The present disclosure relates generally to radiological imaging, and more specifically, to exemplary embodiments of a system, method and computer-accessible medium for ultralow dose computed tomography (“CT”) image reconstruction with pre-log shifted-Poisson model and texture-based Markov Random Field (“MRF”) prior.
Currently, CT operates at high X-ray exposure, in terms of high X-ray flux, for example, 250 mAs, and/or relative high X-ray energy, for example, 120 kVp, to produce diagnostic images for a patient of average size and/or weight. For use in an emergency room, where detection is the major task, the radiation exposure can be reduced to 100 mAs or lower. For pediatric imaging, and preventive screening of smokers for detection of early lung cancer, the preferred radiation exposure is as low as necessary to achieve a specific clinical task. Clinical studies using low-dose CT (“LDCT”) screening, where the radiation exposure can be reduced to around 50 mAs, for lung cancer detection have recently shown a 25% reduction of cancer morbidity.
X-ray CT is a widely-used imaging modality. Prior work has been performed to minimize the radiation-associated risk by both hardware and software innovations. For software, statistical modeling for an accurate cost function and iteratively minimizing the cost function for a smooth convergence toward an optimal reconstruction has shown promising performance in maintaining the image quality as compared to the traditional filtered back-projection (“FBP”) reconstruction, while at much lower dose levels. To gain additional benefits from the statistical image modeling (“SIM”), reconstruction methods under the Bayesian theory have been focused on two components: (i) the statistical properties of noise in the acquired data and (ii) an appropriate a priori model for the to-be-reconstructed image.
Compound Poisson statistics for the X-ray counts, and Gaussian distribution for the electronic background noise, are commonly used to model the statistical properties of the data which lead to an intractable likelihood term in the Bayesian framework. (See e.g., References 1-4). In one approach to overcoming the intractable problem, an approximation can be made by replacing the compound Poisson with the Poisson and further replacing the summation of the Poisson and Gaussian distributions as a shift Poisson distribution. (See e.g., References 5 and 6).
Work has been done on the a priori modeling of the to-be-reconstructed image in a concerned application. The a priori image modeling is frequently referred to as regularization or penalty in the SIM reconstruction methods. Generally applicable non-linear priors, such as Huber, have been proposed to preserve edges while de-noising at the same time. (See e.g., References 7 and 8). More recently, efforts have been devoted to incorporating information from the source, such as the images from previous full-dose or diagnostic CT (“FDCT”) scan (see e.g., References 9-12), and complete data sets (see e.g., References 13 and 14).
There are still challenges in obtaining quality radiological images while using lower, and therefore, safer levels of radiation. Thus, it may be beneficial to provide an exemplary system, method and computer-accessible medium for ultralow dose computed tomography (“ULDCT”) image reconstruction with pre-log shifted-Poisson model and texture-based MRF prior which can overcome at least some of the deficiencies described herein above.
An exemplary system, method, and computer-accessible medium for generating computed tomography (“CT”) image(s) of a subject(s) can be provided which can include receiving low dose CT (LDCT) imaging information for the subject(s), where the LDCT imaging information can be based on a radiation dose of less than about 50 mAs, receiving a priori CT image data, and generating the CT image(s) based on the LDCT imaging information and the a priori CT image data. A further CT images(s) can be generated based on the CT image(s) and the a priori CT image data. The a priori CT image data can include a set of Markov Random Field (“MRF”) coefficients derived from high-dose or full-dose CT (FDCT) image data with radiation dose of greater than 100 mAs.
In one embodiment, the a priori CT image data is represented by a plurality of Markov Random Field (“MRF”) coefficients. The radiation dose can be less than about 20 mAs. The radiation dose can be less than about 15 mAs. The CT image(s) or the further CT image(s) can be generated based on a beta value or an incident flux. The incident flux can be based on further CT information generated while a CT scanner was empty. The Beta value can be between 4,000 and 6,000.
In some exemplary embodiments of the present disclosure, a further CT image(s) can be generated based on the low dose CT imaging information. The beta value can be modified prior to generating the further CT image(s). An initial low dose CT image can be generated using the low dose CT imaging information and the incident flux. The CT image can be generated using the initial low dose CT image(s). The low dose CT information can be generated using, for example, a CT scanner. The MRF coefficients can include coefficients related to a number of known tissue types, such as four tissue types, which can include lung, fat, bone, and muscle.
The MRF coefficients can be generated based on the a priori CT information. The MRF coefficients can be generated by segmenting the full dose CT images into four tissue types, where the four tissue types include lung, fat, bone, and muscle. The low dose CT image(s) can be segmented into four tissue types, which can include lung, fat, bone, and muscle. The MRF coefficients can include texture information for a plurality of pixels in a full dose CT image. The CT image(s) can be stored in a storage arrangement using a std::map object computer procedure. The a priori CT image data can be based on a full dose CT image(s) of the subject(s). The a priori CT image data can be based on a further subject(s), and an attribute(s) of the subject(s) can be matched with the attribute(s) of the further subject(s).
These and other objects, features and advantages of the exemplary embodiments of the present disclosure will become apparent upon reading the following detailed description of the exemplary embodiments of the present disclosure, when taken in conjunction with the appended claims.
Further objects, features and advantages of the present disclosure will become apparent from the following detailed description taken in conjunction with the accompanying Figures showing illustrative embodiments of the present disclosure, in which:
Throughout the drawings, the same reference numerals and characters, unless otherwise stated, are used to denote like features, elements, components, or portions of the illustrated embodiments. Moreover, while the present disclosure will now be described in detail with reference to the figures, it is done so in connection with the illustrative embodiments and is not limited by the particular embodiments illustrated in the figures and the appended claims.
The exemplary system, method and computer-accessible medium can provide ULDCT that can achieve the same detection rate of current systems with substantially lower radiation exposure than typical CT systems (e.g., greater than about 50 mAs). For example, the exemplary system, method, and computer-accessible medium, can generate LDCT images using a radiation exposure of about 50 mAs (e.g., about 55 mAs to about 45 mAs). Additionally, the radiation exposure can be less than about 20 mAs (e.g., about 25 mAs to about 15 mAs). Further, images using further reduced radiation exposure levels can also be achieved (e.g., less than about 15 mAs, less than about 10 mAs, or less than about 5 mAs).
As described herein, a high or full dose level can be about 5.0 mSv (e.g., 120 kVp (X-rat tube voltage) and 100 to 300 mAs (X-ray tube current)), a low-dose level can be about 1.5 mSv (e.g. 120 kVp, 30 to 100 mAs), and an ultralow-dose level can be below about 1.0 mSv (e.g. 120 kVP, 20 mAs or less).
The exemplary system, method, and computer-accessible medium, according to an exemplary embodiment of the present disclosure, can accurately model the data acquisition process and all information can be modeled into a theoretically-solid framework. Furthermore, the properties of the majority of normal tissues can be modeled and the acquired data can be shifted to show the properties of a very small portion of abnormal tissues. The data can be acquired in a task-dependent manner to obtain maximum information to achieve the task. The exemplary ULDCT can include (i) more accurate modeling of the data and acquisition process, and then incorporating the data properties more robustly into a theoretically solid framework and (ii) the properties of the normal tissues (e.g. fat, muscle, lung, bone, etc.), which can be a priori known and can be modeled accurately into the theoretically-solid framework. Thus, the acquired data can be processed to show the properties of a very small region where abnormality or pathologically altered issues can occur.
A tractable shifted Poisson model (see e.g., Reference 15) for data statistics can be combined with an exemplary texture-based Markov random field (“MRF”) extracted from previous full-dose CT scans (see e.g., Reference 12) to gain full power of SIM reconstruction for ultra-low dose CT imaging.
Acquired data from a polychromatic X-ray CT system can be denoted as a vector I∈M×3, where M can be the number of data elements. The effective attenuation map of an object can be denoted by a vector μ∈N×1, with N being the dimension of lexicographically ordered pixels. With X-ray incident flux (e.g., together with detector efficiency) being {Ii0}, where index i can indicate the X-ray toward detector bin i, then the X-ray signals reaching a detector bin i can be expressed as Ii0e−[Aμ], in an average sense (e.g., Ii0 can be determined accurately by system calibration in the absence of body) according to Beer's law. Matrix A can model the linear system relationship, and can be called a projection matrix of size M×N with its element denoted by Amn, and row vector denoted by Am. The acquired data include an additional Gaussian electronic noise, which can assume a mean me,i, and a variance σe,i2, predetermined at each detector bin i.
Exemplary Texture-Preserving Low-Dose Ct Image Reconstruction with Shifted-Poisson Data MRF Texture Model (SP-MRFt)
An artificial random vector can be determined as, for example:
where the mean me and variance σe 2 of the background noise can be assumed to be the same for all detector bins for simplicity.
It can be shown that IA can have its mean and variance both equal to Ii0e−[Aμ]
It can be beneficial to enforce noise reduction in the case of low dose high noise situations. Texture-based MRF regularization can efficiently incorporate information from previous full-dose images, and can gain effective noise reduction with minimal cost of high-resolution details. (See e.g., Reference 12). Thus, for example:
where, Ω(n) can denote neighborhood of pixel n and r indexes a region of certain tissue. Within each region, wnn′r can be shift-invariant MRF coefficients depending on neighboring relationships. Regions can be segmented and classified based on different texture properties. These MRF coefficients can be determined according to previous full-dose images under the principle of, for example:
where, the superscript “FD” can mean full-dose. Eq. (4) can be quadratic, and a close form solution can be obtained as, for example:
where, μΩFD can denote a lexicographically ordered neighboring pixels around pixel n for notation convenience, and wr a corresponding vector of MRF coefficients for region r.
Combining Eqs. (2) and (3) can provide an overall objective function to minimize in the framework of a Bayesian MAP (e.g., maximum a posterior) image reconstruction. Thus, for example:
where β can be a parameter of balancing the data fidelity term of Eq. (2) and the priori term of Eq. (3).
The objective function can be composed of a non-quadratic part and a quadratic part. The situation of low-dose data acquisition can make the condition of the non-quadratic part worse (e.g., very small condition number). Also, it can be a non-separable function of a large number of unknown variables μ. In response, its separable surrogate function can be determined.
As shown in Eq. (2), for each ray path, a function hm(lm) can be denoted as, for example, hm(lm)=(Im0e−+σe2)−ImA ln(Im0ee+σe2) with its 1st derivative being:
In the kth iteration, a global surrogate function for the likelihood term in Eq. (6) can be, for example:
where, μk can denote the current estimation, cm can be chosen to ensure monotonic decreasing. Thus, for example:
Because the function for each ray can be quadratic, for example, convex, Eq. (7) can be further relaxed to a separable function by using convexity property based on, for example:
By choosing
the following can be obtained:
For the regularization term R(μ), because it can be quadratic, a separable surrogate function can be obtained using the following:
(μn−μn′)2≤½(2μn−μnk−μn′k)2+½(2μn′−μnk−μn′k)2,
such that
Eqs. (8) and (9) together can provide an overall separable surrogate function, which can include, for example:
According Newton's algorithm, a parallelizable update formula for the exemplary ultralow-dose image reconstruction procedure can be obtained as, for example:
The implementation of the above SP-MRFt procedure can include:
A set of MRF coefficients corresponding to the lung, bone, fat, and muscle can be generated for each patient image in a database. For example, a database of 1,000 images of 1,000 patients can include generated MRF coefficients for each of the 1,000 patients. Thus, such a database can have 4,000 sets of tissue parameters (e.g., 4 sets of parameters for each of bone, lung, muscle, and fat for each of the 1,000 patients).
At MRF prediction procedure 130, beta can be set to zero. Thus, the MRF term R(w) may not have any effect to the ULDCT image reconstruction of Eq. (6). In the subsequent iterative steps, the beta can be set to a non-zero value such that the stored MRF coefficients (w) can be included in the reconstruction by Eq. (6). The FDCT image can be separated into the four tissue regions (e.g., four tissue masks). Pixels from these images can be used to compute the MRF tissue coefficients. The generated coefficients can be arranged into a 7×7 matrix. For each patient chest, there can be four matrices for lung, fat, muscle, and bone, respectively. When the ULDCT image is constructed at a later time using by Eq. (6) at the kth iteration, the tissue mask of the ULDCT image from the (k−1)-th iteration can be used. Thus, Eq. (6) can be applied to each tissue mask of ULDCT image domain where the corresponding MRF coefficients are included in Eq. (6).
In order to compare neighboring pixels, a 3D volume dataset can be used. Each element in the array can be a cubic voxel. In a 2D image slice, each element can be a square area (e.g., a pixel). Each pixel has a value referred to as an image intensity, and the intensity can be used to determine the MRF coefficients. For example, the intensity differences between from neighboring pixels can be fit into a plurality of equations that can be used to determine the coefficients. The number of equations can be based on the matrix being fit into. Thus, for example, a 7×7 matrix can be generated using 49 equations.
The set of MRF coefficients can be different for each of the four tissue categories (of lung, fat, muscle, and bone), and can be based on an autocorrelation between neighboring pixels (in each tissue type). Each set of coefficients can be fit into a matrix of a particular size depending on the number of coefficients. For example, the MRF coefficients for each region can be lit into a 7×7 matrix. The number of parameters can be chosen by empirically testing neighbors covered by a matrix size of 3×3, 5×5, 7×7, 9×9, or 11×11 format. A 7×7 matrix can be sufficient, because smaller than a 7×7 matrix may not have sufficient information to generate the low dose CT images and greater than a 7×7 may not add sufficient new information to generate a superior image, especially when factoring in the increased processing time that can result from a larger matrix.
For example, given four tissue masks 190 of the LDCT or ULDCT image, corresponding tissue MRF coefficients can be obtained (e.g., received) from the MRF spectrum database 155 to construct the second term of Eq. (6). In particular, if a patient has previously had a full dose CT image performed, that previous full dose image will be in a database, and a set of MRF coefficients particular to that patient based on the full dose image can be previously created. Thus, the set of MRF coefficients used to reconstruct the low dose image for a particular patient can be specifically based on that patient's prior full dose image. However, if a patient has not have a prior full dose image generated, the attributes of the patient can be matched to the attributes of another patient that already had a full dose image to use the corresponding MRF coefficients. Attributes can include, but are not limited to, age, sex, height, weight, body mass index, position during scan (e.g., comparing position of low dose image of one patient to the position of the high dose image of a matching patient). Other suitable attributes to more accurately match a patient can be used. Once a match is found in a database, the specific MRF coefficients for the patient with the prior full dose image can be used as the coefficients for the patient without the full dose image.
The first image for a patient can be constructed by the LD sinogram 165 and the incident flux 170. In this iteration, beta can be set to a suitable non-zero value. The data of obtained from procedures 190, 155, 160, 165 and 170 can be input into the pre-log procedure 175 (e.g., of Eq. (6)). From the pre-log operation (e.g., or minimization of Eq. (6)), an updated LDCT or ULDCT image 180 can be generated or obtained. If this updated image is similar to the previously-iterated image, the iterative procedure can end. If the image is not similar, or not sufficiently similar, a further segmentation 185 can be performed on this updated LDCT or ULDCT image for its corresponding tissue masks 190, and the method 150 can be repeated starting at procedure 175 to further update the image. Thus this iterative procedure can repeat until the current iterated result is sufficiently similar to the previously-iterated result. During the iterative procedure, the beta can increase from small (e.g., 1.0) up to 5,000 or 6,000. The beta value 160 can be an empirically determined value, and can be modified (e.g., increased after the initial run) for each iteration of the method 150. Incident flux 170 can be baseline or calibration information obtained using the CT scanner used to generate low dose sinogram information 165 without the patient in the scanner. Thus, incident flux 170 can be used, for example, to eliminate background noise, or any noise or signals introduced by the scanner itself. The incident flux can be determined immediately prior to, or after, the low dose CT scan of the patient. Alternatively, or in addition, the incident flux can be periodically determined irrespective of a scan of the patient.
After the initial set of low dose CT images 180 are generated, these images can be segmented into the four tissue categories (e.g., lung, fat, muscle, and bone), substantially as described above, to generate a tissue mask 190. However, because the procedure 150 is based on the low dose sinogram information 165, the initial set of low dose CT images 180, and the resulting tissue mask can be of a low quality. In order to increase the quality of the generated images a priori data from high dose CT Image data, for example the MRF coefficients 135 generated in method 100, can be used. For example, the MRF coefficients can be stored in a MRF spectrum database. After the initial generation of the low dose images 180 and the resulting tissue mask 190, method 150 can be repeated (e.g., iterated) using the MRF coefficients stored in MRF spectrum database 160 and the low dose sinogram imaging information 165. Additionally, the beta value 160 and/or the incident flux 170 can also be used in the additional iterations. The beta value 160 can be modified from the initial generation of the CT images, and the beta values 160 can also be modified in any subsequent iterations of method 150. For example, the value of beta can increase by 10 in each iteration, although other increments may be used (e.g., by 20, by 50, by 100, by 1,000 etc.). The number of iterations of method 150 can depend on the quality of the generated low dose CT images 180 and the resulting tissue mask 190. Once a sufficient image quality has been achieved, method 150 can end and no additional iterations are needed. The image quality can be determined subjectively by a person evaluating the images or the quality can be automatically determined objectively using a computing arrangement. For example, the density change between images generated at each iteration can be determined. If the density change is below a particular value (e.g., 10%, 5%, 1%, etc.) then the method can end.
The exemplary image reconstruction (e.g., pre-log procedure 175 shown in
In order to speed up the exemplary pre-log image reconstructions, the exemplary system, method, and computer-accessible medium can include various strategies to decrease the image reconstruction time. For example, one strategy can be to save the projection matrix Aij into the memory for all iterations instead of calculating it at each iteration. The projection matrix can have be large, with millions of elements, and would can take up a few hundreds gigabytes (“GB”) memory space. Thus, accessing all the elements can be time consuming (e.g., in many hours for one iteration). In the exemplary LDCT reconstruction, many of the elements are zero, and only a small fraction (e.g., less than about 5%) of the elements have non-zero elements. By using a std::map object computer operation, the non-zero elements can occupy only about 5 GB of computer space. ((See e.g., Reference 20). Accessing the reduced number of elements and memory space can save approximately 67% of time at each iteration. Additional decreases in processing time can be achieved by the use of the body mask. Only those image pixels inside the body mask may be updated instead of all pixels in the image space. Thus, 45% pixels (e.g., outside the body mask) may not need to be iterated; this can reduce the computing time by 10 to 10% per iteration. Further, when updating μjnew by Eq. (6), Aij can be grouped according to the angular sampling around the body, into a few sub-groups in order to speed up the convergence. This updated process (e.g., referred to as the Gauss-Seidel procedure) can be performed for each angular sample or view. If the iteration number can be reduced to a half, the reconstruction time can be cut a half. Additionally, parallel computing with standard computer operation, Open Multi-Processing (MP), can be used to speed up the computing time per iteration, because the updating on one pixel by Eq. (6) may not affect the updating on another pixel. This can save another 67% of time for each iteration. . . .
The exemplary reconstruction procedure was validated on patient ultra-low dose (“ULD”) CT data with both numerical simulation and artificial ULD data based on real data. In the numerical simulation, a case with 2000 incident photons and standard deviation=30 for electronic noise was modeled. A 2D image of a chest phantom 205 from NCAT as shown in
Tables III and IV below illustrate the results from a lung nodule detection for FBP, spatially-invariant (e.g., Huber), and the exemplary spatially-variant a priori knowledge-based Bayesian image reconstruction (e.g., which can be based on a texture) for primary nodules (Table III) and small nodules (Table IV). The detection gain of nodules is statistically significant, compared to the FBP and Huber methods.
Tables V and VI below illustrate a statistical analysis of lung nodule detection. A comparison was performed among FBP, spatially-invariant (e.g., Huber), and the exemplary spatially-variant a priori knowledge-based Bayesian image reconstruction (e.g., which can be based on a texture) for primary nodules (Table V) and small nodules (Table VI). The detection gain of nodules is statistically significant when compared to the FBP and Huber methods.
In the SP-MRFt implementation, MRF coefficients for lung, bone, fat, and muscle are computed from FBP reconstruction of the full-dose data. (See e.g.,
Ultralow-dose CT imaging is desired in the field. However, artifacts of ultralow-dose images can be caused by the rays of very weak signals. The exemplary system, method, and computer-accessible medium, according to an exemplary embodiment of the present disclosure, can include an iterative reconstruction procedure based on the pre-log shifted Poisson statistical model together with a texture-based MRF prior constraint for ultralow-dose CT imaging to mitigate this problem. The pre-log shifted Poisson model can characterize the rays with photon starvation problem and provide sufficient reconstructions. Moreover, it can easily combine the texture-based prior to further reduce noise-induced artifacts with little lose in resolution.
Table VII below illustrates a statistical analysis of lung nodule characterization. A comparison was performed for FBP, spatially-invariant (e.g., Huber), and the exemplary spatially-variant a priori knowledge-based Bayesian image reconstruction (e.g., which can be based on a texture). The characterization gain of nodules is statistically significant when compared to the FBP and Huber methods.
At procedure 825, a further low dose CT image can be generated using, for example, the a priori MRF coefficients, a beta value, the low dose CT imaging information, an incident flux and the initial low dose CT image. At procedure 830, this further low dose CT image can be segmented into the one of four tissue types. At procedure 835, the beta value can be modified (e.g., increased), and an additional low dose CT image can be generated based on the MRF coefficients, the modified beta value, the low dose CT imaging information, the incident flux and the further low dose CT image. At procedure 845, the additional low dose CT image can be segmented. At procedure 850, a determination can be made as to whether the generated image is of a sufficient quality. If the image is of a sufficient quality, then method 800 can end. If the image is not of a sufficient quality, then method 800 can be iterated starting at procedure 835 by modifying the beta value (e.g., increasing the beta value), and then performing procedures 840 and 850, and performing an image quality check again at procedure 850.
At procedure 885, a low dose CT image can be generated using, for example, the a priori MRF coefficients, where the beta value is greater than zero. At procedure 890, a determination can be made as to whether the generated image is of a sufficient quality. If the image is of a sufficient quality, then method 860 can end. If the image is not of a sufficient quality, then method 860 can be iterated starting at procedure 875 by modifying the beta value (e.g., increasing the beta value), and then performing procedures 880 and 885, and performing an image quality check again at procedure 890.
As shown in
Further, the exemplary processing arrangement 905 can be provided with or include an input/output ports 935, which can include, for example a wired network, a wireless network, the internet, an intranet, a data collection probe, a sensor, etc. As shown in
Exemplary Mathematical Derivations and Analysis
Pre-Log Data Measurement Model
Let the measurement of X-ray detection in computed tomography (CT) be expressed as, for example:
NiM≈Possion{
where the measured data {NiM} can include two parts: (i) the X-ray photon counts, which can follow Poisson statistics with mean
If the Poisson and Gaussian variables are assumed to be statistically independent from each other, the mean and variance of the measured datum, NiM, at each detector bin can be derived as, for example:
A particular quantity or random variable NiA can be added, which can produce, for example:
which can be used to prove that the artificial variable can have its variance equals to its mean. Thus, for example:
An assumption can be made that if a random variable has its variance equals its mean; the random variable can follow Poisson statistics. Thus, for example:
where the condition of NiA=NiM−me+σe2>0 can be satisfied and can be noted by [NiM]+ in Eq. (18).
The corresponding log likelihood function can be expressed, assume that all data {NiA}, i=1, 2, 3, . . . , can be statistically independent from each other (where notation [ ]+ can be ignored for notation simplicity), as, for example:
Exemplary Pre-Log Data Shifted Poisson Model+Texture Prior for Bayesian Ultralow-Dose CT Image Reconstruction
Log-Likelihood with Shifted Poisson Model (with Bear Law)
By the Bear Law of
The maximum likelihood (“ML”) solution can be given by, for example:
Exemplary Objective Function with Texture Prior (by Bayesian Theorem)
For simplicity, let L(μ)=L(NA|μ). By Bayesian theorem, the overall objective function can be, for example:
where β can be the adjustable parameter of balancing the two terms of data fidelity term, L(μ), and prior term, R(μ). The prior term can specify that the body can be composed of regions (with index r) and each region can contain a tissue type. In each tissue type region, a neighborhood Ω(.) can be selected, which can be a square or cubic window including the pixels or voxels surrounding the central pixel or voxel. The maximum a posterior probability (“MAP”) solution can be given by, for example:
Exemplary Numerical Optimization Using the Surrogate Function Approach
The following two exemplary expressions can be utilized:
where li=[Aμ]i can represent the line integral of the attenuation coefficient along projection ray from X-ray tube/source toward detector bin i, as defined above.
Surrogate Function for Data Fidelity Term L(μ)
For each ray path, the following definitions can be used:
The Global surrogate function for data fidelity term can be, for example:
The Global surrogate function can be, for example:
The Separable surrogate for data fidelity term can be, for example:
where
because qi(•;lin) can be convex.
Selecting a specific
can lead to an additive form which can converge more quickly. Thus, the separable surrogate function can be, for example:
Exemplary Separable Surrogate for Prior Term
Utilizing
the separable surrogate function for prior term can be, for example:
which can be in the form of quadratic penalty: φ(μj−μ2)=½(μj−μk)2.
Exemplary Overall Separable Surrogate Function
Combining the above surrogate functions from both data fidelity and prior terms can provide, for example:
According Newton's algorithm, the following equations can be used to calculate the MAP solution:
The resulting iterative calculation algorithm can be expressed as, for example:
While the following strategy is based on the assumption that the patient has both the previous full-dose CT scans, (from which the tissue-specific MRF weighs are extracted), and the current ultralow-dose scans, in practice many of the full-dose CT scans can be collected as a database and can be used for a current ULDCT image reconstruction if this current patient does not have the previous FDCT scan. This iteration method, includes procedures which are pre-reconstruction operations for setting up the tissue texture properties; a procedure for initializing the ultralow-dose reconstruction by two alternative choices; a procedure for updating the ultralow-dose reconstruction by a ML (e.g., maximum likelihood) procedure, for example, setting the beta being zero in Eq. (6), until the image can be reasonably segmented for assignment of the MRF weights in each tissue region; and then performing the MAP (e.g., the beta value be non-zero in Eq. (6)) reconstruction.
The assumption that the patient has the previous full-dose CT scans may not be satisfied in many cases. In such cases, the exemplary system, method, and computer-accessible medium, can modify certain procedures as follows:
From the current patient's physical information, such as body mass index (“BMI”), sex, age, etc., a similar full-dose CT scan can be searched for (e.g., from a FDCT database) to perform the initial procedures on to extract the MRF weights, followed by the remaining procedures. The “similar full-dose CT scan” can be a single scan from a person or a set of scans from a group people who have similar physical information.
Newton's Algorithm
Newton's algorithm can be used to obtain the zero point of one function by using the zero point of its tangential line at n-th iteration. It can be beneficial to obtain the zero point of the overall surrogate function, which can be, for example:
Tangential to
at μjn (e.g., the tangential line), can provide, for example:
μjn can be updated to μjn+1 using the μj where the tangential crosses zero (e.g., the zero point of tangential line). Thus, for example:
Exemplary Data Fidelity Term
Therefore, for example:
With a data fidelity term of
The following can be provided:
Using a prior term of
can produce, for example:
Thus, for example:
Which can also include a prior term of
which can produce, for example:
In the exemplary case, the prior term can be tissue-specific MRF model, particularly the tissue texture of each tissue type is incorporated, where wjk can be a shift-variant coefficient. As for the voxel at the boundary, there can be overlap. For the voxels inside the region,
where the superscript index FD can indicate that the CT image was acquired at full- or normal-dose level, and the solution for the MRF weights are, for example:
There can be two choices for the coefficient of ci(lin). One is fixed and the other is iterated.
For the fixed case, ci(lin) can yields, for example:
q(l,lin)=h(lin)+h′(lin)(l−lin)+½ci(lin)(l−lin)2>h(l),∀l∈[0,+∞) (51)
h(l) can be expanded at ln, which can be, for example:
h(l)=h(lin)+h′(lin)(l−lin)+½hn(lin)(l−lin)2 (52)
if only the second order is considered. Thus, the following can be chosen:
While, this kind of choice can be too conservative to make the convergence rate slow, it is simple to make the coefficient constant.
The coefficient can also be updated at each iteration, which may not add to the computation time too much, but can decrease the iteration procedures significantly as follows:
The foregoing merely illustrates the principles of the disclosure. Various modifications and alterations to the described embodiments will be apparent to those skilled in the art in view of the teachings herein. It will thus be appreciated that those skilled in the art will be able to devise numerous systems, arrangements, and procedures which, although not explicitly shown or described herein, embody the principles of the disclosure and can be thus within the spirit and scope of the disclosure. Various different exemplary embodiments can be used together with one another, as well as interchangeably therewith, as should be understood by those having ordinary skill in the art. In addition, certain terms used in the present disclosure, including the specification, drawings and claims thereof, can be used synonymously in certain instances, including, but not limited to, for example, data and information. It should be understood that, while these words, and/or other words that can be synonymous to one another, can be used synonymously herein, that there can be instances when such words can be intended to not be used synonymously. Further, to the extent that the prior art knowledge has not been explicitly incorporated by reference herein above, it is explicitly incorporated herein in its entirety. All publications referenced are incorporated herein by reference in their entireties.
The following references are hereby incorporated by reference in their entireties.
The present application relates to and claims priority from International Patent Application No. PCT/US2018/037003 filed on Jun. 12, 2018 which published as International Publication No. WO 2018/231757 on Dec. 20, 2018 and claims the benefit of U.S. Provisional Patent Application No. 62/518,282, filed on Jun. 12, 2017, the entire disclosures of which are incorporated herein by reference.
This invention was made with government support under Grant Nos. CA206171 and CA143111, awarded by the National Institutes of Health. The government has certain rights in the invention.
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/US2018/037003 | 6/12/2018 | WO | 00 |
Publishing Document | Publishing Date | Country | Kind |
---|---|---|---|
WO2018/231757 | 12/20/2018 | WO | A |
Number | Name | Date | Kind |
---|---|---|---|
6950493 | Besson | Sep 2005 | B2 |
6999549 | Sabol et al. | Feb 2006 | B2 |
7471765 | Jaffray et al. | Dec 2008 | B2 |
7627079 | Boone | Dec 2009 | B2 |
7689017 | Karl et al. | Mar 2010 | B2 |
7826587 | Langan et al. | Nov 2010 | B1 |
7872235 | Rousso et al. | Jan 2011 | B2 |
7911474 | Li et al. | Mar 2011 | B2 |
8160200 | Tkaczyk et al. | Apr 2012 | B2 |
8416914 | Thibault et al. | Apr 2013 | B2 |
8467497 | Lu et al. | Jun 2013 | B2 |
8483463 | Chen et al. | Jul 2013 | B2 |
8507869 | Asma et al. | Aug 2013 | B2 |
8538776 | Reiner | Sep 2013 | B2 |
8577115 | Gering et al. | Nov 2013 | B2 |
8611626 | Miao et al. | Dec 2013 | B2 |
8615118 | Yi et al. | Dec 2013 | B2 |
8712121 | Wiegert et al. | Apr 2014 | B2 |
8812240 | Yu et al. | Aug 2014 | B2 |
8835858 | Volokh et al. | Sep 2014 | B2 |
8862206 | Wang et al. | Oct 2014 | B2 |
9129044 | Shih et al. | Sep 2015 | B2 |
9237874 | DeMan et al. | Jan 2016 | B2 |
9251606 | Liang et al. | Feb 2016 | B2 |
9373159 | Amroabadi et al. | Jun 2016 | B2 |
9558570 | Liang et al. | Jan 2017 | B2 |
9592022 | Larson | Mar 2017 | B2 |
9700264 | Taguchi et al. | Jul 2017 | B2 |
10092253 | Sodickson et al. | Oct 2018 | B2 |
10121267 | Lin et al. | Nov 2018 | B2 |
10463317 | Tian et al. | Nov 2019 | B2 |
10638993 | Yun et al. | May 2020 | B2 |
10726587 | Zhao et al. | Jul 2020 | B2 |
10765890 | Sun et al. | Sep 2020 | B2 |
10792006 | Zhu et al. | Oct 2020 | B2 |
11062489 | Chen et al. | Jul 2021 | B2 |
20130202177 | Bar-Aviv | Aug 2013 | A1 |
20150131883 | Taguchi et al. | May 2015 | A1 |
20150196265 | Suzuki | Jul 2015 | A1 |
20150287223 | Bresler | Oct 2015 | A1 |
20160078782 | Meidenbauer | Mar 2016 | A1 |
20160328842 | Ye et al. | Nov 2016 | A1 |
20170100078 | Han | Apr 2017 | A1 |
20180228460 | Singh | Aug 2018 | A1 |
20180240219 | Mentl | Aug 2018 | A1 |
Entry |
---|
Kubo et al. (“Low dose chest CT protocols (50mAs) as a routine protocol for comprehensive assessment of intrathoracic abnormality”, European Journal of Radiology Open 3 (2016) 86-94, Apr. 2016). (Year: 2016). |
Low dose chest CT protocol (50 mAs) as a routine protocol forcomprehensive assessment of intrathoracic abnormality (European Journal of Radiology Open 3 (2016) 86-94) (Year: 2016). |
Notification of transmittal of the International Search Report and the Written Opinion dated Sep. 26, 2018 for International Application No. PCT/US2018/37003. |
Zhang, H. et al., “Extracting Information from previous full-dose CT scan for knowledge-based Bayesian reconstruction of currentlow-dose CT Images,” IEEE Trans Med Imaging, vol. 35, No. 3, pp. 860-870, Mar. 2016. |
Xu et al., “Low-Dose X-Ray CT Reconstruction via Dictionary Learning” IEEE Transaction of Medical Imaging, vol. 31, No. 9, pp. 1682-1697, Sep. 2012. |
Wang et al., “Hybrid Pre-Log and Post-Log Image Reconstruction for Computed Tomography” IEEE Transactions on Medical Imaging, vol. 36, No. 12, pp. 2457-2465, Dec. 2017. |
Rui et al., “Ultra-low dose CT attenuation correction for PET/CT: Analysis of Sparse view data acquisition and Reconstruction algorithms” Phys Med Biol, vol. 60, No. 19, pp. 7437-7460, Oct. 7, 2015. |
Wang et al., “An Experimental Study on the Noise Properties of X-ray CT Sinogram Data in Radon Space,” Physics in Medicine and Biology, 2008, 53(12): 3327-3341. |
Snyder et al., “Compensation for readout noise in CCD images,” Journal of Optical Society of America A, 1995, 12(2): 272-283. |
P J Rivière, “Penalized-likelihood sinogram smoothing for low-dose CT,” Medical Physics, 2005, 32(6): 1676-1683. |
Wang et al., “Iterative image reconstruction for CBCT using edge-preserving prior,” Medical physics, 2009, 36(1): 252-260. |
Whiting et al.,“Signal statistics of X-ray Computed Tomography,” Proc. SPIE Medical Imaging, 2002, 4682: 53-60. |
Elbakri et al., “Efficient and accurate likelihood for iterative image reconstruction in x-ray computed tomography,” Proc. SPIE Medical Imaging, 2003, 5032: 1839-1850. |
Lasio et al., “Statistical reconstruction for x-ray computed tomography using energy-integrating detectors,” Physics in Medicine and Biology, 2007, 52(8): 2247. |
Little et al., “Sinogram restoration in computed tomography with an edge-preserving penalty,” Medical Physics, 2015, 42: 1307-1320. |
Nett et al., “Low radiation dose C-arm cone-beam CT based on prior image constrained compressed sensing (PICCS): Including compensation for image vol. mismatch between multiple data acquisitions,” Proc. SPIE Medical Imaging, 2009, 7258: 725-803. |
Ma et al., “Low-dose CT image restoration using previous normal-dose scan,” Medical Physics, 2011, 38: 5713-5731. |
Shen et al., “Multi-energy CT acquisition and reconstruction with a stepped tube potential scan,” Medical Physics, 2015, 42(1): 282-296. |
Ouyang et al., “Noise reduction in low-dose cone beam CT by incorporating prior volumetric image information,” Medical Physics, 2012, 39: 2569-2577. |
Chen et al., “Prior image constrained compressed sensing (PICCS): A method to accurately reconstruct dynamic CT images from highly under-sampled projection data sets,” Medical Physics, 2008, 35: 660-663. |
Yavuz et al., “Statistical image reconstruction methods for randoms-precorrected PET scans,” Medical Image Analysis, 1998, 2(4): 369-378. |
Erdogan et al., “Monotonic algorithms for transmission tomography,” The 5th IEEE EMBS International Summer School, Biomedical Imaging, pp. 1-5, 2002. |
Erdogan et al., “Monotonic algorithms for transmission tomography,” IEEE Transactions on Medical, 1999, 18(9): 801-814. |
Number | Date | Country | |
---|---|---|---|
20200126271 A1 | Apr 2020 | US |
Number | Date | Country | |
---|---|---|---|
62518282 | Jun 2017 | US |