1. Field of the Invention
The present invention relates to a system for calculating and using a probability indicator for the anticipated outcome of an immediately following defibrillator shock on the basis of ECG, patient information and treatment characteristics measured during sudden cardiac arrest and resuscitation.
2. Description of Related Art
Nearly 40% of all those who suffer sudden cardiac arrest could have a chance of survival if they receive good, lifesaving treatment immediately. When treatment is delayed, the chances of survival decrease, cf. the article by Holmberg S, Holmberg M: “National register of sudden cardiac arrest outside of hospitals” 1998 [1]. The treatment primarily consists of cardio-pulmonary resuscitation (CPR), which is administered until a defibrillator is in place. Thereafter, the treatment consists of alternating use of the defibrillator and CPR until resuscitation or until an ALS (ALS=“Advanced Life Support”) team arrives. The latter also includes medication and securing of the respiratory passages as part of the treatment, cf. ILCOR, “Advisory statements of the International Liaison Committee on Resuscitation.” Circulation 1997; 95:2172-2184 [6]
Scientific papers in recent years point out a number of factors that affect the chances of survival:
Quality of Studies show that the quality of the CPR influences the survival.
Timing of A study shows that when the duration of sudden cardiac arrest CPR and exceeds a number of minutes, the chance of survival will defibrillator increase if the ambulance personnel first administer a period of treatment: CPR before the defibrillator is used. (Cf. Cobb L, et al. “Influence of cardiopulmonary resuscitation in patients with out-of hospital ventricular fibrillation”. JAMA, Apr. 7, 1999-Vol 281, No 13 [5]
In the case of sudden cardiac arrest, the electrical activity in the heart (ECG) will indicate the state of the heart. Today's defibrillators measure and analyse ECG in order to classify the rhythm. If the rhythm is classified as Ventricle Tachycardia (VT) or Ventricle Fibrillation (VF), defibrillator treatment may have an effect. VT is often the precursor of VF. VF will as time goes by cause the energy and oxygen reserves of the heart muscle to deplete, and eventually the rhythm will become Asystole, a rhythm characterised by very little or no electrical activity. The purpose of the defibrillator treatment is to restore the organised electrical activity of the heart and the associated blood pressure and blood circulation. This is often denoted ROSC—“Return of Spontaneous Circulation”, and is the first step towards survival.
Only a fraction of the shocks delivered actually result in ROSC. Most shocks today do not give ROSC, cf. the publications Gliner BE et al. “Treatment of out-of hospital cardiac arrest with a Low-Energy Impedance-Compensating Biphasic Waveform Automated External Defibrillator” [7], Sunde K, Eftestøl T, Askenberg C, Steen P A. “Quality evaluation of defibrillation and ALS using the registration module from the defibrillator”. Resuscitation 1999 [14]. In general, it can be said that the chance of ROSC is at its greatest immediately after sudden cardiac arrest, when the heart muscle still possesses energy reserves and oxygen. Many patients achieve ROSC after alternating use of shocks and CPR. The disadvantages of having to give many shocks are several: First of all, no CPR will be given during the shock treatment, a factor that further aggravates the situation for the vital organs, particularly for the brain. Furthermore, it has been shown that the heart muscle is also damaged by the shocks, and that the damage increases with the number of shocks and the amount of energy, cf. the publications Ewy G A, Taren D Bangert J et al. “Comparison of myocardial damage from defibrillator discharges at various dosages.” Medical instrumentation 1980; 14:9-12. [16]. For the patient, the ideal would be to be given only one shock, and for this shock to give ROSC.
Thus, for many patients, it is crucial that the administration of CPR be effective, so as to revitalise the heart through supplying a flow of blood through the heart muscle, cf. the publication Michael J R et al. “Mechanism by which augments cerebral and myocardial perfusion during cardiopulmonary resuscitation in dogs”. Circulation 1984; 69:822-835. [17]. This revitalisation can be indicated through ECG measurements, where ECG characteristics such as form, spectral flatness measurements, frequency, amplitude, energy etc. is seen to change back towards the values that would have existed immediately after the heart action was suspended, cf. the publications Eftestøl T, Aase, S O, Husøy J H. “Spectral flatness measure for characterising changes in cardiac arrhythmias”. Computers in Cardiology, [15] and Noc M, Weil M H, Gazmuri S S, Biscera I and Tang W. “Ventricular fibrillation voltage as a monitor of the effectiveness of cardiopulmonary resuscitation”. J Lab Clin Med, September 1994 [13]. This revitalisation will increase the probability of the next shock resulting in ROSC.
Unfortunately, not everyone survives. For many, the reason behind the sudden cardiac arrest is such that resuscitation is impossible. Furthermore, the time factor and the quality of the treatment will also play a part and affect the chance of survival.
Resuscitation guidelines describe a protocol that is the same for everyone, regardless of sex, race, how long the heart action has been suspended, whether a member of the public has given CPR etc. The means of resuscitation are primarily CPR and defibrillator treatment, and later also medication administered by lifesavers who have been given special training in this area. This protocol is such that if the first three shocks have no effect, CPR is to be given for 1 minute, then three more shocks, and so on. As it takes about one minute to give three shocks, the patient will be without CPR for half of the time.
Literature and other patent applications describe technology, the object of which is to guide the lifesaver in the choice between CPR and defibrillator treatment. Brown et al in U.S. Pat. Nos. 5,683,424 and 5,571,142 [10] describe a system that, based on spectral measures in VF, instructs the lifesaver to either give CPR or give a shock. A separate analysis of this method, where the method has been tested on human VF, yields results that show the method to have a low specificity, i.e. that the method will only to a limited degree reduce the number of unnecessary shocks. Noc M, Weil M H, Tang W, Sun S, Pernat A, Bisera J. “Electrocardiographic prediction of the success of cardiac resuscitation”. Crit Care Med, 1999, Vol 27, No 4 [12] describe a similar system, based on an animal model, which links the mean amplitude and dominant frequency of VF to the outcome of the defibrillator shock. Both of these methods aim to advise against defibrillator use as long as the condition of the heart is such that a shock is assumed not to have an effect, and instead use CPR. Both methods define absolute criteria based on a limited number of observations from a defined group of patients or animals.
The object of the present invention is to seek to constantly optimise the treatment through:
The object of the invention is to contribute towards giving the patient a treatment that is better suited to the individual, and which gives a greater chance of survival. The use of historical data could make it possible to adjust for individual differences and for patient group characteristics and for treatment characteristics. If such historical data were present, the system could have means of providing further input about the patient and the treatment in order to supplement information from the sensors connected to the patient:
The above could result that the calculated probability indicator had different values, depending upon patient data, patient group data or treatment data.
Using PROSC to optimise the treatment may be done in several ways. For advanced users, the most expedient will be to present the indicator graphically versus time, as a trend curve. This will immediately provide a direct indication of the state of the heart, and also indicate the effect of medication and CPR.
For groups who are not trained in relating to this type of information, the most appropriate thing will be to provide automatic decision support in the question of whether or not to give CPR, in which way CPR should be given, or whether shocks should be given. The principle of a simple decision support could be:
The following will describe the invention in greater detail, with reference to the drawings, in which:
The system consists of one, alternatively several, computer(s) 1 in a network that can communicate with a number of positioned analysis units 2. These may either be integrated into equipment (U1, U2 . . . ) such as defibrillators or ECG monitors, or they may occur in or as a support product used during the resuscitation attempt. The analysis units 2 generally operate independently of the computers 1, however after use, the analysis units could deliver field data to the computer 1, and could also receive adjusted algorithms for calculation of property vector and/or PROSC
The analysis unit 2 is normally connected to other subsystems, cf.
Some of these subsystems are standard in equipment such as defibrillators and ECG monitors, and these are as follows:
Electrodes E, which provide input on ECG and impedance as well as means for providing defibrillator energy to the heart, are connected to: An ECG measurement system 3, an impedance measurement system 4, the main function of which is to check if the electrodes are connected to the patient, and circuitry for high voltage generation and shock delivery 5. Further subsystems are: Processing means 6 which can classify the present ECG rhythm as shockable or non-shockable, processing means 7 which is typically a microcontroller with software, memory 8, user interface 9, power supply and battery 10, and communication means 11. Subsystems 3-11 are standard equipment in defibrillators /monitors, and will therefore not be described further in this specification.
Analysis unit 2 could be a standalone subsystem which is connected to sensors S, electrodes E and with means of receiving specific information relating to patient and treatment and having means of communicating the computed property vector and/or probability indicator. The analysis unit could also be integrated with existing input/output, signal analysis instrumentation and processing means, for instance within a defibrillator.
Analysis Unit 2 Includes the Following Units:
Unit 12 for determining one or more properties of the heart that are processed to a property vector and based on this calculate the probability of ROSC, PROSC indicator, for the patient who is connected up. Module 13, if present, for determining the blood flow through the heart, based on the measured impedance and the change of the impedance between the electrodes as a function of the pumping action of the heart and the expansion of the lungs. Module 14 for registering CPR characteristics from chest compression data, e.g. chest compression depth and rate, and ventilation data from sensors S. Module 15 for inputting patient specific information. Module 16 for inputting any medication administered; and a module 17 for correlating positive changes in PROSC or the property vector with information regarding the treatment given, and display or use this information to guide the treatment.
Further Detailed Description of the Analysis Unit 2.
Module 12, comprising an algorithm v(x) for the calculation of a property vector (v) and algorithm for calculating the probability of ROSC, PROSC, as a function of ECG from the patient who is connected up, and further as a function of specific information regarding patient and treatment:
Module 13 for calculating blood flow through the heart based on the measured impedance and the change of the impedance between the electrodes as a function of the pumping action of the heart and the expansion of the lungs:
This formula is universally known, and is used in Impedance Cardiography.
ΔZ is the impedance change, p is the resistivity of the blood, L is the distance between the electrodes, and Zo is the numerical value of the impedance. A simplification of this formula is:
Here, k is a constant. This measurement will indicate to what degree the blood is flowing, and will contribute towards characterising the condition of the heart in VF/VT, This measurement serve as an indicator of ROSC in case of a successful defibrillator shock.
Module 14 for measuring and registering CPR parameters. Relevant CPR parameters are:
Module 15 for indicating patient specific information. This information can be passed to the analysis unit 1 e.g. by dedicated push buttons or it may come in from an external source such as a patient database or a patient journal on a PC/handheld computer. Relevant information is:
Geographical area
Age
Sex
Weight
Race
Module 16 for indicating medication and dosage given. This information can be passed on to the analysis unit 1 through dedicated push buttons, from a patient journal on a PC or other devices that log the use of medication. Relevant medicines are
Epinephrine
Lidocaine
Bretylium
Magnesium sulphate
Procainamide
Vasopressins
Thrombolysis medication
Module 17 for correlating changes in PROSC with information regarding the treatment given, and displaying or using this information to guide the treatment.
In this regard, a principle of this invention is that there is an opportunity to improve the algorithms for the calculation of the property vector (v) and the algorithms for the calculation of the probability indicator PROSC. These algorithms are improved as a function of experience data. Experience data will typically come from a number of uses from a number of different analysis units. The experience data is then communicated from the analysis units to a central computer, which calculates improved algorithms and then communicate the improved algorithms back to the analysis units. The interval at which this is done can vary.
The computer 1 includes the following subsystems:
(a) Hardware, (b) operating system, (c) software and interface for communication in a network (d) database for field data, (e) algorithm for calculation of a property vector, (f) algorithm for calculation of PROSC, (g) algorithm for correlating changes in PROSC with information regarding patient and treatment, and (h) system for delivery and receipt of data from positioned defibrillators.
With regard to the computers, the subsystems of hardware, operating systems, software and interface are of a generic nature, and will not be described in greater detail.
Specific information about computer 1:
(d) The database for field data consists of a large amount of patient lo episode data, and contains:
Patient information: Sex, age, weight, race, medical record etc.
Geographical information
Information regarding each defibrillator shock: Curve shape, energy, timing
versus VF.
For each shock:
Preshock ECG
Preshock CPR data
Preshock medication data
Preshock impedance data
Postshock ECG
Postshock impedance data
Annotation of ROSC/No-ROSC, with outcome rhythm for each shock
(e) The algorithm for calculation of the property vector (v) makes use of mathematical methods in order to characterise the condition of the heart based on a recording of a bio-medical signal (x). The bio-medical signal is preferably ECG.
The algorithm for calculation of the property vector is hereafter denoted as v(x). v(x), which is used on empirical ECG data, provides two sets of property vectors:
A set, V1, containing n1 property vectors where the outcome of the shock is ROSC, and a set, V2, containing n2 property vectors where the outcome of the shock is no-ROSC.
In general, v(x) is defined as an operator that operates on an ECG segment, x, consisting of N samples, which generates a property vector, v, consisting of M vector elements that ideally takes care of the information in x lo that separates the group of x that results in ROSC, X1, from the group of x that results in no-ROSC, X2. The methods of property extraction are innumerable, and the literature describes some of these, which can be roughly divided into time and transform domain methods, where the object is to structure x in a manner that is appropriate for property extraction. Among preferred time domain methods are:
Among preferred methods for transform domain property extractions are:
The relation between V1 and X1, V2 and X2 respectively are as follows: X1 containing a set of n1 ECG segments, which, when used on v(x), provides a number of property vectors V1, which all belong to the outcome class ROSC (w1). X2 containing a set of n2 ECG segments, which, when used on v(x), provides a number of property vectors V1, which all belong to the outcome class no-ROSC (w2).
(f) A system for calculation of the PROSC function is based on pattern recognition theory, and forms the second element of the classification system. In this context, the term classes is defined as the collection of measurements of the condition of the heart that corresponds to
The property vectors of the two classes are statistically described by
In the case of a given measurement, v, one wishes to determine the class allocation w1 or w2. It has been proven that the expected probability of misclassification is minimised by selecting the wi that corresponds to the maximum P(wi/v). It is further possible to define (make an estimated choice of) the cost of all types of misclassification, such that the expected risk of a given misclassification is given by the product of the cost and the a posteriori probability of the true class. The expected risk of misclassification can then be minimised by classification is a class corresponding the product with the smallest value.
In most cases, the statistics of the property vector are not known. These quantities must then be estimated before PROSC(v) can be produced. The pattern recognition theory describes a multitude of methods for this, which are based on measurements (practice data) that are examples from the various wi. Some examples:
We will start defining the quantities:
We have n=n1+n2. Estimate for a priori probability will then be
ˆP(wi)=ni/n, i=1,2.
The local estimates (within hypercube j) for the class specific probability function will then be
ˆp(v|wi)=nji/ni, i=1,2.
The local estimates for a posteriori probabilities is calculated in respect of the Bayes formula inserted estimate for a priori probability and the local class specific probability density functions. See R. J. Schalkoff. Pattern recognition: Statistical, structural and neural approaches. John Wiley & sons, New York (N.Y.), 1992
ˆP(wi|v)=nji/(nj1+nj2), i=1,2
It is important that a given classifier be tested on a set of observations (test set) independently of the practice data (practice set), in order to check that the classifier yields the expected results, and if not, adjust the decision limits of the classifier. The demand is that there is consistency between practice and testing, that the classifier fulfils the requirement of generality (general applicability).
By dividing the empirical data in two parts and letting the one part represent a set of data called practice set and the other part represent a set of data called test set, the generality is defined as follows: The decision limits which, after having been used on all of the property vectors in each set of data for classification of the outcome, which provides approximately the same performance (the sum of sensitivity and specificity) for both sets of data (practice and test sets) fulfils the requirement of generality. These decision limits occur through an iterative process where the practice set is included in the calculation of the decision limit, see
Those measurements v that correspond to the ROSC outcome belong in w1. The probability of a given measurement, v, belonging in w1 is given by P(w1/v). In other words, this probability function expresses the probability indicator PROSC of ROSC for a given measurement v.
PROSC(v)=P(w1|v)
As mentioned previously, different property vectors, v, can be calculated by means of a countless number of methods. Which methods and which dimension, M, is suitable for expressing PROSC(v) is assessed on the basis of the expected risk in the case of misclassification for each method. The method that minimises this risk is the most appropriate for expressing PROSC(v).
(g) The algorithm for correlating changes in PROSC with information regarding the patient and the treatment is mainly for scientific purposes. The defibrillator may later use the results from the correlation to guide the user during lifesaving.
PROSC(V) has been provided as described under points (d) and (e). In this analysis, ECG segments are extracted from the patient material, so that the ECG segments describe a course of treatment that is as uniform as possible. Examples of such a course of treatment may be
In these ECG segments, corresponding PROSC(v) segments are calculated as described under points (d) and (e). Consequently, the change in PROSC(v), DPROSC, is calculated for each segment. DPROSC is grouped on the basis of those treatment characteristics that are of interest with regard to the effect of the treatment. As an example, one can group DPROSC with regard to the following treatment characteristics, singly or in combination:
Where significant differences in DPROSC occur for dissimilar treatment conditions, this information may be used to identify advantageous treatment methods. This information may be utilised through the person giving the treatment being given feedback regarding good and poor treatment.
(h) A system for delivering and receiving data from positioned analysis units. Here, no special requirements apply. The exchange of data can take place directly through use of memory modules such as PCMCIA, cordlessly by means of IR or RF communication, via networks such as the Internet, or by a direct connection between communication ports in the equipment and the computer. The most practical method these days is to have the analysis unit 2 communicate directly with the computer 1 via a local PC that it can communicate with, and to have the local computer pass the data on via the Internet.
The computer 1 contains empirical data from previous resuscitation attempts, where the outcome of the resuscitation attempt is known. The main ingredient in the empirical basis is the ECG and the associated outcome after a shock (ROSC/no-ROSC). Additional empirical data impart nuances to the relationship between outcome, treatment and patient specific factors. This additional data can be patient specific information and treatment specific information. A practical way of expressing this statistical interrelationship is through a PROSC algorithm, which is a substitute for all the empirical data, but which mathematically expresses the same relationship between the property vector and PROSC.
This algorithm is entered into the program code of the analysis unit, so that when this receives a segment of ECG, the analysis unit will first perform the same calculation of the property vector as that performed by the computer, and then use the property vector as input to the PROSC algorithm in order to calculate the probability indicator of an immediately following defibrillator shock giving ROSC. In case information about patient or therapy is available for the analysis unit, these elements can be used as input also for the PROSC algorithm.
The ever-changing forms of treatment and patient characteristics warrant a continuous update of the empirical basis. This is achieved by each analysis unit recording information about the patient, treatment and recorded ECG and CPR, and passing this on to a central computer, where the central computer repeats the grouping of the property vectors, readjusts the PROSC algorithm and passes the result back to the analysis units.
In summary, a large number of ECG segments x, from patients that has been defibrillated, and where the outcome of defibrillation is known to be either ROSC or No-ROSC, is available. These segments are grouped into either a training set or a test set. The training set is then subject to a first algorithm v(x), which computes a property vector (v) from x. The property vector may comprise a number of different properties, computed on x, either from the time domain representation of x of from the frequency domain representation. The property vector v is optimized such that vectors associated with ROSC have minimum overlap with vectors associated with No-ROSC. However, because an overlap is expected, decision regions must be chosen. There are two criteria for decision regions: The first criteria is to discriminate v associated with ROSC w1 from v associated with No-ROSC w2. The second criteria is to adjust the decision regions such that classification performance of vectors originating from the training set has about the same performance as if the vectors originated from the test set. With this criteria, generality is assured. Generality means that the risk of overtraining or over-fitting of the data is reduced. The result of this exercise means that there is an algorithm which translates the information in a segment of ECG into a property vector, and that there are decision regions defined for the classification of that property vector to either ROSC (w1) or No-ROSC (w2).
With further information, for instance information about the patient and/or the treatment, the above exercise can be repeated for each category of information. The above exercise can also be limited to have different decision regions, depending on what kind of data that is available. For instance, the decision regions can be depending on sex, race, geography, the kind of defibrillator in use, the use of drugs, and so. The probability indicator PROSC is then defined, for each value of v, as the occurrence of ROSC to the sum of the occurrence of ROSC+No-ROSC. Moreover, for each value of v, classification is made as ROSC or No-ROSC depending on the decision region. For simplification, for instance if the property vector has only got one dimension, the probability indicator can be set to just the magnitude of the vector itself.
With this in place, a therapy device, for instance a defibrillator/monitor, can now be arranged to measure ECG, input of patient information and input of treatment information. This therapy device is then arranged with the same first algorithm to translate a segment of ECG into a property vector, and a second algorithm to translate the property vector together with information on the patient and/or the treatment into a probability indicator PROSC. The use of the probability indicator can be to present the value, or trend on a display. Further, information about value and trend can be used in a third, decision support algorithm, such that recommended treatment becomes a function of both the condition of the patient and how the patient responds to the treatment. Even further use of the probability indicator is to correlate the trend of the indicator to characteristics of the treatment. When for instance a positive trend of PROSC has been identified, this trend is then correlated with CPR characteristics, and the result is, e.g., used to set target values for a CPR feedback system sent as information to the display.
As found practical and feasible, the above system can be further optimized when more data is available. For this reason, the therapy device is arranged with memory and communication means, as noted above, such that the database of information can be expanded, and the algorithms optimized.
While the invention has been described in connection with what is presently considered to be the most practical and preferred embodiment(s), it is to be understood that the invention is not to be limited to the disclosed embodiment, but on the contrary, is intended to cover various modifications and equivalent arrangements included within the spirit and scope of the appended claims.
Number | Date | Country | Kind |
---|---|---|---|
1999 4344 | Sep 1999 | NO | national |
This application is a continuation of U.S. application Ser. No. 10/070,545, filed Jun. 4, 2002, which was the US national phase of international application PCT/NO00/00289 filed 6 Sep. 2000, which designated the US, the disclosures of which are incorporated herein by this reference.
Number | Date | Country | |
---|---|---|---|
Parent | 10070545 | Jun 2002 | US |
Child | 11405662 | Apr 2006 | US |