Reducing the cost of sequencing is important to enable improved healthcare. A standard for measuring the cost of sequencing is the price of a 30× human genome, defined as 90 gigabases.
The price of a genome dropped significantly from 2007 to 2011 where it stabilized to just under $10,000 per genome. A significant milestone has been the $1,000 genome which was recently achieved. The next major milestone is the $100 genome which is expected to take several years. This invention discusses methods to achieve a $10 genome in a substantially contracted time frame. At this price point, it will be economical to sequence every newborn and will make the cost barrier for disease diagnosis and screening, especially in the area of oncology, significantly more economical.
The major cost components for sequencing systems are primarily the consumables which include biochip and reagents and secondarily the instrument costs.
To reach a $10 30× genome, a 100 fold cost reduction, the amount of data per unit area needs to increase by 100 fold and the amount of reagent per data point needs to drop by 100 fold.
In an example $1,000 genome platform with cluster densities of ten million molecules per square centimeter, each molecule occupies on average 10 um2 of chip area. Thus, the average effective pitch is 3,160 nm. If densities 100 fold higher could be obtained with 100 fold fewer copies, for the same chip area and reagent a 100 fold more information would be obtained resulting in 100 fold reduction in costs. At 100 fold higher density, the new pitch would need to be 320 nm. The number of copies to equalize reagent use is 10 copies per molecule, 100 fold fewer than 1,000 copies per cluster.
Thus, what is needed are optical imaging systems that can resolve optical signals from single molecules spaced apart by around 320 nm. However, this resolution is challenging to achieve due to the diffraction limit of light, which is defined by λ/(2*N.A.), where λ is the wavelength of light, and N.A. is the numerical aperture of the optical imaging system, which is near 1 in aqueous-based systems, such as those useful for sequencing and analyte detection. Thus, for detection of optical signals emitted around 650 nm, the 320 nm spacing is near or below the diffraction limit, which can prevent resolving individual features on such an array.
Although other methods exist that are not constrained by the diffraction limit of optical signals, such as electrical based systems developed by companies such as Ion Torrent (purchased by Thermo Fisher) and Oxford Nanopore, image based sequencing systems currently have the lowest sequencing costs of all existing sequencing technologies. Image based systems achieve low cost through the combination of high throughput imaging optics and low cost consumables.
What is needed, therefore, are optical imaging methods and systems that overcome the diffraction limit to facilitate increased resolution of individual features on a closely-packed substrate, such that resolution below the diffraction limit can be done with high accuracy. These methods and systems can have particular applications in high resolution feature detection, including for use in optical imaging for polynucleotide sequence detection.
Methods and systems for sub-diffraction limited imaging of single molecule analytes immobilized to the surface of a substrate. Substrates include flow cells and the like for performing binding reactions with the analytes. Analytes include biomolecules spaced apart on the surface at discrete locations for single molecule resolution, such as individual polynucleotides or proteins. These can be used for high resolution single molecule detection for such applications as single molecule sequencing by synthesis.
In some embodiments, provided herein is a method for sequencing a plurality of polynucleotides immobilized at high density on a surface of a substrate at a single molecule resolution, comprising: providing a substrate comprising a surface, wherein the surface comprises a plurality of polynucleotides immobilized on the surface at discrete locations, and wherein said surface comprises reagents for sequencing by synthesis; performing a plurality of cycles of single molecule sequencing by synthesis comprising, each cycle comprising: contacting said polynucleotides with a set of reversible terminator nucleotides comprising a detectable label; imaging a field of said surface with an optical system to detect an optical signal from each nucleotide incorporated into said polynucleotides, thereby detecting a plurality of optical signals in said field for said cycle; determining a peak location from each of said plurality of optical signals from images of said field from at least two of said plurality of cycles; overlaying said peak locations for each optical signal and applying an optical distribution model at each cluster of optical signals to determine a relative position of each detected analyte on said surface with improved accuracy; deconvolving said optical signals in each field image from each cycle using said determined relative position and a deconvolution function; identifying said detectable labels incorporated into said polynucleotide for each field and each cycle from said deconvolved optical signals; and sequencing said plurality of polynucleotides immobilized on the surface of the substrate from said identified detectable labels across said plurality of cycles at each polynucleotide position.
In some embodiments, the substrate comprises 1,000 or less, 500 or less, 100 or less, 50 or less 25 or less, 20 or less, 15 or less, or 10 or less clonal copies of a single molecule comprising an identical sequence. In some embodiments, the polynucleotides are DNA concatemers.
In some embodiments, each cycle further comprises washing said surface to remove unbound nucleotides after contacting said surface with said plurality of reversible terminator nucleotides and before imaging said field. In some embodiments, the cycle further comprises cleaving said reversible terminator if another cycle is to be performed. In some embodiments, the cycle further comprises cleaving said detectable label if another cycle is to be performed.
In some embodiments, the set of reversible terminator nucleotides comprises at least two distinct nucleotides each with a distinct detectable label. In some embodiments, the set of reversible terminator nucleotides comprise at least four distinct nucleotides each with a distinct detectable label in some embodiments, the set of reversible terminator nucleotides comprises adenine, cytosine, thymine, and guanine. In some embodiments, the set of reversible terminator nucleotides comprises adenine, cytosine, uracil, and guanine.
In some embodiments, the polynucleotide comprises deoxyribonucleic acid or ribonucleic acid. In some embodiments, the plurality of target polynucleotides have a length of about 1 kb to about 100 kb. In some embodiments, the plurality of target polynucleotides have a length of about 10 kb to about 50 kb. In some embodiments, the polynucleotides bound to the surface are separated by a distance of at least 10 nm.
In some embodiments, the detectable label is bound to the 3′-OH group of said reversible terminator nucleotide. In some embodiments, a blocking group that is not a detectable label is bound to the 3′-OH of said reversible terminator nucleotide.
In some embodiments, the plurality of target polynucleotides are immobilized by binding to capture probes bound to said surface at discrete locations. In some embodiments, the plurality of target polynucleotides are linked to an adaptor comprising a capture sequence that is complementary to a sequence of said capture probe, and a priming sequence that is complementary to a sequence of said sequencing primer. In some embodiments, the capture sequence is from 20 to 50 mer. In some embodiments, the priming sequence is from 20 to 50 mer.
In some embodiments, the method of sequencing further comprises performing previous cycle regression to correct a phasing error by comparing a set of polynucleotides having the same sequence or on the basis of the data itself.
In some embodiments, the deconvolution comprises removing interfering optical signals from neighboring polynucleotides using a center-to-center distance between said neighboring polynucleotides from said determined relative positions, in some embodiments, the deconvolution function comprises nearest neighbor variable regression. In some embodiments, the deconvolution comprises separating overlapping wavelengths from each unique detectable label used in each cycle. In some embodiments, the deconvolution function comprises cross-talk regression. In some embodiments, the deconvolution function comprises nearest neighbor variable regression, smoothing, or cross-talk correction.
Polynucleotides
In some embodiments, the polynucleotides are spaced apart on said substrate for single molecule sequencing by synthesis. In some embodiments, the polynucleotides are densely packed on said substrate such that there is overlap between optical signals emitted by said detectable labels from probes bound to adjacent polynucleotides comprising distinct polynucleotide sequences to be sequenced in some embodiments, the polynucleotides immobilized on said surface are spaced apart on average of less than the diffraction limit of the light emitted by the detectable labels and imaged by the optical system. In some embodiments, at least two of said polynucleotides immobilized on said surface are spaced apart less than the diffraction limit of the light emitted by the detectable labels and imaged by the optical system. In some embodiments, at least 10%, 20%, 30%, 40%, 50%, 60%, 70%, 80%, or 90% of said polynucleotides immobilized on said surface are spaced apart from another immobilized polynucleotide by less than the diffraction limit of the light emitted by the detectable labels and imaged by the optical system.
In some embodiments, the optical system comprises a numerical aperture of between 0.2-2.0. In some embodiments, the optical system comprises a numerical aperture of between 1-1.1. In some embodiments, the wavelength of said emitted light is about 400-450 nm, about 450-500 nm, about 500-550 nm, about 550-600 nm, about 600-650 nm, or about 650-700 nm.
In some embodiments, the immobilized polynucleotides comprises a minimum center-to-center distance between adjacent polynucleotides of less than 600 nm, less than 500 nm, less than 400 nm, less than 300 nm, or less than 200 nm. In some embodiments, the polynucleotides are immobilized on said surface at an average density of about 4-25 molecules per square micron. In some embodiments, the polynucleotides are immobilized on said surface at an average density of more than 4, more than 6, more than 8, more than 10, more than 15, or more than 20 molecules per square micron.
In some embodiments, the imaging of said surface is performed at a resolution greater than the critical sampling rate as determined by the Nyquist limit of the optical system. In some embodiments, the imaging of said surface is performed at a resolution of at least 2× the Nyquist sampling frequency. In some embodiments, the imaging of said surface is performed at a resolution of one pixel per 300 nm or higher along an axis of the image field. In some embodiments, the imaging of said surface is performed at a resolution of about 162.5 nm per pixel along an axis of the image field.
In some embodiments, the sequencing method further comprises generating an oversampled image with a higher pixel density from each of said field images from each cycle. In some embodiments, the oversampled image is generated by applying smoothing to each field image based on an anticipated point spread function for said optical signals. In some embodiments, a data set comprising the location of optical signal peaks from said image is generated from said field image or said oversampled image.
In some embodiments, overlaying said peak locations comprises aligning positions of said optical signal peaks detected in each field for a plurality of said cycles to generate a cluster of optical peak positions for each polynucleotide from said plurality of cycles. In some embodiments, the optical distribution model is a Gaussian distribution. In some embodiments, the optical distribution model is a point spread function.
In some embodiments, the relative position is determined for a plurality of said polynucleotides in said field. In some embodiments, the relative position is determined with an accuracy of within 10 nm RMS.
In some embodiments, the sequencing method further comprises overlaying a plurality of images of said field from different cycles to determine a relative offset with respect to a reference image of said field. In some embodiments, the method comprises generating offset values for each of said fields aligned with said reference field. In some embodiments, the relative position of polynucleotides within each field is determined from said offset values. In some embodiments, the offset determination comprises discarding field images whose alignment is outside of an alignment threshold. In some embodiments, the sequencing method comprises overlaying a plurality of images from said field to determine a relative offset with respect to a reference image of said field, wherein said relative position is determined with an accuracy of within 5 nm RMS.
In some embodiments, the method is capable of resolving optical signals from a surface at a density of ˜4-25 per square micron.
In some embodiments, the detectable labels emit light, and the polynucleotides are immobilized on the surface of said substrate at an average pitch below the diffraction limit of light emitted from said detectable labels.
According to some embodiments, also provided herein is a method for accurately determining a relative position of analytes immobilized on the surface of a densely packed substrate, comprising: providing a substrate comprising a surface, wherein the surface comprises a plurality of analytes immobilized on the surface at discrete locations performing a plurality of cycles of probe binding and signal detection on said surface, (each cycle comprising: contacting said analytes with a plurality of probes from a probe set, wherein said probes comprise a detectable label, wherein each of said probes binds specifically to a target analyte; and imaging a field of said surface with an optical system to detect a plurality of optical signals from individual probes bound to said analytes at discrete locations on said surface); determining a peak location from each of said plurality of optical signals from images of said field from at least two of said plurality of cycles; and overlaying said peak locations for each optical signal and applying an optical distribution model at each cluster of optical signals to determine a relative position of each detected analyte on said surface with improved accuracy.
In some embodiments, the method further comprises: deconvolving said optical signals in each field image from each cycle using said determined relative position and a deconvolution function; and identifying said detectable labels bound to said immobilized analytes for each field and each cycle from said deconvolved optical signals.
In some embodiments, the method further comprises using said detectable label identity for each analyte detected at each cycle to identify a plurality of said analytes on said substrate.
In some embodiments, the deconvolution comprises removing interfering optical signals from neighboring analytes using a center-to-center distance between said neighboring analytes from said determined relative positions of said neighboring analytes.
In some embodiments, the deconvolution function comprises nearest neighbor variable regression. In some embodiments, the deconvolution comprises separating overlapping wavelengths from each unique detectable label used in each cycle. In some embodiments, the deconvolution function comprises cross-talk regression. In some embodiments, the deconvolution function comprises nearest neighbor variable regression, smoothing, or cross-talk correction.
In some embodiments, the analytes are single molecules. In some embodiments, the single molecules are single biomolecules. In some embodiments, the single molecules are polynucleotides.
In some embodiments, the analytes are densely packed on said substrate such that there is overlap between optical signals emitted by said detectable labels from probes bound to adjacent analytes. In some embodiments, the analytes immobilized on said surface are spaced apart on average less than the diffraction limit of the light emitted by the detectable labels and imaged by the optical system. In some embodiments, at least two of said analytes immobilized on said surface are spaced apart less than the diffraction limit of the light emitted by the detectable labels and imaged by the optical system. In some embodiments, at least 10%, 20%, 30%, 40%, 50%, 60%, 70%, 80%, or 90% of said analytes immobilized on said surface are spaced apart from another analyte by less than the diffraction limit of the light emitted by the detectable labels and imaged by the optical system.
In some embodiments, the optical system comprises a numerical aperture of between 0.2-2.0. In some embodiments, the optical system comprises a numerical aperture of between 1-1.1. In some embodiments, the wavelength of said light is about 400-450 nm, about 450-500 nm, about 500-550 nm, about 550-600 nm, about 600-650 nm, or about 650-700 nm.
In some embodiments, the immobilized analytes comprises a minimum center-to-center distance between adjacent analytes of less than 60 nm, less than 500 nm, less than 400 nm, less than 300 nm, or less than 200 nm. In some embodiments, the target analytes are immobilized on said surface at an average density of about 4-25 molecules per square micron. In some embodiments, the target analytes are immobilized on said surface at an average density of more than 4, more than 6, more than 8, more than 10, more than 15, or more than 20 molecules per square micron.
In some embodiments, each cycle further comprises repeating steps i) and ii) using additional probes from said probe set. In some embodiments, each cycle further comprises removing unbound probes from said surface after contacting said surface with said plurality of probes and before imaging said field. In some embodiments, each cycle further comprises removal of bound probes from said surface if another cycle is to be performed.
In some embodiments, at least 5, 10, 15, 20, 25, 30, 35, 40, 50, 60, 70, 80, 90, or 100 cycles are performed. In some embodiments, each cycle comprises imaging a plurality of fields of said surface with said optical system.
In some embodiments, the imaging of said surface is performed at a resolution greater than the critical sampling rate as determined by the Nyquist limit of the optical system. In some embodiments, the imaging of said surface is performed at a resolution of at least 2× the Nyquist sampling frequency. In some embodiments, the imaging of said surface is performed at a resolution of one pixel per 300 nm or higher along an axis of the image field. In some embodiments, the imaging of said surface is performed at a resolution of about 162.5 nm per pixel along an axis of the image field.
In some embodiments, the method further comprises generating an oversampled image with a higher pixel density from each of said field images from each cycle. In some embodiments, the oversampled image is generated by applying smoothing to each field image based on an anticipated point spread function for said optical signals. In some embodiments, the method further comprises generating a data set comprising the location of optical signal peaks from said field image or said oversampled image.
In some embodiments, overlaying said peak locations comprises aligning positions of said optical signal peaks detected in each field for a plurality of said cycles to generate a duster of optical peak positions for each analyte from said plurality of cycles. In some embodiments, the optical distribution model is a Gaussian distribution. In some embodiments, the optical distribution model is a point spread function.
In some embodiments, the relative position is determined for a plurality of said analytes in said field. In some embodiments, the relative position is determined with an accuracy of within 10 nm RMS.
In some embodiments, the method further comprises overlaying a plurality of images of said field from different cycles to determine a relative offset with respect to a reference image of said field. In some embodiments, the method comprises generating offset values for each of said fields aligned with said reference field. In some embodiments, the relative position of analytes within each field is determined from said offset values. In some embodiments, the method further comprises discarding field images whose alignment is outside of an alignment threshold. In some embodiments, the method further comprises overlaying a plurality of images from said field to determine a relative offset with respect to a reference image of said field, wherein said relative position is determined with an accuracy of within 5 nm RMS.
In some embodiments, the method is capable of resolving optical signals from a surface at a density of ˜4-25 per square micron.
In some embodiments, the detectable labels emit light, and wherein the target analytes bound to said array comprises an average pitch below the diffraction limit of light emitted from said detectable labels.
Also provided herein, according to some embodiments, is a method for identifying a plurality of densely packed analytes immobilized on a surface of a substrate, comprising: providing a substrate comprising a surface, wherein the surface comprises a plurality of analytes immobilized on the surface at discrete locations; performing a plurality of cycles of probe binding and signal detection on said surface, (each cycle comprising: contacting said analytes with a plurality of probes from a probe set, wherein said probes comprise a detectable label, wherein each of said probes binds specifically to a target analyte; and imaging a field of said surface with an optical system to detect a plurality of optical signals from individual probes bound to said analytes); determining a peak location from each of said plurality of optical signals from images of said field from at least two of said plurality of cycles; overlaying said peak locations for each optical signal and applying an optical distribution model at each cluster of optical signals to determine a relative position of each detected analyte on said surface with improved accuracy; deconvolving said optical signals in each field image from each cycle using said determined relative position and a deconvolution function; determining the identity of each detectable label in each field and each cycle from said deconvolved optical signals; and using said detectable label identity for each analyte detected at each cycle to identify a plurality of said analytes on said substrate.
Also provided herein, according to some embodiments, is a system for determining the identity of a plurality of analytes, comprising an optical imaging device configured to image a plurality of optical signals from a field of a substrate over a plurality of cycles of probe binding to analytes immobilized on a surface of the substrate; and an image processing module, said module configured to: determine a peak location from each of said plurality of optical signals from images of said field from at least two of said plurality of cycles; determine a relative position of each detected analyte on said surface with improved accuracy by applying an optical distribution model to each cluster of optical signals from said plurality of cycles; and deconvolve said optical signals in each field image from each cycle using said determined relative position and a deconvolution function.
In some embodiments, the image processing module is further configured to determine an identity of said analytes immobilized on said surface using said deconvolved optical signals.
In some embodiments, the analytes are each a polynucleotide molecule and wherein said identity comprises a sequence of said polynucleotide molecules.
In some embodiments, the optical image device comprises a moveable stage defining a scannable area.
In some embodiments, the optical image device comprises a sensor and optical magnification configured to sample a surface of a substrate at below the diffraction limit in said scannable area.
In some embodiments, the optical imaging system further comprising a substrate comprising analytes immobilized to a surface of the substrate at a center-to-center spacing below the diffraction limit.
In some embodiments, the deconvolution comprises removing interfering optical signals from neighboring analytes using a center-to-center distance between said neighboring analytes from said determined relative positions of said neighboring analytes. In some embodiments, the deconvolution function comprises nearest neighbor variable regression. In some embodiments, the deconvolution comprises separating overlapping wavelengths from each unique detectable label used in each cycle. In some embodiments, the deconvolution function comprises cross-talk regression. In some embodiments, the deconvolution function comprises nearest neighbor variable regression, smoothing, or cross-talk correction.
In some embodiments, the analytes are single molecules. In some embodiments, the single molecules are single biomolecules. In some embodiments, the single molecules are polynucleotides.
In some embodiments, the analytes are densely packed on said substrate such that there is overlap between optical signals emitted by said detectable labels from probes bound to adjacent analytes. In some embodiments, the analytes immobilized on said surface are spaced apart on average less than the diffraction limit of the light emitted by the detectable labels and imaged by the optical system. In some embodiments, at least two of said analytes immobilized on said surface are spaced apart less than the diffraction limit of the light emitted by the detectable labels and imaged by the optical system. In some embodiments, at least 10%, 20%, 30%, 40%, 50%, 60%, 70%, 80%, or 90% of said analytes immobilized on said surface are spaced apart from another analyte by less than the diffraction limit of the light emitted by the detectable labels and imaged by the optical system.
In some embodiments, the optical system comprises a numerical aperture of between 0.2-2.0. In some embodiments, the optical system comprises a numerical aperture of between 1-1.1. In some embodiments, the wavelength of said light detected by the optical system is about 400-450 nm, about 450-500 nm, about 500-550 nm, about 550-600 nm, about 600-650 nm, or about 650-700 nm.
In some embodiments, the immobilized analytes comprises a minimum center-to-center distance between adjacent analytes of less than 600 nm, less than 500 nm, less than 400 nm, less than 300 nm, or less than 200 nm. In some embodiments, the analytes are immobilized on said surface at an average density of about 4-25 molecules per square micron. In some embodiments, the analytes are immobilized on said surface at an average density of more than 4, more than 6, more than 8, more than 10, more than 15, or more than 20 molecules per square micron.
In some embodiments, the optical imaging device is configured to image said substrate at a resolution greater than the critical sampling rate as determined by the Nyquist limit of the optical system. In some embodiments, the optical imaging device is configured to image said substrate at a resolution of at least 2× the Nyquist sampling frequency. In some embodiments, the optical imaging device is configured to image said substrate at a resolution of no more than 300 nm per pixel along an axis of the image field. In some embodiments, the optical imaging device is configured to image said substrate at a resolution of about 162.5 nm per pixel along an axis of the image field.
In some embodiments, the image processing module is configured to generate an oversampled image with a higher pixel density from each of said field images from each cycle. In some embodiments, the image processing module is configured to apply smoothing to each field image based on an anticipated point spread function for said optical signals to generate said oversampled image. In some embodiments, the image processing module is configured to generate a data set comprising the location of optical signal peaks from said imaged field.
In some embodiments, the system is capable of resolving optical signals from a surface at a density of ˜4-25 per square micron.
In some embodiments, the target analytes are immobilized on said substrate at an average center-to-center distance below the diffraction limit of light detected by the optical imaging device.
The foregoing and other objects, features and advantages will be apparent from the following description of particular embodiments of the invention, as illustrated in the accompanying drawings in which like reference characters refer to the same parts throughout the different views. The drawings are not necessarily to scale, emphasis instead placed upon illustrating the principles of various embodiments of the invention.
The details of various embodiments of the invention are set forth in the description below. Other features, objects, and advantages of the invention will be apparent from the description and the drawings, and from the claims.
As used herein, the term center-to-center distance refers to a distance between two adjacent molecules as measured by the difference between the average position of each molecule on a substrate. The term average minimum center-to-center distance refers specifically to the average distance between the center of each analyte disposed on the substrate and the center of its nearest neighboring analyte, although the term center-to-center distance refers also to the minimum center-to-center distance in the context of limitations corresponding to the density of analytes on the substrate. As used herein, the term “pitch” or “average effective pitch” is generally used to refer to average minimum center-to-center distance. In the context of regular arrays of analytes, pitch may also be used to determine a center-to-center distance between adjacent molecules along a defined axis.
As used herein, the term “overlaying” (e.g., overlaying images) refers to overlaying images from different cycles to generate a distribution of detected optical signals (e.g., position and intensity, or position of peak) from each analyte over a plurality of cycles. This distribution of detected optical signals can be generated by overlaying images, overlaying artificial processed images, or overlaying datasets comprising positional information. Thus, as used herein, the term “overlaying images” encompasses any of these mechanisms to generate a distribution of position information for optical signals from a single probe bound to a single analyte for each of a plurality of cycles.
A “cycle” is defined by completion of one or more passes and stripping of the detectable label from the substrate. Subsequent cycles of one or more passes per cycle can be performed. For the methods and systems described herein, multiple cycles are performed on a single substrate or sample. For DNA sequencing, multiple cycles requires the use of a reversible terminator and a removable detectable label from an incorporated nucleotide. For proteins, multiple cycles requires that the probe removal (stripping) conditions either maintain proteins folded in their proper configuration, or that the probes used are chosen to bind to peptide sequences so that the binding efficiency is independent of the protein fold configuration.
A“pass” in a detection assay refers to a process where a plurality of probes comprising a detectable label are introduced to the bound analytes, selective binding occurs between the probes and distinct target analytes, and a plurality of signals are detected from the detectable labels. A pass includes introduction of a set of antibodies that bind specifically to a target analyte. A pass can also include introduction of a set of labelled nucleotides for incorporation into the growing strand during sequencing by synthesis. There can be multiple passes of different sets of probes before the substrate is stripped of all detectable labels, or before the detectable label or reversible terminator is removed from an incorporated nucleotide during sequencing. In general, if four nucleotides are used during a pass, a cycle will only consist of a single pass for standard four nucleotide sequencing by synthesis.
As used herein, an image refers to an image of a field taken during a cycle or a pass within a cycle. In some embodiments, a single image is limited to detection of a single color of a detectable label.
As used herein, the term “field” refers to a single region of a substrate that is imaged. During a typical assay a single field is imaged at least once per cycle. For example, for a 20 cycle assay, with 4 colors, there can be 20*4=80 images, all of the same field.
A “target analyte” or “analyte” refers to a single molecule, compound, complex, substance or component that is to be identified, quantified, and otherwise characterized. A target analyte can comprise by way of example, but not limitation to, a single molecule (of any molecular size), a single biomolecule, a polypeptide, a protein (folded or unfolded), a polynucleotide molecule (RNA, cDNA, or DNA), a fragment thereof, a modified molecule thereof, such as a modified nucleic acid, or a combination thereof. In an embodiment, a target polynucleotide comprises a hybridized primer to facilitate sequencing by synthesis. The target analytes are recognized by probes, which can be used to sequence, identify, and quantify the target analytes using optical detection methods described herein.
A “probe” as used herein refers to a molecule that is capable of binding to other molecules (e.g., a complementary labelled nucleotide during sequencing by synthesis, polynucleotides, polypeptides or full-length proteins, etc.), cellular components or structures (lipids, cell walls, etc.), or cells for detecting or assessing the properties of the molecules, cellular components or structures, or cells. The probe comprises a structure or component that binds to the target analyte. In some embodiments, multiple probes may recognize different parts of the same target analyte. Examples of probes include, but are not limited to, a labelled reversible terminator nucleotide, an aptamer, an antibody, a polypeptide, an oligonucleotide (DNA, RNA), or any combination thereof. Antibodies, aptamers, oligonucleotide sequences and combinations thereof as probes are also described in detail below.
The probe can comprise a detectable label that is used to detect the binding of the probe to a target analyte. The probe can be directly or indirectly bound to, hybridized to conjugated to, or covalently linked to the target analyte.
As used herein, the tem detectable label refers to a molecule bound to a probe that is capable of generating a detectable optical signal when the probe is bound to a target analyte and imaged using an optical imaging system. The detectable label can be directly or indirectly bound to, hybridized to, conjugated to, or covalently linked to the probe. In some embodiments, the detectable label is a fluorescent molecule or a chemiluminescent molecule. The probe can be detected optically via the detectable label.
As used herein, the term optical distribution model refers to a statistical distribution of probabilities for light detection from a point source. These include, for example, a Gaussian distribution. The Gaussian distribution can be modified to include anticipated aberrations in detection to generate a point spread function as an optical distribution model.
Provided herein are systems and methods that facilitate optical detection and discrimination of probes bound to tightly packed analytes bound to the surface of a substrate. In part, the methods and systems described herein rely on repeated detection of a plurality of target analytes on the surface of a substrate to improve the accuracy of identification of a relative location of each analyte on the substrate. This information can then be used to perform signal deconvolution on each image of a field of the substrate for each cycle to reliably identify a signal from a probe bound to the target analyte. In some embodiments, this type of deconvolution processing can be used to distinguish between different probes bound to the target analyte that have overlapping emission spectrum when activated by an activating light. In some embodiments, the deconvolution processing can be used to separate optical signals from neighboring analytes. This is especially useful for substrates with analytes having a density wherein optical detection is challenging due to the diffraction limit of optical systems.
In some embodiments, the methods and systems described herein are particularly useful in sequencing. By providing methods and systems that facilitate reliable optical detection on densely packed substrates, costs associated with sequencing, such as reagents, number of clonal molecules used, processing and read time, can all be reduced to greatly advance sequencing technologies, specifically, single molecule sequencing by synthesis using optically detected nucleotides.
Although the systems and methods described herein have important implications for advancing sequencing technology, the methods and systems described herein are generally applicable to optical detection of analytes bound to the surface of a substrate, especially on the single molecule level.
Sequencing Cost Reduction
Sequencing technologies include image based systems developed by companies such as Illumina and Complete Genomics and electrical based systems developed by companies such as Ion Torrent and Oxford Nanopore. Image based sequencing systems currently have the lowest sequencing costs of all existing sequencing technologies. Image based systems achieve low cost through the combination of high throughput imaging optics and low cost consumables. However, prior art optical detection systems have minimum center-to-center spacing between adjacent resolvable molecules at about a micron, in part due to the diffraction limit of optical systems. In some embodiments, described herein are methods for attaining significantly lower costs for an image based sequencing system using existing biochemistries using cycled detection, determination of precise positons of analytes, and use of the positional information for highly accurate deconvolution of imaged signals to accommodate increased packing densities that operate below the diffraction limit.
Provided herein are systems and methods to facilitate imaging of signals from analytes immobilized on a surface with a center-to-center spacing below the diffraction limit. These systems and methods use advanced imaging systems to generate high resolution images, and cycled detection to facilitate positional determination of molecules on the substrate with high accuracy and deconvolution of images to obtain signal identity for each molecule on a densely packed surface with high accuracy. These methods and systems allow single molecule sequencing by synthesis on a densely packed substrate to provide highly efficient and very high throughput polynucleotide sequence determination with high accuracy.
The major cost components for sequencing systems are primarily the consumables which include biochip and reagents and secondarily the instrument costs. To reach a $10 30× genome, a 100 fold cost reduction, the amount of data per unit area needs to increase by 100 fold and the amount of reagent per data point needs to drop by 10 fold.
High Density Distributions of Analytes on a Surface of a Substrate
Imaging Densely Packed Single Biomolecules and the Diffraction Limit
The primary constraint for increased molecular density for an imaging platform is the diffraction limit. The equation for the diffraction limit of an optical system is:
where D is the diffraction limit, λ is the wavelength of light, and NA is the numerical aperture of the optical system. Typical air imaging systems have NA's of 0.6 to 0.8. Using λ=600 nm, the diffraction limit is between 375 nm and 500 nm. For a water immersion system the NA is ˜1.0, giving a diffraction limit of 300 nm.
If features on an array or other substrate surface comprising biomolecules are too close, two optical signals will overlap so substantially so you just see a single blob that cannot be reliably resolved based on the image alone. This can be exacerbated by errors introduced by the optical imaging system, such as blur due to inaccurate tracking of a moving substrate, or optical variations in the light path between the sensor and the surface of a substrate.
The transmitted light or fluorescence emission wavefronts emanating from a point in the specimen plane of the microscope become diffracted at the edges of the objective aperture, effectively spreading the wavefronts to produce an image of the point source that is broadened into a diffraction pattern having a central disk of finite, but larger size than the original point. Therefore, due to diffraction of light, the image of a specimen never perfectly represents the real details present in the specimen because there is a lower limit below which the microscope optical system cannot resolve structural details.
The observation of sub-wavelength structures with microscopes is difficult because of the diffraction limit. A point object in a microscope, such as a fluorescent protein or nucleotide single molecule, generates an image at the intermediate plane that consists of a diffraction pattern created by the action of interference. When highly magnified, the diffraction pattern of the point object is observed to consist of a central spot (diffraction disk) surrounded by a series of diffraction rings. Combined, this point source diffraction pattern is referred to as an Airy disk.
The size of the central spot in the Airy pattern is related to the wavelength of light and the aperture angle of the objective. For a microscope objective, the aperture angle is described by the numerical aperture (NA), which includes the term sin θ, the half angle over which the objective can gather light from the specimen. In terms of resolution, the radius of the diffraction Airy disk in the lateral (x,y) image plane is defined by the following formula Abbe Resolutionxy=λ/2NA, where λ is the average wavelength of illumination in transmitted light or the excitation wavelength band in fluorescence. The objective numerical aperture (NA=n·sin(θ)) is defined by the refractive index of the imaging medium (n; usually air, water, glycerin, or oil) multiplied by the sine of the aperture angle (sin(θ)). As a result of this relationship, the size of the spot created by a point source decreases with decreasing wavelength and increasing numerical aperture, but always remains a disk of finite diameter. The Abbe resolution (i.e., Abbe limit) is also referred to herein as the diffraction limit and defines the resolution limit of the optical system.
If the distance between the two Airy disks or point-spread functions is greater than this value, the two point sources are considered to be resolved (and can readily be distinguished). Otherwise, the Airy disks merge together and are considered not to be resolved.
Thus, light emitted from a single molecule detectable label point source with wavelength λ, traveling in a medium with refractive index n and converging to a spot with half-angle θ will make a diffraction limited spot with a diameter: d=λ/2*NA. Considering green light around 500 nm and a NA (Numerical Aperture) of 1, the diffraction limit is roughly d=λ/2=250 nm (0.25 μm), which limits the density of analytes such as single molecule proteins and nucleotides on a surface able to be imaged by conventional imaging techniques. Even in cases where an optical microscope is equipped with the highest available quality of lens elements, is perfectly aligned, and has the highest numerical aperture, the resolution remains limited to approximately half the wavelength of light in the best case scenario. To increase the resolution, shorter wavelengths can be used such as UV and X-ray microscopes. These techniques offer better resolution but are expensive, suffer from lack of contrast in biological samples and may damage the sample.
Deconvolution
Deconvolution is an algorithm-based process used to reverse the effects of convolution on recorded data. The concept of deconvolution is widely used in the techniques of signal processing and image processing. Because these techniques are in turn widely used in man scientific and engineering disciplines, deconvolution finds many applications.
In optics and imaging, the term “deconvolution” is specifically used to refer to the process of reversing the optical distortion that takes place in an optical microscope, electron microscope, telescope, or other imaging instrument, thus creating clearer images. It is usually done in the digital domain by a software algorithm, as part of a suite of microscope image processing techniques.
The usual method is to assume that the optical path through the instrument is optically perfect, convolved with a point spread function (PSF), that is, a mathematical function that describes the distortion in terms of the pathway a theoretical point source of light (or other waves) takes through the instrument. Usually, such a point source contributes a small area of fuzziness to the final image. If this function can be determined, it is then a matter of computing its inverse or complementary function, and convolving the acquired image with that. Deconvolution maps to division in the Fourier co-domain. This allows deconvolution to be easily applied with experimental data that are subject to a Fourier transform. An example is NMR spectroscopy where the data are recorded in the time domain, but analyzed in the frequency domain. Division of the time-domain data by an exponential function has the effect of reducing the width of Lorenzian lines in the frequency domain. The result is the original, undistorted image.
However, for diffraction limited imaging, deconvolution is also needed to further refine the signals to improve resolution beyond the diffraction limit, even if the point spread function is perfectly known. It is very hard to separate two objects reliably at distances smaller than the Nyquist distance. However, described herein are methods and systems using cycled detection, analyte position determination, alignment, and deconvolution to reliably detect objects separated by distances much smaller than the Nyquist distance.
Sequencing
Optical detection imaging systems are diffraction-limited, and thus have a theoretical maximum resolution of 300 nm with fluorophores typically used in sequencing. To date, the best sequencing Systems have had center-to-center spacings between adjacent polynucleotides of ˜600 nm on their arras, or ˜2× the diffraction limit. This factor of 2× is needed to account for intensity, array & biology variations that can result in errors in position. In order to achieve a $10 genome, an approximately 200 nm center to center spacing is required, which requires sub-diffraction-limited imaging capability.
For sequencing, the purpose of the system and methods described herein are to resolve polynucleotides that are sequenced on a substrate with a center-to-center spacing below the diffraction limit of the optical system.
As described herein, we provide methods and systems to achieve sub-diffraction-limited imaging in part by identifying a position of each analyte with a high accuracy (e.g., 10 nm RMS or less). By comparison, state of the art Super Resolution systems (Harvard/STORM) can only identify location with an accuracy down to 20 nm RMS, 2× worse than this system. Thus, the methods and system disclosed herein enable sub-diffraction limited-imaging to identify densely-packed molecules on a substrate to achieve a high data rate per unit of enzyme, data rate per unit of time, and high data accuracy to achieve a $10 genome. These sub-diffraction limited imaging techniques are broadly applicable to techniques using cycled detection as described herein,
As described herein, each of the detection methods and systems required cycled detection to achieve sub-diffraction limited imaging. Cycled detection includes the binding and imaging or probes, such as antibodies or nucleotides, bound to detectable labels that are capable of emitting a visible light optical signal By using positional information from a series of images of a field from different cycles, deconvolution to resolve signals from densely packed substrates can be used effectively to identify individual optical signals from signals obscured due to the diffraction limit of optical imaging. After multiple cycles the precise location of the molecule will become increasingly more accurate. Using this information additional calculations can be performed to aid in crosstalk correction regarding known asymmetries in the crosstalk matrix occurring due to pixel discretization effects.
Methods and systems using cycled probe binding and optical detection are described in US Publication No. 20150330974, Digital Analysis of Molecular Analytes Using Single Molecule Detection, published Nov. 19, 2015, incorporated herein by reference in its entirety.
In some embodiments, the raw images are obtained using sampling that is at least at the Nyquist limit to facilitate more accurate determination of the oversampled image. Increasing the number of pixels used to represent the image by sampling in excess of the Nyquist limit (oversampling) increases the pixel data available for image processing and display.
Theoretically, a bandwidth-limited signal can be perfectly reconstructed if sampled at the Nyquist rate or above it. The Nyquist rate is defined as twice the highest frequency component in the signal. Oversampling improves resolution, reduces noise and helps avoid aliasing and phase distortion by relaxing anti-aliasing filter performance requirements. A signal is said to be oversampled by a factor of N if it is sampled at N times the Nyquist rate.
Thus, in some embodiments, each image is taken with a pixel size no more than half the wavelength of light being observed. In some embodiments, a pixel size of 162.5 nm×162.5 nm is used in detection to achieve sampling at or above the Nyquist limit. Sampling at a frequency of at least the Nyquist limit during raw imaging of the substrate is preferred to optimize the resolution of the system or methods described herein. This can be done in conjunction with the deconvolution methods and optical systems described herein to resolve features on a substrate below the diffraction limit with high accuracy.
Processing Images from Different Cycles
There are several barriers overcome by the present invention to achieve sub-diffraction limited imaging.
Pixelation error is present in raw images and prevents identification of information present from the optical signals due to pixelation. Sampling at least at the Nyquist frequency and generation of an oversampled image as described herein each assist in overcoming pixilation error.
The point-spread (PSF) of various molecules overlap because the PSF size is greater than the pixel size (below Nyquist) and because the center-to-center spacing is so small that crosstalk due to spatial overlap occurs. Nearest neighbor variable regression (for center-to center crosstalk) can be used to help with deconvolution of multiple overlapping optical signals. But this can be improved if we know the relative location of each analyte on the substrate and have good alignment of images of a field.
After multiple cycles the precise location of the molecule will become increasingly more accurate. Using this information additional calculations can be performed to aid in deconvolution by correcting for known asymmetries in the spatial overlap of optical signals occurring due to pixel discretization effects and the diffraction limit. They can also be used to correct for overlap in emission spectrum from different emission spectrum.
Highly accurate relative positional information for each analyte can be achieved by overlaying images of the same field from different cycles to generate a distribution of measured peaks from optical signals of different probes bound to each analyte. This distribution can then be used to generate a peak signal that corresponds to a single relative location of the analyte Images from a subset of cycles can be used to generate relative location information for each analyte. In some embodiments, this relative position information is provided in a localization file.
The specific area imaged for a field for each cycle may vary from cycle to cycle. Thus, to improve the accuracy of identification of analyte position for each image, an alignment between images of a field across multiple cycles can be performed. From this alignment, offset information compared to a reference file can then be identified and incorporated into the deconvolution algorithms to further increase the accuracy of deconvolution and signal identification for optical signals obscured due to the diffraction limit. In some embodiments, this information is provided in a Field Alignment File.
Signal Detection (Cross-Talk/Nearest Neighbor)
Once relative positional information is accurately determined for analytes on a substrate and field images from each cycle are aligned with this positional information, analysis of each oversampled image using crosstalk and nearest neighbor regression can be used to accurately identify an optical signal from each analyte in each image.
In some embodiments, a plurality of optical signals obscured by the diffraction limit of the optical system are identified for each of a plurality of biomolecules immobilized on a substrate and bound to probes comprising a detectable label. In some embodiments, the probes are incorporated nucleotides and the series of cycles is used to determine a sequence of a polynucleotide immobilized on the array using single molecule sequencing by synthesis.
Simulations of Deconvolution Applied to Images
Molecular densities are limited by crosstalk from neighboring molecules.
The physical size of the molecule will broaden the spot roughly half the size of the binding area. For example, for an 80 nm spot the pitch will be increased by roughly 40 nm. Smaller spot sizes may be used, but this will have the trade-off that fewer copies will be allowed and greater illumination intensity will be required. A single copy provides the simplest sample preparation but requires the greatest illumination intensity.
Methods for sub-diffraction limit imaging discussed to this point involve image processing techniques of oversampling, deconvolution and crosstalk correction. Described herein are methods and systems that incorporate determination of the precise relative location analytes on the substrate using information from multiple cycles of probe optical signal imaging for the analytes. Using this information additional calculations can be performed to aid in crosstalk correction regarding known asymmetries in the crosstalk matrix occurring due to pixel discretization effects.
In some embodiments, as shown in
In some embodiments, as shown in
In some embodiments, as shown in
In some embodiments, optical signals are deconvolved from densely packed substrates as shown in
Under the image analysis steps shown in
Peaks from optical signals detected in each raw image or present in the oversampled image are then identified and intensity and position information for each detected analyte is placed into a peak file for further processing.
In some embodiments, N raw images corresponding to all images detected from each cycle and each field of a substrate or output into N oversampled images and N peak files for each imaged field. The peak file comprises a relative position of each detected analyte for each image. In some embodiments, the peak file also comprises intensity information for each detected analyte. In some embodiments, one peak file is generated for each color and each field in each cycle. In some embodiments, each cycle further comprises multiple passes, such that one peak file can be generated for each color and each field for each pass in each cycle. In some embodiments, the peak file specifies peak locations from optical signals within a single field.
In preferred embodiments, the peak file includes XY position information from each processed oversampled image of a field for each cycle. The XY position information comprises estimated coordinates of the locations of each detected detectable label from a probe (such as a fluorophore) from the oversampled image. The peak file can also include intensity information from the signal from each individual detectable label.
Generation of an oversampled image is used to overcome pixelation error to identify information present that cannot be extracted due to pixelation. Initial processing of the raw image by smoothing and deconvolution helps to provide more accurate information in the peak files so that the position of each analyte can be determined with higher accuracy, and this information subsequently can be used to provide a more accurate determination of signals obscured in diffraction limited imaging.
In some embodiments, the raw images are obtained using sampling that is at least at the Nyquist limit to facilitate more accurate determination of the oversampled image. Increasing the number of pixels used to represent the image by sampling in excess of the Nyquist limit (oversampling) increases the pixel data available for image processing and display.
Theoretically, a bandwidth-limited signal can be perfectly reconstructed if sampled at the Nyquist rate or above it. The Nyquist rate is defined as twice the highest frequency component in the signal. Oversampling improves resolution, reduces noise and helps avoid aliasing and phase distortion by relaxing anti-aliasing filter performance requirements. A signal is said to be oversampled by a factor of N if it is sampled at N times the Nyquist rate.
Thus, in some embodiments, each image is taken with a pixel size no more than half the wavelength of light being observed. In some embodiments, a pixel size of 162.5 nm×162.5 nm is used in detection to achieve sampling at or above the Nyquist limit.
Smoothing uses an approximating function capture important patterns in the data, while leaving out noise or other fine-scale structures/rapid phenomena. In smoothing, the data points of a signal are modified so individual points are reduced, and points that are lower than the adjacent points are increased leading to a smoother signal. Smoothing is used herein to smooth the diffraction limited optical signal detected in each image to better identify peaks and intensities from the signal.
Although each raw image is diffraction limited, described herein are methods that result in collection of multiple signals from the same analyte from different cycles. An embodiment of this method is shown in the flowchart in
As shown in
In some embodiments, the relative position information for each analyte is determined on average to less than a 10 nm standard deviation (i.e., RMS, or root mean square). In some embodiments, the relative position information for each analyte is determined on average to less than a 10 nm 2× standard deviation. In some embodiments, the relative position information for each analyte is determined on average to less than a 10 nm 3× standard deviation. In some embodiments, the relative position information for each analyte is determined to less than a 10 nm median standard deviation. In some embodiments, the relative position information for each analyte is determined to less than a 10 nm median 2× standard deviation. In some embodiments, the relative position information for each analyte is determined to less than a 10 nm median 3× standard deviation.
From a subset of peak files for a field from different cycles, a localization file is generated to determine a location of analytes on the array. As shown in
The peaks determined from at least a subset of position information from the cycles are then recorded in a localization file, which comprises a measure of the relative position of each detected analyte with an accuracy below the diffraction limit. As described, images from only subset of cycles are needed to determine this information.
As shown in
In some embodiments, a field alignment file is generated by alignment of images from a single field by determining offset information relative to a master file from the field. One field alignment file is generated for each field. This file is generated from all images of the field from all cycles, and includes offset information for all images of the field relative to a reference image from the field.
In some embodiments, before alignment, each peak file is normalized with a point spread function, followed by generation of an artificial image from the normalized peak file and Fourier transform of the artificial image. The Fourier transform of the artificial image of the normalized peak file is then convolved with a complex conjugate of the Fourier transform of an artificial image from the normalized localization file for the corresponding field. This is done for each peak file for each cycle. The resulting files then undergo an inverse Fourier transform to regenerate image files, and the image files are aligned relative to the reference file from the field to generate offset information for each image file. In some embodiments, this alignment includes a fine fit relative to a reference file.
The field alignment ile thus contains offset information for each oversampled image, and can be used in conjunction with the localization file for the corresponding field to generate highly accurate relative position for each analyte for use in the subsequent “Extract Intensities” steps.
As an example where 20 cycles are performed on a field, and one image is generated for each of 4 colors to be detected, thus generating 80 images of the field, one Field Alignment file is generated for all 80 images (20 cycles*4 colors) taken of the field. In some embodiments, the field alignment file contents include: the field, the color observed for each image, the step type in the cycled detection (e.g., binding or stripping), and the image offset coordinates relative to the reference image.
In some embodiments, during the alignment process XY “shifts” or “residuals” needed to align 2 images are calculated, and the process is repeated for remaining images, best fit residual to apply to all is calculated.
In some embodiments, residuals that exceed a threshold are thrown out, and best fit is re-calculated. This process is repeated until all individual residuals are within the threshold
Each oversampled image is then deconvolved using the accurate position information from the localization file and the offset information from the field alignment file. An embodiment of the intensity extraction step is shown in
In certain embodiments, emission spectra overlap between different signals (i.e. “cross-talk”). For example, during sequencing by synthesis, the four dyes used in the sequencing process typically have some overlap in emission spectra.
In particular embodiments, a problem of assigning a color (for example, a base call) to different features in a set of images obtained for a cycle when cross talk occurs between different color channels and when the cross talk is different for different sets of images can be solved by cross-talk regression in combination with the localization and field alignment files for each oversampled image to remove overlapping emission spectrums from optical signals from each different detectable label used. This further increases the accuracy of identification of the detectable label identity for each probe bound to each analyte on the substrate.
Thus, in some embodiments, identification of a signal and/or its intensity from a single image of a field from a cycle as disclosed herein uses the following features: 1) Oversampled Image provides intensities and signals at defined locations. 2) Accurate Relative Location—Localization File (provides location information from information from at least a subset of cycles) and Field Alignment File (provides offset/alignment information for all images in a field). 3) Image Processing—Nearest Neighbor Variable Regression (spatial deconvolution) and Cross-talk regression (emission spectra deconvolution) using accurate relative position information for each analyte in a field. Accurate identification of probes (e.g., antibodies for detection or complementary nucleotides for sequencing) for each analyte.
The effects of the methods and systems disclosed herein are illustrated in simulated cross-talk plots shown in
The methods described above and in
In some embodiments, molecules to be sequenced, single or multiple copies, will be bound to the surface using covalent linkages, by hybridizing to capture oligonucleotide on the surface, or by other non-covalent binding. The bound molecules will remain on the surface for hundreds of cycles and can be re-interrogated with different primer sets, following stripping of the initial sequencing primers, to confirm the presence of specific variants.
In one embodiment, the fluorophores and blocking groups may be removed using chemical reactions.
In another embodiment, the fluorescent and blocking groups may be removed using UV light.
In one embodiment, the molecules to be sequenced could be immobilized on reactive surfaces that have 50-100 nM diameters and these areas would be spaced at a pitch of 150-300 nM. These molecules may have barcodes, attached onto them for target deconvolution and a sequencing primer binding region for initiating sequencing. Buffers will contain appropriate amounts of DNA polymerase to enable an extension reaction. These sited could contain 10-100 copies of the target to be sequenced generated by any of the gene amplification methods available (PCR, whole genome amplification etc.)
In another embodiment, single target molecules, tagged with a barcode and a primer annealing site would be immobilized on a 20-50 nM diameter reactive surface spaced with a pitch of 60-150 nM. The molecules would be sequenced individually.
In one embodiment, a primer would bind to the target and would be extended using one dNTP at a time with a single or multiple fluorophore(s); the surface would be imaged, the fluorophore would be removed and washed and the process repeated to generate a second extension. The presence of multiple fluorophores on the same dNTP will enable defining the number of repeats nucleotides present in some regions of the genome (2 to 5 or more).
In a different embodiment, following primer annealing, all four dNTPs with fluorophores and blocked 3′ hydroxyl groups would be used in the polymerase extension reaction, the surface would be imaged and the fluorophore and blocking groups removed and the process repeated for multiple cycles.
In another embodiment, the sequences could be inferred based on ligation reactions that anneal specific probes that ligate based on the presence of a specific nucleotides at a given position.
A random array may be used which will have improved densities over prior art random arras using the techniques outlined above, however random arrays generally have 4× to 10× reduced areal densities of ordered arrays. Advantages of a random array include a uniform, non-patterned surface for the chip and the use of shorter nucleic acid strands because there is no need to rely on the exclusionary properties of longer strands.
Those skilled in the art will recognize, or be able to ascertain using no more than routine experimentation, many equivalents to the specific embodiments in accordance with the invention described herein. The scope of the present invention is not intended to be limited to the above Description, but rather is as set forth in the appended claims.
In the claims, articles such as “a,” “an,” and “the” may mean one or more than one unless indicated to the contrary or otherwise evident from the context. Claims or descriptions that include “or” between one or more members of a group are considered satisfied if one, more than one, or all of the group members are present in, employed in, or otherwise relevant to a given product or process unless indicated to the contrary or otherwise evident from the context. The invention includes embodiments in which exactly one member of the group is present in, employed in, or otherwise relevant to a given product or process. The invention includes embodiments in which more than one, or all of the group members are present in, employed in, or otherwise relevant to a given product or process.
It is also noted that the term “comprising” is intended to be open and permits but does not require the inclusion of additional elements or steps. When the term “comprising” is used herein, the term “consisting of” is thus also encompassed and disclosed.
Where ranges are given, endpoints are included. Furthermore, it is to be understood that unless otherwise indicated or otherwise evident from the context and understanding of one of ordinary skill in the art values that are expressed as ranges can assume any specific value or subrange within the stated ranges in different embodiments of the invention, to the tenth of the unit of the lower limit of the range, unless the context clearly dictates otherwise.
All cited sources, for example, references, publications, databases, database entries, and art cited herein, are incorporated into this application by reference, even if not expressly stated in the citation. In case of conflicting statements of a cited source and the instant application, the statement in the instant application shall control.
Section and table headings are not intended to be limiting.
Below are examples of specific embodiments for carrying out the present invention. The examples are offered for illustrative purposes only, and are not intended to limit the scope of the present invention in anyway. Efforts have been made to ensure accuracy with respect to numbers used (e.g., amounts, temperatures, etc.), but some experimental error and deviation should, of course, be allowed for.
The practice of the present invention will employ, unless otherwise indicated, conventional methods of protein chemistry, biochemistry, recombinant DNA techniques and pharmacology, within the skill of the art. Such techniques are explained fully in the literature. See, e.g., T. E. Creighton, Proteins: Structures and Molecular Properties (W.H. Freeman and Company, 1993); A. L. Lehninger, Biochemistry (Worth Publishers, Inc., current addition); Sambrook, et al., Molecular Cloning: A Laboratory Manual (2nd Edition, 1989); Methods In Enzymology (S. Colowick and N. Kaplan eds., Academic Press, Inc.); Remington's Pharmaceutical Sciences, 18th Edition (Easton, Pa.: Mack Publishing Company, 1990); Carey and Sundberg Advanced Organic Chemistry 3rd Ed. (Plenum Press) Vols A and B (1992).
Methods below will describe how to utilize a square ordered array where the pitch ranges between 200 nm and 333 nm. Additional methods will be described that allow even smaller pitches. An imaging system is described in International Application PCT/US2018/020737, filed Mar. 2, 2018 and incorporated herein by reference, which will be used as a reference system which enables sub-diffraction limit imaging. The optical system can include multiple 2,048 by 2,048 pixel cameras operating up to 100 Hz frames per second (fps) with field size 332.8 um by 332.8 um. This system is capable of measuring as little as a single fluor at and above 90 fps. Using this system with 1-10 copies (or 1-10 fluorophores) per molecule at 85 fps achieves the necessary throughput to image a 63 mm×63 mm slide in under 15 minutes. Biochemistry cycles and imaging are continuously and simultaneously performed, either by using two chips or by dividing a single chip into at least 2 regions.
Single-molecule sequencing using sequencing-by-synthesis approach was evaluated on the Apton System. To test the methodology, single-stranded DNA templates with 5′ phosphate group were first attached to the chip with a tecarbohydrazide activated silicon surface of the flow cell through EDC (1-Ethyl-3-(3-mplate dimethylaminopropyl)carbodiimide) chemistry. The sequencing primer vas the annealed the target immobilized on the surface. The sequencing templates used in our initial studies included synthetic oligonucleotide containing EGFR L858R, EGFR T790M, and BRAF V600E mutations and two cDNA samples reversed transcribed from ERCC 00013 and ERCC 00171 control RNA transcripts. After DNA template immobilization and primer annealing, the flow cell is loaded on the Apton instrument for sequencing reactions, which involves multiple cycles of enzymatic single nucleotide incorporation reaction, imaging to detect fluorescence dye detection, followed by chemical cleavage. Therminator IX DNA Polymerase from NEB was used for single base extension reaction, which is a 9° N™ DNA Polymerase variant with an enhanced ability to incorporate modified dideoxynucleotides. Four dNTPs used in the reaction are labeled with 4 different cleavable fluorescent dyes and blocked at 3′-OH group with a cleavable moiety (dCTP-AF488, dATP-AFCy3, dTTP-TexRed, and dGTP-Cy5 from MyChem). During each sequencing reaction cycle, a single labeled dNTP is incorporated and the reaction is terminated because of the 3′-blocking group on dNTP. After dNTP incorporation, the unincorporated nucleotides are removed from the flow-cell by washing and the incorporated fluorescent dye labeled nucleotide is imaged to identify the base. After the images are captured, the fluorescent dye and blocking moiety are cleaved from the incorporated nucleotide using 100 mM TCEP ((tris(2-carboxyethyl)phosphine), pH9.0), allowing subsequent addition of the next complementary nucleotide in next cycle. This extension, detection and cleavage cycle is then repeated to increase the read length.
The synthetic oligonucleotides used were around 60 nucleotides long. A primer that had a sequence ending one base prior to the mutation in codon 790 was used to enable the extension n reaction. The surface was imaged post incorporation of nucleotides by the DNA polymerase and after the cleavage reaction with TCEP. The yellow circle indicates the location of the template molecule that was aligned using data from 10 consecutive cycles of dye incorporation. Molecules were identified with known color incorporation sequences, following that the actual base incorporations are identified by visual inspections which is labor-intensive.
Dye labeled nucleotides were used to sequence cDNA generated from RNA templates. RNA used was generated by T7 transcription from cloned ERCC control plasmids.
Specifically, cDNA templates corresponding to transcripts generated from the ERCC (External RNA Controls Consortium) control plasmids by T7 transcription were sequenced. The cDNA molecule generated were >350 nucleotides long. The surface was imaged post incorporation of nucleotides by the DNA polymerase and after the cleavage reaction with TCEP. The yellow circle in
To improve accuracy of detection, the molecules undergo successive cycles of probe binding and stripping, in this case 30 cycles. In each round, the image is processed to determine the location of the molecules. The images are background subtracted, oversampled by 2×, after which peaks are identified. Multiple layers of cycles are overlaid on a 20 nm grid. The location variance is the standard deviation or the radius divided by the square root of the number of measurements.
It is to be understood that the words which have been used are words of description rather than limitation, and that changes may be made within the purview of the appended claims without departing from the true scope and spirit of the invention in its broader aspects.
While the present invention has been described at some length and with some particularity with respect to the several described embodiments, it is not intended that it should be limited to any such particulars or embodiments or any particular embodiment, but it is to be construed with references to the appended claims so as to provide the broadest possible interpretation of such claims in view of the prior art and, therefore, to effectively encompass the intended scope of the invention.
All publications, patent applications, patents, and other references mentioned herein are incorporated by reference in their entirety. In case of conflict, the present specification, including definitions, will control. In addition, section headings, the materials, methods, and examples are illustrative only and not intended to be limiting.
This application is a continuation of U.S. application Ser. No. 16/458,977, filed Jul. 1, 2019, which is a continuation of U.S. application Ser. No. 15/925,656, filed Mar. 19, 2018, which claims the benefit of U.S. Provisional Application No. 62/473,163, filed Mar. 17, 2017, the entire disclosures of which are hereby incorporated by reference in their entirety for all purposes.
Number | Date | Country | |
---|---|---|---|
62473163 | Mar 2017 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 16458977 | Jul 2019 | US |
Child | 17084017 | US | |
Parent | 15925656 | Mar 2018 | US |
Child | 16458977 | US |