 |
20110319989 | Lane et al. | Dec 2011 | A1 |
20110319990 | Macoviak et al. | Dec 2011 | A1 |
20120016464 | Seguin | Jan 2012 | A1 |
20120022629 | Perera | Jan 2012 | A1 |
20120022633 | Olson et al. | Jan 2012 | A1 |
20120053680 | Boiling et al. | Mar 2012 | A1 |
20120059458 | Buchbinder et al. | Mar 2012 | A1 |
20120123529 | Levi et al. | May 2012 | A1 |
20120150287 | Forster et al. | Jun 2012 | A1 |
20120215236 | Matsunaga et al. | Aug 2012 | A1 |
20120259409 | Nguyen et al. | Oct 2012 | A1 |
20120283820 | Meng et al. | Nov 2012 | A1 |
20120310328 | Olson et al. | Dec 2012 | A1 |
20120316643 | Keranen | Dec 2012 | A1 |
20120323316 | Chau et al. | Dec 2012 | A1 |
20130006352 | Yaron | Jan 2013 | A1 |
20130023985 | Khairkhahan et al. | Jan 2013 | A1 |
20130114214 | Takeguchi et al. | May 2013 | A1 |
20130190857 | Mitra et al. | Jul 2013 | A1 |
20130190865 | Anderson | Jul 2013 | A1 |
20130274873 | Delaloye et al. | Oct 2013 | A1 |
20130310917 | Richter et al. | Nov 2013 | A1 |
20130310928 | Morriss et al. | Nov 2013 | A1 |
20130317598 | Rowe et al. | Nov 2013 | A1 |
20130325114 | McLean et al. | Dec 2013 | A1 |
20130331929 | Mitra | Dec 2013 | A1 |
20140067054 | Chau et al. | Mar 2014 | A1 |
20140074299 | Endou et al. | Mar 2014 | A1 |
20140081394 | Keranen et al. | Mar 2014 | A1 |
20140155997 | Braido | Jun 2014 | A1 |
20140172070 | Seguin | Jun 2014 | A1 |
20140194981 | Menk et al. | Jul 2014 | A1 |
20140200661 | Pintor et al. | Jul 2014 | A1 |
20140236287 | Clague et al. | Aug 2014 | A1 |
20140277417 | Schraut et al. | Sep 2014 | A1 |
20140277419 | Garde et al. | Sep 2014 | A1 |
20140277424 | Oslund | Sep 2014 | A1 |
20140330372 | Weston et al. | Nov 2014 | A1 |
20140343671 | Yohanan et al. | Nov 2014 | A1 |
20140350667 | Braido et al. | Nov 2014 | A1 |
20140358222 | Gorman, III et al. | Dec 2014 | A1 |
20140358224 | Tegels et al. | Dec 2014 | A1 |
20140379074 | Spence et al. | Dec 2014 | A1 |
20140379076 | Vidlund et al. | Dec 2014 | A1 |
20150025623 | Granada et al. | Jan 2015 | A1 |
20150073545 | Braido | Mar 2015 | A1 |
20150073546 | Braido | Mar 2015 | A1 |
20150230921 | Chau et al. | Aug 2015 | A1 |
20150245910 | Righini et al. | Sep 2015 | A1 |
20150282931 | Brunnett et al. | Oct 2015 | A1 |
20150335428 | Keranen | Nov 2015 | A1 |
20150335430 | Loulmet et al. | Nov 2015 | A1 |
20150374493 | Yaron et al. | Dec 2015 | A1 |
20160015514 | Lashinski et al. | Jan 2016 | A1 |
20160074165 | Spence et al. | Mar 2016 | A1 |
20160095705 | Keranen et al. | Apr 2016 | A1 |
20160143732 | Giimsdaie | May 2016 | A1 |
20160184095 | Spence et al. | Jun 2016 | A1 |
20160199177 | Spence et al. | Jul 2016 | A1 |
20160256276 | Yaron | Sep 2016 | A1 |
20160346080 | Righini et al. | Dec 2016 | A1 |
20170007399 | Keranen | Jan 2017 | A1 |
20170007402 | Zerkowski et al. | Jan 2017 | A1 |
20170217385 | Rinkleff et al. | Aug 2017 | A1 |
20170266005 | McGuckin, Jr. | Sep 2017 | A1 |
20170273788 | O'Carroll et al. | Sep 2017 | A1 |
20170273789 | Yaron et al. | Sep 2017 | A1 |
20170281337 | Campbell | Oct 2017 | A1 |
20180000580 | Wallace et al. | Jan 2018 | A1 |
20180085217 | Lashinski et al. | Mar 2018 | A1 |
20180206074 | Tanasa et al. | Jul 2018 | A1 |
20180289481 | Dolan | Oct 2018 | A1 |
20180303606 | Rothstein et al. | Oct 2018 | A1 |
20180318073 | Tseng et al. | Nov 2018 | A1 |
20180318080 | Quill et al. | Nov 2018 | A1 |
Number | Date | Country |
---|---|---|
1684644 | Oct 2005 | CN |
1714766 | Jan 2006 | CN |
101588771 | Nov 2009 | CN |
102670332 | Sep 2012 | CN |
19532846 | Mar 1997 | DE |
19546692 | Jun 1997 | DE |
19857887 | Jul 2000 | DE |
19907646 | Aug 2000 | DE |
0103546 | Mar 1984 | EP |
0597967 | May 1994 | EP |
0592410 | Oct 1995 | EP |
0850607 | Jul 1998 | EP |
1432369 | Jun 2004 | EP |
1521550 | Apr 2005 | EP |
1796597 | Jun 2007 | EP |
1296618 | Jan 2008 | EP |
2072027 | Jun 2009 | EP |
1827314 31 | Dec 2010 | EP |
2620125 | Jul 2013 | EP |
2726018 | May 2014 | EP |
2806829 | Dec 2014 | EP |
2788217 | Jul 2000 | FR |
2815844 | May 2002 | FR |
2004502494 | Jan 2004 | JP |
2008531185 | Aug 2008 | JP |
1271508 | Nov 1986 | SU |
9117720 | Nov 1991 | WO |
9301768 | Feb 1993 | WO |
9829057 | Jul 1998 | WO |
9940964 | Aug 1999 | WO |
9947075 | Sep 1999 | WO |
0041652 | Jul 2000 | WO |
0047139 | Aug 2000 | WO |
0149213 | Jul 2001 | WO |
0154625 | Aug 2001 | WO |
0162189 | Aug 2001 | WO |
0164137 | Sep 2001 | WO |
0176510 | Oct 2001 | WO |
0203892 | Jan 2002 | WO |
0222054 | Mar 2002 | WO |
0236048 | May 2002 | WO |
0247575 | Jun 2002 | WO |
03028558 | Apr 2003 | WO |
03047468 | Jun 2003 | WO |
2005034812 | Apr 2005 | WO |
2005084595 | Sep 2005 | WO |
2006011127 | Feb 2006 | WO |
2006091163 | Aug 2006 | WO |
2006111391 | Oct 2006 | WO |
2006138173 | Dec 2006 | WO |
2005102015 | Apr 2007 | WO |
2007047488 | Apr 2007 | WO |
2007067942 | Jun 2007 | WO |
2007097983 | Aug 2007 | WO |
2008005405 | Jan 2008 | WO |
2008015257 | Feb 2008 | WO |
2008058940 | May 2008 | WO |
2008091515 | Jul 2008 | WO |
2009033469 | Mar 2009 | WO |
2009080801 | Jul 2009 | WO |
2009134701 | Nov 2009 | WO |
2009155561 | Dec 2009 | WO |
2010057262 | May 2010 | WO |
2010121076 | Oct 2010 | WO |
2012063228 | May 2012 | WO |
2013001339 | Jan 2013 | WO |
2013068542 | May 2013 | WO |
2013110722 | Aug 2013 | WO |
2013114214 | Aug 2013 | WO |
2015023579 | Feb 2015 | WO |
2015023862 | Feb 2015 | WO |
2015125024 | Aug 2015 | WO |
2015127264 | Aug 2015 | WO |
2015198125 | Dec 2015 | WO |
2016038017 | Mar 2016 | WO |
2016040881 | Mar 2016 | WO |
2016130820 | Aug 2016 | WO |
2017103833 | Jun 2017 | WO |
Entry |
---|
Kempfert et al., “Minimallly invasive off-pump valve-in-a-ring implantation: the atrial transcatheter approach for re-operative mitral valve replacement after failed repair,” European Journal of Cardiothoracic Surgery, 2009, 35:965-969. |
Webb et al., “Transcatheter Valve-in-Valve Implantation for Failed Bioprosthetic Heart Valves,” Journal of the American Heart Association, 11, Apr. 27, 2010. |
Webb et al., “Mitral Valve in Valve,” TCT Sep. 2009, Live Case: 30 Minutes, St. Paul's Hospital/University of British Columbia. |
Wenaweser et al., “Percutaneous Aortic Valve Replacement for Severe Aortic Regurgitation in Degenerated Bioprosthesis: The First Valve Procedure Using Corevalve Revaiving System,” Catheterization and Cardiovascular Interventions, 70:760-764, 2007. |
Cheung et al,“Transapicai Transcatheter Mitrai Valve-in-Valve Implantation in a Human,” The Society of Thoracic Surgeons, 2009. |
Cheung et al, Live Case Transmissions, NYHA III CHF, Case Summary, Sep. 23, 2010, St. Paul's Hospital/University of British Columbia. |
Shuto et al., “Percutaneous Transvenous Melody Valve-in-Ring Procedure for Mitral Valve Replacement,” J AM Coll Cardiol, 58(24): 2475-2480, 2011. |
Descoutures et al., “Transcatheter Valve-in-Ring Implantation After Failure of Surgical Mitral Repair,” European Journal of Cardio-Thoracic Surgery 44, e8-e15, 2013. |
Weger et al., “First-in-Man Implantation of a Trans-Catheter Aortic Valve in a Mitral Annuloplasty Ring: Novel Treatment Modality for Failed Mitral Valve Repair,” European Journal of Cardio-Thoracic Surgery 39, 1054-1056, 2011. |
Walther et al., “Human Minimally invasive Off-Pump Valve-in-a-Valve implantation,” Case Reports, The Society of Thoracic Surgeons, 2008. |
Walther et al., “Valve-in-a-Valve Concept for Transcatheter Minimally Invasive Repeat Xenograph Implantation,” Preclinical Studies, Journal of the American College of Cardiology, 2007. |
Himbert et al., “Transseptal Implantation of a Transcatheter Heart Valve in a Mitral Annuloplasty Ring to Treat Mitrai Repair Failure,” Circulation Cardiovascular Interventions, American Heart Association, 2011. |
Himbert, Dominique, “Transvenous Mitral Valve Repair Replacement After Failure of Surgical Ring Annuloplasty,” Research Correspondence, Journal of the American College of Cardiology, 2012. |
Casselman et al., “Reducing Operative Morality in Valvular Reoperations: The “valve in ring” Procedure,” Brief Technique Reports, The Journal of Thoracic and Cardiovascular Surgery, vol. 141, No. 5. May 2011. |
Ma et al., “Double-Crowned Valved Stents for Off-Pump Mitral Valve Replacement,” European Journal of Cardio-Thoracic Surgery, 28, 194-199, 2005. |
Bonhoeffer et al., “Percutaneous Replacement of Pulmonary valve in a Right-Ventricle to Pulmonary-Artery Prosthetic Conduit with Valve Dysfunction,” Early Report, The Lancet, vol. 356, Oct. 21, 2000. |
International Search Report from corresponding PCT case No. PCT/IB2015/000901 dated Feb. 15, 2015. |
Andersen, H.R., et al. “Transluminal implantation of Artificial Heart Valve. Description of a New Expandable Aortic Valve and Initial Results with Implantation by Catheter Technique in Closed Chest Pig” European Heart Journal, No. 13, pp. 704-708, 1992. |
Andersen, H.R. “History of Percutaneous Aortic Valve Prosthesis,” Herz No. 34. pp. 343-346. 2009. |
Pavcnik, et al., “Development and initial Experimental Evaluation of a Prosthetic Aortic Valve for Transcatheter Placement,” Cardiovascular Radiology, vol. 183, No. 1. pp. 151-154. 1992. |
Bailey, S. “Percutaneous Expandable Prosthetic Valves,” Textbook of Interventional Cardiology vol. 2, 2nd Ed. pp. 1268-1276. 1994. |
Al-Khaja, et al. “Eleven Years' Experience with Carpentier-Edwards Biological Valves in Relation to Survival and Complications,” European Journal of Cardiothoracic Surgery, vol. 3. pp. 305-311. 1989. |
Ross, “Aortic Valve Surgery,” At a meeting of the Council on Aug. 4, 1966. pp. 192-197. |
Sabbah, et al. “Mechanical Factors in the Degeneration of Porcine Bioprosthetic Valves: An Overview,” Journal of Cardiac Surgery, vol. 4, No. 4. pp. 302-309. 1989. |
Wheatley, “Valve Prostheses,” Operative Surgery, 4th ed. pp. 415-424. 1986. |
Uchida, “Modifications of Gianturco Expandable Wire Stents,” American Journal of Roentgenology, vol. 150. pp. 1185-1187, 1986. |
Number | Date | Country | |
---|---|---|---|
20190167420 A1 | Jun 2019 | US |
Number | Date | Country | |
---|---|---|---|
61943125 | Feb 2014 | US | |
61942300 | Feb 2014 | US | |
61865657 | Aug 2013 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 14912067 | US | |
Child | 16268474 | US |