Keratoconus, pellucid marginal degeneration, corneal transplants, and other cornea abnormalities lead to a distorted cornea. This can restrict the visual acuity and quality of vision of the affected individuals. Currently it is estimated that about 1 in 375 people have keratoconus. Combined with the other conditions, 1-2% of the population suffers from degraded vision that is the result of a distorted cornea.
With the advent of wavefront-driven refractive surgery (e.g., LASIK and PRK), for some individuals these imperfections can be measured with a wavefront-sensing aberrometer, and the measurements are used to guide the laser's sculpting of the cornea. While this technique can be quite effective, refractive surgery is limited to those patients that have a sufficiently large corneal thickness to support the required depth of ablation. Unfortunately, it is usually because of a thin or weak cornea that the aberrations develop. S0, the very people that need a wavefront-guided surgical approach the most are ineligible for LASIK or PRK.
It has recently become possible to treat these individuals with a wavefront-customized Contact Lens (CL) using the same approach. This can successfully correct for the corneal distortion, even with quite strong aberrations. For this to be effective, however, the contact lens must be very stable on the eye so that the region on the contact lens with the wavefront-guided correction patch remains aligned with the pupil. This has led to the use of large-diameter, scleral contact lenses for this purpose, since these are generally much more stable than cornea rigid gas permeable (RGP) or soft contact lenses.
By design, a scleral contact lens rests on the sclera of the eye and not on the cornea. Most of these lenses are designed to vault over the cornea so as not to touch it in any way. Thus, the fit of the scleral lens depends not as much on the corneal shape as it does on the sclera's shape. Various technologies have been developed to profile the eye, including the sclera, to determine the scleral shape and, hence, the fit of the contact lens. These include stereophotography, Scheimflug profilometry, Optical Coherence Tomography (OCT), impression profilometry, and 3D imaging.
The sclera is not necessarily concentric with the cornea. Also, a scleral lens is generally tilted and offset with respect to the cornea. The light is collected through the pupil, so a special alignment is required between the location of the wavefront-guided correction patch on the customized contact lens and the cornea.
The tilt of a scleral contact lens is primarily determined by where it rests on the sclera, which in turn depends on the eye's scleral shape. Scleral lenses also usually have a deep sagittal height, e.g., S0=5-6 mm, and so any amount of tilt can result in a significant XY offset between the optical center of the pupil and the physical center of the tilted scleral contact lens. Thus, the measured position of the wavefront-guided correction patch may not be adequately determined using the contact lens' edge, or fiducial marks located near the outer edge of the contact lens. This can result in errors in properly placing the wavefront-guided correction patch since the designed correction patch does not align properly with the actual aberrations of the eye. Thus, determining the tilt and offset of a scleral or corneal contact lens is important to achieving optimal wavefront-guided corrections.
A method is disclosed for correcting aberrations of a patient's eye with a customized contact lens, including determining an optimal XY location and an optimal angular orientation, a, for placing correction optics on the customized contact lens by using an optical instrument to measure XY offsets (ΔX, ΔY) and X, Y, Z tilt angles, (θx, θy, α), respectively, of a predicate contact lens while disposed on the patient's eye. The predicate contact lens and the customized contact lens can be a scleral contact lens or a corneal contact lens. The optical instrument can comprise a wavefront aberrometer and/or a corneal topographer. The correction optics include a wavefront-customized contact lens with a built-in, optimized wavefront-guided correction patch. The predicate contact lens can include three or more fiducial marks inside of the pupil, which can be used to determine the geometric center of the predicate contact lens in the XY plane when placed on the patient's eye.
An alternative method (i.e., method B) can also be used that does not require the use of fiducial marks on the predicate contact lens.
A contact lens' physical center is typically different (i.e., offset in the XY plane) from the center of the pupil. Typical amounts of XY offset are less than or equal to about 1 mm. The XY offset (ΔX, ΔY) of the contact lens' physical center from the pupil's optical center is measured. Then, the center of the wavefront-guided correction patch is moved in the XY plane to match the location of the measured optical center (i.e., pupil center). This process works for both wavefront-guided customized contact lenses and for regular (non-customized) contact lenses, as the improved centering technique helps regular contact lenses (even without using wavefront-guided correction techniques). The process can also be applied to both scleral and normal contact lenses.
In an embodiment, a method for correcting one or more aberrations of a patient's eye with a customized contact lens includes determining an optimal XY location and an optimal angular orientation, a, for placing correction optics on the customized contact lens by using an optical instrument to measure XY offsets (ΔX, ΔY) and X, Y, Z tilt angles, (θx, θy, θz), respectively, of a predicate contact lens while disposed on the patient's eye. The optical instrument can be a wavefront aberrometer, a slit lamp camera, or other instrument that can take an image of the eye. Alternatively, the optical instrument can be a corneal topographer, an anterior OCT or a Scheimflug imager.
In an example, the correction optics can include a wavefront-customized contact lens with a built-in wavefront-guided correction patch that is placed at the optimal XY location and the optimal angular orientation of the customized contact lens. Both the predicate contact lens and the wavefront-guided customized contact lens can be scleral contact lenses or corneal contact lenses (e.g., rigid contact lenses). Alternatively, the predicate lens can be a patient's habitual scleral lens.
In some embodiments, the predicate contact lens can include two or more fiducial marks arranged in a known geometric pattern. In general, the fiducial marks can be located at the edge of, or are located inside of, the patient's pupil. In one example, three fiducial marks can be geometrically arranged in a 45/45/90 right isosceles triangular pattern.
The total number of fiducial marks on a predicate CL can comprise 2, 3, 4, 5, 6, 7, 8, 9, or 10 fiducial marks, arranged in known (pre-defined) symmetrical or asymmetrical patterns about the X- and Y-axes.
In an example, where the fiducial marks are arranged in an isosceles triangle, the method can include calculating a real center, (Xc, Yc), of the predicate contact lens by averaging X- and Y-coordinates of the three or more fiducial marks. The real CL center (Xc, Yc) is located at the center of a 45/45/90 right isosceles triangle's hypotenuse and can be easily obtained from the average position of two corner marks.
Alternatively, the method can include calculating a real center, (Xc, Yc), of the predicate contact lens by: (1) providing a Radius of Curvature=R, Sagittal height=S0, and Diameter=d of the predicate contact lens; (2) measuring XY coordinates (Xv, Yv) of a Corneal Vertex Normal by using Corneal Topography (CT) and/or visual Eye Imaging (EI) techniques; (3) determining a virtual center (Xe, Ye) of the predicate CL by using the diameter, d, and by fitting an edge of the predicate contact lens to a circle or ellipse; (4) calculating X- and Y-tilt angles (θx, θy) and obtaining a real contact lens center (Xc, Yc), by using equations (7) and (8), as follows:
and (5) calculating the physical center (Xc, Yc) of the predicate contact lens by using equations (9) and (10), as follows:
X
c
=X
e
+S
0 sin θx (Eq. 9)
and
Y
c
=Y
e
+S
0 sin θy (Eq. 10);
wherein no fiducial marks are placed on the predicate contact lens.
Additionally, the method can include measuring Corneal Vertex coordinates (Xv, Yv) by using Purkinje reflections from the patient's cornea. The method can further include calculating the pair of XY offsets (ΔX, ΔY) by using equations (11, 12, 13, and 14), as follows:
In an embodiment, the predicate contact lens' tilt is measured by determining an optical axis normal to an optical Z-axis of the optical instrument. Additionally, the optical Z-axis normal can be determined by measuring Purkinje reflections from light sources projected onto the contact lens. The optical Z-axis normal position can be corrected for misalignment of the eye when looking into the optical instrument. The optical Z-axis normal position can be found by projecting light through an objective front lens of the optical instrument. The optimum location of the correction optics can be measured by using a projected pattern of fiducial marks disposed on the predicate contact lens.
In one example, a method of correcting aberrations of a patient's eye with a customized contact lens can include: (a) placing a predicate contact lens on the patient's eye; (b) measuring a pair of XY offsets (ΔX, ΔY) of a center of the predicate contact lens, while sitting on the patient's eye; (c) measuring a rotation angle, a, of the predicate contact lens, while sitting on the patient's eye; (d) determining an optimal XY location and an optimal angular orientation of a wavefront-guided correction patch on the customized contact lens by using the measured pair of XY offsets (ΔX, ΔY) and the measured rotation angle, a, of the predicate contact lens; (e) defining a center of the wavefront-guided correction patch; and (f) adjusting a placement of the wavefront-guided correction patch on the customized contact lens by: (1) placing the center of the wavefront-guided correction patch at the optimal XY location by X- and Y-distances equal to the measured ΔX and ΔY offsets, respectively, and by (2) rotating the wavefront-guided correction patch by the rotation angle, a, to the optimal angular orientation.
In a different example, residual Higher Order Aberrations (HOAs) of a patient's eye are minimized by fabricating and using an optimized wavefront-customized contact lens, wherein the method includes: (α) measuring one or more residual HOAs of a patient's eye with an optical instrument; (b) designing a wavefront-guided correction patch for a wavefront-customized contact lens by using the measured HOAs of the patient's eye; (c) placing a predicate contact lens on the patient's eye; (d) measuring a pair of XY offsets (ΔX, ΔY) of a center (Xc, Yc) and a rotation angle, α, of the predicate contact lens, while sitting on the patient's eye; (e) determining an optimal XY location and an optimal angular orientation of the wavefront-guided correction patch on the wavefront-customized contact lens by using the measured pair of XY offsets (ΔX, ΔY) and the measured rotation angle, α, of the predicate contact lens; (f) defining a center of the wavefront-guided correction patch; (g) adjusting a placement of the wavefront-guided correction patch on a contact lens by: (1) placing the center of the standard wavefront-guided correction patch at the optimal XY location using the measured pair of ΔX and ΔY offsets, and (2) rotating the wavefront-guided correction patch to the optimal angular orientation by using the rotation angle, α; (h) fabricating an optimized wavefront-customized contact lens by positioning the wavefront-guided correction patch at the optimal XY location and the optimal angular orientation; and (i) minimizing the HOA's of the patient's eye by using the optimized wavefront-customized contact lens on the patient's eye.
In an embodiment, a predicate scleral contact lens can include two or more fiducial marks disposed on the predicate scleral contact lens. In an alternative embodiment, the predicate lens is a traditional scleral lens that was designed with only sphere and cylinder correction.
Disclosed herein is a method and system for determining the tilt(s) and XY offsets of a contact lens in order to determine the correct optical centration and rotation for manufacturing and fabricating wavefront-guided correction optics. There are several alternative embodiments that can be effective for this process.
The abbreviation “CL” means “Contact Lens”. The abbreviation “EI” means “Eye Image”. The abbreviation “CT” means “Corneal Topography”. The term “Sag” means the sagittal height, S0, of a contact lens. The terms “mis-centered” and “de-centered” and “offset” are all interchangeable, and they all refer to an ΔX and ΔY offset of the contact lens in the XY plane from an eye's pupil center. The word “rotation” refers to a rotational angle, α, of the contact lens around the Z-axis. The word “tilt” refers to the X- and Y-tilt angles (θx and θy) of the contact lens around the X-axis and the Y-axis, respectively. A positive rotation is counter-clockwise, and a negative rotation is in the clockwise direction. The term “optimal location”, as it refers to placement of a wavefront-guided correction patch on a wavefront-customized contact lens, means a calculated location where the Higher Order Aberrations (HOAs) (e.g., residual aberrations) of the patient's eye, when wearing a wavefront-customized contact lens, are minimized when the wavefront-guided correction patch is placed at the optimal XY location and to the optimal angular orientation. The word “normal”, as it refers to a corneal or CL vertex, means a direction (e.g., vector) that is perpendicular to a tangent surface at a predefined point on the surface.
“The word “mismatched”, as it refers to the position and rotation of a contact lens on an eye, means that the contact lens is misaligned with respect to the pupil of the eye. Therefore, a “mismatched” contact lens can have at least three possibilities: (1) a non-zero XY offset from the pupil's center, (2) a non-zero rotation angle, α, wherein the vertical meridian of the CL doesn't align with the vertical Y-axis of the pupil, and (3) both a non-zero X and/or non-zero Y offset from the pupil's center and a non-zero rotation angle, α, from the pupil's vertical Y-axis. Hence, a perfectly aligned contact lens would have a zero XY offset and a zero rotation angle, α.
The term “aberrations” includes both Higher Order Aberrations (HOAs) and Lower Order Aberrations (LOAs) of a patient's eye. Standard predicate CLs can correct many LOAs, but not all aberrations. A wavefront-customized contact lens is designed to correct all aberrations, including both LOAs and HOAs. The word “normal” as it refers to a normal corneal contact lens includes both soft corneal contact lenses and rigid gas permeable (RGP) corneal contact lens. Scleral contact lenses are not normal contact lenses. Scleral lenses have a much larger diameter and significantly greater sagittal height, S0, than normal corneal contact lenses. The words “correction patch” means a 2-dimensional wavefront map that is used to correct imperfections (i.e., aberrations) of an individual's eye.
In some embodiments, a corneal topographer instrument can be combined with an aberrometer instrument to make a combined topographer/aberrometer instrument.
A first embodiment of a method of use (i.e., Method A) is to locate one or more fiducial marks 18 on a predicate contact lens 16 in locations away from the edge of the predicate contact lens 16 and nearer to the CL's physical center 14. This will minimize the impact of contact lens tilt, and the real CL center 14 can be measured from the predefined fiducial marks 18.
Method A provides a straightforward approach to identify the real physical CL center (Xc, Yc) and calculate the XY offset (ΔX, ΔY) and rotation angle α of a contact lens that is displaced away from the optical center of the eye (XP, YP). This determination allows an optical diagnostician to properly place the correcting optics (e.g., a wavefront-guided correction patch) on contact lens 16 to make a wavefront-customized contact lens (not shown). Fiducial marks 18 may be positioned in a region where they might overlap with the pupil and thereby degrade the optical quality of the optical zone. Even if they could be placed outside the pupil, to be centrally placed they are, by necessity, located in front of the iris. Iris structure(s) then makes these more difficult to see and potentially creates difficulties with image processing. Hence, in some embodiments, the positions of fiducial marks 18 can be located on the edge of the pupil 10, or inside of the pupil 10 (See, for example,
Method B comprises measuring a tilt and offset of the contact lens directly with an aberrometer optical instrument. If a location on the contact lens that is normal to the optical Z-axis of the aberrometer can be determined, then the real CL center (Xc, Yc) can be obtained by quantifying the tilt of the contact lens. It starts with measuring the virtual CL center (Xe, Ye) from the image processing technique and lens diameter and edge information, by using the following equations, as follows:
X
e=(Xedge1+Xedge2)/2 (Eq. 1)
and
Y
e=(Yedge1+Yedge2)/2 (Eq. 2),
where Xedge1, Xedge2, Yedge1, and Yedge2 are measured directly from a visual image of the eye using edge detection software algorithms.
The surface S(x,y) of a spherical contact lens, especially inside an optical zone of the CL, can be approximated by the lens surface radius of curvature, R, as follows:
where:
In equations (3) and (4), “d”=diameter of the contact lens, R=radius of curvature of the contact lens, and S0=sagittal height of the contact lens. The tilted CL surface, S′(x,y), is defined by the coordinate transformation for each X- and Y-meridian, according to:
And, similarly for the Y-meridian:
These two coordinate transformations define the surface, S′(x,y), of a new tilted CL surface that is tilted by angles θx and θy.
The location of the corneal vertex normal (which is perpendicular to the cornea's surface at the cornea's center) can be determined optically from centroid measurements of two or more (e.g., four) Purkinje reflections of a visual eye image (EI) in
In addition, the contact lens virtual (edge-identified) center (Xe, Ye) can be determined by fitting the contact lens' edge definition to a circle (or ellipse). From these operations, the virtual contact lens center (Xe, Ye) and the corneal vertex location (Xv, Yv) can be determined. The lens' tilt angles (θx and θy) are thus given by:
The following pair of equations define the real center (Xc, Yc) of the tilted contact lens, as follows:
X
c
=X
e
+S
0 sin θx (Eq. 9)
and
Y
c
=Y
e
+S
0 sin θy (Eq. 10).
There are several ways to make the measurement of the corneal vertex position with a contact lens on the eye. In nearly all aberrometers there is also a system for visually imaging the iris of the eye with a camera. This is the system that is generally used to determine the alignment of the contact lens on the eye during the measurement through the aberrometer instrument's front lens. The illumination of the eye is often made using one or more front light sources (usually LEDs) that are arranged circumferentially around the front camera lens. Since the eye and contact lens are both highly curved convex surfaces, there will always be a Purkinje reflection (glint) from these surfaces that is visible in the camera's visual image. These reflections can be used as a means for finding the position of the surface that is normal to the measurement axis. Note that this method depends on the exact arrangement of the imaging optical system, and thus the contact lens vertex position is related to the center of the pattern of Purkinje reflection spots (i.e., glints).
However, if the eye (and hence the cornea) moves radially away from the main optical axis (Z-axis) of the imaging system (See
When fiducial marks 18 are located inside a dilated (larger) pupil, they might not be able to be distinguished from a visible iris image (it is not consistent; sometimes we can, and sometimes we cannot). Instead, they can always be easily found in the WFS (Wavefront Sensor) image taken by an aberrometer instrument.
In some embodiments, one or more fiducial marks 18 can be located on a predicate CL 16: (α) outside of the pupil, but inwards from the edge of the CL 16 (see
In some embodiments, three fiducial marks can be used on a predicate CL, which can be arranged in an isosceles triangle configuration (see, for example,
In a combined corneal topographer and aberrometer instrument, a part of the light is projected through the collecting lens and is collimated by a Telecentric Stop Aperture (TSA) so that the only rays collected in object space are parallel to the instrument's Z-axis. This is shown in
In some embodiments, the following equations (11, 12, 13, and 14) can be used to calculate the XY offset amounts (ΔX, ΔY) between the physical center (Xc, Yc) of the tilted and/or rotated predicate contact lens and the center of the pupil (Xp, Yp), as follows:
Also,
ΔX′=XP−Xc (Eq. 15)
ΔY′=Yp−Yc (Eq. 16).
The XY location of vertex central point (Xv, Yv) can be obtained from either a calibrated EI image (e.g., centroid of the four glints) or from a true vertex location from a telecentric CT image. In the telecentric corneal topographer, an exact vertex location can be directly determined. Determining (Xv, Yv) from EI measurements requires performing a calibration (see
Methods A and B comprise two different approaches to obtain the real CL center (Xc, Yc), while EI and CT measurements comprise two different ways to find the vertex central point (Xv, Yv).
This application claims the priority benefit of U.S. Provisional Application Ser. No. 63/402,351 filed Aug. 30, 2022, which is incorporated by reference in its entirety.
Number | Date | Country | |
---|---|---|---|
63402351 | Aug 2022 | US |