Fiber testing, such as for length and strength, is typically performed in a temperature and humidity controlled environment. Internationally used standards such as the American Society for Testing and Materials (ASTM) Standard Number D-1776 prescribe standard laboratory conditions for testing textile materials such as fibers at 21° Celsius +/−1° and 65% relative humidity +/−2%.
One reason for controlling the temperature and humidity of the fibers during testing in this manner is that the moisture content of such fibers tends to affect characteristics of the fibers, such as length and strength. For example, fibers with higher moisture content tend to exhibit less crimping, and thus fiber length tests tend to report these fibers as having a greater length. Further, possibly due to increased hydrogen bonding between adjacent water molecules in the space between cellulose sheaths and other effects, strength tests on fibers with higher moisture content tend to report such fibers as having a greater strength.
Thus, all fiber testing is most preferably performed at any standard laboratory conditions. In this manner, tests that are performed in different geographical locations and at different times, and which might otherwise have different temperature and humidity conditions in the laboratory, can be reliably compared one to another. When tested under the standardized ASTM conditions given above, it has been assumed that all cotton fibers will equilibrate to a given moisture content of 8.0%. (Re-arranged Order)
However, most fiber testing is not performed under any standardized laboratory conditions. In such nonstandard laboratory conditions, the moisture in the cotton fiber is in equilibrium with the moisture in the air. As a result, the measurements not only may be at different levels in different laboratories but also vary throughout the day as the conditions in an individual laboratory change.
Previous attempts at moisture corrections to the length and strength data have focused on the use of either external moisture measurements or measurements of temperature and relative humidity. The correction is then performed using a correlation between either the directly measured or estimated moisture content and measurements made at standard laboratory conditions. In this manner, measurements taken in nonstandard laboratory conditions can be compared to measurements taken in standard ASTM laboratory conditions at an assumed moisture content of 8.0%.
Unfortunately, there does not appear to be a good correlation between such corrected nonstandard laboratory condition measurements and any standard laboratory condition measurements. What is needed, therefore, is a method that provides a better correlation between the results obtained for fibers tested at standard laboratory conditions, and the corrected results obtained for fibers tested at nonstandard laboratory conditions.
The above and other needs are met by a method for standardizing a measurement of a fiber sample, including the steps of measuring a moisture content of the fiber sample actually being measured during the time of measurement, measuring the fiber sample, and correcting the measurement to a standardized moisture measurement that adjusts for a difference between the measurement at the measured moisture content of the fiber sample and a standardized moisture measurement. An appropriate standardized moisture content is about 7.5% moisture content for ASTM laboratory conditions. If one's selected standard laboratory conditions are different than ASTM conditions, then one can select a different standardized moisture content. Preferably, the measurement includes at least one of fiber length and fiber strength.
Different cotton samples equilibrate to different moisture contents, depending at least in part upon a number of different factors, as described in more detail hereafter. Measurements at different combinations of temperature and relative humidity on a sample set of approximately forty cottons were used to develop the algorithm which corrects for the difference is moisture content between the measured moisture content and the standardized moisture content. These cottons were chosen to span the range of fiber properties from growth areas throughout the world. It has been determined that the analysis to determine the correction must include the actual equilibrium moisture at standard conditions for each sample in this sample set used to determine the correction rather than the assumed 8.0% moisture content. Further, it has been determined that correcting the measurements that are taken at different moisture contents to standardized measurements based on about 7.5% moisture content provides a more accurate overall estimate of the standardized measurements compared to correcting the measurements to values that are based on standard ASTM laboratory conditions at the assumed 8.0% moisture content.
The correction is performed using a measurement of the moisture content of the sample during the time of measurement. In the preferred embodiment, the moisture of the small specimen fiber sample being tested is measured during the time of the measurement. Since the moisture content of the bulk fiber sample, of which small specimen fiber samples are taken, exhibits a distribution affecting the individual measurements, this will reduce the variations in the individual measurements of the different small specimen fiber samples. It has been found that the sample preparation process may significantly change the moisture content of the sample being measured. For this reason, it is preferred that the moisture measurement be of the moisture content of the sample during the time of measurement. This is preferably accomplished by measuring the moisture content of the small sample being measured or alternatively by proper design of the measuring instrument. The measurements are then adjusted using a correlation between the directly measured moisture content and the standard moisture content as described above in regard to testing in standard laboratory conditions. In this manner, measurements taken in nonstandard laboratory conditions can be compared to measurements taken in standard laboratory conditions.
The step of correcting the measurement is alternately accomplished by at least one of applying an algorithm that correlates measurements at different moisture contents, manipulating the measurement in a mathematical equation that correlates measurements at different moisture contents, and using a chart that correlates measurements at different moisture contents.
In some embodiments, the step of correcting the measurement includes correcting the measurement to a moisture content value other than about 7.5%, where the moisture content value is determined based at least in part on fiber characteristics such as at least one of geographical growth location including country and region, growth conditions including rainfall, sunlight, time of year, growth year, harvesting and ginning methods, fiber color, fiber type, and fiber trash content.
Further advantages of the invention are apparent by reference to the detailed description when considered in conjunction with the figures, which are not to scale so as to more clearly show the details, wherein like reference numbers indicate like elements throughout the several views, and wherein:
Cotton samples gain or lose moisture in response to the moisture concentration in the ambient atmosphere. Relative humidity is defined as the percentage of moisture per liter of air compared to the maximum moisture per liter of air that will not produce condensation at that temperature. Relative humidity tends to be a relatively non-linear function within the range of interest as described herein.
The actual moisture content in the air, such as measured in grams per liter, is determined by multiplying the relative humidity times the maximum value as determined by the temperature. Internationally used standards such as the American Society for Testing and Materials (ASTM) Standard Number D-1776 specify standard laboratory conditions at 21° Celsius +/−1° and 65% relative humidity +/−2%, in order to fix the amount of moisture content in the air during both conditioning and testing of the cotton fibers for characteristics such as length and strength. In order to allow the cotton sample time to acclimate to the laboratory conditions, the sample is required to remain in the laboratory for twenty-four hours before being tested.
In
As mentioned above, the sample moisture content is important in fiber measurement because the physical properties of the fiber change due to the absorbed moisture. Without being bound by theory, it is believed that when the moisture penetrates the fiber, weak hydrogen bonds are formed between adjacent fiber sheaths. This results in increased fiber strength. The natural fiber crimp is also reduced, resulting in increased measured fiber length.
Based on these concepts, prior art methods previously corrected all measurements to measurements made at standard ASTM laboratory conditions with an assumed moisture content of 8.0%. However, different cotton samples equilibrate at different moisture contents under the same laboratory conditions. The histograms depicted in
Using data such as that described above, regression curves can be constructed for each cotton sample and the equilibrium moisture content can be calculated for any standard laboratory conditions. When this is done, the change in fiber measurements can be related to differences between the measured moisture content and the actual equilibrium moisture content for that sample rather than the assumed 8.0% moisture content. The importance of this discrepancy arises when attempting to calculate a group behavior for all samples. Unless a proper group behavior is analyzed, the resulting algorithm will not be robust, resulting in poor correlations between the corrected measurements and the actual measurements measured at standard ASTM laboratory conditions.
However, measurements can be made to correlate to measurements at any standard laboratory conditions with a greater degree of precision if additional characteristics of the cotton fiber sample are accounted for. As mentioned above, it has been determined that cotton samples tend to equilibrate to different moisture contents, even though they are held at the same laboratory conditions in terms of temperature and relative humidity. This indicates that the moisture content of a cotton fiber sample is dependent upon more variables than just temperature and relative humidity. It has been determined that the moisture content of a cotton fiber sample is additionally based on at least one of a variety of other fiber characteristics, including geographical growth location including country and region, growth conditions including rainfall, sunlight, time of year, growth year, harvesting and ginning methods, fiber color, fiber type, and fiber trash content.
This information can be used to more accurately standardize and correct the measurements made on fiber samples. For example, a fiber sample having known characteristics as mentioned above can be acclimated at the standard laboratory temperature and relative humidity. Then the moisture content for the fiber sample can be directly measured during the fiber measurements cycle. By directly measured it is meant that the moisture content is measured by a method or device that does not rely on a correlation to the temperature and relative humidity present in the laboratory. For example, such a method would include a resistance measurement. This process avoids errors due to sample moisture content distributions and changes in sample moisture content due to sample preparation processes.
Once the actual moisture content for the fiber sample is known, measurements taken at nonstandard laboratory conditions are then corrected to values that correlate to the actual moisture content as directly measured, rather than to some assumed moisture content value. By constructing charts in this manner of actual moisture contents based upon the varying characteristics as described above, a more accurate measurement data correction can be constructed. According to the more accurate measurement data correction, more than just the moisture content of the fiber sample is used to correct the measurements. Instead, the other characteristics as mentioned above are additionally used to determine the moisture content value to which the measurements should be corrected. In this manner, measurements taken on samples at standard laboratory conditions will compare more accurately with corrected measurements taken on samples at nonstandard laboratory conditions.
The foregoing description of preferred embodiments for this invention has been presented for purposes of illustration and description. It is not intended to be exhaustive or to limit the invention to the precise form disclosed. Obvious modifications or variations are possible in light of the above teachings. The embodiments are chosen and described in an effort to provide the best illustrations of the principles of the invention and its practical application, and to thereby enable one of ordinary skill in the art to utilize the invention in various embodiments and with various modifications as are suited to the particular use contemplated. All such modifications and variations are within the scope of the invention as determined by the appended claims when interpreted in accordance with the breadth to which they are fairly, legally, and equitably entitled.
This application claims priority on U.S. provisional application Ser. No. 60/727,013 filed Oct. 14, 2005. This invention relates to the field of fiber processing and testing. More particularly, this invention relates to correcting to standardized laboratory conditions the measurements that are taken on fibers at nonstandard laboratory conditions.
Filing Document | Filing Date | Country | Kind | 371c Date |
---|---|---|---|---|
PCT/CH2006/000569 | 10/12/2006 | WO | 00 | 3/19/2008 |
Number | Date | Country | |
---|---|---|---|
60727013 | Oct 2005 | US |