In medical ultrasound imaging, using higher frequencies provides better resolution. (As used herein, frequency refers to the center frequency of the signal that is used to drive the transducer.) However, since attenuation is directly proportional to frequency, increasing the frequency also cuts the depth of penetration. For example, if using a 4 MHz signal to capture an image provides a depth of penetration is 10 cm, increasing the frequency to 8 MHz will cut the depth of penetration to about 5 cm. In other words, there is a trade off between resolution and penetration: operating at lower frequencies provides more penetration and less resolution; and operating at higher frequencies provides more resolution and less penetration.
One prior art solution is to use pulses at two different frequencies for each line in the image, then combine the return echo from those pulses into a single line of the image. More specifically, in shallower portions, where there is plenty of signal power, the return from the higher frequency pulse is used to provide higher resolution. But beyond a certain point, where the signal to noise ratio (SNR) of the higher frequency return is too low to provide a good image, the return from the lower frequency pulse is used. Using this two-pulse-per-line approach however, requires twice as many pulses to obtain each ultrasound image frame (i.e., two pulses for each line in the frame instead of the more standard single pulse per line). This increases the time it takes to capture each frame of the image, increases the total ultrasound energy that is transmitted into the patient to capture the images, and increases the overall complexity of the system.
When an ultrasound transducer is driven by a signal that contains a relatively wide range of frequencies, the frequency-dependent attenuation characteristics of the subject being imaged can be relied on to simultaneously provide, using only a single pulse per line of the image, (a) a return from the deeper portions of the image that is dominated by lower frequencies and (b) a return from the shallower portions of the image that is dominated by higher frequencies. These returns are processed into an image with higher resolution in the shallower parts, and lower resolution (yet still with an adequate SNR) in the deeper parts.
Since shorter durations in the time domain correspond to wider spectra in the frequency domain, one way to obtain the desired broadband signal is to use a short pulse. For example, if the transducer is driven by a pulse 10 that consists of two consecutive sinusoidal waves, as depicted in
Similar results can be obtained with real (i.e., non-hypothetical) transducers as well, and examination of the curves in
It is notable that the design point for the center frequency of the transducer, 5.05 MHz, is significantly lower than the 6.67 MHz frequency (which corresponds to a 0.15 μS period) of the sinusoidal pulse 10 used to excite the transducer. Selecting a transducer with a center frequency that is significantly lower than the frequency of the driving signal works because the attenuation of the higher frequency components is greater than for the lower frequency components, which shifts the average frequency of the received return down towards the lower frequencies. To take advantage of this shift, the frequency of the driving signal should preferably be at least 10% higher than the center frequency of the transducer, and more preferably at least 20% higher than the center frequency of the transducer.
The curves for the hypothetical transducer in
This combination of transducer characteristics and driving waveform characteristics advantageously provides superior resolution for the shallower portions of the image (because the returns from those depths dominated by are higher frequency components), and maintains penetration depth for deeper portions of the image, albeit with lower resolution (because the returns from those depths are dominated by lower frequency components). Notably, both these benefits are obtained simultaneously from a single transmit pulse, without the added complexity inherent in actually transmitting two pulses at different frequencies, then receiving the returns from those two pulses, and then combining those two returns into a single image. This arrangement also reduces the amount of ultrasound energy that is transmitted into the patient (and the thermal benefits associated therewith) as compared to the prior art approach of using more than one pulse for each scan line. It also avoids image artifacts that can be introduced by motion of the subject that might occur between the high frequency pulse and the low frequency pulse, or by the algorithms that assemble the output image from the two raw ultrasound images.
If desired, the return signals from all depths may be processed using the same signal processing algorithm. Alternatively, processing of different regions of the image may be optimized based on a priori knowledge of the expected center frequency contained in each to the different regions. For example, the image may be divided into a few depth bands (e.g., a first band between 0 and 4 cm and a second band beyond 4 cm), and different signal processing parameters may be used in each of those regions (e.g., a first set of filter coefficients that is optimized for higher frequency signals may be used for the first depth band, and a second set of filter coefficients that is optimized for lower frequency signals may be used for the second depth band). As yet another alternative, the change in processing may be varied continuously as a function of depth instead of in discrete steps (e.g., by selecting filter coefficients for each pixel in the image as a function of that pixel's depth).
In alternative embodiments, waveforms other than the two consecutive sinusoidal waves depicted in
While the above-described embodiments work well with existing ultrasound transducers, it is expected that customizing the bandwidth and rolloff characteristics of the ultrasound transducer to take advantage of the effects desired herein should further improve performance. For example, instead of designing the ultrasound transducer to have symmetrical rolloff characteristics (which is a common design goal for conventional ultrasound transducers), performance can be improved if is the ultrasound transducer has a steeper rolloff above the center frequency than below the center frequency. The transducer should preferably be relatively flat from about 0.65 times the nominal center frequency f to about 1.1 times f, that is over the rage 0.65 f to 1. 1 f, and roll off slowly (e.g., 6-9 dB/MHz) for 1 MHz above and below that range.
Of course, persons skilled in the relevant arts will recognize that the scope of the invention in not limited by the numeric examples provided herein (e.g., for center frequencies, rolloff rates, etc.) and that they may be adjusted based on the desired design goals of the particular system that is being implemented.
This application claims the benefit of U.S. provisional application 60/867,922, filed Nov. 30, 2006, which is incorporated herein by reference.
Number | Date | Country | |
---|---|---|---|
60867922 | Nov 2006 | US |