The field of the disclosure relates generally to radiation therapy and, more specifically, a flash radiotherapy (FLASH-RT) system for preclinical research based on a multipixel x-ray source. The field of disclosure also relates generally to a micro-CT imaging device for preclinical research based on multipixel x-ray source.
Radiation therapy (RT), or radiotherapy, is one of the most widely-used and successful cancer treatment modalities, with 52% of cancer patients receiving at least one course of RT. However, there are also many scenarios in which durable local-regional control is compromised by normal tissue complications, e.g., locally-advanced tumors of the head and neck, lung, cervix, brain, and prostate. The need for improved therapeutic ratio in RT has driven important advances in ultra-high dose rate RT or Flash-RT technology that has been proven the advantage of improved therapeutic ratio. A small animal FLASH-RT system is needed for preclinical of FLASH-RT.
Conventional small animal irradiator systems have provided an ultra-high dose rate using a commercially available imaging x-ray tube. This type of single source x-ray tube is barely useful in real biological study since its fast dose falloff and thus a very small volume of FLASH radiation. It will be hard for many comparative studies which require much larger and deeper FLASH irradiation volume and for the commercial x-ray tubes.
In another design, combining two commercially available and rotatable x-ray tubes with rotating anode facing each other was provided. It required the phantom being very close to the sources. The FLASH dose region had a more uniform dose rate distribution compared with the above-mentioned system.
One example of a commercial system is the Small Animal Radiation Research Platform (Xstrahl Inc., Suwanee, GA) that can produce ˜0.05 Gy/s dose rate at the isocenter located at 35 cm nominal source-to-surface distance (SSD). Bazalova-Carter and Esplen studied feasibility of achieving UHDR in a commercially available x-ray tube with a fixed anode. In a later development of a kV FLASH tube for in vitro irradiations, x-ray beams were controlled by a shutter to deliver FLASH irradiations in pulses. This design is not practical for in vivo studies as UHDR is only achieved within a few mm of the surface due to the small SSD that results in fast dose falloff due to beam divergence. Rezaee et al. proposed a kV FLASH design consisting of two commercial diagnostic x-ray tubes with rotating anodes facing each other. This design requires the phantom to be attached to the surfaces of both tubes. Pulsed photon FLASH beam could be delivered uniformly in a water phantom in 20 mm spacing between two tube surfaces. However, the conformity of the dose to the target is subpar due to limited beam angles. Also, none of the mentioned methods can deliver x-ray beams in either well-defined macro-pulses or continuous current (DC) mode of operation. The lack of small animal FLASH irradiator that can mimic clinical FLASH-RT greatly hinders the biological study of FLASH-RT and clinical translations. What is needed is a system to reach a FLASH dose rate that is also flexible to control an inversely optimized dose distribution.
The present embodiments may relate to, inter alia, systems and methods for a flash radiotherapy (FLASH-RT) device to provide treatment. The present embodiments provide treatment by a FLASH-RT device that spares normal tissue when killing cancer cells. This may be performed by using the FLASH-RT device to deliver radiation at an ultra-high dose rate by irradiating the target simultaneously by all sources. Additionally, or alternatively, the FLASH-RT device may deliver radiation at a conventional dose rate. For example, the FLASH-RT device may provide a conventional dose rate by sequentially scanning the sources or adjusting the beam current.
In one embodiment, a flash radiotherapy (RT) irradiator system is described. The flash RT irradiator system including a multipixel x-ray source, a flash RT device connected to the multipixel x-ray source. The flash RT device is configured to scan an anatomy of an animal, and deliver, using the multipixel x-ray source, radiation at an ultra-high dose rate to a treatment location of the scanned anatomy.
In another embodiment, a method is described for delivering radiation to an animal. The method includes scanning, using a flash RT device, an anatomy of an animal and delivering, using a multipixel x-ray source, radiation at an ultra-high dose rate to a treatment location of the scanned anatomy.
In another embodiment, an inverse-geometry micro-CT system is described including a multipixel circular x-ray source array, a rotatable detector array, and a rotatable multi-aperture collimator configured to direct x-ray beams from the multipixel x-ray source to the rotatable detector array, wherein the multi-aperture collimator is an arc shaped collimator including a plurality of slot openings configured to rotate to scan an animal.
In yet another embodiment, a method of scanning small animal is described. The method includes scanning a plurality of partial focal spots of a circular x-ray source array, collimating, with a rotatable multi-aperture collimator, the x-ray source array to a detector, rotating the collimator and detector to another angle, repeating scanning the plurality of partial focal spots at the other angle, and reconstructing a plurality of 3D C T images using data measured by the detector.
Advantages will become more apparent to those skilled in the art from the following description of the preferred embodiments which have been shown and described by way of illustration. As will be realized, the present embodiments may be capable of other and different embodiments, and their details are capable of modification in various respects. Accordingly, the drawings and description are to be regarded as illustrative in nature and not as restrictive.
The patent or application file contains at least one drawing executed in color. Copies of this patent or patent application publication with color drawing(s) will be provided by the Office upon request and payment of the necessary fee.
The figures described below depict various aspects of the systems and methods disclosed therein. Each figure depicts an embodiment of a particular aspect of the disclosed systems and methods, and that each of the Figures is intended to accord with a possible embodiment thereof. Further, wherever possible, the following description refers to the reference numerals included in the following Figures, in which features depicted in multiple Figures are designated with consistent reference numerals.
There are shown in the drawings arrangements which are presently discussed, it being understood, however, that the present embodiments are not limited to the precise arrangements and are instrumentalities shown.
The figures depict preferred embodiments for purposes of illustration only. One skilled in the art will readily recognize from the following discussion that alternative embodiments of the systems and methods illustrated herein may be employed without departing from the principles described herein.
The present embodiments are related to, inter alia, systems and methods for providing a flash radiotherapy (RT) irradiator device using a multipixel x-ray source. In one exemplary embodiment, a method may be provided that uses a flash radiotherapy (RT) irradiator device that may use a multipixel x-ray source to scan an anatomy of an animal, determine a treatment location of the animal, and deliver an ultra-high dose of radiation to the animal's treatment location. The multipixel x-ray source may include a ring of focal spots capable of irradiating x-ray beams from around fifty beam angles. The x-ray beams may be collimated with a 3D printed collimator of the animal's anatomy.
One discovery in radiation oncology is the normal tissue complication-probability (NTCP) reduction effect of FLASH-RT. Comparing with conventional dose rate (CDR), ultra-high dose rates (UHDR) correspond to dose rates greater than 40 Gy/s and substantially improve normal tissue sparing while maintaining high tumor control probability (TCP). In some embodiments, FLASH-RT has the potential to significantly widen the TCP-NTCP therapeutic window, allowing dose intensification via hypofractionation. Proton is an example FLASH-RT modality, especially for small targets.
In at least one embodiment, a small animal irradiation platform is provided to deliver the same conformal dose distribution with different temporal structures including pulsed and continuous in CDR and UHDR modes. In some embodiments, platforms for preclinical FLASH investigations include but are not limited to (a) UHDR 8-15 MeV electron beams from clinical linear accelerators (linacs) (x-ray beam currents with target and flattening filter removed) and (b) clinical proton beam platforms (with enhanced beam currents and stationary beam modulators) able to produce UHDR spread-out Bragg peaks (SOBP) for both passively scattered and pencil-beam scanning (PBS) proton fields.
In some embodiments, a small animal FLASH irradiator based on distributed x-ray source technology is provided. In this exemplary embodiment, a multi-pixel thermionic emission x-ray (MPTEX) source is provided. An MPTEX source is based on thermionic cathodes. Even with as many as 50 cathodes, the power consumed by thermionic cathodes is only a fraction of the power produced by focal spots. An example application of an MPTEX source is a Tetrahedron Beam CT (TBCT), an imaging technique that can produce superior image quality in a similar footprint to Cone Beam CT (CBCT) imaging.
In another example, small animal FLASH irradiator (SAFI) is based on an MPTEX source. Based on a linear MPTEX source for TBCT imaging, a SAFI system comprises a circular array of x-ray sources. By orienting all sources toward the center of the focal spot circle, conformal UHDR treatments can be realized in application to mouse RT. The same circular source array can also be used for inverse-geometry micro-CT (IGMCT) onboard imaging. As many as 51 x-ray sources can be generated on a circle of 10 cm radius. The FLASH-RT treatment plan can be optimized analogously to clinical Intensity Modulated RT (IMRT) treatment plans.
The SAFI system, in some embodiments, is integrated with an IGMCT. Besides used for treatment planning simulation and setup for small animal irradiation, micro-CT may also be used as an in vivo imaging tool for biomedical research in general. X-ray exposure is a major concern for diagnostic CT. According to Rose's model, the radiation exposure increases cubically with the image resolution. The radiation dose by a micro-CT scan with resolution better than 100 micron is often too high for live animals. In one example, the annular-array x-ray source of the SAFI system is an in-plane fluence modulation supported by IGMCT. In this example, the system only needs to image a small region-of-interest (ROI) in the animal body. Conventional CT and micro-CT have to deliver an almost uniform dose to the whole body in order to avoid data truncation problem. The disclosed IGMCT can modulate in-plane x-ray intensity so that only the beams passing through the ROI have higher flux. IGMCT with x-ray fluence modulation will minimize radiation exposure to peripheral organs and tissues and achieve low-dose micro-CT imaging. Further, the example IGMCT overcomes the fundamental dose barrier in micro-CT high-resolution imaging. Further, IGMCT can be used as an important in vivo imaging modality for biology research beyond radiation oncology.
In reference to
In some embodiments, the SAFI is integrated with inverse-geometry micro-CT based on circular x-ray source array. The SAFI includes a donut-shaped copper anode with 10 degree anode angle and a circular array of cathodes. In this example, the cathodes may generate 51 focal spots on the tungsten targets that are brazed on the copper anode. X-ray beams exit the inner wall of vacuum housing, then are collimated by a full-ring multi-aperture collimator (MAC) designed in treatment planning. A partial-ring MAC and a small area detector will replace the full-ring treatment MAC and rotate about the animal during imaging. All sources are activated simultaneously for FLASH-RT treatments or sequentially for IGMCT imaging. The SAFI achieves UHDR by delivering x-ray beams from 51 angles. Further, IGMCT is realized by rotating a compact small area while activating the sources in FOV sequentially.
In some embodiments, a design factor may include a distance between the center of a single source and the geometric center dsc. In this example design, 100 mm is a quite minimal choice for the irradiator. It may be necessary for a proper-size slot collimator (40 mm) or imaging detector (50 mm), room for tube housing (30 mm) and fit in with a small animal (20 mm). The x-ray fluence may be proportional to the inverse square of the distance from the source.
Additionally, or alternatively, the x-ray fluence at the center of the irradiator may be proportional to the inverse square of the ring source radius, but the number of sources may be proportional to the radius. There may be a tradeoff between stronger x-ray intensity and a larger irradiating volume. For FLASH purpose, the more compact the sources, the higher dose rate may be achieved. From this aspect, if collimation is not considered, there is a minimum number of sources if the distance between the sources to center is minimal. For example, 30 mm tube house and 20 mm small animal radius may allow more than 60 sources with 4 mm spacing.
During a FLASH-RT treatment, all beams are to be turned on simultaneously with flux intensities optimized in treatment planning. A full ring MAC will be custom designed based on the actual geometry of the target and fabricated by 3D printing techniques. The MAC will conform the beams 300 to the target while avoiding unnecessary leakage x-rays 302 produced by adjacent sources as shown in
Inverse-geometry micro-CT (IGMCT): Micro-CT is broadly used in preclinical research. Small animal irradiators need onboard imaging for treatment planning and animal positioning. As shown in
Additionally, or alternatively, the distributed x-ray source allows in-plane modulation of beam intensity. Traditionally, a physical bowtie filter is used to reduce the intensity of x-ray away from the central axis. But the position of the ROI varies with projection angles. IGMCT allows dynamically modulating the in-plane x-ray intensities based on the ROI positions. As shown in
In reference to
In view of
Using Equation (1), the dose rate per mA current along the vertical axis of a 40-mm water phantom may be estimated.
The exemplary dose rate per mA current vs depth from the surface of cylindrical water phantom is plotted in
The required cathode current IFLASH may then be calculated by Equation (2).
Required single cathode current IFLASH for N sources to obtain the intended dose rate DFLASH=100 Gy/s at the center of phantom is simply 2985/N mA. In the exemplary design, FLASH dose-rate requirement DFLASH may be 100 Gy/s at the center of phantom, and there may be 51 sources. The required cathode current for each source IFLASH may be, for example, 58.5 mA.
In some embodiments, example radiation beam properties from an ideal point source are plotted in
In some embodiments, IGMCT is used to realize fluence-modulated scanning with the SAFI system. The system may, for example, acquire a low-resolution image with static source positions and object and high-resolution (˜100 micron) images by rotating either the source or object in 4-6 steps in addition to detector rotation. The image quality and resolution may then be evaluated.
In one embodiment, inverse-geometry micro-CT imaging may be provided. The imaging MAC can be machined using electrical discharge machining (EDM) or 3D printing. Brass may be sufficient for collimating 60 kVp x-rays. Imaging dose is the fundamental limitation of micro-CT image resolution. With each traditional in-plane fan-beam divided into multiple beam segments, the beam intensity can be modulated so that the beams intercepting the ROI will have the highest flux, thus greatly reducing the imaging dose to the volume outside the ROI. A low-resolution image may be scanned with a fixed source position first, then ROI is designed to have an image with higher quality. There are a few approaches to modulate beam intensity on the SAFI system including: 1) changing filament temperature; 2) applying gate voltage; or 3) modulating pulse width. Among them, only the pulse width can be modulated rapidly. The beam control FPGA program may be modified to optimize the x-ray flux and program the beam scan based on the optimized beam sequence. When field size is narrowly collimated, longer exposure time may be necessary as geometric efficiency is reduced. Treatment planning solutions can be used to calculate the imaging dose, which can be incorporated into the RT prescription. The image quality may be evaluated with image phantoms and dose may be measured with film.
A treatment planning solution may be needed for FLASH-RT and CDR-RT treatment of rodents with the SAFI. The SAFI is flexible in optimization parameters: 1) The beam weight can be modulated by controlling the cathode current or pulse width; 2) The aperture shapes can also be optimized to achieve optimal dose conformity; 3) In-field beam intensity can be modulated by physical compensators manufactured by 3D printing; and 4) The beam kVp can be modulated for optimal depth dose distribution. Dose calculation and optimization algorithms can be used, including the use of scripting functionality. In some embodiments, an optimization algorithm may include a simple inverse optimization algorithm based on conjugate gradient algorithm. Although the SAFI system is capable of modulating the beam fluence with a compensator, a simple beam weight optimization algorithm may be sufficient. The target and OARs may be contoured and 51 beams (can be any existing photon or proton beams) may be placed by a beam template and the blocks for the beams may be generated with user designed block margin. Then the block shape may be exported for 3D printing and dose calculation. The dose distribution of each beam may be calculated in micro-CT images, and then the beam weight may be optimized based on user defined objectives. The solution including beam weight and total dose may be evaluated.
One technical improvement of the disclosed embodiments includes a small animal FLASH-RT irradiator for emulating clinically-realistic CDR-RT and FLASH-RT treatments. Radiobiologists and radiation oncologists may use the reliable, low-cost, and compact small animal irradiator for pre-clinical study of FLASH effects. The disclosed SAFI system is based on circular distributed x-ray source technology and provides a small animal irradiator that can achieve intensity-modulated and conformal UHDR treatments to animal subjects. Further, it is a SARR irradiator based on distributed x-ray source technology.
Another technical improvement of the disclosed embodiments includes a dose efficient inverse-geometry micro-CT (IGMCT). Conventional CT and CBCT employ a physical bowtie filter to reduce imaging dose and avoid saturation of detectors. The imaging subject receives a relatively uniform radiation dose in the whole body within FOV. The modulation by bowtie filter is fixed and cannot be change based on animal's body profile and projection angle. The inverse-geometry CT allows “virtual bowtie” that can dynamically modulate the in-plane x-ray flux. Reduction of imaging dose for micro-CT is highly desired due to the ultra-fine image resolution. In CT scan, a high image resolution and contrast within an ROI. IGMCT supports dynamic x-ray fluence modulation based on the ROI positions so that only the x-ray beams intercepting the ROI have higher x-ray flux. Dynamic in-plane intensity modulation based on ROI has not been exploited before on both clinical diagnostic CT and pre-clinical micro-CT. It may greatly reduce unnecessary radiation exposure of CT imaging. SAFI provides an implementation in small animal imaging that overcomes the fundamental dose limitation on image resolution.
Definitions and methods described herein are provided to better define the present disclosure and to guide those of ordinary skill in the art in the practice of the present disclosure. Unless otherwise noted, terms are to be understood according to conventional usage by those of ordinary skill in the relevant art.
In some embodiments, the terms “a” and “an” and “the” and similar references used in the context of describing a particular embodiment (especially in the context of certain of the following claims) are construed to cover both the singular and the plural, unless specifically noted otherwise. In some embodiments, the term “or” as used herein, including the claims, is used to mean “and/or” unless explicitly indicated to refer to alternatives only or to refer to the alternatives that are mutually exclusive.
The terms “comprise,” “have” and “include” are open-ended linking verbs. Any forms or tenses of one or more of these verbs, such as “comprises,” “comprising,” “has,” “having,” “includes” and “including,” are also open-ended. For example, any method that “comprises,” “has” or “includes” one or more steps is not limited to possessing only those one or more steps and may also cover other unlisted steps. Similarly, any composition or device that “comprises,” “has” or “includes” one or more features is not limited to possessing only those one or more features and may cover other unlisted features.
All methods described herein are performed in any suitable order unless otherwise indicated herein or otherwise clearly contradicted by context. The use of any and all examples, or exemplary language (e.g. “such as”) provided with respect to certain embodiments herein is intended merely to better illuminate the present disclosure and does not pose a limitation on the scope of the present disclosure otherwise claimed. No language in the specification should be construed as indicating any non-claimed element essential to the practice of the present disclosure.
Groupings of alternative elements or embodiments of the present disclosure disclosed herein are not to be construed as limitations. Each group member is referred to and claimed individually or in any combination with other members of the group or other elements found herein. One or more members of a group are included in, or deleted from, a group for reasons of convenience or patentability. When any such inclusion or deletion occurs, the specification is herein deemed to contain the group as modified thus fulfilling the written description of all Markush groups used in the appended claims.
To facilitate the understanding of the embodiments described herein, a number of terms are defined below. The terms defined herein have meanings as commonly understood by a person of ordinary skill in the areas relevant to the present disclosure. Terms such as “a,” “an,” and “the” are not intended to refer to only a singular entity, but rather include the general class of which a specific example may be used for illustration. The terminology herein is used to describe specific embodiments of the disclosure, but their usage does not delimit the disclosure, except as outlined in the claims.
All of the compositions and/or methods disclosed and claimed herein may be made and/or executed without undue experimentation in light of the present disclosure. While the compositions and methods of this disclosure have been described in terms of the embodiments included herein, it will be apparent to those of ordinary skill in the art that variations may be applied to the compositions and/or methods and in the steps or in the sequence of steps of the method described herein without departing from the concept, spirit, and scope of the disclosure. All such similar substitutes and modifications apparent to those skilled in the art are deemed to be within the spirit, scope, and concept of the disclosure as defined by the appended claims.
This written description uses examples to disclose the disclosure, including the best mode, and also to enable any person skilled in the art to practice the disclosure, including making and using any devices or systems and performing any incorporated methods. The patentable scope of the disclosure is defined by the claims, and may include other examples that occur to those skilled in the art. Such other examples are intended to be within the scope of the claims if they have structural elements that do not differ from the literal language of the claims, or if they include equivalent structural elements with insubstantial differences from the literal language of the claims.
This application claims priority to U.S. Provisional Patent Application Ser. No. 63/333,776, filed Apr. 22, 2022, the entire disclosure of which is hereby incorporated by reference in its entirety.
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/US2023/019367 | 4/21/2023 | WO |
Number | Date | Country | |
---|---|---|---|
63333776 | Apr 2022 | US |