The present disclosure generally relates to systems and methods for radiation therapy. More particularly, the present disclosure generally relates to systems and methods for designing radiation treatment plans for radiation therapy.
Volumetric modulated arc therapy (“VMAT”) is becoming a widely adopted radiation treatment technique due to its ability to achieve highly conformal dose distributions when treating patients using radiation. The theoretical framework of VMAT is based on an earlier technique known as intensity modulated arc therapy (“IMAT”). In IMAT, radiation dose is delivered inside a targeted volume by directing radiation from a number of angles along a given arc, and using various aperture configurations at each beam angle along the arc. However, to achieve satisfactory dose distributions in a targeted volume using IMAT, multiple arcs were often required. Therefore, more practical algorithms, such as VMAT algorithms, employing one or two arcs were subsequently developed.
Compared to the more traditional static beam intensity modulated radiation therapy (“IMRT”), VMAT is significantly more efficient in both treatment time and total monitor units (“MU”) for similar dose distributions. However, unlike IMRT, the arc optimization problem presented by VMAT is significantly more complex, due to the substantially increased number beam orientations and mechanical constraints of the gantry and collimator of the treatment machines.
To avoid the computational complexity for optimization beam configurations in VMAT, several approaches have developed. One example is the multiresolution method, in which a set of sparse beams are first optimized, and new beams are progressively inserted between sparsely sampled beams using interpolation. Then MLC aperture shapes and weights are randomly sampled using simulated annealing. Typically, VMAT often attempts to optimize about 180 or more beams around a patient. The process would then starts with a small set of equally distributed beams, for example, six beams. An inverse optimization is then performed without much computational burden. Once solved, new beams are interpolated and feasible solutions are searched in the vicinity of the previous solutions. The process is then repeated until the desired number of beams is reached. Although this method has been effective in reducing the complexity of the optimization problem, such greedy heuristic methods do not guarantee optimality. In addition, results are often difficult to reproduce and generate inferior dosimetry due to the lack of global optimization. Furthermore, computational burden increases as more beams are added for multiple arcs and non-coplanar arcs.
Despite the potential for using a single arc, in practice, two or more arcs are often utilized with VMAT in order to achieve the desired dosimetry. This is because achieving optimal results often requires applying different initial conditions. Also, in addition to the stochastic nature of simulated annealing, optimization results depend highly on the order and timing in which the parameters, such as weights and penalties, are being applied. As a result, reproducing an existing plan is difficult, if not impossible, and multiple iterations are often required to achieve optimality.
Due to the importance that VMAT plays in today's radiotherapy practice, there is a strong need to overcome the existing limitations of previous technologies.
The present disclosure overcomes the drawbacks of previous technologies by providing a system and method for generating a radiotherapy treatment for volumetric modulated arc therapy (“VMAT”). In particular, a novel optimization framework is introduced that solves an optimization problem as a global problem. As will be described, in this framework, a single optimization function is utilized that includes a level set function to regularize aperture shapes without relying on a present aperture library. The optimization may then be solved using a proximal-class primal-dual algorithm, which is much more robust than stochastic methods currently available in VMAT solutions.
In accordance with one aspect of the present disclosure, a method for generating a volumetric modulated arc therapy (“VMAT”) plan is provided. The method includes receiving a representation of a patient comprising information related to target and non-target volumes of interest, and generating an objective function based on the representation of the patient. The method also includes performing an iterative optimization process using the objective function to generate a VMAT plan, and generating a report according to the VMAT plan.
In accordance with another aspect of the present disclosure, a volumetric modulated arc therapy (“VMAT”) system is provided. The system includes a radiation source configured to generate and direct radiation to a patient, and a gantry housing the radiation source and configured to rotate about an axis of rotation. The system also includes a control mechanism configured to control the rotation of the gantry and the delivery of radiation from the radiation source to a target volume in the patient, and a computer in communication with the control mechanism. The computer is configured to receive a representation of a patient comprising information related to target and non-target volumes of interest, and generate an objective function based on the representation of the patient. The computer is also configured to perform an iterative optimization process using the objective function to generate a VMAT plan, generate and provide control signals, corresponding to the VMAT plan, to the control mechanism to irradiate the patient.
The foregoing and other aspects and advantages of the invention will appear from the following description. In the description, reference is made to the accompanying drawings which form a part hereof, and in which there is shown by way of illustration a preferred embodiment of the invention. Such embodiment does not necessarily represent the full scope of the invention, however, and reference is made therefore to the claims and herein for interpreting the scope of the invention.
The patent or application file contains at least one drawing executed in color. Copies of this patent or patent application publication with color drawing(s) will be provided by the Office upon request and payment of the necessary fee.
Volumetric modulated arc therapy (“VMAT”) is a widely employed radiation therapy technique that can achieve dosimetry comparable with static beam intensity modulated radiation therapy (“IMRT”) while reducing monitor units and treatment time. Current methods for generating VMAT plans often utilize algorithms that do not guarantee optimality and produce inconsistent results. For instance, the multiresolution method optimizes a set of sparsely selected beam angles, and progressively inserts new beams using interpolation. This greedy heuristic approach does not guarantee optimality, and can produce inferior dosimetry.
To avoid stochastically simulated annealing and direct optimization approach the multiresolution method, one group developed a column-generation-based VMAT algorithm. This iterative technique begins by selecting an aperture for a given beam angle from a set of apertures based on its contribution to an objective function. The optimization then proceeds to the next beam angle in a set of densely sampled arc beams and selects the next aperture, imposing potential mechanical constraints based on the previous aperture. Such greedy heuristic approach is limited, however, in that it solves a sub-problem at each step, and does not simultaneously optimize all possible beam angles. Also, the number of possible aperture shapes increases combinatorially with each iteration. Furthermore, using a complete aperture set (e.g. for a large tumor, or high resolution dose modulation) quickly becomes mathematically intractable.
Another group introduced a binary level-set model in the optimization VMAT. While efficient, the binary level-set is discontinuous, and derivatives at the boundary do not exist, which can lead to poor accuracy. To avoid the progressive sampling issue, yet another group developed an approach that starts with an IMRT optimization every 2° of an arc, and then creates apertures by merging and simplifying adjacent fluence maps. However, this leads to another problem that is common in inverse treatment planning, which is that heuristic conversion from a fluence map to multi-leaf collimator (“MLC”) segments typically introduces noticeable and unpredictable dosimetric quality degradations.
As appreciated from the above, VMAT optimization methods to date have a number of drawbacks and produce empirical solutions, which are limited in consistency and quality. Therefore, the present disclosure provides a system and method for generating a radiotherapy treatment for volumetric modulated arc therapy (“VMAT”) that overcome the limitations of these previous technologies. Specifically, a novel direct aperture optimization framework is introduced that is non-progressive and provides significantly improved results. Termed herein as comprehensive VMAT (“comVMAT”), the present framework utilizes an optimization function that may include a dose fidelity term, an anisotropic total variation term, and a level set function.
Referring particularly to
In some implementations, steps of the process 100 may be performed using at least one processor configured to execute programming or instructions stored in non-transitory computer readable media. The processor may be a general-purpose processor. Alternatively, the processor may be a dedicated or application-specific processor having non-transitory programming or executable instructions hardwired therein. For example, steps of the process 100 may be carried out using an optimizer or a processing module of a computer.
The process 100 may begin at process block 102 with receiving a representation of a patient. The representation, received from an imaging system, database or data server, may include images, or image data, acquired from a patient, for instance, during a treatment simulation protocol. For example, the images may include computed tomography (“CT”) images, magnetic resonance (“MR”) images, positron emission tomography (“PET”) images, and others. The images may be one-dimensional (“1D”), two-dimensional (“2D”), three-dimensional (“3D”) and four-dimensional (“4D”) images.
In some aspects, the representation at process block 102 may also include additional information related to target and non-target volumes of interest (“VOI”). This may include contours of diseased and normal tissue structures, as well as dosing requirements or dose constraints based upon predetermined dose prescriptions. Alternatively, a user or clinician may provide such information. For instance, a clinician may select VOIs, either manually or with the help of automated or semi-automated segmentation algorithms, and provide specific dose constraints for the selected VOIs.
Then, an optimization framework may then be applied to generate a VMAT plan. As indicated by process block 104, this includes generating an objective function based the received representation. As discussed in more detail below, the optimization function may include an L2-norm fidelity term that is used to penalize the difference between an optimized dose and a prescribed dose. The optimization function may also include an anisotropic total variation term that can promote piecewise continuity in the fluence maps, and allowing for direct aperture optimization. In addition, a level set function may be further included in the optimization function in order to describe the aperture shapes, with the difference between aperture shapes at adjacent angles being penalized to control multi-leaf collimator (“MLC”) motion range.
Then, at process block 106, the objective function may be used in an iterative optimization process that is carried out to generate the VMAT plan. In particular, the VMAT plan may reflect a set of optimized apertures corresponding to various beam angles in an arc used to irradiate the patient, as well as duration of exposure for each aperture. In some implementations, a block minimization algorithm may be adopted to solve the optimization function in the optimization process. An alternating optimization strategy may also be used to solve the fluence intensity and aperture shapes simultaneously.
A report may then be generated based on the VMAT plan generated, as indicated by process block 108. The report may be in any form and include a variety of information, including information describing beam characteristics and delivery configurations optimized using an optimization framework, in accordance with the present disclosure. For example, the report may be provided as output to a display.
In some aspects, report may include instructions suitable for execution by a therapy system, such as the VMAT system described with reference to
The present optimization framework will now be described. In this framework, an objective function can be generated using the following form:
where fθ, cθ, and Φθ are the optimization variables. In particular, fθ is the vectorized fluence map, cθ is a value that f approaches within an aperture, and Φθ is the level set function, defined as positive where the aperture exists and negative elsewhere. The level set {(x,y)|Φθ(x,y)=0} describes the aperture boundary. Beam angles are indexed by θ, which ranges from 1 to n, and x and y are indices for a beamlet at a given θ. The fluence to dose transformation matrix is denoted by A, and the desired dose, d0, is set as the prescription dose at the PTV and zero elsewhere. The diagonal weighting matrix, W, weighs the structures of interest. The derivative matrices, D1 and D2, take the derivative of the fluence in both directions parallel and orthogonal to the MLC leaf movement. H is the Heaviside function,
Essentially, H(Φθ(x,y)) equals one inside the aperture and zero elsewhere. fθx,y is a scalar value representing a single beamlet at a given beam angle θ and an x and y location on the beam, while f0 is a vector of all the fluences at a specific beam angles. c0 is a scalar quantity and only has one value per beam at a given time.
Intuitively, the dose fidelity term, in Eq. (1), attempts to push the final dose as close as possible to the desired dose. Term set 1 is the anisotropic total variation (“TV”) regularization, which has been shown to successfully encourage piecewise continuity on the fluence maps. The TV regularization term considers the entire fluence map of the beam, so the term ultimately controls the segment size and shape, abating irregularities and holes in the aperture shape. Soft regulation of the minimal leaf gap and the max leaf interdigitation can be accomplished by independently adjusting the weightings λ1 and λ2, respectively. Term set 2 is pushing f toward c where the aperture is defined and zero elsewhere. Term set 3 encourages adjacent beam angles to be similar to regulate leaf movement between beam angles. For the first and nth θ, the Φθ−1(x,y) and Φθ+1(x,y) are equal to their respective Φθ(x,y).
As mentioned, a block minimization algorithm may be used to solve the objective function in Eq. (1). This may be achieved by alternatingly updating the fluence fθ, aperture intensity cθ, and aperture shape Φθ, while holding the other two variables constant. The algorithm may be broken down into 3 modules, as described below. Each iteration of the algorithm runs module 1 through 3, and the process is repeated until a satisfactory convergence rate is achieved. Convergence of the alternating approach is assured, as long as each module is able to find a minimum for its respective variable, while holding the other variables constant.
In particular, module 1 minimizes Eq. (1) with respect to fθ while holding cθ and Φθ constant. This subproblem can be rewritten as
where I is the identity matrix, and HΦ
minimize S(Kx)+R(x). (4)
Here, S and R are lower semicontinuous functions, and K is a matrix. Equation (3) can be written in the form of Eq. (4) by defining the following:
The overrelaxed Chambolle-Pock algorithm may be used to solve this formulation, with the iteration:
z
fn+1=p
zn+1=p
where τ and σ are step sizes that satisfy the constraint τσ∥K∥2≤1, and p is the overrelaxation parameter ranging from 0 to 2. The operator norm of K is estimated with the power iteration method. The proximity operator, or “proxy” operator, is defined as
and S* is the convex conjugate of S, defined as
Evaluation of these operations yield closed form, low cost calculations for the Chambolle-Pock algorithm,
and a separable sum rule allows for
With respect to the second module 2, Eq. (1) is minimized with respect to cθ while Φθ and fθ are held constant, which is provided by the closed-form solution
This calculation takes an average of the beamlet intensities that are defined as part of the aperture for each beam angle.
With respect to the second module 3, Eq. (1) is minimized with respect to Φθ while fθ and cθ are held constant,
Φθ is iteratively updated by the expression
where
was derived as
Practically, a sigmoid function and its derivative can be used to approximate the Heaviside and the Dirac delta function,
where q is some constant. A larger value of q allows for the sigmoid function to more closely resemble the Heaviside function. The derivation for
can be obtained using the following formulation:
The formulation can be solved using gradient descent, where Φθ can be updated by the equation. To iteratively determine Φθ, we change Φθ by a small step dΦθ(x,y)/dt, which changes Eq. (15) to
The first order Taylor expansion of Eq. (16) is
Gathering all of the terms that include dΦθ(x,y)/dt results in
Simplifying Eq. (18) yields
The first set of terms that include dΦθ(x,y)/dt can be rewritten as the inner product
The terms fθ
Once the algorithm has converged, and the apertures shapes no longer change, a final polishing step may be taken to ensure superior plan quality. The formulation locks in the solved aperture shapes and solves for the fluence of each beam angle without the aperture regularization constraints,
The optimization variable, b, contains one intensity value for each beam angle. F is a binary matrix containing all of the aperture information from H(Φθ(x,y)) for all of the beam angles. These two variables are related to the fluence via the equation f=Fb. This optimization can be easily solved with the Chambolle-Pock algorithm. By solving the optimization in Eq. (22) as the last step, only the dose difference is penalized, ensuring that the regularization and aperture constraints are not hindering the final dosimetric outcome.
Although plan delivery time is not explicitly controlled by the objective function in Eqn. (1), it may be indirectly maintained by weights of terms in the objective function. For instance, the total variation term, term set 1, which minimizes the number of deliverable segments, and the aperture similarity term, term set 3, may be selectively controlled to limit the MLC leaf motion between each angle, and hence delivery time. However, a trade-off between plan quality and delivery time would likely be considered since relaxing the weights on these terms to allow more segments and greater changes between apertures will increase treatment time and offset the benefit of using a single arc.
To test the efficacy of the present approach, single arc comVMAT plans were generated for a glioblastoma multiforme (“GBM”) case, a lung (“LNG”) case, and two head and neck cases—one with three PTVs (H&N3PTV) and one with four PTVs (H&N4PTV). The plans were optimized using an alternating optimization strategy and utilized 180 beams with 2° angular resolution. Other beam numbers of angular resolution may also be possible. The plans were compared against the traditional clinical VMAT (“clnVMAT”) plans utilizing two overlapping coplanar arcs for treatment. A summary of prescription doses and PTV volumes for the four patients is summarized in Table I.
As appreciated from description herein, the present framework utilizes a non-greedy algorithm can simultaneously optimize all beams in an arc and directly generate deliverable apertures to generating comVMAT plans. This algorithm can take advantage of a linear accelerator's full digital capability in dose rate and gantry rotation speed modulation. Furthermore, the algorithm generates plans that are superior to those obtained using existing VMAT algorithms.
Specifically, results herein show that the present optimization framework can be used to generate comVMAT plans that can converge within 600 iterations using a block minimization algorithm. In addition, comVMAT plans are able to consistently reduce the dose to all organs-at-risk (“OARs”), as compared to plans obtained using previous techniques. On average, comVMAT plans reduced the max and mean OAR dose by 6.59% and 7.45%, respectively, of the prescription dose. Reductions in max dose and mean dose were as high as 14.5 Gy in the LNG case and 15.3 Gy in the H&N3PTV case. PTV coverages measured by D95 (95% of the PTV dose), D98 (98% of the PTV dose), and D99 (99% of the PTV dose) were within 0.25% of the prescription dose. By comprehensively optimizing all beams, beams with higher intensities may be utilized, resulting in dose distributions that resemble those from static beam IMRT plans with beam orientation optimization.
To evaluate the present framework, a convolution/superposition code was used with a 6 MV x-ray polyenergetic kernel. Beamlet dose was calculated for 180 equally spaced coplanar beam angles around the patient. Beamlet size was chosen to be 0.5×0.5 cm2, and the dose matrix resolution was 0.25×0.25×0.25 cm3. The resulting dose was stored in a dose matrix A for optimization. A 5 cm ring structure was added to the optimization to minimize dose spillage. Plans for each patient were then optimized using the comVMAT algorithm, and W was adjusted until a desirable dose was achieved.
The comVMAT plans were compared to the patients' respective clnVMAT plans. Specifically, the clnVMAT plans were generated on an Eclipse treatment planning system using two superimposing 360° coplanar arcs with 90° collimator rotation. The PTV D95, D98, D99, Dmax, and PTV homogeneity, defined as D95/D5, were evaluated. The Dmax and Dmean for the OARs were also assessed. Max dose was defined as the dose at 2% of the structure volume, D2, which is recommended by the ICRU-83 report.
During optimization of each case, aperture shapes converged within about 600 iterations, with a relative convergence of 10−1. This degree of convergence has been shown to produce plans that are dosimetrically equivalent to other plans that have tighter convergences. An example convergence plot for the GBM case is shown in
As mentioned, the comVMAT algorithm was capable of optimizing all 180 beams in the coplanar arc simultaneously for all tested cases.
Referring now to
The two H&N cases, having multiple PTVs and matching the dose coverage to 95% of the PTVs (
On average, the PTV D95, D98, and D99 changed by −0.01%, +0.02%, and −0.23% of the prescription dose, indicating virtually identical dose coverage between comVMAT and clnVMAT. The average calculations included all of the PTVs from the H&N cases. However, PTV Dmax increased, on average, by 1.40% of the prescription dose. This change was associated to the two H&N cases with multiple PTVs. The GBM case had a reduced max dose to the PTV, while the LNG case minimally increased the PTV max dose by 0.08% of the prescription dose. As summarized in Table III, comVMAT was able to decrease dose for the OARs for all cases, with the largest valued dose differences being negative. Specifically, comVMAT spared the OARs max and mean dose by an average of 6.59% and 7.45% of the prescription dose, respectively. Comparing all the cases, the LNG case had the single largest sparing in max dose for the OARs. Specifically, the proximal bronchus was spared by 14.5 Gy of max dose. Likewise, the single largest reduction in the mean dose to an OAR was achieved in the H&N3PTV case, sparing the larynx by 15.3 Gy of mean dose.
Referring now
In using a proximal-class primal-dual algorithm, such as the Chambolle-Pock algorithm, the present framework, by way of L2-norm fidelity terms and L1-norm regularization terms, introduces fluence map simplification into the dose domain optimization. In addition, the optimization problem can be iteratively solved without greedy heuristic solutions, as implemented by previous technique. This allows all VMAT beams to be optimized at the same time without progressive sampling, providing a number of theoretical and practical advantages.
At the theoretical level, the optimization function introduced herein represents a simple yet complete description of the physical problem. Specifically, an L2-norm fidelity term was used to minimize the dose distribution of the prescription dose, and an anisotropic total variation regularization term was used to piecewise smooth the fluence map. The level set function term shaped the fluence to fit the aperture and encouraged the continuity in the aperture shapes between adjacent beams.
Unlike commonly observed solutions utilizing heuristic algorithms, the present optimization framework produces a finalized solution that does not require additional “patches.” As described, the Chambolle-Pock algorithm was used to manage optimization, given the size of the optimization problem and that the cost function is not differentiable. This algorithm proved to remarkably fast at solving this type of optimization problem because it did not require solving system of linear equations involving the fluence to dose transformation matrix, A, at every iteration. Unlike first order methods or alternating direction method of multipliers, the Chambolle-Pock algorithm only requires the multiplication of the matrix and its transpose at each iteration, resulting in substantial speedup.
The present framework offers several advantages compared to existing VMAT methods. First, comVMAT optimizes all beam apertures and beam intensities together, providing greater flexibility to reaching targeted or ideal dosing. For instance, a difference in isodose distribution can be appreciated in
As described, the present framework may produce apertures that are deliverable using more than one segment. In the example of
Multi-segment apertures obtained using the present framework may also provide the opportunity to generate hybrid static beam IMRT and VMAT plans. For instance, by further relaxing the TV regularization, a dosimetric improvement is expected, with more beams needing two or more segments. The gantry speed and LINAC output could then be modulated to deliver these beams, adding more static beam flavor to comVMAT for superior dosimetry.
At the practical level, the single arc comVMAT described herein has been shown to be superior to the current commercial implementation of clnVMAT that uses two superimposing arcs. Also, comVMAT may be potentially advantageous in knowledge based planning because it is more robust to optimization history. By contrast, clnVMAT depends on the entire history of optimization parameter set up, which is impossible to track and incorporate in knowledge based planning. Furthermore, comVMAT plans are reproducible, and provide a single set of optimization parameters for future learning.
In addition, the present formalism does not depend on greedy heuristics and produces dosimetry using a single arc that is at least comparable to the multiple arc VMAT planning techniques. Herein, the dose calculation code was fine-tuned to match the percent depth dose (“PDD”) and penumbra of the Eclipse system beams. However, the TPC commissioning process is not entirely transparent in order to achieve an exact reproduction. As such, the present approach may be modified in accordance with specific TPS vendors to achieve an exact comparison.
As appreciated from description herein, a new approach for the comprehensive VMAT optimization was demonstrated. The optimization framework described formulates the VMAT problem using a single optimization function, which includes using a level set function to regularize the MLC aperture shapes without relying on a preset aperture library. The optimization function was solved using a proximal-class primal-dual algorithm, which is more robust than stochastic method used in the existing VMAT solutions. The results showed that the new comVMAT using a single arc was consistently superior in OAR sparing to the clinical VMAT using two arcs, while keeping a similar PTV dosimetry. As a VMAT approach, the particular direct aperture optimization (“DAO”) formulation described herein was specifically tailored to produce one segment per beam in most cases. However, the formulation may be generalized using a number of mathematical tools for applications including static beam IMRT may involve multiple levels of segmentation. For instance, it is envisioned that the present approach can be generalized to a libraryless DAO formulation for both static beam and VMAT IMRT problems.
Referring now to
The system 600 also include a computer 616 that receives commands and scanning parameters from an operator via a console 618, or from a memory or other suitable storage medium. An associated display 620 allows the operator to observe data from the computer 616, including images or representations of the patient 610 that may be used to review or modify the treatment plan, and to position the patient 610 by way of appropriately adjusting the position of the patient table 612. The operator supplied commands and parameters may also be used by the computer 616 to provide control signals and information to the control mechanism 614.
The radiation source 602 produces a radiation beam 622, or “field,” that is modulated by a collimator 624. The collimator 624 may include a multileaf collimator (“MLC”) that is composed of a plurality of independently adjustable collimator leaves. In such a configuration, each leaf in the collimator 624 is composed of an appropriate material that inhibits the transmission of radiation, such as a dense radioopaque material, and may include lead, tungsten, cerium, tantalum, or related alloys.
The radiation source 602 mounted on the gantry 604 rotates about a rotation axis 606 so that the radiation beam 622 may irradiate the target volume 608 in the patient 610 from a variety of gantry angles, θi. The radiation source 602 is controlled by a radiation controller 626 that forms a part of the control mechanism 614, and which provides power and timing signals to the radiation source 602.
A collimator controller 628, which forms a part of the control mechanism 614, controls the movement of each of the collimator leaves in and out of its corresponding sleeve. The collimator controller 628 moves the collimator leaves rapidly between their open and closed states to adjust the aperture shape of the collimator 624 and, therefore, the shape and fluence of the radiation beam 622. The collimator controller 628 receives instructions from the computer 616 to allow program control of the collimator 624.
A gantry controller 630, which forms a part of the control mechanism 614, provides the signals necessary to rotate the gantry 604 and, hence, to change the position of the radiation source 602 and the gantry angle, θi, of the radiation beam 622 for the radiation therapy. The gantry controller 630 connects with the computer 616 so that the gantry 604 may be rotated under computer control, and also to provide the computer 616 with a signals indicating the gantry angle, θi, to assist in that control. The position of the patient table 612 may also be adjusted to change the position of the target volume 608 with respect to the radiation source 602 by way of a table motion controller 632, which is in communication with the computer 616.
The system 600 may also include an imager 632 that extends from the gantry 604, which is configured to acquire image data from a patient, for instance, during patient setup. To image the patient, the control mechanism 614 controls the extension of the imager 632 and radiation provided by the radiation source 602. Data acquired by the imager 632 may then be relayed to the computer 616, which then generate images viewable on the display 620. The images may then be used to correct the positioning of the patient prior to treatment.
During operation of the system 600 to deliver treatment, the collimator controller 628 receives, from the computer 616, segmentation information indicating the aperture shape to be used for each gantry angle, θi, during each sweep of the radiation source 602, in accordance with the treatment plan. The segmentations describe the intensity of the radiation beam 622 that is desired for each gantry angle, θi. The aperture number, shapes and segmentations may be optimized using an optimization framework, as described.
In some aspects, the computer 616 may be configured to generate a VMAT plan using programming or instructions stored in non-transitory computer readable media. To this end, the computer 616 may include an optimizer or a processing module configured specifically to carry out steps for generating the VMAT plan. Alternatively, treatment planning may be performed using an external computer, such as a treatment planning workstation, in communication with the computer 616.
Among other steps, the computer 616 may be configured to receive a representation of a patient using an input, and generate an objective function based on the representation of the patient. As described, the representation may include various images, or image data, acquired from a patient, for instance, during a treatment simulation protocol. To this end, the computer 616 may be configured to control the imager 632 to acquire the imaging data. In addition, as described, the representation received the computer 616 may also include other information, such contours of diseased and normal tissue structures, dosing requirements or dose constraints based upon predetermined dose prescriptions, and so forth.
The computer 616 may then generate an objective function based on the representation of the patient, and generate a VMAT plan by performing an iterative optimization process using the objective function. In some implementations, the VMAT plan generated by the computer 616 may be configured in accordance with the mechanical and operational specifications of the system 600. To this end, the computer 616 may have such information stored in a memory, or may acquire such information from the control mechanism 614 or other system, or may determine such information.
The computer 616 may then report results associated with the VMAT plan generated to a user via the display 620. In addition, the computer 616 may also generate and provide control signals and information to the control mechanism 614 to execute the VMAT plan and treat the patient.
The present invention has been described in terms of one or more preferred embodiments, and it should be appreciated that many equivalents, alternatives, variations, and modifications, aside from those expressly stated, are possible and within the scope of the invention.
This application is based on, claims priority to, and incorporates herein by reference in its entirety U.S. Provisional Application 62/479,296, filed Mar. 30, 2017, and entitled “SYSTEM AND METHOD FOR NON-PROGRESSING SAMPLING IN VOLUMETRIC MODULATED ARC THERAPY.”
This invention was made with government support under R43CA183390 and R01CA1883300 awarded by the National Institutes of Health. The government has certain rights in the invention.
Number | Name | Date | Kind |
---|---|---|---|
9764162 | Willcut | Sep 2017 | B1 |
20120136194 | Zhang | May 2012 | A1 |
Entry |
---|
Bedford J.L., “Treatment planning for volumetric modulated arc therapy,” Med. Phys. 36, 5128-5138 (2009). |
Bortfeld et al, “Single-arc IMRT?,” Phys. Med. Biol. 54, N9-N20 (2009). |
Bratengeier K., “2-Step IMAT and 2-Step IMRT in three dimensions,” Med. Phys. 32, 3849-3861 (2005). |
Cao, D. et al, “A generalized inverse planning tool for volumetric-modulated arc therapy,” Phys. Med. Biol. 54, 6725-6738 (2009). |
Chambolle et al, “A first-order primal-dual algorithm for convex problems with applications to imaging,” J. Math. Imaging Vision 40, 120-145 (2011). |
Cheng, L.-T. et al, “Binary level-set shape optimization model and algorithm for volumetric modulated arc ther-apy in radiotherapy treatment,” SIAM J. Sci. Comput. 35, B1321-B1340 (2013). |
Condat, L. “A Primal-Dual Splitting Method for Convex Optimization Involving Lipschitzian, Proximable and Linear Composite Terms,” Journal of Optimization Theory and Applications 158, 460-479 (2013). |
Craft, et al, “Multicriteria VMAT optimization,” Med. Phys. 39, 686-696 (2012). |
Crooks, et al, “Aperture modulated arc therapy,” Phys. Med. Biol. 48, 1333-1344 (2003). |
Dong, P. et al, “4pi noncoplanar stereotactic body radiation therapy for centrally located or larger lung tumors,” Int. J. Radiat. Oncol., Biol., Phys. 86, 407-413 (2013). |
Earl, M., et al, “Inverse planning for intensity-modulated arc therapy using direct aperture optimization,” Phys. Med. Biol. 48, 1075-1089 (2003). |
Gorski, J., et al, “Biconvex sets and optimization with biconvex functions: A survey and extensions,” Math. Methods Oper. Res. 66, 373-407 (2007). |
Grégoire et al, “State of the art on dose prescription, reporting and recording in Intensity-Modulated Radiation Therapy (ICRU report No. 83),” Cancer/Radiothér. 15, 555-559 (2011). |
Jin, J.-Y., et al, “Advances in treatment techniques: Arc-based and other intensity modulated therapies,” Cancer J. 17, 166-176 (2011). |
Kawashima, M, et al, “Comparison of total MU and segment areas in VMAT and step-and-shoot IMRT plans,” Radiol. Phys. Technol. 6, 14-20 (2013). |
Kim, H, et al, “Dose optimization with first-order total-variation minimization for dense angularly sampled and sparse intensity modulated radiation therapy (DASSIM-RT),” Med. Phys. 39, 4316-4327 (2012). |
Luan, S. et al. “A new MLC segmentation algorithm/software for step-and-shoot IMRT delivery,” Med. Phys. 31, 695-707 (2004). |
Men, et al, “An exact approach to direct aperture optimization in IMRT treatment planning,” Phys. Med. Biol. 52, 7333-7352 (2007). |
Neylon, J., et al, “A nonvoxel-based dose convolution/superposition algorithm optimized for scalable GPU architectures,” Med. Phys. 41, 101711 (15pp.) (2014). |
Nguyen D, et al, “A novel software and conceptual design of the hardware platform for intensity modulated radiation therapy,” Med. Phys. 43, 917-929 (2016). |
Nguyen D, et al, “Dose domain regularization of MLC leaf patterns for highly complex IMRT plans,” Med. Phys. 42, 1858-1870 (2015). |
Osher et al, Level set methods and dynamic implicit surfaces. (Springer Science & Business Media, 2006). |
Otto, K. “Volumetric modulated arc therapy: IMRT in a single gantry arc,” Med. Phys. 35, 310-317 (2008). |
Peng, F., et al, “A new column-generation-based algorithm for VMAT treatment plan optimization,” Phys. Med. Biol. 57, 4569-4588 (2012). |
Que, W, “Comparison of algorithms for multileaf collimator field segmentation,” Med. Phys. 26, 2390-2396 (1999). |
Rao, M., et al, “Comparison of Elekta VMAT with helical tomotherapy and fixed field IMRT: Plan quality, delivery efficiency and accuracy,” Med. Phys. 37, 1350-1359 (2010). |
Salari et al, “A column-generation-based method for multi-criteria direct aperture optimization,” Phys. Med. Biol. 58, 621-639 (2013). |
Shepard D. M., et al., “An arc-sequencing algorithm for intensity modulated arc therapy,” Med. Phys. 34, 464-470 (2007). |
Verbakel W.F.A.R., et al, “Volumetric intensity-modulated arc therapy vs conventional IMRT in head-and-neck cancer: A comparative planning and dosimetric study,” Int. J. Radiat. Oncol., Biol., Phys. 74, 252-259 (2009). |
Xia, P, et al. “A leaf sequencing algorithm to enlarge treatment field length in IMRT,” Med. Phys. 29, 991-998 (2002). |
Yu, C.X., “Intensity-modulated arc therapy with dynamic multileaf collimation: An alternative to tomotherapy,” Phys. Med. Biol. 40, 1435-1449 (1995). |
Zhu, L. et al, “Using total-variation regularization for intensity modulated radiation therapy inverse planning with field-specific numbers of segments,” Physics in medicine and biology 53, 6653 (2008). |
Number | Date | Country | |
---|---|---|---|
20180280725 A1 | Oct 2018 | US |
Number | Date | Country | |
---|---|---|---|
62479296 | Mar 2017 | US |