The present disclosure relates generally to stimulated echo acquisition mode spectroscopy, and more specifically, to exemplary embodiments of systems, methods and computer-accessible mediums for providing a simultaneous acquisition of double-spin and stimulated echoes
Proton magnetic resonance (“MR”) spectroscopy (“1H-MRS”) uses selective excitation of a volume of interest (“VOI”) for localization, and to prevent aliasing in spectroscopic imaging (“1H-MRSI”). (See, e.g., References 1-5). Two known and commonly used VOI excitation modules use three orthogonal spatially-selective pulses: 90°-180°-180° Point Resolved Spectroscopy (“PRESS”) (see, e.g., Reference 1), or 90°-90°-90° Stimulated Echo Acquisition Mode (“STEAM”). (See, e.g., Reference 2). The three pulses generate nine transverse coherence pathways: three free induction decays (“FID”), four spin echoes (“SE”) and two stimulated echoes (“STE”). (See, e.g., References 6 and 7). Judicious use of gradient spoiling can dephase all the pathways that originate from spins outside the VOI, which can result in an excellent rejection of outer volume contamination. (See, e.g., References 8, 9). PRESS and STEAM each refocus a single coherence pathway: the double spin echo (e.g., in PRESS) or one of the stimulated echoes (e.g., in STEAM), both of which originate from spins within the VOI.
For a given peak radio-frequency (“RF”) power and specific absorption rate (“SAR”), STEAM's 90° pulses can offer several advantages over PRESS's 180° pulses: (i) SAR that can facilitate the reduction of the recycle time (“TR”) to improve the signal-to-noise-ratio (“SNR”)/unit-time (see, e.g., Reference 10); (ii) higher pulse bandwidths, which can lead to smaller chemical shift displacement (“CSD”) (see, e.g., Reference 11); (iii) shorter echo times (“TE”), suffering less T2 and J-coupling losses; and (iv) improved immunity to B1+ variations. These, however, can come at a cost of about half the SNR per scan. (See, e.g., Reference 9). Since the metabolites' signals can be 4-5 orders of magnitude weaker than the water used in magnetic resonance imaging (“MRI”), the SNR advantage of PRESS can trump STEAM's advantages, and can be the method of choice for 1H-MRS experiments. PRESS's higher CSD in the two 180° pulse directions can be mitigated with high bandwidth refocusing pulses designed using, for example, optimal control theory (see, e.g., References 12 and 13), or the Shinnar-Le Roux procedure(see, e.g., Reference 14), but at a cost of increased SAR, which can become prohibitive when attempting to keep the same CSD. The spatial B1+ variability can be addressed with adiabatic pulses, for example, SADLOVE (see, e.g., Reference 15) or LASER (see, e.g., References 16, 17 and 18), but their double-refocusing can prolong the minimum TE, and can increase SAR. The longer TE can attenuate signals of J-coupled metabolites that can often have short T2s, while their higher SAR can prolong the repetition time, thereby reducing the SNR/unit-time.
While both PRESS and STEAM can select a single, particular coherence, pathway, there can be compelling advantages to combining multiple pathways. For example, solid and liquid state nuclear magnetic resonance (“NMR”), pulse sequences such as HNCA, TOCSY-HSQC and MQMAS, can combine different pathways for increased sensitivity (see, e.g., References 19-21), whereas multi-pulse MR imaging sequences can refocus multiple coherence pathways to generate contrast and boost sensitivity (see, e.g., Reference 22), or to simultaneously obtain maps of several parameters. (See, e.g., Reference 23). Similarly, refocusing the spin and stimulated echo pathways for efficient T1 mapping has been suggested recently for echo planar 1H-MRSI. (See, e.g., Reference 24).
An exemplary three-pulse 1H-MRS localization paradigm can be modified by adjusting the three RF pulses' phases, flip angles, timings, and coherence selection gradients, in order to acquire both the double-spin and one of the stimulated echo pathways, this exemplary procedure can be denoted as STRESS (e.g., STEAM+PRESS). Utilizing the maximal B1+ immunity can yield a family of STRESS sequences, parameterized by a (e.g., the flip angle of the first RF pulse). The B1+ immunity can be shown to be superior to that of both PRESS and STEAM for most values of α. Results can be shown in a phantom and a healthy volunteer at 3 T for the 110°-108°-108° member of the STRESS family, which retains 83%-100% of the PRESS SNR (e.g., metabolite dependent) at only 75% of its SAR, and about 33% of its in-plane CSD for equal TE and TR, while also offering a shorter minimum echo time of, for example, TE=26.6 ms compared to PRESS's TE=40 ms.
Thus, it may be beneficial to provide an exemplary system, method and computer-accessible medium that can be used to, for example, simultaneously acquire a stimulated echo and a double spin echo, and which can overcome at least some of the deficiencies described herein above.
To that end, an exemplary method, system and non-transitory computer-accessible medium according to exemplary embodiments of the present disclosure can be provided for performing magnetic resonance spectroscopy for at least one structure. For example, with such system, method and/or computer-accessible medium, it can be possible to receive information related to a substantially simultaneous acquisition of a stimulated echo pathway and a double spin echo associated with at least one portion of the at least one structure.
For example, the information can include a constructive interference of the stimulated echo pathway and the double spin echo. The simulated echo pathway can have a first weight and the double spin echo can have a second weight which can be substantially the same as the first weight. In addition and/or alternatively, the simulated echo pathway can have a first phase, and the double spin echo can have a second phase which can be substantially the same as the first phase.
Further and/or alternatively, the simulated echo pathway and the double spin echo can have a first angle, a second angle and a third angle therebetween, according to an exemplary embodiment of the present disclosure. The first angle can be between about 90 degrees to about 140 degrees, and/or the second angle and the third angle can be substantially the same. The second angle and the third angle can depend on the first angle. In another exemplary embodiment of the present disclosure, the simulated echo pathway and the double spin echo can be substantially insensitive to inhomogeneities.
According to still another exemplary embodiment of the present disclosure, it can be possible to adjust the timing between respective pulses of the simulated echo pathway and the double spin echo, for example, such that the respective stimulated echo pathway pulse and the respective double spin echo pulse refocus at substantially the same time. In yet another exemplary embodiment of the present disclosure, it can be possible to adjust a gradient amplitude between respective pulses of the simulated echo pathway and the double spin. According to a further exemplary embodiment of the present disclosure, it can be possible to adjust a flip angle of respective pulses of the simulated echo pathway and the double spin. Further, it can be possible to generate at least one image of the portion(s) based on the information.
In some exemplary embodiments of the present disclosure, the stimulated echo pathway and the double spin echo can have different relaxation times. The information can further include an application of a pre-emphasis gradient pulse or a field mapping applied to the double spin echo to prevent a resonant frequency drift of the double spin echo. The stimulated echo pathway can include two stimulated echoes pulses, and the double spin echo can include four spin echo pulses.
These and other objects, features and advantages of the exemplary embodiments of the present disclosure will become apparent upon reading the following detailed description of the exemplary embodiments of the present disclosure, when taken in conjunction with the appended claims.
Further objects, features and advantages of the present disclosure will become apparent from the following detailed description taken in conjunction with the accompanying Figures showing illustrative embodiments of the present disclosure, in which:
Throughout the drawings, the same reference numerals and characters, unless otherwise stated, are used to denote like features, elements, components or portions of the illustrated embodiments. Moreover, while the present disclosure will now be described in detail with reference to the figures, it is done so in connection with the illustrative embodiments and is not limited by the particular embodiments illustrated in the figures or in the appended claims.
Three pulses having flip angles α, β, γ, can give rise to a total of nine transverse coherence pathways: (i) three free induction decays (e.g., FID1, FID2, FID3), (ii) four spin echoes (e.g., SE12, SE13, SE23, SE123) and (iii) two stimulated echoes (e.g., STE+, STE−), as illustrated in
while STE−'s can be proportional to 1/2 sin(α)sin(β)sin(γ).
For example, any three pulses of α, β, γ nutation can generate nine transverse coherence pathways: (i) three FIDs following each pulse, (ii) four spin echoes (SE12, SE13, SE23, SE123) and (iii) two stimulated echoes (STE+, STE−). FIDi can stand for the FID created by the ith pulse; SEij can stand for the spin echo excited by the ith pulse and can be refocused by the jth pulse; and SE123 can be the doubly refocused spin echo observed in the “standard” PRESS sequence. If a gradient having moment kj can be played after the jth pulse, each pathway can experience a total gradient moment shown in the third column. Also shown are two spoiling schemes, for example: (i) a “standard” PRESS procedure (see, e.g., Reference 6), designed to spoil all coherence pathways but SE123, and a (ii) “modified” CPMG-like procedure, which can refocus both SE123 and one of the stimulated echoes (STE). Refocused pathways are shaded. These two exemplary procedure can be used in the STRESS and PRESS sequences shown in
To simultaneously refocus both SE123 and STE−, while dephasing all other pathways, for example, four changes can be proposed to the basic three-pulse localization procedure: (i) Set the inter-pulse gradient spoiler moments to a 1:2:1 ratio (see e.g., Table 1 herein), (ii) Set the inter-pulse delays (e.g., center-to-center) to the same 1:2:1 ratio: τ=TE/4 between the excitation and first second pulse, 2τ=TE/2 between the second and third pulses and τ=TE/4 between the last pulse and acquisition, (iii) implement a “Carr-Purcell-Meiboom-Gill (“CPMG”) condition”, where the second and third pulses' phases' can be incremented by 90° relative to the first and (iv) modify the flip angles of the three pulses as detailed below, and indicated in
Table 2 indicates relative signal amplitudes in homogeneous (η=1.0) and inhomogeneous (η=0.75) B1+ field, for different STRESS family members (Eq. (3) with S0=1 and TR=∞), with PRESS and STEAM for comparison. Relaxation can be neglected in the third and fourth columns, and can be included in the fifth and sixth columns. Further shown in the fourth and sixth columns, in parenthesis, is the percent decline in the signal relative to the case with no B1+ inhomogeneity. For example, α, β and γ can be the flip angles of the first, second and third VOI localization pulses. As a can increase, the signal amplitude and the immunity to B1+ can decrease. Additionally, β and γ also can decrease, which can reduce the relative SAR, and increase the pulses' bandwidths (reducing in-plane CSD). Thus, there can be a trade-off of SAR and CSD and SNR and B1+-immunity. The α>115° cases begin performing unfavorably and should generally be avoided. When α=90°, the STRESS sequence coincides with PRESS, but has the additional B1+ immunity of the CPMG condition, making it considerably more robust to B1+ inhomogeneities than a standard PRESS. (See e.g.,
Exemplary procedures can be understood by examining the standard CPMG sequence used for T2 relaxometry (see, e.g., Reference 26), where, for example:
90°0°−τ−180°90°−τ−[echo]−τ−180°90°−τ−[echo]− . . .
where only T2* decay can be accumulated between the pulses. It can be assumed that T2* can be caused by a spatially linear B0-field inhomogeneity; the effect can be analogous to the 1:2:1 gradient-moment ratio suggested above. For example, the inter-pulse timing modification can assure SE123 and STE− can refocus at the same time, as can also be seen from
If the relaxation is not considered, the combined amplitude of SE123+STE− can be
In one exemplary case, the VOI can often be square in-plane, and the β=γ case can be examined, although the results derived below can be valid regardless of the VOI's dimensions. In an exemplary general case, according to another exemplary embodiment of the present disclosure, it may not be necessary for β and γ to be equal. In the event of B1+ inhomogeneity, the signal's dependence on η can be determined, and the scaling of the B1 field can be, for example:
Fixing α, and demanding both maximal immunity to B1+ variations at nominal B1+ (η=1) and maximal signal intensity can produce, for example:
Where an implicit relation between the flip angles α, β, γ can be obtained. Numerically solving for γ as a function of α and β can yield a family of exemplary STRESS pulse sequences. This is shown in
The individual SE123 (e.g., line 305) and STE− (e.g., line 310), and combined SE123+STE− (e.g., line 315) signal strengths are shown in the graphs of in
The STE− can undergo T1 relaxation for 2τ=TE/2 between the first and second pulses, while the SE123 undergo T2 relaxation. The signal at TE can, therefore, be weighted by both pathways, for example, for the STRESS−α members of the STRESS family. Thus, for example:
The signal can be proportional to S0, which can be obtained in an exemplary pulse-acquire experiment, where B1−, can be the receiver profile and (1−e−TR/T
The differences between Eqs. (3) and (4) can be referred to during absolute or ratio quantification. Even if TR>>T1, the magnitude of the exemplary STRESS echo (see e.g., Eq. (3)) can contain some minor T1 dependence from the STE− component. However, for most cases, TE<<T1, hence exp(−TE/2T1)≈1 with a negligible effect on quantification. Furthermore, since the STE−'s “true TE” (e.g., time it evolves in the transverse plane) can be TE/2, it can be less affected by T2 relaxation compared to the SE123 pathway, which can reduce STRESS' SNR penalty relative to PRESS for TE>0. This can be seen quantitatively from Table 2 above, which compares the signals in a homogeneous B1+ field of PRESS with STRESS for TE=40 ms, T2=150 ms and, T1=1200 ms. At TE=40, PRESS's signal can decay to 77% of the full amplitude (e.g., at TE=0), while, for example, that of STRESS−100 (e.g., α/β/γ=100°/131°/131° can almost be identical at 76%. For example, the STRESS−100 SNR can even surpass PRESS by about 20% if approximately 25% B1+ inhomogeneity can be introduced. Thus, although the theoretical SNR of PRESS in the absence of relaxation and B1+ variation can be higher than STRESS for α>90°, near parity can be expected under realistic conditions at fields ≧3 T. The longer TE, the more pronounced the advantage: for TE=100 ms, T2=150 ms, T1=1200 ms, S0=B1−=1, TR=∞ and η=1, Eqs. (3) and (4) illustrate STRESS−100's SNR to be 54% the full spin echo magnitude (e.g., at TE=0), higher than PRESS's 51%.
Phase cycling the pulses' and receiver's phases among different averages can be a common technique for providing additional suppression of unwanted coherence pathways. (See, e.g., References 38 and 39). Exemplary phase cycling procedures for PRESS and STEAM can be translated to STRESS, since a special phase relationship can be maintained between the different pulses to ensure the SE123 and STE pathways add constructively. One such exemplary phase relationship, φ1=0°, φ2=φ3=90° (e.g., the so-called “CPMG phase relation”), can be provided, where φi can be the phase of the ith pulse (i=1,2,3). This may not be the only such phase relationship, according to an exemplary embodiment of the present disclosure. For example, setting φ1=180°, φ2=φ3=90° and consulting Table 1, which lists the exemplary phases of the different echoes, SE123 and STE− can be seen as being once more in phase. As long as the phase of SE123 and STE− can be kept equal, the above discussion of STRESS' applicability and advantages holds. According to the exemplary system, method and computer-accessible medium, it can be possible to provide four possible phase cycling schemes, outlined in Table 3 below, which have been labeled C1, C2, C3 and C4. The first exemplary procedure, C1, can be used to alternate the overall phase of the pulse train and can eliminate receiver artifacts. The second exemplary procedure, C2, can suppress SE23, FID2 and FID3. This can be useful, since SE23 and FID3 can be responsible for particularly troublesome artifacts (see, e.g., Reference 29), and can also be seen from Table 1 to be the most weakly spoiled pathways in STRESS. The third exemplary procedure, C3, can be used to suppress most, if not all, pathways except FID1 and DC artifacts. The fourth exemplary procedure, C4, can suppress all unwanted coherence pathways, but not DC artifacts. Different cycles can be combined via nesting (see, e.g., Reference 39). However, special care must be taken when nesting different phase cycles. This can involve adding their phases modulo 360°, which may not conserve their exemplary special phase relationships. For example, two STRESS experiments having (φ1, φ2, φ3)=(0°,90°,90° and (φ1, φ2, φ3)=(180°,90°,90°) can be considered. Although each of these can yield a valid STRESS signal, in which STE− and SE123 can both be in phase, the experiment including phases, which can be the direct sum of both, (φ1, φ2, φ3)=(180°,180°,180°), would not and can be inadmissible in a phase cycle.
Table 3 illustrates different phase cycling procedure for the exemplary STRESS experiments. Here φ1, φ2, φ3 can be the phases of pulses 1, 2 and 3, respectively, and φADC can be the receiver's phase, given in degrees. For each combination φ1, φ2, φ3, φADC), the Table 3 shows the phase of each of the pathways, as calculated from Table 1. DC stands for the phase of any DC artifacts in the signal, and can equal the receiver's phase, φADC. The shaded pathways can be those which cancel out upon addition of the different cycles. For example, all pathways can retain SE123 and STE− in-phase (e.g., 90°), which can be beneficial component used by most or all exemplary STRESS experiments.
Exemplary experiments were performed in a 3 T whole body MR imager (e.g., Tim-Trio, Siemens AG, Erlangen Germany), using a transmit-receive head-coil capable of producing 24 μT (1 kHz) of B1+. In-vivo experiments were done on a 31 year old healthy female.
Exemplary experiments were also performed on a spherical phantom containing water doped with copper sulfate (e.g., T1/T2=660/320 ms), and metabolite concentrations approximating in-vivo human brain: N-Acetyl-L-aspartate (e.g., NAA, 12.5 mM, T1/T2=605/465 ms), choline-chloride (e.g., Cho, 3 mM, T1/T2=235/200 ms), creatine-hydrate (e.g., Cr, 10 mM, T1/T2=336/288 ms), myo-inositol (e.g., 7.5 mM), L-glutamic acid (e.g., 12.5 mM) and DL-lactic acid (e.g., about 5.0 mM). Following an initial localization MRI, a 2D CSI slice was positioned in its center and shimmed to about 5.0 Hz.
An exemplary STRESS−110 sequence (see, e.g.,
Rectilinear k-space sampling was employed with a 2D 20×20 acquisition matrix, 20 mm slice thickness, 80×80 mm2 VOI and 160×160 mm2 FOV (see, e.g.,
Once obtained, the FIDs from each scan were zero filled to twice their length, arranged in a three dimensional kx-ky-t matrix and Fourier transformed over all axes. Automated 0th order phasing was performed in each voxel using the NAA peak as a reference. Gaussian apodization was applied in the time domain to improve the SNR. “SNR maps” of the NAA (e.g., about 2.01 ppm), Cho (e.g., about 3.2 ppm) and CH2 peak of creatine (e.g., about 3.9 ppm) were obtained by dividing their peak height by the standard deviation of the noise over the metabolite-free 8-10 ppm range in each voxel.
Due to, for example, dielectric resonances at 3 T, non-negligible B1+ variation were encountered within the VOI. To monitor them, and also demonstrate insensitivity of the exemplary STRESS to them, a 2D Actual Flip Angle imaging sequence was used to obtain B1+ maps at an isotropic 5×5×5 mm3 spatial resolution, with TR1/TR2/TE=20/100/10 ms for a 3:12 minutes total acquisition time.
The exemplary experiments were repeated in a healthy, 31 year old female. A localizer was followed by an about 5:36 minute 3D MP-RAGE (e.g., 256×256×192 mm3 FOV, 1 mm3 isotropic). An in-house auto-shim routine then optimized the whole-head B0 homogeneity in ˜5 minutes to 24 Hz. (See, e.g., Reference 29). This was followed by two 2D CSI from a supra-ventricular slice using PRESS and the exemplary STRESS−110 (e.g., VOI shim: 17 Hz). Acquisition parameters for both were: TR/TE=1000/40 ms, 16×16 matrix, 160×160 mm2 FOV, 80×80 mm2 VOI, 15 mm slice thickness, 4 averages for a total of about 17:08 minutes. As with the phantom above, the about 40 ms TE was the minimum for PRESS, given the amount of spoiling and pulse lengths, although a TE as low as about 26.6 ms was attainable with STRESS−110. A 3D flip angle imaging sequence followed to obtain B1+ maps for the 1H-MRSI slice. The protocol took 49 minutes. Post processing was, for example, identical to the phantom, including the SNR maps.
Sample STRESS−110 and PRESS spectra from a column of voxels through the phantom's center, are shown
Averaging over the middle 5×5 voxels can reveal that the mean STRESS SNR for NAA can be about 85% that of PRESS, while for Cho it can be 86%, in close agreement with the 83% theoretical predictions for STRESS−110 in Table 2. The slight increase can be explained by the finite T2s and B1+ variations. This can be supported by the ±10% variation in the nominal B1+ mapped across the slice.
The exemplary spectra (see, e.g.,
The same sequences were evaluated in a 2D supra-ventricular slice from the brain of a volunteer (see e.g.,
Both the spectra and 1D SNR plots substantially confirm the theoretical predictions, that, for example: (i) the STRESS averaged NAA SNR in the central 4×4 group of voxels can be 83% that of PRESS, in agreement with the predictions of Table 2 and Eq. (1); and (ii) substantial CSD can be observed in PRESS, with Cr and Cho peaks appearing outside the VOI (see, e.g., arrows provided in
The 1D Cr SNR map (see, e.g.,
For example, two exemplary popular 1H-MRS localization modules, for example, PRESS and STEAM, can be designed to acquire a specific coherence pathway, while dephasing all others. By acquiring SE123 and STE− simultaneously, for example, the exemplary system, method and computer-accessible medium according to an exemplary embodiment of the present disclosure, can be used to trade some SNR for significantly reduced TE, CSD and SAR, as well as increased immunity to B1+ variations. As shown in-vivo for Cr, this SNR loss can be expected to be small, approximately 12%, or even be a gain under realistic acquisition conditions at 3 T and up (e.g., short T2s, substantial B1+ inhomogeneities). The exemplary STRESS family of sequences can therefore be thought of as a hybrid STEAM/PRESS sequence, parameterized by the flip angle of the first pulse. Unlike STEAM, which can be designed to refocus either STE− or STE+ (see, e.g., Reference 25), STRESS can refocus STE− since the spoiling gradients may not be adjusted to refocus both SE123 and STE+.
Two exemplary practical considerations can be noted: (i) phase encoding in 1H-MRSI can be applied after the third pulse, as shown in
Simultaneous acquisition of spin and stimulated echoes can be inherent to one of the earlier pulse sequences (e.g., albeit for a different purpose): the Meiboom-Gill variant (see, e.g., Reference 26) of the Carr-Purcell echo train (see, e.g., Reference 32), used in NMR and MRI. A recent coherence pathways analysis pinpointed the refocused STEs as the cause of its immunity to B1+ variations. (See, e.g., Reference 27). A “standard” PRESS sequence cannot benefit from this immunity since STE+ and STE− can be normally spoiled. The exemplary system, method and computer-accessible medium does not spoil the STE− and when α/β/γ=90°/180°/180°, it can be similar to PRESS, but with added benefit of B1+ variation immunity due to the incorporated CPMG condition (see, e.g.,
The exemplary system, method and computer-accessible medium according to an exemplary embodiment of the present disclosure—with its short, high bandwidth pulses—can share many of STEAM's advantages, including small CSD, low SAR and shorter available TEs. While short TEs can also be achieved with PRESS by shortening the duration of 180°s, short refocusing pulses in PRESS can often suffer from much wider transition bands compared to lower flip angle pulses of equal duration. This can severely impact the spectral quality at the VOI's edges in-plane. The 25%-lower SAR of the exemplary system, method and computer-accessible medium facilitates shortening of the TR for (e.g., ˜12% here) higher SNR/unit-time or larger spatial coverage in 3D 1H-MRSI. (See, e.g., Reference 10).
Another exemplary feature common to STEAM and the exemplary system, method and computer-accessible medium, according to an exemplary embodiment of the present disclosure, can be a slightly wider (e.g., approximately 10%) and sharper slice profile, due to the STE− component, compared to PRESS with pulses of similar peak B1 and bandwidths. (See, e.g., Reference 9). For example, even with the same pulses (e.g., 90°-180°-180°), a PRESS VOI can be slightly narrower than the exemplary STRESS−90. To match VOI profiles, the selective gradients can be increased by approximately 10%, yielding additional CSD reduction for the exemplary STRESS−90. Full numerical 3D simulations of the expected VOI profiles were done for the PRESS and the exemplary STRESS−110, and the selection gradients were calibrated to equalize VOI widths.
The exemplary SE123 signal can remain transverse between the second and third pulses, while STE− can be stored longitudinally for 2τ=TE/2. The two pathways, therefore, can differ in their relaxation and evolution during that period, which can create a more complex dependency on T1, T2 and B1+ inhomogeneity than in STEAM or PRESS, as seen by comparing Eqs. (3) and (4). Knowledge of T1 and T2 can be used for absolute quantification in either case. However, B1 dependency in PRESS can be cancelled out by taking metabolite ratios between, for example, NAA and a reference water scan, where, for example:
with E1,NAA=e−TR/T
where e−TE/2T
Its magnitude can depend on TR, TE, T2 and T1 of the two species, η and β. For example, for the exemplary STRESS−100 (e.g., α/β/γ=100°/131°/131°), TE=40 ms, T2, NAA=150 ms, T2, water=80 ms, T1,NAA=1200 ms, T1, water=800 ms, TR=1000 ms and 20% B1+ inhomogeneity (e.g., η=0.8), E=2%. E can rarely exceed 5% for these parameters' range in-vivo, on par with other confounds in ratio quantification. (See, e.g., Reference 33).
Due to the presence of both SE123 and STE− in the signal of the exemplary system, method and computer-accessible medium, spectral fitting can account for the true evolution of the spins, using sequence-specific basis sets. This, however, for example, may not be different from PRESS and STEAM, that can also use customized basis sets for quantification, nor are these sets a priori more complicated. Note, however, that at long TEs, some interference can occur between the SE123 and STE− signals for J-coupled metabolites that may not occur at short TEs, as seen from the lactate and myo-inositol peaks in the phantom (see e.g.,
As T1 and T2 relaxation can be introduced, the maximum of the signal curve (see e.g., Eq. (3)) as a function of the B1+ inhomogeneity, η, can shift slightly from the nominal B1+, such that Eq. (2) may no longer hold true. However, the optimal η for the ranges of B1+, T1 and T2 in vivo was found to be very close to about 1.0, with deviations of only a few percent. Since the exemplary system, method and computer-accessible medium already benefits from significant immunity B1+ variations, this shift can be negligible in practice.
Other sources of artifacts must be kept in mind. For example, a time-dependent resonant frequency drift due to, for example, eddy currents, can affect the transverse SE123 pathway between the second and third pulses, but can leave the longitudinal STE− untouched. Magnets susceptible to eddy currents can, therefore, apply pre-emphasis gradient pulses (see, e.g., Reference 34), or use field mapping for quality assurance. (See, e.g., Reference 35). The exemplary spectra line shapes in
Although the exemplary system, method an computer-accessible medium according to an exemplary embodiment of the present disclosure, can reduce the in-plane mutations considerably, it can be used to increase the first pulse's flip angle, α, compared to PRESS and STEAM, leading to increased CSD in that direction, and higher SAR. However, since α can be kept closer to 90° than 180°, the increases can be small. For example, using a peak B1 of about 1.0 kHz, the FWHM bandwidth of the SLR-designed first pulse of the exemplary STRESS−110 (e.g., α=110°) was about 3.2 kHz, versus about 3.4 kHz of the PRESS SLR 90°; increasing the slice CSD by only about 6%. Although the SAR of the exemplary STRESS' 110°slice selection pulse was about 20% higher than that of the 90° used in PRESS, the exemplary STRESS' total SAR was about 25% lower than PRESS in practice.
Simultaneous acquisition of the spin- and stimulated-echo pathways, according to an exemplary embodiment of the present disclosure, can generate a family of sequences, dubbed STRESS that can combine the advantages of STEAM (e.g., low SAR, CSD, short TE) with the higher signal intensities normally associated with PRESS and the superior B1+ immunity associated with a CPMG train. When α=90°, β=γ=180°, the exemplary STRESS can coincide with PRESS, but can also properly combine the CPMG condition for greater B1+ immunity. When 90°<α<136°, the sequence can sacrifice some of the full spin-echo amplitude for a considerable reduction in SAR and in-plane CSD compared to PRESS. Although the results presented here focused on 1H-MRSI, the exemplary system, method and computer-accessible medium, according to an exemplary embodiment of the present disclosure can be equally and/or similarly applicable and effective for single-voxel 1H-MRS. The modification suggested can be simple, subtle and incorporate easily into existing PRESS or STEAM sequences.
As shown in
Further, the exemplary processing arrangement 902 can be provided with or include an input/output arrangement 914, which can include, for example, a wired network, a wireless network, the internet, an intranet, a data collection probe, a sensor, etc. As shown in
The foregoing merely illustrates the principles of the disclosure. Various modifications and alterations to the described embodiments will be apparent to those skilled in the art in view of the teachings herein. It will thus be appreciated that those skilled in the art will be able to devise numerous systems, arrangements, and procedures which, although not explicitly shown or described herein, embody the principles of the disclosure and can be thus within the spirit and scope of the disclosure. Various different exemplary embodiments can be used together with one another, as well as interchangeably therewith, as should be understood by those having ordinary skill in the art. In addition, certain terms used in the present disclosure, including the specification, drawings and claims thereof, can be used synonymously in certain instances, including, but not limited to, for example, data and information. It should be understood that, while these words, and/or other words that can be synonymous to one another, can be used synonymously herein, that there can be instances when such words can be intended to not be used synonymously. Further, to the extent that the prior art knowledge has not been explicitly incorporated by reference herein above, it is explicitly incorporated herein in its entirety. All publications referenced are incorporated herein by reference in their entireties. Additional exemplary embodiments of the present disclosure are provided and described in the enclosed Appendix and drawing(s).
The following references are hereby incorporated by reference in their entireties.
This application relates to and claims priority from U.S. Patent Application No. 61/974,401, filed on Apr. 2, 2014, the entire disclosure of which is incorporated herein by reference.
This invention was made with Government support under grant from National Institute of Health under Grant numbers NS050520 and EB01015. The Government has certain rights therein.
Number | Date | Country | |
---|---|---|---|
61974401 | Apr 2014 | US |