The drawings described herein are for illustration purposes only and are not intended to limit the scope of the present disclosure in any way.
Electromagnetic radiation offers many advantages over chemicals or heat as a decontaminant. Radiation is generally much less disturbing to the object being sterilized than either reactive oxidizers like chlorine or high temperatures. In addition, radiation can be applied with less labor and hence involve less risk. Unfortunately, germicidal ultraviolet radiation will not penetrate many common materials such as paper, plastics, fibers or metals. In contrast, high energy gamma rays will penetrate many objects, but require very large doses due to the small probability of interaction with the biological pathogens of interest, thereby further requiring massive shielding for safe use. X-ray radiation has been found to be a suitable decontaminant, is penetrating, and can be controlled simply and safely.
Next, the x-ray spectrum is tailored at 14 to match the absorption characteristics of the contaminated environment. In particular, the photon energy for the x-radiation is selected based on the x-ray transmission of the absorption materials identified in the contaminated environment. In addition, the dose of radiation needed to kill a biological hazard found in the environment is determined. For instance, the amount of electrical power and the duration of the radiation for a given area are computed as further described below. X-ray radiation is then applied to the contaminated environment. More specifically, the x-ray radiation having a diffused radiation angle is generated 16 and then directed towards 18 the contaminated area.
The practicality of this concept was demonstrated with a feasibility experiment. Samples of 106 spores of Bacillus subtilis, which is a non-hazardous surrogate for Bacillus anthracis, were first placed in a test environment and exposed to a dose of x-ray radiation from a copper anode source having photon energies primarily around 8 keV. Irradiated and control samples were then individually incubated in soy broth at 35° C. for a week. Samples with one or more viable spores produce a cloudy infusion, while a completely sterilized sample remains clear. Our results showed that at delivered doses of over 4.5 J/cm2, all samples were completely sterilized. The highest dose delivered to a sample that remained incompletely sterilized was 330 mJ/cm2. Hence the 8 keV x-ray kill dose for 106 spores of our surrogate fell somewhere between those two values.
Design of the x-ray source for decontamination applications is qualitatively different than for conventional x-ray tubes used for imaging. Importantly, the x-ray emitting area needs to be large so that sharp shadows in the illuminated volume are avoided. If sharp, high contrast shadows occur, microscopic pathogens could escape from the irradiation and circumvent the desired sterilization. Since x-rays are emitted from the outermost few microns of anode material which receives electron bombardment, the electron beam must be tailored to impinge over the full surface of the anode to achieve the largest effective source size. To this end, the electric field guiding the electrons must be crafted to diverge from the cathode and intersect the anode uniformly, to the greatest extent possible. This technique of tailoring the electric field distribution in the x-ray source is referred to herein as “field sculpting”.
Traditional x-ray sources used for imaging applications are designed as point-source emitters as shown in
To make a diffuse x-ray lamp, it is necessary for a large area of the anode surface to emit x-rays. This requires the electron current to be spread wide, avoiding focusing effects. A modified x-ray source design is shown in
Additionally, this x-ray source may be configured to irradiate over a very wide angle by positioning the output window as close as possible to the anode. X-rays are generated in the first few micrometers of the anode surface that is bombarded with electron current. Any location in the irradiated zone in a clear line of sight to the active anode surface will receive x-rays. The design and location of the output window can be configured to transmit close to a full 2π steradians of irradiated solid angle.
Furthermore, the radiation should thoroughly penetrate the materials covering, surrounding or otherwise obstructing the biological hazard. The x-ray radiation should not pass through the contaminated materials having failed to interact with the biological hazard. High energy x-ray photons will penetrate denser materials, but the resultant scattering cross-section of the photon is reduced. Therefore, a larger flux of x-ray photons is required, leading to longer exposure times to achieve a sufficient kill dose. This is the reason it is advantageous to choose the x-ray photon energy consistent with the materials needing to be decontaminated
X-ray radiation of different photon energies penetrates a contaminated environment to different depths. The x-ray transmission, T, of the absorption material in the contaminated environment is given by
T=e−94 nL,
Where σ is the absorption material's atomic cross section, n is the number density (atoms per cubic centimeter), and L is the path length that the x-rays follow through the absorption material. The material's atomic cross section is a function of the photon energy. Above the K-shell binding energy, the cross section varies as the inverse square of the photon energy. This strong relationship results in a wide range of transmission T versus energy. We choose to use an energy at which T=e−1.
The photon energies produced by an x-ray source can be scaled through the judicious choice of the anode materials. This is understood through Moseley's empirical formula for k-alpha x-rays. The formula shows the x-ray photon energy is dependent on the square of the atomic number of an element
EKα(Z−1)2
Where EK is the x-ray photon energy and Z is the atomic number of the anode material. For instance, an x-ray source having a molybdenum (Z=42) anode will generate radiation having a photon energy of 18 keV. In comparison, a silver (Z=47) anode can generate radiation having a photon energy of 22 keV. It is envisioned that x-ray sources will be fabricated with different anode materials to ensure penetration through various material compositions providing decontamination radiation inside an object in addition to surface decontamination.
The above description is merely exemplary in nature and is not intended to limit the present disclosure, application, or uses. The decontamination and sterilization technique described above can be transitioned to a number of different applications, including but not limited to portable biological pathogen decontamination for civilian and military uses, remotely-deployable payload for delivery by unmanned platforms, mail sorting sterilization, baggage sterilization, battlefield medical sterilizer and mold spore eradication.