The heel-to-shin test is a test of lower limb coordination and position sense, and is often used to evaluate the integrity of the cerebellum. In the test, a supine or seated patient is requested to place the heel of one foot at the top of the shin just below the knee of the opposite leg, slide the heel down the center of the shin toward the ankle, and then drag the heel back up to the knee. The patient may perform this action several times with both heels, each sliding along the shin of the opposite leg. As the patient performs the test, a medical practitioner (e.g., physician, nurse, or physical therapist) identifies the number of times that the patient's heel deviates from the center of the shin, this number being indicative of the health of the cerebellum.
While this form of testing can provide the practitioner with an idea of the functioning of the patient's cerebellum, it is imprecise as it relies on the practitioner's subjective impressions as to whether or not the heel deviates from the center of the shin to an extent to which the occurrence should actually be counted as a deviation. As one can imagine, one practitioner may count slight movements from the center of the shin as deviations that should be noted, while another practitioner may only count larger movements from the center of the shin as deviations that should be noted. Because of this subjectivity, the results of the test are of less value than they could be and are not very useful for comparison with the results of other heel-to-shin tests administered by other practitioners.
In view of the subjectivity of the current heel-to-shin testing, it can be appreciated that it would be desirable to have systems and methods for heel-to-shin testing that are more precise and yield more useful results.
The present disclosure may be better understood with reference to the following figures. Matching reference numerals designate corresponding parts throughout the figures, which are not necessarily drawn to scale.
As described above, it would be desirable to have systems and methods for heel-to-shin testing that are precise and yield useful results. Disclosed herein are examples of such systems and methods. In one embodiment, a system for heel-to-shin testing comprises a patient interface that can be applied to a patient's shin. The patient interface includes at least one elongated touch sensor that can be positioned over the medial portion of the patient's shin from the knee to the ankle. The system can further include a control module that is in electrical communication with the touch sensor and that can collect data from it. The data that is collected depends upon the nature of the touch sensor. In cases in which the touch sensor is only configured to detect contact, the data includes the number of times the heel deviates from the medial shin, which can be identified from the number instances in which contact is interrupted. In cases in which the touch sensor detects contact as well as the position at which the contact occurs, the data includes the number of times the heel deviates from the shin as well as the axial positions at which such deviations occurred. In such a case, the nature of the deviations can be identified and the percentage of the total distance and/or time that the heel is or is not correctly aligned with the shin can be determined.
In the following disclosure, various specific embodiments are described. It is to be understood that those embodiments are example implementations of the disclosed inventions and that alternative embodiments are possible. Such alternative embodiments include hybrid embodiments that include features from different disclosed embodiments. All such embodiments are intended to fall within the scope of this disclosure.
The disclosed systems and methods enable more precise heel-to-shin testing by electronically sensing contact between the patient's heel and the shin of the patient's opposite leg. Such sensing removes the opportunity for judgement errors or subjectivity on the part of the individual (e.g., medical practitioner) administering the test. Accordingly, more accurate results can be obtained and, therefore, more accurate assessments can be made about the functioning of the patient's cerebellum.
The touch sensors comprised by the patient interface can take a variety of forms. In some embodiments, the touch sensor 26 comprises at least one force transducer that, for example, comprises one or more strain gauges, piezoelectric elements, resistive elements, or a capacitive elements. The touch sensors can register contact by the heel when the heel is pressed against the sensor with a force that meets or exceeds a predetermined threshold.
With reference back to
Also shown in
In order to conduct heel-to-shin testing, the patient interface 12 must be applied to the shin. To prevent slippage, it is desirable to attach the patient interface 12 to the shin. Such attachment can be achieved in a variety of ways. In one embodiment, the patient interface 12 can be secured to the shin with adhesive tape that lines its back side. In another embodiment, the patient interface 12 can be secured to the shin using flexible, possibly elastic, bands 52, as depicted in
In yet another embodiment, the patient interface 12 can be secured to the shin using a stocking 54, as depicted in
Also shown in
The system 10 can be used to conduct heel-to-shin tests similar to those performed in the prior art, but with much greater precision and accuracy. To conduct such a test, the control module 14 is powered on, for example, by pressing one of the buttons 44 (e.g., a “power” button). The patient interface 12 is then (or previously) applied to the shin of the patient in a manner in which the interface extends from the knee to the ankle along the centerline of the shin.
Once the patient interface 12 has been applied to the patient's shin, an evaluation session can be conducted. The patient, who may be lying supine or sitting, places the heel of one foot against the top of the shin of the opposite leg. Assuming the patient is able to place the heel in the correct position, a touch sensor of the patient interface 12 detects contact between the heel and the shin. Which touch sensor detects this contact depends upon the configuration of the patient interface 12. If the patient interface 12 is one that comprises a separate top sensor, such as sensor 22 shown in
Next, the patient slowly slides his or her heel down the shin while trying to keep the heel in the centerline of the shin, where the medial touch sensor is located.
If the patient deviates from the centerline of the shin but corrects and returns the shin to the centerline of the shin to continue the test, this re-established contact is detected by the medial touch sensor. Any further deviations from the centerline of the shin will again be detected in similar manner to that described above. Accordingly, both in the downward stroke and upward stroke of the heel, the control module 14 can count the number of times the heel deviates from the centerline of the shin with great precision and accuracy. In addition, the control module 14 can determine the positions along the shin at which contact is made or lost and, potentially, in what direction each deviation occurred. Furthermore, the control module 14 can calculate the total percentage of distance along the shin or time during the evaluation that the heel either is or is not in proper alignment with the centerline of the shin. Accordingly, the system 10 collects all the information normally collected by the prior art heel-to-shin test as well as other information that may be valuable to the medical practitioner. The information gleaned by the system can be displayed in the display 46, stored within memory 62, and/or transmitted to the separate computing device.
It is also noted that the disclosed system can further include components that are used to conduct other evaluations relevant to a patient's brain or neurological system health. For example, appropriate vibration elements can be integrated into the stocking to evaluate the patient's ability to sense vibrations for the purpose of performing a neurologic vibratory sense evaluation.
Number | Name | Date | Kind |
---|---|---|---|
4062355 | Kaye | Dec 1977 | A |
5476103 | Nahsner | Dec 1995 | A |
6050962 | Kramer | Apr 2000 | A |
6692449 | Brown | Feb 2004 | B1 |
8827718 | Chiu | Sep 2014 | B2 |
9050199 | Reed | Jun 2015 | B1 |
9763848 | Handzic | Sep 2017 | B1 |
9990333 | Reed | Jun 2018 | B1 |
20070275957 | Weiner | Nov 2007 | A1 |
20080306407 | Taylor | Dec 2008 | A1 |
20090024062 | Einarsson | Jan 2009 | A1 |
20090043170 | Sulkin | Feb 2009 | A1 |
20090299227 | Thaler | Dec 2009 | A1 |
20110055753 | Horodezky | Mar 2011 | A1 |
20110208444 | Solinsky | Aug 2011 | A1 |
20110319787 | Lamoise | Dec 2011 | A1 |
20130192071 | Esposito | Aug 2013 | A1 |
20130276317 | Smirman | Oct 2013 | A1 |
20140213929 | Dunbar | Jul 2014 | A1 |
20140288383 | Barnett | Sep 2014 | A1 |
20140296749 | Reid, Jr. | Oct 2014 | A1 |
20150297132 | Bichel | Oct 2015 | A1 |
20160120733 | Ishikawa | May 2016 | A1 |
20160157779 | Baxi | Jun 2016 | A1 |
20160338621 | Kanchan | Nov 2016 | A1 |
20170025033 | Rath | Jan 2017 | A1 |
20170071816 | Vain | Mar 2017 | A1 |
20170079868 | Reid, Jr. | Mar 2017 | A1 |
20170089782 | Hirt | Mar 2017 | A1 |
20170100300 | Rapp | Apr 2017 | A1 |
20170265810 | Van De Vyver | Sep 2017 | A1 |
20170273623 | Chen | Sep 2017 | A1 |
20180028096 | Katz | Feb 2018 | A1 |
20180160940 | Kim | Jun 2018 | A1 |
20180220966 | Cohen | Aug 2018 | A1 |
20190117156 | Howard | Apr 2019 | A1 |
20190200915 | Baker | Jul 2019 | A1 |
20200178849 | Cheng | Jun 2020 | A1 |
20200367823 | Chahine | Nov 2020 | A1 |
20200371675 | Sung | Nov 2020 | A1 |
20210022666 | Malawey | Jan 2021 | A1 |
Number | Date | Country |
---|---|---|
101338049 | Nov 2012 | KR |
Entry |
---|
NYU School of Medicine, The Precise Neurological Exam, Coordinate, Gait and Rhomberg Test, https://web.archive.org/web/20191205075405/https://informatics.med.nyu.edu/modules/pub/neurosurgery/coordination.html (Year: 2018). |