Clinicians can use pressure-volume (P-V) loops and parameter estimations of the respiratory system (e.g., resistance and compliance) to monitor the patient status and to offer safe ventilation therapy to the patient. Most accurate results are obtained using the transpulmonary pressure, which is accessible with esophageal catheter measurements. However, often a catheter is not present, in which case the above estimations are impacted by the presence of a patient effort and the chest wall mechanics. Modern ventilators, therefore, avoid a patient effort and ignore chest wall mechanics when determining the transpulmonary pressure.
Respiratory therapists can use a P-V loop as guidance for choosing the right ventilator settings to avoid atelectasis and overdistension (VILI). Ventilators can display P-V loops on their screen. Ventilators should use the transpulmonary pressure to construct the P-V loop accurately, since the transpulmonary pressure is responsible for the deformation of lung tissue, (i.e., the parenchyma tissue with alveoli). The transpulmonary pressure Pl is the difference between a pleural pressure and an alveolar pressure.
However, the transpulmonary pressure is difficult to measure as state-of-the-art requires the use of catheters (see, e.g., Umbrello, M. and Chiumello, D., 2018, “Interpretation of the transpulmonary pressure in the critically ill patient”, Ann Transl Med 2018; 6(19):383). For that reason, ventilators use the measured pressure at the airway opening, Pwye, and the derived alveolar pressure Palv, to construct the P-V loop. This method is accurate if there is no respiratory effort from the patient. The bigger the patient's respiratory effort, the more significant the difference between Palv and Pl, resulting in a less accurate P-V loop.
Therefore, setting the right pressure and volume is especially relevant for patients with damaged lungs which are ventilated with pressure support ventilation (PSV), proportional assist ventilation (PAV, PAV+) and other ventilation modalities such as NAVA where the patient makes an effort.
However, it can be difficult to measure or determine the transpulmonary pressure in a non-invasive way, such that safe ventilation can be offered in a more accurate and convenient way (i.e., without catheters and without avoid or circumvent a patient effort in determining the transpulmonary pressure).
The following discloses certain improvements to overcome these problems and others.
In one aspect, a mechanical ventilation device includes at least one electronic controller configured to receive imaging data related to a dimension of a diaphragm of a patient during inspiration and expiration while the patient undergoes mechanical ventilation therapy with an associated mechanical ventilator; calculate a pressure value of a chest of the patient based on at least the imaging data; and when the calculated pressure value does not satisfy an acceptance criterion, at least one of output an alert indicative of the calculated pressure value failing to satisfy the acceptance criterion; and output a recommended adjustment to one or more parameters of the mechanical ventilation therapy delivered to the patient.
In another aspect, a mechanical ventilation method includes, with at least one electronic controller, receiving imaging data related to a dimension of a diaphragm of a patient during inspiration and expiration while the patient undergoes mechanical ventilation therapy with an associated mechanical ventilator; calculating a pressure value of a chest of the patient based on at least the imaging data; and when the calculated pressure does not satisfy an acceptance criterion, at least one of outputting an alert indicative of the calculated pressure value failing to satisfy the acceptance criterion; and outputting a recommended adjustment to one or more parameters of the mechanical ventilation therapy delivered to the patient.
One advantage resides in preventing VILI in patients undergoing mechanical ventilation therapy.
Another advantage resides in determining a transpulmonary pressure of a patient undergoing mechanical ventilation therapy.
Another advantage resides in adjusting settings of a ventilator in delivering mechanical ventilation therapy to a patient based on a calculated a transpulmonary pressure of the patient.
Another advantage resides in automatically adjusting settings of a mechanical ventilator to help wean patients off mechanical ventilation therapy.
Another advantage resides in providing mechanical ventilation therapy without the use of invasive catheters or dedicated ventilation maneuvers for measuring a transpulmonary pressure of the patient.
Another advantage resides in using a detected thickening fraction of the diaphragm to wean a patient off of mechanical ventilation therapy.
Another advantage resides in using ultrasound to non-invasively measure a diaphragm response.
A given embodiment may provide none, one, two, more, or all of the foregoing advantages, and/or may provide other advantages as will become apparent to one of ordinary skill in the art upon reading and understanding the present disclosure.
The disclosure may take form in various components and arrangements of components, and in various steps and arrangements of steps. The drawings are only for purposes of illustrating the preferred embodiments and are not to be construed as limiting the disclosure.
As used herein, the singular form of “a”, “an”, and “the” include plural references unless the context clearly dictates otherwise. As used herein, statements that two or more parts or components are “coupled,” “connected,” or “engaged” shall mean that the parts are joined, operate, or co-act together either directly or indirectly, i.e., through one or more intermediate parts or components, so long as a link occurs. Directional phrases used herein, such as, for example and without limitation, top, bottom, left, right, upper, lower, front, back, and derivatives thereof, relate to the orientation of the elements shown in the drawings and are not limiting upon the scope of the claimed invention unless expressly recited therein. The word “comprising” or “including” does not exclude the presence of elements or steps other than those described herein and/or listed in a claim. In a device comprised of several means, several of these means may be embodied by one and the same item of hardware.
With reference to
In some embodiments, the medical imaging device 18 can comprise a wearable US imaging device 18. In a more particular example, the medical imaging device 18 includes an ultrasound transducer 20 that is wearable by the patient P (e.g., on the abdomen or chest of the patient P in position to image the diaphragm of the patient, as shown in
In some embodiments, an additional imaging device (e.g., a CT imaging device 26 as shown in
In some embodiments, a database 30 can store previously-acquired CT (or X-ray) images 28. These images 28 can be retrieved from the database 30 for processing by the electronic controller 13. Similarly to the CT imaging device 26, the database 30 may not be located in the same room, or even the same department, as the mechanical ventilator 2 (or in the same room or department as the CT imaging device 26). As diagrammatically indicated in
In some embodiments, previously-acquired CT images 28 of the patient P can be used to generate a biomechanical model 32 of thoracic structures (including lungs) and including a chest wall of the patient P. This model 32 can be used as a reference point for analyzing current CT images 28 of the patient P. The model 32 can be stored in the database 30 (or in the non-transitory computer readable medium 15 of the mechanical ventilator 2).
The non-transitory computer readable medium 15 can store instructions executable by the electronic controller 13 to perform a mechanical assistance method or process 100 for monitoring the patient P during mechanical ventilation therapy using the mechanical ventilator 2. With reference to
At an operation 102, imaging data is received by the electronic controller 13. The imaging data can include data related to a dimension of the diaphragm of a patient P during inspiration and expiration while the patient undergoes mechanical ventilation therapy with the mechanical ventilator 2. The received imaging data includes the US imaging data 24 acquired by the ultrasound transducer 20, and includes a thickness, a thickness change, and/or a position of the diaphragm of the patient P during inspiration and expiration.
In some embodiments, the received imaging data includes the CT images 28 acquired by the CT imaging device 26, and the CT images 28 include at least a chest wall of the patient P. In some embodiments, the received imaging data can comprise previously-acquired images stored in the database 30, and/or can further include retrieving the model 32 of the patient P.
In some embodiments, the electronic controller 13 can also receive data obtained by the mechanical ventilator 2. This ventilator data can include, for example, an airflow during inhalation of the patient P, an airflow pressure in an airway of the patient P during mechanical ventilation therapy, and so forth.
At an operation 104, the electronic controller 13 is configured to calculate a pressure value (i.e., a transpulmonary pressure (Pl), a tidal variation in the transpulmonary pressure (DPl), and so forth) of the chest of the patient P based on the received data. In one example, the transpulmonary pressure (Pl) can be calculated based on the US imaging data 24, the CT images 28, the data received from the mechanical ventilator 2, and so forth. In a particular embodiment, the electronic controller 13 is configured to (i) determine a diaphragmatic muscle pressure (Pmus) from the US imaging data 24 at an operation 101; and (ii) determine a chest wall elastance (Ecw) from the CT images 28 (or the previously-acquired images stored in the database 30, or from the biomechanical model 32) at an operation 103. The diaphragmatic muscle pressure (Pmus) and the chest wall elastance (Ecw), along with the airway pressure/airflow data from the mechanical ventilator 2, can be used to calculate the transpulmonary pressure (Pl) of the chest of the patient P.
In
P
wye
−P
mus
=Q
air
R
rs
+V
T
/C
l
+V
T
/C
cw (1)
where VT is the tidal volume, which is the integral of the air flow Qair. Another equation describing a pressure drop in the airway due to flow resistance can be defined according to Equation 2:
P
wye
=P
alv
+Q
air
R
rs
By using Equation (2) to solve for Pwye, equation (1) can be re-written into equation (3):
P
alv
−P
mus
=V
T
/C
l
+V
T
/C
cw (3)
The alveolar pressure can be determined from the mechanical ventilator 2 according to Equation (2): Palv=Pwye−Qair Rrs.
The term VT/Cl in Equation (1) represents an elastic work (i.e., a pressure change) that is needed to deform the lung from its functional residual capacity (FRC) to the volume at the end of inhalation according to Equation (4): Vinhale=FRC+VT. This elastic pressure change is the change (i.e., variation) in the transpulmonary pressure, DPl=VT/Cl during breathing. To determine the absolute value of the transpulmonary pressure, a pre-tension (i.e., an elastic preload) value needs to be determined according to Equation (5): P=Pl.0+DPl. The pre-tension or preload Pl.0 is the tension that keeps the lungs inflated at the end of exhalation when there is no action of the respiratory muscles. The pre-tension is the result of the balance between the chest wall which wants to expand (spring-out), and the lungs which want to recoil. This is the reason why the pleural pressure is negative even when the muscle effort is zero in the absence of ventilator support. The nature of this pre-tension becomes clear when there is a puncture in the pleural sac in which case the lung(s) may collapse (i.e., the pneumothorax).
This pre-tension value can be used as a further input to calculate the transpulmonary pressure (Pl) of the chest of the patient P, as shown in
The term VT/Ccw in Equation (1) represents the elastic recoil (i.e., pressure) of the chest wall. During quite tidal breathing this term is negative because of the chest wall tends to “spring out”, in which case the chest wall elasticity works with the diaphragm muscle during a volume expansion of the lungs.
Pmus and Ccw can be determined with imaging data of the patient P. Once these parameters are known, and having VT and Palv continuously available from the mechanical ventilator 2, it is possible to calculate the variation in the transpulmonary pressure on a breath-by-breath basis according to (rewriting Equation (1) as Equation (6)):
DP
l
=V
T
/C
l
=P
alv
−P
mus
−V
T
/C
cw (6)
The inspiratory muscle pressure Pmus can be determined from the US imaging data 24 acquired by the ultrasound transducer 20. The diaphragmatic muscle pressure Pmus can be estimated from the diaphragmatic ultrasound imaging data 24 in different ways. In one example, Ers a patient specific biomechanical model 32 is applied to calculate Pmus. The biomechanical model 32 takes as input the diaphragmatic muscle excursion xd obtained from the ultrasound measurements 24 and the pressure from the mechanical ventilator 2, and provides as output the muscle pressure Pmus which is exerted by the diaphragm on the lungs and the chest wall. An advantage of the biomechanical model 32 is that the patient specific geometries are taken into account in the estimation of Pmus. The evaluation of the biomechanical model 32 can be done off-line. The biomechanical model 32 simulates Pmus as a function of the diaphragm excursion xd. The output Pmus as a function of xd is stored in a lookup table (not shown) in the non-transitory computer readable medium 15. The lookup table is then used real time.
In another example, a skeleton muscle model with a force-length relation for muscle fibers can be used (see, e.g., Zhang et al. BioMed Eng OnLine (2016) 15:18, “Biomechanical simulation of thorax deformation using finite element approach”). In such a model, the force generated by a muscle fiber can be determined from its contraction (force-length relationship). The contraction of the diaphragm muscle during an inhalation effort by the patient (thickness and length change) is measured with ultrasound (e.g., the diaphragmatic thickening fraction TFDI and the diaphragm excursion). The muscle force is converted to a muscle pressure by dividing by the projected surface area of the diaphragm, Pmus=Fmus/Ad (i.e., a piston model).
The chest wall compliance Ccw can be determined, for example, from the CT images 28 acquired by the CT imaging device 26. The chest wall compliance Ccw during inspiration can be determined in different ways. In one example, an indirect measurement using a Positive End Expiratory Pressure (PEEP) step method (PSM) can be performed by the electronic controller 13 (see, e.g., Persson, P. et al., 2018, “Evaluation of lung and chest wall mechanics during anaesthesia using the PEEP-step method”, British Journal of Anaesthesia, 120 (4): 860e867 (2018)). The PSM provides the lung compliance Cl of a fully sedated patient when Pmus=0. The inverse chest wall compliance is the difference between the inverse respiratory system compliance and the inverse lung compliance according to 1/Crs=1/Cl+1/Ccw. The chest wall compliance is less sensitive to disease than the lung compliance. The PSM procedure can be performed at an early ICU stage when the patient is still fully sedated. The measured Ccw can then be used later when there is a respiratory effort.
In another example, a direct measurement process can be performed using a mouthpiece which requires cooperation from the patient (see, e.g., Gideon, E. A. et al., 2021, “The effect of estimating chest wall compliance on the work of breathing during exercise as determined via the modified Campbell diagram”, Am J Physiol Regul Integr Comp Physiol 320: R268-R275). In another example, a look-up table comprising chest wall stiffness values based on patient information such as age and gender can be used (see, e.g., Gideon).
In another example, an algorithm to compute the chest wall stiffness based on a CT scan. A finite element model of the thorax is constructed based on a segmentation of the CT scan (see, e.g., Zhang). Mechanical properties of the structures (intercostal muscle, diaphragm, bone, cartilage, tendons) can be known. The effective chest wall stiffness (inverse compliance) can be calculated by imposing a pressure ΔP in a direction perpendicular to the chest wall (as a boundary condition), and subsequently determine the simulated volume change ΔV from the model output, Ecw=1/Ccw=ΔP/ΔV. Optionally the finite element model can take into account gravity to evaluate the of tissue (i.e., fat) around the chest or belly, patient position (e.g., prone, supine), or abdominal pressure.
In similar ways the lung compliance can be determined. However, the chest wall compliance is less sensitive to disease properties and the chest wall structure and material properties are less sensitive to person-to-person variations (bone stiffness is known to be less variable than parenchymal tissue). Therefore, it is advantageous to focus on the chest wall compliance.
In another example, a machine learning algorithm to compute the chest wall stiffness (i.e., compliance). The model is trained with imaging (e.g., CT, X-ray, and so forth), data from the mechanical ventilator 2 (VT, Palv) and catheter data (esophageal pressure, Pes). The trained model takes as input the imaging and mechanical ventilator data and provides as output Pes. Subsequently the lung and chest wall compliance can be calculated, subsequently, Cl=VT/(Palv−Pes) and 1/Ccw=1/Crs+1/Cl.
Referring back to
In another example embodiment, at an operation 110, a recommended adjustment to one or more parameters of the mechanical ventilation therapy delivered to the patient P is output. Again, this can be done by displaying a message on the display device 14 of the mechanical ventilator 2, thereby indicating to a medical professional that the calculated pressure value (Pl, DPl) is not satisfactory.
In a further example embodiment, at an operation 112, the mechanical ventilator 2 is controlled to adjust one or more parameters of the mechanical ventilation therapy delivered to the patient P. If the operation 112 is performed, then the pressure value (Pl, DPl) can be re-calculated, and this re-calculated pressure value (Pl, DPl) can be analyzed to determine if the acceptance criterion is met. It will be appreciated that more than one of the operations 108, 110, and 112 can be performed (e.g., the alert can be displayed and the settings of the mechanical ventilator 2 can be adjusted). In some embodiments, the operations 102-106 and at least one of operations 108-112 can be repeated iteratively to provide feedback control of the mechanical ventilator 2 based at least on whether the calculated pressure value (Pl DPl) satisfies the acceptance criterion.
In a particular embodiment, the display device 14 of the mechanical ventilator 2 is configured to display a pressure-volume (P-V) curve 30 of lungs of the patient P (shown schematically in
Modern ventilators can measure the global compliance of the respiratory system. In the measurement procedures a patient effort is avoided (e.g., by taking a measurement at zero flow or during exhalation). However, it is difficult to do this in a reliable and accurate manner since the slopes in the flow curves are high. Consequently, a small misalignment leads to a big error. The diaphragmatic ultrasound imaging data 24 to detect when there is no patient effort, besides looking at the P-V curve 30. To do so, the mechanical ventilator 2 can be synchronized with the ultrasound transducer 20. Pmus (or a surrogate such as TFDI) can be used to plot the P-V curve 30 for display on the display device 14. The compliance of the respiratory system, Crs can be determined from the region with neither Pmus nor flow (determined from the volume)
The disclosure has been described with reference to the preferred embodiments. Modifications and alterations may occur to others upon reading and understanding the preceding detailed description. It is intended that the exemplary embodiment be construed as including all such modifications and alterations insofar as they come within the scope of the appended claims or the equivalents thereof.
This patent application claims the priority benefit under 35 U.S.C. § 119(e) of U.S. Provisional Application No. 63/407,772, filed on Sep. 19, 2022, the contents of which are herein incorporated by reference. The following relates generally to the respiratory therapy arts, mechanical ventilation arts, mechanical ventilation weaning arts, ventilator induced lung injury (VILI) arts, and related arts.
Number | Date | Country | |
---|---|---|---|
63407772 | Sep 2022 | US |