The present disclosure relates to orthopedic knee prosthetics and, more specifically, to orthopedic knee prosthetics for use with total knee arthroplasty procedures.
The knee is the largest joint in the body. Normal knee function is required to perform most everyday activities. The knee is made up of the lower end of the femur, which rotates on the upper end of the tibia, and the patella, which slides in a groove on the end of the femur. Large ligaments attach to the femur and tibia to provide stability. The long thigh muscles give the knee strength and produces knee motion.
The joint surfaces where these three bones touch are covered with articular cartilage, a smooth substance that cushions the bones and enables them to move easily. The condition of this cartilage lining the knee joint is a key aspect of normal knee function and is important to the physician when evaluating a potential need for a knee joint replacement.
All remaining surfaces of the knee are covered by a thin, smooth tissue liner called the synovial membrane. This membrane releases a special fluid that lubricates the knee, reducing friction to nearly zero in a healthy knee.
Normally, all of these components work in harmony. But disease or injury can disrupt this harmony, resulting in pain, muscle weakness, and reduced function.
In addition to the smooth cartilage lining on the joint surfaces, there are two smooth discs of cartilage that cushion the space between the bone ends. The inner disc is called the medial meniscus, while the disc on the outer side of the knee joint is called the lateral meniscus. The role of the menisci is to increase the conformity of the joint between the femur and the tibia. The menisci also play an important function as joint shock absorbers by distributing weight-bearing forces, and in reducing friction between the joint segments.
There are also four major ligaments that play an important part in stability of the knee joint. The Medial Collateral Ligament (MCL) and the Lateral Collateral Ligament (LCL) are located on opposing sides on the outside of the joint. The Anterior Cruciate Ligament (ACL) and the Posterior Cruciate Ligament (PCL) are more centrally located ligaments within the joint. The ACL attaches to the knee end of the femur, at the back of the joint and passes down through the knee joint to the front of the flat upper surface of the Tibia. The ACL contacts the femur on the inner lateral condyle. When disrupted, this allows for laxity to occur on the lateral side of the knee. The ACL passes across the knee joint in a diagonal direction and with the PCL passing in the opposite direction, forms a cross shape, hence the name cruciate ligaments.
Total knee replacement (TKR), also referred to as total knee arthroplasty (TKA), is a surgical procedure where worm, diseased, or damaged surfaces of a knee joint are removed and replaced with artificial surfaces. Materials used for resurfacing of the joint are not only strong and durable but also optimal for joint function as they produce as little friction as possible.
The “artificial joint or prosthesis” generally has three components: (1) a distal femoral component usually made of a biocompatible material such as metal alloys of cobalt-chrome or titanium; (2) a proximal tibial component also made of cobalt chrome or titanium alloy; and a bearing component disposed there between usually formed of a plastic material like polyethylene.
In total knee arthroplasty (TKA) there are three main types of implants: The first main type is the posterior cruciate retaining (PCR) total knee arthroplasty, where the surgeon retains the posterior cruciate ligament and sacrifices the anterior cruciate ligament. The second main type is the posterior stabilizing (PS) total knee arthroplasty, where the surgeon sacrifices both the anterior cruciate ligament (ACL) and the posterior cruciate ligament (PCL). With a PS TKA posterior stabilization is introduced into the TKA by using a cam/post mechanism. The third main type is the posterior cruciate sacrificing (PCS) TKA where the surgeon sacrifices both the ACL and the PCL, but does not use a cam/post mechanism for posterior stabilization. Rather, this TKA type uses constraint in the polyethylene to stabilize the anteroposterior movement.
Any of the above three main types of TKA implant can have a fixed bearing (FB) design or a mobile bearing (MB) design. With the fixed bearing design, the polyethylene insert is either compression molded or fixed in the tibial tray using a locking mechanism. In a mobile bearing design, the polyethylene insert is free to either rotate, translate or both rotate and translate.
While knee arthroplasty is known as one of the most consistently successful surgeries offered, there is room for improvement. For example, the ACL is sacrificed during the installation of a total knee arthroplasty system, and doing so can have a negative clinical impact for some patients.
The role of the ACL is to pull the femur in the anterior direction at terminal extension and at full extension. The ACL, attached to the lateral condyle of the femur also works as a tether and keeps the lateral condyle in contact with the lateral meniscus. The PCL pulls the femur in the posterior direction with increasing flexion. The PCL also acts as a tether on the medical condyle of the femur, keeping the medial condyle in contact with the medial meniscus. Together these two ligaments are vitally important to the stability of the knee joint, especially in contact sports and those that involve fast changes in direction and twisting and pivoting movements. Therefore, a torn or absent ACL has serious implications for the stability and function for the knee joint. In other orthopedic fields, surgeons usually recommend ACL replacement surgery for a torn ACL because without the ACL, the femorotibial joint becomes unstable. It is assumed that this instability leads to meniscus and cartilage damage. Unfortunately, the ACL is sacrificed in TKA.
Attempts have been made to design a TKA that retains the ACL, but these procedures are often very difficult to perform and the function of the ACL is often compromised. Fluoroscopic studies have been conducted on previous ACL retaining TKA designs and they have reported that these patients have difficulty achieving full extension and often experience a very tight knee at 90 degrees of knee flexion, under weight-bearing conditions. This is probably due to the knee joint becoming overly constrained due to the retention of the cruciate ligaments, but the patient's geometrical condylar shapes being altered. Sacrificing the ACL contributes to laxity in the joint that allows the femur freedom of motion due to the changes in their condylar shapes.
Known TKA implants, such as PS and PCR TKA, provide for posterior stabilization, but not anterior stabilization. What is needed, therefore, is a TKA implant that provides for anterior stabilization in the absence of a surgically removed ACL while also accommodating a retained PCL.
According to one aspect of the disclosure, a total knee implant prosthesis comprises a femoral component including a pair of condyles and a cam positioned between the pair of condyles. The cam has a convex curved surface including a center point that is laterally offset from a center line of the femoral component when the femoral component is viewed in a first plane. It should be appreciated that the first plane may correspond to a traverse plane of a patient's body. The total knee implant prosthesis further comprises a tibial component including a medial bearing surface, a lateral bearing surface, and a post positioned between the medial bearing surface and the lateral bearing surface. The post has a curved surface that is angled to face toward the medial bearing surface and away from the lateral bearing surface when the tibial component is viewed in the first plane. The femoral component is configured to rotate relative to the tibial component between a full extension position and a full flexion position, and the cam and the post are sized, shaped, and positioned so that the cam engages the post at a contact point on the curved surface of the post when the femoral component is in the full extension position. When the femoral component is rotated from the full extension position toward the full flexion position, the cam and the post are sized, shaped, and positioned so that the contact point moves laterally along the curved surface of the post. The cam and the post are also sized, shaped, and positioned so that the cam is disengaged from the post when the femoral component is in the full flexion position.
In some embodiments, the tibial component may have a medial-lateral center line when the tibial component is viewed in the first plane, and the post may have a medial-lateral center line that is laterally offset from the medial-lateral center line of the tibial component when the tibial component is viewed in the first plane.
In some embodiments, the curved surface of the post may define an arced line having a center point that lies on the medial-lateral center line of the post when the tibial component is viewed in the first plane. In some embodiments, the arced line is convex. In some embodiments, the arced line may have a radius extending from an origin that is offset in a lateral direction from the medial-lateral center line of the post. The radius may be offset by a distance equal to less than 6 mm. In other embodiments, the distance may be less than or equal to 12 mm.
Additionally, in some embodiments, the medial bearing surface may include a distal-most point, and a distance may be defined in an anterior-posterior direction between the center point and the distal-most point of the medial bearing surface. The distance may be greater than 0 mm and less than or equal to about 10 mm. In other embodiments, the distance may be greater than 0 mm and less than or equal to about 15 mm.
In some embodiments, the medial-lateral center line of the post may be offset in a lateral direction from the medial-lateral center line of the tibial component by a distance that is equal to less than 6 mm. In other embodiments, the distance may be less than or equal to 12 mm.
In some embodiments, when the tibial component is viewed in a second plane extending orthogonal to the first plane, the curved surface of the post may define a concave curved line. It should be appreciated that the second plane may correspond to the sagittal plane of the patient's body. In some embodiments, the concave curved line may be defined by a radius that is in a range of 3 mm to 25 mm. In other embodiments, the distance may be in a range of 3 mm to 30 mm. In some embodiments, when the tibial component is viewed in the first plane, the curved surface may define a convex curved line.
In some embodiments, the medial bearing surface and the lateral bearing surface may be asymmetrical. Additionally, in some embodiments, the lateral bearing surface may be flatter than the medial bearing surface.
According to another aspect, a total knee implant prosthesis comprises a tibial component including a pair of bearing surfaces and a post positioned between the bearing surfaces, and a femoral component configured to rotate relative to the tibial component. The femoral component includes a pair of condyles sized and shaped to articulate on the bearing surfaces and an anterior cam positioned between the pair of condyles. The cam engages the post at a first contact point when the femoral component is at 0 degrees of flexion, and the cam engages the post at a second contact point located lateral of the first contact point when the femoral component is at a first degree of flexion greater than 0 degrees. Additionally, the cam is disengaged from the post when the femoral component is at a second degree of flexion greater than the first degree of flexion.
In some embodiments, the post may have a medial-lateral center line when the tibial component is viewed in a first plane, the first contact point may be located medial of the medial-lateral center line, and the second contact point may be located lateral of the medial-lateral center line.
In some embodiments, the tibial component may have a medial-lateral center line when the tibial component is viewed in the first plane. The medial-lateral center line of the post may be laterally offset from the medial-lateral center line of the tibial component when the tibial component is viewed in the first plane.
Additionally, in some embodiments, the cam may include a posterior surface configured to engage an anterior surface of the post at the first contact point and the second contact point. The posterior surface of the cam may define a convex curved line when the femoral component is viewed in a first plane, and the anterior surface of the post may define a convex curved line when the femoral component is viewed in the first plane.
In some embodiments, the anterior surface of the post may define a concave curved line when the tibial component is viewed in a second plane positioned orthogonal to the first plane. In some embodiments, the convex curved line that is defined by the cam may have a center point that is laterally offset from a center line of the femoral component when the femoral component is viewed in a first plane.
In some embodiments, the cam may be configured to engage an anterior surface of the post that is angled to face toward a medial bearing surface of the pair of bearing surfaces and away from a lateral bearing surface of the pair of bearing surfaces.
The detailed description particularly refers to the following figures, in which:
While the concepts of the present disclosure are susceptible to various modifications and alternative forms, specific exemplary embodiments thereof have been shown by way of example in the drawings and will herein be described in detail. It should be understood, however, that there is no intent to limit the concepts of the present disclosure to the particular forms disclosed, but on the contrary, the intention is to cover all modifications, equivalents, and alternatives falling within the spirit and scope of the invention as defined by the appended claims.
Terms representing anatomical references, such as anterior, posterior, medial, lateral, superior, inferior, etcetera, may be used throughout the specification in reference to the orthopaedic implants and orthopaedic surgical instruments described herein as well as in reference to the patient's natural anatomy. Such terms have well-understood meanings in both the study of anatomy and the field of orthopaedics. Use of such anatomical reference terms in the written description and claims is intended to be consistent with their well-understood meanings unless noted otherwise.
The exemplary embodiments of the present disclosure are described and illustrated below to encompass prosthetic knee joints and knee joint components, as well as methods of implanting and reconstructing knee joints. Of course, it will be apparent to those of ordinary skill in the art that the preferred embodiments discussed below are exemplary in nature and may be reconfigured without departing from the scope and spirit of the present invention. However, for clarity and precision, the exemplary embodiments as discussed below may include optional steps, methods, and features that one of ordinary skill should recognize as not being a requisite to fall within the scope of the present invention.
Referring now to
The femoral component 12 is illustratively formed from a metallic material such as cobalt-chromium or titanium, but may be formed from other materials, such as a ceramic material, a polymer material, a bio-engineered material, or the like, in other embodiments. The tibial tray insert 18 is illustratively formed from a polymer material such as an ultra-high molecular weight polyethylene (UHMWPE), but may be formed from other materials, such as a ceramic material, a metallic material, a bio-engineered material, or the like, in other embodiments.
As shown in
The exemplary femoral component 12 includes a pair of condyles 22, 24, each of which has an arcuate shape in order to allow for smooth rotation of the femur with respect to the tibia. In general, the femoral component includes an anterior portion 26 and a posterior portion 28 that are shown by the dotted line imaginary boundary line 29 in
The front exterior face 30 of the femoral component 12 includes an articulation surface 48 that is configured to engage a corresponding surface of a patella component. The articulation surface 48 defines the depression 32 and includes the arcuate bridge 34. At the arcuate bridge, the articulation surface 48 separates into a medial articulation surface 50 of the medial condyle 22 and a lateral articulation surface 52 of the lateral condyle 24. The surfaces 50, 52 are configured to engage with and articulate on corresponding bearing surfaces 54, 56, respectively, of the tibial component 14. The articulation surfaces 50, 52 of the condyles 22, 24 flatten out and do not exhibit a uniform arcuate shape from anterior to posterior. Additionally, as illustrated in
As shown in
When viewed in the plane of
As shown in
The superior-inferior distance 86 between the distal-most point 68 and the origin 82 (and also the posterior-most point 72 of the cam 36) is equal to about 12.25 mm in the illustrative embodiment. In other embodiments, the distance 86 may be in a range of about 5 mm to about 20 mm. The radius 80 of the surface 60 is illustratively equal to about 3 mm, but, in other embodiments, the radius 80 may be in a range of about 1 mm to about 6 mm. In still other embodiments, the radius 80 may be greater than 6 mm. It should be appreciated that in other embodiments the radius 80 and the distances 84, 86 may be greater or less than these ranges depending on the physical requirements of a particular patient.
Referring now to
As shown in
As shown in
In the illustrative embodiment, the posterior-most point 72 is the medial-lateral mid-point of the surface 60 of the cam 36. As shown in
In the illustrative embodiment, the center line of the gap 20 is also offset by the same distance 98 from the center line 96 of the femoral component 12. In that way, the gap 20 is laterally offset in the femoral component 12. It should be appreciated that in other embodiments the distance 98 may be greater or less than these ranges depending on the physical requirements of a particular patient. In other embodiments, the center line of the gap 20 offset from the center line by a different distance than the other structures of the femoral component 12. In still other embodiments, the center line of the gap 20 may not be offset at all.
Returning to
The post 38 has an anterior surface or wall 100 that is configured to engage the posterior surface 60 of the cam 36 of the femoral component 12 when the implant 10 (and hence the knee) is at full extension and over part of flexion. As shown in
The bearing surfaces 54, 56 are illustratively concave surfaces. Additionally, as shown in
Referring now to
As shown in
In the illustrative embodiment, the anterior wall 100 of the post 38 is arcuate or rounded when the post 38 is viewed in a transverse plane. As shown in
The arced line 140 (and hence the anterior wall 100 in the transverse plane) has a radius 142 that extends from an origin 144. As shown in
Due to the combination of the distances 120, 146, the origin 144 is offset laterally from the central line 116 of the tibial insert 18 by about 4.2 mm. In other embodiments, the origin 144 may be offset in a range of about 0 mm to about 12 mm.
As shown in
In the illustrative embodiment, the anterior wall 100 of the post 38 is angled toward the medial bearing surface 54. As shown in
Referring now to
The curved line 160 (and hence the anterior wall 100) has a radius 162 that extends from an origin 164. The radius 162 of the anterior wall 100 is illustratively equal to about 25 mm, but, in other embodiments, the radius 162 may be in a range of about 3 mm to about 25 mm. In still other embodiments, the radius may be greater than 25 mm. It should be appreciated that in other embodiments the radius may be greater or less than these ranges depending on the physical requirements of a particular patient.
Referring now to
The anterior cam 36 of the femoral component 12 is illustrated in contact with the anterior wall 100 of the tibial post 38 at about 0 degrees of flexion in
As the femoral component 12 is articulated between about 0 degrees of flexion and about 7.5 degrees of flexion, the femoral component 12 rotates laterally relative to the tibial insert 18, and the contact point between the cam 36 and the post 38 moves laterally during flexion along the anterior wall 100, as shown in
As the femoral component 12 is articulated between about 7.5 degrees of flexion and 15 degrees of flexion, the femoral component 12 continues to rotate laterally relative to the tibial insert 18, and the contact point between the cam 36 and the post 38 moves laterally along the anterior wall 100 during flexion, as shown in
As described above, the femoral component 12 rotates relative to the tibial insert 18 in the direction indicated by arrow 210 in
As described above, the location where the cam 36 contacts the post 38 moves laterally as the femoral component 12 is articulated from about 0 degrees of flexion to about 15 degrees of flexion. As shown in
Referring now to
As described above, the cam of the femoral component and the post of the tibial component or insert are offset in the lateral direction from the respective center lines of those components. It should be appreciated that in other embodiments, the cam of the femoral component may be centered on the center line of the femoral component with the post of the tibial component offset in the lateral direction. In such embodiments, the cam width is greater than the post width.
Following from the above description and invention summaries, it should be apparent to those of ordinary skill in the art that, while the methods and apparatuses herein described constitute exemplary embodiments of the present invention, the invention contained herein is not limited to this precise embodiment and that changes may be made to such embodiments without departing from the scope of the invention as defined by the claims. Additionally, it is to be understood that the invention is defined by the claims and it is not intended that any limitations or elements describing the exemplary embodiments set forth herein are to be incorporated into the interpretation of any claim element unless such limitation or element is explicitly stated. Likewise, it is to be understood that it is not necessary to meet any or all of the identified advantages or objects of the invention disclosed herein in order to fall within the scope of any claims, since the invention is defined by the claims and since inherent and/or unforeseen advantages of the present invention may exist even though they may not have been explicitly discussed herein.
This application is a continuation of U.S. patent application Ser. No. 16/246,910, filed Jan. 14, 2019, which is a continuation of U.S. patent application Ser. No. 15/222,862, now U.S. Pat. No. 10,179,052, filed Jul. 28, 2016. Both of the above-identified prior applications are expressly incorporated herein by reference.
Number | Name | Date | Kind |
---|---|---|---|
3765033 | Goldberg et al. | Oct 1973 | A |
3869731 | Waugh et al. | Mar 1975 | A |
4209861 | Walker et al. | Jul 1980 | A |
4215439 | Gold et al. | Aug 1980 | A |
4262368 | Lacey | Apr 1981 | A |
4340978 | Buechel et al. | Jul 1982 | A |
4470158 | Pappas et al. | Sep 1984 | A |
4888021 | Forte et al. | Dec 1989 | A |
5007933 | Sidebotham et al. | Apr 1991 | A |
5071438 | Jones et al. | Dec 1991 | A |
5133758 | Hollister | Jul 1992 | A |
5147405 | Van Zile et al. | Sep 1992 | A |
5219362 | Tuke et al. | Jun 1993 | A |
5326361 | Hollister | Jul 1994 | A |
5330533 | Walker | Jul 1994 | A |
5344460 | Turanyi et al. | Sep 1994 | A |
5358527 | Forte | Oct 1994 | A |
5370699 | Hood et al. | Dec 1994 | A |
5387240 | Pottenger et al. | Feb 1995 | A |
5395401 | Bahler | Mar 1995 | A |
5405396 | Heldreth et al. | Apr 1995 | A |
5413604 | Hodge | May 1995 | A |
5549686 | Johnson et al. | Aug 1996 | A |
5571194 | Gabriel | Nov 1996 | A |
5609639 | Walker | Mar 1997 | A |
5609643 | Colleran et al. | Mar 1997 | A |
5639279 | Burkinshaw et al. | Jun 1997 | A |
5658342 | Draganich et al. | Aug 1997 | A |
5681354 | Eckhoff | Oct 1997 | A |
5683468 | Pappas | Nov 1997 | A |
5702458 | Burstein et al. | Dec 1997 | A |
5702466 | Pappas et al. | Dec 1997 | A |
5755801 | Walker et al. | May 1998 | A |
5776201 | Colleran et al. | Jul 1998 | A |
5778537 | Leini | Jul 1998 | A |
5800552 | Forte | Sep 1998 | A |
5811543 | Hao et al. | Sep 1998 | A |
5824100 | Kester et al. | Oct 1998 | A |
5824102 | Buscayret | Oct 1998 | A |
5871543 | Hofmann | Feb 1999 | A |
5871546 | Colleran et al. | Feb 1999 | A |
5879392 | McMinn | Mar 1999 | A |
5906643 | Walker | May 1999 | A |
5935173 | Roger et al. | Aug 1999 | A |
5964808 | Blaha et al. | Oct 1999 | A |
5997577 | Herrington et al. | Dec 1999 | A |
6004351 | Tomita et al. | Dec 1999 | A |
6013103 | Kaufman et al. | Jan 2000 | A |
6039764 | Pottenger et al. | Mar 2000 | A |
6056779 | Noyer et al. | May 2000 | A |
6080195 | Colleran et al. | Jun 2000 | A |
6123729 | Insall et al. | Sep 2000 | A |
6206926 | Pappas | Mar 2001 | B1 |
6264697 | Walker | Jul 2001 | B1 |
6299646 | Chambat et al. | Oct 2001 | B1 |
6325828 | Dennis et al. | Dec 2001 | B1 |
6344059 | Krakovits et al. | Feb 2002 | B1 |
6379388 | Ensign et al. | Apr 2002 | B1 |
6443991 | Running | Sep 2002 | B1 |
6475241 | Pappas | Nov 2002 | B2 |
6491726 | Pappas | Dec 2002 | B2 |
6540787 | Biegun et al. | Apr 2003 | B2 |
6558426 | Masini | May 2003 | B1 |
6582469 | Tornier | Jun 2003 | B1 |
6589283 | Metzger et al. | Jul 2003 | B1 |
6730128 | Burstein | May 2004 | B2 |
6764516 | Pappas | Jul 2004 | B2 |
6770099 | Andriacchi et al. | Aug 2004 | B2 |
6797005 | Pappas | Sep 2004 | B2 |
6846329 | McMinn | Jan 2005 | B2 |
6893388 | Reising et al. | May 2005 | B2 |
6893467 | Bercovy | May 2005 | B1 |
6916340 | Metzger et al. | Jul 2005 | B2 |
6926738 | Wyss | Aug 2005 | B2 |
6972039 | Metzger et al. | Dec 2005 | B2 |
6986791 | Metzger | Jan 2006 | B1 |
7025788 | Metzger et al. | Apr 2006 | B2 |
7066963 | Naegerl | Jun 2006 | B2 |
7081137 | Servidio | Jul 2006 | B1 |
7105027 | Lipman et al. | Sep 2006 | B2 |
7160330 | Axelson, Jr. et al. | Jan 2007 | B2 |
7261740 | Tuttle et al. | Aug 2007 | B2 |
7326252 | Otto et al. | Feb 2008 | B2 |
7422605 | Burstein et al. | Sep 2008 | B2 |
7572292 | Crabtree et al. | Aug 2009 | B2 |
7658767 | Wyss | Feb 2010 | B2 |
7678152 | Suguro et al. | Mar 2010 | B2 |
7842093 | Peters et al. | Nov 2010 | B2 |
7875081 | Lipman et al. | Jan 2011 | B2 |
8915965 | Komistek | Dec 2014 | B2 |
9132014 | Sanford et al. | Sep 2015 | B2 |
10179052 | Clary et al. | Jan 2019 | B2 |
20030009232 | Metzger et al. | Jan 2003 | A1 |
20040243244 | Otto et al. | Dec 2004 | A1 |
20040243245 | Plumet et al. | Dec 2004 | A1 |
20050096747 | Tuttle et al. | May 2005 | A1 |
20050143832 | Carson | Jun 2005 | A1 |
20050154472 | Afriat | Jul 2005 | A1 |
20050209701 | Suguro et al. | Sep 2005 | A1 |
20060015185 | Chambat et al. | Jan 2006 | A1 |
20060178749 | Pendleton et al. | Aug 2006 | A1 |
20070135926 | Walker | Jun 2007 | A1 |
20080021566 | Peters et al. | Jan 2008 | A1 |
20080119940 | Otto et al. | May 2008 | A1 |
20080243258 | Sancheti | Oct 2008 | A1 |
20080269596 | Revie et al. | Oct 2008 | A1 |
20090043396 | Komistek | Feb 2009 | A1 |
20090306785 | Farrar et al. | Dec 2009 | A1 |
20090319047 | Walker | Dec 2009 | A1 |
20090326663 | Dun | Dec 2009 | A1 |
20100016977 | Masini | Jan 2010 | A1 |
20100036500 | Heldreth et al. | Feb 2010 | A1 |
20100042224 | Otto et al. | Feb 2010 | A1 |
20110118847 | Lipman et al. | May 2011 | A1 |
20140257502 | Masini et al. | Sep 2014 | A1 |
20150134067 | Qu et al. | May 2015 | A1 |
20150182344 | McKinnon et al. | Jul 2015 | A1 |
Number | Date | Country |
---|---|---|
201438977 | Apr 2010 | CN |
101879099 | Nov 2010 | CN |
103126787 | Jun 2013 | CN |
103327937 | Sep 2013 | CN |
19529824 | Feb 1997 | DE |
636352 | Feb 1995 | EP |
732091 | Sep 1996 | EP |
1440675 | Jul 2004 | EP |
1591082 | Nov 2005 | EP |
2248488 | Nov 2010 | EP |
2254519 | Dec 2010 | EP |
2786727 | Oct 2014 | EP |
2417971 | Sep 1979 | FR |
2621243 | Apr 1989 | FR |
2787012 | Jun 2000 | FR |
2835178 | Aug 2003 | FR |
8-500992 | Feb 1996 | JP |
2004167255 | Jun 2004 | JP |
2010259808 | Nov 2010 | JP |
2014510562 | May 2014 | JP |
0209624 | Feb 2002 | WO |
2004058108 | Jul 2004 | WO |
2005072657 | Aug 2005 | WO |
2007108804 | Sep 2007 | WO |
2007108933 | Sep 2007 | WO |
2007119173 | Oct 2007 | WO |
2015118517 | Aug 2015 | WO |
2016048800 | Mar 2016 | WO |
Entry |
---|
Yoshiya et al., “In Vivo Kinematic Comparison of Posterior Cruciate-Retaining and Posterior-Stabilized Total Knee Arthroplasties Under Passive and Weight-Bearing Conditions,” J Arthroplasty, vol. 20, No. 6, 2005, 7 Pgs. |
Kurosawa, et al., “Geometry and Motion of the Knee for Implant and Orthotic Design”. The Journal of Biomechanics vol. 18 No. 7(1985), pp. 487-499, 12 Pages. |
Japanese SR for Corresponding Patent Application No. 2009-501393, Dated Oct. 26, 2010, 5 Pages (DEP6251). |
EPO Search Report From Corresponding EPO Patent App No. 12164381.1-1654, Dated Mar. 7, 2013, 3 Pages (DEP6251 EPETD2). |
Uvehammer, et al., “In Vivo Kinematics of Total Knee Arthroplasty: Concave Versus Posterior-Stabilised Tibial Joint Surface,” Journal of Bone & Joint Surgery, vol. 82-B, No. 4, May 2000, pp. 499-505. |
Uvehammer et al., “In Vivo Kinematics of Total Knee Arthroplasty: Flat Compared With Concave Tibial Joint Surface,” J Orthop Res 18(6):856-64, 2000. |
European Search Report in EPO App. No. 09164478.1-2310 Dated Apr. 28, 2010, 11 Pages (DEP6123). |
European Search Report for Furopean Patent Application No. 17178067.9-1664, Dated Jan. 3, 2018. |
English translation of Japanese Search Report for Application No. 2017-145208, Mar. 30, 2021, 2 pages. |
Chinese First Office Action and Search Report, Chinese Application No. 201710629741.X, dated Jul. 27, 2020, 8 pages. |
Chinese Search Report for Chinese Application No. 201010173717.8, 2 Pages (DEP6299CNNP). |
European Search Report From Corresponding EPO App No. 06739287.8 Dated Mar. 16, 2010, 3 Pages (DEP6251). |
European Search Report in Corresponding EPO App No. 10162138.1 Dated Aug. 30, 2010, 7 Pages (DEP6299). |
European Search Report From Corresponding EPO App. No. 12164381.1-2310, Dated May 18, 2012, 4 Pages. |
European Search Report in Corresponding EPOapp. No. 12181217.6-2310, Dated Jan. 15, 2013, 6 Pages. |
European Search Report in Corresponding App. (I.E., 09164478.1-2310), Mailed Oct. 20, 2009, 6 Pages (DEP6123). |
Australian Search Report for Corresponding App.. No. 2006340364, Dec. 11, 2009, 2 Pages (DEP6251AU). |
Shaw, et al., “The Longitudinal Axis of the Knee & The Role of the Cruciate Ligaments in Controlling Transverse Rotation”, J. Bone Joint Surg. Am. 1974;56:1603-1609. |
PCT Notification Concerning Transmittal of International Prel. Report for Corresponding International App. No. PCT/US2006/010431, Oct. 2, 2008, 6 Pages (DEP6251WOPCT). |
PCT Notification Concerning Transmittal of International Prel. Report for Corresponding International App. No. PCT/US2006/010431, Jun. 5, 2007, 8 Pages (DEP6251 WOPCT). |
Andriacchi, T.P., “The Effect of Knee Kinematics, Gait and Wear On the Short and Long-Term Outcomes of Primary Knee Replacement,” NIH Consensus Development Conference on Total Knee Replacement, pp. 61-64, Dec. 8-10, 2003, (4 Pgs). |
Asano et al. “In Vivo Three-Dimensional Knee Kinematics Using a Biplanar Image-Matching Technique,” Clin Orthop Rel Res, 388: 157-166, 2001 (10 Pgs). |
Barnes, C.L., et al, Kneeling is Safe for Patients Implanted With Medial-Pivot Total Knee Arthroplasty Designs, Journal of Arthroplasty, vol. 00, No. 0 2010, 1-6, 6 Pgs. |
Bertin et al., “In Vivo Determination of Posterior Femoral Rollback for Subjects Having a Nexgen Posterior Cruciate-Retaining Total Knee Arthroplasty,” J Arthroplasty, vol. 17, No. 8, 2002, 9 Pages. |
Blaha, et al., “Kinematics of the Human Knee Using an Open Chain Cadaver Model”, Clinical Orthopaedics and Related Research, vol. 410 (2003); 25-34. |
Clary et al., “Kinematics of Posterior Stabilized and Cruciate Retaining Knee Implants During an in Vitro Deep Knee Bend,” 54th Annual Meeting of the Orthopaedic Research Society, Poster No. 1983, Mar. 2008. |
D'Lima et al., “Quadriceps Moment Arm and Quadriceps Forces After Total Knee Arthroplasty,” Clin Orthop Rel Res 392:213-20, 2001. |
Dennis, et al, “A Multicenter Analysis of Axial Femorotibial Rotation After Total Knee Arthroplasty”, Clinical Orthopaedics 428 (2004); 180-89. |
Dennis, et al., “In Vivo Anteroposterior Femorotibial Translation of Total Knee Arthroplasty: A Mul Ticenter Analysis,” Clin Orthop Rel Res, 356: 47-57, 1998. |
Dennis et al., “In Vivo Determination of Normal and Anterior Cruciate Ligament-Deficient Knee Kinematics,” J Biomechanics, 38, 241-253, 2005, 13 Pgs. |
Dennis et al “Multicenter Determination of in Vivo Kinematics After Total Knee Arthroplasty,” Clin Orthop Rel Res., 416, 37-57, 21 Pgs. |
Fan, Cheng-Yu, et al, Primitive Results After Medial-Pivot Knee Arthroplasties: A Minimum 5-Year Follow-Up Study, The Journal of Arthroplasty, vol. 25, No. 3 2010, 492-496, 5 Pgs. |
Ferris, “Matching Observed Spiral Form Curves To Equations of Spirals In 2-0 Images,” The First Japanese-Australian Joint Seminar, 7 Pgs. |
Freeman, Mar., et al, The Movement of the Normal Tibio-Femoral Joint, the Journal of Biomechanics 38 (2005) (2), pp. 197-208, 12 pgs. |
Fuller, et al., “A Comparison of Lower-Extremity Skeletal Kinematics Measured Using Skin and Pin Mounted Markers”, Human Movement Science 16 (1997) 219-242. |
Goodfellow et al., “The Mechanics of the Knee and Prosthesis Design,” The Journal of Bone and Joint Surgery, vol. 60-B, No. 3, 12 Pgs. |
Hill, et al., “Tibiofemoral Movement 2: The Loaded and Unloaded Living Knee Studied By MRI”, The Journal of Bone & Joint Surgery, vol. 82-B, No. 8 (Nov. 2000) 1196-1198. |
P. Johal et al, “Tibio-Femoral Movement in the Living Knee. a Study of Weight Bearing and Non-Weight Bearing Knee Kinematics Using ‘Interventional’ MRI,” Journal of Biomechanics, vol. 38, Issue 2, Feb. 2005, pp. 269-276 (8 Pgs). |
Karachalios, et al., “A Mid-Term Clinical Outcome Study of the Advance Medial Pivot Knee Arthroplasty,” www.sciencedirect.com, The Knee 16 (2009); 484-488. |
Kessler et al., “Sagittal Curvature of Total Knee Replacements Predicts in Vivo Kinematics,” Clinical Biomechanics 22(1): 52-58, 2007. |
Komistek, et al., “In Vivo Fluoroscopic Analysis of the Normal Human Knee”, Clinical Orthopaedics 410 (2003): 69-81. |
Komistek, et al., “In Vivio Polyethylene Bearing Mobility is Maintained in Posterior Stabilized Total Knee Arthroplasty”, Clinical Orthopaedics 428 (2004): 207-213. |
Koo, et al., “The Knee Joint Center of Rotation is Predominantly On the Lateral Side During Normal Walking”, Journal of Biomechanics, vol. 41 (2008); 1269-1273. |
Li et al., “Anterior Cruciate Ligament Deficiency Alters the in Vivo Motion of the Tibiofemoral Cartilage Contact Points in Both Anteroposterior and Mediolateral Directions,” JBJS-AM, vol. 88, No. 8, Aug. 2006, 9 Pgs. |
Mannan, et al., “The Medical Rotation Total Knee Replacement: a Clinical and Radiological Review at a Mean Follow-Up of Six Years”, The Journal of Bone and Joint Surgery, vol. 91-B, No. 6 (Jun. 2009): 750-756. |
Moonot, et al., “Correlation Between the Oxford Knee and American Knee Society Scores At Mid-Term Folow-Up”, The Journal of Knee Surgery, vol. 22, No. 3 (Jul. 2009), 226-230. |
Murphy, Michael Charles, “Geometry and the Kinematics of the Normal Human Knee”, Submitted To Massachusetts Institute of Technology (1990). |
Nakagawa, et al., “Tibiofemoral Movement 3: Full Flexion in the Living Knee Studied By MRI”, The Journal of Bone and Joint Surgery, vol. 82-B, No. 8 (Nov. 2000): 1199-1200. |
“Nexgen Complete Knee Solution Cruciate Retaining Knee (CR),” Zimmer, Available at: http://zimmer.com.au/ctl?template=PC&op=global&action=&template=PC&id=356, downloaded on Feb. 18, 2009, (1 page). |
Omori, et al., “The Effect of Geometry of the Tibial Polyethylene Insert on the Tibiofemoral Contact Kinematics in Advance Medical Pivot Total Knee Arthroplasty”, The Journal of Orthopaedics Science (2009) 14: 754-760. |
Ranawat, “Design May Be Counterproductive for Optimizing Flexion After TKR,” Clin Orthop Rel Res 416: 174-6, 2003. |
Ries, “Effect of ACL Sacrifice, Retention or Substitution on K After TKA,” http://www.orthosupersite.comniew.asp?rid=23134, Aug. 2007, 5 Pgs. |
Saari et al., “The Effect of Tibial Insert Design on Rising From a Chair, Motion Analysis After Total Knee Replacement,” Clin Biomech 19(9); 951-6, 2004. |
“Scorpio Knee TS Single Axis Revision Knee System,” Stryker Orthopaedics, http://www.stryker.com/stellenUgroups/public/documents/web prod/023609.pdf, (6 pgs). |
Shakespeare, et al., “Flexion After Total Knee Replacement. a Comparison Between the Medial Pivot Knee and a Posterior Stabilised Implant,” www.sciencedirect.com, The Knee 13 (2006): 371-372. |
Suggs et al., “Three-Dimensional Tibiofemoral Articular Contact Kinematics of a Cruciate-Retaining Total Knee Arthroplasty,” JBJS-AM, vol. 88-A, No. 2, 2006, 9 Pgs. |
Vanguard Complete Knee System, Biomet, Available at http://www.biomet.com/patientsnanguard_complete.cfm, downloaded on Feb. 2009, 3 pgs. |
Walker, et al., “Motion of a Mobile Bearing Knee Allowing Translation and Rotation”, Journal of Arthroplasty 17 (2002): 11-19. |
Wang et al., “A Biomechanical Comparison Between the Single-Axis and Multi-Axis Total Knee Arthroplasty Systems for Stand-To-Sit Movement,” Clin Biomech 20(4); 428-33, 2005. |
Wang et al., “Biomechanical Differences Exhibited During Sit-To-Stand Between Total Knee Arthroplasty Designs of Varying Radii,” J Arthroplasty 21 (8): 1193-1199, 2006. |
Number | Date | Country | |
---|---|---|---|
20210228367 A1 | Jul 2021 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 16246910 | Jan 2019 | US |
Child | 17230376 | US | |
Parent | 15222862 | Jul 2016 | US |
Child | 16246910 | US |