The present invention relates to spirometers and more particularly to spirometers which are coupled to manual resuscitators. The invention also relates to spirometers which optionally include carbon dioxide (CO2) sensors and/or carbon monoxide (CO) sensors.
Millions of adult manual resuscitators are used each year in emergency situations to temporarily augment ventilation, both inside and outside of hospitals. These devices are sometimes used by personnel with limited training, as has been pointed out in the introduction to ISO8382; the International Standard that specifies the minimum performance and safety requirements for the type of resuscitator with which our spirometer is intended to be used.
Because of the limited training of some of the clinicians which may use these resuscitators, and the importance of their providing correct ventilatory parameters (tidal and minute volumes) in critical situations, we believe the use of the spirometer to be of great importance as it will give the clinician a good indication of the amount of gas the patient is actually receiving by measuring the amount of gas the patient is exhaling. However, the value of the spirometer extends beyond this and can provide valuable information to even a well-trained clinician. For example, the clinician may assume that the patient is receiving the amount of gas determined by the frequency and amount of compression applied to the bag. Under ideal conditions this is true, but particularly when the patient's lungs are non-compliant or a mucous plug is present, leakage can occur. Potential sites of leakage are around the face mask or endotracheal tube cuff, or through a pressure vent valve, at tubing connections, through a malfunctioning air inlet valve, and elsewhere. In reality, the patient may be receiving little or no ventilation. The instructions typically supplied by the manufacturers of the manual resuscitators, for the clinician to observe the rise and fall of the chest and listen for exhalation to assure proper ventilization, is often not practical, especially at prehospital emergency sites. Factors such as high background noise and difficulty in seeing chest wall movement due to rigid chest walls, obesity and bulky clothing can render these instructions meaningless. This has been frequently pointed out in the literature.
Furthermore, ISO 8382 recommends that a functional test of the resuscitator be carried out immediately prior to use. This is indeed a good idea but often overlooked when a patient needs immediate intervention to initiate respiration.
Spirometers are devices used to measure the volume and flow rate of gas exhaled and/or inhaled by a user or patient. The information displayed by the spirometer addresses the problems outlined above by giving the clinician the needed feedback as to how well the resuscitator is functioning and whether or not the desired ventilation parameters are being achieved.
The present invention provides a spirometer having a housing and an exhale air tube. The exhale air tube includes a first section, a contiguous second section and a third section. The first section has a converging channel in which the exhaled air/gas velocity is increased thereby diminishing static pressure. The second section has a laminar air flow pressure drop.
A pressure transducer connected to measure the pressure drop across the first and second sections provides an indication of the pressure drop across the two sections. The third section of the exhale air tube has a gradually diverging flow diameter to act as a pressure recovery diffuser. A feature of the invention is that the air tube has an adaptor coupling member for easily coupling the spirometer to the manual resuscitator. The gas inlet of the adapter has complementary conical dimensions to the conical expiratory port of the manual resuscitator.
Advantageously, optional CO2 and/or CO sensors may be included in the instrument. From the pressure transducer voltage output, the instrument derives and displays the expiratory flow, expiratory tidal volume, expiratory minute volumes and ventilator frequency. In addition, the exhaled CO2 and/or CO levels are displayed giving additional valuable information to the clinician, as will be discussed in more detail below.
The above and other objects, advantages and features of the invention will become more apparent when considered with the accompanying drawings wherein:
Referring to
Referring to
Expiratory Flow—The flow is displayed graphically, giving the clinician a visual representation of expiratory flow. This can help the clinician avoid initiating the next inspiratory cycle before the patient's lungs empty.
Expiratory Tidal Volume—Each expired breath is displayed in milliliters. The clinician, typically, keys the tidal volume to the patient's size.
Expiratory Minute Volume—This is a predicted volume based on the last several breaths. It is an important parameter as it tells the clinician how much ventilatory gas the patient is receiving over time. The clinician typically will also attempt to key this to the patient's size.
Ventilatory Frequency—The ventilatory frequency is also a calculated value based on the last several breaths. It, along with the average tidal volume, determines the minute volume.
The availability of these data permit the clinician to vary the different ventilation parameters to meet specific patient needs. For example, should the clinician wish to provide a high ventilatory frequency, the tidal volume can be adjusted to obtain the desired minute volume and avoid hyperventilating the patient; and, by observing the expiratory flow, avoid over-inflating the patient's lungs.
Bar graph BG is a visual representation of the expiratory flow, the number of bars energized being a quantitative indication of the instantaneous expiratory flow rate.
The ETV (Expiratory Tidal Volume) display turns on as soon as the EXP FLOW (Expiratory Flow) display goes to zero, following an exhalation. It turns off as soon as the next exhalation flow commences. The volume is presented in milliliters, rounded off to the nearest 10 milliliters.
The EXP MIN VOL (Expiratory Minute Volume) and RESP/MIN (Respirations Per Minute) displays alternate every 2½ seconds with updated information.
The means of measuring gas flow is to measure the pressure change (ΔP) across a flow restrictor. Flow restrictors can be designed to impose a pressure drop which is proportional to the density of the gas or to the gas viscosity, or to a combination of the two. The preferred embodiment of the present invention uses a combination of the two.
Preferably, the laminar flow restrictor, which imposes a pressure drop proportional to the viscosity, consists of, in effect, a bundle of capillary tubes CT. In order to design a laminar flow restrictor with workable dimensions, it is desirable to substantially reduce the flow diameter of the restrictor from the flow diameter of the expiratory port of the resuscitator. As the flow diameter is decreased (between A and B of
As the ΔP between the high pressure sensing port P1 (point A) and the entrance to the laminar flow restrictor (point B) is proportional to the flow volume squared, multiplied by the gas density, and the pressure drop between the entrance of the laminar restrictor (point B) and the low pressure sensing port P2 (point C) is linearly proportional to the flow volume multiplied by the viscosity, we get an overall ΔP which is proportional to the flow volume to something in the order of the 1.5 power of the flow, and a function of both the gas density and viscosity.
The resulting pressure measurements, based on both the density and viscosity have several desirable characteristics. For example, the flow sensor will normally be calibrated with air. However, when used to measure the expiratory flow from a patient, it will normally contain some CO2; perhaps as high as seven percent. The CO2 causes the density of the gas to increase and the viscosity to decrease. Therefore, by combining the two, we essentially negate the error in flow measurement that would otherwise be introduced by the presence of the CO2.
The diffuser section DS (between C and D) is designed to recover a substantial portion of the static pressure drop between. A and B. Therefore, the ΔP available for measuring the expiratory gas flow is actually greater than the overall pressure drop caused by the spirometer and gives an adequate signal for reliable flow measurements, even at low flows. In order to minimize the degrading effect of moisture on the performance of the laminar flow restrictor II, filter 13 may be provided as illustrated in
In summary, the ΔP measured by the pressure transducer is a function of the expiratory gas flow rate. The voltage produced by the pressure transducer versus time is used to calculate all of the displayed expiratory parameters.
Referring now to the block diagram of
In addition to sensing the expiratory gas volume parameters, as already described, one embodiment of the spirometer will contain carbon dioxide (CO2) sensors, as shown in
A further embodiment of the spirometer will contain a carbon monoxide (CO) sensor, located where the CO2 sensor is shown in
The dimensions for the standard tapered 30 mm conical connectors are specified in ISO535601 and their use with the manual resuscitator, with which the spirometer is intended to be used, is specified in ISO 8382.
If a CO2 or CO sensor is incorporated in the spirometer, the displays will be modified.
While the invention has been described in relation to preferred embodiments of the invention, it will be appreciated that other embodiments, adaptations and modifications of the invention will be apparent to those skilled in the art.
Number | Name | Date | Kind |
---|---|---|---|
3081766 | Felkel et al. | Mar 1963 | A |
4440177 | Anderson et al. | Apr 1984 | A |
4774941 | Cook | Oct 1988 | A |
4884460 | Nowacki et al. | Dec 1989 | A |
4914720 | Knodle et al. | Apr 1990 | A |
5038621 | Stupecky | Aug 1991 | A |
5137026 | Waterson et al. | Aug 1992 | A |
5518002 | Wolf et al. | May 1996 | A |
5592934 | Thwaites | Jan 1997 | A |
5743270 | Gazzara et al. | Apr 1998 | A |
5789660 | Kofoed et al. | Aug 1998 | A |
5997483 | Johnson | Dec 1999 | A |
6090049 | Cha | Jul 2000 | A |
6128963 | Brömster | Oct 2000 | A |
6159147 | Lichter et al. | Dec 2000 | A |
6322519 | Moulin | Nov 2001 | B1 |
6435183 | Farman | Aug 2002 | B1 |
6502573 | Ratner | Jan 2003 | B1 |
6543449 | Woodring et al. | Apr 2003 | B1 |