This application claims priority to Greek patent application no. 20230101087, filed Dec. 29, 2023, which is incorporated herein by reference in its entirety.
The present disclosure relates to hydrogels suitable for use as conductive compositions, methods of making these compositions, and the use of these compositions with medical electrodes, standalone transmitters or the like.
Epidermal electronics have emerged as non-invasive, compact, and ergonomic tools to monitor an individual's health status in resting and during dynamic conditions such as sweating or exercising [1]. The commercial electrophysiology devices are typically metal electrodes such as silver/silver chloride (Ag/AgCl), which although highly conductive, are flat and rigid hence not adequately conforming to human skin. This mismatch often leads to skin irritation, electrode degradation and sweat-induced failure, all of which hampers signal quality in the long term [2, 3]. Furthermore, those electrodes have a finite shelf-life [4] due to the gradual conversion of Ag to AgCl over time, even during storage, severely impacting the stability and reliability of the electrodes.
Conducting hydrogels combine the soft and stretchable nature, water-absorption and retention properties of hydrogels with the conductivity typically provided by conducting polymers or carbon-based materials, making them highly suitable for advanced and comfortable wearable devices [5]. These hydrogels possess inherent adhesiveness, superior skin conformability, and exhibit mixed conduction (both ionic and electronic), along with remarkable mechanical properties, surpassing the performance of conventional metal electrodes. As such, they serve as a more biomimetic alternative for sensing, stimulating, and modulating electrical activity in biological tissues [6]. Furthermore, conducting hydrogels conform to the skin reducing the gap between the electrode and the skin, minimizing skin-electrode contact impedance hence enhancing the signal-to-noise ratio (SNR) of electrophysiological measurements [7, 8].
However, several aspects need to be addressed for clinical and commercial translation to fully realize the potential of conducting hydrogels. The main challenge is optimizing the electrical conductivity to be comparable with metal electrodes while ensuring long-term stability and biocompatibility. To render hydrogels electrically conducting, standard methods include doping them with ionic liquids, metallic nanoparticles, carbon-based conductive fillers or conducting polymers [9-11]. Furthermore, developing scalable fabrication methods while keeping the cost of materials and processes as low as possible is another critical requirement [12].
Hydrogels can be synthesized using various methods, such as polymerization [13], crosslinking [11, 14], solution mixing [11, 15], radiation [16, 22], etc. They can be made from synthetic materials, including polyvinyl alcohol, polyacrylamide and polyethylene glycol [1, 5, 13], natural materials including gelatin, chitosan and alginate [17-21], or hybrids [17, 22, 23].
While synthetic hydrogels might seem superior in terms of mechanical properties they often suffer in terms of biocompatibility, biodegradability, and sustainability. Natural hydrogels are arguably more desirable especially given their high resemblance to the skin, low toxicity, sustainability and facile processing [17, 24]. However, they still need the addition of chemical crosslinkers, often toxic, as they tend to degrade in water and suffer in terms of mechanical robustness and stability.
There is thus a need for new eco-friendly, natural and biocompatible hydrogels, and a method for making the same, that can be used, for instance, for the manufacturing of medical electrodes (e.g., ECG, EMG, EEG) or as a standalone transmitter for various signals.
The shortcomings of the prior art are generally mitigated by a new system and method for pre-heating an electrolytic cell typically used for the electrolytic production of aluminum.
According to a first aspect, it is disclosed a thermoreversible electrically and/or ionically conducting hydrogel comprising: a mixture of glycerol, chitosan and gelatin; a given amount of an electrically conducting material, and an aqueous solvent; wherein the conducting hydrogel is free of crosslinking agent.
According to a preferred embodiment, for 12 mL of said aqueous solvent, such as water, the hydrogel comprises:
According to a preferred embodiment, for 12 mL of said aqueous solvent, such as water, the hydrogel comprises:
More preferably, wherein for 12 mL of said aqueous solvent, such as water, the hydrogel comprises:
According to a preferred embodiment, said conducting material comprises poly(3,4-ethylenedioxythiophene) polystyrene sulfonate (PEDOT:PSS), thermally reduced graphene (TRG), or Ti3C2Tx (MXene).
According to a preferred embodiment, the hydrogel may be for use for the manufacturing of a medical electrode, a standalone transmitter for various signals, and/or wearable electronics for recording/stimulation. Preferably, the medical electrode or standalone transmitter is manufactured using 3D printing, casting or any sorts of molding techniques. More preferably, the medical electrode is used for electrophysiological signal comprising, but not limited to, electrocardiogramalectromyography (EMG), electroencephalography (EEG), electrogastrography (EGG) or electrooculography (EOG).
According to another aspect, it is disclosed a process for the making of the electrically and/or ionically conducting hydrogel as disclosed herein, the process being a “one-pot” process and comprising the steps of:
According to a preferred embodiment, the temperature of the aqueous solvent is about 50-60°, more preferably about 56° C. (i.e. hot plate setting).
According to a preferred embodiment, the aqueous solvent has a pH between about 4 and 6.5.
According to a third aspect, it is disclosed a medical electrode comprising:
According to a preferred embodiment, the shape of the hydrogel is obtained by molding said hydrogel in its sol state at a given temperature, the hydrogel being then cooled until jellification/stretchable (Gel state).
Preferably, the shape of the hydrogel may be obtained by 3D printing said hydrogel in its sol state at a given temperature.
According to a preferred embodiment, the medical electrode as disclosed herein is for use for electrophysiological signal comprising, but not limited to, electrocardiography (ECG), electromyography (EMG), electroencephalography (EEG), electrogastrography (EGG) or electrooculography (EOG).
The resulting conducting hydrogels exhibit remarkable robustness, do not require crosslinking agents, and possess a unique thermo-reversible property, simplifying the fabrication process and ensuring enhanced long-term stability. Moreover, their fabrication is sustainable, employing environmental-friendly materials and processes, while retaining their skin-friendly characteristics.
It is also disclosed hydrogel electrodes that were tested for ECG signal acquisition and outperformed the commercial electrodes. The hydrogel-based electrodes deliver high-quality ECG signals, boasting a superior signal-to-noise ratio (SNR) and remarkable resilience against interference and motion artifacts, compared to their commercial counterparts.
Other advantages of the invention will be disclosed herein after.
The above and other aspects, features and advantages of the invention will become more readily apparent from the following description, reference being made to the accompanying drawings in which:
A novel hydrogel, method of making and their use as medical electrodes or the like will be described hereinafter. Although the invention is described in terms of specific illustrative embodiments, it is to be understood that the embodiments described herein are by way of example only and that the scope of the invention is not intended to be limited thereby.
The terminology used herein is in accordance with definitions set out below.
As used herein % or wt. % means weight % unless otherwise indicated. When used herein % refers to weight % as compared to the total weight percent of the phase or composition that is being discussed.
By “about”, it is meant that the value of data disclosed herein can vary within a certain range depending on the margin of error of the method or device used to evaluate or measure such data. A margin of ±10% is generally accepted.
The description which follows, and the embodiments described therein are provided by way of illustration of an example of particular embodiments of principles and aspects of the present invention. These examples are provided for the purposes of explanation and not of limitation, of those principles of the invention. In the description that follows, like parts and/or steps are marked throughout the specification and the drawing with the same respective reference numerals.
As aforesaid, it is first disclosed a hydrogel comprising gelatin (or hydrolyzed collagen), chitosan (polysaccharide processed from chitin) and glycerol.
Gelatin, that can be represented by the Formula I below, is a form of hydrolyzed collagen, a protein that constitutes the majority of the human skin composition.
Being both biocompatible and biodegradable, cheap and relatively highly abundant, gelatin has been widely used in several biological applications. For example in wearables, as it possesses a variety of functional groups such as amino (R—NH2), hydroxyl (R—OH) and carboxyl (R—COOH), it facilitates the formation of polyelectrolyte complexes (PECs) and adherence to the skin [21]. Several studies have hence focused their attention on gelatin-based hydrogel electronics, endowing the hydrogel with electrical conductivity for example by sandwiching it with 2D materials such as polypyrrole, and reduced graphene oxide [17].
Chitosan, that can be represented by the Formula Il below, is the second-most abundant polysaccharide [26].
Chitosan is a derivative of chitin, easily obtained from chitin, which is very similar in the chemical structure, and chitosan is bio-adhesive. Similarly, to gelatin, chitosan is also biocompatible and biodegradable, exhibits ionic conductivity, and the ability to form intermolecular interactions such as PECs due to its functional groups [18, 19, 27].
Finally, glycerol, is a triol that is represented by Formula Ill below.
Glycerol is a natural plasticizer employed in several hydrogel electrodes [5, 28]. For example, increasing the ratio of bound water molecules against the free water molecules using its hydroxyl groups has shown to promote hydrogen bonding [28], thereby reducing the gelation time, making the resulting hydrogel more elastic and robust with non-drying and anti-freezing behavior [5, 16, 28].
The hydrogels according to the present invention combine gelatin, chitosan and glycerol, benefiting from their synergistic properties through a one-pot fabrication approach as shown in
Golde, the resulting hydrogel, depicted in
It is believed that the combination of chitosan, being a cationic polysaccharide, and gelatin, being an amphoteric protein, at certain pH, enables the formation of electrostatically entangled PECs forming polymeric architectures with relatively high tensile strength for a natural hydrogel.
The hydrogel according to the present invention demonstrates several attractive features: robustness, highly stretchable, inherently adhesive and thermo-reversible. Thermo-reversibility facilitates the repair of the hydrogel upon drying or cracking, which makes it reusable (see
To enhance its conducting properties, three conducting materials have been tested as shown in
All the three resulting hydrogel electrodes are robust, with mechanical strength that outperforms state-of-the-art natural and hybrid-based hydrogels (SI-Table 1 for comparison) and can conform to the skin, with the potential for high-quality ECG recordings. Owing to the inherent adhesiveness of the hydrogel as demonstrated in
According to a preferred embodiment, three natural hydrogel materials have been combined to create a robust base hydrogel, namely Golde, which refrains from the use of chemical crosslinkers while featuring tissue-like mechanical properties and thermo-reversibility, which is important for synthesis and post-synthesis modification. Typically, mechanical stability and integrity are imparted via crosslinking i.e., formation of chemical or physical bonds between the polymer chains within a hydrogel network. Crosslinking creates a three-dimensional network that prevents the dissolution or disintegration of the hydrogel in an aqueous environment allowing the hydrogel to maintain its shape, strength, withstand mechanical forces and maintain its functionality. Here, chitosan creates a PEC (polyelectrolyte complex) with gelatin, resulting in a tough natural hydrogel with a relatively high tensile strength without the addition of any external crosslinking agent. In addition, no mold or fouling was observed in the hydrogel samples after 3 months of storage in ambient room conditions, possibly attributed to chitosan's antimicrobial properties [26]. Furthermore, the integration of glycerol, is found to improve the hydrogel's elasticity while inhibiting its dehydration by acting as a plasticizer and a humectant
According to a preferred embodiment, the hydrogel as disclosed herein is made electrically conducting and its electrode impedance is lowered by blending the base material (Golde) with a conducting material. Three different conducting materials have been tested: (Poly(3,4-ethylenedioxythiophene) polystyrene sulfonate) also known as PEDOT:PSS, MXenes and thermally reduced graphene (TRG). All three materials were processed from water suspensions and mixed directly with Golde solution (as shown in
PEDOT:PSS is a conducting polymer that has been widely used in wearable electronics [30], typically interacts with other charged polymers, chitosan herein, creating complexes that may improve the overall hydrogel properties.
MXene, is a composite of a metallic carbonitride, in the present case titanium carbide, in the following notation (Ti3C2Tx) where T represents the surface functional group (—OH, —F, —O). MXene (Ti3C2Tx), exhibits metallic conductivity as well as tunable surface functional groups and negatively charged hydrophilic surface enabling strong interactions with polymer networks such as electrostatic and hydrogen bonds [32].
Thermally reduced graphene, or TRG, features a large surface area and alongside improving the conductivity, its (nano) structure could act as mechanical reinforcement to the hydrogel providing stability and conducting paths through interpolymeric interactions [17].
The surface morphology of the different hydrogels is shown in
XRD data are shown in
Prior to testing the mechanical performance of the hydrogels, a swelling test (see
Table A presents a comparison of the hydrogel electrodes according to the present invention (4 last lines in bold) and other recent relevant works for electrophysiological acquisition in terms of tensile strength and elongation at break with some using chemical or physical cross-linking agent. Other works in the literature are either not tested using the same method or not designed for our application. All materials abbreviations are shown below.
Abbreviations: Poly-vinyl alcohol (PVA), Polypyrrole (Ppy), reduced graphene oxide (rGO), Acrylic acid (AA), (PNIPAM), Acrylamide (Am), Polyacrylamide (PAAm), lauryl methacrylate (LMA), Lithium salt (LS), Thermally reduced graphene (TRG), N,N′-methylenediacrylamide (MBA, MBAm), Guar gum (GG) and phenylboronic acid grafted sodium alginate (Alg-PBA), Gelatin (Gel), Chitosan (CS), silver nanoparticles (AgNPs), Carboxymethyl cellulose (CMC), polyethylene-3,4-dioxythiophene: sodium polystyrene sulfonate (PEDOT:PSS).
As shown in equation 1, the reactance Xc is inversely proportional to the frequency (f) of the current passing through the material with capacitance (C).
Another contributing element is the frequency independent real element (resistance, R) which is a scalar. As shown by equation 2, the total impedance (Ztotal) is the vector sum of its resistance and reactance, as frequency decreases the reactance contribution becomes the dominating element [34]. Therefore at the highest frequencies the impedance is stable.
The EIS (Electrochemical impedance spectroscopy) and CV (Cyclic Voltammetry) curves of the electrodes are shown in
The stability of the hydrogel electrodes (in terms of their electronic properties) was evaluated by monitoring the change in gain over frequency (
The hydrogel electrodes' low electrode impedance (see
The long-term stability of the electrodes was tested after 3 months of storage under ambient room conditions with a decrease in SNR of 18.2% compared to a freshly prepared sample (see
To further evaluate the “standalone” performance and skin-adhering properties of the hydrogel electrodes, G-PEDOT:PSS electrode was fabricated as a proof of concept on an all-flexible electrode setup using copper tape for the connection, as shown in
It has been developed a self-adhering, and reusable conducting hydrogel that does not require any crosslinker, using a simple one-pot synthesis method. The hydrogel is composed of a natural biopolymer blend (gelatin, chitosan and glycerol) and is infused with three types of electrically conducting materials: PEDOT:PSS, MXene and TRG. The resulting hydrogel electrodes namely Golde (biopolymer blend), G-PEDOT:PSS, G-MXene and G-TRG were characterized according to their structure, surface composition, and mechanical and electrical performance.
The mechanical properties of the hydrogels match those of the skin (0.5-500 kPa) [25] while exhibiting stretchability and robustness, outperforming similar natural hydrogels in literature. Morphological and structural analysis revealed favorable interactions between the conducting materials and the biopolymer network, further corroborated by the improvement in the mechanical and electrical properties of the conducting material-infused hydrogels compared to the base material, Golde.
The hydrogel variants as disclosed herein were used as an electrode alternative to commercial Ag/AgCl for electrophysiology recording, herein ECG. The recorded SNR of the hydrogel electrodes was considerably higher than that of the commercial ones and showed no skin irritation and pain-free removal after 30 mins of ECG recording. Furthermore, benefitting from the inherent adhesiveness of the hydrogel an all-flexible electrode setup was developed using copper tape for the connection. The ECG recordings showed less noise and motion artifact susceptibility compared to the commercially available Ag/AgCl. Moreover, although other variants might have shown better performance in some aspects, G-PEDOT:PSS variant was overall the best taking into account the cost and ease in processability alongside SNR performance and skin friendliness.
The hydrogels as disclosed herein outperforms the commercial one; hence, it could be easily integrated into wearable setups, possibly enhanced with the augmentation of Al for advanced health monitoring and analytics [38].
Although herein demonstrated for ECG, it Is believed that other types of measurements would show similarly superior performance. Further, the presence of glycerol in the base material and the low sol-gel (glassy state) temperature i.e., thermo-reversibility of the hydrogel, give the hydrogel a thixotropic nature rendering it a potentially good candidate for extrusion-based hydrogel 3D printing or other applications.
Gelatin (Bovine Skin, Type B, G9391, USA), Chitosan (high molecular weight, deacetylated chitin, 419419, Ireland), Acetic acid 100% and Glycerol 99%, Ethylene glycol (EG), 4-Dodecylbenzenesulfonic acid (DBSA) and 3-glycidyloxypropyl) trimethoxysilane (GOPS) were obtained from Sigma-Aldrich Co. Aqueous PEDOT/PSS dispersion (CLEVIOS PH 1000™, 1L, Germany) from Heraeus GmbH & Co. MAX phase (Ti3C2Al powder, ≤40 μm, Ukraine) was obtained from Carbon-Ukraine Ltd. Graphite powder (Sigma-Aldrich, 10 mesh), sulfuric acid (Sigma-Aldrich, ACS reagent, 95.0-98.0%), hydrochloric acid (Sigma-Aldrich, ACS reagent, 37%), Lithium Fluoride salt (Sigma-Aldrich), potassium permanganate (Fischer Scientific, C99%), hydrogen peroxide (Sigma-Aldrich, 30 wt % in H2O) were used. Graphite oxide was prepared from natural graphite by using improved synthesis proposed by Tour et al.
Different chitosan (CS) solutions have been prepared as indicated in Table 1 below:
More preferably, the chitosan solution of 0.5 wt % was prepared by measuring 10 mL of distilled water at 85° C. into a 50 mL borosilicate beaker stirred at 250 rpm. 0.05 g of the chitosan powder is added, and 100 μL of Acetic acid (dropwise) is added upon dispersion into the solvent. The solution is stirred for preferably 2 hours until the chitosan dissolves completely. The solution is preferably transparent and thick without visible fibers.
Different hydrogel solutions have been prepared as indicated in Tables 2 and 3 below:
More preferably, following a simple one-pot process, a volume of 12 mL of distilled water was added to a 50 ml beaker and stirred at 160 rpm at no more than 55° C. In this order, 3 mL of glycerol were added followed by 3 mL of the prepared chitosan solution and wait until the solution is mixed thoroughly (not more than 5 min). Afterwards, slowly increase the stirring speed to 350 rpm and add 3.75 g of the gelatin powder to avoid aggregation. After 20 minutes the solution becomes homogenous and transparent with golden/amber color. To prepare larger amounts of the Golde hydrogel solution, all the ratios must be consistent.
5% ethylene glycol (EG), 0.5% 4-Dodecylbenzenesulfonic acid (DBSA), 1% 3-glycidyloxypropyl)trimethoxysilane (GOPS) added to the aqueous PEDOT/PSS dispersion. Using a digital scale in this order EG, DBSA, GOPS were added to the falcon tube. After adding each additive, the solution was sonicated for 5 minutes (I.e., EG, sonicate, DBSA, etc.) Finally, using a syringe filter of 0.8 μm pore size to complete the final mixture that would be used. The GOPS was added to the PEDOT:PSS mix later (just before mixing it with the hydrogel solution) or not at all to avoid having clumps that prevents the PEDOT:PSS from uniformly dispersing within the hydrogel solution.
The MAX phase substance, Ti3C2Al powder, with a size of up to 40 μm, was utilized to produce MXene (Ti3C2Tx) following a carefully refined minimally intensive layer delamination (MILD) process. This technique involves the targeted removal of aluminum through the in-situ application of lithium-fluoride salt (LiF), adopting a top-down strategy. In the initial step, the etching agent dissolves 3.2 grams of LiF in 40 mL of 9 M HCl. Following this, 2 g of the MAX powder was gently introduced to the etching agent and agitated for 24 hours at room temperature. Next, the MAX powder containing the etching agent was purified numerous times with DI water. This was accomplished by performing numerous centrifugation steps (each lasting 6 minutes at a speed of 3500 rpm) and sonication, with the slurry discarded post-centrifugation in every wash cycle. This procedure concluded once the solution attained a pH higher than 6 and observed a dark-green colloidal suspension of singular or few-layered (Ti3C2Tx) flakes. Obtained MXene colloidal solution was then preserved in a freezer at −20° C. to maintain its stability for future application. A free-standing MXene film was produced using vacuum-assisted filtration to process the Ti3C2Tx-rich colloidal solution through a filtration membrane (specifically, Celgard™ 3501 from Celgard LLC, USA). After being air-dried for 24 hours, the MXene (Ti3C2Tx) film was carefully detached from the filtration membrane. The concentration was found for MXene colloidal solution by dividing the weight of the air-dried MXene film by the volume of the colloidal solution. Finally, the dilution was performed to obtain a concentration of (1 mg/mL) of MXene colloidal solution using autoclaved double distilled water (18 MΩ).
TRG is made by using Graphene oxide as a raw material and then thermally reducing it into a TRG powder. Graphene Oxide (GO) was synthesized using the Tour method: 300 mL of con·H2SO4 and 75 mL of orthophosphoric acid were mixed in a 500 ml beaker. Subsequently, 10.0 gm of graphite powder was added to the above solution and stir the mixture for 45 min at 30° C. Then KMnO4 was slowly (58.0g) added to the mixture. The mixture was allowed to stir at 25-30° C. for 72 h. Then 150 ml of 5% H2O2 solution was added to the mixture and stirred for 2 h at 30° C. The content was allowed to settle and centrifuge to remove the superannuated solution. The residue was washed with 1M HCl followed by DI water. The residue was further purified using cellulose membrane for 1 week in DI water medium. The purified residue was dried under the freeze drier, graphite oxide (GO) was obtained as a solid powder.
Thermally reduced graphene (TRG) was prepared by placing GO (300 mg) into a 25-mm i.d., 1.3-m long quartz tube. The opening end of the quartz tube was closed using a rubber stopper. A nitrogen inlet was then inserted through the rubber stopper. The sample was flushed with nitrogen for 15 min, and the quartz tube was quickly inserted into tube furnace, preheated to 1000° C. and held in the furnace for 30 s.
Golde hydrogel solution in sol-state was placed on a hot-plate stirrer at 55° C. and stirred at 500 rpm. At a ratio of 10 v/v %, 0.5 mL of the conducting material solution was added (dropwise) into 5 mL of the Golde solution. The resulting solution was stirred for 10 minutes. At the same temperature, the stirring speed was lowered to 90 rpm to move bubbles to the surface (if any) for 10 minutes.
A 3D mold of the flexible electrode was designed using TinkerCAD™ and 3D printed into a flexible polyurethane mold using ELEEGOT Mars 3 pro 3D printer. The conducting hydrogel electrode was attached to a copper tape having a metal stud to connect to the ADInstruments Power lab™ 26T bio-amplifier, with only the head (active site) of the flexible electrode exposed to the skin, while the tail is covered to achieve a comparable result to the previously tested electrodes. Furthermore, in future tests and real-life applications, the setup would be furtherly improved and the exposed hydrogel area would be maximized to include the entire hydrogel area, thereby improving the signal quality. ADInstruments LabChart application was used to record and analyze the Lead I ECG signal.
To observe the hydrogel surface morphology, SEM imaging was conducted. To assess the available functional groups found throughout the hydrogel, Raman spectroscopy was used instead of FTIR since it is less sensitive to perturbations caused by the water molecules present in the hydrogel samples. Finally, since the material organization (i.e., amorphous, semicrystalline) in this hydrogel composite electrode affects surface morphology and, thereby its interactions with the skin (ex., wettability and contact area) in addition to its mechanical properties, XRD analysis was performed to improve our understanding of this material structure.
Sample preparation for analysis: In its sol-state the hydrogel samples were cast into a 3D printed mold of 6 mm diameter and 1.3 mm of thickness (refer to Figure S3). Next, the hydrogel is placed in a −20° C. freezer for 2 hours. Then the frozen samples are then rapidly moved into a freeze drier and placed there for 24 hours (until the sample is completely freeze-dried). Finally, the samples surfaces are gold coated to make them visible in the SEM scan. To perform the SEM imaging (JEOL JSM-7610F) SEM was used.
Hydrogel sample preparation: As distilled water is the main solvent of the hydrogels, all hydrogel samples were cast into a 3 cm×2 cm mold and soft-baked at 58° C. for around 10 minutes (an estimate based on observation) this is done to reduce the amount of the free water from the hydrogel matrix and in turn stabilizing it. This abovementioned temperature is not mandatory, however we observed that at lower temperatures the hydrogels was taking long time to be ready. At higher temperatures the hydrogels lost water rapidly (less controlled) and because it is believed that it could cause some of the conducting materials to start degrading. Then, they were left to cool down on a tabletop for 5 minutes. The tensile testing of the hydrogels as disclosed herein was performed in ambient room conditions to simulate the conditions that they will be exposed to during their application.
Tensile testing Using the Instron™ 5966 universal testing machine with a 10 kN load cell and an extension rate of 0.30 mm/s and 5N of maximum load, the hydrogels are fixed with a clamp sample holder. A small piece of tissue paper and duct tape were used to firmly attach the sample to the clamp without destroying the samples. The dog-bone-shaped samples were stretched to their breaking point while their stress/strain response was plotted.
Electrochemical impedance spectroscopy (EIS) was carried out using a electrochemical impedance analyzer (AutoLab PGSTAT204, Sweden) with a sine wave with an amplitude of 10 mV and a frequency ranging from 0.1 Hz to 100 kHz with a focus window of 1 Hz to 100 kHz. The hydrogel sample was 16×14 mm and a thickness of 1.5 mm.
Cyclic voltammetry (CV) was performed on the same hydrogel samples at a scan rate of 0.05 and 0.1 V/s and a potential window of (±0.6 V). Specific capacitance was calculated from the CV data.
Gain response setup The gain response vs frequency of both the commercial and the hydrogel electrodes was evaluated using National Instrument tools, namely the NI ELVIS II+ board. The experimental protocol employed a sinusoidal signal of peak amplitude 0.1V and a frequency sweep starting from 1 Hz to-100 kHz as the stimulus.
Swelling ratio the hydrogels were placed in a pettri dish in ambient room conditions with an adequate amount of deionized water to submerge the hydrogels. Afterwards, the change in their weights was recorded using a balance and finally, the swelling ratio was estimated, this was performed in triplets (n=3)/variant over the course of 0.5, 1, 3, 5, 9, 12, 24 and 48 hours.
Adhesive test chicken skin preparation: chicken thighs were de-skinned, the skin was washed with 70% ethanol and dried off its oil and placed on a piece of aluminum foil.
While illustrative and presently preferred embodiments of the invention have been described in detail hereinabove, it is to be understood that the inventive concepts may be otherwise variously embodied and employed and that the appended claims are intended to be construed to include such variations except insofar as limited by the prior art.
Number | Date | Country | Kind |
---|---|---|---|
20230101087 | Dec 2023 | GR | national |