This application relates to methods of administrating medicinal therapies and in particular to a method of administering insulin according to a protocol to rapidly and safely reach a targeted blood glucose level.
Insulin is a particularly difficult medicine to administer. Not only does the proper dose of insulin vary from person to person, the proper dose also varies from day to day and dosing time to dosing time depending on the food to be ingested and the amount of physical exertion the person has expended during the course of the day. Hence the amount of insulin to be administered at any given time must be determined using a number of different factors. The correct amount varies from a low of a few units in any one administration, which may be give up to 4 times per day to over 150 units per day. This situation is further exacerbated by the fact that insulin is not bioavailable when taken by mouth. Since its discovery in 1922, only two methods of administration of insulin have proved practical. The first method is by syringe and vial, where the person withdraws the desired amount of insulin from a vial into a syringe and injects the desired dose under the skin. The second method, available only for the last ten to fifteen years, is the insulin pen where the insulin is contained in the pen and the user injects a selected dose under the skin by selecting the desired dose using a dial on the pen. In both of these methods, the amount of insulin is selectable by the unit, that is, over the range of 10 units to 150 units, the dose is selected to the nearest unit.
In selecting a dose of insulin, either the patient or the physician in charge of the management of the patient's diabetes must make a determination of the proper size of a dose. With rare exception, on a day to day basis, the patient is the person who makes this decision. While the physician may be best qualified, he neither has the time himself nor other resource to make this decision for the patient on a day to day basis. Nor is the physician compensated for making this dosing decision other than a periodic fee for an office visit. The patient may or may not be motivated to make a dosing decision, and he is usually the least skilled or qualified. Methods to help the patient decide on a proper dose include counting carbohydrates and making calculations based on measurements of blood glucose. For the large part, though, these calculations are difficult to make, error prone, and even when done properly do not provide sufficiently accurate control of blood glucose levels. Proper dosing of insulin for a person with diabetes to adequately control their blood glucose is far from routine and compliance to any regimen is poor.
Over the past several years an approach generally called “Treat to Target” has been studied for use primarily in persons with Type II or non-insulin dependent diabetes. An example of such a study is found in Alvarsson, M et. al. “Beneficial Effects of Insulin versus sulphonylurea on insulin secretion and metabolic control in recently diagnosed Type II Diabetic Patients”, Diabetes Care 26, 2231-2237, 2003.
In one common version of this method a person measures their fasting blood glucose in the morning and injects a dose of long acting insulin, typically Lantus (Hoechst-Aventis) or Levemir (Novo) based on that glucose measurement. The doses of insulin are increased periodically until the fasting blood glucose reaches a target, typically 100 mgm % (5.5 mMolar). Once that target is reached, the dose is then held constant until there is some further change in the fasting blood glucose level. Large, multi-center clinical studies have shown that this method is very effective in reducing fasting blood glucose, and does so more quickly and safely compared to other therapeutic regimens (for example, the combination of a sulphonylurea and metformin). Further, the “Treat to Target” method, when used in newly diagnosed persons with diabetes and who are treatment naïve, restores a large fraction of the person's normal insulin secretion capability (see: The effect of intensive insulin therapy on pancreatic beta cell function in newly diagnosed Type 2 diabetic patients—Poster, Annual American Diabetes Association Meeting, Chicago, 2007). In studies of this method, though, the interaction between the person with diabetes and a healthcare professional is intense with an interaction at least once a week and often twice a week to encourage compliance and to monitor use of the protocol. To treat the tens of millions of person with Type II diabetes in the world today, there is insufficient manpower to implement a Treat to Target method. Hence there is a clear need for a method of implementing a Treat to Target protocol that does not require intensive interaction with a healthcare professional and does not require a high level of skill from the person with diabetes.
In one aspect of the invention a method of implementing a Treat to Target protocol for managing blood glucose levels that requires minimal time of the healthcare professional and a minimum of skill from the person with diabetes or the person caring for the person with diabetes is provided. The only requirement of the healthcare professional is the diagnosis of diabetes and the providing of a prescription for insulin. The only requirement of the person with diabetes or the person caring for the person with diabetes is the ability to obtain a measurement of glycemic control when requested and to enter that measurement value, either by using a keyboard, or by saying the glucose value (to be interpreted by a voice recognition and verification system), or by using a cable or wireless transmission from the glucose meter. The measurement of glycemic control may be a fasting blood glucose value, or a fasting measurement of glucose concentration in interstitial fluid as is now performed by several continuous glucose monitoring systems such as the Navigator™ manufactured by Abbott Laboratories. The measurement of glycemic control may also be a fructosamine value, a hemoglobin A1C value, or any value shown to adequately represent the level of glycemic control in the person with diabetes who will receive the insulin dose calculated by this method. In this “Treat to Target” method the measurement of glycemic control is provided to a protocol manager for calculation of an appropriate insulin dose. This protocol manager may be one or more persons or may be completely under the control of software on a computer. The protocol manager uses a Treat to Target protocol and the measurements of glycemic control, which include a recent measurement taken since the last calculated dose of insulin and may include measurements taken on earlier occasions, to calculate a dose of insulin for the person to administer. This insulin dose may be administered one or more times per day depending on the details of the “Treat to target” protocol.
In another aspect of the invention an insulin dispensing unit is provided to fill an insulin administration device with the calculated dose of insulin. The insulin administration device may be a syringe or an insulin pen as is known in the art, or an insulin pump such as the Medtronic Paradigm operating in a “basal only” mode, or an insulin administration patch as described in copending application 60/931,758 dated May 29, 2007 entitled Passive Drug Delivery system, the contents of which are incorporated herein in their entirely by reference. In one aspect of the invention the filling device comprises a container of insulin and a container of diluent. In other aspects of the invention the filling device may be an insulin cartridge with a fixed insulin concentration. Depending on the calculated dose of insulin, varying amounts from each of these containers will be delivered to the insulin administration device such that the volume of fluid put into the insulin administration device has the desired number of units of insulin.
In another aspect of the invention this insulin dispensing device further comprises a communication link from the user to the protocol manager. This communication link may be a land line telephone or a cell phone or an internet connection or any other link capable of accurately communicating the measurement of glycemic control to the protocol manager and to communicate the calculated dose of insulin back from the protocol manager to the insulin dispensing device such that the insulin dispensing device can fill the insulin administration device with the calculated dose of insulin. Alternatively, the protocol manager may comprise computer software that is resident in the insulin dispensing device.
In a further aspect of the invention the insulin dispensing device keeps a detailed record of the amount of insulin dispensed which may or may not be communicated to the protocol manager. By keeping the detailed record of dispensed insulin, the insulin dispensing device is capable of calculating when there is insufficient insulin remaining in the insulin container to provide the calculated dose. In this event, the insulin dispensing device makes the user aware of this lack. The insulin dispensing device may fill the insulin administration device with a partial dose and, when the insulin dispensing device has been resupplied with a new container of insulin, finish filling the insulin administration device, or the insulin dispensing device may simply wait to fill the insulin administration device after the insulin dispensing device has been resupplied with a fresh container of concentrated insulin. Similarly the dispensing device can keep track of the use of the diluent to insure that sufficient diluent is available to fill an administration device and to determine when fresh diluent must be provided. To reduce the chances of infection, each insulin and diluent container system is closed so that the insulin administration device is aseptically filled. Since the protocol manager has a complete record of the doses of insulin that the user has administered, it may also automatically order additional insulin and diluent for the user such that these supplies arrive as they are needed by the user.
In this embodiment the glycemic measurements are sent to an entity herein called the protocol manager. The protocol manager may be a person or group of persons, or may be a device such as a computer. This protocol manager is capable of receiving the glycemic measurements and, in one embodiment of the invention, using the glycemic measurements to calculate an appropriate dose of insulin for administration to the patient. The calculated dose may be a basal dose to be administered at a constant rate over a set period of time, for example over the course of an hour, or day, or week. Alternatively, the calculated dose to be administered may be a dose to be given over a very short period of time, for example, a bolus dose of insulin. Or, the dose may be a combination of a basal dose and a bolus dose such that the amount and time of administered insulin has been determined. In this latter case the calculated dose is clearly a dosing regimen, that is, amounts of insulin administered as a function of time.
The calculated dosing information is then sent to a dispensing device. The patient places an administration device such as a patch as described later, or an insulin syringe or an insulin pen. The dispenser then fills the administration device with the appropriate dose as calculated by the protocol manager. As shown in
Insulin dispensing device 100 further comprises a display and keyboard 13 for displaying instructions and entering information. As shown, insulin dispensing device 100 is requesting a glycemic control value which in this case is a glucose reading. Typically this glucose reading will be obtained using a glucose meter and strip as is well known in the art. This glucose reading is typically a fasting glucose reading obtained first thing in the morning but could be a glucose reading obtained at any other time as required by the details of the Treat to Target protocol as implemented by the protocol manager. Alternatively the glycemic control value could be a fructosamine value or a hemoglobin A1C reading or any other value required by the Treat to Target protocol.
Insulin dispensing device 100 further comprises button 14 which may be used to initiate operation of insulin dispensing device 100, and button 15 which may be used to select menu options to be displayed on display 13.
Insulin patch 200 further comprises activation button 22. After the adhesive release liner has been removed, the device is place on the skin and activation button 22 is pressed. Pressing the button causes a thin needle (not shown) to be deployed into the skin and causes the insulin dose to start flowing into the skin.
In one embodiment of the invention the user follows the flow chart shown in
In
Since the protocol manager has a record of the doses administered by the user and the amount of medication contained in the medication cartridge when it was first used, the protocol manager can estimate when the user is likely to need a new medication cartridge. In this case the protocol manager can notify an appropriate organization, such as the shipping portion of the company providing the treat to target system, that it is time to resupply the user. The shipping portion of the organization can then fill the resupply order and ship fresh medication cartridges to the user as necessary.
After the patch has been filled, the patch is presented back to the user (Box just below dotted line box A in
The process shown in
In the above-case the insulin administration patch was capable of delivering insulin at a constant rate over the life of the patch. In this case the insulin which would be placed in the patch would be a relatively rapid acting insulin such as regular insulin, or even more rapidly acting insulin such as Novolog (Novo) or Humalog (Lilly) or Apidra (Aventis).
The insulin dispensing device then fills the syringe with the calculated dose of appropriate insulin (bottom box in dotted line box B in
A third embodiment of the invention is shown in
This application claims subject matter disclosed in copending application No. 60/931,758 dated May 29, 2007 entitled Passive Drug Delivery system, the contents of which are incorporated herein in their entirely by reference.
Number | Date | Country | |
---|---|---|---|
60931758 | May 2007 | US |