The invention is generally directed to low-input chromatin immunoprecipitation followed by sequencing (ChIP-seq) technologies, in particular low-input ChIP-seq technologies allowing high-throughput and simultaneous processing of multiple samples. Exemplary applications include automated and parallel analysis of histone modifications.
The primary tool utilized for examining histone modifications is chromatin immunoprecipitation (ChIP), which applies immunoassay to capture chromatin fragments by targeting the specific type of modified histone bound to them. These chromatin fragments yield DNA fragments that can be sequenced (i.e. ChIP-seq) and provide a genome-wide map of histone binding. ChIP-seq can also be used to profile bindings of molecules such as RNA polymerases and transcription factors to the genome.
There are a couple of critical limitations associated with conventional ChIP-seq assays. First, they typically require a large number of cells (˜107-108 cells). In contrast, the sample amount generated by lab animals and patients is very limited. For example, a core needle biopsy generates a total of 104-105 cells. Circulating tumor cells are present by the frequency of 1-10 per ml of whole blood in patients with metastatic cancer. The sensitivity issue has hindered clinical research using patient materials and made patient stratification based on epigenomics impractical. Second, these assays typically require manual procedures for a duration of 3-4 days (not including library preparation and sequencing) and are not suitable for high-throughput data production. Rapid characterization of a large number of samples is vital for use of epigenomic knowledge in clinical research and patient stratification, because 1) epigenomic profiles vary among individual subjects, cell/tissue types and disease/developmental stages; 2) there are also tens to hundreds of histone marks of interests.
Microfluidics has been shown to provide a powerful platform for conducting low-input genomic, transcriptomic, proteomic and epigenomic analysis. There have been a few different attempts to utilize microfluidics to improve the ChIP process. Wu et al. developed an automated microfluidic device capable of performing ChIP-qPCR analysis using only 2000 cells, and was able to scale their platform to 16 simultaneous reactions (Wu, A. R. et al. Automated microfluidic chromatin immunoprecipitation from 2,000 cells. (Lab on a Chip 9, 1365, doi:10.1039/b819648f (2009); Wu, A. R. et al. High throughput automated chromatin immunoprecipitation as a platform for drug screening and antibody validation. Lab on a Chip 12, 2190-2198, doi:10.1039/c21c21290k (2012)). In work by inventors of the instant application, microfluidic ChIP-qPCR sensitivity was further reduced to 50 cells (Geng, T. et al. Histone modification analysis by chromatin immunoprecipitation from a low number of cells on a microfluidic platform. Lab Chip 11, 2842-2848, doi:10.1039/c11c20253g (2011)). Recently, combining a packed bed with a shear stress inducing oscillating washing step, it was demonstrated generating high-quality ChIP-seq data using as few as 100 cells (Cao, Z., Chen, C., He, B., Tan, K. & Lu, C. A microfluidic device for epigenomic profiling using 100 cells. Nat Methods 12, 959-962, doi:10.1038/nmeth.3488 (2015)).
Rotem et al., using a droplet microfluidic platform, was able to barcode chromatin from single cells and then perform ChIP-seq (Drop-ChIP)(Rotem, A. et al. Single-cell ChIP-seq reveals cell subpopulations defined by chromatin state. Nat Biotechnol 33, 1165-1172, doi:10.1038/nbt.3383 (2015)). While useful for understanding single cell heterogeneity, Drop-ChIP yielded a low number of reads per single cell (1000 unique reads per cell) and required a large pooling of single cells (on the order of 103 cells) to generate datasets of sufficient quality as reference epigenomes. Other state-of-the-art ChIP technologies (iChIP, FARP-ChIP, LiNDA, nano-ChIP) all work at 500-10,000 cells per assay range. While high throughput analysis does exist in the form of AHT-ChIP, the analysis requires 100 million cells per assay, more than traditional ChIP-seq.
In previous works, an inventor of the instant invention designed a packed bed of immunomagnetic beads for highly efficiently collection of ChIP DNA that enabled ChIP-seq using as few as 100 cells. In spite of the high adsorption efficiency at the theoretical limit, a packed bed of beads creates substantial pressure drop in the microfluidic structure, and this problem would be further confounded were running multiple units in parallel attempted.
According to an aspect of some embodiments of the invention, the invention demonstrates Low Input Fluidized bed Enabled ChIP (LIFE-ChIP), a microfluidic platform for running multiple parallel ChIP assays simultaneously by utilizing microfluidic fluidized beds. Our device permitted running 4 ChIP-seq assays in one run with as few as 50 cells per assay. The supporting data has high reproducibility among various individual units and devices. An exemplary LIFE-ChIP assay could be finished in as little as 1 hour. The technology paves the way to high-throughput epigenomic profiling required by precision medicine.
There are several distinct advantages associated with LIFE-ChIP-seq. First, this fluidized-bed based ChIP technology permits running multiple parallel assays simultaneously. Such a feature is critical for high-throughput processing of samples needed for the examination of patient materials in the precision medicine settings. Furthermore, the fluidized bed technology effectively alleviates pressure building associated with bead manipulation in a microfluidic system.
Exemplary embodiments may permit production of multiple replicates from one sample and/or examination of multiple samples, all with a single device. A prototype embodiment detailed in the Examples demonstrates producing 4 replicates on one type of histone modification or 2 replicates each on two types of histone modifications from one sample.
LIFE-ChIP-seq is also a low-input technology. A prototype embodiment produced high-quality data with 100 or less cells per assay. This is comparable to or better than other state-of-the-art ChIP-seq technologies such as MOWChIP-seq and iChIP (Cao, Z., Chen, C., He, B., Tan, K. & Lu, C. A microfluidic device for epigenomic profiling using 100 cells. Nat Methods 12, 959-962, doi:10.1038/nmeth.3488 (2015); Lara-Astiaso, D. et al. Immunogenetics. Chromatin state dynamics during blood formation. Science 345, 943-949, doi:10.1126/science.1256271 (2014)). The low-input characteristic of exemplary embodiments is important for profiling primary cell types with low abundance.
An exemplary microfluidic process of LIFE-ChIP-seq may be largely automated. There is very little input from the operator during the process. Thus the technology may largely eliminate human errors and save labor.
Generally, a processing of using an exemplary microfluidic device comprises on-chip steps of loading of antibody-coated beads into a LIFE-ChIP-seq platform containing multiple parallel unit; flowing of the chromatin fragments through the fluidized beds (i.e. chromatin immunoprecipitation), and flowing of washing buffer through the fluidized beds for removing nonspecifically bound chromatin. Rotating schemes may be applied during loading and washing steps to keep even distribution of reagents among chambers and good fluidization of the beads.
An exemplary microfluidic device may comprise a plurality of chambers configured to accommodate fluidized beds for chromatin immunoprecipitation assays; one or more inlet ports configured or configurable to be in fluid communication with the plurality of chambers; a plurality of micromechanical valves actuatable in different combinations to control and change which of the plurality of chambers are in fluid communication with the one or more inlet ports and which of the plurality of chambers are not in fluid communication with the one or more inlet ports at different stages of operation; a magnetic field generating device for generating a magnetic field in the plurality of reaction chambers for manipulating magnetic beads; and a plurality of outlet ports for recovering contents from the chambers.
An exemplary method of performing chromatin immunoprecipitation (ChIP) for multiple parallel assays may comprise loading a plurality of parallel units with antibody-coated immunomagnetic beads; flowing chromatin molecules from an inlet into fluidized beds containing the immunomagnetic beads to form chromatin-bead conjugates; washing the chromatin-bead conjugates with a plurality of sequential wash buffers flowed through the fluidized beds; and collecting washed chromatin-bead conjugates for DNA release.
a. 2D top view schematic of an exemplary microfluidic LIFE-ChIP device. A glass slide that is bonded to the PDMS part is not shown.
c. 3D perspective view of the microfluidic networks within the LIFE-ChIP device with a two-layer structure. The valves 102 are all in the control layer and the rest of the structures are in the fluidic layer. A glass slide that is bonded to the PDMS part is not shown.
The microfluidic device 100 comprises two layers: a control layer which can be pressurized and a fluidic layer which allows the flow of reagents, molecules and particles. The fluidic layer comprises the inlet ports 101, channels 103, reaction chambers 105, and outlet ports 107. The control layer comprises the valves 102, which in some exemplary embodiments may be micromechanical valves. Together the two layers (fluidic layer and control layer) allow for fluid/particle manipulation within the channels 102 and reaction chambers 105 via the micromechanical valves 102. The control layer is adjacent and in a fixed position with respect to the fluidic layer. As non-limiting exemplary dimensions, structures of control layer may be, for example, 50 μm to 20 μm in depth.
As illustrated by
A microfluidic device 100 is configured for performing an assay or part thereof, in particular a ChIP-seq assay or a part thereof. Generally, a ChIP-seq assay for examining histone modifications typically involves several steps: i) cross-linking to fix histones to DNA sequences that they interact with; ii) sonication or enzymatic digestion to generate chromatin fragments; iii) chromatin immunoprecipitation (ChIP) to adsorb chromatin fragments containing the histone of interest on immunomagnetic beads that are functionalized with an antibody targeting a specifically modified histone; iv) release of the DNA fragments (i.e. ChIP DNA) from bead surface; v) sequencing of the DNA fragments and establishing genome-wide profile for the histone modification. Exemplary microfluidic devices according to the invention may perform one more of these steps. In particular, an exemplary microfluidic device 100 may perform at least step (iii), namely chromatin immunoprecipitation. Other steps (e.g., (i), (ii), (iv), and/or (v)) may be perform off-chip or on-chip. Variations of the microfluidic device 100 depicted in the figures may be made within the spirit and scope of the invention to include one or more of the remaining ChIP-seq steps on-chip or off-chip with respect to the chromatin immunoprecipitation step.
Exemplary operation of a microfluidic device 100 is demonstrated in
First, as shown in
The beads may be referred to as immunomagnetic beads, or magnetic immunoprecipitation beads. For the sake of uniformity, most of this description uses the term “immunomagnetic beads,” but different types of beads may be employed in different embodiments.
To use specific conditions as a non-limiting example, a bead suspension (6 μg/μl) may be flowed into one of 4 reaction chambers at a flow rate of 8 μl/min while the valves alternated flow among the chambers in the order of chamber C1-C3-C2-C4 at 4 s intervals to ensure even loading. In a prototype embodiment using the conditions just identified, each loading cycle took 16 s (4 sec for each of the 4 chambers) and all 4 chambers were finished loading in 20 cycles. During the loading process a magnet 207 is situated adjacent to the reaction chambers. The magnet may be repositionable with respect to the plurality of reaction chambers. As depicted in
Next, as depicted by
The chromatin loading step allows targeted chromatin fragments to adsorb on the beads' surfaces (i.e. ChIP). Sufficient time is allowed for immunoprecipitation, e.g., 30 to 120 minutes. Broadly speaking, immunoprecipitation time may be 10-180 minutes. A narrower exemplary range is 30-60 minutes. For example, immunoprecipitation time may be no longer than 10, 15, 20, 25, 30, 35, 40, 45, 50, 55, or 60 minutes from start to completion. An exemplary time breakdown for a microfluidic device with 4 reaction chambers is as follows. Chromatin solution is directed into one of 4 reaction chambers while the on-chip valves alternated flow among the chambers in the order of chamber C1-C3-C2-C4 with intervals varying from 8 to 0.25 s for a total of 104 s (cycle 1: 8 s per chamber, cycle 2: 6 s per chamber, cycle 3: 4 s per chamber, cycle 4: 3 s per chamber, cycle 5: 2 s per chamber, cycle 6: 1 s per chamber, cycle 7-8: 0.5 s per chamber, cycle 9-12: 0.25 s per chamber). After these 12 loading cycles (i.e. rotating scheme 2), all chambers are opened to load chromatin into them simultaneously (
After the chromatin loading and adsorption, wash buffers are flowed sequentially through the fluidized beds of the reaction chambers, as depicted in
After performing all washing steps, the beads may be collected from the device outlet ports 107. This may be accomplished by removing the magnet 207, opening the valve 203 (if not already open) configured for closing off the reaction chambers from the outlet ports, and flowing an IP buffer to flush the beads from the reaction chambers. Exemplary conditions for this collection step are, for example, a flushing rate of 400 μl/min for 15 s.
The washed and collected chromatin-bead conjugates may then be subject to ChIP DNA release (off-chip) and sequencing library preparation. These steps may be performed according to known best practices. One example is detailed below in the Examples section.
A microfluidic device according to the invention may comprise a controller or be an element of a system which further includes a controller. The controller is attached or attachable to the control layer of the microfluidic device to control the valves and may also control the liquid delivery into the inlets and liquid removal from the outlets.
Exemplary systems, devices, and methods may be characterized according to a number of parameters, including input volume flow rate (of respective reagents, solutions, suspension, buffers, etc. admitted to the fluidic layer through one or more inlet ports), magnet location or locations (at one or more stages of the process), washing time, and overall time of the LIFE-ChIP assay, among others. Exemplary operational conditions encourage high levels of fluidization for washing and a dense packing of beads for the chromatin loading step, all while avoiding bead loss in each chamber.
Operational conditions including flow rate and magnetic field location are summarized in
Of the three magnet positions identified in
Washing time may be characterized as the duration for flowing each wash buffer through fluidized bed to remove nonspecifically adsorbed chromatin fragments. Washing removes nonspecifically adsorbed chromatin and improves enrichment of ChIP DNA. However, excessive washing can decrease the amount of ChIP DNA collected thus require additional amplification and result in lower-quality sequencing library. Washing times may be for, example, between 0 and 30 min with 1000 cells per assay (unit), more preferably 10 to 30 min. No washing, while possible, is generally disadvantageous because it generally yields lower-quality data. By contrast washing times from 10 to 30 min generally yield high-quality data. Similarities between data obtained with washing times 10-30 min suggested that the fluidized bed washing was effective for removing nonspecific chromatin molecules (see EXAMPLES below). A fluidized bed implemented in LIFE-ChIP decreases the possibility for physical trapping as compared to prior packed bed reactors.
Depending on the configuration of specific parameters discussed above for individual alternative embodiments, the total time of a LIFE-ChIP assay may vary. However, exemplary total on-chip times range from 1 hour to 3 hours, for example, with acceptable quality of resulting ChIP-seq data. The 1 to 3 hr duration includes multiple parallel assays, e.g., two, four, or more than four. An exemplary microfluidic device is scalable to tens (e.g., at least 10, 20, 30, 40, 50, 60, 70, 80, 90), hundreds (e.g., at least 100, 200, 300, 400, 500, 600, 700, 800, 900), or more parallel assays. Generally speaking, an exemplary microfluidic device is scalable to any size.
An exemplary method of manufacturing a fluidic device like device 100 shown in the figures is by multi-layer soft lithography. The fluidic layer, the control layer, or both layers may be monolithic, individually or collectively. Suitable materials include, for example, polymers (e.g., polydimethylsiloxane, PDMS). A specific example manufacturing method employed with a test prototype is described below in the EXAMPLES.
Different types of beads may be used with embodiments of the invention. In particular, two or more (e.g., 2, 3, 4, 5, 10, 20, 50, or more) types of beads with different antibody-coatings may be used in a single run on a single microfluidic device 100.
Following is a non-limiting example demonstrating manufacture, use, and performance of a prototype microfluidic device consistent with the foregoing descriptions and in particular the figures. The example is demonstrative that an exemplary microfluidic device performs well with low-input (e.g., <100 cells per assay) and high-throughput (e.g., four assay in an hour). LIFE-ChIP-seq data are deposited under accession number GSE 102932 at Gene Expression Omnibus (https://www.ncbi.nlm.nih.gov/geo/query/acc.cgi?acc=GSE102932).
The microfluidic system consisted of 4 reaction chambers (units) connected to 7 input channels with successive splits. There were two-layered micromechanical valves (Unger, M. A., Chou, H.-P., Thorsen, T., Scherer, A. & Quake, S. R. Monolithic Microfabricated Valves and Pumps by Multilayer Soft Lithography. Science 288, 113-116, (2000).) to close/open each and every inlet and split. The bell-shaped reaction chambers had a major axis of 9 mm and a minor axis of 3.5 mm and included microscale pillars (with a diameter of 140 μm) to prevent collapse. The reaction chambers had a depth of 50 um, while the parts that formed fully closed valves had a depth of 25 μm.
The chip was manufactured using two-layer soft lithography (Unger, M. A., Chou, H.-P., Thorsen, T., Scherer, A. & Quake, S. R. Monolithic Microfabricated Valves and Pumps by Multilayer Soft Lithography. Science 288, 113-116, (2000).) in polydimethylsiloxane (PDMS, RTV615, Momentive Advanced Materials). Photomasks were prepared using LayoutEditor and printed on Mylar transparencies at high resolution (10,160 DPI) by FineLine Imaging. The photomasks were used to fabricate 2 separate master molds using photolithography on silicon wafers (3 inch, university wafer). The control layer and a part of the fluidic layer (some channels and reaction chambers) were fabricated using SU-8 2025 (MicroChem) to create rectangular channels (depth ˜50 um). The fully closed valves of the fluidic layer were fabricated by spinning 2 layers of AZ 9260 (EMD Performance Materials) each at 1500 RPM for 30 seconds, with a 10 min wait between the spins to allow evaporation of solvent (depth ˜20 um).
The fluidic layer master was then heated to 130° C. for 60 s to round the AZ channels. Prepolymer PDMS was mixed in a ratio of 5:1 prepolymer to crosslinking agent, degassed, and cast as a bulk layer (˜5 mm thick) on the fluidic layer mold, while a degassed 20:1 ratio of PDMS was spun onto the control layer mold at 1750 RPM for 30 s. After partial curing at 75° C. for 12 min, the fluidic layer casting was removed from the silicon master and inlet and outlet ports were punched using a 1.5 mm core punch (Harris Uni-Core). The fluidic layer was laid on top of the control layer (that was still attached to its master), aligned and cured at 75° C. for 1 h. The multilayer device was then removed from the mold and punched, using a 1.5 mm core punch. The PDMS device and a glass slide (50 mm×75 mm, Corning), precleaned with dish soap, were treated by plasma (Harrick Plasma) then bonded together irreversibly. The glass bound PDMS chips were cured at 75° C. for 30 min to strengthen bonding and reduce air bubbles before being used.
GM 12878 cells were obtained and cultured as described in Cao, Z., Chen, C., He, B., Tan, K. & Lu, C. A microfluidic device for epigenomic profiling using 100 cells. Nat Methods 12, 959-962, doi:10.1038/nmeth.3488 (2015). GM cells were cultured in RPMI 1640 supplemented with 15% FBS, 100 U penicillin, and 100 mg/mL streptomycin at 37° C. in a humid incubator with 5% CO2. The cells were passaged every 2 d in order to maintain log phase growth.
Sonicated chromatin was prepared as described in Cao, Z., Chen, C., He, B., Tan, K. & Lu, C. A microfluidic device for epigenomic profiling using 100 cells. Nat Methods 12, 959-962, doi:10.1038/nmeth.3488 (2015). Sample containing 106 cells was used as the starting material and centrifuged at 1,600 g for 5 min and washed twice with 1 ml cold PBS. Cells were crosslinked by adding 1 ml freshly made 1% formaldehyde and incubated at room temperature for 5 min on a shaker. Crosslinking was promptly terminated by adding 50 μl of 2.5 M glycine and shaking for 5 min. The pellet was resuspended in 130 μl Covaris buffer (10 mM Tris-HCl, pH 8.1, 1 mM EDTA, 0.1% SDS and 1× protease inhibitor cocktail) and sonicated using the following conditions. Using Covaris M 220 (Covaris), the sample was sheared at 4° C. with 75 W peak incident power, 5% duty factor, and 200 cycles per burst for 12 min. The sheared sample was centrifuged in a 4° C. centrifuge at 14,000 g for 10 min, and the supernatant was transferred to a new tube. Sheared chromatin was diluted with IP buffer (20 mM Tris-HCl, pH 8.0, 140 mM NaCl, 1 mM EDTA, 0.5 mM EGTA, 0.1% (w/v) sodium deoxycholate, 0.1% SDS, 1% (v/v) Triton X-100, with 1% freshly added PMSF and PIC) and aliquoted into 25 μl samples containing 250, 500, 1500, and 5,000 cell samples and stored at
−80° C. Selective chromatin samples were purified by ethanol precipitation for fragment analysis using HS D1000 tapes on Tapestation 2200 (Agilent) (
Protein A coated superparamagnetic Dynabeads (Invitrogen) were utilized for immunoprecipitation of sheared chromatin. 10 μl of bead suspension was washed twice with 100 μl freshly prepared blocking buffer (made by mixing 467 μl filtered PBS and 33 μl of 7.5% Bovine Albumin Fraction V solution) in each wash. The beads were incubated for 2 h on a rotator at 4° C. in 300 μl blocking buffer containing 5, 10, or 20 μl of H3K4me3 antibody (07-473, Millipore) for 50, 100-300, and 1000 cells per CUP assay, respectively. The beads were then washed twice in 100 μl IP buffer, before being resuspended in 50 μl IP buffer for on-chip use.
A microfluidic system corresponding to system 300 in
CUP and input chromatin samples (beads or 10 μl solution, respectively) were incubated overnight at 65° C. with 200 μl (190 μl for input) elution buffer (200 mM NaCl, 50 mM Tris-HCl, 10 mM EDTA, 1% SDS, 0.1M NaHCO3) supplemented with 2 μl of 10 μg/μl proteinase K. After incubation, DNA was isolated from protein debris and beads by adding 200 μl of Phenol:Chloroform:Isoamyl Alcohol 25:24:1 to the elution mix and vortexing. After a 5 min centrifuge at 16,100 g, the supernatant (200 μl) was removed and mixed with 750 μl 100% ethanol, 50 μl of 10 M ammonium acetate, and 2 μl of 5 μg/μl glycogen. The solution was vortexed and incubated at −80° C. for 2 h. After incubation, the samples were centrifuged at 16,100 g at 4° C. for 5 min. The supernatant was removed, and 500 μl of 70% ethanol was added to the tube without disturbing the pellet. After a 5 min centrifuge under the same conditions, the solution was removed, and the pellet air dried. The pellet was then dissolved into 40 μl of water for library preparation.
Library preparation was performed using the Swift Bio S2 library preparation kit (Swift Biosciences) using 40 μl of purified DNA. The manufacturer's instructions were followed with minor modifications. Specifically, 2.5 μl of 20× EvaGreen were added into the 50 μl amplification reaction mix, and amplification was terminated after samples saw a >3000 RFU increase (in a BioRad CFX Connect). After DNA purification with SPRI beads, DNA was then eluted into 7 μl low EDTA TE buffer where 2 μl could be used for qPCR analysis for preliminary quality control analysis, Kappa DNA quantification, and Tapestation fragment size analysis and the other 5 μl for library pooling. Libraries were pooled at 10 nM for sequencing by Illumina HiSeq 4000 with single-end 50 nt read.
Sequencing reads were first trimmed using Trim Galore before being aligned to the hg19 genome using bowtie v2.4 with default settings. Normalized reads were computed across the genome in 25 bp bins according to the following equation:
For visualization, the reads were elongated to 250 bp reads to more accurately describe the read DNA before normalization and were normalized according to an input sample also with elongated reads. The normalized samples were converted to bigwig format using UCSC bedGraphtoBigWig.
Uniquely mapped un-normalized reads were used for peak calling. Peak calling was performed using MACS using a P value <10−5 with default settings.
Correlation analysis was carried out in order to examine the consistency among different datasets. Promoter regions were defined as 2 kb upstream and downstream from the transcription start sites extracted from RefSeq data. The average normalized signal over each promoter region was quantified and the correlation was calculated using a custom Perl script.
The prototype LIFE-ChIP-seq platform consisted of 7 inlet ports connected to 4 bell-shaped reaction chambers through a series of splits (as depicted in
The operational conditions (flow rate and magnet location) summarized in
The platform was operated in several steps (see
Using the above-described platform, examination was made of tri-methylation of lysine 4 on histone H3 (H3K4me3) in GM12878 cells (a human B-Lymphocyte cell line). All data are summarized in Table 1.
Samples were tested containing various numbers of cells 4000, 1200, 400, and 200 cells (equivalent to 1000, 300, 100, and 50 cells per ChIP-seq assay) (
With a parallel system, it is also important to examine reproducibility across different devices.
The LIFE-ChIP-seq platform operation was optimized on a number of different parameters. The input volume was optimized at the flow rate of 1 μl/min, which was selected from the fluidized bed behavior described in
Washing time was another optimized parameter. Washing times were explored varying between 0 and 30 min with 1000 cells per assay (unit). Our LIFE-ChIP-seq data showed that no washing yielded lower-quality data (with average correlation among units of 0.917) which, nevertheless, yielded an average correlation of 0.915 with ENCODE data (
Finally, how the overall time of the LIFE-ChIP assay affected the results was also examined. Using the minimal conditions determined through the optimizations described above, a LIFE-ChIP-seq experiment was performed using 30 min of loading and 10 min of washing per buffer, which equated to a 1 h on-chip assay compared to the 3 h assay (2 h loading and 20 min per washing step) that was set as a baseline. The 1 h assay showed a slightly higher average inter-unit correlation than the 3 h assay (0.964 vs. 0.956) (
Finally, using a single device, two histone marks (H3K4me3 and H3K27ac) were profiled in one run. Two types of antibody-coated beads (one coated with anti-H3K4me3 and the other with anti-H3K27ac) from two separate inlets were loaded sequentially while avoiding cross contamination (
While the invention has been described in terms of exemplary embodiments, those skilled in the art will recognize that the invention can be practiced with modification within the spirit and scope of the appended claims. Accordingly, the present invention should not be limited to the embodiments as described above, but should further include all modifications and equivalents thereof within the spirit and scope of the description provided herein.
This application claims the benefit of U.S. provisional patent application No. 62/573,957, filed Oct. 18, 2017, the complete contents of which are herein incorporated by reference.
This invention was made with government support under grant numbers EB017235 and CA214176 awarded by the National Institutes of Health. The government has certain rights in the invention.
Number | Date | Country | |
---|---|---|---|
62573957 | Oct 2017 | US |