The present disclosure relates to ultrasonography, and more particularly to systems and methods for providing augmented ultrasonography with implanted electronic device(s), including Complementary Metal Oxide Semiconductor (CMOS) mote(s).
There have been decades of technological advances in the field of ultrasound medical diagnostic imaging. With pioneering work1 dated back to the 1950s, medical ultrasound, or ultrasonography, has evolved to become a routine imaging procedure for real-time tomography. Compared to other medical tomographic techniques, it has many advantages because of the unique properties of ultrasound waves. In addition to being non-ionizing, the low acoustic energy loss in tissue (approximately 0.5˜1 dB/cm/MHz2,3) allows for significant imaging depth while high spatial resolution is possible because of the short wavelength characteristic of acoustic energy in tissue (about 0.5 mm at a typical carrier frequency of 3 MHz). Imaging frame rates are typically in excess of 25 frames per second (fps) due to advances in imaging hardware and software.
At the same time and motivated by many of the same properties that make it attractive for diagnostic imaging, ultrasound is emerging as a means for powering and communicating with implanted medical devices4-8. Wireless powering is attractive because it eliminates the need for batteries, which introduce extra safety concerns and consume considerable volume. What is particularly advantageous for ultrasound in these applications is that attenuation in tissue is significantly less for ultrasound when compared with electromagnetic waves at comparable wavelengths9. Wavelength is important because it determines minimum antenna sizes and, consequently, the size of the implanted devices themselves. The low absorption of ultrasound in tissue also supports average power densities as high as 7.2 mW/mm2 without deleterious effects10. For electromagnetic radiation, the specific absorption ratio (SAR) limits incident average power densities to less than 0.22 mW at 1.6 GHz, for example, in the case where an antenna is positioned at the tissue surface11.
To achieve μW-level power budget, previous devices based on ultrasound for wireless powering rely on focused, almost continuous ultrasound “beams”. Traditional digital modulation techniques can be then implemented on top of such beams for data transmission, like on-off keying (OOK)12, and pulse-width modulated amplitude-shift keying (PWM-ASK)6,13. Uplink data can be captured in the acoustic domain once the device is properly powered, in the form of either backscattered incident ultrasound5,14,15 or active transmission of ultrasound energy from the implant12. In all of these implementations, accurate knowledge of the implant location is required for power delivery and communication. This location information can be obtained either through an additional imaging session prior to the operation of the implant or a designed localization process on top of normal operation. One such designed process uses a sharp ultrasound pulse emitted from the implant to back calculate the delays in each transmitter element for focusing, but the implant needs to harvest some power before the chirp can be emitted16-18. Another example extracts this delay information from higher order harmonics of the implant's reflected acoustic waves19; but without an active signature, harmonics in the tissue, long used in ultrasound harmonic imaging20, can potentially shadow this information. Even if the implant can be localized, in many in vivo settings, continuous movement due to muscle activity, heartbeat, and respiration can perturb device location, requiring frequent recharacterization to maintain focus.
Thus, it may be beneficial to provide exemplary cavity-separated multi-nanopore (CSMP) device and method which can facilitate protein sequencing, which can overcome at least some of the deficiencies described herein above.
The following is intended to be a brief summary of the exemplary embodiments of the present disclosure, and is not intended to limit the scope of the exemplary embodiments.
According to various exemplary embodiments of the present disclosure, the operation of the exemplary specifically designed ultrasound mote21 can be integrated and/or utilized with medical sonography, facilitating such combination to be biogeographically located in the image, to be powered by the ultrasound imager, and to communicate back through the image. For example, the mote (e.g., the implantable device) can stay continuously powered when located within the ultrasound imaging sensor's field of view (FoV). Bi-directional data transmission is established synchronously with the imaging frame rate and the uplink information (from implant to the imager) can be retrieved during image reconstruction. Multiple motes can be deployed within a given FoV with non-interfering parallel operation, as digital uplink data signatures are spatially separated in the reconstructed image. Such exemplary digital data signatures give the motes contrast relative to the surrounding, continuously moving biological environment and can be used to localize them at the resolution of the resulting B-mode image. These exemplary sensors facilitate an “augmented ultrasonography”, delivering real-time, biogeographically related physiological information from multiple locations on top of a traditional ultrasound tomographic imaging.
The use of imaging-mode ultrasound can impose a restricted power budget on the motes, typically nano-Watts (nW) or below, as well as a data rate limited to the imaging frame rate, which is typically in the range of 10s of Hz. Fortunately, these data rates are sufficient for most physiological parameters, such as temperature, blood pressure, pH, and most biomarkers. There are many CMOS sensor systems available that operate with nW or sub-nW power consumption that would be compatible with this power envelope at the bandwidths of interest14,22-24. Even in the absence of integrated sensing, motes such as these transmitting unique identifiers can find application in real-time tracking of surgical sites, such as in intramedullary nailing25,26.
Indeed, modern clinical practice benefits significantly from imaging technologies and much effort is directed toward making this imaging more informative through the addition of contrast agents or reporters. Here, an exemplary design of a battery-less integrated circuit mote acting as an electronic reporter during medical ultrasound imaging can be provided. When implanted within the field-of-view of a brightness-mode (B-mode) ultrasound imager, this exemplary mote can transmit information from its location through backscattered acoustic energy which is captured within the ultrasound image itself. It is possible to characterize the operation of such motes in vitro and in vivo. Performing with a static power consumption of less than 57 pW, the motes operate at duty cycles for receiving acoustic energy as low as 50 ppm. Motes within the same field-of-view during imaging have demonstrated signal-to-noise ratios of more than 19.1 dB at depths of up to 40 mm in lossy phantom. Physiological information acquired through such motes, which is beyond what is measurable with endogenous ultrasound backscatter and which is biogeographically located within an image, has the potential to provide an augmented ultrasonography.
To that end, an exemplary system according to the exemplary embodiments of the present disclosure can be provided for use with ultrasound imaging. The exemplary apparatus can include at least one implantable device which is configured to (i) communicate with an ultrasound imaging system, (ii) be located by the ultrasound imaging system using ultrasound signals generated thereby to generate a location of the implantable device(s), and (iii) transmit data to the ultrasound imaging system from the location of the implantable device(s) that was located by the ultrasound imaging system.
Further, an exemplary method can be provided for performing ultrasound imaging according to the exemplary embodiments of the present disclosure. With the exemplary method, it is possible to implant, into a structure, at least one device which is responsive to ultrasound signals generated by the ultrasound imaging system, locate the device(s) within the structure by the ultrasound imaging system using the ultrasound signals to generate a location of the device(s) within the structure, and transmit data to the ultrasound imaging system from the location of the device(s) that was located by the ultrasound imaging system.
In further exemplary embodiment of the present invention, it is possible to power the implantable device(s) by the ultrasound imaging system. The implantable device(s) can be continuously powered by the ultrasound imaging system when the implantable device(s) is located within a field of view of the ultrasound imaging system. In addition, it is possible to transmit the data to the ultrasound imaging system by the implantable device(s) using modulation. The implantable device(s) can be further configured to be implanted into the structure or a body through an injection. The implantable device(s) can be sized and configured to be injected into the structure of the body using a syringe.
According to still further exemplary embodiments of the present disclosure, it is possible to receive ultrasound signals from the ultrasound imaging system by the implantable device(s) so that the ultrasound imaging system ascertains the location of the implantable device(s). The transmission of the data to the ultrasound imaging system and the receipt of the ultrasound signals can be synchronous. The receipt of the ultrasound signals by the implantable device(s) from the ultrasound imaging system can be performed using a further modulation. The further modulation of the received ultrasound signals can be performed using at least one amplitude shift key, and the modulation of the transmission of the data can be performed using at least one load shift key back.
In a further exemplary embodiment of the present disclosure, the implantable device(s) can be a CMOS device or a monolithically-integrated device (e.g., a monolithically-integrated piezoelectric transducer). There can also be a plurality of implantable devices. It is possible to deploy each of the plurality of the implantable devices within a field of view of the ultrasound imaging system and providing a non-interfering parallel operation. Additionally or alternatively, each of the plurality of the implantable devices can be implanted at different locations within the structure.
According to yet an additional exemplary embodiment of the present disclosure, the transmission of the data from the e implantable device to the ultrasound imaging system can be performed using a backscattered acoustic energy.
These and other objects, features and advantages of the exemplary embodiments of the present disclosure will become apparent upon reading the following detailed description of the exemplary embodiments of the present disclosure, when taken in conjunction with the appended claims.
Further objects, features and advantages of the present disclosure will become apparent from the following detailed description taken in conjunction with the accompanying Figures showing illustrative embodiments of the present disclosure, in which:
Throughout the drawings, the same reference numerals and characters, unless otherwise stated, are used to denote like features, elements, components or portions of the illustrated embodiments. Moreover, while the present disclosure will now be described in detail with reference to the figures, it is done so in connection with the illustrative embodiments and is not limited by the particular embodiments illustrated in the figures and the appended paragraphs.
Exemplary B-Mode Sonography and Challenges in the Physical Layer.
Harvesting energy in the context of B-mode sonography can be performed from an ultrasound energy that can be very sparse in time. For the frame rate of 50 fps employed in the exemplary case according to the exemplary embodiment of the present disclosure, each 1-μs-long pulse provides power to the mote with a duty cycle of 50 parts per million (ppm). The energy that can be available for power harvesting scales as approximately 0.33 pJ/(kPa)2 for each pulse with energy at a MHz-level center frequency. The energy harvesting circuits should be able to respond to these frequencies without significant static power consumption from biasing and leakage currents. Energy storage in the form of decoupling capacitors can be required to maintain continuous powering. At a 50-ppm duty cycle and a minimum incident pressure of 400 kPa, which is typical for the imaging studies here, e.g., the decoupling capacitance should be at least 100 pF. Making the decoupling capacitance larger than minimum value, however, extends the start-up time, the time interval between when the imaging probe first find the mote to the time the mote generates a distinguishable data signature, since a supply voltage can reach at least 1.2 V before the mote can begin to operate.
An exemplary data transmission from the mote to the imager can take advantage of the way B-mode imaging is performed. For example,
This can be verified in the measured hydrophone data which provides a measured pulse envelope at different distances from the source linear array that is shown in the exemplary graphs of
To ensure that the mote can establish a reliable data link when positioned anywhere within the FoV, the downlink data as transmitted by the imaging transducer can be maintained constant for the entire frame. At the mote side, backscattered data can also be maintained constant for each imaging frame. This can reduce or eliminate any dependency data transmission may have on the exact shape of the pulse packet, since every pulse within each packet carries the same downlink or uplink bit. The data rate in this case is determined by the frame rate.
Exemplary System Level Design.
Such exemplary received pulses can be used to recover a clock 330 of
Turning to
The Downlink example of
Exemplary Rectifier Design and Power Regulation. To harvest heavily duty-cycled ultrasound energy, it is possible use an active rectifier (show in detail in
Many previous efforts have relied on operating the piezoelectric transducer at the resonance of the piezoelectric crystal for maximum power delivery28,29. However, those approaches introduce two problems. First, a resonant power delivery requires a high-quality mechanical resonance, which, in turn, demands highly mismatched boundary conditions around the piezo transducer of interest. This can be achieved by adding an air-pocket-based backing structure underneath the piezo transducer12, at the cost of a more complicated packaging as well as an increased mote volume. Second, significant impedance changes typically occur within the approximately 100-kHz bandwidth around the package-defined resonance frequency. As a result, the amount of power delivered strongly depends on the frequency tuning resolution of the imaging transducer array, making the motes difficult to power in practice.
For these reasons, according to exemplary embodiments of the present disclosure, it is possible to assume a capacitive input impedance, a typical case when the piezo transducer is surrounded by non-optimized mechanical boundary conditions. Resonance is still possible in the electrical domain with the help of an inductor for conjugate matching, although it is possible to choose not to do this because it would significantly increase implant volume. Instead, to boost the power harvesting without resonance, it is possible to employ a “switch-only” rectifier, a technique commonly used in low-frequency mechanical energy harvesting circuits30 but adapted here for carrier frequencies in the MHz regime for reliable, nW-level power harvesting. Such an approach consumes little on-chip area while relaxing the requirement for a high-quality mechanical resonance from the piezoelectric element. These exemplary benefits together can facilitate an efficient off-resonance power harvesting and the potential to scale to sub-0.1-mm3 range implant volumes with piezoelectric transducer integration14,31.
The exemplary rectifier according to exemplary embodiments of the present disclosure can generate, e.g., an 1.2-V dc supply, VCC, which can be clamped by the voltage limiter, and can be used to power circuits which should have enough performance to track the MHz-level ultrasound carrier. A 0.5-V VDD is generated from VCC with a linear voltage regulator 385 (shown in
Exemplary Clock Synchronization and Bi-Directional Data Telemetry. Once sufficient energy is harvested for start-up, the mote establishes synchronized bi-directional data transmission with the imager.
To support the data downlink, it can be important and/or beneficial to measure the duration of one of the pulses in the pulse packet. For example, it is possible to count the number of cycles with the counter 392 in the first pulse that meet the threshold for clock recovery within each pulse packet (shown in
The uplink data appears as a frame-to-frame change in intensity within the reconstructed image, which results from a change in the backscattered echo. To implement this exemplary backscatter ASK, the exemplary circuits can modulate the effective acoustic impedance of the transducer by changing the electrical load attached to it. For example, it is possible to maximize the modulation depth by shorting the piezoelectric element through MBS (see
This exemplary approach provides three issues which can be addressed in the exemplary design according to the exemplary embodiments of the present disclosure. First, when the transducer is shorted, no clock generation occurs from the pulse packet in the given frame. This missing clock cycle provides no mechanism to synchronously recover from sending a 0. To rectify this issue, the backscatter modulator according to the exemplary embodiment of the present disclosure can asynchronously reset any bit 0 to 1. This exemplary approach extends each bit 0 to take two clock cycles. To maintain a data-independent frame length, each bit 1 is also artificially extended to take two cycles, such that the uplink data rate is data-independent at one half of the frame rate. Second, no downlink data can be recovered when an uplink 0 is transmitted; this precludes full-duplex operation. Third, the power harvesting is uplink-data dependent. As a result, the implant should be able to operate in the worst case at half the maximum power that would otherwise be harvestable for the case of a continuous string of 0's being sent in the uplink channel.
Exemplary Link Layer and Application Layer. To implement the digital control of the chip, a low-leakage standard-cell library according to an exemplary embodiment of the present disclosure can be employed. To facilitate a unique downlink communication to multiple implants within the same FoV, an identification generator can be embedded in the application layer that assigns a unique identification sequence to each mote, implemented with, e.g., an eight-bit fuse array.
Exemplary Post Processing and Packaging. The exemplary mote ASIC 405 can be fabricated or otherwise provided using, e.g., a TSMC 180 nm MSRFG 1P6M process, occupying a volume of, e.g., 830 μm×740 μm×300 μm (shown in
Exemplary Performance Characterization. Initial electrical testing can be performed to verify the functionality of the ASIC. To stay fully functioning, the exemplary chip can use preferably, e.g., 57 pW of static power when the dc output of the rectifier reaches 1.2 V. Achieving such ultra-low power consumption can facilitate a higher acoustic energy attenuation, supporting mote operation in deeper tissue.
For example, for ultrasonography, a Verasonics Vantage 256 research ultrasound imaging systems can be connected to an L12-3V (Verasonics Inc.), a 192-element linear array probe.
To harvest sufficient power and deliver detectable uplink modulation depth, the mote can require the focused pulse amplitude in a pulse packet to reach approximately, e.g., 400 kPa, at which point each pulse provides an input energy of approximately, e.g., 52.8 nJ. For example, motes can start-up after power regulation is complete, which can typically utilize about 235 frames or 4.7 seconds. In the exemplary reconstructed floating-point-valued B-mode image 510s, the uplink data sent from the mote can lead to a significant, regular brightness change, which aids in mote localization. The ratio between the signal power and the noise power, or signal-to-noise ratio (SNR), varies as the mote moves away from the source transducer, shown qualitatively in
With, e.g., a fixed 15-mm focal depth, the mote can stay fully functioning with an uplink SNR of at least 19.1 dB up to 40 mm away from the transducer array. The SNR for the uplink data as rendered in the reconstructed image is higher than that produced by measuring the echo at a single transducer because the delay-and-sum algorithm utilizes the spatial correlation of echo data receive from multiple elements. The 40-mm depth is practically limited by the maximum voltage that can be applied to the piezoelectric elements on the L12-3V probe to generate ultrasound pressure waves. Still, this maximum pressure is much lower than the safety limit for human use as regulated by the “mechanical index” (MI) threshold of 1.9 kPa/√{square root over (Hz)}10, or 3.8 MPa at 4 MHz, beyond which tissue damage can occur. The corresponding ultrasound power is only 9.2 mW/cm2, much lower than the 720 mW/cm2 safety limit10,12.
Exemplary Ultrasonography with Phantom. To demonstrate the functioning of the motes within an imaging application, chicken breasts are hand sliced into about 1-cm-thick layers and soaked in castor oil to act as a tissue phantom, with two motes between the chicken-breast layers (see
For example, in this exemplary case, it can be difficult to visually identify the location of the motes directly from the B-mode movie, as this miniaturized implant is not easily distinguished from other structures in this complicated mechanical environment. However, as active devices, they can present temporal modulation far stronger than the baseline noise level, with a greater than 23.4-dB SNR from the mote placed 15.1 mm away from the transducer, and a greater than 22.1-dB SNR from the mote placed 26.5 mm deep, for all received uplink data. This exemplary property can facilitate an easy location of multiple motes, as well as robust communication with multiple motes simultaneously, as shown in
Exemplary In Vivo Demonstration. To demonstrate the functioning of the mote in vivo, the exemplary devices can be implanted into the lower hind limb of a mouse, while the same B-mode imaging system and imaging configuration used for the phantom experiment can be used to communicate with the mote (as shown in
In this description, numerous specific details have been set forth. It is to be understood, however, that implementations of the disclosed technology can be practiced without these specific details. In other instances, well-known methods, structures and techniques have not been shown in detail in order not to obscure an understanding of this description. References to “some examples,” “other examples,” “one example,” “an example,” “various examples,” “one embodiment,” “an embodiment,” “some embodiments,” “example embodiment,” “various embodiments,” “one implementation,” “an implementation,” “example implementation,” “various implementations,” “some implementations,” etc., indicate that the implementation(s) of the disclosed technology so described may include a particular feature, structure, or characteristic, but not every implementation necessarily includes the particular feature, structure, or characteristic. Further, repeated use of the phrases “in one example,” “in one exemplary embodiment,” or “in one implementation” does not necessarily refer to the same example, exemplary embodiment, or implementation, although it may.
As used herein, unless otherwise specified the use of the ordinal adjectives “first,” “second,” “third,” etc., to describe a common object, merely indicate that different instances of like objects are being referred to, and are not intended to imply that the objects so described must be in a given sequence, either temporally, spatially, in ranking, or in any other manner.
While certain implementations of the disclosed technology have been described in connection with what is presently considered to be the most practical and various implementations, it is to be understood that the disclosed technology is not to be limited to the disclosed implementations, but on the contrary, is intended to cover various modifications and equivalent arrangements included within the scope of the appended claims. Although specific terms are employed herein, they are used in a generic and descriptive sense only and not for purposes of limitation.
The foregoing merely illustrates the principles of the disclosure. Various modifications and alterations to the described embodiments will be apparent to those skilled in the art in view of the teachings herein. It will thus be appreciated that those skilled in the art will be able to devise numerous systems, arrangements, and procedures which, although not explicitly shown or described herein, embody the principles of the disclosure and can be thus within the spirit and scope of the disclosure. Various different exemplary embodiments can be used together with one another, as well as interchangeably therewith, as should be understood by those having ordinary skill in the art. In addition, certain terms used in the present disclosure, including the specification and drawings, can be used synonymously in certain instances, including, but not limited to, for example, data and information. It should be understood that, while these words, and/or other words that can be synonymous to one another, can be used synonymously herein, that there can be instances when such words can be intended to not be used synonymously. Further, to the extent that the prior art knowledge has not been explicitly incorporated by reference herein above, it is explicitly incorporated herein in its entirety. All publications referenced are incorporated herein by reference in their entireties.
Throughout the disclosure, the following terms take at least the meanings explicitly associated herein, unless the context clearly dictates otherwise. The term “or” is intended to mean an inclusive “or.” Further, the terms “a,” “an,” and “the” are intended to mean one or more unless specified otherwise or clear from the context to be directed to a singular form.
This written description uses examples to disclose certain implementations of the disclosed technology, including the best mode, and also to enable any person skilled in the art to practice certain implementations of the disclosed technology, including making and using any devices or systems and performing any incorporated methods. The patentable scope of certain implementations of the disclosed technology is defined in the claims, and may include other examples that occur to those skilled in the art. Such other examples are intended to be within the scope of the claims if they have structural elements that do not differ from the literal language of the claims, or if they include equivalent structural elements with insubstantial differences from the literal language of the claims.
The following reference is hereby incorporated by references, in their entireties:
This application relates to and claims the benefit of priority from U.S. Provisional Patent Application No. 63/354,076, filed Jun. 21, 2022, the entire disclosure of which is incorporated herein by reference in its entirety.
This invention was made with government support under HR0011-15-2-0054 and D20AC00004 awarded by the Defense Advanced Research Projects Agency. The government has certain rights in the invention.
Number | Date | Country | |
---|---|---|---|
63354076 | Jun 2022 | US |
Number | Date | Country | |
---|---|---|---|
Parent | PCT/US2023/025861 | Jun 2023 | WO |
Child | 18990861 | US |