N/A
Increasing fat content for some subjects may lead to liver steatosis, which may progress to fibrosis, cirrhosis, liver failure, or even hepatocellular carcinoma. One of the methods to evaluate fat content in the liver is proton density fat fraction (PDFF) acquired with magnetic resonance imaging (MRI) which was often used as a benchmarking standard. However, one of the limitations of MRI is low accessibility, thus prohibiting frequent follow-up exams. Ultrasound attenuation coefficient has also been shown to have the potential to quantify fat content in the human liver. Ultrasound attenuation coefficient estimation (ACE) has been reported using different approaches, such as spectral shift technique, reference phantom-based methods, and the reference frequency method (RFM). However, in ultrasound imaging the ultrasound pulses undergo multi-path reflections (for example, within the fat and muscle layers of body wall) that will reveal reverberation clutter artifacts at different locations. The reverberation clutters superimposed on liver echoes may bias the value of the attenuation coefficient, thus effective reverberation clutter suppression is needed for ACE.
Several methods have been proposed to suppress the reverberation clutter signals by transforming the received ultrasound signals to predefined bases that are orthonormal and independent such as Discrete Fourier Transform and Discrete Wavelet Transform, and then the pre-defined bases are filtered/discarded to mitigate the reverberation clutter signals. Although these approaches may be computation-effective, they may produce poor results when reverberation clutter and tissue characteristics are overlapped. Furthermore, physiological differences among patients imply large variability of signal characteristics, results in difficulty to choose appropriate thresholds/bases to suppress reverberation clutter signals. Also, conventional methods for selecting bases depends on the actual data.
Therefore, there remains a need for effective, adaptive methods to overcome these limitations in reverberation clutter suppression.
The present disclosure addresses the aforementioned drawbacks by providing systems and methods to suppress reverberation clutter signals adaptively. In some configurations, a robust principal component analysis (RPCA) may be used to separate a static (e.g. low dimension) background from sparse moving (e.g. high dimension) objects with the presence of outliers. Principal component analysis (PCA) may include computing a set of linearly uncorrelated variables which is called principal components based on the covariance characteristics of the data. While PCA may be used to suppress reverberation clutter signals adaptively, PCA may be sensitive to data outliers, and thus degrades the reverberation clutter suppression capability. RPCA, by contrast, may effectively suppress reverberation clutter signals in the presence of outliers.
In some configurations, RPCA may assume the sparsity nature of the high dimensional moving objects signals, and a transformation may be used to ensure the tissue signal satisfies the sparsity property. To handle the sparsity requirement of RPCA, the tissue signal may be transformed to the wavelet domain to fulfill the sparsity condition. In accordance with the present disclosure, the use of the RPCA combined with wavelet kernels, may be used to suppress reverberation clutter signals to achieve robust ACE.
In one configuration, a method is provided for reverberation signal suppression in ultrasound imaging of a subject. The method includes accessing or acquiring ultrasound imaging data of a subject that includes a plurality of frames at different times and including tissue signals and reverberation signals. The method also includes generating a region of interest (ROI) frames subset by determining a ROI for each frame in the plurality of frames and generating a spatiotemporal matrix from the ROI frames subset. The method also includes separating tissue signals from reverberation signals in the spatiotemporal matrix using an adaptive method. The method also includes generating an image of the subject with the reverberation signals suppressed by subtracting the separated reverberation signals from the ultrasound imaging data.
In one configuration, a system is provided for reverberation signal suppression in ultrasound imaging of a subject. The system includes a computer system configured to: i) access ultrasound imaging data of a subject that includes a plurality of frames at different times and including tissue signals and reverberation signals; ii) generate a region of interest (ROI) frames subset by determining a ROI for each frame in the plurality of frames; iii) generate a spatiotemporal matrix from the ROI frames subset; iv) separate tissue signals from reverberation signals in the spatiotemporal matrix using an adaptive method; and v) generate an image of the subject with the reverberation signals suppressed by subtracting the separated reverberation signals from the ultrasound imaging data.
The foregoing and other aspects and advantages of the present disclosure will appear from the following description. In the description, reference is made to the accompanying drawings that form a part hereof, and in which there is shown by way of illustration a preferred embodiment. This embodiment does not necessarily represent the full scope of the invention, however, and reference is therefore made to the claims and herein for interpreting the scope of the invention. Like reference numerals will be used to refer to like parts from Figure to Figure in the following description.
Systems and methods are provided to adaptively suppress reverberation clutter signals in ultrasound imaging. In some configurations, a robust principal component analysis (RPCA) may be used to separate a static or low-dimension background signal from sparse, moving or high-dimension objects in the presence of outliers. To handle sparsity requirements of RPCA, a tissue signal may be transformed to the wavelet domain, or another suitable sparse domain, to fulfill the sparsity condition. The use of the RPCA combined with wavelet kernels may be used to suppress reverberation clutter signals to achieve robust ultrasound attenuation coefficient estimation.
Referring to
To differentiate the tissue signals from the reverberation clutter signals, the tissue signals may have large motion variability as compared with the reverberation signals across frames. In a non-limiting example of liver imaging, to obtain moving tissue signals of liver multiple ultrasound frames may be acquired with subjects breathing freely or breathing heavily so that the liver moves significantly during real-time in-vivo scanning. To achieve nearly static reverberation clutter signals which may originate from the body wall, the ultrasound probe may be held tightly upon the subject's body surface so that the clutter signals do not change significantly during breathing. In another non-limiting example, the beating heart may present a moving tissue signal to facilitate separation from static clutter signals from the body wall. Since the moving tissue signals possess high dimensional signals and the static reverberation clutter signals possess low dimensional signals, the reverberation clutter signals can be separated from the tissue signals using adaptive methods.
In some configurations, RPCA may be used to estimate the reverberation clutter signals from the received signals. Once the reverberation clutter signals are estimated, the tissue signals can be estimated by subtracting the received signals from the estimated reverberation clutter signals. Any appropriate separation method may be used. In non-limiting examples, a model-based method (e.g. principal component analysis, singular value decomposition), non-parametric-based method, or blind source method (e.g. independent component analysis), and the like may be used for the separation method.
Referring to
In a non-limiting example, the ROI data-points may be transformed from 3-dimensional Cartesian coordinates to 2-dimensional Casorati co-ordinates, where each row and column represents the spatial and temporal data-points, respectively. The matrix can be quantitatively summarized by statistics (e.g., mean, median) to measure performance. Such performance metrics can be provided on a range of 0-1, 0%-100%, or another suitable range.
The high-dimensional tissue signals and low-dimensional reverberation signals may be identified and separated by processing the spatiotemporal matrix at step 208 using methods in accordance with the present disclosure. Reverberation clutter signals may then be suppressed in the region of interest at step 210. An image of the subject with suppressed reverberation clutter signals may also be generated at step 212.
A signal model for the observed received spatiotemporal IQ data can be expressed as:
Where X, S, L and N are received signals, tissue signals, reverberation clutter artifact, and noise. In (1), the reverberation (L) possesses static low-dimensional (low-rank) signals, while the tissue(S) possesses high-dimensional (high-rank) signals.
RPCA may be used to recover low-rank components and to reduce the impact of corrupted data. An RPCA technique may be expressed as follows:
Where ∥L∥*, ∥S∥1, and ∥P∥F2 represent the nuclear norm, L1-norm and Frobenius norm of L, S and P. λ1 and λ2 are the regularization parameters which affect the estimated L and S. The method may be used in separating sparse dynamic data from static data, and may exploit the sparse property of signal S. In some configurations, the tissue signals S may not be sparse in the spatiotemporal domain, thus, making it difficult to meet the sparsity conditions. In a non-limiting example, the sparsity of the tissue signal in the wavelet-domain may be used to address sparsity condition issues instead of the spatiotemporal domain. Such an optimization problem can be expressed as
Where WH is the adjoint 2D wavelet transformation. Any appropriate wavelet kernel may be used, and any wavelet kernel can be used for the 2D wavelet transformation and adjoint 2D wavelet transformation.
In some configurations, an Alternating Direction Method of Multiplier (ADMM) may be used as a possible optimization method to solve eq. (3). Any appropriate optimization algorithm may be used, such as such as augmented Lagrange multiplier, fast alternating minimization, iteratively reweighted least squares, and the like can be used to solve eq. (3). When using ADMM, two auxiliary variables U, and V may be introduced and eq. (3) may be rewritten as
Removing the linear equality constraints in (4) with the Lagrangian method generates the following objective cost function:
Where μ1 and μ2 are the ADMM optimization parameters to control the convergence of the algorithm. At each iteration k, ADMM may perform the following four steps to minimize the cost function
Eqs. (6) and (7) are convex problems possessing closed-form solutions: singular value thresholding (SVT) and soft thresholding (ST), respectively. Additionally, eqs. (8) and (9) can be solved by differentiating with respect to Lk and Sk, summarized as follows:
Where E1 and E2 can be computed using a gradient decent method. A non-limiting example method is summarized as the Table I below:
In some configurations, the RPCA algorithm may involve repeated computations of the singular value decomposition (SVD) and thresholding of matrices during the singular value thresholding process. This repeated computation of the SVD may be a bottleneck of computational complexity, but as one non-limiting example singular value thresholding may be used to alleviate this complexity by computing fewer singular values as those singular values that lie above a specified threshold. Speeding up the algorithms that involve thresholding of singular values may be accomplished by using a truncated SVD approach to compute only those singular values of interest. The truncated SVD approach is one non-limiting example of the possible methods for the fast computation of SVD, and any appropriate computation of SVD processes may be used.
Referring to
Referring to
Referring to
RPCA may be used to mitigate the reverberation clutter artifacts in ultrasound attenuation coefficient estimation. The methods in accordance with the present disclosure provide accurate and robust ultrasound attenuation coefficient estimation. In the non-limiting examples above, tissue moves while clutters are static. The methods in accordance with the present disclosure may also be used in situations where the unwanted clutter signals move while the desired tissue signal are static: in such cases, tissue signal will be low dimensional (low rank) and clutter signal will be high dimensional (high rank). Therefore, clutters can still be isolated and subtracted from the received signal to obtain a cleaner tissue signal to achieve clutter suppression.
When energized by a transmitter 606, a given transducer element 604 produces a burst of ultrasonic energy. The ultrasonic energy reflected back to the transducer array 602 (e.g., an echo) from the object or subject under study is converted to an electrical signal (e.g., an echo signal) by each transducer element 604 and can be applied separately to a receiver 608 through a set of switches 610. The transmitter 606, receiver 608, and switches 610 are operated under the control of a controller 612, which may include one or more processors. As one example, the controller 612 can include a computer system.
The transmitter 606 can be programmed to transmit unfocused or focused ultrasound waves. In some configurations, the transmitter 606 can also be programmed to transmit diverged waves, spherical waves, cylindrical waves, plane waves, or combinations thereof. Furthermore, the transmitter 606 can be programmed to transmit spatially or temporally encoded pulses.
The receiver 608 can be programmed to implement a suitable detection sequence for the imaging task at hand. In some embodiments, the detection sequence can include one or more of line-by-line scanning, compounding plane wave imaging, synthetic aperture imaging, and compounding diverging beam imaging.
In some configurations, the transmitter 606 and the receiver 608 can be programmed to implement a high frame rate. For instance, a frame rate associated with an acquisition pulse repetition frequency (“PRF”) of at least 100 Hz can be implemented. In some configurations, the ultrasound system 600 can sample and store at least one hundred ensembles of echo signals in the temporal direction.
The controller 612 can be programmed to implement an imaging sequence using the techniques described in the present disclosure, or as otherwise known in the art. In some embodiments, the controller 612 receives user inputs defining various factors used in the implementation of the imaging sequence.
A scan can be performed by setting the switches 610 to their transmit position, thereby directing the transmitter 606 to be turned on momentarily to energize transducer elements 604 during a single transmission event. The switches 610 can then be set to their receive position and the subsequent echo signals produced by the transducer elements 604 in response to one or more detected echoes are measured and applied to the receiver 608. The separate echo signals from the transducer elements 604 can be combined in the receiver 608 to produce a single echo signal.
The echo signals are communicated to a processing unit 614, which may be implemented by a hardware processor and memory, to process echo signals or images generated from echo signals. As an example, the processing unit 614 can suppress reverberation signal clutter noise/artifacts using the methods described in the present disclosure. Images produced from the echo signals by the processing unit 614 can be displayed on a display system 616.
The present disclosure has described one or more preferred embodiments, and it should be appreciated that many equivalents, alternatives, variations, and modifications, aside from those expressly stated, are possible and within the scope of the invention.
This application claims the benefit of U.S. Provisional Patent Application Ser. No. 63/214,002 filed on Jun. 23, 2021 and entitled “Systems and Methods for Reverberation Clutter Artifacts Suppression in Ultrasound Imaging,” which is incorporated herein by reference as if set forth in its entirety for all purposes.
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/US2022/034182 | 6/20/2022 | WO |
Number | Date | Country | |
---|---|---|---|
63214002 | Jun 2021 | US | |
62214002 | Sep 2015 | US |