The disclosure is related to diagnosing and monitoring tissue using optical biopsy, and treating tissue in vivo, without extracting the tissue for biopsy.
Up to eight-five percent of all human cancers start in the epithelial tissue. As shown in Table 1 below, some of these cancers, such as melanoma of the skin for example, are easier to detect and to treat, resulting in better five-year survival rates, although there is still need for improved detection and treatments. Others, particularly in the esophagus, colon, and lung are difficult to find at an early stage, have low survival rates if found early, and have extremely low survival rates if found at later stages. Furthermore, some patient populations have a higher risk of cancer occurrence based on other factors.
In general, the course of care for most cancers involves a procedure to acquire data (typically tissue). The acquired tissue is typically sent off to a laboratory outside of the context of the tissue acquisition procedure. Depending on the circumstances, this analysis may take several hours, days, or weeks. After the analysis is returned, the physician may make a diagnosis, and if treatment is necessary, a treatment procedure may be employed. Because of the time required for analysis of the acquired tissue, the treatment procedure is performed during a separate patient procedure or examination, and typically during a patient visit days to weeks later. Treatment may then be repeated at various time points subsequent during separate patient procedures to verify that the cancer has been eliminated and has not returned. As one example, a dermatologist may visually inspect the skin. If a suspicious mole is found, a piece of tissue may be cut out and sent to a pathology lab for analysis. Based on the pathology information, the patient may undergo a Moh's surgery on the mole where successive layers of tissue are sliced off and sent for immediate pathology analysis until a layer with no cancer cells is obtained. The patient will probably undergo follow-up visits to visually inspect that spot and verify that the cancer has not returned. Similar procedures will be followed for other cancers, but with the disadvantage that is it difficult to accurately track the tissue location when inside the body in places such as the colon, esophagus, bladder, cervix, oral cavity, and others.
As another example, patients with Gastroesophageal Reflux Disease (GERD) may progress to Barrett's Esophagus (BE), at which point they have a 30 to 150 times greater chance of getting esophageal cancer than the general population. As a result, the current standard of care is for these patients to undergo a random biopsy surveillance procedure on a periodic basis. The biopsy procedure consists of a four-quadrant biopsy taken every centimeter through the affected portion of the esophagus (the Seattle Protocol). These biopsies are sent to a pathology lab and, based on the results, the patient comes back for the next round of surveillance or further treatment occurs such as an oral drug, or in cases of high grade dysplasia or cancer, an esophagectomy.
There are significant issues with this current approach to detection and treatment of numerous cancer types including lack of coverage of tissue, lack of sufficient detection at early stages of the disease, time lag between sample acquisition and treatment procedures due to the inability to acquire and diagnose tissue quickly during the same procedure or patient examination, and need for multiple procedures. Because the diagnosis occurs later in time after the tissue acquisition, it is also difficult to return to the exact location of the biopsy for further monitoring and treatment. Misdiagnosis by the pathologist, and lack of effective treatment options can occur as a result.
Advances by the applicant in low coherence interferometry (LCI), including angle-resolved LCI (a/LCI) and Fourier domain LCI (f/LCI) (referred to collectively as “f/a/LCI”) enable in vivo diagnosis of epithelial tissue health, specifically if tissue is normal, pre-cancerous, cancerous, diseased, or abnormal. This opens up new opportunities, the most significant described of which in the invention to follow is the potential to diagnosis, treat, and monitor tissue in vivo, employing methods, processes, techniques, and systems that use real-time optical biopsy systems, including f/a/LCI systems, for examining and monitoring tissue during the course of the same or concomitant medical procedure to determine if a therapeutic should be applied to the tissue.
Embodiments in the detailed description cover methods, processes, techniques, and systems that use real-time optical biopsy systems for examining and monitoring tissue during the course of the same or concomitant medical procedure to determine if a therapeutic should be applied to the tissue. The real-time optical biopsy systems disclosed herein are systems based on low coherence interferometer (LCI) detection of light scattered from a sample that can obtain structural and/or depth-resolved information regarding in vivo tissue in a single data collection event and which permits diagnosis in connection with the data collection. New therapeutic procedures and techniques can be implemented as a result. Specifically, tissue can be diagnosed and treated during the same or concomitant medical procedure or examination. This is an improvement over traditional biopsy techniques where diagnosis of the tissue cannot be performed until the biopsy procedure is completed and the biopsy results are received after the procedure thereby delaying treatment. Further, the location of the analyzed tissue is known thereby allowing localized treatment of the tissue, or the location may be returned to for follow up monitoring.
These methods, processes, techniques, and systems disclosed herein offer an opportunity to significantly improve the standard of care for patients and decrease overall health care costs by diagnosing and treating tissue conditions, including pre-cancerous and cancerous conditions, in vivo. The methods, processes, and techniques disclosed herein effectively reduce the treatment time to the time of a first medical procedure on the patient, thus providing earlier treatment and potentially better and more timely results at a lower cost. This also provides more accurate diagnosis and determination of treatment effectiveness since the monitoring is performed on a localized level with the ability to diagnose, treatment, and monitor the affected tissue during the same or concomitant medical procedure or examination. The above-described methods, processes, techniques, and systems also enable more efficient diagnosis, treatment, and monitoring, or throughput of patients. This may be particularly important where health facilities and appointments are a limited resource.
In disclosed embodiments, real-time optical biopsy systems include Fourier domain and/or angle-resolved low coherence interferometry (LCI) optical biopsy technologies (hereinafter referred to collectively and generically as “f/a/LCI”). During the same or concomitant medical procedure or examination, a physician or other health care professional will be able to scan tissue in vivo on a localized level using a real-time f/a/LCI system, monitor the scan, diagnose tissue status as normal, pre-cancerous, cancerous, abnormal, diseased or the like, and administer a therapeutic based on the tissue status, if desired or needed. Because the scan of the tissue can be performed in real-time using the real-time f/a/LCI system, which collects depth-resolved and/or structural information in a single data collection event, monitoring of the treated tissue can also be performed in real-time and during the same or concomitant medical procedure or tissue examination. In the same regard, diagnosis of the tissue can also be performed during the same or concomitant medical procedure or tissue examination. A therapeutic can also be administered during the same or concomitant procedure or tissue examination. If desired, multiple medical procedures at different time points can then be used to monitor the status of tissue in vivo over time to determine tissue status, health or response to treatment. This allows physicians or other clinicians to fully maximize the information opportunity provided by the real-time f/a/LCI system and vastly improve the quality of care for the patient.
In one embodiment, a method for examining and monitoring tissue to determine if a therapeutic should be applied to the tissue during a same or concomitant medical procedure is provided. The method includes optically examining using a real time f/a/LCI system a tissue to detect tissues that are cancerous, abnormal, diseased or the like which conditions are generally not perceptible to the human eye. Real-time feedback information is monitored regarding the examination of the tissue from the real-time f/a/LCI system. Based on the real-time feedback information, a diagnosis is made as to whether a treatment should be applied to the tissue. If a treatment is to be applied, a selected therapy or combination of therapies is applied during the same or concomitant medical procedure.
The new methods, processes, techniques, and systems address the shortcoming of the current approaches. For example, since real-time optical biopsy systems can acquire data points in short periods of time (e.g., in a few seconds or minutes), it is possible to scan much larger areas of the tissue during a same or concomitant medical procedure. Furthermore, real-time f/a//LCI systems can detect tissue changes at an earlier stage in the disease. A therapeutic can be delivered immediately to a localized area where the real-time f/a/LCI system detected pre-cancerous, cancerous, abnormal, diseased tissue, or to a general area during the same or concomitant medical procedure. Subsequent scans can be taken to verify the treatment outcome and monitor tissue health over time. Information from the real-time optical biopsy systems described herein can be used to determine dosing levels or which choice of multiple treatment options to use. A standardized database in the computer can be employed to allow consistent analysis of tissue based on a database of tissue characteristics versus tissue health by detecting anomalies in tissue which may be pre-cancerous, cancerous, abnormal, diseased or the like.
Some implementations include the integration of a real-time optical biopsy system with an endoscope and/or therapeutic system. This integration results in a system with the capability to both diagnose and treat tissue in vivo. Several architectures are described including the use of an endoscopic probe, where a real-time optical biopsy system probe and the endoscopic light probe share or occupy one or more channels. Several architectures are also described including the use of multi-channel endoscopes where the real-time optical biopsy system probe occupies one channel and a therapeutic applicator can occupy another channel. The therapeutic system may be manually controlled or computer-controlled. There are a wide range of possible therapeutics including, but not limited to, elements, compounds, drugs, liquids, heat, cold, radio-frequency (RF) ablation, photodynamic therapy, and radiation. Another architecture example uses a single channel endoscope where the real-time optical biopsy system probe and the therapeutic system occupy the same fiber or fiber bundle channel. Yet another implementation uses a scanning real-time optical biopsy system where multiple points are scanned in an automated or semi-automated fashion.
In addition to clinical activities, a real time optical biopsy such as f/a/LCI can be used in research activities, particularly those that track tissue health over time, such as in the study of chemo-preventatives. Real time f/a/LCI could be used to scan a tissue sample or cell culture at various points in time to assess changes in the status of the tissue or cells. For example a cell culture of cancer cells could be scanned and then treated with a chemo-preventative and then scanned at subsequent time points to see if the cancer cells were killed (such as by apoptosis) or not.
The embodiments set forth below represent the necessary information to enable those skilled in the art to practice the invention and illustrate the best mode of practicing the invention. Upon reading the following description in light of the accompanying drawing figures, those skilled in the art will understand the concepts of the invention and will recognize applications of these concepts not particularly addressed herein. It should be understood that these concepts and applications fall within the scope of the disclosure and the accompanying claims.
Embodiments in the detailed description cover methods, processes, techniques, and systems that use real-time optical biopsy systems for examining and monitoring tissue during the course of the same or concomitant medical procedure to determine if a therapeutic should be applied to the tissue. The real-time optical biopsy systems disclosed herein are systems based on low coherence interferometer (LCI) detection of light scattered from a sample that can obtain structural and/or depth-resolved information regarding in vivo tissue in a single data collection event and which permits diagnosis in connection with the data collection. New therapeutic procedures and techniques can be implemented as a result. Specifically, tissue can be diagnosed and treated during the same or concomitant medical procedure or examination. This is an improvement over traditional biopsy techniques where diagnosis of the tissue cannot be performed until the biopsy procedure is completed and the biopsy results are received after the procedure thereby delaying treatment. Further, the location of the analyzed tissue is known thereby allowing localized treatment of the tissue, or the location may be returned to for follow up monitoring.
These methods, processes, techniques, and systems disclosed herein offer an opportunity to significantly improve the standard of care for patients and decrease overall health care costs by diagnosing and treating tissue conditions, including pre-cancerous and cancerous conditions, in vivo. The methods, processes, and techniques disclosed herein effectively reduce the treatment time to the time of a first medical procedure on the patient, thus providing earlier treatment and potentially better and more timely results at a lower cost. This also provides more accurate diagnosis and determination of treatment effectiveness since the monitoring is performed on a localized level with the ability to diagnose, treatment, and monitor the affected tissue during the same or concomitant medical procedure or examination. The above-described methods, processes, techniques, and systems also enable more efficient diagnosis, treatment, and monitoring, or throughput of patients. This may be particularly important where health facilities and appointments are a limited resource.
In disclosed embodiments, real-time optical biopsy systems include Fourier domain and/or angle-resolved low coherence interferometry (LCI) optical biopsy technologies (hereinafter referred to collectively and generically as “f/a/LCI”). During the same or concomitant medical procedure or examination, a physician or other health care professional will be able to scan tissue in vivo on a localized level using a real-time f/a/LCI system, monitor the scan, diagnose tissue status as normal, pre-cancerous, cancerous, abnormal, diseased or the like, and administer a therapeutic based on the tissue status, if desired or needed. Because the scan of the tissue can be performed in real-time using the real-time f/a/LCI system, which collects depth-resolved and/or structural information in a single data collection event, monitoring of the treated tissue can also be performed in real-time and during the same or concomitant medical procedure or tissue examination. In the same regard, diagnosis of the tissue can also be performed during the same or concomitant medical procedure or tissue examination. A therapeutic can also be administered during the same or concomitant procedure or tissue examination. If desired, multiple medical procedures at different time points can then be used to monitor the status of tissue in vivo over time to determine tissue status, health or response to treatment. This allows physicians or other clinicians to fully maximize the information opportunity provided by the real-time f/a/LCI system and vastly improve the quality of care for the patient.
The new methods, processes, techniques, and systems address the shortcoming of the current approaches. For example, since real-time optical biopsy systems can acquire data points in short periods of time (e.g., in a few seconds or minutes), it is possible to scan much larger areas of the tissue during a same or concomitant medical procedure. Furthermore, real-time f/a//LCI systems can detect tissue changes at an earlier stage in the disease. A therapeutic can be delivered immediately to a localized area where the real-time f/a/LCI system detected pre-cancerous, cancerous, abnormal, diseased tissue, or to a general area during the same or concomitant medical procedure. Subsequent scans can be taken to verify the treatment outcome and monitor tissue health over time. Information from the real-time optical biopsy systems described herein can be used to determine dosing levels or which choice of multiple treatment options to use. A standardized database in the computer can be employed to allow consistent analysis of tissue based on a database of tissue characteristics versus tissue health by detecting anomalies in tissue which may be pre-cancerous, cancerous, abnormal, diseased or the like.
Because of timely acquisition of tissue information, real-time feedback information regarding the tissue is provided by the real-time f/a/LCI system and can be monitored by a physician or clinician in real-time and during the same or concomitant medical procedure or examination, thereby minimizing time, inconvenience, and/or discomfort to the patient (block 14). Further, a timely diagnosis of the results can be performed. A diagnosis of the tissue information from the real-time f/a/LCI system can be performed to determine if treatment of the examined tissue is necessary or desired. If necessary or desired, the treatment can be undertaken during the same or concomitant medical procedure or examination, and without having to wait for biopsy results or only after lengthy scans are performed (block 16). If treatment is required, a general, local, or combination of general and local treatment can be performed on the tissue in the same localized area of examined by the real-time f/a/LCI system with accuracy and during the same or concomitant medical procedure or examination of the patient (block 18).
Thereafter, it can be determined if further monitoring of the affected tissue is desired or needed (block 20). This further monitoring can be performed during the same or concomitant medical procedure or examination of the patient or during a subsequent medical procedure or examination of the patient. If further monitoring is needed, the overall process can be performed again (block 10) wherein an optical biopsy of the treated tissue can be performed (block 12). If further monitoring is not required, or it is not required or possible to see results during the same or concomitant medical procedure or examination of the patient, the process ends (block 22). Likewise, if no treatment is desired or needed (block 16), and further monitoring is not required or desired (block 24), the process ends (block 22). If further monitoring is required even though treatment is not required or desired after an optical biopsy (block 24), the process can be repeated (block 10) and another optical biopsy performed (block 12).
In this regard, the above-described methods and processes can reduce the number of medical procedures required to achieve a therapeutic result. If a traditional biopsy is performed, a diagnosis of the tissue cannot be performed until the biopsy results are received. Therapy, if needed or desired, can only be performed during a subsequent medical procedure or examination of the patient. The above-described methods and processes also allow monitoring of the effectiveness of the therapy during the same or concomitant medical procedure if desired, because the information regarding the tissue can be obtained and analyzed during the same or concomitant medical procedure and after therapy has been administered. This effectively reduces the application of treatment to the time of a first medical procedure on the patient, thus providing earlier treatment and potentially better and more timely results at a lower cost. This also provides more accurate diagnosis and determination of treatment effectiveness since the monitoring is performed on a localized level with the ability to diagnose, monitor, and treat the affected tissue during the same or concomitant medical procedure or examination. The above-described methods and processes also enable more efficient diagnosis, treatment, and monitoring, or throughput of patients. This may be particularly important where health facilities and appointments are a limited resource.
As an example, a tissue examination procedure may be an esophageal endoscopy performed on patients with risk of esophageal cancer (such as those with Barrett's Esophagus). In the prior method, a physical biopsy of the esophagus is taken and sent to a pathological laboratory for analysis. It may take one week or so for a laboratory technician to analyze the extracted tissue sample and provide the information regarding the results to the attending physician. If, for example, it is determined that dysplasia is present, the patient is then scheduled for another medical procedure or examination in the future. An esophageal endoscopy is then performed again where a radio frequency (RF) ablation or other treatment may be performed. The monitoring of the treatment cannot be determined during the second medical procedure either. A biopsy must be performed in yet a subsequent medical procedure or examination, and the process repeated, thus adding substantial delay between the patient's first procedure of a biopsy and analysis of the effectiveness of the treatment.
With the methods, processes, techniques, and systems disclosed herein, the physician uses real-time f/a/LCI to scan tissue. Because the information regarding the scan is provided on a localized level and in real-time, the physician can treat any precancerous, cancerous, diseased, or abnormal areas concomitantly with the scanning. Alternately, the physician might first scan the tissue and then go back and ablate any areas of concern during the same or concomitant medical procedure. With the embodiments disclosed herein, there is the possibility of scanning, diagnosis and treatment in the same or concomitant medical procedure. Follow up might then consist of repeating this procedure at certain time intervals with additional treatment as necessary.
The remainder of this section focuses on system designs that allow these new methods, processes and techniques to be carried out in the process of examining and treating patients. Additional embodiments of the methods, processes and techniques disclosed include medical procedures using real-time f/a/LCI, examples of which are described in more detail below. Various systems may be implemented and used to carry out the methods, processes and techniques. Examples of these new systems and methods, processes, and techniques are described below in more detail in this application. These systems are not an exhaustive list, but illustrate examples enabled by the present invention to diagnose, monitor, and treat cancer using f/LCI, a/LCI, or f/a/LCI.
In one embodiment, the system that can be employed to carry out the medical procedure or examination can consist of: (1) a real-time f/a/LCI optical biopsy tissue diagnosis system, (2) an endoscope, and (3) a therapeutic that can be delivered via the endoscope. This integrated system will then allow the operator to assess the tissue health and apply the appropriate therapeutic to tissue meeting certain criteria. A typical biopsy endoscope 26 is illustrated in
Before discussing various embodiments of real-time f/a/LCI systems and endoscope systems that may be used to examine, diagnose, and administer treatment to a patient's tissue, more information regarding real-time f/a/LCI systems is provided.
Since this angle-resolved, cross-correlated signal is spectrally dispersed, the new data acquisition scheme is significant as it permits data to be obtained in seconds or minutes, a threshold determined to be necessary for acquiring data from in vivo tissues. Information about all depths of the sample at each of the multitude of different scattering angles on the sample can be obtained with one scan on the order of approximately 40 milliseconds. From the spatial, cross-correlated reference signal, structural (size) information can also be obtained using techniques that allow size information of scatterers to be obtained from angle-resolved data.
The faLCI technique in
The key advances of the present invention can be broken down into three components: (1) new rapid data acquisition methods, (2) fiber probe designs, and (3) data analysis schemes. Thus, the present invention is described in this matter for convenience in its understanding.
An exemplary apparatus, as well as the steps involved in the process of obtaining angle and depth-resolved distribution data scattered from a sample, are also set forth in
Lenses L3 (71) and L4 (78) are arranged to produce a collimated pencil beam 70 incident on the sample 48 (step 106 in
The light 80 scattered by the sample 58 is collected by lens L4 (78) and relayed by a 4f imaging system comprised of lenses L5 (83) and L6 (84) such that the Fourier plane of lens L4 (78) is reproduced in phase and amplitude at the spectrograph slit 88 (block 108 in
The mixed scattered light signal 86 is a function of vertical position on the spectrograph slit 88, y, and wavelength λ once the light is dispersed by the spectrograph 69. The detected signal at pixel (m, n) can be related to the scattered light 80 and reference input beam 56 (Es, Er) as:
I(λm,yn)=Er(λm,yn)|2+Es(λm,yn)|2+2ReEs(λm,yn)E*r(λm,yn) cos φ (1)
where φ is the phase difference between the two beams 70, 56 and · denotes an ensemble average in time. The interference term is extracted by measuring the intensity of the signal 70 and reference beams 56 independently and subtracting them from the total intensity.
In order to obtain depth-resolved information, the wavelength spectrum at each scattering angle is interpolated into a wavenumber (k=2π/λ) spectrum and Fourier transformed to give a spatial cross correlation, ΓSR(z) for each vertical pixel yn:
ΕSR(z,yn)=∫dkeikzEs(k,yn)E*r(k,yn) cos φ, (2)
The reference beam 54 takes the form:
E
r(k)=Eoexp[−((k−ko)/Δk)2]exp[−((y−yo)/Δy)2]exp[ikΔl] (3)
where ko (yo and Δk (Δy) represent the center and width of the Gaussian wave vector (spatial) distribution and Δl is the selected path length difference. The scattered light 80 takes the form
E
s(k,θ)=ΣjEoexp[−((k−ko)/Δk)2]exp[iklj]Sj(k,θ) (4)
where Sj represents the amplitude distribution of the scattering originating from the jth interface, located at depth lj. The angular distribution of the scattered light 80 is converted into a position distribution in the Fourier image plane of lens L4 through the relationship y=f4θ. For the pixel size of the CCD 90 (e.g., 20 μm), this yields an angular resolution (e.g., 0.57 mrad) and an expected angular range (e.g., 228 mrad).
Inserting Equations (3) and (4) into Equation (2) and noting the uniformity of the reference field 54 (Δy>>slit height) yields the spatial cross correlation at the nth vertical position on the imaging spectrograph 69:
Evaluating this equation for a single interface yields:
∇SR(z,yn)=|Eo|2exp[−((z−Δl+lj)Δk)2/8]Sj(ko,θn=yn/f4)cos φ. (6)
Here, it is assumed that the scattering amplitude S does not vary appreciably over the bandwidth of the source light 52. This expression shows that we obtain a depth resolved profile of the scattering distribution 80 is obtained with each vertical pixel corresponding to a scattering angle.
In the experiments that produced the depth-resolved profile of the sample 58 illustrated in
In addition to obtaining depth-resolved information about the sample 58, the scattering distribution data (i.e., a/LCI data) obtained from the sample 58 using the disclosed data acquisition scheme can also be used to make a size determination of the nucleus using the Mie theory. A scattering distribution 114 of the sample 58 is illustrated in
In order to fit the scattered data 114 to Mie theory, the a/LCI signals are processed to extract the oscillatory component which is characteristic of the nucleus size. The smoothed a/LCI data 114 is fit to a low-order polynomial (4th order was used for example herein, but later studies use a lower 2nd order), which is then subtracted from the distribution 116 to remove the background trend. The resulting oscillatory component is then compared to a database of theoretical predictions obtained using Mie theory 118 from which the slowly varying features were similarly removed for analysis.
A direct comparison between the filtered a/LCI data 116 and Mie theory data 118 may not possible, as the chi-squared fitting algorithm tends to match the background slope rather than the characteristic oscillations. The calculated theoretical predictions include a Gaussian distribution of sizes characterized by a mean diameter (d) and standard deviation (6D) as well as a distribution of wavelengths, to accurately model the broad bandwidth source.
The best fit (
As an alternative to processing the a/LCI data and comparing to Mie theory, there are several other approaches which could yield diagnostic information. These include analyzing the angular data using a Fourier transform to identify periodic oscillations characteristic of cell nuclei. The periodic oscillations can be correlated with nuclear size and thus will possess diagnostic value. Another approach to analyzing a/LCI data is to compare the data to a database of angular scattering distributions generated with finite element method (FEM) or T-Matrix calculations. Such calculations may offer superior analysis as they are not subject to the same limitations as Mie theory. For example, FEM or T-Matrix calculations can model non-spherical scatterers and scatterers with inclusions while Mie theory can only model homogenous spheres.
As an alternative embodiment, the present invention can also employ optical fibers to deliver and collect light from the sample of interest to use in the a/LCI system for endoscopic applications, such as that illustrated in
The fiber optic a/LCI scheme for this alternative embodiment makes use of the Fourier transform properties of a lens. This property states that when an object is placed in the front focal plane of a lens, the image at the conjugate image plane is the Fourier transform of that object. The Fourier transform of a spatial distribution (object or image) is given by the distribution of spatial frequencies, which is the representation of the image's information content in terms of cycles per mm. In an optical image of elastically scattered light, the wavelength retains its fixed, original value and the spatial frequency representation is simply a scaled version of the angular distribution of scattered light.
In the fiber optic a/LCI scheme, the angular distribution is captured by locating the distal end of the fiber bundle in a conjugate Fourier transform plane of the sample using a collecting lens. This angular distribution is then conveyed to the distal end of the fiber bundle where it is imaged using a 4f system onto the entrance slit of an imaging spectrograph. A beamsplitter is used to overlap the scattered field with a reference field prior to entering the slit so that low coherence interferometry can also be used to obtain depth-resolved measurements.
Turning now to
Light in the reference field 54′ emerges from fiber F1 and is collimated by lens L11(124) which is mounted on a translation stage 126 to allow gross alignment of the reference arm path length. This path length is not scanned during operation but may be varied during alignment. A collimated beam 128 is arranged to be equal in dimension to the end 131 of fiber bundle F3 (130) so that the collimated beam 128 illuminates all fibers in F3 (130) with equal intensity. The reference field 54′ emerging from the distal tip of F3 (130) is collimated with lens L3 (132) in order to overlap with the scattered field conveyed by fiber F4 (134). In an alternative embodiment, light 54′ emerging from fiber F1 is collimated then expanded using a lens system to produce a broad beam.
The scattered field is detected using a coherent fiber bundle. The scattered field is generated using light in the signal arm 122, which is directed toward the sample 58′ of interest using lens L2 (138). As with the free space system, lens L2 (138) is displaced laterally from the center of single-mode fiber F2 such that a collimated beam is produced which is traveling at an angle relative to the optical axis. The fact that the incident beam strikes the sample 58′ at an oblique angle is essential in separating the elastic scattering information from specular reflections. The light scattered by the sample 58′ is collected by a fiber bundle consisting of an array of coherent single mode or multi-mode fibers. The distal tip of the fiber is maintained one focal length away from lens L2 (138) to image the angular distribution of scattered light. In the embodiment shown in
As illustrated in
It is expected that the above-described a/LCI fiber-optic probe will collect the angular distribution over a 0.45 radian range (approx. 30 degrees) and will acquire the complete depth resolved scattering distribution 114 in a fraction of a second.
There are several possible schemes for creating the fiber probe which are the same from an optical engineering point of view. One possible implementation would be a linear array of single mode fibers in both the signal and reference arms. Alternatively, the reference arm 136 could be composed of an individual single mode fiber with the signal arm 122 consisting of either a coherent fiber bundle or linear fiber array.
The fiber probe tip can also have several implementations which are substantially equivalent. These would include the use of a drum or ball lens in place of lens L2 (138). A side-viewing probe could be created using a combination of a lens and a mirror or prism or through the use of a convex mirror to replace the lens-mirror combination. Finally, the entire probe can be made to rotate radially in order to provide a circumferential scan of the probed area.
Yet another data acquisition embodiment of the present invention could be a faLCI system is based on a modified Mach-Zehnder interferometer as illustrated in
Scattered light 162 from the sample is collected by lens L1 (155) and, via the Fourier transform property of the lens L1 (155, the angular distribution of the scattered field 162 is converted into a spatial distribution at the distal face of the multimode coherent fiber bundle 156 (e.g., Schott North America, Inc., length=840 mm, pixel size=8.2 μm, pixel count=13.5K) which is located at the Fourier image plane of lens L1 (155). The relationship between vertical position on the fiber bundle, y′, and scattering angle, θ is given by y′=f1θ. As an illustration, the optical path of light scattered 162 at three selected scattering angles is shown in
The angular distribution exiting a proximal end 164 of the fiber bundle 156 is relayed by the 4f imaging system of L2 (138) and L3 (132) (f2=3.0 cm, f3=20.0 cm) to the input slit 88″ of the imaging spectrograph 69″ (e.g., Acton Research, InSpectrum 150). The theoretical magnification of the 4f imaging system is (f3/f2) 6.67 in this example. Experimentally, the magnification was measured to be M=7.0 in this example with the discrepancy most likely due to the position of the proximal end 164 of the fiber bundle 156 with relation to lens L2 (166). The resulting relationship between vertical position on the spectrograph slit 88″, y, and θ is y=Mf1(θ−θmin). The optical path length of the reference arm is matched to that of the fundamental mode of the sample arm. Light 167 exiting the reference fiber 54″ is collimated by lens L4 (168) (e.g., f=3.5 cm, spot size=8.4 mm) to match the phase front curvature of the sample light and to produce even illumination across the slit 88″ of the imaging spectrograph 69″. A reference field 170 may be attenuated by a neutral density filter 172 and mixed with the angular scattering distribution at beamsplitter BS (174). Mixed fields 176 are dispersed with a high resolution grating (e.g., 1200 lines/mm) and detected using an integrated, cooled CCD (not shown) (e.g., 1024×252, 24 μm×24 μm pixels, 0.1 nm resolution) covering a spectral range of 99 nm centered at 840 nm, for example.
The mixed fields 176, a function of wavelength, λ, and θ, can be related to the signal and reference fields (Es, Er) as:
I(λm,θn)=Er(λm,θn)|2+Es(λm,θn)|2+2ReEs(λm,θn)E*r(λm,θn)cos(φ), (7)
where φ is the phase difference between the two fields, (m,n) denotes a pixel on the CCD, and . . . denotes a temporal average. I(λm,θn) is uploaded to a personal computer (PC) using LabVIEW software manufactured by National Instruments and processed in 320 ms to produce a depth and angle-resolved contour plot of scattered intensity. The processing of the angle-resolved scattered field to obtain depth and size information described above, and in particular reference to the data acquisition apparatus of
This disclosure expands the capability of one or more therapeutics to the system. The system may or may not be used to collect biopsy samples.
In this example, the endoscopic probe 180 interfaces with an endoscope control box 192, which is the source of anything passing into the endoscopic probe 180 and the receiver for visual information returning from the endoscopic probe 180. In many cases, the visual image of the tissue 48 is displayed on a screen allowing the operator to see inside the patient without using the eyepiece 188. In this regard, the endoscope control box 192 may be under the control of the computer 43 via a communications line(s) 193 to provide control and for receiving images of the patient's tissue if the endoscopic probe 180 employs a camera.
Note that the endoscopic probe 180, the real-time f/a/LCI system 40, and therapeutic functions are shown as independent connections and control boxes in
Another class of therapeutics is applied substances. The therapeutic substance could take the form of a liquid, gel, aerosol, or gas, as examples. This could include, but is not limited to, drugs, compounds, and/or elements that cause a chemical reaction at the tissue site and/or substances that affect the tissue in a physical manner such as hot or cold liquids or acids or bases. Collectively, these administered therapeutics will be referred to as “substances.”
Another therapeutic that can be used is photodynamic therapy. Here, the patient is given a drug called a photosensitizer and then exposed to a particular type (wavelength) of light, for example, the light from a Nd:YAG laser at a wavelength of 630 micrometers. Numerous photosensitizers are known in the art, including but not limited to porfimer sodium, chlorins, bacteriochlorins, purpurins, benzoporphyrins, texaphyrins, etiopurpurins, naphthalocyanines and phthalocyanines. The drug interacts with the light and produces a form of oxygen that kills nearby cells. The photosensitizer is typically injected into the blood, and between 24 and 72 hours later, the tumor is exposed to light. This time window is set by the fact that the photosensitizer remains in the cancer cells longer than in other cells in the body. The photodynamic therapy has several side affects including damage to tissue near the tumor and sensitizing the skin and eyes to light for up to six weeks after the treatment. A photodynamic therapy system can be integrated with the real-time f/a/LCI system 40 and the endoscopic probe 180. One possible implementation of an integrated photodynamic therapy system with a real-time f/a/LCI system is illustrated in
As illustrated in
The photodynamic therapy system 202 may be controlled by the computer 43 via a communications line(s) 203. The real-time f/a/LCI system 40 can provide guidance information that will help pinpoint where to use the photodynamic therapy on tissue 48. An advantage of guiding the photodynamic therapy should be reduced damage to nearby, non-cancerous tissue. Care would need to be taken to ensure that the light used for the real-time f/a/LCI system 40 does not activate the photodynamic therapy system 202 in a harmful manner. Some possible solutions include using low enough power levels for the real-time f/a/LCI system 40 as not to activate the photodynamic therapy system 202 to a harmful level or use a wavelength for the real-time f/a/LCI system 40 that is out of the range of the activation wavelength(s) for the photodynamic therapy system 202.
The endoscopic probe 180 may employ single or multi-instrument channels. A dual instrument channel variation is illustrated in
Another variation on this integrated system is the use of a hot or cold therapeutic to ablate or kill the abnormal tissue. The tissue can be locally heated or burned to destroy the cells. Alternately, the tissue could be chilled or frozen to achieve the same effect. There are numerous system implementations that will achieve this effect. A partial list includes placing a small heating coil at or near the end of the endoscopic probe 180 that is controlled by heater control unit 220 that in turn is controlled by the computer 43, as illustrated in
Another class of therapeutics involves removal of the non-normal (pre-cancerous or cancerous) tissue. This could be done via a variety of methods including cutting, scraping, using a punch biopsy, using an alligator clip biopsy and many others. One possible implementation is shown in
Another implementation is illustrated in
Note that the high power source therapeutic system 240 can either be continuous wave (CW) in operation or pulsed. Any wavelength can be used conceptually, selection will be driven by availability of sources and which wavelength(s) provide the best interaction with tissue to ablate abnormal tissue while minimizing effects on adjacent healthy tissue. Also, the multiple boxes shown for the computer 43, real-time f/a/LCI system 40, high power light source 240, and optical switching device 246 may be consolidated into fewer packages or devices.
The real-time f/a/LCI system 40 may also be used in conjunction with nanoparticles to modify the signal generated by the interaction with the sample and/or treat a condition within the sample. As an example, nanoparticles might be used to increase the optical contrast between the cell and the cell nuclei to increase the signal strength generated by the real-time f/a/LCI system 40. This may enable deeper penetration in the sample, which would be advantageous in many applications including the detection of skin cancer. Skin cancer is not normally detectable by f/a/LCI because the precancers or cancers start about one (1.0) millimeter below the surface and insufficient light reaches that depth and is scattered back. Increasing contrast may reduce the amount of light required to generate an f/a/LCI signal enabling deeper penetration in the tissue. Another application of f/a/LCI with nanoparticles is in the treatment of precancers or cancers. Nanoparticles can be used in a variety of treatment options for cancers, including using the nanoparticles which are toxic or carry toxic substances to kill precancerous or cancerous cells or tissue or using nanoparticles for photodynamic therapy where the nanoparticles absorb a light (perhaps from a specific wavelength or wavelength range) and heat up, thereby killing cells. For example, a real-time a/f/LCI system can be used to identify and diagnose the presence of pre-cancerous or cancerous tissue, and then during the same or concomitant medical procedure the physician can treat the tissue with the nanoparticles. Several such uses or therapies utilizing nanoparticles are known in the art as shown by the following references each of which is incorporated herein: O'Neal et al. Photo-thermal tumor ablation in mice using near infrared-absorbing nanoparticles, Cancer Letters 209:171-176 (2004); Gu et al., Targeted Nanoparticles for Cancer Therapy, NanoToday 2:14-21 (2007); Loo et al., Nanoshell-Enabled Photonics-Based Imaging and Therapy of Cancer, Technology in Cancer Research & Treatment, 3: 33-40 (2004).
Another embodiment is to use a standalone real-time f/a/LCI system 40 to provide monitoring of an area of tissue 48 with a therapeutic provided separately. This is illustrated by example in
Another implementation of the real-time f/a/LCI system 40 can be used in conjunction with an endoscope and scanning mechanism that permits the real-time f/a/LCI system 40 to scan more than one spot on the tissue 48.
The exemplary systems illustrated thus far have shown an independent computer 43 as part of the system. This is not a requirement. However, the processing of the f/a/LCI information regarding the tissue could be done by any type of computer, such as a laptop, desktop, remote computer (including one connected by a wireless network), or other. There may be varying levels of physical integration.
There is a range of automation that can be achieved with this system and all levels are intended to be covered by the present invention. As examples, low automation might be the case where the real-time f/a/LCI system generates information and displays it to the screen. Using this information, the operator delivers some dosage of some therapeutic to the tissue. In this case, there may be no electronic connection between the computer and the endoscope or the therapeutic control. A middle level of automation might be the case where there is a connection between the computer and the therapeutic delivery system and the computer determines the dosage level based on information from the real-time f/a/LCI system and internal algorithms. The operator would control when the therapeutic is delivered, but the dosage is determined via software. A very high level of automation might be the case where the therapeutic is delivered independent of operator control. As the tissue is being scanning (either manually or automatically), the computer can control the delivery of the therapeutic based on information received from the real-time f/a/LCI system and internal algorithms.
There are numerous possible configurations of the real-time f/a/LCI system 40 and therapeutic delivery techniques described above.
Some of the therapies discussed have been localized or regional in nature. f/a/LCI offers an advantage here by pinpointing the location(s) to apply one or more of these therapies. f/a/LCI and the information generated by real-time f/a/LCI systems may also be used to guide or determine the use of other therapies which may involve the whole body or areas outside the location where the pre-cancer or cancer may be found. This included many of the therapies used today including radiation, stereotactic radiosurgery or therapy (which uses multiple radiation beams to irradiate small targets with minimum impact to adjacent tissue, also known as gamma knife), surgery (including, but not limited to general, Moh's surgery, laparoscopic or minimally invasive surgery (MIS) and robotically assisted MIS), and chemotherapies (including both oral and injected chemotherapies). The real-time f/a/LCI may be used as part of a procedure for gating the use of these therapies.
In addition, the f/a/LCI systems disclosed herein can be used to detect in tissue the margin or boundary between pre-cancerous, cancerous or diseased cells and normal cells. Repeated application of real-time f/a/LCI is then used to direct the serial surgical removal of all or nearly all the pre-cancerous or cancerous cells in the same or concomitant medical procedure. Such combination of real-time f/a/LCI optical biopsy and surgical removal of pre-cancerous, cancerous or diseased issue can be applied to any organ of tissue of the body using the methods, processes, techniques and systems of the present inventions. As another option, these therapies (in particular, the chemotherapies) may be used in conjunction with one or more of the localized treatment options. As an example, a location in the esophagus may be identified as pre-cancerous by an f/a/LCI system leading to an RF ablation treatment for that area of the esophagus and a course of chemotherapy.
Early detection by the f/a/LCI may enable not only the use of chemotherapies, but also chemopreventatives that have been developed or are under development. These chemopreventatives have not been widely deployed because there may be no good way to identify pre-cancerous conditions at an early enough stage, and because of the difficulty in identifying and testing potential chemopreventatives because of the issues in identifying pre-cancerous conditions at an early enough stage and conducting longitudinal testing to validate the effectiveness of these chemopreventatives. One possible example of this would be the identification of a pre-cancerous lesion in an esophagus where the patient then undergoes a course of injected (or oral) chemopreventative followed by f/a/LCI monitoring exams at one or more time points to verify the reduction or elimination of the pre-cancerous lesion.
With the above backdrop, more detail regarding possible aspects of the systems are now described. In certain systems illustrated and previously described above, an endoscope probe tip 250 is shown in certain embodiments as a protective cover. The probe tip 250 may be disposable as a convenient means to keep the tip end 224 of the endoscope shaft 184 sterile so it can be used for multiple patients. In this regard,
As illustrated in
One function of the probe tip 250 can be to create a fixed geometry between an optical fiber probe 45, an imaging element, and the tissue 48 under examination. Thus, a first component that can comprise the probe tip 250 is a means to locate an imaging element, such as a lens 282, relative to the fiber optic or bundle probe 45.
If the probe tip 250 is employed in a real-time f/a/LCI system 40, the lens 282 can be placed approximately one focal length away from the fiber probe 45. This may be required for the lens 282 to properly capture the reflected angular distribution of light from the tissue for analysis. In alternate embodiments, the lens 282 can be positioned such that an individual single or multimode fiber or an array of such fibers is maintained at the focus of the lens 282. In other embodiments, the imaging lens 282 can be positioned at other distances from the fiber optic probe 45, which are different than the focal length of the lens 282.
The function of the optical window 294 is also to position the tissue relative to the lens 282 a proper distance from the tissue due to the rigid form of the cylindrical-shaped removable sheath member 288. The abutment of the optical window 294 to the tissue surface provides a fixed distance between the tissue surface and the lens 282 in the fixed sheath 284. This may be necessary to properly capture reflected light from the tissue on the lens 282. Maintaining the relationships between the tissue (via the optical window 294) and the lens 282, and between the lens 282 and the fiber probe 45 can be important in properly capturing reflected light from a tissue to analyze characteristics about its surface and/or underlying cell structures.
The optical window 294 may be perpendicular with respect to the longitudinal axis of the probe tip 250, as illustrated in
In an application of the probe tip 250 designed for a real-time f/a/LCI system, the optical window 294 is designed on the disposable removable sheath member 288 to be located approximately at the focal length of the lens 282. Providing the optical window 294 approximately one focal length away from the lens 282 allows the proper capture of the angular distributions of reflected light in the Fourier domain.
In alternative embodiments, the lens 282 may be integrated into the removable sheath member 288 as opposed to being integrated into the fixed sheath 284. Other alternative embodiments allow for different positioning of the optical window 294 relative to the lens 282.
In order to allow the removable sheath member 288 to be placed onto the probe tip 250 and removed after endoscopic application, a locking mechanism may also be included. This prevents having to wash the fixed sheath 284 after each endoscopic application since the fixed sheath 284 and the lens 282 are not exposed when protected by the removable sheath member 288. In this regard, the removable sheath member 288 is first placed onto the fixed sheath 284 prior to application. Thereafter, it may be locked into place to prevent the removable sheath member 288 from coming loose during application. After the probe tip 250 is removed from the endoscopic application, the removable sheath member 288 can be unlocked and removed for disposal. In this manner, the fixed sheath 284 and exposed lens 282, which may be one of the more expensive components of the probe tip 250, are never exposed to the tissue and do not have to be washed.
In the embodiments shown in
While the removable sheath member 288 described above will prevent direct contamination of the distal face of the fiber probe 45, it is possible that fluids could penetrate through the locking pin channel 298 or come in contact with the portion of the fiber probe 45, which is not covered by the removable sheath member 288. For this reason, the probe tip 250 can be designed to additionally incorporate a deployable sterile skirt 302 which can prevent such contamination.
In the illustrated embodiment, the skirt 302 is attached to the removable sheath member 288 at a point distal to the locking pin 296 and locking pin channel 298. The skirt 302 can be composed of a plastic or latex material, suitable for preventing fluid from reaching the channel or bundle. The skirt 302 may be lubricated with any type of lubricant desired before being attached to the removable sheath member 298 and/or prior to endoscopic application. Prior to deployment, the skirt 302 may be coiled or otherwise collapsed to allow for facile manipulation of the locking pin 296 within the locking pin channel 298, as illustrated in
The remainder of the present application provides additional embodiments of real-time f/a/LCI systems that may be employed in the same or concomitant procedures described above. A Fourier domain optical biopsy system is possible that is not angle-resolved. These systems are referred to as fLCI systems. One exemplary embodiment of a fLCI system 320 is shown in
This illumination scheme achieves Kohler illumination in that the fiber acts as a field stop, resulting in the proper alignment of incident or illuminating light and thereby achieving critical illumination of the sample. In the fLCI system 320, the white light beam is split by the beamsplitter 406 (BS) into a reference beam 405 and an input beam 407 to the sample 408. The light scattered by the sample 408 is recombined at the BS 406 with light reflected by the reference mirror 414 (M).
The reference beam 405 in conjunction with the reference mirror 414 forms a portion of a reference arm that receives a first reference light and outputs a second reference light. The input beam 407 and the sample 408 form a portion of a sample arm that receives a first sample light and outputs a second sample light.
Those skilled in the art will appreciate that the light beam can be split into a plurality of reference beams and input beams (e.g., N reference beams and N input beams) without departing from the spirit and scope of the present invention. Further, the splitting of the beams may be accomplished with a beamsplitter or a fiber splitter in the case of an optical fiber implementation of an exemplary embodiment of the present invention.
In the exemplary embodiment of the present invention shown in
The detected signal is linearly related to the intensity as a function of wavelength I(λ), which can be related to the signal and reference fields (Es, Er) as:
I(λ)=Es(λ)|2+Er(λ)|2+2ReEs(λ)E*r(λ) cos φ (8)
where φ is the phase difference between the two fields and < . . . > denotes an ensemble average.
The interference term is extracted by measuring the intensity of the signal and reference beams independently and subtracting them from the total intensity.
The axial spatial cross-correlation function, ΓSR(z) between the sample and reference fields is obtained by resealing the wavelength spectrum into a wavenumber (k=2π/λ) spectrum then Fourier transforming:
ΓSR(z)=∫dkeikzEs(k)E*r(k) cos φ (9)
This term is labeled as an axial spatial cross-correlation as it is related to the temporal or longitudinal coherence of the two fields.
Another exemplary embodiment of an fLCI scheme is shown in
In
The reference light in reference fiber 425, in conjunction with a lens 426 (preferably an aspheric lens) and the reference mirror 428, forms a portion of a reference arm that receives a first reference light and outputs a second reference light. Specifically, reference light in reference fiber 425 is directed to the reference mirror 428 by lens 426, and the reference light reflected by the reference mirror 428 (second reference light) is coupled back into the reference fiber 425 with lens 426. The sample light in sample fiber 427 and the sample 430 form a portion of a sample arm that receives a first sample light and outputs a second sample light. Specifically, sample light in sample fiber 427 is directed to the sample 430 by lens 434 (preferably as aspheric lens), and at least a portion of the sample light scattered by the sample 430 is coupled into the sample fiber 427 by lens 431. In the exemplary embodiment shown in
At least a portion of the reflected reference light in reference fiber 425 and at least a portion of the scattered sample light on sample fiber 427 are coupled into a detector fiber 433 by the FS 424. The output of detector fiber 433 coincides with the input of a miniature spectrograph 432, where the light is spectrally dispersed and detected.
From this autocorrelation, the coherence length of the field, lc=1.21 μm is determined. This is slightly larger than the calculated width of lc=2/Δk1/c=0.98 μm, with any discrepancy most likely attributed to uncompensated dispersion effects. Note that rescaling the field into wavenumber space is a nonlinear process which can skew the spectrum if not properly executed.
In data processing, a fitting algorithm is applied (e.g., a cubic spline fit) to the rescaled wavenumber spectrum and then resampled (e.g., resample with even spacing). The resampled spectrum is then Fourier transformed to yield the spatial correlation of the sample. Those skilled in the art will appreciate that other frequency-based algorithms or combinations of algorithms can be used in place of the Fourier transform to yield spatial correlation. One example of a software tool that can be used to accomplish this processing in real time or near real time is to use LabView™ software.
In one exemplary embodiment of the present invention, the sample consists of a glass coverslip (e.g., thickness, d˜200 μm) with polystyrene beads which have been dried from suspension onto the back surface (1.55 μm mean diameter, 3% variance). Thus, the field scattered by the sample can be expressed as:
E
s(k)=Efront(k)eik
In Equation 10, Efront and Eback denote the field scattered by the front and back surfaces of the coverslip, and δz is the difference between the path length of the reference beam and that of the light reflected from the front surface and n the index of refraction of the glass. The effect of the microspheres will appear in the Eback term as the beads are small and attached closely to the back surface. Upon substituting Equation 10 into Equation 9, a two peak distribution with the width of the peaks given by the coherence length of the source is obtained.
In order to obtain spectroscopic information, a Gaussian window is applied to the interference term before performing the Fourier transform operation. Those skilled in the art will appreciate that other probabilistic windowing methodologies may be applied without departing from the spirit and scope of the invention. This makes it possible to recover spectral information about light scattered at a particular depth.
The windowed interference term takes the form:
E
s(k)E*r(k)exp[−((k−kw)/Δkw)2]. (1)
The proper sizing of a windowed interference term can facilitate the processing operation. For example, by selecting a relatively narrow window (Δkw small) compared to the features of Es and Ek, we effectively obtain <Es(kw)E*r(kw)>. In processing the data below, we use Δkw=0.12 μm−1 which degrades the coherence length by a factor of 16.7. This exemplary window setting enables the scattering at 50 different wavenumbers over the 6 μm−1 span of usable spectrum.
In
Note that the correlation function is symmetric about z=0, resulting in a superposed mirror image of the scan. Since these are represented as cross-correlation functions, the plots are symmetric about z=0. Thus, the front surface reflection for z>0 is paired with the back surface reflection for z<0, and vice versa.
In
To obtain the spectrum of the scattered light, we repeatedly apply the Gaussian window and increase the center wavenumber by 0.12 μm−1 between successive applications. As mentioned above, Δkw=0.12 μm−1 is used to degrade the coherence length by a factor of 16.7. This results in the generation of a spectroscopic depth-resolved reflection profile.
In
In
For comparison, the same ratio for the front surface reflections (dashed line in
There are many applications of the various exemplary embodiments of the present invention. One exemplary application of fLCI is in determining the size of cell organelles, in particular the cell nucleus, in epithelial tissues. In biological media, for example, the relative refractive indices are lower for organelles compared to microspheres and thus, smaller scattering signals are expected. The use of a higher power light source will permit the smaller signals to be detected. Other examples include detection of sub-surface defects in manufactured parts, including fabricated integrated circuits, detection of airborne aerosols, such as nerve agents or biotoxins, and detection of exposure to such aerosols by examining epithelial tissues within the respiratory tract.
Additionally, the larger the size of the nucleus (compared to the microspheres in this experiment), the higher the frequency modulation of the spectrum. Those skilled in the art will appreciate that higher frequency oscillations are detected at a lower efficiency in Fourier transform biopsy techniques. Therefore, in order to detect these higher frequency oscillations, a higher resolution spectrograph is used.
Embodiments disclosed herein also involve new low-coherence interferometry (LCI) techniques which enable acquisition of structural and depth information regarding a sample of interest at rapid rates. A sample can be tissue or any other cellular-based structure. The acquisition rate is sufficiently rapid to make in vivo applications feasible. Measuring cellular morphology in tissues and in vivo as well as diagnosing intraepithelial neoplasia and assessing the efficacy of chemopreventive and chemotherapeutic agents are possible applications. Prospectively grading tissue samples without tissue processing is also possible, demonstrating the potential of the technique as a biomedical diagnostic.
In one embodiment, a “swept-source” (SS) light source is used in LCI to obtain structural and depth information about a sample. The swept-source light source is used to generate a reference signal and a signal directed towards a sample. Light scattered from the sample is returned as a result and mixed with the reference signal to achieve interference and thus provide structural and depth-resolved information regarding the sample. With a “swept-source” light source, the light source is controlled or varied to sweep the center wavelength of a narrow band of emitted light over a given range of wavelengths, thus synthesizing a broad band source. Because the light is emitted in particular wavelengths or narrower ranges of wavelengths during emission, scattered light returned from the sample is known to be in response to a particular wavelength or range of wavelengths. Thus, the returned scattered light is spectrally-resolved and depth-resolved, because the returned light is in response to the light source emitted light over a narrow spectral range. This is opposed to a wider or light source that generates all wavelengths of light in one light emission in time, wherein the returned scattered light from the sample contains scattered light at a broad range of wavelengths. In this instance, a spectrometer is used to spectrally-resolve the returned scattered light. However, when using a swept-source light source, the series of returned scattered lights from the sample at each wavelength are already in the spectral domain to provide spectrally-resolved information about the sample. The spectrally-resolved information about the sample can be detected.
Another embodiment involves using a swept-source light source in angle-resolved low-coherence interferometry (a/LCI), referred to herein as “swept-source Fourier domain a/LCI,” or “SS a/LCI.” The data acquisition time for SS a/LCI can be less than one second, a threshold which is desirable for acquiring data from in vivo tissues. The swept-source light source is employed to generate a reference signal and a signal directed towards a sample over the swept range of wavelengths or ranges of wavelengths. The light is either directed to strike the sample at an angle, or the light source or another component in the system (e.g., a lens) is moved to direct light onto the sample at an angle or plurality of angles (i.e., two or more angles), which may include a multitude of angles (i.e., more than two angles). This causes a set of scattered light to be returned from the sample at a plurality of angles, thereby representing spectrally-resolved and angle-resolved (also referred to herein as “spectral and angle-resolved”) scattered information about the sample from a plurality of points on the sample. The spectral and angle-resolved scattered information about the sample can be detected. This SS a/LCI embodiment can also use the Fourier domain concept to acquire depth-resolved information. It has recently been shown that improvements in signal-to-noise ratio, and commensurate reductions in data acquisition time are possible by recording the depth scan in the Fourier (or spectral) domain. In this embodiment, the SS a/LCI system can combine the Fourier domain concept with the use of a swept-source light source, such as a swept-source laser, and a detector, such as a line scan array or camera, to record the angular distribution of returned scattered light from the sample in parallel and the frequency distribution in time.
Lenses (L3) 1028 and (L4) 1030 are arranged to produce a collimated pencil beam 1032 incident on the sample 1017 (block 66,
The light scattered by the sample 1017 is collected by lens (L4) 1030 (block 1068,
In this embodiment, the detector device 1026 is a one-dimensional detection device in the form of a line scan array, which is comprised of a plurality of detectors. This allows the detector device 1026 to receive light at the plurality of scatterer angles from the sample 1017 and mixed with the reference beam 1014 at the same time or essentially the same time to receive spectral information about the sample 1017. Providing the line scan array 1026 allows detection of the angular distribution of the combined beams 1044, or said another way, at multiple scatter angles. Each detector in the detector device 1026 receives scattered light from the sample 1017 at a given angle at the same time or essentially the same time.
Because the emitted light from the swept-source light source 1012 is broken up into particular wavelengths or narrower ranges of wavelengths during emission, returned scattered light 1034 from the sample 1017 is known to be in response to a particular wavelength or range of wavelengths. Thus, the returned scattered light 1034 is spectrally-resolved, because the returned scattered light 1034 is in response to the light source emitted light over a spectral domain. This is opposed to a wider or broadband light source that generates all wavelengths of light in one light emission at the same time, wherein the returned scattered light from the sample contains scattered light at all wavelengths. In this instance, a spectrometer is used to spectrally-resolve the returned scattered light. However, when using the swept-source light source 1012, the series of returned scattered light 1034 from the sample 1017 at each wavelength is already in the spectral domain to provide spectrally-resolved information about the sample.
I(λm,yn)=|Er(λm,yn)|2+Es(λm,yn)|2+2ReEs(λm,yn)E*r(λm,yn) cos φ (12)
where Φ is the phase difference between the two fields and . . . denotes an ensemble average in time. The interference term is extracted by measuring the intensity of the scattered light 1034 and reference beam 1014 independently and subtracting them from the total intensity. In one method of obtaining depth-resolved information about the sample 1017, the wavelength spectrum at each scattering angle is interpolated into a wavenumber (k=2π/λ) spectrum and Fourier transformed to give a spatial cross correlation, ΕSR(z) for each vertical pixel yn:
ΕSR(z,yn)=∫dkeikzEs(k,yn)E*r(k,yn) cos φ (13)
The reference field takes the form:
E
r(k)=Eoexp[−((k−ko)/Δk)2]exp[−((y−yo)/Δy)2]exp[ikΔl] (14)
where ko (yo and Δk (Δy) represent the center and width of the Gaussian wavevector (spatial) distribution and Δl is the selected path length difference. The scattered sample field takes the form:
E
s(k,θ)=ΣjEoexp[−((k−ko)/Δk)2]exp[iklj]Sj(k,θ) (15)
where Sj represents the amplitude distribution of the scattering originating from the jth interface, located at depth lj. The angular distribution of the scattered sample field is converted into a position distribution in the Fourier image plane of lens (L4) 1030 through the relationship y=f4θ. For the exemplary pixel size of the line scan array 1026 of eight (8) to twelve (12) micrometers (μm), this yields an angular resolution of 0.00028 to 0.00034 mradians and an expected angular range of 286 to 430 mradians for a 1024 element array. Inserting Equations (14) and (15) into Equation (13) and noting the uniformity of the reference field (Δy>>camera height) yields the spatial cross correlation at the nth vertical position on the detector:
Evaluating this equation for a single interface yields:
ΕSR(z,yn)=|Eo|2exp[−((z−Δl+lj)Δk)2/8]Sj(ko,θn=yn/f4)cos φ (17)
Here, it is assumed that the scattering amplitude S does not vary appreciably over the bandwidth of the source. This expression shows obtaining a depth-resolved profile of the scattering distribution with each vertical pixel corresponding to a scattering angle. The techniques described in U.S. patent application Ser. No. 11/548,468 entitled “Systems and Methods for Endoscopic Angle-Resolved Low Coherence Interferometry,” which is incorporated herein by reference in its entirety, may be used for obtaining structural and depth-resolved information regarding scattered light from a sample.
To obtain the same or similar data set as is obtained from a single frame capture from an imaging spectrometer using a broadband light source, the SS a/LCI apparatus and system 1010 can capture a series of data acquisitions from the line scan array 1026 at each wavelength and combine them. In this embodiment, the data acquisition rate of the line scan arrays 1026 is less than the sweep rate of the swept-source light source 1012. If one were to assume that 1000 wavelength (frequency) points are needed (and thus points in time for the swept-source), ten (10) to twenty (20) data acquisitions of scattered information from the sample 1017 may be recovered per second using a line scan array. For example, this scenario could yield a time per acquisition of 50 to 100 milliseconds, which is satisfactory for clinical and commercial viability.
Line scan arrays and camera detector devices are widely available for both the visible and the near infrared wavelengths. Visible line scan arrays can operate from approximately ˜400 nm to ˜900 nm, for example, and may be based on silicon technology. Near infrared line scan arrays may operate from approximately ˜900 nm to ˜1700 nm or further. Table 2 below gives some typical specifications from several manufacturers as examples.
As previously discussed above, a swept-source laser may be employed as the swept-source light source 1012. Some examples are provided in Table 3 below.
Faster acquisition times are possible. Swept-source light sources at shorter wavelengths will allow use of a high speed detector 1026, such as silicon detectors for example. For example, some Atmel® silicon-based cameras can achieve 100,000 lines per second, potentially allowing 100 data point acquisitions per second or 10 milliseconds per acquisition. Alternately, as another example, the line scan array 1026 may be based on InGaAs technology and may be faster, reaching readout rates of 50,000 to 100,000 lines per second and thus reducing the acquisition time to 10 milliseconds. It is expected that the sweep rate, power, wavelength range, and other performance characteristics of the swept-source light sources can enable high performance versions of the a/LCI apparatuses and systems, including the SS a/LCI apparatus and system 1010 of
In addition to obtaining depth-resolved information about the sample 1017, the scattering distribution data (i.e., a/LCI data) obtained from the sample 1017 using the disclosed data acquisition scheme can also be used to make a size determination of the nucleus using the Mie theory, as previously discussed. A filtered curve is determined using the scattered data. Comparison of the filtered scattering distribution curve (i.e., a representation of the scattered data) to the prediction of Mie theory enables a size determination to be made.
In order to fit the scattered data to Mie theory, the a/LCI signals are processed to extract the oscillatory component which is characteristic of the nucleus size. The smoothed data is fit to a low-order polynomial (2nd order is typically used but higher order polynomials, such as 4th order, may also be used), which is then subtracted from the distribution to remove the background trend. The resulting oscillatory component can then be compared to a database of theoretical predictions obtained using Mie theory from which the slowly varying features were similarly removed for analysis.
A direct comparison between the filtered a/LCI data and Mie theory data may not be possible, as the Chi-squared fitting algorithm tends to match the background slope rather than the characteristic oscillations. The calculated theoretical predictions include a Gaussian distribution of sizes characterized by a mean diameter (d) and standard deviation as well as a distribution of wavelengths to accurately model the broad bandwidth source.
The best fit can be determined by minimizing the Chi-squared between the data and Mie theory, yielding a size of 10.2.+/−.1.7 μm, in excellent agreement with the true size. The measurement error is larger than the variance of the bead size, most likely due to the limited range of angles recorded in the measurement.
As an alternative to processing the a/LCI data and comparing to Mie theory, there are several other approaches which could yield diagnostic information. These include analyzing the angular data using a Fourier transform to identify periodic oscillations characteristic of cell nuclei. The periodic oscillations can be correlated with nuclear size and thus will possess diagnostic value. Another approach to analyzing a/LCI data is to compare the data to a database of angular scattering distributions generated with finite element method (FEM) or T-Matrix calculations. Such calculations offer superior analysis as they are not subject to the same limitations as Mie theory. For example, FEM or T-Matrix calculations can model non-spherical scatterers and scatterers with inclusions while Mie theory can only model homogenous spheres. Other techniques are described in U.S. Pat. No. 7,102,758 entitled “Fourier Domain Low-Coherence Interferometry for Light Scattering Spectroscopy Apparatus and Method,” which is incorporated herein by reference in its entirety.
In another embodiment of the invention, an SS a/LCI apparatus and system can be provided, including for endoscopic applications, by using optical fibers to deliver and collect light from the sample of interest. These alternative embodiments are illustrated in
Turning now to
Turning to
The scattered sample field is detected using a coherent fiber bundle. The scattered sample field is generated using light in the signal arm 1082 which is directed toward the sample of interest using lens (L2) 1098. As with the free space system, lens (L2) 1098 is displaced laterally from the center of single-mode fiber (F2) such that a collimated beam is produced which is traveling at an angle relative to the optical axis. The fact that the incident beam strikes the sample at an oblique angle is essential in separating the elastic scattering information from specular reflections. The scattered light 1034′ is collected by a fiber bundle consisting of an array of coherent single mode or multi-mode fibers. The distal tip of the fiber is maintained one focal length away from lens (L2) 1098 to image the angular distribution of scattered light. In the embodiment shown in
As illustrated in
There are several possible schemes for creating the fiber probe which are the same from an optical engineering point of view. One possible implementation would be a linear array of single mode fibers in both the signal and reference arms. Alternatively, a reference arm 1096 could be composed of an individual single mode fiber with the signal arm 1082 consisting of either a coherent fiber bundle or linear fiber array.
The probe 1093 can also have several implementations which are substantially equivalent. These would include the use of a drum or ball lens in place of lens (L2) 1098. A side-viewing probe could be created using a combination of a lens and a mirror or prism or through the use of a convex mirror to replace the lens-mirror combination. Finally, the entire probe can be made to rotate radially in order to provide a circumferential scan of the probed area.
Another exemplary embodiment of a fiber optic SS a/LCI system is the illustrated a/LCI system 1010″ in
As illustrated in
The use of a swept-source light source also opens up the possibility of another system architecture that has the capability to acquire scattering information from more than one scattering plane from a sample. This implementation is referred to as a “Multiple Angle Swept-source a/LCI” system or MA SS a/LCI. An example of an MA SS a/LCI system 1010′″ is illustrated in
The MA SS a/LCI system 1010′″ is exemplified in
The MA SS a/LCI system 1010′″ may also be implemented using a broadband light source, such as a superluminescent diode (SLD), and using a spectrometer detection device. In either case, whether using a broadband light source or swept-source light source 1012, in the fiber optic embodiment of a MA SS a/LCI system 1010′″, the fiber bundle 1094 that receives the combined beams 1044 from the sample 1017 can be captured by a plurality of optical fibers 1119 in the fiber bundle 1094, as illustrated in
One possible distribution of the scattering angles across the CCD camera 1026″ is shown in
Potential components for the CCD camera 1026″ include but are not limited to a Cascade:Photometrics™ 650 CCD camera as the image detector. For the light source, the Thorlabs INTUN™ continuously tunable laser is an example of one of many suitable sources. This example would be useful because the center wavelength is 780 nm, which is compatible with standard NIR optical elements, including the Cascade camera, and offers a tuning range of 15 mm, which is comparable to the line width used in SS a/LCI systems previously described. The tuning speed of 30 nm/s for this source is optimal for synchronization with the Cascade CCD camera as better than 0.1 nm resolution can be achieved based on the 300 Hz frame rate which can be realized when using a region of interest with the Cascade CCD. The SS a/LCI scheme will improve acquisition time and upgrade the a/LCI system to a state-of-the-art technology for studies of cell mechanics at faster time scales.
The data acquisition may be limited by the frame rate of the CCD camera 1026″ and not by the sweep speed of the swept-source light source 1012. Table 4 below lists exemplary CCD cameras. The fastest listed is only 1000 frames per second, so if 1000 wavelength points are required, a full scan will take approximately 1 second. It may be possible to scan faster if fewer pixels are needed in this example, or if fewer points in the wavelength can be used. Several of these cameras will let the user target specific regions of interest to acquire images, thus speeding up the frame rate. For example, with the Atmel® camera, if one uses a region of interest that is 100×100 pixels for a total of 10,000 pixels, then the frame rate might be as high as 15,000 frames per second allowing a scan time of 70 milliseconds for 1000 wavelength points. It is expected that the speed of the CCD cameras will increase over time and the increased camera speed will translate into higher performance of the MA SS a/LCI system.
In addition to the SS a/LCI and MA SS a/LCI implementations described herein, a time-domain a/LCI implementation is also possible. An example of this a/LCI system 1130 implementation is shown by example in
The system 1130 uses photodiode arrays #1 and #21132, 1134 to collect angular scattered light from the sample (not shown). The system 1130 provides a swept-source light source 1136 in the form of a Ti:Sapphire laser operating in a pulsed mode in this embodiment. The swept-source light source 1136 directs light 1138 to a beam splitter (BS1) 1140, which splits the light 1138 into a reference signal 1141 and sample signal 1142. The reference signal 1141 goes through acousto optic modulator (AOM) 1144 with w+10 MHz, and then through retroreflector (RR) 1154 mounted on a reference arm 1153, wherein the retroreflector (RR) 1154 is moved by a distance, δz to change the depth in the sample to perform depth scans. The sample signal 1142 goes through AOM 1146 with frequency ‘ω’ and then through imaging optics 1148. Imaging optics 1148 shine collimated light onto the sample and then collects the angular scattered light from the sample. The reference signal 1141 and the angular scattered light are combined at beamsplitter (BS2) 1152 and then imaged onto the photodiode arrays #1 and #21132, 1134. Signals 1135, 1137 from each photodiode 1132 or 1134 are subtracted from the photodiode in the other array 1132 or 1134 which corresponds to the same angular location. A multi-channel demodulator 1160 is used on a subtracted signal 1139. All signals then go to a computer 1162 for processing. Processing of the time-domain depth information from the subtracted signal 1139 and received by the multi-channel demodulator 1160 can be performed just as previously described in above for this embodiment, as possible examples or methods.
Even though the systems 1130 illustrated in
Note that this system uses some means of subtracting the signals 1135, 1137 on the photodiodes arrays 1132, 1134 on a photodiode basis and then demodulating each channel. This may be accomplished in a serial or parallel fashion. One implementation would be to digitally acquire data from the photodiode arrays (as in the case of a line scan camera) and then use a digital signal processor (DSP) chip or similar to subtract and demodulate the data. This may require that the offset frequency between the two AOMs be less than the line rate of the line scan arrays. Since line scan arrays that receive signal data up to 100,000 lines/second exist, an offset of <50 KHz may be acceptable.
A second implementation would be to use the photodiode arrays 1132, 1134 and perform the subtraction in an analog basis. It may be the case that the two photodiode arrays are actually two sections of the same two-dimensional array. There also may then be a dedicated demodulator for each photodiode pair or, again, a digitizer and appropriate digital signal processor (DSP) chips.
In another embodiment and approach to collecting information about a sample of interest, a step forward from time domain a/LCI systems is taken to still collect the angular information in a serial fashion. However, depth information is collected from a sample of interest using a Fourier domain approach. The light source that may be used can include a broadband light source in combination with a spectrometer to process spectrally-resolved information about the sample. Alternatively, a swept-source light source with a photodiode or another implementation may be used.
Since this system 1170 does not use a time domain data acquisition approach, the AOMs 1144, 1146 and the moving retroreflector (RR) 1154 in the reference arm 1153, as provided in the systems 1130 in
As illustrated in
Another implementation of a/LCI is a multi-spectral a/LCI system. Embodiments of multi-spectral a/LCI systems 1210, 1210′ are illustrated in
The system 1210 of
The super-continuum light source 1213 directs light 1212 to a beam splitter (BS1) 1215, which splits the light 1216 into a reference signal 1217 and sample signal 1218. The reference signal 1217 goes through AOM 1221, and then through retroreflector (RR) 1219 mounted on a reference arm 1220, wherein the retroreflector (RR) 1219 is moved by the reference arm 1220 to change the depth in the sample to perform depth scans. The sample signal 1218 goes through AOM 1222 with frequency ‘ω’ and then through imaging optics 1223. Imaging optics 1223 shine light from the super-continuum light source 1213 onto a sample and then collects the angular scattered light from the sample. The reference signal 1217 and the angular scattered light are combined at beamsplitter (BS2) 1224 and then imaged onto the photodiode arrays #1 and #21211, 1212. Signals 1225, 1226 from each photodiode array 1211 or 1212 are subtracted from the photodiode in the other array 1211 or 1212 which corresponds to the same angular location. A multi-channel demodulator 1228 is used on the resulting subtracted signal 1227. The subtracted signal 1227 travels to a computer 1230 for processing.
Another approach to the multi-spectral a/LCI system 1210 in
As illustrated in
It is possible to provide this system 1210′ with one spectrometer, although the combination of multiple spectrometers allows for high spectral resolution for the Fourier domain depth detection and the broad range of wavelengths needed to acquire the multi-spectral information. The system 1210′ can be expanded to as many sections of the optical spectrum as needed. Fiber implementations based on fiber couplers and fiber filters are also possible.
The system 1210′ may also be provided with a broadband swept-source light source for the acquisition of depth information and the acquisition of multi-spectral information. Another approach is to multiplex together multiple sources at different wavelengths to obtain the multi-spectral information. For example, an 830 nm center wavelength, 20 nm 3 dB width SLD could be multiplexed together with a 650 nm center wavelength, 15 nm 3 dB width SLD to obtain a/LCI information at two wavelengths. Further, as the various wavelengths become farther apart, it may be necessary to put in compensation components to account for the variation in index of refraction at the different wavelengths. For example, if one is using a 400 nm and an 800 nm wavelength, it may be the case that when the interferometer arms are path length matching for the 400 nm wavelength, they are mismatched for the 800 m wavelength by more than the imaging depth available with the spectrometer (typically 1 to 2 mm).
The f/a/LCI systems and methods described herein can be clinically viable methods for assessing tissue health without the need for tissue extraction via biopsy or subsequent histopathological evaluation. The f/a/LCI systems and methods described herein can be applied for a number of purposes: for example, early detection and screening for dysplastic tissues, disease staging, monitoring of therapeutic action, and guiding the clinician to biopsy or surgery sites. The non-invasive, non-ionizing nature of the optical biopsy based on an f/a/LCI probe means that it can be applied frequently without adverse affect. The potential of f/a/LCI to provide rapid results will greatly enhance its widespread applicability for disease screening.
Nuclear morphology measurement is also possible using the f/a/LCI systems and methods described herein. Nuclear morphology is a necessary junction between a cell's topographical environment and its gene expression. One application of the f/a/LCI systems and methods is to connect topographical cues to stem cell function by investigating nuclear morphology. In one embodiment, the f/a/LCI systems and methods use a swept-source light source approach described herein and create and implement light scattering models. The second is to provide nuclear morphology as a function of nanotopography. Finally, by connecting nuclear morphology with gene expression, the structure-function relationship of stem cells, e.g., human mesenchymal stem cells (hMSC), under the influence of nanotopographic cues can be established.
The f/a/LCI methods, processes, techniques, and systems described herein can also be used for cell biology applications and medical treatment based on such applications. Accurate measurements of nuclear deformation, i.e., structural changes of the nucleus in response to environmental stimuli, are important for signal transduction studies. Traditionally, these measurements require labeling and imaging, and then nuclear measurement using image analysis. This approach is time-consuming, invasive, and unavoidably perturbs cellular systems. The f/a/LCI techniques described herein offer an alternative for probing physical characteristics of living systems. The f/a/LCI techniques disclosed herein can be used to quantify nuclear morphology for early cancer detection, diagnosis and treatment, as well as for noninvasively measuring small changes in nuclear morphology in response to environmental stimuli. With the f/a/LCI methods, processes, techniques, and systems provided herein, high-throughput measurements and probing aspherical nuclei can be accomplished. This is demonstrated for both cell and tissue engineering research. Structural changes in cell nuclei or mitochondria due to subtle environmental stimuli, including substrate topography and osmotic pressure, are profiled rapidly without disrupting the cells or introducing artifacts associated with traditional measurements. Accuracy of better than 3% can be obtained over a range of nuclear geometries, with the greatest deviations occurring for the more complex geometries.
In one embodiment disclosed herein, the f/a/LCI systems and methods described herein are used to assess nuclear deformation due to osmotic pressure. Cells are seeded at high density in chambered coverglasses and equilibrated with 500, 400 and 330 mOsm saline solution, in that order. Nuclear diameters are measured in micrometers to obtain the mean value +/− the standard error within a 95% confidence interval. Changes in nuclear size are detected as a function of osmotic pressure, indicating that the f/a/LCI systems and methods disclosed herein can be used to detect cellular changes in response to factors which affect cell environment. One skilled in the art would recognize that many biochemical and physiological factors can affect cell environment, including disease, exposure to therapeutic agents, and environmental stresses.
To assess nuclear changes in response to nanotopography, cells are grown on nanopatterned substrates which create an elongation of the cells along the axis of the finely ruled pattern. The f/a/LCI systems and processes disclosed herein are applied to measure the major and minor axes of the oriented spheroidal scatterers in micrometers through repeated measurements with varying orientation and polarization. A full characterization of the cell nuclei is achieved, and both the major axis and minor axis of the nuclei is determined, yielding an aspect ratio (ratio of minor to major axes).
The f/a/LCI systems and methods disclosed herein can also be used for monitoring therapy. In this regard, the f/a/LCI systems and methods are used to assess nuclear morphology and subcellular structure within cells (e.g., mitochondria) at several time points following treatment with chemotherapeutic agents. The light scattering signal reveals a change in the organization of subcellular structures that is interpreted using a fractal dimension formalism. The fractal dimension of sub-cellular structures in cells treated with paclitaxel and doxorubicin is observed to increase significantly compared to that of control cells. The fractal dimension will vary with time upon exposure to therapeutic agents, e.g., paclitaxel, doxorubicin and the like, demonstrating that structural changes associated with apoptosis are occurring. Using T-matrix theory-based light scattering analysis and an inverse light scattering algorithm, the size and shape of cell nuclei and mitochondria are determined. Using the f/a/LCI systems and methods disclosed herein, changes in sub-cellular structure (e.g., mitochondria) and nuclear substructure, including changes caused by apoptosis, can be detected. Accordingly, the f/a/LCI systems and processes described herein have utility in detecting early apoptotic events for both clinical and basic science applications.
Although embodiments disclosed herein have been illustrated and described herein with reference to preferred embodiments and specific examples thereof, it will be readily apparent to those of ordinary skill in the art that other embodiments and examples can perform similar functions and/or achieve like results. The previous description of the disclosure is provided to enable any person skilled in the art to make or use the disclosure. Various modifications to the disclosure will be readily apparent to those skilled in the art, and the generic principles defined herein may be applied to other variations without departing from the spirit or scope of the disclosure. All such equivalent embodiments and examples are within the spirit and scope of the present invention and are intended to be covered by the appended claims.
It will also be apparent to those skilled in the art that various modifications and variations can be made to the present invention without departing from the spirit and scope of the invention. Thus, the disclosure is not intended to be limited to the examples and designs described herein, but is to be accorded the widest scope consistent with the principles and novel features disclosed herein. For example, the present invention is not limited to a particular Fourier domain or angle-resolved optical biopsy system, tissue type examined, therapy or therapeutic, an endoscope or endoscopic probe, control systems or interfaces, or methods, processes, techniques disclosed herein and their order.
The embodiments set forth above represent the necessary information to enable those skilled in the art to practice the invention and illustrate the best mode of practicing the invention. Upon reading the following description in light if the accompanying drawings figures, those skilled in the art will understand the concepts of the invention and will recognize applications of these concepts not particularly addressed herein. It should be understood that these concepts and applications fall within the scope of the disclosure and the claims that follow.
This patent application claims priority to U.S. Provisional Patent Application Ser. No. 61/019,662, filed on Jan. 8, 2008 and entitled “Systems and Methods for Tissue Diagnostic, Monitoring, and/or Therapy,” which is incorporated herein by reference in its entirety. This patent application is related to U.S. Pat. No. 7,102,758, filed on May 6, 2003 and entitled “Fourier Domain Low-Coherence Interferometry for Light Scattering Spectroscopy Apparatus and Method,” which is incorporated herein by reference in its entirety. This patent application is also related to U.S. patent application Ser. No. 11/548,468, filed on Oct. 11, 2006 and entitled “Systems and Methods for Endoscopic Angle-Resolved Low Coherence Interferometry,” which is incorporated herein by reference in its entirety. This patent application is also related to U.S. patent application Ser. No. 12/210,620, filed on Sep. 15, 2008 and entitled “Apparatuses, Systems, and Methods for Low-Coherence Interferometry (LCI),” which is incorporated herein by reference in its entirety. This patent application is also related to U.S. patent application Ser. No. 11/780,879, filed on Jul. 20, 2007 and entitled “Protective Probe Tip, Particularly for Use on a Fiber-Optic Probe Used in an Endoscopic Application,” which is incorporated herein by reference in its entirety.
Number | Date | Country | |
---|---|---|---|
61019662 | Jan 2008 | US |