This invention relates to a system for evaluating surface characteristics by measuring a fluorescence emission spectrum of a voltage sensitive chromophore affixed to the surface.
Voltage sensitive dyes (VSD), also known as potentiometric dyes or voltage sensitive chromophores, have been used to determine membrane potentials for biological cells. Such dyes have been designed to become embedded within a cell membrane orthogonal to the membrane surface where they are then exposed to the membrane electric field. U.S. Patent Application Publication 2010/0074847 to Madden et al., describes the use of chromophore probes for optical imaging of biological materials. The use of voltage sensitive dyes, is also discussed by L. Loew in an article titled “Potentiometric dyes: Imaging electrical activity in cell membranes” (Pure & Appl. Chem., Vol. 68, pp. 1405-1409, 1996). The use is very similar to that described U. S. Patent Application Publication 2010/0074847.
The absorbance or fluorescence emission of a VSD can be measured for the membrane in both the polarized state and un-polarized state. The shift in the resulting spectrum gives the electrical potential difference between the two states.
Two types of VSDs, which vary in their response mechanism, are known. What are considered “slow” VSDs partition within the cell, and their fluorescence intensity is a Nernstian-concentration-dependent response. What are considered “fast” VSD's do not depend on partitioning and instead respond to the electric field directly via the Stark effect.
Experimental strategies for measuring electric field strength (intensity) at charged solid surfaces using a fluorescent dye as a probe particularly in bulk behavior are described by J. Pope et al. in an article entitled “Measurement of electric fields at rough metal surfaces by electrochromism of fluorescent probe molecules embedded in self-assembled monolayers” (J. Am. Chem. Soc., Vol. 114, pp. 10085-10086, 1992), as well as by J. Pope et al. in an article entitled “Measurements of the potential dependence of electric field magnitudes at an electrode using fluorescent probes in a self-assembled monolayer” (J. Electroanalytical Chem., Vol. 498, pp. 75-86, 2001).
U.S. Pat. No. 5,156,918 No. (Marks et al.) describes the use of a poly(phenylene ether) to which a pyridine-terminated chromophore is attached and the resulting nonlinear optical material can be covalently attached to solid surfaces as a monolayer for various purposes.
Surface potential is a key parameter in colloidal and biological sciences governing interactions between materials such as the attractive or repulsive interaction between materials. Thus, adhesion forces between particles or polymers and a surface that affects particle patterning are directly influenced by surface potentials. Such particles and polymers cover a wide range of materials ranging from biological to inorganic materials. Currently there are two prominent methods to determine surface potentials. However, each of these methods create at least one problem. For example, the method of using streaming potentials described by H. Xie et al. in the article “Zeta potential of ion-conductive membranes by streaming current measurements” (Langmuir, Vol. 27, pp. 4721-4727, 2011) requires expensive equipment and the method is damaging to tested samples. Moreover, the use of atomic force microscopy as described by S. Li et al. in the article “Excluding contact electrification in surface potential measurement using Kelvin probe force microscopy” (ACS Nano, Vol. 10, pp. 2528-2535, 2016) is restricted to very small areas and requires long scanning times.
There remains a need for less burdensome and more versatile methods for determining solid-liquid interface electrical field intensities, and for determining the characteristics of liquids.
The present invention represents a surface evaluation system for evaluating a surface, including:
a chromophore application system for applying a monolayer of a voltage sensitive chromophore to the surface, wherein the applied voltage sensitive chromophore is covalently bound to the surface and has a fluorescence emission spectrum which varies in accordance with a characteristic of the surface;
an irradiation source that irradiates the monolayer of the covalently bound voltage sensitive chromophore with actinic radiation;
a fluorescence sensing system for measuring a fluorescence emission spectrum from the irradiated monolayer of the covalently bound voltage sensitive chromophore; and
an analysis system that analyzes the measured fluorescence emission spectrum to evaluate the characteristic of the surface.
This invention has the advantage that characteristics of a surface can be conveniently measured by measuring the fluorescence emission spectrum of a voltage sensitive chromophore attached to the surface.
It has the additional advantage that defects and features on the surface that are not detectable with visible radiation can be inspected.
It is to be understood that the attached drawings are for purposes of illustrating the concepts of the invention and may not be to scale. Identical reference numerals have been used, where possible, to designate identical features that are common to the figures.
The following discussion is directed to various embodiments of the present invention and while some embodiments can be desirable for specific uses, the disclosed embodiments should not be interpreted or otherwise considered to limit the scope of the present invention, as claimed below. In addition, one skilled in the art will understand that the following disclosure has broader application than is explicitly described in the discussion of any embodiment.
The invention is inclusive of combinations of the embodiments described herein. References to “a particular embodiment” and the like refer to features that are present in at least one embodiment of the invention. Separate references to “an embodiment” or “particular embodiments” or the like do not necessarily refer to the same embodiment or embodiments; however, such embodiments are not mutually exclusive, unless so indicated or as are readily apparent to one of skill in the art. The use of singular or plural in referring to the “method” or “methods” and the like is not limiting. It should be noted that, unless otherwise explicitly noted or required by context, the word “or” is used in this disclosure in a non-exclusive sense.
Each term that is not explicitly defined in the present application is to be understood to have a meaning that is commonly accepted by those skilled in the art. If the construction of a term would render it meaningless or essentially meaningless in its context, the term definition should be taken from a standard dictionary.
The use of numerical values in the various ranges specified herein, unless otherwise expressly indicated otherwise, are considered to be approximations as though the minimum and maximum values within the stated ranges were both preceded by the word “about”. In this manner, slight variations above and below the stated ranges can be used to achieve substantially the same results as the values within the ranges. In addition, the disclosure of these ranges is intended as a continuous range including every value between the minimum and maximum values.
When a solid is immersed in a fluid, an electric field is generated at the interface. This interfacial electric field can also be characterized by a corresponding interfacial potential difference. In accordance with embodiments of the invention, the interfacial electric field (or the corresponding interfacial potential difference) can be quantified by measuring a wavelength shift of the fluorescence emission spectrum between a fast VSD present in the liquid and a monolayer of the VSD affixed to a solid surface in contact with the same liquid (or via a probe with bound VSD immersed in the liquid). Given the same solvent shell structure and reorganization on excitation for bulk and surface VSD bound at the interface, the fluorescence wavelength shift upon excitation is due to the extra energy needed to flip the chromophore dipole in the interfacial electric field. Chromophore dipole moments for a series of voltage sensitive dyes are available in the literature (for example, see P. Fromherz, “Monopole-dipole model for symmetrical solvatochromism of hemicyanine Dyes,” J. Phys. Chem., Vol. 99, pp. 7188-7192, 1995). The solvatochromic shift of the VSD is cancelled out since both fluorescence spectra (that is, the bound VSD monolayer and dissolved VSD in the liquid) are recorded in the same solvent.
In various embodiments, the shift in fluorescence emission spectrum of a bound VSD at a solid/liquid interface can be used to perform various functions such as calculating the electric field strength at the interface between a liquid and a surface, identifying or characterizing liquid, determining characteristic differences between a plurality of liquids, and evaluating surface characteristics/uniformity.
For other types of surfaces 100, linking groups 105 having other appropriate coupling agents 147 can be used. For example, polymers have carboxylic acid groups (—COOH) that can react with an alcohol (C—OH) coupling group to form an ester (—COOC) linkage. Surfaces with carboxylic acid groups can also couple with an amino (—NH2) coupling group to form an amide (—CONH—) linkage. Alternatively, carboxylic acid groups on surfaces can be coupled to amine coupling groups to form an amino acid linkage. In other examples, isocyanate coupling groups (—NCO) can be used to form a linkage to surfaces with carboxylic acids, alcohols, and amines. Other reactive functional groups include aromatic chloromethyl, amide, hydrazide, aldehyde, hydroxyl, thiol and epoxy. Clearly, similar reactions can be used for many types of surfaces 100, as long as the surfaces 100 have functional groups, or can be modified with functional groups, that will react with the coupling agent 147 that is included in the linking group 105.
Returning to a discussion of
In an exemplary embodiment, the voltage sensitive chromophore precursor 120 includes a p-substituted dialkylamino aryl group 180 that is linked to a terminal N-containing heterocyclic aromatic group 185 with a conjugate linkage 182 as illustrated in
The parts of the voltage sensitive chromophore precursor 120 (i.e., the p-substituted dialkylamino aryl group 180 and the N-containing heterocyclic aromatic group 185) are conjugatively linked and are designed to be reacted with the chromophore component 145 to yield the voltage sensitive chromophore 152. One skilled in the art will recognize that a conjugatively-linked system (also known as a “conjugated system”) is a system of bonded atoms including pi orbitals that have been delocalized through bonding. The pi electrons in these orbitals are delocalized over the atoms that participate in the pi orbital bonding. This normally lowers the total energy of the system, and is favored in such bonding as alternating single and double bonds, certain functional groups attached to aromatic rings, alternating single and double bonds attached to aromatic rings, etc.
In alternate embodiments, other types of molecules can be chosen for the two chromophore precursors (i.e., chromophore component 145 and the voltage sensitive chromophore precursor 120), depending on the particular voltage sensitive chromophore that is used (e.g., any of the voltage sensitive chromophores in the aforementioned article by P. Fromherz, entitled “Monopole-dipole model for symmetrical solvatochromism of hemicyanine Dyes”). Typically, the resulting voltage sensitive chromophore is a conjugated system, and is formed from any number of alternating single and double bonds and aromatic rings, including functional groups on the aromatic rings. The requirement is that the precursors are hindered from rotation around internal linkages, so as to maintain the largest dipole moment that reverses upon excitation with actinic light, and that the precursors can be conjugatively linked to form the final chromophore.
The result of the first and second reactions 110, 125 is to provide a monolayer of the voltage sensitive chromophore that is covalently bound to the surface 100 (i.e., the voltage sensitive chromophore surface 130). In accordance with the method of
As illustrated in
After, the voltage sensitive chromophore surface 130 and the voltage sensitive chromophore solution 155 have been prepared, fluorescence emission spectra are measured and compared. A measure surface fluorescence emission spectrum step 135 is used to measure a surface fluorescence emission spectrum 140 of the voltage sensitive chromophore surface 130 while it is in contact with a sample of the liquid 132. This involves irradiating the voltage sensitive chromophore surface 130 with actinic radiation provided by an irradiation source while it is in contact with the liquid and using a fluorescence sensing system to measure the emission spectrum of the resulting fluorescent radiation. In this context, “actinic radiation” is radiation that is adapted to stimulate fluorescence of the voltage sensitive chromophore. Typically, the actinic radiation for many voltage sensitive chromophores will be ultraviolet (UV) radiation or short wavelength visible radiation. For example, in the case of the exemplary voltage sensitive chromophore 152 of
In some configurations, the solution fluorescence emission spectrum 165 can be predetermined and stored for later comparison with the surface fluorescence emission spectrum 140. For example, a number of different surfaces 100 can be evaluated using the method of
Once the surface fluorescence emission spectrum 140 and the solution fluorescence emission spectrum 165 are determined, they are compared using a determine interfacial electric field intensity step 170 to determine a measured interfacial electric field intensity 175. As will be discussed below, in an exemplary arrangement, the interfacial electric field intensity 175 is determined responsive to a Stark shift in the peaks of the fluorescence emission spectra.
It can be seen that the surface fluorescence emission spectrum 140 measured for the voltage sensitive chromophore surface 130 (
Those skilled in the art will recognize that it will typically be desirable to smooth the measured fluorescence emission spectra using an appropriate smoothing algorithm to smooth out any measurement noise. Any appropriate curve smoothing algorithm known in the art can be used in accordance with the present invention. Example curve smoothing algorithms would include low-pass filter algorithms, smoothing spline algorithms, and curve-fitting algorithms that fit a standard functional form (e.g., a Gaussian function) to the measured data. A peak-finding algorithm can then be used to identify the peaks of the fluorescence emission spectra in order to determine the Stark shift. Peak-finding algorithms are well-known in the data processing art, and any appropriate algorithm can be used in accordance with the present invention.
ΔEsensor=hc{tilde over (ν)}sensor (1)
where h=6.626×10−34 J·s is Planck's constant, c=3.00×108 m/s is the speed of light, and {tilde over (ν)}sensor is the wave number (1/wavelength) of the emitted photon.
The excited state energy level 364 associated with the voltage sensitive chromophore 152 when it is in solution is lower than the excited state energy level 362 for the sensor where the voltage sensitive chromophore 152 is bound to the surface 100, such that the emitted photons have a lower energy ΔEsolution given by:
ΔEsolution=hc{tilde over (ν)}solution (2)
where {tilde over (ν)}solution is the wave number of the emitted photon.
The difference between the energies of the excited state energy levels 362, 364 (ΔEinterface) corresponds to the Stark shift:
where the Stark shift Δ{tilde over (ν)}Stark={acute over (ν)}sensor−{tilde over (ν)}solution is the difference between the wave numbers of the emitted photons in the sensor and solution configurations. The Stark shift can also be related to a corresponding difference in the wavelengths:
where λsensor and λsolution are the wavelengths of the emitted photons in the sensor and solution configurations, respectively.
The energy difference ΔEinterface corresponds to the energy difference between the two orientations of the chromophore dipole in the presence of the interfacial electric field, which is directly related to the magnitude of interfacial electric field {right arrow over (E)} and the chromophore dipole moment {right arrow over (P)}, and the angle θ between them:
ΔEinterface=2|{right arrow over (E)}∥{right arrow over (P)}|cos θ (5)
Combining Eqs. (3) and (5) and solving for the magnitude of the interfacial electric field {right arrow over (E)} gives:
Therefore, experimentally measuring the Stark shift Δ{tilde over (ν)}Stark enables the magnitude of the interfacial electric field |{right arrow over (E)}| to be determined given estimates of the dipole moment and the angle that the chromophore dipole makes with the interfacial electric field at the interface. The cosine function is relatively close to 1.0 for a fairly large range of angles around θ=0°, so that it can be neglected in many cases without significantly impacting the calculated result. Even if θ is as large as 60°, the calculation of the electric field would only be in error by a factor of about 2×. Note that any solvatochromic shift associated with the particular liquid is nulled out given that both spectra are taken in the same liquid.
The dipole moment of the voltage sensitive chromophore 152 of
{right arrow over (P)}=e·δ/2=(1.60×10−19 C)(0.217 nm)/2=1.74×10−29 C·m (7)
where e=1.60×10−19 C is the elementary charge, and δ=0.217 nm is the estimated distance that the charge is displaced when the polarity of the dipole is flipped. Using this value, the magnitudes of the interfacial electric fields for each case can be calculated using Eq. (6). The results are summarized in Table 1, where the dipole angle was estimated to be θ=45°.
From Table 1, it can be seen that for a given surface (e.g., quartz with a monolayer of a particular voltage sensitive chromophore 152), the magnitude of the interfacial electric field, and therefore the Stark shift, vary according to the characteristics of the liquid. In some embodiments this fact can be used to provide a sensor for determining characteristics of an unknown liquid. One of the primary liquid characteristics that will influence the interfacial electric field is the dielectric constant of the liquid.
In an exemplary arrangement, the determine characteristic of liquid step 200 computes the Stark shift, and then uses the curve of
In some embodiments, the determine characteristic of liquid step 200 determines the liquid characteristic 205 based on only the surface fluorescence emission spectrum 140, without measuring the solution fluorescence emission spectrum 165. For example, surface fluorescence emission spectra 140 can be measured for a library of common liquids using the voltage sensitive chromophore surface 130. A surface fluorescence emission spectrum 140 can then be measured for an unknown liquid 132. The resulting surface fluorescence emission spectrum 140 for the unknown liquid can then be compared to the library of surface fluorescence emission spectra 140 for the common liquids to determine whether there is a match. If so, it can be determined that the unknown liquid is likely to be the matching liquid from the library. In this example, the determined characteristic 205 is the identity of the liquid 132.
A variety of applications can be envisioned for the method of
While the method of
The determine interfacial electric field intensity step 170 of
In an exemplary class of spectrum comparison methods, a first wavelength 430 and a second wavelength 435 are defined. In the illustrated example, the first wavelength 430 is 650 nm and the second wavelength 435 is 560 nm. Preferably, the first and second fluorescence emission spectra 225, 235 overlap with both the first and second wavelengths 430, 435. A first intensity ratio R1=A1/B1 is determined for the first fluorescence emission spectrum 225, where A1 is the value of the first fluorescence emission spectrum 225 at the first wavelength 430, and B1 is the value of the first fluorescence emission spectrum 225 at the second wavelength 435. Likewise, a second intensity ratio R2=A2/B2 is determined for the second fluorescence emission spectrum 235, where A2 is the value of the second fluorescence emission spectrum 235 at the first wavelength 430, and B2 is the value of the second fluorescence emission spectrum 235 at the second wavelength 435.
If a fluorescence emission spectrum is centered between the first and second wavelengths 430, 435, then the intensity ratio will be approximately equal to 1.0. If the spectrum is to the left of center, then the intensity ratio will be smaller than 1.0, and if the spectrum is to the right of center, then the intensity ratio will be greater than 1.0. Therefore, the intensity ratio will be a measure of the location of the spectral peak for the spectrum, and the difference in the intensity ratios will be a measure of the shift in the spectral peaks.
Table 2 shows exemplary ratios determined for the fluorescence emission spectra of
In some embodiments, the intensity ratios can be related to the quantity being measured (e.g., the interfacial electric field intensity 175 (
In other types of spectrum comparison methods more than two wavelengths can be compared. For example, additional spectra ratios can be computed between additional wavelengths pairs to provide additional information about the shape and position of the spectrum. In some cases, a spectral difference can be determined by subtracting the two spectra on a wavelength-by-wavelength basis. The spectral difference can then be analyzed to characterize the difference between the spectra. For example, an RMS spectral difference can be computed between the spectra measured for two liquids to determine if the spectra are statistically indistinguishable, indicating that the two liquids may be the same. In some configurations, the spectra can be analyzed to determine corresponding color values (e.g., in the well-known L*a*b* color space), and color differences can be computed (e.g., ΔE* values) to characterize the difference between two spectra.
In some embodiments, many spectra are compared to determine which liquid is present at the surface. In such cases, any classification and discrimination analysis methods known in the statistics and chemometrics fields can be used. These methods include the use of the Mahalanobis distance, K nearest neighbor, discriminant analysis, cluster analysis, factor analysis, and principal component analysis (PCA) and obtaining spectral contrast angles between pairs of spectra. As the number of compared spectra increases, the more useful these methods become, especially PCA and cluster analysis, leading to full pattern recognition of the spectral shape of a given liquid/chromophore spectral response. Many of these techniques can be extended to spectral modeling or deconvolution, allowing a liquid to be identified by all of its wavelength responses.
An irradiation source 510 is positioned to irradiate the voltage sensitive chromophore surface 130 with actinic radiation that stimulates fluorescence of the voltage sensitive chromophore 152. A fluorescence sensing system 515 (e.g., a spectrophotometer) measures a surface fluorescence emission spectrum 140 which is emitted by the irradiated voltage sensitive chromophore surface 130. In some embodiments, the fluorescence sensing system 515 is a micro-spectrophotometer which is adapted to sense the surface fluorescence emission spectrum 140 emitted from a localized area voltage sensitive chromophore surface 130. In this case, the irradiation source 510 can illuminate the same localized area (as shown in
Generally, the surface fluorescence emission spectrum 140 will be measured while the voltage sensitive chromophore surface 130 is in contact with some fluid (e.g., a liquid or a gas). In some embodiments, the surface fluorescence emission spectrum 140 can be measured while the voltage sensitive chromophore surface 130 is exposed to air (or some other gas). In other embodiments, the surface fluorescence emission spectrum 140 can be measured while the voltage sensitive chromophore surface 130 is in contact with a liquid 132. In this case, a liquid applicator (not shown) can be used to bring the liquid 132 into contact with the voltage sensitive chromophore surface 130. For example, the liquid 130 can be added into the recessed region in the surface holder 505 of
An analysis system 520 is used to analyze the surface fluorescence emission spectrum 140 to evaluate a characteristic of the voltage sensitive chromophore surface 130. In an exemplary embodiment, the characteristic is the interfacial electric field intensity 175 (
In some embodiments, the fluorescence sensing system 515 measures a plurality of surface fluorescence emission spectra 140 at a lattice of spatial positions 525 on the voltage sensitive chromophore surface 130. In this case, either the surface holder 505 or the fluorescence sensing system 515 can be moved in a scanning path in order to sense the surface fluorescence emission spectra 140 at the different positions. In such configurations, the analysis system 520 can be used to analyze the plurality of surface fluorescence emission spectra 140 to evaluate a uniformity of a characteristic of the voltage sensitive chromophore surface 130. This can be useful for a number of applications, including detecting surface defects or inspecting features on the surface.
There are a wide variety of applications that involve depositing a pattern of features onto a surface, including the fabrication of electrical components or devices (e.g., on a silicon wafer or on a glass or polymer substrate). In some cases, it can be difficult to inspect such components, particularly when they may include features that are fabricated using transparent, or partially transparent, materials. Such features may not be easily detected by conventional machine vision inspection techniques that involve capturing an image of the surface using visible light. However, the presence of the features will produce local perturbations in the interfacial electric field of the surface. These perturbations can be detected using methods in accordance with embodiments of the present invention by applying a monolayer of voltage sensitive chromophore to the surface and measuring the fluorescence emission spectrum. The presence of edges and surface textures on the surface can also affect the amplitude and direction of the emitted fluorescent light. Such perturbations can be detectable with appropriate types of sensors.
In one such embodiment, the surface 100 having the features thereon is treated to attach a monolayer of the voltage sensitive chromophore 152 (
The analysis system 520 can analyze the measured pattern of fluorescent light in order to inspect the surface. In some embodiments, features in the measured pattern of fluorescent light can be compared to a reference pattern associated with the device being fabricated, for example to verify that there are no shorts or voids in a pattern of micro-wires. In some embodiments, the measured pattern of fluorescent light can be analyzed to detect surface defects (e.g., non-uniformities in a surface region that was supposed to be uniform).
When the surface evaluation system 500 is being used to measure a pattern of fluorescent light, it may not be necessary to measure the fluorescent light while the voltage sensitive chromophore surface 130 is in contact with a liquid 132, particularly if the goal is to detect non-uniformities or other defects rather than determining quantitative interfacial electric field intensities. In such cases, the pattern of fluorescent light can be measured when the voltage sensitive chromophore surface 130 is exposed to air (or some other gas).
The methods described herein have many advantages compared to other methods for using voltage sensitive chromophores currently used in the art. The describes methods provide a chromophore that is covalently attached to the surface. In contrast, other methods, such as those described in the aforementioned articles by Loew and Pope et al., use chromophores embedded in membranes or self-assembled monolayers. In such arrangements, the location of the chromophore is not known precisely, nor is the measured fluorescence emission spectra guaranteed to be from a chromophore in a known position with respect to the surface or membrane, but can be mixed with the spectra of unembedded chromophore molecules and chromophore molecules at various distances from the surface. In the methods of the present invention, the voltage sensitive chromophore is attached to the surface in a known way, and is created only when it is attached to the surface, so the position of the chromophore is known precisely.
Another advantage of the present invention is that it can be used with a wide variety of different surfaces including silica, polymer, metal, inorganic, and other surfaces. Other methods have restrictions on the sample surface, such as requiring a biological membrane a metal surface or a metal electrode. For example, the method described in the aforementioned articles by Pope et al. require that the surface be a metal electrode, and specifically a silver electrode, in order to enhance the fluorescence of the chromophore. Our method has been shown to work without metal-enhanced fluorescence.
In other methods, such as those described in the aforementioned articles by Pope et al., external fields need to be applied (for example, to an electrode surface or a membrane surface) to provide a change in the electric field to induce a Stark shift, which is measured by a shift in the fluorescence emission spectrum. In methods of the present invention, the emission of the chromophore on the surface is compared to the emission of the chromophore in solution, which enables the Stark shift to be determined and the interfacial electric field to be calculated without the use of any applied field. This makes the measurement dependent only on the voltage sensitive chromophore, the surface, and the liquid, and is not dependent on any applied fields.
Because the methods in this application do not require applied fields, the methods are simpler and do not require the use of more complicated instrumentation. Additionally, the methods can be applied over a large area, and need not be confined to smaller areas, such as in atomic force microscopy.
Furthermore, the methods are non-destructive for the sample, and only require that the surface be modified to include the voltage sensitive chromophore. In other methods for determining an interfacial electric field, such as the streaming potential method described in the aforementioned article by Xie et al. and the atomic force microscopy method described in the aforementioned article by Li et al., the sample is significantly changed, and may not be used for any other testing.
Generally, the methods of the present invention are simpler, more precise, more accurate, and more versatile than prior art methods for measuring interfacial electric fields. Furthermore, the present methods can be used to characterize large areas of a surface, and to provide a sensor for characterizing a liquid.
The invention has been described in detail with particular reference to certain preferred embodiments thereof, but it will be understood that variations and modifications can be effected within the spirit and scope of the invention.
Reference is made to commonly assigned, co-pending U.S. patent application Ser. No. ______ (Docket K002020), entitled: “Method for measuring solid-liquid interfacial potentials,” by A. Wexler et al.; to commonly assigned, co-pending U.S. patent application Ser. No. ______ (Docket K002101), entitled: “Method for characterizing a liquid,” by A. Wexler et al.; and to commonly assigned, co-pending U.S. patent application Ser. No. ______ (Docket K002102), entitled: “Method for determining a characteristics difference between liquids,” by A. Wexler et al., each of which is incorporated herein by reference.