The heart has four native valves, including the aortic valve, pulmonary valve, mitral valve (also known as the left atrioventricular valve), and the tricuspid valve (also known as the right atrioventricular valve). When these valves begin to fail, for example by not fully coapting and allowing retrograde blood flow (or regurgitation) across the valve, it may be desirable to repair or replace the valve. Prosthetic replacement heart valves may be surgically implanted via an open chest and open-heart procedure while the patient is on cardiopulmonary bypass. However, such procedures are extremely invasive, and frail patients, who may be the most likely to need a prosthetic heart valve, may not be likely to survive such a procedure. More recently, prosthetic heart valves have been trending to less invasive procedures, including collapsible and expandable heart valves that can be delivered through the vasculature in a transcatheter procedure.
Unless otherwise indicated, as used herein, the term “tricuspid valve” refers to the right atrioventricular valve, as opposed to just a generic term for a three-leaflet valve. Initial human trials to replace the native tricuspid valve in a transcatheter procedure (e.g., via the femoral vein) have shown promising results, with patients experiencing significant improvements in quality of life after the prosthetic valve implantation. It is thought that important characteristics of a successful transcatheter prosthetic tricuspid valve device include not only a good clinical outcome for the patient, but the ease of use of the tricuspid valve, including for example having a small enough size to be able to avoid a surgical cut down of the patient's femoral vein for delivery.
According to one aspect of the disclosure, a prosthetic heart valve system is for replacing a native right atrioventricular valve. The system may include a collapsible and expandable anchor frame. The anchor frame may include a support frame that has a central portion, and atrial and ventricular portions each flared radially outwardly from the central portion, the atrial and ventricular portions sized to sandwich or clamp an annulus of the right atrioventricular valve therebetween. An atrial sheet may be coupled to the atrial portion of the support frame and may extend radially inwardly to a central aperture in the atrial sheet, the atrial sheet being formed of a material that is substantially impermeable to blood. The system may include a generally cylindrical leaflet support structure, an inflow end of the leaflet support structure coupled to the atrial sheet. The atrial sheet may define an opening positioned radially outside of the leaflet support structure so that blood flowing around an exterior of the leaflet support structure can pass through the opening.
The leaflet support structure may be the central portion of the support frame, and an outer sealing fabric may extend between the atrial and ventricular portions of the support frame. In another embodiment, the support frame is an outer stent, and the leaflet support structure is an inner stent. In a further embodiment, the support frame is an outer stent, and the leaflet support structure is a fabric tube. The system may further include a collapsible and expandable prosthetic heart valve including a stent and a plurality of prosthetic leaflets, the prosthetic heart valve configured to be expanded into and received within the fabric tube of the anchor frame. In another embodiment, a plurality of prosthetic leaflets may be directly coupled to the leaflet support structure, the prosthetic leaflets forming a valve that allows blood to flow from the inflow end of the leaflet support structure to the outflow end of the leaflet support structure, but generally blocks blood from flowing from the outflow end of the leaflet support structure to the inflow end of the leaflet support structure. A radiopaque marker may be coupled to the atrial sheet adjacent to the opening.
The system may further include a ventricular sheet coupled to the ventricular portion of the support frame and extending radially inwardly to a central aperture in the ventricular sheet, wherein the ventricular sheet defines an opening positioned radially outside of the leaflet support structure, a tube extending from the opening in the ventricular sheet to the opening in the atrial sheet to form a shunt. The system may include an expandable occluder configured to be received within the shunt and to seal blood flow across the shunt, the occluder being formed as a braided mesh of metal strands. The system may also include an expandable stent configured to be received within the shunt and to limit blood flow across the shunt, the stent having an hourglass shape.
A closure member may be coupled to the shunt, the closure member including a bioabsorbable member maintaining the closure member in an open condition in which blood is free to flow through the shunt. In one embodiment, the closure member is formed of a shape-memory material, the closure member including two apices and two “C”-shaped sides, the two “C”-shaped sides nesting with each other in the absence of applied forces. The bioabsorbable member may be a wire connected to the two apices of the closure member, the wire maintaining the closure member in an open condition in which the two “C”-shaped sides form a generally circular passageway, the tube of the shunt passing through the generally circular passageway. In another embodiment, the closure member is formed of a shape-memory material, the closure member having a closed condition in the absence of applied forces, the closure member forming a spiral shape with an interior diameter when in the closed condition. In the open condition of the closure member, two ends of the closure member overlap, the two ends being coupled together by the bioabsorbable member so that an interior diameter of the closure member in the open condition is larger than the interior diameter of the closure member in the closed condition. The bioabsorbable member may be a sheath that receives the two ends of the closure member within the sheath, or alternatively it may be a wire that wraps around the two ends of the closure member within the sheath.
The tube forming the shunt may be cylindrical, or it may be bean-shaped in transverse cross-section. A semi-permeable membrane may be positioned within the shunt.
According to another aspect of the disclosure, a prosthetic heart valve system is for replacing a native right atrioventricular valve. The system may include a collapsible and expandable anchor frame. The anchor frame may include a support frame that has a central portion, and atrial and ventricular portions each flared radially outwardly from the central portion, the atrial and ventricular portions sized to sandwich or clamp an annulus of the right atrioventricular valve therebetween. The anchor frame may include an atrial sheet coupled to the atrial portion of the support frame and extending radially inwardly to a central aperture in the atrial sheet, and a ventricular sheet coupled to the ventricular portion of the support frame and extending radially inwardly to a central aperture in the ventricular sheet. The anchor frame may include a generally cylindrical leaflet support structure, an inflow end of the leaflet support structure coupled to the atrial sheet, and an outflow end of the leaflet support structure coupled to the ventricular sheet to provide a conduit from the central aperture in the atrial sheet to the central aperture in the ventricular sheet through the leaflet support structure. A first radiopaque marker may be positioned on the atrial sheet, and a second radiopaque marker may be positioned on the ventricular sheet. In the expanded condition of the support frame, a line of safe passage (e.g., a line of clear or unobstructed) extends between the first radiopaque marker and the second radiopaque marker, the line of safe passage extending along a gap between the leaflet support structure and the central portion of the support frame.
According to a further aspect of the disclosure, a prosthetic heart valve system is for replacing a native right atrioventricular valve. The system may include a collapsible and expandable anchor frame. The anchor frame may include a support frame that has a central portion, and atrial and ventricular portions each flared radially outwardly from the central portion, the atrial and ventricular portions sized to sandwich or clamp an annulus of the right atrioventricular valve therebetween. The anchor frame may include an atrial sheet coupled to the atrial portion of the support structure and extending radially inwardly to a central aperture in the atrial sheet, and a ventricular sheet coupled to the ventricular portion of the support structure and extending radially inwardly to a central aperture in the ventricular sheet. The anchor frame may include a generally cylindrical leaflet support structure, an inflow end of the leaflet support structure coupled to the atrial sheet, and an outflow end of the leaflet support structure coupled to the ventricular sheet to provide a conduit from the central aperture in the atrial sheet to the central aperture in the ventricular sheet through the leaflet support structure. A guidewire may be pre-assembled to the prosthetic heart valve, such that in an expanded condition of the support frame, the guidewire extends through the atrial sheet and through the ventricular sheet along a pathway positioned radially outside of the leaflet support structure. The guidewire may have a first end extending beyond the atrial sheet and a second end extending beyond the ventricular sheet, an intermediate portion of the guidewire extending from the atrial sheet to the ventricular sheet. The system may include a catheter having a lumen sized and shaped to ride over the guidewire.
One important aspect of transcatheter tricuspid valve implantations is the size of components introduced into the patient's body. For example, it may be preferable for the delivery device that is used to deliver the transcatheter tricuspid valve to be small enough to be delivered through the femoral vein without requiring a surgical cut down of the vein. One desirable size example is about 30 French (10 mm) or below. Further, given the requirement for a transcatheter valve to be delivered through the patient's vasculature, the components of the system must be small enough to pass through the patient's vasculature safely. This size objective may be at odds with the objective of achieving strong anchoring, as more stent material may generally provide the ability for better anchoring, but also result in a larger device. In other words, the inclusion of a relatively large amount of anchoring structures in a prosthetic tricuspid valve may help provide better anchoring, but at a cost of increasing the overall size of the device (and thus the components, such as a delivery device, which will house the prosthetic heart valve).
Counterintuitively, it may also be preferable that a prosthetic tricuspid valve does not immediately resolve all tricuspid regurgitation. For example, in patients with severe or torrential tricuspid regurgitation, there may actually be a danger of transitioning between extreme tricuspid regurgitation to zero regurgitation in what is effectively an instantaneous change following implantation of a prosthetic tricuspid valve. This potential danger lies, at least in part, in the fact that the flow dynamics change nearly instantaneously, while the heart muscle cannot acclimate instantaneously. As an example, a patient with severe or torrential tricuspid regurgitation may have a relatively thin right ventricle due to dilation from heart failure. However, after tricuspid regurgitation is resolved, pressures within the right ventricle may increase, putting the patient at risk of right ventricular rupture or further dilation. This may place undue strain on the right ventricle. One solution to this issue, as described in greater detail below, is to include a shunt within the prosthetic tricuspid valve. The shunt, which may be temporary, may allow a certain amount of regurgitation from the right ventricle to the right atrium, even when the prosthetic leaflets of the prosthetic tricuspid valve are fully coapted. This shunt may reduce the pressure within the right ventricle, compared to what would be expected with zero regurgitation, but increase pressure within the right ventricle compared to the patient's disease state prior to the implantation. In other words, the combination of the coapting prosthetic leaflets and the shunt may decrease, but not eliminate, regurgitation. With this decrease in regurgitation, resistance is increased when the right ventricle contracts, which may allow the right ventricle to strengthen and/or thicken over a period of days, weeks, or months. During this period of days, weeks, or months, the shunt may be configured to naturally close to eliminate any remaining regurgitation, or otherwise may be actively closed during a secondary procedure. By the time the shunt closes, the right ventricle preferably has already acclimated to the new hemodynamics, so that upon complete (or substantially complete) elimination of regurgitation, the right ventricle has already gotten stronger and is more “ready” for the change in hemodynamics that occurs upon complete (or substantially complete) elimination of regurgitation.
Furthermore, in some patients, it may be advisable or necessary to implant a pacemaker along with (or shortly after) a prosthetic tricuspid valve implantation. Prosthetic heart valves may expand into contact with the AV Node, or otherwise cause interference with the natural conduction system of the heart. Thus, pacemakers may be implanted during, or after, a prosthetic heart valve implantation procedure to manage the patient's heart rhythm. If a shunt is included, the shunt may provide a passageway to assist with the placement of one or more pacemaker leads. And even if a shunt is not provided, other features are described below to assist in providing a passageway to assist with the placement of one or more pacemaker leads. Traditionally, pacemaker leads are passed through the center of the valve member of the prosthetic heart valve. In other words, pacemaker leads typically need to be delivered through an area where tissue leaflets (or in some cases fabric leaflets) are actively opening and closing as they perform their valve functionality. After implantation of the pacemaker leads, the leads may pass through the conduit formed by the prosthetic valve leaflets, but the pacemaker leads may create an obstruction that “pins” the prosthetic leaflets or otherwise inhibits the prosthetic leaflets from properly opening and/or closing. If the pacemaker leads inhibit proper coaptation of the prosthetic leaflets, regurgitation may occur. There is also a potential danger that the delivery device containing the pacemaker leads could damage the prosthetic leaflets if the delivery device is passed through the center of the prosthetic valve. Thus, the shunts described herein may provide for safe passageway and positioning of pacemaker leads such that the pacemaker leads (and/or a delivery device for same) do not need to pass through or ever be positioned within the area where the prosthetic leaflets are closing and opening. And as is described in greater detail below, even if a shunt is not included to provide a ready-made conduit for pacemaker implantation, other features may be provided to achieve a similar goal. Although general embodiments have been described above, more specific embodiments are disclosed below.
The atrial portion 102 may be formed as a portion of a stent or other support structure that includes or is formed by a plurality of generally diamond-shaped cells, although other suitable cell shapes, such as triangular, quadrilateral, or polygonal may be appropriate. In some examples, the atrial portion 102 may be formed as a braided mesh, as a portion of a unitary stent, or a combination thereof. According to one example, the stent that includes the atrial portion 102 may be laser cut from a tube of Nitinol and heat-treated to the desired shape so that the stent, including atrial portion 102, is collapsible for delivery, and re-expandable to the set shape during deployment. The atrial portion 102 may be heat treated into a suitable shape to conform to the native anatomy of the tricuspid valve annulus to help provide a seal and/or anchoring between the atrial portion 102 and the native tricuspid valve annulus.
The atrial portion 102 may include features for connecting the atrial portion to a delivery system. For example, the atrial portion 102 may include pins or tabs 122 around which sutures (or suture loops) of the delivery system may wrap so that while the suture loops are wrapped around the pins or tabs 122, the anchor frame 101 maintains a connection to the delivery device. However, in some embodiments, these pins or tabs 122 may be omitted.
The ventricular portion 104 may also be formed as a portion of a stent or other support structure that includes or is formed of a plurality of diamond-shaped cells, although other suitable cell shapes, such as triangular, quadrilateral, or polygonal may be appropriate. In some examples, the ventricular portion 104 may be formed as a braided mesh, as a portion of a unitary stent, or a combination thereof. According to one example, the stent that includes the ventricular portion 104 may be laser cut from a tube of Nitinol and heat-treated to the desired shape so that the ventricular portion 104 is collapsible for delivery, and re-expandable to the set shape during deployment. It should be understood that the atrial portion 102 and ventricular portion 104 may be formed as portions of a single support structure, such as a single stent or braided mesh. However, in other embodiments, the atrial portion 102 and ventricular portion 104 may be formed separately and coupled with one another. It should also be understood that the atrial portion 102 (which may be referred to as a disk) and the ventricular portion 104 (which may also be referred to as a disk) may each be substantially symmetric, or otherwise may each be asymmetric. As an example, there is only minimal annulus structure close to the atrial septum, and thus it may be preferable to have smaller stent cells and/or a smaller disk diameter in this region of the atrial disk, resulting in an asymmetric atrial disk. This may also mean that the anchoring stent 101 should be loaded in a certain orientation within the delivery device, or that the anchoring stent 101 can be rotated or will self-align so that the anchoring stent 101 is deployed in the desired rotational orientation relative to the patient's anatomy. And even if the atrial or ventricular disks are cut in a symmetric pattern, it might be desirable to heat set one or both structures so that the disks are asymmetric in their pre-set shape, for example curving more up or down compared to the rest of the structure. Also, it may be desirable that the cells in the atrial disk that are to be positioned closest to the atrial septum are softer to avoid any damage to the septal wall.
The anchor frame 101 may be configured to expand circumferentially (and radially) and foreshorten axially as the anchor frame 101 expands from the collapsed delivery configuration to the expanded deployed configuration. In the particular illustrated example, the anchor frame 101 includes or defines a plurality of atrial cells 111a in one circumferential row and a plurality of ventricular cells 111b in two circumferential rows. Each of the plurality of cells 111a, 111b may be configured to expand circumferentially and foreshorten axially upon expansion of the anchor frame 101. As shown, the cells 111a-b may each be diamond-shaped. In addition, a third plurality of cells 111c may be provided in additional circumferential rows, in this embodiment three additional center rows, forming the central portion 103.
Still referring to
Anchor frame 201 may include one or more fabric components that may provide one or more functions, for example sealing. In the illustrated embodiment, the anchor frame 201 may include a sealing skirt 220 on a luminal and/or abluminal surface thereof. This sealing skirt 220 may be generally similar to other sealing skirts provided on stents or frames of transcatheter prosthetic heart valves. This luminal and/or abluminal sealing skirt 220 may be formed of any suitable material, including biomaterials such as bovine pericardium, or biocompatible polymers such as ultra-high molecular weight polyethylene, woven polyethylene terephthalate (“PET”), expanded polytetrafluoroethylene (“ePTFE”), or combinations thereof. The sealing skirt 220, particularly if positioned on the abluminal surface of the anchor frame 201, may include a “bump” (or gasket or ring) portion to enhance sealing, similar to that described in U.S. patent application Ser. No. 17/548,984, the disclosure of which is hereby incorporated by reference herein. In addition to sealing skirt 220, the anchor frame 201 may include an atrial sheet 230 and a ventricular sheet 240. In some embodiments, any combination of sealing skirt 220, atrial sheet 230, and ventricular sheet 240 may be formed as an integral member, although in other embodiments, the atrial sheet 230 and ventricular sheet 240 are formed of different materials that provide different functionality and are not formed as integral members.
Referring still to
A valve-receiving member 250, which may be generally cylindrical, may have a first inflow end coupled to the atrial sheet 230 so that the first inflow end is substantially coextensive with the central aperture of the atrial sheet 230, and a second outflow end coupled to the ventricular sheet 240 so that the second outflow end is substantially coextensive with the central aperture of the ventricular sheet 240. The valve-receiving member 250 is preferably formed of a fabric, such as PTFE, UBMWPE, Kevlar braid, Dacron, or biomaterials such as tissue. In some embodiments, the valve-receiving member 250 may be formed of thin wires of Nitinol, stainless steel, or other biocompatible metals or metal alloys formed into a braided, knitted, or woven structure. The inflow and outflow end of the valve-receiving member 250 may be coupled to the atrial sheet 230 and ventricular sheet 240, respectively, by any suitable means including sutures. At least in part because the valve-receiving member 250 is suspended within the anchor frame 201 via atrial sheet 230 and ventricular sheet 240, the valve component eventually received within the valve-receiving member 250 will retain its shape, even if the native tricuspid valve deforms the shape of the anchor frame 201, for example from forces as the heart contracts. In other words, the shape of the valve component within the valve-receiving member 250 is substantially independent of the shape of the anchor frame 201. This may be desirable because the valve component will typically have a circular shape or profile, and it is desirable to maintain that circular shape or profile to help ensure that the prosthetic leaflets of the valve component are able to coapt to prevent regurgitation across the prosthetic leaflets. If forces on the anchor frame 201 were transmitted to the valve component in the valve-receiving member 250, and the valve component was to be deformed, the prosthetic leaflets of the valve component may not be able to coapt correctly.
Although not required, it may be desirable to include radiopaque markers 270 on both the inflow end of the valve-receiving member 250 and the outflow end of the valve-receiving member 250. These radiopaque markers 270 may be formed from any biocompatible substance that is easily visualized during imaging and have any desired configuration. For example, the radiopaque markers 270 may be small pieces of biocompatible metal coupled to the valve-receiving member or radiopaque threads that are sutured into the valve-receiving member 250, although other configurations may be appropriate. These radiopaque markers 270 may readily show, during imaging, where the ends of the valve-receiving member 250 are located so that the valve component may be reliably and desirably positioned within the valve-receiving member 250 prior to expansion of the valve component. And, as described in greater detail below, the radiopaque markers 270 may also readily show the position of the shunt 280, if such a shunt is included. Examples of materials that may be used to form the radiopaque markers 270 may include materials with high atomic mass, such as gold, tungsten, platinum, and iridium as well as combinations or alloys of these materials. In some embodiments, it may be preferable to form the radiopaque markers 270 from a material that will enhance ingrowth into the material in order to more rapidly close shunt 280, a concept which is described in greater detail below.
Although the valve-receiving member 250 may be configured to receive a balloon-expandable or self-expandable prosthetic heart valve in a secondary procedure, in other embodiments the valve-receiving member 250 may include prosthetic leaflets directly coupled to the valve-receiving member 250. For example, bioprosthetic tissue (e.g., bovine or porcine pericardium) or synthetic fabric leaflets may be directly sutured to the interior of the valve-receiving member 250 so that the prosthetic heart valve may be implanted in a single step, instead of a two-step procedure. In embodiments in which the prosthetic leaflets are directly coupled to the valve-receiving member 250, it is preferable that the valve-receiving member 250 is formed of fabric or tissue to reduce the overall collapsed profile of the device, but in some embodiments, the valve-receiving member 250 may take the form of an inner metal stent that is collapsible and expandable. The shunt 280 described below may work equally well whether the prosthetic heart valve is designed as a two-stage or single-stage implant.
To expand on the point above, although the shunts are described herein along with illustrations and descriptions of a two-stage implant, the invention is not limited to two-stage implants. For example, the size of the inner valve of a prosthetic tricuspid valve may be generally between about 20 mm and about 36 mm in diameter (preferably between about 27 mm and about 33 mm), whereas the size of the native tricuspid valve annulus (in a patient with severe or torrential regurgitation) may be on average somewhere around 52 mm (e.g. between a range of about 36 mm to about 70 mm). Some single-stage prosthetic heart valves include two stents—an inner stent to support the prosthetic leaflets, and an outer stent for anchoring. In these embodiments, although the inner stent is coupled to the outer stent (e.g., at an atrial side or ventricular side of the prosthetic heart valve), the size difference noted above results in a gap space being available between the outside of the inner stent and waist of the outer anchoring stent that is designed to contact the native valve annulus. The shunts described herein may be positioned within or along that gap space in substantially the same manner as described in connection with the exemplary two-stage implants shown herein.
To expand on the point above even further, some recently developed prosthetic tricuspid heart valves include a single stent that has a relatively small (e.g., between about 27 mm and about 33 mm) central diameter to directly house the prosthetic leaflets, and relatively large (e.g., 50 mm or more) atrial and ventricular disks. Examples of those recently developed prosthetic tricuspid heart valves are described in greater detail in U.S. Patent Application No. 63/341,702 filed May 13, 2022, the disclosure of which is hereby incorporated by reference herein. Those recently developed prosthetic tricuspid valves include a fabric covering extending between the large atrial and ventricular disks, and it is that fabric covering that directly presses against the native tricuspid valve annulus. In these recently developed prosthetic valves, there is similarly a gap space between the outside of the center portion of the stent that houses the prosthetic leaflets, and the fabric that presses against the native tricuspid valve annulus upon implantation. That gap space may be used either to place a shunt, or to otherwise allow for shunt-like activity, as described in greater detail in connection with
The shunt 280 may be a tube that provides a pathway for blood to flow in an unrestricted fashion (at least initially). The shunt may have a first open end coupled to the atrial sheet 230, and a second open end coupled to the ventricular sheet 240. Preferably, the material forming the tube of the shunt 280 is substantially impermeable to blood, so that blood only flows between the two ends of the shunt 280. For example, when the right atrium contracts, blood may flow through both the shunt 280 and through the open leaflets positioned within valve-receiving member 250. When the right ventricle contracts, the leaflets positioned within the valve-receiving member 250 completely (or substantially completely) block blood from flowing in the retrograde direction through the valve-receiving member 250, but the shunt 280 allows for the unimpeded flow of blood in the retrograde direction through a flow opening 282 of the shunt 280, limited mainly by the size of the lumen of the shunt 280. In some examples, the shunt 280 may have a diameter of between about 2 mm and about 10 mm. In other examples, the shunt 280 may have a diameter of between about 4 mm and about 8 mm. However, it should be understood that these diameters are merely exemplary. Prior to the shunt 280 being closed, whether in a secondary procedure or as the result of the shunt 280 being designed to self-close over time, the shunt 280 may also be used as a lumen through which a pacemaker lead may be placed during a pacemaker implantation procedure. And although the shunt 280 is described as being substantially impermeable to blood, in some embodiments, the shunt 280 may be formed of a material that is permeable to blood. For example, as is described in greater detail below, it may be preferable to form the shunt 280 from a material that promotes ingrowth so that the shunt 280 seals over time. In these embodiments, it may be desirable to form the shunt 280 from a material that is porous and/or permeable to blood as such materials may enhance ingrowth.
In one embodiment, the shunt 280 may be formed as a substantially cylindrical tube of fabric, which may be formed of any of the materials described in connection with atrial sheet 230 or ventricular sheet 240. The atrial sheet 230 and ventricular sheet 240 may include openings where the respective ends of the shunt 280 are coupled to the two sheets, in a similar way as described in relation to the connection of the valve-receiving member 250 to the atrial sheet 230 and ventricular sheet 240. Also, one or both ends of the shunt 280 may include radiopaque markers 270, in a similar or identical fashion as described in connection with valve-receiving member 250. The main body of the shunt 280 may be positioned radially outside of the valve-receiving member 250, and radially inside the central portion 203 of the stent of the anchor frame 201. In other embodiments, described in greater detail, the shunt is instead positioned radially outside of an inner stent, and radially inside either an outer anchoring stent, or radially inside an outer fabric that directly contacts the native valve annulus.
In one exemplary use of anchoring frame 201, it may first be transitioned to a collapsed condition and placed within a sheath of a delivery device, the sheath maintaining the anchoring frame in the collapsed condition. The sheath of the delivery device may have an outer diameter of up to about 38 French (12.67 mm), although it is preferably smaller and has a diameter of between about 30 French (10 mm) and 38 French (12.67 mm), and it is most preferably even smaller with a diameter of between about 20 French (6.67 mm) and about 28 French (about 9.33 mm). After the anchoring frame 201 is within the delivery device, it may be introduced into the femoral vein. If the device is small enough (e.g. 33 French (11 mm) or smaller, preferably 30 French (10 mm) or smaller), it may be introduced into the femoral vein without the need for a surgical cut down of the femoral vein. The delivery device may be advanced through the patient's vasculature, through the inferior vena cava, into the right atrium, and may be oriented (for example via a steering mechanism) so that the distal end of the sheath is within or adjacent to the native tricuspid valve annulus. The distal end of the sheath may be withdrawn relative to the anchoring frame 201, removing the constraint on the anchoring frame 201 and allowing the anchoring frame 201 to begin to self-expand into the native tricuspid valve annulus, with the ventricular portion 204 abutting the ventricular side of the valve annulus, the atrial portion 202 abutting the atrial side of the valve annulus, and the valve annulus received within the waisted central portion 203. When deployed, the anchor frame 201 may provide reliable anchoring via (i) the pinching of the native tricuspid valve annulus by the atrial portion 202 and the ventricular portion 204; (ii) the oversizing of the central waist portion 203 relative to the native tricuspid valve annulus; and/or (iii) anchoring tines or barbs, if included. However, it should be understood that the anchor frame 201 may be designed so that oversizing of the central waist portion 203 relative to the native annulus is not needed. In fact, although the three above-described modalities of anchoring may be provided in a single device, any combination of anchoring mechanisms (i), (ii), and/or (iii) listed above may be sufficient. In some embodiments, the oversizing of the ventricular portion 204 relative to the native valve annulus may help the valve remain in place by providing a mechanism for resisting migration into the right atrium. While the anchor frame 201 is deployed, and before the valve component is deployed into the valve-receiving member 250 (which may be referred to as the first stage of implantation), the valve-receiving member 250 provides an open conduit between the right ventricle and the right atrium. If the anchor frame 201 is designed as a single-stage implant with prosthetic leaflets already attached to the valve-receiving member 250, the initial valve implantation may be complete at this point. If the anchor frame 201 is designed as a two-stage implant, the valve prosthesis may be implanted next as outlined below.
Once the first stage of implantation has been completed and the anchor frame 201 is deployed within the native tricuspid valve annulus, the valve-receiving member 250 creates an open conduit between the right atrium and the right ventricle. Preferably, there is little or no delay between completing the first stage of implantation and beginning the second stage of implantation. In the second stage of implantation (if required), a valve component 300 is deployed into the anchor frame 201.
The prosthetic leaflets 340 may have, in the aggregate, a generally cylindrical profile, with three leaflets total, each leaflet coupled to an adjacent leaflet at a commissure feature of the stent 320. However, more or fewer prosthetic leaflets 340 may be provided as desired. The inner cuff 360 may be formed of biocompatible tissue or synthetic material, such as PTFE, PET, or ultra-high molecular weight polyethylene (UHMWPE). The outer cuff 380 may similarly be formed of biocompatible tissue or synthetic material, such as PTFE, PET, or UHMWPE. In the illustrated configuration, the outer cuff 380 has an inflow edge that is fixed (e.g., via suturing) to the inner cuff 360 and/or the stent 320, with an outflow edge that is coupled to the inner cuff 360 and/or stent 320 at spaced apart circumferential locations, to create one or more openings between the outer cuff 380 and the inner cuff 360 into which blood may flow. If blood flows into these openings, e.g., during retrograde blood flow, the outer cuff 380 may billow outwardly to help ensure there is no regurgitation around the outside of the valve component 300. Additional details that may be relevant for use with valve component 300 are described in greater detail in U.S. Pat. No. 10,548,722, the disclosure of which is hereby incorporated by reference herein. However, because valve component 300 is, in this embodiment, intended to be received within valve-receiving member 250, it should be understood, that the inner cuff 360 and/or outer cuff 380 may be omitted, with the valve-receiving member 250 itself providing cuff/skirt functionality.
The valve component 300 may be collapsed to a small diameter and positioned within the sheath of a delivery device and introduced into the right heart via any suitable means and delivery route. For example, the valve component 300 may be delivered via the femoral vein, similar to anchor frame 201, via a transapical access route through the chest and through the right ventricle, via a transjugular delivery route and through the superior vena cava, or any other desirable delivery route that leads to the tricuspid valve. In some circumstances, particularly if the second stage implantation is being performed immediately after the first stage, the same catheter may even be used for each stage of implantation, although this is not required.
Regardless of the delivery route, the distal end of the delivery catheter housing the collapsed valve component 300 is positioned adjacent to, or within, the valve-receiving member 250 of the anchor frame 201.
If valve component 300 is self-expandable, the catheter sheath may be withdrawn or advanced to uncover the valve component 300, removing the constriction maintaining the valve component 300 in the collapsed condition, thus allowing the valve component 300 to self-expand into the valve-receiving member 250. If the stent 320 of valve component 300 includes an outwardly flared outflow section, the outward flare may protrude beyond the outflow end of the valve-receiving member 250 and provide additional anchoring force to resist migration toward the right atrium. However, in other embodiments, the outwardly flared outflow section may be omitted, with the stent being generally cylindrical.
If valve component 300 is balloon-expandable, the valve component 300 may be crimped over an uninflated balloon when the valve component 300 is mounted to the delivery catheter in the collapsed condition. However, if the valve component 300 is balloon-expandable it may or may not be covered by a catheter sheath during the second stage implant. Rather than withdrawing a sheath to allow the valve component 300 to self-expand, the balloon is inflated (e.g., by pushing saline through the delivery device into the balloon) to force the valve component 300 to expand into the valve-receiving member 250 of the anchor frame 201. Regardless of whether the valve component 300 is self-expanding or balloon-expandable, the radiopaque markers 270 on the inflow and/or outflow end of the valve-receiving member 250 (if included) may be referenced with imaging to confirm that the valve component 300 is in the desired position relative to the valve-receiving member 250 prior to expansion of the valve component 300.
Whether self-expanding or balloon-expandable, it is preferable that the valve component has a diameter (at least at the portion housing the prosthetic leaflets 340) of between about 25 mm and about 35 mm, including between about 28 mm to about 32 mm, including about 29 mm, about 30 mm, and about 31 mm.
After the valve component 300 is deployed within the anchor frame 201, the prosthetic leaflets 340 may take on the function of the previously failing native tricuspid valve leaflets. At this point, any remaining catheters or other accessories within the patient may be removed, and the procedure completed. As noted above, as the patient recovers from the prosthetic heart valve implantation, the intentional regurgitation of blood across the shunt 280 may allow the heart to acclimate to the new hemodynamics without overloading the right ventricle. Thus, over time, the patient's right ventricle may become stronger and, after the patient has recovered from the prosthetic heart valve implantation, the shunt 280 may be closed in a separate procedure. For example, an occluder may be implanted into the shunt 280 to occlude the passageway, eliminating any regurgitation of blood across the shunt 280. The occluder may take any suitable form that is configured to be positioned within and/or through the shunt 280 and to either immediately or over time occlude the shunt 280 so that blood can no longer flow through the shunt 280. In one embodiment, as shown in
As shown in
In the embodiment described above, the shunt 280 is closed as part of a separate, second procedure. In other embodiments, the shunt 280 may be designed so that a separate procedure is not necessary to close the shunt 280, but rather the shunt 280 will naturally close over time by virtue of its design.
One example of a self-closing shunt 480 is illustrated in
Referring to
Although various materials may be suitable for forming the biodegradable or bioabsorbable wires or sutures described above, typically these sutures are made from Polylactide acid, Polyglycolide acid, or copolymers of these materials. These biodegradable sutures may be designed to have any desired range of degradation time, with the degradation time being influenced by factors including the makeup of the material forming the sutures, as well as the molecular weight of the material.
As described above, the various shunts described herein may be closed via a secondary procedure or intervention, or otherwise may be self-closing. Instead of mechanical action to close the shunts, as described in connection with
In all of the embodiments described above, the shunts are pre-assembled to the prosthetic heart valve so that the shunts are functioning immediately upon implantation of the anchor frame, allowing for regurgitation through the shunt as soon as the anchor frame is implanted. However, in some embodiments, it may be desirable to exclude such a shunt. For example, if a patient does not appear to be at risk of right ventricular overload, the inclusion of a shunt may not be desirable. However, those patients still may require a pacemaker to be implanted after the prosthetic heart valve is implanted. In the embodiments with shunts described above, the shunts may act as a convenient delivery pathway that the pacemaker leads may be passed through during implantation, prior to closure of the shunt, while eliminating the risk that the delivery of the leads will damage the prosthetic leaflets and the risk that the lead placement may interfere with proper coaptation of the prosthetic leaflets. However, if no shunt is included, features may be provided to assist with pacemaker implantation. For example,
In another embodiment, the designs of
As noted above, a “docking” station-style anchor frame that is part of a two-stage implantation procedure is described herein to provide context for the shunts and pacemaker lead pathway features of this disclosure. But the invention is not limited to this type of prosthetic heart valve system and is instead equally applicable to single-stage prosthetic heart valve implants, including those with a two-stent configuration or a single-stent configuration.
Still referring to
Although the invention herein has been described with reference to particular embodiments, it is to be understood that these embodiments are merely illustrative of the principles and applications of the present invention. It is therefore to be understood that numerous modifications may be made to the illustrative embodiments and that other arrangements may be devised without departing from the spirit and scope of the present invention as defined by the appended claims.
This application claims priority to the filing date of U.S. Provisional Patent Application No. 63/350,447, filed Jun. 9, 2022, the disclosure of which is hereby incorporated by reference herein.
Number | Date | Country | |
---|---|---|---|
63350447 | Jun 2022 | US |