This invention relates generally to magnetic resonance imaging (MRI), and more particularly the invention relates to mapping the magnitude of RF fields in a MRI pulse sequence.
Magnetic resonance imaging (MRI) requires placing an object to be imaged in a static magnetic field (B0), exciting nuclear spins in the object with a RF magnetic field (B1), and then detecting signals emitted by the excited spins as they precess within the magnetic field (B0). Through the use of magnetic gradient and phase encoding of the excited magnetization, detected signals can be spatially localized in three dimensions.
For in vivo MRI at high field (≧3 T) it is essential to consider the homogeneity of the active B1 field (B1+), particularly if surface coils are used for RF transmission. The B1+ field is the transverse, circularly polarized component of B1 that is rotating in the same sense as the magnetization. When exciting or manipulating large collections of spins, nonuniformity in B1+ results in nonuniform treatment of spins. This leads to spatially varying image signal and image contrast and to difficulty in image interpretation and image-based quantification. The B1+ field experienced by spins within the body is influenced by several factors including the distance from the RF transmit coil, tissue dielectric constant, and factors related to the body size and RF wavelength. In high-field (≧3 T) abdominal, cardiac, and neuro imaging, B1+ inhomogeneity on the order of 30-50% has been predicted and observed. When using surface coil transmission, even greater variations in B1+ can be observed over typical imaging FOVs.
There are several existing B1+ mapping methods based on measurements at progressively increasing flip angles, stimulated echoes, or signal ratios. The most simple and straightforward of these methods is the double-angle method, which involves acquiring images with two flip angles α and 2α, where TR>>T1 such that image signal is proportional to sin(α) and sin(2α), respectively. The B1+ field is derived from the ratio of signal magnitudes. Previous double-angle approaches have been limited by the requirement of long TRs and therefore long imaging times and motion compensation issues. While accurate in static body regions, these methods are not practical in areas of the body that experience motion, such as the chest and abdomen.
The invention combines a double-angle method with a B1-insensitive magnetization reset sequence at the end of each data acquisition, which lifts the dependence of TR on T1 and enables rapid imaging, and interleaved spiral readouts, enabling multi-slice acquisition of B1+ maps in a short time (a single breathhold). Phantom and in vivo validation data are presented along with B1+ homogeneity measurements in the heart at 3 T.
The invention and objects and features thereof will be more readily apparent from the following detailed description and appended claims when taken with the drawings.
a)-1f) illustrate Bloch equation simulations showing the performance of B1-insensitive saturation pulses.
a)-2d) illustrate SDAM pulse sequence with spiral acquisitions.
a)-3d) illustrate a phantom validation study.
a)-4c) illustrate validation in a human leg at 1.5 T.
a)-5c) illustrate validation in a human head at 3 T.
a)-6b) illustrate Abdominal B1+ maps in a healthy volunteer at 3 T with short repetition times.
a)-8d) illustrate Cardiac B1+ maps in a healthy volunteer at 3 T.
The proposed method uses an adaptation of the double angle method (DAM) described previously. Such methods allow calculation of a flip-angle map, which is an indirect measure of the B1+ field. Two images are acquired: I1 with prescribed tip angle α1 and I2 with prescribed tip angle α2=2α1. All other signal-affecting sequence parameters are kept constant. For each voxel, the ratio of magnitude images satisfies
where r represents spatial position and α1(r) and α2(r) are tip angles that vary with the spatially varying B1+ field. If the effects of T1 and T2 relaxation can be neglected, then the actual tip angles as a function of spatial position satisfy
A long repetition time (TR>5 T1) is typically used with the double-angle methods so that there is no T1 dependence in either I1 or I2 (i.e., f1(T1,TR)=f2(T1,TR)=1.0). Instead, the proposed method includes a magnetization-reset sequence after each data acquisition with the goal of putting the spin population in the same state regardless of whether the α or 2α excitation was used for the preceding acquisition (i.e., f1(T1,TR)=f2(T1,TR)≠1.0).
The RF pulse used for saturation of the magnetization must destroy Mz over a range of B1+ values since this sequence is intended for use in areas with inhomogeneous B1+. Two classes of RF pulses that meet this requirement were investigated: composite pulse trains and BIR4 pulses. The performance of these waveforms as saturation pulses is shown in
Imaging Sequence—
Because B1+ profiles are usually quite smooth, in cases where SNR was lacking, we opted to use only the first 4 or 8 ms of each readout, which maintains the same FOV, but reduces the spatial resolution (Δx Δy) and reduces the total readout time Tread. The net effect on SNR is an increase. To reduce ringing artifacts caused by k-space truncation, a radial Hamming window was applied to gridded k-space data before inverse Fourier-transforming.
As shown in
Experimental Methods—Studies were performed on GE Signa 1.5 T and 3 T scanners (GE Healthcare, Waukesha, Wis., USA) with EXCITE hardware. Both scanners were equipped with gradients capable of 40 mT/m amplitude and 150 T/m/s slew rate and a receiver capable of 2-μs sampling (±250 kHz). The composite saturation pulse tested in
To validate the accuracy of the method, B1+ measurements were made in a phantom and normal volunteers. In the phantom study, the new method was used with TR ranging from 1 s to 100 ms. This was compared with a reference double-angle method consisting of two sequential gradient-echo scans using a long TR of 3 s and Cartesian k-space sampling. The phantom was spherical with 18 cm diameter and consisted of distilled water doped with copper sulfate. The T1 of the phantom was 345 ms as measured using an inversion-recovery sequence. A quadrature transmit/receive head coil was used, with the phantom positioned halfway out the end of the coil to produce an inhomogeneous B1+. A linear transmit/receive extremity coil was used to produced a B1+ map in the leg of a normal volunteer, both with the new method and the reference double-angle method with a long TR (3 s). The B1+ field was inhomogeneous where the leg extended beyond the coil. Last, for comparison with previous studies (4), measurements were made of the B1+ homogeneity in a human head at 3 T using a standard birdcage head coil for RF transmission and signal reception. Maps made with TRs ranging from 3000 to 400 ms were compared.
Abdominal B1+ maps covering the lower liver were acquired in two healthy volunteers at 3 T. Scan planes were localized using a standard three-plane localizer, and six axial slices were prescribed spanning the liver. Fat saturation was used in both subjects. B1+ maps of six slices were acquired in a 16-s scan with a 1-s TR (8 spiral interleaves×2 flip angles). This was repeated with a 500-ms TR (8-s scan) and 250-ms TR (4-s scan).
Cardiac B1+ maps were acquired in four healthy volunteers at 3 T. Short-axis and long-axis views were localized using the GE I-drive real-time system. In each volunteer, eight parallel short axis slices were prescribed spanning the left ventricle from base to apex (base was slice 1 and apex was slice 8). Fat saturation was used in all subjects. Acquisitions were cardiac gated (using a plethysmograph), with imaging occurring in mid-diastole, and were performed during breath-holds. Whole-heart coverage was achieved in a single breath-hold of 20 R-R intervals (14 to 25 s with heart rates of 48 to 85 beats/min).
Static Validation—
Breath-Held Measurements—Representative abdominal B1+ maps from one volunteer are shown in
Representative cardiac B1+ maps are shown in
The level of inhomogeneity observed (100%) and the general shape of B1+ variation are consistent with published references, and with SDAM, were acquired in substantially shorter scan time.
SNR of the acquired images is dependent on the saturation recovery time TSR, with image SNR α1−e−TSR/T1. As a natural consequence, shorter TRs result in lower TSR (see
The key component of the proposed sequence is the robust magnetization saturation. The effectiveness of the saturation used in these experiments was demonstrated by varying the TR and this showed negligible change in the B1+ maps. Although a short composite pulse was used here, there are many possibilities for saturation pulses that are insensitive to B1 and B0 inhomogeneity and the appropriate pulse can be chosen for different applications. It is likely that the most robust method would be a long train of adiabatic refocusing pulses; however, this would come at a price in minimum TR and RF heating. The basic idea of resetting the magnetization to remove T1 weighting could be extended to other methods for measuring B1+ as well, such as phase-based methods in which the flip angle is converted into phase in the transverse plane by applying a 90° (B1+ insensitive) plane-rotation pulse.
With very short TR, such as when TR<<T2, there is a possibility that stimulated echoes could form. These can be avoided by modulating the size of the dephaser after the saturation pulse. Given that the total number of applications of the saturation pulse is not large (twice the number of spiral interleaves), dephasers with relative areas that are geometrically increased could be used to remove the possibility of stimulated echoes.
The current multislice acquisition may be affected by flow between acquisitions. The use of a spatially nonselective saturation pulse alleviates this problem to some degree, as all magnetization within the sensitive region of the transmit coil (and within the parameters of the particular saturation pulse; see
For cardiac applications, there is the consideration of arrhythmia, or otherwise irregular R-R intervals. This will cause unequal amounts of saturation recovery time among k-space segments, which can result in image artifacts, and errors in the B1+ maps that have not been yet been analyzed. Also, our initial implementation requires a breathhold, which some patients may not be able to perform. However, with the introduction of a navigated acquisition, these limitations can be alleviated.
There are numerous applications for B1+ information that are anticipated to be useful. One possibility is adjusting the RF power dynamically to achieve the desired flip angle in regions of interest. Also, it is possible to design RF pulses “on-the-fly” to compensate for smoothly varying B1+ components. For some applications, such as transmit SENSE, a flip-angle dynamic range may be required that is larger than that supported with the arccos relationship of Eq. [2] (about 20 dB). In this case, the method could be applied multiple times, with the flip angle increased upon successive measurements to improve the dynamic range. The speed of the method makes such multiple acquisition schemes feasible. Besides the flip angle, the phase of the B1+ field may also be desired. To do this, an additional data acquisition could be performed using an adiabatic half-passage for the excitation pulse. The x- and y-components of the resulting magnetization would correspond to the x- and y-components of B1+.
B1+ mapping is becoming an integral part of prescan calibration in high-field MRI and MRI with surface transmit coils. The SDAM in accordance with the invention can yield volumetric B1+ maps with adequate spatial resolution in just a few seconds. This has substantial implications for high-field neuroimaging and for the first time enables rapid cardiac and abdominal B1+ mapping within the duration of a breath-hold (necessary for time-efficient motion suppression).
While the invention has been described with reference to specific embodiments, the description is illustrative of the invention and is not to be construed as limiting the invention. Various modifications and applications may occur to those skilled in the art without departing from the spirit and scope of the invention as defined by the appended claims.
The U.S. Government has rights in the disclosed invention pursuant to NIH Grant No. HL074332 to Stanford University and the University of Southern California, and NIH Grant No. EB002992 to Stanford University.