Interest in epicardial (outer wall of the heart) treatment of ventricular cardiac arrhythmias has grown significantly within electrophysiology. The thickness of the myocardial wall makes it difficult to treat all heart rhythm problems endocardially (from inside the heart). Although early (pre-reperfusion era) data suggested that epicardial ventricular tachycardia (VT) occurred in a minority of patients, recent (non-ischemic VT and rapid reperfusion era) data suggests that about 70% of VT patients have epicardial substrates for the disease. Several studies have suggested that epicardial ablation procedures are not only a viable second line of defense when endocardial methods fail, but that epicardial procedures should be performed in concert with all endocardial treatments to guarantee the greatest probability of treatment success of both ventricular tachycardia (which kills 500,000 Americans per year) and atrial fibrillation (which is the largest cause of strokes in the U.S.). The epicardial surface is also considered to be an important potential therapeutic location for drug and cell delivery and as well as for heart failure therapy.
The ability to gain minimally invasive access to the heart's outer wall for ablation and other therapies has done much to promulgate the adoption of epicardial strategies.
However, conventional types of guidance methods lead to an unacceptable and high risk of perforations of the RV tissue. For instance, Sosa and Scanavacca had an initial perforation rate of 8%, which decreased to 4.5% with experience. Others using this approach have experienced 12% unsuccessful pericardial access, with pericardial effusion in 13 of 35 cases and one death from complications. Still another study suggested that 10 to 20% of all patients undergoing pericardial access for epicardial ablation using the methods described above experienced effusions due to some level of ventricular perforation.
Therefore, it is a goal of an aspect of an embodiment of the present invention to, among other things, improve the method of access and reduce the risks of its use.
Aspects of the present invention may find applicability in systems and methods such as those described in, for example, epicardial electrophysiology and other procedures in which the location of an access needle may be inferred according to the detection of different pressure frequencies in separate organs, or different locations, in the body of a subject.
An aspect of an embodiment of the present invention may provide, among other things, continuous measurements of the pressure-frequency characteristics across the different regions of the thorax. As discussed further herein, data of that type may be used, for example, to determine how the thoracic pressure-frequency waveform changes upon gaining proximity of and then traversing through the parietal pericardial membrane. The inventors have found that it can further be determined if the transition to a two-component (intubation+heartbeat) signal was smooth and gradual, or more abrupt in nature.
Aspects of the present invention may further provide, among other things, improved instrumentation to collect continuous segments of pericardial and non-pericardial pressure-frequency data in vivo, in contrast to the discrete-location measurements. In particular, an exemplary approach incorporates a high precision fiber-optic sensor into the distal tip of the access needle, which may replace the strain gauge sensor acted on by a fluid column within a stationary sheath in an alternative embodiment.
Further aspects of the invention may provide, among other things, an analysis algorithm, method, technique and system designed to process pressure-frequency data so as to identify when the needle's tip had safely entered, for example, the pericardial space.
An aspect of an embodiment of the present invention may, among other objects, reduce the clinical risks associated with minimally invasive subxiphoid access, and thus improving the reliability, safety and efficacy of it in the epicardial treatment of cardiac arrhythmias and other clinical treatment paths involving the pericardium and epicardial surface.
According to aspects of the invention, methods for inferring the location of a needle in a subject may include one or more steps of inserting a needle including a first sensor into a body of a subject, receiving cardiac waveform information of the subject from a second sensor, and receiving pressure frequency information from the first sensor. Embodiments may include distinguishing, by a computer processor, a current location of the needle from another location and/or distinguishing the transition of the needle from a first location to a second location, based on an algorithm including the pressure frequency information and the cardiac waveform information.
Embodiments may include determining a reference phase based on the cardiac waveform information and/or determining a test phase based on the pressure frequency information. In embodiments, the distinguishing of location, and/or movement, of the needle may be based on the pressure frequency information from the test phase and the cardiac waveform information from the reference phase.
In embodiments, the distinguishing of location, and/or movement, of the needle may include comparing the algorithm results of the pressure frequency information and the cardiac waveform information to a plurality of predetermined threshold values.
In embodiments, the combining of the pressure frequency information and the cardiac waveform information may include phase sensitive detection and matched filtering.
In embodiments, the combining of the pressure frequency information and the cardiac waveform information may include integrating the pressure frequency information.
In embodiments, the reference phase may be a cardiac phase of the subject immediately preceding the test phase. In embodiments, the cardiac waveform information may include information derived from a plurality of cardiac phases of the subject.
In embodiments, the cardiac waveform information may include at least one of ventricle pressure, arterial pressure, pulse oximetry, and electrocardiogram (ECG) signals.
In embodiments, the current location may be a non-pericardial location and the other location may be a pericardial location, or vice-versa. In embodiments, at least one of the current location and the other location may be a thorax of the subject.
Embodiments may also include distinguishing between a location remote from the pericardium, a location close to or in contact with the pericardium, and a location inside the pericardium.
Embodiments may also include identifying a location within ventricular tissue of the subject or inside the interior of the heart of the subject.
According to further aspects of the invention systems for accessing one or more locations of a subject may also be provided. Such systems may include, for example, a needle having a distal end and a proximal end and a first sensor in communication with the needle for sensing pressure frequency in the one or more locations of the subject. Embodiments may further include a processor configured to perform various steps, such as those discussed above. For example, in embodiments, a processor may be configured to receive cardiac waveform information of the subject from a second sensor and receive pressure frequency information from the first sensor. The processor may be further configured to distinguish a current location of the needle from another location, and/or the movement of the needle from one location to another, based on an algorithm including the pressure frequency information and the cardiac waveform information.
The processor may be further configured to determine a reference phase based on the cardiac waveform information, and/or determine a test phase based on the pressure frequency information. The processor may be further configured to distinguish the location and/or movement of the needle based on the pressure frequency information from the test phase and the cardiac waveform information from the reference phase.
In embodiments, the processor may be configured to compare the algorithm results of the pressure frequency information and the cardiac waveform information to a plurality of predetermined threshold values.
In embodiments, the processor may be configured to combine the pressure frequency information and the cardiac waveform information using phase sensitive detection and matched filtering.
In embodiments, the processor may be configured to integrate the pressure frequency information as part of the combining of the pressure frequency information and the cardiac waveform information.
In embodiments, the processor may be configured such that the reference phase is a cardiac phase of the subject immediately preceding the test phase.
In embodiments, the processor may be configured such that the cardiac waveform information includes information derived from a plurality of cardiac phases of the subject.
In embodiments, the processor may be configured to receive cardiac waveform information including at least one of ventricle pressure, arterial pressure, pulse oximetry, and electrocardiogram (ECG) voltages.
In embodiments, the processor may be configured to distinguish between a non-pericardial location and a pericardial location, a pericardial location and a thorax of the subject, and/or between a pericardial location and a location within ventricular tissue of the subject or inside the interior of the heart of the subject.
An aspect of an embodiment of the present invention provides a system for an access needle sensor device that serves as a guideway for introducing other devices into the pericardium, for instance sheath catheters that might subsequently be employed for procedures in the pericardium and the epicardium of the heart. Other devices that the present invention device may accommodate include, but not limited thereto, the following: ablation catheters, guidewires, pacing leads, pacing catheters, pacemakers, visualization and recording devices, drugs, lumens, steering devices or systems, drug or cell delivery catheters, fiber endoscopes, suctioning devices, irrigation devices, electrode catheters, needles, optical fiber sensors, sources of illumination, vital signs sensors, and the like. These devices may be deployed for procedures in an integral body part or space.
An aspect of an embodiment of the present invention provides a system for an access needle sensor device that serves as a guideway for introducing other devices into the pericardium, for instance sheath catheters that might subsequently be employed for procedures in the pericardium and the epicardium of the heart. Other devices that the present invention device may accommodate include, but not limited thereto, the following: ablation catheters, guidewires, pacing leads, pacing catheters, pacemakers, visualization and recording devices, drugs, lumens, steering devices or systems, drug or cell delivery catheters, fiber endoscopes, suctioning devices, irrigation devices, electrode catheters, needles, optical fiber sensors, sources of illumination, vital signs sensors, and the like. These devices may be deployed for procedures in an integral body part or space.
According to further aspects of the invention, a device for inferring the location of a needle in a subject may be provided including a first input configured to receive pressure frequency information from a first sensor and a second input configured to receive cardiac waveform information of the subject from a second sensor. In embodiments, the first sensor may be disposed proximate to a distal end of the needle. Embodiments may further include a processor in communication with the sensors, the processor configured to perform steps such as those discussed above. For example, the processor may be configured to receive cardiac waveform information of the subject from the second sensor via the second input and/or to receive pressure frequency information from the first sensor via the first input. In embodiments, the processor may be configured to distinguish a current location of the needle from another location, and/or distinguish movement of the needle from a first location to a second location, based on an algorithm including the pressure frequency information and the cardiac waveform information.
In embodiments, the processor may be further configured to determine a reference phase based on the cardiac waveform information, and determine a test phase based on the pressure frequency information. The distinguishing may be based on the pressure frequency information from the test phase and the cardiac waveform information from the reference phase.
Additional features, advantages, and embodiments of the invention may be set forth or apparent from consideration of the following detailed description, drawings, and claims. Moreover, it is to be understood that both the foregoing summary of the invention and the following detailed description are exemplary and intended to provide further explanation without limiting the scope of the invention claimed. The detailed description and the specific examples, however, indicate only preferred embodiments of the invention. Various changes and modifications within the spirit and scope of the invention will become apparent to those skilled in the art from this detailed description.
The accompanying drawings, which are included to provide a further understanding of the invention, are incorporated in and constitute a part of this specification, illustrate embodiments of the invention and together with the detailed description serve to explain the principles of the invention. No attempt is made to show structural details of the invention in more detail than may be necessary for a fundamental understanding of the invention and various ways in which it may be practiced. In the drawings:
It is understood that the invention is not limited to the particular methodology, protocols, and configurations, etc., described herein, as these may vary as the skilled artisan will recognize. It is also to be understood that the terminology used herein is used for the purpose of describing particular embodiments only, and is not intended to limit the scope of the invention. It also is be noted that as used herein and in the appended claims, the singular forms “a,” “an,” and “the” include the plural reference unless the context clearly dictates otherwise. Thus, for example, a reference to “a sensor” is a reference to one or more sensors and equivalents thereof known to those skilled in the art.
Unless defined otherwise, all technical and scientific terms used herein have the same meanings as commonly understood by one of ordinary skill in the art to which the invention pertains. The embodiments of the invention and the various features and advantageous details thereof are explained more fully with reference to the non-limiting embodiments and examples that are described and/or illustrated in the accompanying drawings and detailed in the following description. It should be noted that the features illustrated in the drawings are not necessarily drawn to scale, and features of one embodiment may be employed with other embodiments as the skilled artisan would recognize, even if not explicitly stated herein. Descriptions of well-known components and processing techniques may be omitted so as to not unnecessarily obscure the embodiments of the invention. The examples used herein are intended merely to facilitate an understanding of ways in which the invention may be practiced and to further enable those of skill in the art to practice the embodiments of the invention. Accordingly, the examples and embodiments herein should not be construed as limiting the scope of the invention, which is defined solely by the appended claims and applicable law. Moreover, it is noted that like reference numerals reference similar parts throughout the several views of the drawings.
Although various embodiments may be described in the context of a subxiphoid access, and epicardial treatment, for clarity, the invention encompasses and may be applied to other types of treatment, particularly those taking place in, proximate to, and/or or penetrating tissue or organs of the body including different pressure frequency characteristics.
An aspect of an embodiment of the present invention system (and related method) includes a precision fiber-optic pressure sensor and a novel signal analysis algorithm, method, technique and system for identifying pressure-frequency signatures which, in the clinical setting, may allow for safer access to, for example, the pericardial space.
As discussed further below, aspects of the invention include instrumentation constituting the improved subxiphoid access system, as well as the structure of the data analysis algorithm, method, technique and system. Test results of the use of such systems and methods are also provided based on a series of institutionally approved in vivo trials of the intra-thoracic navigation of the tip of a pressure-frequency sensing Tuohy needle in adult canines. The inventors have further assessed the ability to limit the number of residual false positive and false negative identifications of needle location within the pericardium, as compared with the empirical method as employed in the trials by a practicing electrophysiologist.
A series of ACUC approved animal trials were performed in the University of Virginia vivarium by a practicing electrophysiologist on ten canines (>22 kg), using the standard-of-care pericardial access techniques currently employed in the clinical Electrophysiology Lab. Anesthesia was induced using fentanyl and etomidate, and maintained with isoflurane. Canines were mechanically ventilated at a constant rate for the duration of the trials, between 13 and 16 breaths per minute. For each animal, a minimum of 4 pericardial access procedures were performed with a standard 17 gauge epidural (Tuohy) needle. The clinician guided it to the pericardium by fluoroscopy and injection of contrast agent. In selected procedures the final pericardial location was verified by placing a guide wire through the access needle's lumen and observing its location under fluoroscopy. Hydrodynamic pressure data from the access needle (>10 s records) were acquired during each procedure at locations intra-thoracic prior to the diaphragm, intra-thoracic after puncturing the diaphragm, and intra-pericardial. The location of the needle, and therefore our determination of each signal's pericardial or non-pericardial nature, was identified by the clinician's judgment. Some of the data were taken with the needle held in static positions, and the rest while it was being moved from the thoracic cavity into the pericardium, as well as upon withdrawal. Certain pressure recordings were made with ventilation held to anatomically remove breathing signals from the waveform.
During the access procedures, four simultaneous measurements were acquired on a laboratory computer using LabVIEW SignalExpress™ (National Instruments, Austin, Tex.). The four measurements included the pressure as monitored by the sensor in the needle's tip, the left ventricle (LV) pressure, the right femoral artery (A-line) pressure, and the electrocardiogram (ECG) voltages.
In an initial study of pericardial pressure dynamics, a strain gauge on the proximal end of the access needle was used to sense the pressures at the distal tip via a column of fluid between it. Although such instrumentation is highly effective when utilized properly for reading static pressure measurements from a single location, it is difficult to employ for reading small-amplitude hydrodynamic pressures when the fluid-filled conduit (be it a sheath, catheter, or needle) is moving. This is because the fluid filled column has weight, and hence the fluid's hydrostatic pressure contributes inertial artifacts to the amplitude of the pressure read at the proximal end of the fluid conduit. If the needle's orientation relative to gravity is not maintained unchanged while it is being advanced towards the pericardium, the resulting fluctuations in the force transduced by the strain gauge introduce substantial noise onto the signal. This makes the technique very difficult to employ in the clinical setting during an access procedure.
Therefore, a more viable method of monitoring pressures at the tip of a dynamically shifting access needle is desired. In embodiments, exemplary systems may include a solid-state, or optical, pressure sensor. As used herein, “pressure sensors” may preferably include sensors that are capable of registering not only steady-state pressure, but also pressure frequency over time, which may require rapid responsiveness and accuracy. In embodiments, such sensors may also be understood as including, and/or in communication with, a computer processor configured to determine pressure frequency from pressure signals. Varying pressure information received over a period of time may also be referred to generally as pressure frequency information.
A fully solid-state sensor was implemented by the inventors including a fiber optic (Fabry-Perot resonator) pressure transducer (FISO, model FOP-MIV-BA-C1-F2-M2-R1-SC, Quebec, Canada), and was introduced into the lumen of an access needle, with the transducer tip placed just within the edge of the distal tip of the needle so that it did not protrude into the tissue. The sensor at the distal tip of the fiber had a diameter of 550 μm. This was followed by a 20-mm segment of bare fiber of roughly half the sensor's diameter, with the remaining length encased in a PTFE sheath all the way to the connector at the proximal end. The resolution was ≈0.3 mm Hg, with a residual thermal drift of <−0.05% ° C.−1. In embodiments, a pressure sensor, such as an optical sensor, may be affixed to, or proximate to, an end of an access needle. As discussed further below, securing the pressure sensor may be advantageous in reducing unwanted noise, but is not necessarily required.
The optical fiber itself was fixed in place at the proximal end of the needle by passing it through a Tuohy-Borst adapter which was coupled onto the needle's Luer lock. In this instance, the transducer was not fixed at the tip of the needle, since it had to be removed from the inner lumen so that a guide wire could be passed through the needle to verify the location inside the pericardium. This was accomplished by decoupling the Tuohy-Borst fitting from the needle's Luer lock and withdrawing the sensor. During use, the fiber optic cable was connected to its mating light source/signal conditioning device (FISO, model EVO/FPI-HR, Quebec, Canada). The voltages from the analog output board on the signal conditioner were read by an analog-to-digital data converter (National Instruments, model USB-6009, Austin, Tex.) and stored for subsequent analysis.
As mentioned above, three standard measures of cardiac dynamics were acquired in synchrony with the needle pressures during each trial. First, the LV pressures were monitored via a clinical pigtail catheter, which was inserted into the left ventricle from the animal's carotid artery. The pigtail catheter was flushed and filled with saline, and the pressure in it was monitored by a clinical flush-through transducer (Hospira, model Transpac™, Lake Forest, Ill.). Second, central A-line pressures were obtained via a 10 French sheath, placed in the femoral artery of the canine. The sheath was flushed and filled with saline, and monitored by a transducer identical to that used in the LV pigtail catheter. Both the LV and A-line transducers were connected to a data acquisition and signal conditioning board (National Instruments, model USB-9237, Austin, Tex.) via a custom-built cable that bridged the RJ11 jack of the transducer wiring to the RJ50 input plug on the board. These resulting data were useful in assessing the effect of cardiac dynamics on pericardial pressures.
Third, electrophysiological data were taken during the procedures. The bipolar electrical signal between leads on the right arm (RA) and left arm (LA) was monitored by a custom-built ECG reading system. The ionic potentials were transduced to electronic potentials via standard clinical pediatric ECG electrodes (Ambu®, model Blue Sensor M, Glen Burnie, Md.). The RA and LA leads were connected to a custom-built, differential-amplifier signal conditioning circuit (Burr-Brown/TI model OPA2227 P operational amplifiers, Dallas, Tex.) to buffer, amplify, and bandpass-filter the incoming signal. The output signal from the circuit was connected to the second port of the A/D data acquisition device. The signals from both data acquisition units were transferred to the laboratory computer via USB 2.0 connections at a rate of ˜1.6 kHz by the LabVIEW SignalExpress™ software.
Fast Fourier Transform Analysis. For each canine, a set of non-pericardial (pre-diaphragm and post-diaphragm thoracic signals) and intra-pericardial signals were collected. These signals were examined for segments with minimal noise, and suitable samples 10 s long were extracted for further analysis. A hanning window was applied to the segment, and a linear-peak Fast Fourier Transform (FFT) was carried out in LabVIEW. Using the embedded virtual instrument functions on the same 10 s segment of any of the cardiodynamic reference signals, the average heartbeat frequency for that segment was calculated, and the magnitude of the FFT at that specific frequency point was identified as the signal magnitude of interest at the heart rate. Altogether, 98 non-pericardial and 112 pericardial signals were collected and analyzed in this manner.
Following the collection of signal-magnitude data from all 210 signals, an automated search was performed to determine how well a threshold value of the cardiac signal strength at the heart-rate frequency could separate pericardial from non-pericardial locations of the needle tip. A custom program in MATLAB® was written and used to numerically evaluate the magnitude data. The frequency component at the heart rate is significantly less in the non-pericardial signal when compared to the pericardial signal. The goal then was to define a threshold pressure which indicated the best obtainable separation between non-pericardial and pericardial signals by inferring the presence of false positives and false negatives (i.e., non-pericardial signals above the threshold, denoted by falsep; pericardial signals below the threshold, denoted by falsen, respectively). Possible threshold values were tested from the range of 0 to 0.6 mmHg (0 to 80 Pa), in increments of 0.0001 mmHg (0.013 Pa). For each possible threshold value within that range, the number of non-pericardial signal magnitudes above that threshold (false positives) and the number of pericardial signal magnitudes below that threshold (false negatives) were counted. The best obtainable threshold was then found by minimizing the score
Score=Wp*falsep+Wn*falsen (1)
where the scoring weights Wp and Wn indicated the relative importance of false positives and false negatives. In a given calculation for best obtainable threshold, Wp and Wn remain constant. A best obtainable threshold was found for every combination of values of Wp and Wn ranging from 0.1 to 1.0 in increments of 0.1. The nature and application of the findings is discussed below in the Results Section.
Custom Algorithm and Method Analysis. Additional investigations were conducted for determining if a cardiac signal was present in the needle-tip signal, taking into account the inconsistency of the heart rate in many of the experiments and, in general, the spectral complexity of the heartbeat signal. For example, heart-rate variability could not only cause FFT spectral leakage, but even separate the FFT cardiac peak into distinct sub-peaks of lower magnitude, making FFT-based measurements inconsistent. Using the ECG as a reference signal, the needle sensor's waveform was segmented between consecutive R waves in the QRS complex of the signal. A custom-synthesized algorithm that combined aspects of phase sensitive detection (PSD) and matched filtering then analyzed it to ascertain the presence of a consistent waveform with a fundamental frequency at that of the heart rate, but with the ability to accommodate beat-to-beat variations in heart rate. A flow chart for this algorithm is shown in
In
For the algorithm and method analysis, the same signals from the FFT analysis were used for the algorithm and method analysis, and an automated program in LabVIEW performed it on each of the needle sensor's waveforms. After reviewing the spectral structure of all the types of waveforms that were captured, none of which contained signal components ≧100 Hz (which was taken to be the Nyquist frequency), the signals were down-sampled to 2×100 Hz=200 Hz to allow for more precise filtering. All filtering routines were performed in a zero-phase implementation, infinite impulse response (IIR) mode, which filtered the forward signal and then reversed and refiltered it, and reversed it again in order to remove all phase-sensitive filtering transients. Moreover, all of the simple high- and low-pass filters were of elliptic structure in order to minimize the order of the filter, thus both minimizing start- and end-signal transients and maximizing the computational efficiency.
The ECG waveform served as the reference for the algorithm and method in identifying the timing of cardiac cycles in the animal. The ECG signal segment was notch-filtered at 60 Hz (Q≈50) to eliminate any residual power-line noise that the signal conditioner did not fully remove. The signal was then high pass-filtered at a cutoff frequency of 20 Hz, to maintain only the high frequency elements that compose the QRS complex. Then, the strength of the remaining high-frequency noise components were attenuated with an undecimated wavelet transform using a biorthogonal 4_4 (FBI) wavelet to insure that only the QRS peak of interest remained. To provide for consistency in subsequent mathematical steps, the signal terms were then rectified by their absolute value and the remaining detectable but extraneous peaks were smoothed with a moving average filter. As shown in
The 10 s records of the needle's hydrodynamic pressure signal were high-pass filtered with a cutoff frequency of 1 Hz to attenuate low frequency breathing components. The resulting signal was further separated into cardiac segments according to the timing points found using the ECG analysis above. Only cardiac segments containing a full cardiac cycle (a start and an end point) were analyzed. A goal of the algorithm and method was to quantitatively compare the waveform between consecutive cardiac segments. In many implementations of phase-sensitive detection, the input waveform is multiplied by a reference signal in order to evaluate whether the waveform is of the desired frequency and phase. If the input waveform's characteristics match those of the reference waveform, then the multiplied signal will be perfectly rectified. For reasons discussed below, in place of a common reference waveform which would be used for that multiplication, the algorithm instead used an interpolated version of the previous cardiac segment as the reference waveform for the currently analyzed cardiac segment. Examples of both pericardial and non-pericardial signals are shown in
This routine combines features of both match filtering and PSD. If the main signal component which is present is fundamentally matched to cardiac dynamics, then the signal should repeat between cardiac segments, regardless of noise. It is assumed that if the needle is in the pericardial space, then there should be good agreement between the current and last cardiac segment of the waveform. However, instead of convolving the signals or performing Fourier multiplication of the signals, which is computationally inefficient or requires a large window of consistent repeatable signal, respectively, the signal may be integrated.
Another significant point is that when integrating a high-noise signal, as more time points are integrated, the integral of white noise tends to zero. Hence, this algorithm and method also functions efficiently in high noise scenarios. In summary, this approach, which employs the ECG as the reference waveform, is a more elaborate but robust way of ascertaining the presence and magnitude of a cardiac signal pattern in the needle sensor's pressure signal, which can take into account and adjust for irregular or inconsistent heart rates.
The signal magnitude at the heart rate frequency of the FFT of the 98 non-pericardial waveforms was (0.09±0.08) mmHg, and it was (0.55±0.31) mmHg for the 112 pericardial signals. Unpaired, one-tailed t-tests revealed a statistically significant difference in means of the two groups (p<0.01). Purposeful ventricular perforation was successful in 5 animals, and the mean signal at the heart rate from these 5 measurements was (24.98±9.34) mmHg, which was significantly greater than the magnitudes for both the non-pericardial and pericardial groups (p<0.01).
The average, dimensionless signal output from the algorithm for the 98 non-pericardial waveforms was (0.024±0.03), and (0.21±0.14) for the 112 pericardial signals. Average signal output for ventricular signals was (14.74±5.97). As shown in
Using Eq. (1) on both the FFT and algorithm data, a number of thresholds were found which separated the pericardial from the non-pericardial data. However, this analysis also revealed some overlap of non-pericardial and pericardial signals, i.e., regions of non-separation of the signals (transition zone in
After analyzing both the FFT and algorithm data and method, the values for thresholds Ta and Tb were selected. The number of associated false positives and false negatives are displayed in
The separation of non-pericardial from pericardial signals is the main focus in what follows. This is because the separation of ventricular from non-ventricular signals is easily achieved for Tc without any false outputs. The performance of both the FFT and algorithm analyses (and related methods) of the data is shown in
False positives are more of a procedural nuisance than a safety concern. On the other hand, false negatives are dangerous, because the clinician might continue to advance the access needle into the ventricle if the needle is actually pericardial but the algorithm indicated that it was instead non-pericardial. Therefore, one object of the invention is to not only minimize the total number of false outputs (of both kinds), but to make the minimization of false negatives the highest priority. The custom-synthesized algorithm is able to minimize the number of false negatives with much fewer false positives (see
The resulting threshold values in the algorithm are 0.0405 for Ta. 0.077 for Tb, and 4 for Tc. Of the 210 non-ventricular pressure measurements, 87.14% of the acquired signals fell in the appropriate zone upon analysis with the algorithm, with 1.43% of the signals identified as false negatives, 1.90% of the signals identified as false positives, and 9.52% of the signals in the transition zone between Ta and Tb. All 5 (100%) ventricular measurements fell above Tc, with no false negatives or false positives regarding signals falling on the wrong side of Tc.
It is important to address the pericardial signal outputs which fall beneath both Ta and Tb (i.e., the false negatives shown in
Thus, it was noted that it is possible that solitary PVC beats may cause an error in the algorithm and method if only one previous beat is used for comparison to the current cardiac segment. In embodiments, such errors may be avoided, for example, by adapting the algorithm to compare, for example, the previous two to five beat segments. There are several ways in which it might be accomplished, e.g. by acquiring and averaging the parameters of two neighboring PVC beats and using that result to drive the segmentation of the needle waveform. Other criteria may be applied to disregard a reference phase. Such criteria may rely, for example, on an appropriate statistical measure used to identify a significantly irregular beat pattern (e.g., a sudden change in the frequency band in which the beat occurred, indicating the occurrence of what in nonlinear dynamics is called a “transition to chaos”).
In embodiments, the algorithm may be adapted to focus primarily on capturing the first cardiac segment once the needle is positioned pericardially. If one PVC beat were to occur at that first segment, then the algorithm would simply have a delayed response of one extra cardiac segment, and the probability of a PVC occurring at such a precisely defined moment is very low.
Pericardial access can be critical for curing several cardiac conditions but it is fraught with a high risk of both procedural failure and ventricular perforation. To address these issues, the inventors measured the pressure-frequency signals generated by a solid state (fiber optic) sensor at the tip of a pericardial access needle, and collected ECG signals in synchrony. By employing a novel algorithm, method, technique and system that contains characteristics of both matched filtering and phase-sensitive detection to process those data, the location of the needle's tip was distinguished accurately. In analogy with phase sensitive detection, the present subject matter uses measures of cardiac dynamics to separate the incoming needle's signal into distinct sections. Also, just as matched filtering checks an incoming signal according to a known outgoing signal, exemplary algorithms according to aspects of the invention may be configured to compare the current cardiac section of the signal to the previous section, to search for a similar pattern. Such algorithms, and associated methods, techniques and systems, can distinguish pericardial from non-pericardial waveforms regardless of signal dynamics and structure, as long as a signal with the fundamental frequency (equal to the cardiac frequency) is present at any given time. This is significant, because as different parts of the pericardium are accessed at different needle insertion angles, the signal structure can change.
It was found that the exemplary algorithms, and associated methods, techniques and systems, presented herein provided a better approach than tracking an FFT peak for several reasons. First, FFT peaks take time to establish in a signal window, since they sample global (and not local) frequency content. However, use of an exemplary algorithm in systems such as those described is expected to decrease the time lag needed to establishing the needle's presence in the pericardium to just one cardiac cycle (i.e., ˜1 s).
In order to achieve such results, it is expected that a sensor fixed, or locked, proximate to the end of the needle would be advantageous. That is, configurations in which the fiber optic sensor is left un-fixed within the tip of the needle (e.g. so that it can be easily extracted whenever a guide wire had to be passed into the needle's lumen), may result in some mechanical noise being imposed on the signal based on the residual motion of the sensor's tip. In order to reduce the overall effect of such noise, the algorithm's output may also be averaged over an extended window.
Also, as the patient's heart rate undergoes normal shifts, both spectral leakage and separation of FFT peaks is a concern. Since the systems and methods described herein track the patient's cardiac dynamics directly, embodiments may also provide means of tracking heart rate during accelerations into tachycardia, decelerations into bradycardia, and inconsistencies as occur in atrial fibrillation. Also, embodiments of the invention may be useful for patients with low cardiac signal strength in the pericardial waveform because of previous cardiac surgery and the resulting pericardial adhesions.
Another important aspect of the invention is to quantify the pressure waveform at the needle's tip as it breaches the pericardial membrane, going from non-pericardial to pericardial anatomy. In studying this, the inventors looked at several examples of dynamic pressure measurements made while the needle both breached into the pericardium and was pulled out of it were acquired, and an example is shown in
According to aspects of the invention, embodiments may also be adapted to allow for the analysis to occur in real time. Such adaptation may include, among other features, real-time data analysis during signal acquisition, as opposed to post-processing of data segments. Also, a statistical analysis of this and other pericardial and non-pericardial signals may be analyzed to determine not clinical thresholds between anatomical zones, as well as determining confidence intervals for those thresholds. This would allow for systems and methods that tell the clinician which anatomical zone they are in with statistical certainty. The needle itself is also an important part of the access system that can be improved to provide real-time or near real-time analysis. The device used by the inventors was a Tuohy epidural needle, but with the fiber optic sensor threaded through the lumen to the tip. However, the pressure sensor was neither fixed to the tip of the needle, nor designed to be otherwise housed by the Tuohy needle, and this allowed the fiber optic tip to move in certain situations, thus increasing noise and decreasing signal fidelity. A needle with, for example, a fixed or locked pressure sensor would be expected to produce readings with a larger signal-to-noise ratio.
As discussed above, following in vivo studies on 10 adult canine models, the inventors analyzed 215 pressure-frequency measurements made at the distal tip of the access needle, of which 98 were from non-pericardial, 112 were from pericardial and 5 were from ventricular locations. The needle locations as identified by the exemplary systems and methods were significantly different from each other (p<0.01), and systems and methods showed improved performance when compared to a standard FFT analysis of the same data. Moreover, the structure of the algorithm, method, technique and system can be advantageously used to minimize, or overcome, the time lags intrinsic to FFT analysis, such that the needle's location may be determined in near-real time.
An advantage accruing from the use of the means and method of the invention is, but not limited thereto, the ability to allow for pacing of the epicardium itself with relative ease.
It should be appreciated that as discussed herein, a subject may be a human or any animal. It should be appreciated that an animal may be a variety of any applicable type, including, but not limited thereto, mammal, veterinarian animal, livestock animal or pet type animal, etc. As an example, the animal may be a laboratory animal specifically selected to have certain characteristics similar to human (e.g. rat, dog, pig, monkey), etc. It should be appreciated that the subject may be any applicable human patient, for example.
In an aspect of an embodiment of the invention, the access needle 100 is used for accessing the thorax 51 and pericardium of a subject 50, wherein the access needle comprises a pressure frequency sensor or system for sensing pressure frequency in the thorax, the pericardium or other tissue of the heart. However, it should be appreciated that various embodiments of the present invention device or system and method are not necessarily limited to accessing the thorax and pericardium of a subject. It may also be used in the organ structures or tubular structures in the thorax as well as other locations or regions in the body. An organ includes, for example, a solid organ, a hollow organ, parenchymal tissue (e.g., stomach, brain, esophagus, colon, rectum, kidneys, liver, etc.) and/or stromal tissue. Hollow organ structures includes, for example, stomach, esophagus, colon, rectum, and ducts, or the like. A tubular structure may include a blood vessel. A blood vessel may include one or more of the following: vein, venule, artery, arterial, or capillary.
Turning to
The computer system 900 may also include a main memory 908, preferably random access memory (RAM), and may also include a secondary memory 910. The secondary memory 910 may include, for example, a hard disk drive 912 and/or a removable storage drive 914, representing a floppy disk drive, a magnetic tape drive, an optical disk drive, a flash memory, etc. The removable storage drive 914 reads from and/or writes to a removable storage unit 918 in a well known manner. Removable storage unit 918, represents a floppy disk, magnetic tape, optical disk, etc. which is read by and written to by removable storage drive 914. As will be appreciated, the removable storage unit 918 includes a computer usable storage medium having stored therein computer software and/or data.
In alternative embodiments, secondary memory 910 may include other means for allowing computer programs or other instructions to be loaded into computer system 900. Such means may include, for example, a removable storage unit 922 and an interface 920. Examples of such removable storage units/interfaces include a program cartridge and cartridge interface (such as that found in video game devices), a removable memory chip (such as a ROM, PROM, EPROM or EEPROM) and associated socket, and other removable storage units 922 and interfaces 920 which allow software and data to be transferred from the removable storage unit 922 to computer system 900.
The computer system 900 may also include a communications interface 924. Communications interface 124 allows software and data to be transferred between computer system 900 and external devices, including, for example, pressure sensors as described herein, ECGs, etc. The communications interface 924 may include a plurality of physical and/or virtual input/output ports configured to communicate with different sensors, etc. Examples of communications interface 924 may include a modem, a network interface (such as an Ethernet card), a communications port (e.g., serial or parallel, etc.), a PCMCIA slot and card, a modem, wife, Bluetooth, etc. Software and data transferred via communications interface 924 are in the form of signals 928 which may be electronic, electromagnetic, optical or other signals capable of being received by communications interface 924. Signals 928 are provided to communications interface 924 via a communications path (i.e., channel) 926. Channel 926 (or any other communication means or channel disclosed herein) carries signals 928 and may be implemented using wire or cable, fiber optics, blue tooth, a phone line, a cellular phone link, an RF link, an infrared link, wireless link or connection and other communications channels.
In this document, the terms “computer program medium” and “computer usable medium” are used to generally refer to media or medium such as various software, firmware, disks, drives, removable storage drive 914, a hard disk installed in hard disk drive 912, and signals 928. These computer program products (“computer program medium” and “computer usable medium”) are means for providing software to computer system 900. The computer program product may comprise a computer useable medium having computer program logic thereon. The invention includes such computer program products. The “computer program product” and “computer useable medium” may be any computer readable medium having computer logic thereon.
Computer programs (also called computer control logic or computer program logic) are may be stored in main memory 908 and/or secondary memory 910. Computer programs may also be received via communications interface 924. Such computer programs, when executed, enable computer system 900 to perform the features of the present invention as discussed herein. In particular, the computer programs, when executed, enable processor 904 to perform the functions of the present invention. Accordingly, such computer programs represent controllers of computer system 900.
In an embodiment where the invention is implemented using software, the software may be stored in a computer program product and loaded into computer system 900 using removable storage drive 914, hard drive 912 or communications interface 924. The control logic (software or computer program logic), when executed by the processor 904, causes the processor 904 to perform the functions of the invention as described herein.
In another embodiment, the invention is implemented primarily in hardware using, for example, hardware components such as application specific integrated circuits (ASICs). Implementation of the hardware state machine to perform the functions described herein will be apparent to persons skilled in the relevant art(s).
In yet another embodiment, the invention is implemented using a combination of both hardware and software.
In an example software embodiment of the invention, the methods described above may be implemented in SPSS control language or C++ programming language, but could be implemented in other various programs, computer simulation and computer-aided design, computer simulation environment, MATLAB, or any other software platform or program, windows interface or operating system (or other operating system) or other programs known or available to those skilled in the art.
The description given above is merely illustrative and is not meant to be an exhaustive list of all possible embodiments, applications or modifications of the invention. Thus, various modifications and variations of the described methods and systems of the invention will be apparent to those skilled in the art without departing from the scope and spirit of the invention. Although the invention has been described in connection with specific embodiments, it should be understood that the invention as claimed should not be unduly limited to such specific embodiments. Indeed, various modifications of the described modes for carrying out the invention which are obvious to those skilled in the memory circuit design, memory circuit manufacture or related fields are intended to be within the scope of the appended claims.
The following patents, applications and publications as listed below and throughout this document are hereby incorporated by reference in their entirety herein.
PCT International Application Serial No. PCT/US2008/056643, filed Mar. 12, 2008, entitled, “Access Needle Pressure Sensor Device and Method of Use” and corresponding U.S. patent application Ser. No. 12/530,830 filed Sep. 11, 2009.
PCT International Application Serial No. PCT/US2008/056816, filed Mar. 13, 2008, entitled, “Epicardial Ablation Catheter and Method of Use” and corresponding U.S. patent application Ser. No. 12/530,938 filed Sep. 11, 2009.
PCT International Application Serial No. PCT/US2008/057626, filed Mar. 20, 2008, entitled, “Electrode Catheter for Ablation Purposes and Related Method Thereof” and corresponding U.S. patent application Ser. No. 12/532,233 filed Sep. 21, 2009.
PCT International Application Serial No. PCT/US2010/033189, filed Apr. 30, 2010, entitled “Access Trocar and Related Method Thereof”. PCT International Application Serial No. PCT/US2008/082835, filed Nov. 7, 2008, entitled, “Steerable Epicardial Pacing Catheter System Placed Via the Subxiphoid Process,” and corresponding U.S. patent application Ser. No. 12/741,710 filed May 6, 2010.
PCT International Application Serial No. PCT/US2010/061413, filed Dec. 21, 2010, entitled “System For Femoral Vasculature Catheterization and Related Method.
PCT International Application Serial No. PCT/US2011/025470, filed Feb. 18, 2011.
a. U.S. patent application Ser. No. 12/741,710 entitled “Steerable Epicardial Pacing Catheter System Placed via the Subxiphoid Process,” filed May 6, 2010, U.S. Patent Application Publication No. 2010/0241185, Sep. 23, 2010;
b. International Patent Application No. PCT/US2008/082835 entitled “Steerable Epicardial Pacing Catheter System Placed via the Subxiphoid Process,” filed Nov. 7, 2008, International Patent Application Publication No. WO 2009/062061, May 14, 2009;
c. International Patent Application No. PCT/US2010/033189 entitled “Access Trocar and Related Method Thereof,” filed Apr. 30, 2010;
d. U.S. patent application Ser. No. 12/760,837 entitled “Coaxial Catheter Systems for Transference of Medium,” filed Apr. 15, 2010;
e. U.S. patent application Ser. No. 11/191,676 entitled “Coaxial Catheter Systems for Transference of Medium,” filed Jul. 28, 2005; U.S. Pat. No. 7,727,225 Jun. 1, 2010;
f. International Patent Application No. PCT/US2005/026738 entitled “Coaxial Catheter Systems for Transference of Medium,” filed Jul. 28, 2005, International Patent Application Publication No. WO06/15091, Feb. 9, 2006;
g. U.S. patent application Ser. No. 12/625,153 entitled “Blood Flow Bypass Catheters and Methods for the Delivery of Medium to the Vasculature and Body Ducts,” filed Nov. 24, 2009;
h. U.S. patent application Ser. No. 11/884,421 entitled “Blood Flow Bypass Catheters and Methods for the Delivery of Medium to the Vasculature and Body Ducts,” filed Aug. 15, 2007, U.S. Patent Application Publication No. 2008/0262467, Oct. 23, 2008;
i. International Patent Application No. US2006/005876 entitled “Blood Flow Bypass Catheters and Methods for the Delivery of Medium to the Vasculature and Body Ducts,” filed Feb. 16, 2006, International Patent Application Publication No. WO06/089243, Aug. 24, 2006;
j. U.S. patent application Ser. No. 12/532,233 entitled “Electrode Catheter for Ablation Purposes and Related Method Thereof,” filed Sep. 21, 2009, U.S. Patent Application Publication No. 2010/0211064, Aug. 19, 2010;
k. International Patent Application No. PCT/US2008/057626 entitled “Electrode Catheter for Ablation Purposes and Related Method Thereof,” filed Mar. 20, 2008, International Patent Application Publication No. WO 2008/118737, Oct. 2, 2008;
l. U.S. patent application Ser. No. 12/530,938 entitled “Epicardial Ablation Catheter and Method of Use,” filed Sep. 11, 2009, U.S. Patent Application Publication No. 2010/0114093, May 6, 2010;
m. International Patent Application No. PCT/US2008/056816 entitled “Epicardial Ablation Catheter and Method of Use,” filed Mar. 13, 2008, International Patent Application Publication No. WO 2008/112870, Sep. 18, 2008;
n. U.S. patent application Ser. No. 12/530,830 entitled “Access Needle Pressure Sensor Device and Method of Use,” filed Sep. 11, 2009, U.S. Patent Application Publication No. 2010/0094143, Apr. 15, 2010;
o. International Patent Application No. PCT/US2008/056643 entitled “Access Needle Pressure Sensor Device and Method of Use,” filed Mar. 12, 2008, International Patent Application Publication No. WO 2008/115745, Sep. 25, 2008;
p. U.S. patent application Ser. No. 12/304,801 entitled “Closure Device for Skull Plates and Related Method Thereof,” filed May 18, 2009, U.S. Patent Application Publication No. 2010/0042158, Feb. 18, 2010;
q. International Patent Application No. PCT/US2007/014881 entitled “Closure Device for Skull Plates and Related Method Thereof,” filed Jun. 26, 2007, International Patent Application Publication No. WO 2008/02595, Jan. 3, 2008;
r. U.S. patent application Ser. No. 12/513,258 entitled “Means and Methods for Cytometric Therapies,” filed May 1, 2009, U.S. Patent Application Publication No. 2010/0210927, Aug. 19, 2010;
s. International Patent Application No. PCT/US2007/023047 entitled “Means and Methods for Cytometric Therapies,” filed Nov. 1, 2007, International Patent Application No. WO 2008/057370, May 15, 2008;
t. U.S. patent application Ser. No. 12/375,139 entitled “System and Method for Intracranial Implantation of Therapeutic or Diagnostic Agents,” filed Jan. 27, 2009, U.S. Patent Application Publication No. 2009/0192487, Jul. 30, 2009;
u. International Patent Application No. PCT/US2007/016256 entitled “System and Method for Intracranial Implantation of Therapeutic or Diagnostic Agents,” filed Jul. 18, 2007, International Patent Application No. WO 2008/013709, Jan. 31, 2008;
v. U.S. patent application Ser. No. 12/160,378 entitled “Multi-Port Catheter System with Medium Control and Measurement Systems for Therapy and Diagnosis Delivery,” filed Aug. 1, 2008, U.S. Patent Application Publication No. 2009/0048577, Feb. 19, 2009;
w. International Patent Application No. PCT/US2007/000353 entitled “Multi-Port Catheter System with Medium Control and Measurement Systems for Therapy and Diagnosis Delivery,” filed Jan. 9, 2007, International Patent Application Publication No. WO 2007/081842, Jul. 19, 2007;
x. International Patent Application No. US2006/013621 entitled “Catheter Systems for Delivery of Agents and Related Method Thereof,” filed Apr. 12, 2006, International Patent Application Publication No. WO06113267, Oct. 26, 2006;
y. U.S. patent application Ser. No. 11/105,166 entitled “Catheter Systems for Delivery of Agents and Related Method Thereof,” filed Apr. 13, 2005, U.S. Patent Application Publication No. 2005/0245896, Nov. 3, 2005; U.S. Pat. No. 7,670,327, issued Mar. 2, 2010;
z. U.S. patent application Ser. No. 10/985,340 entitled “Catheter Navigation within an MR Imaging Device,” filed Nov. 10, 2004, U.S. Patent Application Publication No. 2005/0119556, Jun. 2, 2005;
aa. U.S. patent application Ser. No. 10/429,524 entitled “Catheter Navigation within an MR Imaging Device,” filed May 5, 2003, U.S. Patent Application Publication No. 2003/0195412, Oct. 16, 2003; U.S. Pat. No. 6,834,201, issued Dec. 21, 2004;
ab. International Patent Application No. PCT/US2002/002363 entitled “Catheter Navigation within an MR Imaging Device,” filed Jan. 28, 2002;
ac. U.S. patent application Ser. No. 09/772,188 entitled “Catheter Navigation within an MR Imaging Device,” filed Jan. 29, 2001;
ad. U.S. patent application Ser. No. 10/444,884 entitled “Cell Delivery Catheter and Method,” filed May 23, 2003, U.S. Patent Application Publication No. 2003/0204171, Oct. 30, 2003;
ae. U.S. patent application Ser. No. 09/574,857 entitled “Cell Delivery Catheter and Method,” filed May 19, 2000, U.S. Pat. No. 6,599,274, issued Jul. 29, 2003;
af. U.S. patent application Ser. No. 09/548,110 entitled “Multi-Probe System,” filed Apr. 12, 2000, U.S. Pat. No. 6,626,902, issued Sep. 30, 2003;
ag. International Patent Application No. US99/24253 entitled “Mri And Magnetic Stereotaxis Surgical System,” filed Oct. 15, 1999, International Patent Application Publication No. WO00/23000, Apr. 27, 2000;
ah. U.S. patent application Ser. No. 09/174,189 entitled “Combined Magnetic Resonance Imaging and Magnetic Stereotaxis Surgical Apparatus and Processes,” filed Oc. 16, 1998, U.S. Pat. No. 6,298,259, issued Oct. 2, 2001;
ai. International Patent Application No. US99/17880 entitled “MR-Visible Device for Magnetic Stereotaxis Neurological Interventions,” filed Aug. 6, 1999, International Patent Application No. WO00/07652, Feb. 17, 2000;
aj. U.S. patent application Ser. No. 09/131,031 entitled “MR-Visible Medical Device for Neurological Interventions Using Nonlinear Magnetic Stereotaxis and a Method Imaging,” filed Aug. 7, 1998, U.S. Pat. No. 6,272,370, issued Aug. 7, 2001;
ak. U.S. patent application Ser. No. 09/114,414 entitled “Magnetic Stereotactic System for Treatment Delivery,” filed Jul. 13, 1998, U.S. Pat. No. 6,216,030, Apr. 10, 2001;
al. U.S. patent application Ser. No. 08/464,279 entitled “Magnetic Stereotactic System for Treatment Delivery,” filed Jun. 5, 1995, U.S. Pat. No. 5,707,335, issued Jan. 13, 1998;
am. U.S. patent application Ser. No. 08/096,214 entitled “Magnetic Stereotactic System for Treatment Delivery,” filed Jul. 19, 1993, U.S. Pat. No. 5,779,694, filed Jul. 14, 1998;
an. U.S. patent application Ser. No. 07/904,032 entitled “Magnetic Stereotactic System For Treatment Delivery,” filed Jun. 25, 1992;
ao. U.S. patent application Ser. No. 07/463,340 entitled “Magnetic Stereotactic System for Treatment Delivery,” filed Jan. 10, 1990, U.S. Pat. No. 5,125,888 Jun. 30, 1992.
The following patents, applications and publications as listed below and throughout this document are hereby incorporated by reference in their entirety herein.
WO95/10319, Fleischman, et al., “Electrodes for Creating Lesions in Body Tissue”, 4/1995.
EP 1 129 681 A1, Pezzola, A., “Phacoemulsification Tip”, 9/2001.
E. Aliot, W. Stevenson, J. Almendral-Garrote, F. Bogun, C. Calkins, E. Delacretaz, P. Bella, G. Hindricks, P. Jais, M. Josephson, J. Kautzner, G. Kay, K. Kuck, B. Lerman, F. Marchlinski, V. Reddy, M. Schalij, R. Schilling, L. Soejima, and D. Wilber, “EHRA/HRS expert consensus on catheter albation of ventricular arrhythmias,” Europace, vol. 11, no. 6, pp. 771-817, 2009.
E. Sosa, M. Scanavacca, A. d'Avila, and F. Pilleggi, “A new technique to perform epicardial mapping in the electrophysiology laboratory,” J. Cardiovasc. Electrophysiol., vol. 7, no. 6, pp. 531-6, 1996.
U. Tedrow and W. Stevenson, “Strategies for epicardial mapping and ablation of ventricular tachycardia,” J. Cardiovasc. Electrophysiol., vol. 20, no. 6, pp. 710-3, 2009.
The disclosures of all references and publications cited above are expressly incorporated by reference in their entireties to the same extent as if each were incorporated by reference individually.
This application claims priority under 37 CFR §1.78(a) to U.S. Provisional Application Ser. No. 61/482,527 filed on May 4, 2011, and is a continuation of U.S. application Ser. No. 13/464,762 filed May 4, 2012. U.S. application Ser. No. 13/464,762 is a continuation in-part of U.S. application Ser. No. 12/530,830 filed on Sep. 11, 2009, now U.S. Pat. No. 8,282,565, which is a National Stage Entry of PCT/US2008/056643 filed on Mar. 12, 2008, which claims priority under 37 CFR §1.78(a) to U.S. Provisional Application Ser. No. 60/918,782 filed on Mar. 19, 2007, the contents of all of which are incorporated herein by reference in their entireties. U.S. application Ser. No. 13/464,762 is also a bypass continuation-in-part of PCT/US2011/025470 filed Feb. 18, 2011, which claims priority to U.S. Provisional Application Ser. No. 61/305,560 filed Feb. 18, 2010, and U.S. Provisional Application Ser. No. 61/442,836 filed Feb. 15, 2011, the contents of all of which are incorporated herein by reference in their entireties. The present application is related to the following applications, of which all of the disclosures of the following applications are hereby incorporated by reference herein in their entireties: PCT International Application Serial No. PCT/US2008/056643, filed Mar. 12, 2008, entitled, “Access Needle Pressure Sensor Device and Method of Use” and corresponding U.S. patent application Ser. No. 12/530,830 filed Sep. 11, 2009; PCT International Application Serial No. PCT/US2008/056816, filed Mar. 13, 2008, entitled, “Epicardial Ablation Catheter and Method of Use” and corresponding U.S. patent application Ser. No. 12/530,938 filed Sep. 11, 2009; PCT International Application Serial No. PCT/US2008/057626, filed Mar. 20, 2008, entitled, “Electrode Catheter for Ablation Purposes and Related Method Thereof” and corresponding U.S. patent application Ser. No. 12/532,233 filed Sep. 21, 2009; PCT International Application Serial No. PCT/US2010/033189, filed Apr. 30, 2010, entitled “Access Trocar and Related Method Thereof”; PCT International Application Serial No. PCT/US2008/082835, filed Nov. 7, 2008, entitled, “Steerable Epicardial Pacing Catheter System Placed Via the Subxiphoid Process,” and corresponding U.S. patent application Ser. No. 12/741,710 filed May 6, 2010; PCT International Application Serial No. PCT/US2010/061413, filed Dec. 21, 2010, entitled “System For Femoral Vasculature Catheterization and Related Method; and PCT International Application Serial No. PCT/US2011/025470, filed Feb. 18, 2011.
Number | Date | Country | |
---|---|---|---|
61482527 | May 2011 | US | |
61305560 | Feb 2010 | US | |
61442836 | Feb 2011 | US | |
60918782 | Mar 2007 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 13464762 | May 2012 | US |
Child | 15295102 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 12530830 | Sep 2009 | US |
Child | 13464762 | US |