Pulsed light provided by sources such as flashlamps and lasers have been used for several years for the removal of unwanted hair, and for treatments such as removal of pigmentation or vascular lesions.
Pulsed light sources may act through a number of mechanisms, including photomechanical disruption, photothermal ablation, photochemical ablation and photothermal coagulation/denaturation. Some of these light sources act by purely non-thermal means (i.e. mode-locked or Q-switched short pulse lasers) and others use a thermal effect to produce ablation or disruption of tissue. The most widely used and successful light sources for depilation or removal of pigmented and vascular lesions employ a photothermal interaction in which the mechanism of action is pulsed light used to locally heat the lesion or hair follicle with one or more closely spaced pulses in order to raise the temperature of the targeted tissue during or immediately following a single pulse to a temperature which will cause the death, damage, or reduced function, or reduced viability of the targeted tissue. The higher absorption of the target tissue will elevate its temperature with respect to the surrounding tissue, resulting in selective thermal destruction of the lesion or follicle, while the surrounding tissue will not be raised to a temperature which causes thermal destruction. Prior devices relied on relatively high power pulses of light (electromagnetic radiation) from a light source, such as a laser or a flashlamp. For example, typical prior devices provide energy pulses in the range of 10-40 J/cm2 for a treatment provided to a selected area. Some of these prior devices provide for supplying a single pulse providing the treatment energy. Other prior devices utilize a series of closely spaced lower energy pulses, where the frequency of the pulses may be in the range of 10 hz or higher, which corresponds to a time between pulses of 100 ms or less.
One issue with present systems is that because a relatively large amount of fluence and power applied to the skin, an incorrect setting in terms of the fluence or time of treatment can cause unwanted results, and in some cases can result in scarring, hypopigmentation or hyperpigmentation. For example, consider a situation in which a fluence of 40 J/cm2 is delivered to a treatment area during a 10 ms pulse width, the peak intensity for this treatment would be 4,000 W/cm2. Because the treatment time is so short, there is a possibility that a patient would receive too much energy in too short of a time period which could result in a treatment area receiving an excess amount of energy and might cause unwanted damage to the tissue. Further, because the treatment energy is applied over such a short time period, the patient's natural pain reflex may not be fast enough to allow a patient to discontinue the treatment, prior to severe overheating of the tissue. Thus, these prior systems typically require trained and qualified medical personnel to apply the treatment. Such devices are therefore not safely or reasonably used directly by a patient or consumer.
Another issue with present systems is that they require relatively complex and expensive power supplies. This is because the appropriate energy must be delivered over a very short time, necessitating a highly rated high voltage power supply. In the past, the complexity of the power supply required that the power supply be housed in a module which was separated from the treatment device housing the flashlamp that provides light energy to the treatment area.
Additionally, because of the high treatment power used in prior systems, these systems must provide for a relatively sophisticated cooling system for keeping the temperature of critical system components sufficiently low, so that the energy being dissipated over a short time period will not cause the application device to become overheated.
The systems and methods herein provide a different approach for using light to provide dermatologic treatments. Instead of using one pulse, or a series of closely spaced pulses, to provide a high amount of energy to a treatment area over a short amount time, the system and method herein deliver an amount of energy necessary for a treatment over a relatively long time period. As with prior systems, the system herein can be used to treat unwanted hair follicles, and to treat pigmentation and vascular lesions.
The system herein provides for a series of relatively low energy pulses, which provides a lower rate of temperature rise of target tissue than in prior systems. It has been recognized by the inventors herein that this lower rate temperature rise can provide an effective method for treatment of follicles and pigment. In fact, the temperature rise of targeted tissue by low energy pulse exposures can fall into at least three classes:
The system and method use an approach of stacking or delivering multiple low power pulses on a treatment area over a relatively long period of time. The total fluence required to reach the desired treatment effect on the tissue may be larger than in existing methods. However, the economy and reduction in system complexity resulting from the delivery of many smaller, lower fluence pulses, rather than a smaller number of larger fluence pulses, is significant. In addition, a lower fluence, multipulse treatment may be a less painful treatment than single pulse treatments.
In some prior small air cooled systems a major cost is in the circuitry and components necessary to provide electrical pulses to drive the flashlamp to provide relatively high energy per pulse (12-60 j/cm2). Utilizing a method which provides for a treatment with lower energy pulses (1-6 j/cm2) allows for the development of significantly different systems for providing thermal treatments. One area where significant cost savings can be realized is in the capacitor bank that is used in the high voltage power supply of the system, and in other components of the power supply.
The present system also utilizes longer time intervals spacing between pulses which further allows for providing a power supply with reduced demands on the capacitor bank which is discharged through the flashlamp. An embodiment of the present system provides a power in a range from 0.5 to tens of watts of average power per square cm of treatment spot size. This contrasts with some prior systems which provide for power in the range of hundereds to thousands of watts per square cm of treatment area. In one embodiment of the present system pulses are separated by at least 0.1 seconds which allows for unassisted cooling or thermal relaxation of the skin between the successive pulses. In one embodiment treatments will provide for between 2 to 20 pulses per spot.
Another aspect of an embodiment of the invention is the contact detector. This is a switch which detects the contact of the treatment window with the skin. If the window is not in contact with the skin a treatment pulse will not be delivered. The contact switch may be a mechanical switch or a capacitive sensing switch or a switch that detects the resistance change when electrodes are applied to skin. Another implementation may use an optical detector to sense the presence of skin at the treatment aperture. This will result in enhanced eye safety where the device by only allowing firing when the window is contacting a surface.
Another aspect of an embodiment of the invention is the pulse formatting. The first pulse generated by the light flashlamp can be a safety pulse which is smaller than the following pulses. This smaller initial pulse will be safer for the eyes than one large pulse, or series of equally sized pulses. The spacing from the initial pulse to the second pulse can be about 300 ms to allow the blink reflexes (about 250 ms) to protect the eyes from successive pulses.
The system and method provide for multiple pulses of light spaced apart over significant time periods. This allows for a relatively small and inexpensive power supply as compared with prior power supplies used in connection with driving light sources for providing for light treatment of skin. In the system herein the charging circuit has plenty of time to get the capacitive storage element charged back up without complexity and without heating the charging components, and because the discharging circuit and the lamp itself do not have to be aggressively cooled.
Generally light therapy device manufacturers have not offered pulsed light based devices with broadband or blue light due to the complications it can cause. At high peak power and pulse energies, the strong absorption in hemoglobin and melanin can result in large temperature rises of target and associated tissue. Targeting appropriate tissue with broadband or blue pulsed light is made more difficult due to the fact that short wavelength light penetrates tissue shallowly. The result is that epidermal damage is produced with such sources even in the most lightly pigmented skin for single pulse treatments. By multiple pulsing at relatively low fluence/low peak powers and long interpulse spacings, and by using a large spot size, epidermal damage can be minimized and the treatment benefits of broad band or the blue end of pulsed broadband light can be realized. One benefit may be that the follicular stem cells responsible for generating hair are shallowly located in the follicle and can be effectively reached by said broadband pulsed light or even narrow band blue pulsed light during all stages of follicle growth. Epidermal cooling can also increase the safety margin of this device.
Applications which may benefit from pulsed broadband light having appreciable spectral power below 600 nm and above 400 nm include temporary hair removal, hair stunting, delayed hair growth, epidermal pigment removal, temporary or permanent removal or damage of sebaceous glands, the treatment of inflammatory acne vulgaris, reduction of sweat glands or sweat gland function, and the general destruction of follicular stem cells. Such pulsed broadband sources may operate from 400-1200 nm, which includes the 400-550 nm part of flashlamp spectral output generally avoided or filtered out, for the reasons outlined above. Some applications may in fact exclusively use this part of the spectrum, as it is clear from demonstration clinical experiments that this part of the spectrum plays a previously unknown or misunderstood role in dermatologic treatments, particularly in temporary hair reduction. One experiment was performed to study the effect of the blue-green part of the pulsed lamp spectrum. In this experiment, a light source was used to generate low fluence, short pulse operation parameters as described herein. Treatments performed with sub-millisecond pulse durations (in the range of 0.3 ms to 0.5 ms were used), and fluences between 1.7 and 4.0 J/cm2 with multiple pulses (for the 1.7 J/cm^2, the number of pulses was 60, for the 2.5 J/cm^2, the number was 30, and for 4.0 J/cm^2, it was 15), widely spaced in time (in the range of 10 seconds was used) demonstrated that filtering out the spectral band between 350 and 590 nm (while maintaining the overall fluence) results in a greatly reduced efficacy for temporary hair reduction.
Short wavelength (e.g. in the range of 400-600 nm) or broadband (e.g. in the range of 400-1200 nm) hair reduction acts differently than longer wavelength hair reduction (e.g. in the range of 600-1200 nm). These treatments appear to deactivate all the follicles in the treatment area, not just the follicles in a particular hair-growth phase as is the case with longer wavelength devices. Complete and long lasting hair reduction or removal is possible in a single treatment on some individuals. In some individuals this fact allows users to shave only once—just before the first treatment—and prevent the regrowth of hair through maintenance treatments. In two individuals undergoing weekly maintenance treatments, near 100% hair clearance has been observed in excess of 6 weeks from the initial treatment. These treatments are an example of a new method for hair or acne management. In the case of hair, the subject initially shaves, then on a weekly, or even daily basis, is treated or treats himself with low fluences and multiple exposures per treatment area. Using an embodiment of a system and method herein it has been demonstrated that treatments of 3 J/cm^2, with pulse width 0.5-10 ms, of broadband light with wavelengths between 400 and 1200 nm, applied to a particular site 2-10 shots per treatment, for example, can prevent hair regrowth without the need for any shaving beyond the initial session. In the case of acne vulgaris, an initial “knock-down” (e.g. an initial decrease in bacterial population, or decrease in severity or level of infection) of the affected areas may be necessary, followed by regular treatments which serve to maintain a reduced functionality for the sebaceous glands and the bacterial populations that reside in them.
Fluences as low as 0.1-1.0 J/cm2 may be effective for maintaining hair reduction levels, and for maintaining sebaceous gland reduced function for the treatment of acne and seborrhea. Experiments demonstrating near-100% maintenance of hair reduction on a weekly treatment basis suggest that low fluences (<3 J/cm2) using all of the visible pulsed light spectrum may be effective, and that further, treatments in which only the 400-600 nm part of the pulsed light spectrum is delivered to skin may allow sub-Joule/cm2 fluences to be efficaciously used.
A power cord 102 can be provided to supply AC power to the high power voltage of the system. The AC power can be supplied to a capacitor charging circuit 104. The capacitor charging circuit 104 can use any of a wide range of different designs which are widely known to those of skill in the art. The capacitor charging circuit 104 receives AC power from the power cord 102. The capacitor charging circuit 104 converts the input AC power to a higher voltage DC circuit appropriate for charging the capacitor 106 to a treatment voltage. In one embodiment the capacitor would be of a value in the range of 200-400 μF and would be charged to a treatment voltage of approximately 350 V. In one embodiment the capacitor would be a photoflash capacitor, similar to a capacitor used to drive a flashlamp for a camera, such capacitors are often aluminum electrolytic capacitors, but as one of skill in the art will appreciate a range of different capacitors, or banks of multiple capacitors could be used. The capacitor 106 would discharge a stored charge through the flashlamp 114, and this discharge would cause the flashlamp to output a pulse of light. The discharge of the capacitor 106 is initiated by a trigger 110 which provides an initial triggering event (i.e. supplies, electrical energy) which causes the gas in the flashlamp to become conductive, at which point a charge stored on the capacitor 106 is discharged through the flashlamp. The trigger could be a wire or conducting stripe mounted to the reflector 116, proximal to the lamp body, or it could be a conducting stripe on the lamp etc. As is known in the art a trigger for a flashlamp is in general some electrical device such as a coil, which generates a electromagnetic field in response to a triggering current. The trigger conductor is positioned in proximity to the flashlamp, such that the electromagnetic field energizes the gas in the flashlamp, and causes it to become conductive, such that the charge stored in the capacitor is discharged through the flashlamp. In one embodiment the flashlamp would have a length of about 2 cm, and a 1 cm arc length, and the diameter of the flashlamp would be in the range of 5 mm. The trigger will activate the flashlamp 114 when the capacitor is fully charged and when the window 118 is pressed against the treatment area.
Element 116 is a reflector which reflects light generated by the flashlamp through the window 118 and toward the skin which is pressed against the window 118. The reflector 116 can consist of a molded plastic material covered with a reflective coating, or alternatively may be a stamped sheet metal part covered with a reflective coating.
As shown in
The sensor device could be implemented in number of different ways. One embodiment would provide for a circuit, wherein when the window 118 is physically pressed against the area skin to be treated, the window 118 would be pressed more firmly against the housing 120. The depression of window 118 would close a circuit, wherein when the circuit is initially closed the trigger 110 would be activated, which would cause the gas in the flashlamp 114 to become conductive, and the charged capacitor 106 would be discharged through the flashlamp 114.
The pulse width of the charge discharged through the flashlamp 114, and the result pulse width of the light output by the flashlamp is determined by the characteristics of the flashlamp and the capacitor. The power supply and the trigger can also be coupled to a timing sequence circuit 122 wherein following the initial pulse of the flashlamp, in response to the sensing device determining that the window 118 has been pressed against the skin, the capacitor is charged and then discharged through the flashlamp a predetermined number of times, to apply a treatment to the area of skin adjacent to the window 118. If the sensing device 112 determines that the window 118 has been pulled away from the skin then the sequence of generating pulses of light will stop. In between treatments to different areas a moist sponge or washcloth can wiped on the window 118 to cool the device.
The system 100 of
In another embodiment, the lamp envelope material may be fabricated from more expensive quartz-based materials, such as cerium-doped quartz, in order to more selectively control the ultraviolet radiation which may otherwise pass to the treatment area.
Another advantage of the system 100 is that because its operation is relatively simple, the device does not need to have a microprocessor. The fact that the device is low average power, its operation is relatively simple, and simplicity of the treatment sensing circuitry all combine to provide a very safe device which does not require microprocessors and sophisticated timing circuits.
Another alternative embodiment of system 100 provide for using a 4 AA batteries to supply power to the capacitor charging circuitry, voltage provided by the batteries would then be converted to the voltage needed to charge the capacitor 106. Because of the low power of the device 100, the window 118 can be made of plastic rather than glass, as is the case in prior higher power systems.
A further embodiment of system 100 uses material for the window 118 which may act as a selective wavelength filter. In the case of a plastic window, said filtering may consist solely of blocking ultraviolet (UV) radiation from the lamp. Alternatively, the window may consist of partially absorbing material which acts to transmit only selected therapeutic spectral bands.
Another type of sensing device may be a capacitance sensor which detects the capacitance of the human body when the device is applied to the treatment area.
Another type of sensor would be an optical sensor that detects the presence of skin at the treatment window.
Another type of sensing device 212 can sense the stray AC electrical field which is present in close proximity to a person's skin, the microprocessor can monitor a signal from the sensing device and when it indicates that the stray AC field is consistent with the sensing device being in close proximity to the skin the microprocessor can fire the trigger. As described above in connection with system 100, the pulse width can be determined by the characteristics of the capacitor 206 and the flashlamp 214. In addition further control circuitry can be provided to further control the pulse width. For example a solid-state device, such as an IGBT, could be connected between one of the electrodes of the flashlamp and ground. The controller can control the opening or the closing of the solid state device, such that the length of time of the charge being discharged through the flashlamp can be controlled in part by the opening and closing of the solid state device. In order to enhance the safety of the system 200 a second controller could be provided which monitors the operation of the controller 222. If the controller 222 fails, then the second controller could either take over the operation of the system 200, and may in some embodiments cause the operation of system to stop operation and generate an error message on the user interface device 224.
In one embodiment of system 200 the area of skin treated (spot size) is about 1 cm2. The pulse width of the light generated by the device is in the range of 0.5 to 2 ms. The fluence per pulse is in the range of 4-6 j/cm2 and the total fluence applied to a treatment area is in the range of 20-60 j/cm2. This higher fluence corresponds to a higher power than provided by the system 100. The system 200 can provide for a wider range of treatments than provided by the system 100. The system 200 can provide treatments for permanent hair reduction or removal, reduction of pigmentation, larger area hair stunting such as the underarm area. Further it should be recognized that the systems herein may also provide for treatment of acne vulgaris and for target sebaceous glands that are associated with acne. It should also be recognized that the treatments provided could include either killing or in some cases stunting tissue growth, or causing to behave different fashion. Generally in the range of 3-6 j/cm2 the operation would provide for killing or damaging hair follicles, and would provide for reducing pigment. In the range 1-3 j/cm2 the effect would tend to be more towards arresting hair but very little pigment reduction effect.
The user interface device 224 provides a range of user control. The user interface can display when the system 200 is ready to apply light pulses, and indicate when the device is in standby mode. The user can also put the device into a ready mode or a standby mode. The user interface can also allow the user to apply a treatment by pressing a contact fire button. The user interface can allow the user to adjust the amount of power, or fluence per pulse. The user interface can also provide a beeper to provide information to the user. The user interface can also provide a fan control, and a power level indicator. The power level is such that in the system 200 the flashlamp 214 will be a quartz xenon lamp. In one embodiment the lamp input power will be 20-30 watts. The system 200 can also include a forced air cooling system. In the forced a cooling system a fan not shown will circulate air 226 around the flashlamp 214.
As shown above in systems 100 and 200 the high voltage power supply and the flashlamp are both contained in a relatively small housing of a handheld device. This is in contrast with many prior system which require much more complex and expensive power supply to drive the flashlamp to output much higher levels fluence.
Attached hereto as
In one embodiment, the microprocessor 222 described in connection with
In another embodiment which is not shown the system would provide a higher amount of power over a larger treatment area. This requires a larger capacitor to discharge a larger amount of energy through the flashlamp. As a result of the larger energy requirements the components of the power supply will be larger. Thus, in some embodiments it may be desirable to provide the power supply outside of the housing of the hand held device. This high power system can provide treatments for permanent hair reduction, pigment reduction and high speed large area hair stunting. The power supply can be a separate unit which is connected to hand piece with the flashlamp. The power supply would supply power to drive the flashlamp. Controls for the operation of the high power system could be provided on either the hand piece or on the power supply unit. The area of treatment of the high power system can be in the range of 3 cm2. The pulse width for can be in the range 0.1-5 ms. The fluence for this system can be 4-6 j/cm2 per pulse and 20-60 j/cm2 per treatment to a given area. A user interface can provide controls and information to the user such as ready, standby, contact fire, power level, beeper, fan, and power level indicator. A microprocessor/controller would be coupled to the user interface and other components of the system to control the overall operation of the system and to receive information from and provide information to a user. The flashlamp for the high power system could be a quartz xenon lamp. The lamp input power can be in the range of 50-80 watts. The window material for the high power system can be glass. In one embodiment high power system would be water-cooled. In another embodiment it could be air-cooled.
Another embodiment could use a thermally conductive potting material to conduct heat away from the lamp.
Another embodiment which could provide a variation for any of the above systems provides for a replaceable module for the flashlamp of the system. The replaceable module could be implemented in a number of ways. In one embodiment the window which is pressed again the skin would be able to be removed and the flashlamp could be removed from coupling devices which couple the electrodes of the flashlamp to connectors through which the charge from the electrode is received. In another variation the treatment window and the flashlamp could be part of one replacement module. The replacement module could be provided with a plug which is plugged into a receptacle in the handheld device. The charge from the capacitor is then provided through the plug to the electrodes of the flashlamp. One in skilled in the art would recognize that this replaceable module could be implemented in a number of other ways.
Testing has been done which shows that pulses of approximately 1 ms in the range of 5 j/cm2 at a frequency of in the range of 1 hz can provide effective treatment for removal of hair and pigments. In one test application of 5 J/cm2 pulses at 1 Hz and in the range of 5 pulses provided both hair removal and pigment treatment. Such testing suggests that different skin types can benefit from different types of filtering. For example for light skin very little filtering is advantageous. Testing has also shown that it is advantageous to build a thermal pedestal at the treatment area. This means it is beneficial to allow the heat to build up at the treatment area through an accumulation of pulses from the flashlamp. In some application it may also be possible to achieve significant vascular effects.
Treatments for acne vulgaris using low fluence, multiple exposure, broadband pulsed light have been described above. However, it may be useful to shift the spectral shape of the light source. Some evidence has been recently presented suggesting that yellow-green light from pulsed dye lasers used at sub-purpuric fluences (up to ˜3 J/cm2) may be useful in the treatment of acne either directly (“Efficacy of low-fluence pulsed laser in treatment of inflammatory acne vulgaris: a randomized controlled study”, E D Seaton et alia, poster session at the American Academy of Dermatology, March 2003, which is incorporated by reference herein), or. in conjunction with photosensitizing agents (“Treatment of acne vulgaris with laser-assisted photodynamic therapy”, M R Alexiades-Armenakas et alia, presentation at the American Society for Lasers in Dermatology and Medicine, April 2003, which is incorporated by reference herein). Operating low fluence broadband flashlamps for these treatments may allow for similar results with greatly reduced cost and complexity of the light-based device. It has been suggested that the specific waveband associated with pulsed dye laser treatments may be key for acne treatments. Wavelength shifting of the broadband output using fluorophores and phosphors can produce efficient yellow-green output in the 1 J/cm2 range, comparable with the PDL treatment parameters reported for acne treatment. An example material is 0.1-1.0% cerium-doped YAG crystals used as a filter. The 400-500 nm band emitted from the flashlamp would be converted to 500-650 nm light.
Still another application is the treatment of fine vasculature, diffuse redness and persistent erythema. The pulsed light bands from 400-600 nm are particularly important for these treatments since hemoglobin absorption is strong in this region. The short (millisecond) range pulses should be effective in targeting the small scale vascular structures associated with these conditions. Repeated small fluence exposure treatments will be required to produce the same clinical effects seen with large single pulse treatments using pulsed and filtered flashlamp light sources in prior devices.
There are a wide range of different possible embodiments for systems herein. For example one type of system could be provided for hair stunting treatments on small areas of skin. A simple inexpensive handheld device for this type of system would provide a spot size (treatment area) of about 1 cm2. The capacitor and flashlamp would provide a fixed pulse width of between around 0.5 to 2 ms. The fluence per pulse would be about 1-2 J/cm2, and a treatment application for an area would apply multiple pulses to provide an energy of about 4-10 J/cm2 per treatment. This system would provide a contact fire switch. The lamp type would be xenon filled pyrex envelop, and the lamp input power would be 4-5 watts. In this simple system no microprocessor would be required. The power source could provide a user with the option of using 4 AA alkaline batteries, or using a power adapter to receive 120 V AC power. The lamp cooling could be convection air or conduction, and water and/or ice could optionally be used. The treatment window could be cooled using a wet sponge or washcloth, or optionally ice. Given the low energy and low cost requirements of the system a plastic treatment window would be used.
Another system herein could be provided for hair stunting treatments on larger areas of skin, and for larger diameter follicles, and would provide a larger spot size than that described in the system above. For example, in one embodiment the spot size would be about 3 cm2. The capacitor and flashlamp would provide a fixed pulse width of between around 2 to 3 ms. The fluence per pulse would be about 1-2 J/cm2, and a treatment application for an area would apply multiple pulses to provide an energy of about 4-10 J/cm2 per treatment. This system would provide a contact fire switch. The lamp type would be pyrex envelope filled with a xenon gas, and the lamp input average power would be about 20 watts. In this simple system no microprocessor would be required. The power source could be either a power adapter to receive 120 V AC power, or an AC cord with power adapter built into the handpiece could be used. The lamp cooling could be forced air or conduction to potting material. The treatment window could be cooled using a wet sponge or washcloth, or optionally ice. Given the low energy and low cost of the system a plastic or glass treatment window would be used.
Another embodiment of a system herein provides for permanent hair reduction, pigment reduction and stunting growth of larger diameter hair. This system is a handheld device. The spot size is in the range of 1 cm2. The pulse width is in the range of 0.5-2.0 ms. The fluence per pulse would be about 4-6 J/cm2, and a treatment application for an area would apply multiple pulses to provide an energy of about 30-60 J/cm2 per treatment. Due to the higher power and possible broader range of applications, additional controls may be desired, such as user interface which indicates when the device is in a ready mode, or stand by mode. The power level output can be adjusted by varying the voltage on the capacitor, and a power level indicator can be provided. The lamp type is quartz envelope filled with a xenon gas. The lamp input power is 20-30 watts. This system would be provided with a main controller and back up controller to provide for enhanced safety of operation due to the issues associated with providing for variable power control, and possible clinical risks associate with higher energy pulses, particularly for darker skin types. The power supply can provide for using an ac power cord. The lamp cooling is provided using forced air. Window cooling is provided using wet sponge of towlette. The window material is preferably glass, but for lower power operation plastic could be used.
Another embodiment of a system herein provides for permanent hair reduction, pigment reduction and high speed large area hair stunting. This system, due to its higher power of operation, would likely be implemented using a handpiece which houses the flashlamp, and the power supply could be implemented in a separate console which would be connected to the handpiece and the flashlamp via a umbilical cable carrying control signals and power. The spot size for the system would be in the range of 3 cm2. The pulse width would be fixed somewhere in the range of 2-5 ms. The fluence per pulse would be about 4-6 J/cm2, and a treatment application for an area would appl multiple pulses to provide an energy of about 30-60 J/cm2 per treatment. Due to the higher power and possible broader range of applications, additional controls may be desired, such as a user interface which indicates when the device is in a ready mode, or in a stand by mode. The power level output can be adjusted by varying the voltage on the capacitor, and a power level indicator can be provided. The lamp type is quartz envelope filled with a xenon gas. The lamp input average power is 50-80 watts. This system would be provided with a main controller and back up controller to provide for enhanced safety of operation due to the issues associated with providing for variable power control, and possible risks associate with higher energy pulses. The power supply can be provided for using a ac power cord. The lamp cooling is provided using forced air, and window which is in contact with skin, and through which light is transmitted, can be cooled by application of an wet cloth, or antiseptic towlette. The window material is preferably glass.
In considering different implementations of the method and system herein it is constructive to consider a number of different combinations and operations which can be provided in different embodiments of such a system or method. Below is an illustrative, but non-exhaustive identification of some combinations, which could be utilized in accordance with the present system and method.
A Broad band 400-1200 nm multiple pulse light treatment using a light source with pulse spacing greater than 100 ms and less than 15 seconds between each pulse, where multiple pulses are applied to each treatment spot.
A Blue band 400 nm-550 nm light source for dermatologic treatment for applying light treatments to skin.
A dermatologic treatment device with lamp current waveshape determined primarily by capacitor and lamp characteristic.
A dermatologic treatment device with lamp current waveshape determined primarily by the impedance of the storage capacitor, lamp and an inductor (coil) circuit.
A dermatologic treatment device in which the energy for the treatment pulse is stored in a capacitor in a handpiece, which also holds a light source for applying the treatment.
A dermatologic treatment device with broadband fluence 400-1200 nm from 0.5-8 j/cm^2 in each pulse and delivering from 2-50 pulses per treatment spot.
A dermatologic treatment device with 400-600 nm fluence from 0.1-5 j/cm2 in each pulse and delivering from 2-50 pulses per treatment spot.
A dermatologic treatment device with capacitive contact sensor to prevent treatment when window is not in contact with the skin.
A dermatologic treatment device with multiple capacitive contact sensors.
A dermatologic treatment device with mechanical contact sensor to prevent treatment when window is not in contact with the skin.
A dermatologic treatment device with multiple mechanical contact sensors A dermatologic treatment device with resistive contact sensor to prevent treatment when window is not in contact with the skin.
A dermatologic treatment device with multiple resistive contact sensors A dermatologic treatment device with stray field contact sensor to prevent treatment when window is not in contact with the skin.
A dermatologic treatment device with multiple stray field contact sensors.
A dermatologic treatment device with contact sensors that must “break and make” between single site/multi-pulse treatments to ensure that the user moves to the next treatment site. This means that the contact sensors must first indicate contact with the skin prior to enabling the firing sequence of light pulses, and before the NEXT sequence can begin, the sensors must change state, indicating that the device is no longer in contact with the skin.
A multiple pulse dernatologic treatment device that emits a smaller initial light pulse in a treatment sequence
A multiple pulse dermatologic treatment device that emits a series of increasingly larger pulses as part of treatment sequence.
A multiple pulse dermatologic treatment device that has cooling cryo chilled contact window, gel, forced air, to protect the epidermal pigment during a treatment. The cryo can be a cryogen—liquid or gas used to cool an object. In one embodiment this can be a compressed/refrigerated liquid is sprayed on skin. As the liquid strikes the skin, it absorbs heat and cools the epidermis, and to some extent, the dermis.
A dermatologic treatment device with optical sensors to measure quantity or quality of reflected light from treatment area to determine the appropriate treatment parameters for that area.
A multiple pulse dermatologic treatment device with a lamp, reflector and a filter which cuts of radiation below 400 nm.
A multiple pulsed dermatologic treatment device with wavelength output from 400 nm-600 nm.
A multiple pulsed dermatologic treatment device with wavelength output from 400 nm-1200 nm.
A multiple pulsed dermatologic treatment device with wavelength output from 400 nm-500 nm.
A dermatologic treatment where a first pulse is small, and reflection is sensed by a detector, if the reflection is too small or too large then the following pulses in the sequence are inhibited. (providing eye safety).
A dermatologic treatment where a first pulse is small, and reflection is sensed by a detector, the following pulse or pulses are modified based upon the measured reflected light. (for safety and efficacy).
A dermatologic treatment device that consists of a lamp, reflector, filter, and that applies 2 or more pulses to the treatment area with pulse spacing and pulse width such that it causes multiple thermal excursions of the target tissue resulting in the destruction of the target by repeated temperature cycles.
While the above description provides details for a number of embodiments herein, it should be recognized that other possible modifications and variations are possible, and that other systems and methods not expressly described above are still within the scope of the present invention. Thus, the scope of the inventions are defined by the attached claims and their equivalents.
The present application claims benefit from U.S. Provisional Patent Application Ser. No. 60/456,848, Mar. 21, 2003, entitled SYSTEM AND METHOD FOR LOW AVERAGE DERMATOLOGIC LIGHT TREATMENT DEVICE, which is incorporated herein by reference. The present application also claims benefit from U.S. Provisional Patent Application Ser. No. 60/488,252, Jul. 18, 2003, entitled SYSTEM AND METHOD FOR LOW AVERAGE POWER DERMATOLOGIC LIGHT TREATMENT DEVICE, which is incorporated herein by reference.
Number | Name | Date | Kind |
---|---|---|---|
634113 | Riley | Mar 1950 | A |
2699771 | Rüttger-Pelli | Jan 1955 | A |
3327712 | Kaufman | Jun 1967 | A |
3538919 | Meyer | Nov 1970 | A |
3648706 | Holzer | Mar 1972 | A |
3693623 | Harte et al. | Sep 1972 | A |
3834391 | Block | Sep 1974 | A |
3867948 | Kellenbom | Feb 1975 | A |
3900034 | Katz et al. | Aug 1975 | A |
4020383 | Labadini et al. | Apr 1977 | A |
4022534 | Kishner | May 1977 | A |
4122853 | Smith | Oct 1978 | A |
4233493 | Nath | Nov 1980 | A |
4298005 | Mutzhas | Nov 1981 | A |
4388924 | Weissman et al. | Jun 1983 | A |
4461294 | Baron | Jul 1984 | A |
4505545 | Salia-Munoz | Mar 1985 | A |
4539987 | Nath et al. | Sep 1985 | A |
4608978 | Rohr | Sep 1986 | A |
4608990 | Elings | Sep 1986 | A |
4617926 | Sutton | Oct 1986 | A |
4658823 | Beddoe et al. | Apr 1987 | A |
4667658 | Guibert | May 1987 | A |
4686986 | Fenyö et al. | Aug 1987 | A |
4733660 | Itzkan | Mar 1988 | A |
4747660 | Nishioka et al. | May 1988 | A |
4757431 | Cross et al. | Jul 1988 | A |
4784135 | Blum et al. | Nov 1988 | A |
4813412 | Yamazaki et al. | Mar 1989 | A |
4819669 | Politzer | Apr 1989 | A |
4829262 | Furumoto | May 1989 | A |
4860172 | Schlager et al. | Aug 1989 | A |
4884568 | Hahn | Dec 1989 | A |
4917084 | Sinofsky | Apr 1990 | A |
4930504 | Diamantopoulos et al. | Jun 1990 | A |
4950880 | Hayner | Aug 1990 | A |
4976709 | Sand | Dec 1990 | A |
5000752 | Hoskin et al. | Mar 1991 | A |
5057104 | Chess | Oct 1991 | A |
5059192 | Zaias | Oct 1991 | A |
5139494 | Freiberg | Aug 1992 | A |
5161526 | Hellwing et al. | Nov 1992 | A |
5182857 | Simon | Feb 1993 | A |
5207671 | Franken et al. | May 1993 | A |
5217455 | Tan | Jun 1993 | A |
5226907 | Tankovich | Jul 1993 | A |
5258989 | Raven | Nov 1993 | A |
5259380 | Mendes et al. | Nov 1993 | A |
5282797 | Chess | Feb 1994 | A |
5290273 | Tan | Mar 1994 | A |
5304169 | Sand | Apr 1994 | A |
5304170 | Green | Apr 1994 | A |
5312395 | Tan et al. | May 1994 | A |
5320618 | Gustafsson | Jun 1994 | A |
5336217 | Buys et al. | Aug 1994 | A |
5337741 | Diamond | Aug 1994 | A |
5344418 | Ghaffari | Sep 1994 | A |
5344434 | Talmore | Sep 1994 | A |
5397327 | Koop et al. | Mar 1995 | A |
5405368 | Eckhouse | Apr 1995 | A |
5409479 | Dew et al. | Apr 1995 | A |
5425728 | Tankovich | Jun 1995 | A |
5441531 | Zarate et al. | Aug 1995 | A |
5458596 | Lax et al. | Oct 1995 | A |
5474549 | Ortiz et al. | Dec 1995 | A |
5486172 | Chess | Jan 1996 | A |
5511563 | Diamond | Apr 1996 | A |
5522813 | Trelles | Jun 1996 | A |
5527350 | Grove et al. | Jun 1996 | A |
5569979 | Scott et al. | Oct 1996 | A |
5572091 | Langer et al. | Nov 1996 | A |
5591157 | Hennings et al. | Jan 1997 | A |
5595568 | Anderson et al. | Jan 1997 | A |
5611795 | Slatkine et al. | Mar 1997 | A |
5620478 | Eckhouse | Apr 1997 | A |
5660836 | Knowlton | Aug 1997 | A |
5683380 | Eckhouse et al. | Nov 1997 | A |
5725522 | Sinofsky | Mar 1998 | A |
5735844 | Anderson et al. | Apr 1998 | A |
5755753 | Knowlton | May 1998 | A |
5769844 | Ghaffari | Jun 1998 | A |
5769878 | Kamei | Jun 1998 | A |
5782895 | Zarate et al. | Jul 1998 | A |
5807261 | Benaron et al. | Sep 1998 | A |
5810801 | Anderson et al. | Sep 1998 | A |
5814040 | Nelson et al. | Sep 1998 | A |
5820625 | Izawa et al. | Oct 1998 | A |
5830208 | Muller | Nov 1998 | A |
5843074 | Cocilovo | Dec 1998 | A |
5843143 | Whitehurst | Dec 1998 | A |
5885274 | Fullmer et al. | Mar 1999 | A |
5919219 | Knowlton | Jul 1999 | A |
5964749 | Eckhouse et al. | Oct 1999 | A |
5989283 | Wilkens | Nov 1999 | A |
6015404 | Altshuler et al. | Jan 2000 | A |
6050990 | Tankovich et al. | Apr 2000 | A |
6080146 | Altshuler et al. | Jun 2000 | A |
6080147 | Tobinick | Jun 2000 | A |
6096066 | Chen et al. | Aug 2000 | A |
6120497 | Anderson et al. | Sep 2000 | A |
6168590 | Neev | Jan 2001 | B1 |
6171332 | Whitehurst | Jan 2001 | B1 |
6228074 | Almeida | May 2001 | B1 |
6235015 | Mead, III et al. | May 2001 | B1 |
6241753 | Knowlton | Jun 2001 | B1 |
6273884 | Altshuler et al. | Aug 2001 | B1 |
6280438 | Eckhouse et al. | Aug 2001 | B1 |
6319273 | Chen et al. | Nov 2001 | B1 |
6374265 | Chen et al. | Apr 2002 | B1 |
6377855 | Knowlton | Apr 2002 | B1 |
6381498 | Knowlton | Apr 2002 | B1 |
6383176 | Connors et al. | May 2002 | B1 |
6387089 | Kreindel et al. | May 2002 | B1 |
6405090 | Knowlton | Jun 2002 | B1 |
6413253 | Koop et al. | Jul 2002 | B1 |
6443978 | Zharov | Sep 2002 | B1 |
6453202 | Knowlton | Sep 2002 | B1 |
6461866 | Whitehurst | Oct 2002 | B1 |
6482199 | Neev | Nov 2002 | B1 |
6485484 | Connors et al. | Nov 2002 | B1 |
6508813 | Altshuler | Jan 2003 | B1 |
6511475 | Altshuler et al. | Jan 2003 | B1 |
6517532 | Altshuler et al. | Feb 2003 | B1 |
6524329 | Benedict | Feb 2003 | B1 |
6558372 | Altshuler | May 2003 | B1 |
6569155 | Connors et al. | May 2003 | B1 |
6595986 | Almeida | Jul 2003 | B2 |
6602275 | Sullivan | Aug 2003 | B1 |
6605080 | Altshuler et al. | Aug 2003 | B1 |
6648904 | Altshuler et al. | Nov 2003 | B2 |
6653618 | Zenzie | Nov 2003 | B2 |
6663620 | Altshuler et al. | Dec 2003 | B2 |
6663659 | McDaniel | Dec 2003 | B2 |
6723090 | Altshuler et al. | Apr 2004 | B2 |
6730113 | Eckhardt et al. | May 2004 | B2 |
6743222 | Durkin et al. | Jun 2004 | B2 |
6749624 | Knowlton | Jun 2004 | B2 |
6981970 | Karni | Jan 2006 | B2 |
20010051766 | Gazdzinski | Dec 2001 | A1 |
20020055092 | Hochman | May 2002 | A1 |
20020091377 | Anderson et al. | Jul 2002 | A1 |
20020128635 | Altshuler et al. | Sep 2002 | A1 |
20020161357 | Anderson et al. | Oct 2002 | A1 |
20020173780 | Altshuler et al. | Nov 2002 | A1 |
20020198575 | Sullivan | Dec 2002 | A1 |
20030004499 | McDaniel | Jan 2003 | A1 |
20030004501 | Wilkens et al. | Jan 2003 | A1 |
20030004556 | McDaniel | Jan 2003 | A1 |
20030023283 | McDaniel | Jan 2003 | A1 |
20030032900 | Ella | Feb 2003 | A1 |
20030032950 | Altshuler et al. | Feb 2003 | A1 |
20030045916 | Anderson et al. | Mar 2003 | A1 |
20030055413 | Altshuler et al. | Mar 2003 | A1 |
20030055414 | Altshuler et al. | Mar 2003 | A1 |
20030057875 | Inochkin et al. | Mar 2003 | A1 |
20030065313 | Koop et al. | Apr 2003 | A1 |
20030065314 | Altshuler et al. | Apr 2003 | A1 |
20030125788 | Long | Jul 2003 | A1 |
20030130709 | Haber et al. | Jul 2003 | A1 |
20030195494 | Altshuler et al. | Oct 2003 | A1 |
20030199859 | Altshuler et al. | Oct 2003 | A1 |
20040010298 | Altshuler et al. | Jan 2004 | A1 |
20040024388 | Altshuler | Feb 2004 | A1 |
20040034319 | Anderson et al. | Feb 2004 | A1 |
20040092916 | Jay | May 2004 | A1 |
20040093042 | Altshuler et al. | May 2004 | A1 |
20040147985 | MacFarland et al. | Jul 2004 | A1 |
20040147986 | Baumgardner et al. | Jul 2004 | A1 |
20050085878 | Wilkens et al. | Apr 2005 | A1 |
Number | Date | Country |
---|---|---|
968854 | Jun 1975 | CA |
1041610 | Oct 1978 | CA |
33 47 730 | Jul 1985 | DE |
39 06 860 | Sep 1989 | DE |
0 565 331 | Apr 1993 | EP |
2 360 946 | Oct 2001 | GB |
4-98795 | Mar 1992 | JP |
WO 8602783 | May 1986 | WO |
WO 8900871 | Feb 1989 | WO |
WO 9515725 | Jun 1995 | WO |
WO 9622813 | Aug 1996 | WO |
WO 9737723 | Oct 1997 | WO |
WO 9824514 | Jun 1998 | WO |
WO 9838933 | Sep 1998 | WO |
WO 9851235 | Nov 1998 | WO |
WO 9907438 | Feb 1999 | WO |
WO 9911324 | Mar 1999 | WO |
WO 0054685 | Sep 2000 | WO |
WO 0054685 | Sep 2000 | WO |
Number | Date | Country | |
---|---|---|---|
20040230260 A1 | Nov 2004 | US |
Number | Date | Country | |
---|---|---|---|
60456848 | Mar 2003 | US | |
60488252 | Jul 2003 | US |