SYSTEM AND METHOD FOR MEASURING SULPHITE IN FOOD SAMPLES USING AN AMPEROMETRIC BIOSENSOR AND THE USE OF SAID BIOSENSOR

Abstract
A system for measuring sulphite in food samples is provided using a biosensor, a method for determining sulphite in food samples using the biosensor, and the biosensor for measuring the sulphite values in food samples. The system uses an amperometric biosensor having a bi-enzymatic system, a working electrode, a counter electrode and a reference electrode. The bi-enzymatic system has the human sulphite oxidase and peroxidase enzymes, and the immobilisation of the enzymes in the working electrode is carried out via simple physical retention. The mixture of the enzymes is mixed with a chemical mediator by applying a sample to the amperometric biosensor; changing the electrical current upon coming into contact with the analyte to be detected; and amperometrically measuring said electrical current in solutions with continuous agitation, using a constant potential.
Description
STATE OF THE ART

Sulphites are food additives with antioxidant and antimicrobial properties that are used in the food industry to improve quality, appearance and to extend the shelf life of foods and drinks. In crustaceans, sulphites are applied after the harvest and during the entire production process to prevent melanosis. The addition of sulphites inhibits the formation of black stains caused by the enzymatic oxidation of the monophenols to melanin. Melanosis does not affect the taste of crustaceans and is not harmful for consumers. Nevertheless, the aforementioned stains can drastically affect the acceptability of the products by the consumer and significantly decrease their market value.


The community legislation on food additives is based on the principle that only the additives that are explicitly authorized can be used. The majority of the food additives can only be used in limited quantities in specific food products and they are listed in Regulation No. 1129/2011, which modifies the CE no. 1333/2008. This Regulation lists the food additives with maximum levels allowed different from quantum satis. For fresh and frozen crustaceans and cephalopods the maximum concentration of sulphite (E220-E228) allowed is 150 mg/kg.


On the other hand, the Commission of the Codex Alimentarius names sulphites as priority allergens and indicates that the control of the use of sulphite is important to protect the consumers who are sensitive to sulphite.


In food and drinks, sulphite is bonded reversibly in the form of adducts with carbonyl compounds and hydroxy sulphonates. These adducts are stable to low and intermediate pH and disassociate completely free sulphite above pH 9. The free and reversible forms of sulphite bond are referred to as total sulphite. Both in free and total form, both are of interest for the food industry.


A critical control point in the processing of crustaceans is established to guarantee that the final product does not contain an excessive concentration of sulphite. The critical limits must be established in the entire process to ensure that the final product complies with the regulatory limits and the periodic supervision of the effectiveness of the treatment and the controls must be required. Therefore, the monitoring of sulphite must be an integral part of a program of the Firm of Hazard Analysis and Critical Control Points (HACCP).


The HACCP needs to be rapid and optimal to be applied daily, but the current methods for the determination of sulphite are complicated, are time-consuming and are subjective.


Said methods such as test strips, enzymatic methods or methods based on titration have great limitations, false positives, and little reproducibility and accuracy.


The optimized Monier-Williams method (AOAC 962.16) has been fully utilized to determine the total concentration of sulphites in food products. Nevertheless, this method involves much time in analysis, labour, it requires technical personnel and shows false positives.


Therefore, one of the great challenges of the industry of crustaceans is to achieve a system that is highly sensitive, fast and portable for the determination of sulphite in crustaceans in order to prevent contaminated products from reaching the market. Enzymatic biosensors are suitable solutions that will be applied for the determination of sulphites in food, more specifically in crustaceans, as simple, precise and fast cost tools.


OBJECT OF THE INVENTION

The object of the invention is a system for measuring sulphite in food samples, simply and quickly by the use of an enzymatic amperometric biosensor. This system is bi-enzymatic, composed of two enzymes: human sulphite oxidase (EC 1.3.8.1) and peroxidase, with simple physical retention on a conductive base, of a liquid aliquot of these enzymes with a dialysis membrane, together with a chemical mediator.


The enzymatic biosensor for the determination of sulphite in food samples is also object of the invention.


The method for the determination of sulphite in food samples using the aforementioned biosensor is also object of the invention.


The use of the system described to measure the values of sulphite in food samples is also object of the invention.


In particular, the food samples are crustaceans.


The resulting measurement device, which has shown satisfactory selectivity and stability to sulphite as a substrate in matrices such as crustaceans, is also object of the invention.


Today, a commercial biosensor of sulphite applied to the agro-food industry does not exist, and no commercial enzyme is available.


The system which is the object of the invention offers results which are as reliable, or more so, than the method of (Monier-Williams) used today, but with advantageous characteristics/particularities with respect to it which are reflected in summarized form in the following table:
















Optimized Monier-




Williams method
Object of the



(AOAC 962.16)
invention


















Procedure
Acid distillation
Enzymatic oxidation



followed by titration
of the sulphite,




amperometric detection


Time of analysis
1 hour
3 Minutes


Simplicity
Low
Medium


Interpretation
Comparison with the
Result on screen



standards at the start



of the titration


Limitation
Limited quantification
Daily calibration












DESCRIPTION OF THE DRAWINGS

To better understand the object of the invention, a preferential form of embodiment is represented in the attached figures, subject to accessory changes that do not essentially alter it. In this case:



FIG. 1 represents the reaction scheme for the determination of sulphite.



FIG. 2 represents the calibration curve to see the range of response of the biosensor to sulphite.





DETAILED DESCRIPTION OF THE INVENTION

This invention represents a system for measuring sulphite in food samples by biosensor; a method for the determination of sulphite in food samples using the aforementioned biosensor; and the use of the aforementioned biosensor to measure the values of sulphite in food samples.


In particular, the food samples are crustaceans.


The biosensor used is an amperometric biosensor that comprises a bi-enzymatic system, which is composed by two enzymes: humane sulphite oxidase (EC 1.3.8.1) and peroxidase.


The aforementioned amperometric biosensor is composed of a substrate, a working electrode, a counter electrode and a reference electrode. Preferably, the biosensor is composed of a gold working electrode, a stainless-steel counter electrode and of a silver/silver chloride (Ag/AgCl). reference electrode


The enzymes are deposited on the surface of the working electrode with simple physical retention. Before depositing them, on the surface of the working electrode the chemical mediator which can comprise mediators such as Methylene Blue, Meldola's Blue, Tetrathiafulvalene, Ferricyanide and Ferrocene, dissolved in organic solvents such as isopropyl alcohol, propanol, methanol and chloroform. The chemical mediator is preferably Ferrocene dissolved in isopropyl alcohol.


The method which is the object of the invention comprises the phases of


a) application of a sample to the amperometric biosensor;


b) change of electric current on entering into contact with the analyte to be detected; and


c) measuring amperometrically said electric current in solutions with continuous stirring, using a constant potential.


Before depositing the enzymes on the working electrode, these are dissolved in a carbonate/bicarbonate buffer and/or CAPS and/or CAPSO 0.1M in a pHs range of 8.0-11.1, the optimal pH being 10. For said preparation, sulphite oxidase:peroxidase ratios of 1:1, 1:2, 1:3 and 1:4 are worked with in terms of enzymatic units, the optimal ration being 1:3.


Said bi-enzymatic combination, in turn, comprises a drying solution composed of trehalose and/or monosodium glutamate in a range of 10-40%, the optimal value being 30%.


The enzymes deposited on the working electrode are covered with a permeable or osmotic membrane such as those used in dialysis, previously hydrated, with a nominal molecular weight limit in the range of 5-50 kilodaltons (kDa), the optimal range being 6-10 kDa.


Subsequently, for its use, the biosensor is hydrated in the measuring solution that corresponds to the carbonate/bicarbonate buffer 0.1M in a pHs range of 8.0-11.1, the optimal pH being 10.


The biocatalytic scheme for the determination of sulphite in crustaceans would be as follows:


At pH 10 all the sulphur dioxide present in the sample (both free and combined) is in the form of SO3 (sulphite form).


The average enzyme sulphite oxidase in the oxidation of the sulphite to sulphate generating hydrogen peroxide in presence of oxygen. The subsequent catalytic reduction of the peroxide generated by the peroxidase (HRP), is the indirect measure of the oxidized sulphite. The ferrocene is used as a chemical mediator for this last enzymatic reaction; therefore, the amperometric signal that is used for the reading of this scheme of reactions corresponds to the reduction of said mediator, as indicated in FIG. 1, which represents a reaction scheme for the determination of sulphite.


The level of sulphite in a sample is determined applying the sample to the biosensor and measuring the signal of current to a potential that is found in the range of 0 to 50 mV. Preferably, the voltage of the electric current used for the determination of sulphite is approximately 0 mV.


The pH for the determination of sulphite is in the range of 8.0 to 11.1. Preferably, the pH for the determination of sulphite is approximately 10.0.


Said biosensor together with a measuring cuvette and a stirrer is what we call an electrochemical cell which consists of the following elements:

    • Working electrode;
    • Counter electrode;
    • Reference electrode;
    • Measuring cuvette; and
    • Stirrer


The biosensor has a range of response of 50-300 ppm of sulphite; and in addition, the linear response is total in said range, as shown in FIG. 2, which represents the range of response of the biosensor (calibration curve of 50 to 300 mg/kg). Thus, the ranged marked by Regulation No. 1129/2011 is covered, which lists the food additives with maximum levels allowed for fresh and frozen crustaceans and cephalopods; said value being 150 ppm.


A calibration line is constructed of 5 points (not including the white), and of 3 replicas for each concentration. The linearity is evaluated, checking the regression coefficient r2 and the Response/Concentration Factor.


Once the linearity is verified in the range of 50 to 300 mg/kg, it is studied whether routine calibrations made by the biosensor (2 points) are stable enough to ensure accuracy and precision during the lifetime of the biosensor. For this purpose, 23 calibrations were made with 5 different biosensors, so that the stability of the slope (m), of the coefficient of linear regression (r2) was checked. The criteria of acceptance proposed are as follows:

    • The % RSD of the slope will not exceed 15
    • No straight line will have a r2<0.995 (R<0.99)


The results obtained are shown on the following table:




















SLOPE
CUT-OFF POINT



BIOSENSOR
r2
(m)
(b)





















Biosensor 1
0.9986
−268451.72
−9.19



Biosensor 1
0.9998
−216795.02
−2.08



Biosensor 1
0.9997
−212064.04
−3.3



Biosensor 1
0.9997
−213370.52
−3.38



Biosensor 1
0.9999
−216305.44
−1.96



Biosensor 2
0.9999
−252625.28
−1.21



Biosensor 2
0.9997
−238793.58
−3.21



Biosensor 2
0.9997
−260319.96
−3.89



Biosensor 2
0.9999
−160476.95
−0.16



Biosensor 2
0.9998
−162145.8
1.92



Biosensor 3
0.9994
−158693.02
−3.36



Biosensor 3
0.9998
−199188.24
−2.5



Biosensor 3
0.9999
−200175.32
−1.59



Biosensor 3
0.9987
−207896.84
−6.96



Biosensor 3
0.9994
−175763.08
3.96



Biosensor 4
0.9999
−165459.56
0.9



Biosensor 4
0.9999
−241312.66
−0.55



Biosensor 4
0.9999
−224968.14
1.14



Biosensor 4
0.9999
−225537.96
−0.47



Biosensor 5
0.9998
−254172.28
−2.75



Biosensor 5
0.9998
−240394.24
−3.07



Biosensor 5
0.9992
−226618.6
−5.68



Biosensor 5
0.9989
−217246.88
−6.48



AVERAGE
0.9996
−214729.35
−2.34



SD
0.0004
32738.83
3



RSD (%)
0.04
15




MAXIMUM
0.9999
−158693.02
3.96



MINIMUM
0.9986
−268451.72
−9.19










The method is considered precise in the entire range of work, from 50 to 300 mg/kg and in the matrices that are the test object (raw and cooked prawn), according to the criteria of RSD HORWITZ


During the lifetime of the biosensor, the stability of the straight lines of calibration is considered suitable.


The accuracy is the degree of concordance between the result of the test and an accepted reference value. Said accuracy is obtained by the study of recoveries, in the entire range of work, from 50 to 300 mg/kg, so that at each level of concentration, the two matrices of study are analysed (raw and cooked prawn). The method is considered of good accuracy if the average recovery at all the levels of concentration, meets the criteria established by AOAC. The following table shows the results:













PLANNING AND RESULTS
ACCURACY









REFERENCE

RECOVERY












VALUE

No.
% AVG.




(mg/kg)
MATRIX
Replicas
REC.
CRITERIA
Acceptance















50
RAW PRAWN
4
84
80-110
YES



COOKED PRAWN
4


100
RAW PRAWN
4
96
90-107
YES



COOKED PRAWN
4


200
RAW PRAWN
4
92
90-107
YES



COOKED PRAWN
4


300
RAW PRAWN
4
96
90-107
YES



COOKED PRAWN
4









The method is considered accurate in the entire range of concentration, from 50 to 300 mg/kg, in the matrices which are the test object (raw and cooked prawn), according to the acceptable intervals of recovery established by AOAC.


On the other hand, the degree of concordance is determined between independent test results obtained under predetermined conditions. Said precision is usually expressed as imprecision by the calculation of relative standard deviations and/or the Horrat ratio. To obtain the maximum imprecision at all the levels of concentration, the maximum variability of matrices (species, fresh and processed products, . . . ), measured throughout the life of the biosensor, etc. has been considered


The following table shows the results:


















Modified
Interval of





Monier-
uncertainty Modified


Sample
Biosensor
Williams
Monier-Williams
Acceptance



















1
180.9
214
166-262
Yes


2
195.5
224
176-272
Yes


3
216
236
184-288
Yes


4
104.76
115
 90-140
Yes


5
112.06
128
 100- 156
Yes


6
137.06
127
 99-155
Yes


7
131.5
117
 92-142
Yes


8
183.82
194
150-238
Yes


9
81
96
 74-118
Yes


10
68
81
 62-100
Yes


11
156
165
130-200
Yes


12
155
190
146-234
Yes


13
97
121
 96-146
Yes


14
121
143
112-171
Yes


15
120
131
103-159
Yes


16
226
222
174-270
Yes


17
181
241
189-293
No


18
96
122
 94-150
Yes


19
155
176
137-215
Yes


20
117
163
128-198
No


21
127
162
127-197
Yes


22
165.96
202
158-246
Yes


23
124.16
154
119-189
Yes


24
117.16
130
102-158
Yes


25
107
135
107-163
Yes


26
60.28
55
42-68
Yes


27
75.28
67
51-83
Yes


28
75.18
92
 73-111
Yes


29
294.8
337
265-409
Yes


30
86.12
84
 65-103
Yes


31
58.16
63
50-76
Yes


32
61.02
75
59-91
Yes


33
88.02
102
 80-124
Yes


34
22.34
21
16-26
Yes


35
41.58
46
35-57
Yes


36
76.1
84
 65-103
Yes


37
132.2
166
131-201
Yes


38
172.54
206
162-250
Yes


39
70.6
90
 71-109
Yes


40
72.52
84
 65-103
Yes


41
65.26
75
59-91
Yes


42
64.6
78
62-94
Yes


43
79.84
88
 69-107
Yes


44
70
64
51-77
Yes









The method is considered precise in the whole range of work, from 50 to 300 mg/kg and in the matrices which are the test object (raw and cooked prawn), according to the criteria of RSD HORWITZ.


To ensure that the results obtained are suitable, a comparative study was performed with the modified Monier Williams method for determination of sulphites in samples in which other volatile sulphur compounds are present. It is based on the transformation of the sulphites in SO2. The SO2 is oxidized to H2SO4 which finally is titrated with a standardized dissolution of NaOH. The modified Monier Williams method is based on the method of the AOAC, Sulphurous Acid (Total) in Food, Optimized Monier-Williams Method. 990.28, pp 29, 1995. This method is used frequently in the official control of sulphites in food.


For this purpose, a total of 44 samples of cooked and raw prawn were analysed both by the modified Monier-Williams method and by the method which is the object of the invention. The results obtained are shown in the foregoing table.


The comparative study with the modified Monier-Williams method, shows a reliability value of 95%.


Lastly, the lifetime of the biosensor is determined in relation to its storage and number of measures it can support. For this purpose, a series of biosensors are tested with 0, 15, 30, 60 and 90 days of drying stored at 3-8° C., and hydrated on day 1, 7 and 15 after its drying. The results are shown on the following table:

















H1
H7
H15



















T0











batch
Biosensor 1
Biosensor 1
Biosensor 1


calibrat
R = 0.9996
R = 0.9999
R = 0.9999


no. of cumulative
110
105
108


measures










T15











batch
Biosensor 2
Biosensor 2
Biosensor 2


calibration
R = 0.9999
R = 0.9999
R = 0.9998


no. of cumulative
107
110
104


measures










T30











batch
Biosensor 3
Biosensor 3
Biosensor 3


calibration
R = 0.9999
R = 0.9999
R = 0.9998


no. of cumulative
104
110
120


measures










T60











batch
Biosensor 4
Biosensor 4
Biosensor 4


calibration
R = 0.9993
R = 0.9999
R = 0.9993


no. of cumulative
105
105
110


measures










T90











batch
Biosensor 2
Biosensor 2
Biosensor 2


calibration
R = 0.9999
R = 0.9999
R = 0.9999


no. of cumulative
115
105
108


measures





where:


T0 = 0 days of dry storage


T15 = 15 days of dry storage


T30 = 30 days of dry storage


T60 = 60 days of dry storage


T90 = 90 days of dry storage


H1 = 1 day hydrated


H7 = 7 days hydrated


H15 = 15 days hydrated






The biosensors are able to support a minimum of 100 measures in any of the aforementioned conditions showing perfect calibration.


Examples of Preferential Embodiment

For a better understanding of this invention, the following examples are shown, described in detail, that must be understood non-limiting of the scope of this invention.


Example 1. Determination of Total Sulphite Total in Raw Prawn

Clean and discard the shell, the cephalothorax and other non-edible parts of the exoskeleton and visible digestive tract. Then, homogenise the sample with the aid of a mincer. Weigh 2 gr of the homogenised sample and mix it with 18 ml of extraction solution (buffer 0.1M CAPS pH 9.5). Stir vigorously 30 seconds and apply high-intensity ultrasound for 20 minutes with 20-second pulses and an amplitude of 60%; allow to reach room temperature before proceeding to analysis.


Add 1 mL of the homogenised sample to the electrochemical cell with 10 mL of buffer 0.1M CAPS pH 9.5, with constant stirring and start the amperometric measurement in the biosensor at a potential of 0 mV.


Example 2. Determination of Total Sulphite in Cooked Prawn

Clean and discard the shell, the cephalothorax and other non-edible parts of the exoskeleton and visible digestive tract. Then, homogenise the sample with the aid of a mincer. Weigh 5 gr of the homogenised sample and mix it with 40 ml of extraction solution (buffer 0.1M CAPSO pH 10.0). Stir vigorously 30 seconds and apply low-intensity ultrasound for 30 minutes at 45° C.; allow to reach room temperature before proceeding to analysis.


Add 1 mL of the homogenised sample to the electrochemical cell with 10 mL of buffer 0.1M CAPSO pH 10.0, with constant stirring and start the amperometric measurement in the biosensor at a potential of 0 mV.


Example 3. Determination of Total Sulphite in Raw Prawn

Clean and discard the shell, the cephalothorax and other non-edible parts of the exoskeleton and visible digestive tract. Then, homogenise the sample with the aid of a mincer. Weigh 2 gr of the homogenised sample and mix it with 10 ml of extraction solution (buffer 0.1M carbonate/bicarbonate pH 10.0). Stir vigorously 30 seconds and apply low-intensity ultrasound for 20 minutes at 40° C.; allow to reach room temperature before proceeding to analysis. To eliminate the traces of sample in suspension, proceed to filter/centrifuge the sample before injecting it in the equipment.


Add 1 mL of the homogenised sample to the electrochemical cell con 10 mL of buffer 0.1M carbonate/bicarbonate pH 10.0, with constant stirring and start the amperometric measurement in the biosensor at a potential of 0 mV.

Claims
  • 1. System to measure sulphite in food samples by biosensor comprising an amperometric biosensor that comprises: a) a bi-enzymatic system;b) a working electrode;c) a counter electrode; andd) a reference electrode.
  • 2. The biosensor of claim 1, wherein the working electrode is gold.
  • 3. The biosensor of claim 1, wherein the counter electrode is stainless steel.
  • 4. The biosensor of claim 1, wherein the reference electrode is a silver chloride (Ag/AgCI) electrode.
  • 5. The biosensor of claim 1, wherein the bi-enzymatic system is composed of the enzymes human sulphite oxidase and peroxidase.
  • 6. The biosensor of claim 5, wherein the immobilisation of the enzymes in the working electrode is carried out by simple physical retention.
  • 7. The biosensor of claim 5, wherein the mixture of the enzymes is in turn, mixed with a chemical mediator.
  • 8. The biosensor of claim 7, wherein the chemical mediator is ferrocene dissolved in isopropyl alcohol.
  • 9. Method for the determination of sulphite in food samples using the amperometric biosensor of claim 1, comprising the phases of: applying a sample to the biosensor amperometric;changing electric current on entering into contact with the analyte to be detected; andamperometrically measuring said electric current in solutions with continuous stirring, using a constant potential.
  • 10. The method of claim 9, wherein the voltage of the electric current used for the determination of sulphite is in the range of 0 to 50 mV.
  • 11. The method of claim 10, wherein the voltage of the electric current used for the determination of sulphite is approximately 0 mV.
  • 12. The method of claim 9, wherein a pH for the determination of sulphite is in the range of 8.0 a 11.1.
  • 13. The method of claim 12, wherein the pH for the determination of sulphite is approximately 10.0.
  • 14. Use of the biosensor of claim 1 to measure the values of sulphite in food samples.
Priority Claims (1)
Number Date Country Kind
P201530995 Jul 2015 ES national
PCT Information
Filing Document Filing Date Country Kind
PCT/ES2016/070279 4/19/2016 WO 00