In laparoscopic and robotic surgery, wristed and articulating instruments provide additional dexterity for the surgeon—enabling access to tissue in small, constrained spaces.
Commercially available articulating surgical instruments have overall diameters in the range of 5 to 10 mm and have a variety of mechanical features to enable articulation away from the primary instrument shaft. Some of these instruments include one discreet joint for each articulation direction, while others use multiple joints for a given direction. Articulation is controlled through tendons or rods extending axially through the primary instrument shaft such that a combination of pulling, pushing or rotating the rods or tendons results in movement at the instrument end effector.
There are very some small diameter instruments, also known as “mini-lap” instruments, used in surgery. Mini-lap instruments in use today are fabricated in a 3 mm (or less) diameter that is less invasive to the patient that larger diameter instruments. In fact, some mini-lap instruments do not require surgical incision closure and can be used with percutaneous entry into the abdominal cavity. These instruments are typically “straight-stick” instruments that replicate the functions of their larger, 5 mm counterparts.
There is a strong desire to miniaturize surgical instrumentation as a means to reduce patient tissue trauma. However, size reductions typically result in a reduction in capability and/or an increase in the cost of fabrication. This application describes instrument designs that allow for manufacture of wristed or articulating instruments in a 3 mm outer diameter profile while minimizing the overall impact to cost and capability. This provides surgeons with instruments having excellent dexterity in combination with a diameter that can minimize tissue trauma relative to more conventionally-sized instruments.
This application shows and describes surgical instruments having instrument shafts shafts and end effectors positioned at the distal end of an instrument shaft. The surgical instruments may be ones configured for manual operation using a proximal drive mechanism in the form of a manually actuated instrument handle at the proximal end of the instrument shaft. Alternatively, for robotically assisted operation, the instrument's proximal drive mechanism receives motion from robotically controlled actuators operating in accordance with surgeon input to a surgical robotic system.
The described embodiments make use of laser cut tubing technology to create the pivots of the shaft in-line with the outer diameter. This will maximize the internal diameters of the instruments, which will be occupied with tendons to create the articulating motion.
The distal joint is designed to allow the distal-most segment of the tube (comprising the jaw members) to not only angle about the pivot point at pivot 14, but also to dilate about the same pivot point. This is because the distal-most segment is split along its axis—enabling one section to bend in different directions and at different amounts than the opposite section. The dilation is a flowering type of dilation as opposed to a more symmetrical type of dilation that occurs when a stent expands Referring to
Co-pending application Ser. No. 15/003,656, incorporated herein by reference, describes articulating tubes fabricated with laser cutting methods. In those embodiments, multiple joints are laser cut into to the tube. These joints allow the tube to bend about an axis where the distal portion of the bend and the proximal portion of the bend are still tubular in nature.
Note that as described in Co-pending application Ser. No. 15/003,656, the instrument shaft may include an inner core 16 (
As shown in
The ability to cut different jaw shapes into the outer tube with a laser could reduce the overall cost of the instrument and enable the inner core to be a common part amongst all instruments. Alternatively, components of the jaw profiles could also exist on the inner core if desired. This may open up opportunities to create multi-function instruments where the external tube structure can have one purpose and the inner core can serve a second purpose.
The described instrument shaft design is beneficial in that it enables a very small diameter instrument (e.g. one with a 3 mm outer diameter) to be configured to articulate by maximizing the space within the outer tube lumen so that tendons may pass through the lumen. It may also decrease the costs for machining surgical end effectors by using laser cutting technology, and allow for a reduction in the distance from the articulation joint to the end effector tip compared with other instrument designs.