The present invention pertains generally to laser systems for performing ophthalmic laser surgeries. More particularly, the present invention pertains to laser systems that can be customized to perform a particular surgical laser procedure on a specifically identified ophthalmic tissue. The present invention is particularly, but not exclusively, useful as a system and method for customizing a same laser system to selectively perform different ophthalmic surgeries by specifying a laser beam configuration, establishing a base reference datum for the system, and identifying a scanning mode for each particular surgical procedure.
Any ophthalmic laser surgical procedure requires a combination of precision, accuracy and efficacy that must be maintained within the range of the operational capabilities of the laser system that is being used. Thus, the system itself, or a particular subcomponent of the system, can be a limiting factor for a surgical procedure. Moreover, there are many different operational requirements, and many different operational factors that are unique for each ophthalmic laser procedure. Stated differently, the surgical requirements for a laser system will change depending on the particular procedure that is to be followed and on the specific part of the eye where the surgery is to be performed. A consequence of all this is the need for a versatile laser system.
With recent improvements in so-called “femtosecond” laser systems, as well as in imaging techniques such as Optical Coherence Tomography (OCT), it is now possible to perform laser surgical procedures deep into the eye. No longer is the cornea the only ophthalmic tissue of interest for laser surgery. Specifically, laser surgical procedures involving the crystalline lens, trabecular meshwork, sclera, vitreous and retina are now being considered for laser surgery. Not surprisingly, each of these different tissues in the eye has its own unique set of operational and anatomical issues.
As is well known, the Laser Induced Optical Breakdown (LIOB) of ophthalmic tissue can be efficaciously employed using a pulsed femtosecond laser beam in many different surgical procedures. Alternative tissue interactions with the related effects may use nanosecond or picosecond laser pulses. These procedures include LASIK flap, corneal surgery (e.g. Keretoplasty), refractive corrections, cataract related surgeries, tissue relaxing incisions, glaucoma surgery, and posterior/retinal surgeries. As implied above, each case for LIOB is unique and each procedure has its own particular requirements. Consequently, although similar considerations for the employment of LIOB may be required in each of these different cases/procedures, they will inevitably lead to different operational conclusions. Primarily, the required considerations include: reference datum, laser patient interface, laser beam configuration, focal point placement, and focal point scanning. Importantly, these considerations are interrelated and must be evaluated in the context of other considerations.
With the above in mind, the specification for a Maser beam configuration involves the selection of value ranges for operational parameters that will best accomplish the required alteration of a selected target tissue. In particular, based on the specific anatomical features that will be involved, the selection of operational parameters typically includes: 1) choosing an appropriate wavelength for the laser beam; 2) determining “fluence” (energy density) at specific locations along the beam path; 3) selecting a pulse energy; 4) establishing a pulse rate; 5) setting a pulse duration; 6) selecting an appropriate patient interface; and 7) setting a focusing means.
Insofar as the focal point placement is concerned, in addition to the anatomical considerations for the type of tissue that is being targeted, it may also be necessary to consider the proximity of the target tissue to a different type of tissue (i.e. the location of a tissue interface). In the event, within permissible cost restraints, the placement of a focal point will preferably be accomplished with the greatest possible precision. In all cases this means that an appropriate, cost effective, base reference datum needs to be established. And, the focal point of the laser beam must be accurately located relative to the base reference datum. For ophthalmic surgeries, such a base reference datum may be a point, a line (e.g. an axis), or a surface. Further, the precision required for establishing the reference datum may need to involve the use of high-level mathematical computations (e.g. expansion series), and/or sophisticated imaging techniques such as OCT.
In addition to laser beam configuration and focal point placement, all ophthalmic laser surgeries require the selection of an appropriate scanning operation. Specifically, this selection will generally involve the identification of a focal point path, along with the determinations of the rate of focal point movements and the spacing between adjacent focal points. This may also involve calibration requirements for the entire laser beam generating unit.
As an additional consideration for customizing an appropriate laser system for performing an ophthalmic surgery, it is necessary to evaluate the particular patient interface that is to be used. In particular, any distortions of the eye that may be caused when stabilizing the eye with a patient interface (i.e. a contact lens) needs to be accounted for. In some instances the possible adverse effects may be minimal and, therefore, generally acceptable. On the other hand, such distortions may degrade optical qualities of the laser beam to the point where the laser beam becomes ineffective. Suffice to say, the patient interface may be a critical factor for consideration with creating an efficacious surgical laser system.
In light of the above, it is an object of the present invention to provide a system and method for performing ophthalmic laser surgery on selected tissue in an eye which allows for customizing the system based on considerations of the specific requirements of a particular surgical procedure. Another object of the present invention is to provide a system and method for performing an ophthalmic laser surgical procedure with a specific patient interface, which allows a user to selectively specify a laser beam configuration, in combination with a selectively established base reference datum, and a selectively identified scanning mode. Yet another object of the present invention is to provide a system and method for performing an ophthalmic laser surgical procedure which is easy to use, is simple to establish and is cost effective.
In accordance with the present invention, a multi-mode operable laser system can be customized to perform ophthalmic laser surgery on selected tissue inside an eye. Structurally, the system includes a laser unit, for generating a laser beam, and for focusing the laser beam to a focal point. Preferably, the laser beam is a pulsed femtosecond laser beam that can be configured to perform a Laser Induced Optical Breakdown (LIOB) procedure on the selected target tissue.
A mode selector is operationally connected with the laser unit to establish and define an operational mode for the system. This is, done by selecting system functionalities, based on compatibility considerations, that are best suited to perform the required surgical procedure on the selected target tissue. In detail, the operational mode includes:
1) a specification of the configuration parameters that are to be used for the laser beam, to include the wavelength of the laser beam, fluence measurements along the laser beam, pulse power, pulse rate and pulse duration;
2) a process for establishing a base reference datum in the eye that uses point selection, axis selection, expansion series and/or real-time imaging or detector techniques; and
3) a plan for moving the focal point of the laser beam in accordance with a defined scanning procedure, wherein the scanning procedure is based on considerations of the rate of focal spot movement, spacing between focal points, pattern description (e.g. raster or spiral) and calibration points.
In addition to the laser unit and the mode selector, the system of the present invention also includes a computer for controlling the laser unit. More specifically, the computer is connected to both the laser unit and to the mode selector. With these connections, the computer is responsive to the mode selector to control the laser unit for the purpose of implementing the operational mode.
An additional feature of the system involves considerations for the incorporation of a patient interface. Whenever such an interface is used, its purpose is to stabilize a patient's eye during ophthalmic surgery. It can happen, however, that the use of an interface can introduce optical distortions into the eye. As a practical consideration, patient interfaces can have different levels of effect, on different areas of the eye, depending on whether they are a planar contact lens, a curved contact lens, a conformable (e.g. water filled) contact lens, or contact free. In the event, they may affect other functional aspects of a customized laser system and, therefore, they must be considered and accounted for in the system set-up and subsequent operations.
Due to the plethora of set-up possibilities for different operational modes, and due to the distinct possibility that one particular operational mode may be satisfactory for a a sequence of surgical procedures on respectively different patients, the present invention provides for a so-called default operational mode. Accordingly, the functionalities that constitute the operational mode are preselected. In detail, the configuration parameters for the laser beam are preset, the process for establishing the base reference datum is preset, and the plan for scanning the laser beam's focal point is also preset. Flexibility for the system, however, is provided by allowing the system user to move from the default status of any functionality, as desired.
With the above disclosure in mind, it is to be appreciated that the present invention envisions an operation of the system which includes switching between two different operational modes. For example, consider an integrated operation that requires both a cataract surgery and corneal incisions. In such an integrated operation, the particular procedure that will be deeper in the eye (i.e. the cataract surgery) will likely be done first. In any event, the set-up for both modes of operation can be accomplished prior to starting the integrated operation.
In line with earlier disclosure, each procedure (cataract surgery and corneal incisions) will be accomplished separately, and may require different operational modes. In this case, the system set-up initially requires that compatibility considerations be made for an operational mode to perform cataract surgery (i.e. a first operational mode). This involves determining the appropriate operational parameters for the laser beam (e.g. fluence and spot size), establishing a base reference datum, prescribing scanning patterns, and employing a patient interface (e.g. pressure on the eye) for a cataract surgery (e.g. capsulotomy and lens fragmentation).
Following the establishment of the first mode of operation (e.g. cataract surgery) a second mode of operation (e.g. corneal incisions) is established by essentially repeating the same compatibility considerations made for the first mode. This requires additional evaluations and appropriate adjustments of the laser beam's operational parameters and scanning patterns, possibly the selection of a new base reference datum, and a possible modulation of the patient interface. More specifically, insofar as the patient interface is concerned, it can likely happen that only very little pressure needs to be exerted on the eye during the cataract surgery portion of the integrated operation. This minimal pressure is generally preferable in order to minimize posterior and other corneal deformations that could otherwise introduce unwanted distortions of the laser beam. On the other hand, greater pressures are likely required during the corneal incisions portion of the integrated operation. This is so because of the need to stabilize the cornea during corneal incisions, and the fact that, in this portion of the integrated operation, posterior and other corneal deformations will have minimal effect on the procedure.
The novel features of this invention, as well as the invention itself, both as to its structure and its operation, will be best understood from the accompanying drawings, taken in conjunction with the accompanying description, in which similar reference characters refer to similar parts, and in which:
Referring initially to
For purposes of the present invention, the laser unit 12 will generate a so-called femtosecond laser that is capable of performing Laser Induced Optical Breakdown (LIOB) on selected tissues inside the eye 26 of the patient 24. Further, the imaging unit 18 is preferably of a type that is capable of creating three dimensional images of different tissues inside the eye 26 (e.g. an Optical Coherence Tomography (OCT) device). As envisioned for the system 10, the computer 16 will use input from the imaging unit 18 in its control of the laser unit 12 during laser ophthalmic surgeries in the eye 26.
In
It will be appreciated by the skilled artisan that all tissues in the eye 26, and specifically including those tissues mentioned above, will have their own unique response to LIOB. Also, depending on the location of the particular tissue in the eye 26 (i.e. the posterior depth of the tissue in the eye 26), the laser beam 22 can be operationally affected. Moreover, the operational requirements for particular surgical procedures will differ from one procedure to another. With this in mind, it is clear there is a need to customize the system 10 in a way that will account for all operational requirements (functionalities) each time the system 10 is to be used.
Inquiry block 58 in
As a final consideration for the set-up of an operational mode for the system 10, inquiry block 64 requires identifying a particular scanning procedure for the focal point of the laser beam 22. In essence, this scanning procedure will establish a path for the focal point of laser beam 22, and define how it moves along this path. If a preset scanning procedure is not to be used as a default, set-up block 66 requires that settings be made for the rate of focal point movement, and focal point spacing on the focal point path. Additionally, calibration requirements can be attended to.
For a final check on the set up of system 10, action block 68 requires a verification of the operational mode before the particular surgical procedure is to be performed (see action block 70). As a practical matter, this verification requires confirmation as to whether the default status is to be used. If the default status is not to be used, the user/operator has the opportunity at this time to recheck and verify that the system 10 has been properly customized for the particular surgical procedure.
In accordance with the present invention, the system 10 will be effectively controlled by the computer program product 16 during a surgical procedure. The computer program product 16, however, will also prompt the user/operator during the set-up of an operational mode, as has been discussed above with reference to
In the event a default operational mode is not to be followed by the system 10, action block 84 of the computer program 72 requires that a configuration for the laser beam 22 be specified. Once the laser beam 22 has been properly configured (inquiry block 86), the computer program 72 then requires a base reference datum be established (action block 88 and inquiry block 90). Next, a scanning plan is identified (action block 92). After all of this, inquiry block 94 then effectively determines whether an operational mode has been established for the computer program 72. If so, the computer program 72 moves to action block 82 and, thereafter, the computer 16 controls the laser unit 12 in its customized operational mode for, performance of the selected surgical procedure.
In accordance with the present invention, a computer program product 72 for performing ophthalmic laser surgery on selected tissue in an eye 26 is provided wherein the computer program product 72 comprises program sections for respectively: incorporating a laser unit 12 for generating a laser beam 22; defining an operational mode for the system 10; specifying, value ranges for configuration parameters for the laser beam 22 for implementation of the operational mode; establishing a base reference datum in the eye 26 for implementation of the operational mode; identifying a scanning procedure for the focal point of the laser beam 22 for implementation of the operational mode; and focusing the laser beam 22 to a focal point, to implement the operational mode by performing Laser Induced Optical Breakdown (LIOB) on the selected tissue. Additionally, the computer program product 72 can include program sections for: choosing a surgical procedure; and selecting a patient interface 28 from a group comprising a planar contact lens, a curved contact lens, a conformable contact lens, a fluid interface, a suction element, or air (no patient interface).
While the particular Surgical Laser Unit with Variable Modes of Operation as herein shown and disclosed in detail is fully capable of obtaining the objects and providing the advantages herein before stated, it is to be understood that it is merely illustrative of the presently preferred embodiments of the invention and that no limitations are intended to the details of construction or design herein shown other than as described in the appended claims.
Number | Name | Date | Kind |
---|---|---|---|
3169459 | Friedberg | Feb 1965 | A |
4169664 | Bailey, Jr. | Oct 1979 | A |
4573778 | Shapiro | Mar 1986 | A |
4846172 | Berlin | Jul 1989 | A |
4881808 | Bille | Nov 1989 | A |
4891043 | Zeimer et al. | Jan 1990 | A |
4907586 | Bille et al. | Mar 1990 | A |
4995715 | Cohen | Feb 1991 | A |
5098426 | Sklar et al. | Mar 1992 | A |
5246435 | Bille | Sep 1993 | A |
5439462 | Bille | Aug 1995 | A |
5480396 | Simon et al. | Jan 1996 | A |
5720894 | Neev et al. | Feb 1998 | A |
5779696 | Berry et al. | Jul 1998 | A |
5984916 | Lai | Nov 1999 | A |
5993438 | Juhasz et al. | Nov 1999 | A |
6004314 | Wei et al. | Dec 1999 | A |
6010497 | Tang et al. | Jan 2000 | A |
6059772 | Hsia et al. | May 2000 | A |
6099522 | Knopp et al. | Aug 2000 | A |
6322556 | Gwon et al. | Nov 2001 | B1 |
6391020 | Kurtz et al. | May 2002 | B1 |
6454761 | Freedman | Sep 2002 | B1 |
6467906 | Alpins | Oct 2002 | B1 |
6485413 | Boppart et al. | Nov 2002 | B1 |
6514241 | Hsia et al. | Feb 2003 | B1 |
6579282 | Bille | Jun 2003 | B2 |
6966905 | Bille | Nov 2005 | B2 |
7027233 | Goldstein et al. | Apr 2006 | B2 |
7232436 | Bille | Jun 2007 | B2 |
7458380 | Jones et al. | Dec 2008 | B2 |
7618415 | Kessler et al. | Nov 2009 | B2 |
7655002 | Myers | Feb 2010 | B2 |
7699467 | Dick et al. | Apr 2010 | B2 |
7703923 | Bille | Apr 2010 | B2 |
7731362 | Gerlach | Jun 2010 | B2 |
7789910 | Knox et al. | Sep 2010 | B2 |
7800760 | Bille | Sep 2010 | B2 |
7950398 | Schroeder et al. | May 2011 | B2 |
8088124 | Loesel et al. | Jan 2012 | B2 |
8231221 | Donitzky et al. | Jul 2012 | B2 |
20020173778 | Knopp et al. | Nov 2002 | A1 |
20020198516 | Knopp et al. | Dec 2002 | A1 |
20030074150 | Goldstein et al. | Apr 2003 | A1 |
20040199149 | Myers et al. | Oct 2004 | A1 |
20050165387 | Lubatschowski et al. | Jul 2005 | A1 |
20050192562 | Loesel et al. | Sep 2005 | A1 |
20060195076 | Blumenkranz et al. | Aug 2006 | A1 |
20080009922 | Bille | Jan 2008 | A1 |
20080025351 | Loesel | Jan 2008 | A1 |
20080177256 | Loesel et al. | Jul 2008 | A1 |
20080281303 | Culbertson et al. | Nov 2008 | A1 |
20090069794 | Kurtz | Mar 2009 | A1 |
20090137993 | Kurtz | May 2009 | A1 |
20090143772 | Kurtz | Jun 2009 | A1 |
20090149841 | Kurtz | Jun 2009 | A1 |
20100004643 | Frey et al. | Jan 2010 | A1 |
20100022994 | Frey et al. | Jan 2010 | A1 |
20100022995 | Frey et al. | Jan 2010 | A1 |
20100022996 | Frey et al. | Jan 2010 | A1 |
20100076417 | Suckewer et al. | Mar 2010 | A1 |
20100082017 | Zickler et al. | Apr 2010 | A1 |
20100114079 | Myers et al. | May 2010 | A1 |
20100191230 | Dick et al. | Jul 2010 | A1 |
20100290007 | Van de Velde | Nov 2010 | A1 |
20100324542 | Kurtz | Dec 2010 | A1 |
20100324543 | Kurtz et al. | Dec 2010 | A1 |
20110040293 | Bor | Feb 2011 | A1 |
20110040376 | Christie et al. | Feb 2011 | A1 |
20110118712 | Lubatschowski et al. | May 2011 | A1 |
20110144628 | Vogler | Jun 2011 | A1 |
20110160710 | Frey et al. | Jun 2011 | A1 |
20110166557 | Naranjo-Tackman et al. | Jul 2011 | A1 |
20110190741 | Deisinger et al. | Aug 2011 | A1 |
20110202044 | Goldshleger et al. | Aug 2011 | A1 |
20110202046 | Angeley et al. | Aug 2011 | A1 |
20110319875 | Loesel et al. | Dec 2011 | A1 |
20120078240 | Spooner | Mar 2012 | A1 |
20120078241 | Gooding | Mar 2012 | A1 |
20120150160 | Vogler et al. | Jun 2012 | A1 |
20120172854 | Raymond et al. | Jul 2012 | A1 |
Number | Date | Country |
---|---|---|
0765648 | Apr 1997 | EP |
0798987 | Feb 2001 | EP |
1252872 | Feb 2002 | EP |
WO2008112292 | Sep 2008 | WO |
WO2010076799 | Jul 2010 | WO |
2010129916 | Nov 2010 | WO |
WO 2011011202 | Jan 2011 | WO |
2012038011 | Mar 2012 | WO |
WO2013053367 | Apr 2013 | WO |
Entry |
---|
Sacks et al., High Precision Subsurface Photodisruption in Human Sclera,Journal of Biomedical Optics, Jul. 2002, pp. 442-450, vol. 7, No. 3. |
Chai et al., Aqueous Humor Outflow Effects of Partial Thickness Channel Created by a Femtosecond Laser in Ex Vivo Human Eyes, Optical Interactions with Tissue and Cells XVIII, 2007, pp. 1-8, Proc of SPIE, vol. 6435. |
PCT International Search Report, Application No. PCT/IB2013/000681, Apr. 15, 2013. |
Number | Date | Country | |
---|---|---|---|
20130281991 A1 | Oct 2013 | US |