The embodiments described herein are related to a stretcher for transporting a patient, and more particularly, to a stretcher having a reconfigurable patient support surface and multi-purpose blower.
Generally, healthcare facilities use stretchers to transport patients throughout the facility. When a patient is transported to particular areas of the healthcare facility, for example the operating room, the patient is typically transferred to an operating table. An operating table typically has a width that is less than a width of a stretcher. The width off the operating table is configured to allow as much access for the doctor to the patient as possible. A need remains for a stretcher that is reconfigurable for use throughout the entire healthcare facility. Another need remains for implementing multiple uses of a blower positioned in or on the stretcher.
A stretcher is used across hospital units and for many clinical objectives. Traditionally, stretchers have been used primarily for temporary supporting and transporting a patient throughout a healthcare facility and occupied by the patient for relatively short periods of time. Stretchers have evolved to be used for supporting patients for longer periods of time as the patient waits for admission or treatment, transporting patients throughout a healthcare facility, and supporting patients during procedures. Traditionally, stretchers have had simple mattresses, also called a support surfaces, which are generally not powered and do not provide active therapy for patients. Traditional hospital beds have evolved to having powered surfaces, which are often equipped with air sources such as blowers or compressors that serve a specific purpose such as removing moisture from the interface between the patient and the support surface, inflating and deflating support bladders, or providing a turn function to turn a patient on the support surface. As patients spend more extended periods of time on a stretcher, as opposed to a typical hospital bed, the patients are subject to additional risks that they will develop bed sores, also known as pressure ulcers. This risk has not normally be considered significant in the traditional stretcher use case, but as stretchers are used for more extended periods of time, these risks have increased.
As healthcare delivery has evolved to provide additional care in an ambulatory, rather than inpatient approach, stretchers have also evolved to provide support to patients during lower acuity medical procedures. When stretchers are used to support patients during lower acuity surgical procedures, surgical smoke can be generated as a result of the procedures. Surgical smoke can irritate the respiratory tract and has a high concentration of carcinogens. Despite this harm, surgeons struggle to protect themselves from surgical smoke.
This migration of the traditional stretcher from being used primarily for intra-facility transportation to a patient support apparatus that may be used for patient care and longer term patient support creates challenges that require the stretcher to perform some functions of a traditional hospital bed or surgical table, while still providing ease of mobility throughout the facility. This creates the challenges of having systems that provide some of the functionality of hospital beds or surgical tables while maintaining the mobility and adaptability to several different use cases.
The present disclosure includes one or more of the features recited in the appended claims and/or the following features which, alone or in any combination, may comprise patentable subject matter.
According to a first aspect of the disclosed embodiments, a stretcher for transporting a patient includes a frame and a patient support surface positioned on the frame. The frame includes a center section extending along a length of the frame. At least one side panel is positioned adjacent the center section. At least one surgical accessory rail is provided. The at least one side panel is moveable relative to the center section to expose the surgical accessory rail.
In some embodiments of the first aspect, the at least one side panel can include a right side panel and a left side panel. The right side panel can move independently of the left side panel. The at least one side panel can rotate relative to the center section to expose the surgical accessory rail. The at least one side panel can be slid inward to expose the surgical accessory rail. The at least one side panel can include a head panel positioned in a head section of the frame. A thigh panel can be positioned in a thigh section of the frame. The head panel can move independently of the thigh panel.
Optionally, in the first aspect, the patient support surface can include at least one bolster and a center section. The at least one bolster can be moveable relative to the center section when the at least one side panel is moved. The at least one bolster can include a head bolster positioned on the head panel. A thigh bolster can be positioned on the thigh panel. The head bolster can move independently from the thigh bolster.
It may be desired, in the first aspect, that the patient support surface can include a collapsible section joining the center section to the at least one bolster. The collapsible section can be configured to collapse to expose the surgical accessory rail. The collapsible section can be configured to collapse so that the bolster is moved inward toward the center section. The collapsible section can include collapsible foam. The collapsible section can include diamond cut foam.
It may be contemplated, in the first aspect, that the at least one side panel can be moved to reduce the surface area of the patient support surface. A siderail can be is moved away from the at least one panel prior to moving the at least one panel. The siderail can be folded under the frame prior to moving the at least one panel. The frame can be moved into a flat configuration prior to moving the at least one panel.
In some embodiments of the first aspect a blower can be provided. At least one valve can direct air or suction from the blower through the support surface to at least one of a surgical smoke vacuum inlet, a ventilation and air filtration system to create an air curtain of upwardly flowing filtered air around at least a portion of a perimeter of the patient support surface, and an x-ray bladder of the patient support surface to open an x-ray cassette pocket to facilitate insertion and removal of an x-ray cassette. The blower can be positioned in the patient support surface. The blower can be positioned on the frame. The at least one valve can direct air or suction to at least one of a microclimate management system, a turn assist bladder, a heating and cooling system, an air assisted transfer system, a patient boosting bladder, and a surface width extension and retraction system.
According to a second aspect of the disclosed embodiments, a stretcher for transporting a patient includes a frame and a patient support surface positioned on the frame. The frame includes a center section extending along a length of the frame. A right panel has a right head panel and a right thigh panel. A left panel has a left head panel and a left thigh panel. A right surgical accessory rail and a left surgical accessory rail are provided. The right panel is moveable relative to the center section to expose the right surgical accessory rail. The left panel is moveable relative to the center section to expose the left surgical accessory rail.
In some embodiments of the second aspect, the right panel can move independently of the left panel. The right head panel can move independently of the right thigh panel. The left head panel can move independently of the left thigh panel. Each of the right panel and the left panel can be folded upward to expose the respective surgical accessory rail. Each of the right panel and the left panel can be slid inward to expose the respective surgical accessory rail. Each of the right panel and the left panel can be moveable relative to the center section to reduce a width of the patient support surface.
Optionally, in the second aspect, the patient support surface can include a center section. A right bolster can be positioned on the right panel. A left bolster can be positioned on the left panel. The right bolster can includes a right head bolster positioned on the right head panel. A right thigh bolster can be positioned on the right thigh panel. The left bolster can include a left head bolster positioned on the left head panel. A left thigh bolster can be positioned on the left thigh panel.
It may be desired, in the second aspect, that the patient support surface can include collapsible sections joining the center section to each of the right bolster and the left bolster. The collapsible sections can be configured to collapse to expose the surgical accessory rail. The collapsible sections can be configured to collapse so that the right bolster and the left bolster are moved inward toward the center section. The collapsible sections can include collapsible foam. The collapsible sections can include diamond cut foam.
It may be desired, in the second aspect, that a blower can be provided. At least one valve can direct air or suction from the blower through the support surface to at least one of a surgical smoke vacuum inlet, a ventilation and air filtration system to create an air curtain of upwardly flowing filtered air around at least a portion of a perimeter of the patient support surface, and an x-ray bladder of the patient support surface to open an x-ray cassette pocket to facilitate insertion and removal of an x-ray cassette. The blower can be positioned in the patient support surface. The blower can be positioned on the frame. The at least one valve can direct air or suction to at least one of a microclimate management system, a turn assist bladder, a heating and cooling system, an air assisted transfer system, a patient boosting bladder, and a surface width extension and retraction system.
According to a third aspect of the disclosed embodiments, a stretcher for transporting a patient includes a frame, a patient support surface positioned on the frame, and a blower. At least one valve directs air or suction from the blower through the support surface to at least one of a surgical smoke vacuum inlet, a ventilation and air filtration system to create an air curtain of upwardly flowing filtered air around at least a portion of a perimeter of the patient support surface, and an x-ray bladder of the patient support surface to open an x-ray cassette pocket to facilitate insertion and removal of an x-ray cassette.
In some embodiment of the third aspect, the blower can be positioned in the patient support surface. The blower can be positioned on the frame. The at least one valve can direct air or suction to at least one of a microclimate management system, a turn assist bladder, a heating and cooling system, an air assisted transfer system, a patient boosting bladder, and a surface width extension and retraction system. The frame can include at least one panel that is moveable to expose a surgical accessory rail.
Additional features, which alone or in combination with any other feature(s), such as those listed above and those listed in the claims, may comprise patentable subject matter and will become apparent to those skilled in the art upon consideration of the following detailed description of various embodiments exemplifying the best mode of carrying out the embodiments as presently perceived.
The detailed description particularly refers to the accompanying figures in which:
Referring now to
A lift column 40 extends upward from the lower frame 30 and is operable to raise and lower an upper deck 50 vertically relative to the floor. In the illustrated embodiment, the column 40 includes a plurality of telescoping sections 42 that enable the column to extend and retract to raise and lower an upper deck 50 coupled to a top 44 of the column 40. The lift column 40 is powered by a hydraulic system such as that disclosed in U.S. Patent Publication No. 2005/0198737A1, published on Sep. 15, 2005 and titled HYDRAULIC LIFT APPARATUS FOR A PATIENT SUPPORT,” which is incorporated herein by reference for the disclosure of an applicable hydraulic lift system. Movement of the column 40 between a retracted and extended position is controlled with foot pedals (not shown), in some embodiments. In other embodiments, the column 40 may include an electric motor or manually actuable hydraulic system. In some embodiments, the operation of the column 40 may be controlled by foot pedals. Other mechanisms for moving the column 40 between the retracted position and the extended position are contemplated.
The upper deck 50 extends between the head end 16 and the foot end 18. The upper deck 50 includes a head section 52 positioned adjacent the head end 16, a thigh section 54 adjacent the head section 52 and supported on the column 40, and a leg section 56 extending from the thigh section 54 adjacent the foot end 18. The thigh section 54 extends between the head section 52 and the leg section 56. The thigh section 54 is pivotably coupled to the top 44 of the column 40 and movable relative to the thigh section 54 as suggested by
Referring to
As illustrated best in
A left side panel 92 (shown in
A left side bolster 112 extends along a left edge of the center frame 70 along the left side 14 of the stretcher 10. The left side bolster 112 includes a left head bolster 114 and a left thigh bolster 116. The left head bolster 114 extends along the head section 82. The left head bolster 114 is positioned on the left head panel 94 and moves with the left head panel 94 between the extended and retracted positions. The left thigh bolster 116 extends along the thigh section 86. The left thigh bolster 116 is positioned on the left thigh panel 96 and moves with the left thigh panel 96 between the extended and retracted positions. The left head bolster 114 and the left thigh bolster 116 move independently of one another, in some embodiments. The left side bolster 112 and the right side bolster 102 move independently of one another, in some embodiments.
Each of the bolsters 104, 106, 114, and 116 are moveable relative to the center section 100 to a retracted position to reduce a surface area of the patient support surface 80. The panels 74, 76, 94, and 96 and the respective bolsters 104, 106, 114, and 116 are moveable relative to the center section 100 to reduce a width 120 of the patient support surface 80 between a right edge 122 and a left edge 124 of the patient support surface 80. The right edge 122 of the patient support surface 80 extends along the right side 12 of the stretcher 10, and the left edge 124 of the patient support surface 80 extends along the left side 14 of the stretcher 10. In some embodiments, the upper deck 50 is moved into the flat configuration (shown in
A right siderail 160 extends along the right side 12 of the stretcher 10. A left siderail 162 extends along the left side 14 of the stretcher 10.
Referring to
In the retracted position, the upper deck 50 and the patient support surface 80 are effectively narrower to provide a surgeon better patient access during surgery. It will also be appreciated that, in the retracted position, the bolsters 104, 106, 114, and 116 also serve as barriers to position and immobilize a patient during surgical procedures. Further, in the retracted position, a system of surgical accessory rails 170 are exposed on an underside 190 of the right side panel 72 and the left side panel 92. The system of surgical accessory rails 170 are exposed for accessories such as arm boards, leg stirrups, hip pads, and the like to clamp onto during surgery. In the illustrated embodiment, the surgical accessory rails 170 include a pair of right head accessory rails 174 on the underside 190 of the right head panel 74 and a right thigh accessory rail 176 on the right thigh panel 76. It will be appreciated that, in some embodiments, the stretcher 10 includes any number of right head accessory rails 174 and right thigh accessory rails 176. At least one left head accessory rail (not shown, but similar to the right head accessory rails 174) is positioned on an underside of the left head panel 94. At least one left thigh accessory rail (not shown, but identical to the right thigh accessory rail 176) is positioned on an underside of the left thigh panel 96.
Referring now to
The upper deck 250 extends between the head end 216 and the foot end 218. The upper deck 250 includes a head section 252 positioned adjacent the head end 216, a thigh section 254 extending from the head section 252, and a leg section 256 extending from the thigh section 254 adjacent the foot end 218. The thigh section 254 extends between the head section 252 and the leg section 256. The head section 252 articulates relative to the thigh section 254 between a raised position and a lowered position. Movement of the head section 252 between the raised position and the lowered position is controlled with foot pedals (not shown), in some embodiments. Movement of the head section 252 between the raised position and the lowered position is controlled with user inputs (not shown), in some embodiments. In some embodiments, movement of the head section 252 between the raised position and the lowered position is manually controlled. Other mechanisms for moving the head section 252 between the raised position and the lowered position are contemplated. The leg section 256 articulates relative to the thigh section 254 between a raised position and a lowered position. Movement of the leg section 256 between the raised position and the lowered position is controlled with foot pedals (not shown), in some embodiments. Movement of the leg section 256 between the raised position and the lowered position is controlled with user inputs (not shown), in some embodiments. In some embodiments, movement of the leg section 256 between the raised position and the lowered position is manually controlled. Other mechanisms for moving the leg section 256 between the raised position and the lowered position are contemplated. When the head section 252 is in the lowered position and the leg section 256 is in the raised position, the upper deck 250 is in a flat configuration.
The upper deck 250 includes a center frame 270 extending a length of stretcher 210 from the head end 216 to the foot end 218. A left side panel 272 extends along a left edge of the center frame 270 along the left side 214 of the stretcher 210. The left side panel 272 includes a left head panel 274 and a left thigh panel 276. The left head panel 274 is positioned in the head section 252 of the upper deck 250, and the left thigh panel 276 is positioned in the thigh section 254 of the upper deck 250 adjacent the left head panel 274. The left head panel 274 and the left thigh panel 276 are slidably coupled to the center frame 270 so that the left head panel 274 and the left thigh panel 276 move relative to the center frame 270 between an extended position (shown in
A right side panel (not shown, but identical to the left side panel 272) extends along a right edge of the center frame 270 along the right side 212 of the stretcher 210. The right side panel includes a right head panel and a right thigh panel. The right head panel is positioned in the head section 252 of the upper deck 250, and the right thigh panel is positioned in the thigh section 254 of the upper deck 250 adjacent the right head panel. The right head panel and the right thigh panel are slidably coupled to the center frame 270 so that the right head panel and the right thigh panel move relative to the center frame 270 between an extended position (shown in
A patient support surface 280 is positioned on the upper deck 250 and extends from the head end 216 to the foot end 218. A head section 282 of the patient support surface 280 is positioned on the head section 252 and moves with the head section 252. A thigh section 284 of the patient support surface 280 extends from the head section 282. The thigh section 284 is positioned on the thigh section 254 and moves with the thigh section 254. A leg section 286 extends from the thigh section 284. The leg section 286 is positioned on the leg section 256 and moves with the leg section 256. The thigh section 284 extends between the head section 282 and the leg section 286.
The patient support surface 280 includes a center section 300 extending from the head end 216 to the foot end 218. A left side bolster 302 extends along a left edge of the center section 300 along the left side 214 of the stretcher 210. The left side bolster 302 includes a left head bolster 304 and a left thigh bolster 306. The left head bolster 304 extends along the head section 282. The left head bolster 304 is positioned on the left head panel 274 and moves with the left head panel 274 between the extended and retracted positions. The left thigh bolster 306 extends along the thigh section 286. The left thigh bolster 306 is positioned on the left thigh panel 276 and moves with the left thigh panel 276 between the extended and retracted positions. The left head bolster 304 and the left thigh bolster 306 move independently of one another, in some embodiments.
A right side bolster 312 extends along a left edge of the center frame 270 along the right side 212 of the stretcher 210. The right side bolster 312 includes a right head bolster 314 and a right thigh bolster 316. The right head bolster 314 extends along the head section 282. The right head bolster 314 is positioned on the right head panel and moves with the right head panel between the extended and retracted positions. The right thigh bolster 316 extends along the thigh section 286. The right thigh bolster 316 is positioned on the right thigh panel and moves with the right thigh panel between the extended and retracted positions. The right head bolster 314 and the right thigh bolster 316 move independently of one another, in some embodiments. The right side bolster 312 and left right side bolster 302 move independently of one another, in some embodiments.
Each of the bolsters 304, 306, 314, and 316 are moveable relative to the center section 300 to a retracted position to reduce a surface area of the patient support surface 280. The bolsters 304, 306, 314, and 316 are moveable relative to the center section 300 to reduce a width 320 of the patient support surface 280 between a right edge 322 and a left edge 324 of the patient support surface 280. The right edge 322 of the patient support surface 280 extends along the right side 212 of the stretcher 210, and the left edge 324 of the patient support surface 280 extends along the left side 214 of the stretcher 210.
The patient support surface 280 includes collapsible sections 330 joining the center section 300 to each bolster 304, 306, 314, and 316. The collapsible sections 330 are configured to collapse to reduce the surface area of the patient support surface 280. The collapsible sections 330 are configured to collapse so that the respective bolster 304, 306, 314, and 316 is moved inward toward the center section 300. In some embodiments, the collapsible sections 330 include a collapsible foam. In some embodiments, as shown in
A right siderail 360 extends along the right side 212 of the stretcher 210. A left siderail 362 extends along the left side 214 of the stretcher 210.
Referring to
In the retracted position, the upper deck 250 and the patient support surface 280 are effectively narrower to provide a surgeon better patient access during surgery. Further, in the retracted position, a plurality of surgical accessory rails 370 are exposed under the left side panel 272 and the right side panel. The surgical accessory rails 370 are exposed for accessories such as arm boards, leg stirrups, hip pads, and the like to clamp onto during surgery. In the illustrated embodiment, the surgical accessory rails 370 include a pair of left head accessory rails 374 under the left head panel 274 and a left thigh accessory rail 376 under the left thigh panel 276. It will be appreciated that, in some embodiments, the stretcher 210 includes any number of left head accessory rails 374 and left thigh accessory rails 376. At least one right head accessory rail (not shown, but substantially identical to the left head accessory rails 374) is positioned on the right head panel. At least one right thigh accessory rail (not shown, but identical to the left thigh accessory rail 376) is positioned on the right thigh panel.
Referring now to
A control circuitry 410 is electronically coupled to the air source 402 and configured to operate the air source 402. The control circuitry 410 includes a processor and a memory, wherein the memory stores instructions that are operable by the processor to control the air source 402. The control circuitry 410 includes a temperature sensor 412 and a pressure sensor 414. The pressure sensor 414 compares a pressure in the pneumatic system 400 (for example, in the air source 402) to an atmospheric pressure 420. In some embodiments, the air source 402 is controlled based on feedback from the temperature sensor 412 and the pressure sensor 414. The control circuitry 410 includes a port 416 for coupling to a user interface 418. The port 416 is wired or wireless, in some embodiments. The user interface 418 includes user inputs for controlling the air source 402. The user interface 418 is positioned on the stretcher, in some embodiments. The user interface 418 is separate from the stretcher, in some embodiments, and communicates with the control circuitry 410 via a wired or wireless connection. The control circuitry 410 also includes a port 422, for example, a universal serial bus port, for coupling external devices to the control circuitry 410.
A hose 430 extends from the air source 402 to at least one air inlet 432. The illustrated embodiment includes a pair of air inlets 432. In some embodiments, any number of air inlets 432 are provided. In an embodiment where the air source 402 is positioned in the patient support surface, the hose 430 extends through the patient support surface and the air inlets 432 are positioned at an outer edge or surface of the patient support surface. The air inlets 430 and the hose 430 pull air into the air source 402, when the air source 402 is operational. An interface connector 434 discharges the air from the air source 402 to a microclimate management system 436. The microclimate management system 436 discharges cooling air throughout the patient support surface. The air is discharged through vents 438 in the patient support surface. In some embodiments, the patient support surface includes any number of vents.
Referring now to
A control circuitry 510 is electronically coupled to the air source 502 and configured to operate the air source 502. The control circuitry 510 includes a processor and a memory, wherein the memory stores instructions that are operable by the processor to control the air source 502. The control circuitry 510 includes a temperature sensor 512 and a pressure sensor 514. The pressure sensor 514 compares a pressure in the pneumatic system 500 (for example, in the air source 502) to an atmospheric pressure 520. In some embodiments, the air source 502 is controlled based on feedback from the temperature sensor 512 and the pressure sensor 514. The control circuitry 510 includes a port 516 for coupling to a user interface 518. The port 516 is wired or wireless, in some embodiments. The user interface 518 includes user inputs for controlling the air source 502. The user interface 518 is positioned on the stretcher, in some embodiments. The user interface 518 is separate from the stretcher, in some embodiments, and communicates with the control circuitry 510 via a wired or wireless connection. The control circuitry 510 also includes a port 522, for example, a universal serial bus port, for coupling external devices to the control circuitry 510.
A hose 530 extends from the air source 502 to at least one air inlet 532. The illustrated embodiment includes a pair of air inlets 532. In some embodiments, any number of air inlets 532 are provided. In an embodiment where the air source 502 is positioned in the patient support surface, the hose 530 extends through the patient support surface and the air inlets 532 are positioned at an outer edge or surface of the patient support surface. The air inlets 530 and the hose 530 pull air into the air source 502, when the air source 502 is operational. An interface connector 534 discharges the air from the air source 502 to a microclimate management system 536. The microclimate management system 536 discharges cooling air throughout the patient support surface. The air is discharged through vents 538 in the patient support surface. In some embodiments, the patient support surface includes any number of vents.
A valve 540 is positioned in the hose 530. The valve 540 is coupled to a hose 542. By operating the valve 540 the air in the hose 530 is directed into the hose 542 so that the air can be fed to another system, for example at least one of a microclimate management system, a turn assist bladder, a surgical smoke vacuum, a ventilation and air filtration system, a heating and cooling system, an air assisted transfer system, an x-ray bladder, a patient boosting bladder, and a surface width extension and retraction system.
Additionally, a valve 550 is positioned between the interface connector 534 and the microclimate management system 536. The valve 550 is coupled to a hose 552. By operating the valve 550 the air in the interface connector 534 is directed into the hose 552 so that the air can be fed to another system, for example at least one of a microclimate management system, a turn assist bladder, a surgical smoke vacuum, a ventilation and air filtration system, a heating and cooling system, an air assisted transfer system, an x-ray bladder, a patient boosting bladder, and a surface width extension and retraction system.
Referring now to
A control circuitry 610 is electronically coupled to the air source 602 and configured to operate the air source 602. The control circuitry 610 includes a processor and a memory, wherein the memory stores instructions that are operable by the processor to control the air source 602. The control circuitry 610 includes a temperature sensor 612 and a pressure sensor 614. The pressure sensor 614 compares a pressure in the pneumatic system 600 (for example, in the air source 602) to an atmospheric pressure 620. In some embodiments, the air source 602 is controlled based on feedback from the temperature sensor 612 and the pressure sensor 614. The control circuitry 610 includes a port 616 for coupling to a user interface 618. The port 616 is wired or wireless, in some embodiments. The user interface 618 includes user inputs for controlling the air source 602. The user interface 618 is positioned on the stretcher, in some embodiments. The user interface 618 is separate from the stretcher, in some embodiments, and communicates with the control circuitry 610 via a wired or wireless connection. The control circuitry 610 also includes a port 622, for example, a universal serial bus port, for coupling external devices to the control circuitry 610.
A hose 630 extends from the air source 602 to at least one air inlet 632. The illustrated embodiment includes a pair of air inlets 632. In some embodiments, any number of air inlets 632 are provided. In an embodiment where the air source 602 is positioned in the patient support surface, the hose 630 extends through the patient support surface and the air inlets 632 are positioned at an outer edge or surface of the patient support surface. The air inlets 630 and the hose 630 pull air into the air source 602, when the air source 602 is operational. An interface connector 634 discharges the air from the air source 602 to a microclimate management system 636. The microclimate management system 636 discharges cooling air throughout the patient support surface. The air is discharged through vents 638 in the patient support surface. In some embodiments, the patient support surface includes any number of vents.
A valve 640 is positioned in the hose 630. The valve 640 is coupled to a hose 642. By operating the valve 640 the air in the hose 630 is directed into the hose 642 so that the air can be fed to another system, for example at least one of a microclimate management system, a turn assist bladder, a surgical smoke vacuum, a ventilation and air filtration system, a heating and cooling system, an air assisted transfer system, an x-ray bladder, a patient boosting bladder, and a surface width extension and retraction system. A filter 644 is positionable at the end of the hose 642.
Additionally, a valve 650 is positioned between the interface connector 634 and the microclimate management system 636. The valve 650 is coupled to a hose 652. By operating the valve 650 the air in the interface connector 634 is directed into the hose 652 so that the air can be fed to another system, for example at least one of a microclimate management system, a turn assist bladder, a surgical smoke vacuum, a ventilation and air filtration system, a heating and cooling system, an air assisted transfer system, an x-ray bladder, a patient boosting bladder, and a surface width extension and retraction system. In some embodiments, the reversible air source 602 is set to a vacuum function so that air is drawn into the vents 638 and discharged to the other system through the hoses 642 and 652.
In the embodiments of
Microclimate management (MCM) includes blowing air through a spacer material to cool and dry the skin. This can be increasingly important for stretchers as patient remain on them for long durations as staffing is more limited in the emergency department. This can be the initiation of pressure injuries that subsequently appear in inpatient units later in the patient's stay. Operating turn assist to help turn patients includes using inflating support bladders to provide optimal immersion and minimize peak pressure.
An additional need that the blowers of
Another need that the blowers of
Yet another need that the blowers of
A further need that the blowers of
Yet a further need that the blowers of
Another need that the blowers of
At least one need that the blowers of
The embodiments of
Any theory, mechanism of operation, proof, or finding stated herein is meant to further enhance understanding of principles of the present disclosure and is not intended to make the present disclosure in any way dependent upon such theory, mechanism of operation, illustrative embodiment, proof, or finding. It should be understood that while the use of the word preferable, preferably or preferred in the description above indicates that the feature so described can be more desirable, it nonetheless cannot be necessary and embodiments lacking the same can be contemplated as within the scope of the disclosure, that scope being defined by the claims that follow.
In reading the claims it is intended that when words such as “a,” “an,” “at least one,” “at least a portion” are used there is no intention to limit the claim to only one item unless specifically stated to the contrary in the claim. When the language “at least a portion” and/or “a portion” is used, the item can include a portion and/or the entire item unless specifically stated to the contrary.
It should be understood that only selected embodiments have been shown and described and that all possible alternatives, modifications, aspects, combinations, principles, variations, and equivalents that come within the spirit of the disclosure as defined herein or by any of the following claims are desired to be protected. While embodiments of the disclosure have been illustrated and described in detail in the drawings and foregoing description, the same are to be considered as illustrative and not intended to be exhaustive or to limit the disclosure to the precise forms disclosed. Additional alternatives, modifications and variations can be apparent to those skilled in the art. Also, while multiple inventive aspects and principles have been presented, they need not be utilized in combination, and many combinations of aspects and principles are possible in light of the various embodiments provided above.
This application claims priority under 35 U.S.C. § 119 (e) to U.S. Provisional Application No. 63/510,666, filed Jun. 28, 2023, which is expressly incorporated by reference herein.
Number | Date | Country | |
---|---|---|---|
63510666 | Jun 2023 | US |