This disclosure relates to the analysis of petroleum reservoirs using tracers.
A petroleum reservoir is an underground pool of hydrocarbon compounds contained in porous or fractured rock formations. The petroleum in the reservoir is accessed through one or more borings in the earth that penetrate the material above the reservoir and enable transport of the petroleum to the surface. Water flooding is used, for example, to increase the pressure within the reservoir, thereby increasing oil production rates; and to displace hydrocarbons with the reservoir. Water is ideal for flooding reservoirs due to its ready availability and immiscibility with hydrocarbons. Determining the presence of fluid flow paths between oil wells, and the flow capacity between them, allows for a more detailed description of reservoir heterogeneity and facilitates water flood rate management.
Cross-well tracers, also referred to as inter-well tracers, can be used to obtain information about reservoir fluid flow patterns by injecting the tracer at an injection location, and subsequently retrieving and analyzing a quantity of the injected tracer at a producing location downstream from the injection location. Tracer test data can be used to validate or update existing reservoir/geological models leading to history matching and production optimization with much greater fidelity.
Nonetheless, despite the advantages associated with using tracer test data for reservoir management, much of the potential of full-field tracer implementation has not yet been realized. One impediment to such implementation is the conventional data collection methodology for breakthrough monitoring, which involves manual collection of samples from each well head, and transportation of the collected samples to a laboratory for work up that typically includes pre-concentration, purification, chemical derivatization (to ensure vaporization and ionizability), and analysis in a gas chromatography triple-quadrupole mass spectrometer (GC-QqQ-MS). Even when an automatic sampler is used to collect the fluids from the producer wells, the vials are still transported manually to the laboratory for work-up.
For modern reservoir management operations, autonomous and near real-time monitoring of full-field tracer campaigns can be accomplished by flow-based optical detection modalities. In these tracer tests, tens or hundreds of uniquely identifiable, optically detectable barcodes are introduced into injector wells and expected to be collected from different producer wells. These tracers are detected at trace/ultra-trace levels and their origins of injection unambiguously determined. By limiting manual intervention in this measurement workflow, some of the major shortcomings of conventional molecular tracers, such as fluorobenzoic acids and naphthalenesulfonic acids, can be overcome. In particular, current measurement methods involving detection of these tracers at ultra-trace levels employ complex and expensive processes carried out by personnel on specialized equipment (GC-QqQ-MS).
To implement this type of tracer detection, this disclosure features methods and systems that can be realized in a flow-through workflow in which a groundwater sample with multiple organic tracer species from a petroleum containing reservoir is separated into multiple components, each of which corresponds to a different tracer species, and then combined with an excess of one or more lanthanide elements, such that the lanthanide element concentration is several times larger than the concentration of the tracer species. In this manner, each lanthanide element will bind to one tracer molecule or no tracer molecules. Washing away or removing unbound lanthanide elements yields a solution in which 1:1 complexes of lanthanide elements and tracer species are present; it has been observed that the limits of detection can be reduced if excess lanthanide ions are removed after tracer-lanthanide binding. Measuring fluorescence emission from such complexes yields a signal which varies approximately linearly with the concentration of the tracer species in the groundwater sample, allowing relative concentrations of each of the multiple organic tracer species to be determined.
In a first aspect, this application features methods that include obtaining a groundwater sample that includes multiple organic tracer species from a petroleum containing reservoir, isolating an aqueous fraction of the groundwater sample, separating the aqueous fraction into a plurality of components, where each component corresponds to a different one of the organic tracer species, combining each of the separated components with at least one lanthanide element to form a plurality of component solutions, where a ratio of the at least one lanthanide element to the separated component in each component solution is 10:1 or greater, and analyzing each component solution to determine a relative amount of each organic tracer species in the groundwater sample.
Embodiments of the methods can include any one or more of the following features.
The multiple organic tracer species can include at least twenty organic tracer species, for example, at least thirty organic tracer species, or at least fifty organic tracer species. The multiple organic tracer species can include 2,6-pyridine dicarboxylic acid. The multiple organic tracer species can include at least one derivative of 2,6-pyridine dicarboxylic acid. The multiple organic tracer species can include 4,7-bis(sulfonatophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid. The multiple organic tracer species can include at least one derivative of 4,7-bis(sulfonatophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid.
Isolating the aqueous fraction of the groundwater sample can include extracting solid material from the groundwater sample. The methods can include extracting the solid material by adsorbing the solid material onto an adsorbent material. The methods can include extracting the solid material by sedimentation.
Isolating the aqueous fraction of the solid material can include removing one or more ionic species from the aqueous fraction. The one or more ionic species can include ions that are complexed by one or more of the multiple organic tracers species. Isolating the aqueous fraction can include reducing an amount of one or more chromogenic species from the aqueous fraction. Isolating the aqueous fraction can include increasing a concentration of at least one of the multiple organic tracer species in the aqueous fraction, relative to a concentration of the at least one species in the groundwater sample.
Separating the aqueous fraction into a plurality of components can include chromatographically separating the multiple organic tracer species. The methods can include measuring retention times of the multiple organic tracers species on a chromatography column, and identifying at least one of the multiple organic tracer species based on its corresponding retention time.
Separating the aqueous fraction into a plurality of components can include performing a first chromatographic separation to separate the aqueous fraction into a first plurality of chromatographic fractions, and performing a second chromatographic separation of at least some members of the first plurality of chromatographic fractions to obtain the plurality of components. At least one of the first and second chromatographic separations can include a reversed-phase chromatographic separation, an ion-exchange chromatographic separation, or a hydrophilic interaction chromatographic separation.
The first chromatographic separation can be performed on a first chromatography column and the second chromatographic separation can be performed on a second chromatography column having a chemical selectivity that differs from a chemical selectivity of the first chromatography column. An output fluid stream from the first chromatography column can be directed into an inlet of the second chromatography column.
Combining each of the separated components with at least one lanthanide element can include, for each component, combining the component with a an aqueous solution that includes the at least one lanthanide element.
The at least one lanthanide element can include two or more lanthanide elements, for example, five or more lanthanide elements. The at least one lanthanide element can include terbium. The at least one lanthanide element can include europium.
At least one of the component solutions can include a complex formed of terbium and 2,6-pyridine dicarboxylic acid or a derivative of 2,6-pyridine dicarboxylic acid. The at least one of the component solutions featuring the complex formed of terbium and 2,6-pyridine dicarboxylic acid or a derivative of 2,6-pyridine dicarboxylic acid can include an ammonium acetate buffer. The ratio of terbium to 2,6-pyridine dicarboxylic acid or a derivative of 2,6-pyridine dicarboxylic acid in the at least one of the component solutions can be 10:1 or greater, for example, 20:1 or greater, 100:1 or greater.
A first component solution can include a complex formed of terbium and a first derivative of 2,6-pyridine dicarboxylic acid, and a second component solution comprises a complex formed of terbium and 2,6-pyridine dicarboxylic acid or a second derivative of 2,6-pyridine dicarboxylic acid different from the first derivative. At least one of the component solutions can include a complex formed of europium and 4,7-bis(sulfonatophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid or a derivative of 4,7-bis(sulfonatophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid. The ratio of europium to 4,7-bis(sulfonatophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid or a derivative of europium to 4,7-bis(sulfonatophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid in the at least one of the component solutions can be 10:1 or greater (for example, 20:1 or greater, 100:1 or greater).
A first component solution can include a complex formed of europium and a first derivative of 4,7-bis(sulfonatophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid, and a second component solution can include a complex formed of europium and 4,7-bis(sulfonatophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid or a second derivative of 4,7-bis(sulfonatophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid different from the first derivative.
The methods can include, prior to analyzing each component solution, purifying each component solution to remove an excess quantity of the at least one lanthanide element from each component solution. The purifying can include, for each component solution, chromatographically separating the excess quantity of the at least one lanthanide element from the organic tracer species of the component solution. The methods can include identifying the organic tracer species of each component solution based on a retention time of each component solution on a chromatography column used to perform the chromatographic separation.
Analyzing each component solution can include measuring a quantity of fluorescence emission from a complex of the organic tracer species of the component solution and one of the at least one lanthanide elements, and determining an estimate of an amount of the organic tracer species in the groundwater sample based on the quantity of fluorescence emission. Measuring the quantity of fluorescence emission can include exposing the component solution to illumination radiation and terminating the exposure at an initial time t0, and at a time tm>t0, detecting the fluorescence emission from the component solution, where (tm−t0) is 2 microseconds or more (for example, 5 microseconds or more). At least one of the component solutions can include a first complex formed by a terbium ion and 2,6-pyridine dicarboxylic acid or a derivative of 2,6-pyridine dicarboxylic acid, and an ammonium acetate buffer solution. At least one of the component solutions can include a second complex formed by a terbium ion and 2,6-pyridine dicarboxylic acid or a derivative of 2,6-pyridine dicarboxylic acid, and an ammonium acetate buffer solution, where the second complex is different from the first complex. At least one of the component solutions can include a first complex formed by a europium ion and 4,7-bis(sulfonatophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid or a derivative of 4,7-bis(sulfonatophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid, and an ammonium acetate buffer solution. At least one of the component solutions can include a second complex formed by a europium ion and 4,7-bis(sulfonatophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid or a derivative of 4,7-bis(sulfonatophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid, and an ammonium acetate buffer solution, where the second complex is different from the first complex. At least one of the component solutions can include a first complex formed by a terbium ion and 2,6-pyridine dicarboxylic acid or a derivative of 2,6-pyridine dicarboxylic acid, and an ammonium acetate buffer solution, and at least one of the component solutions can include a second complex formed by a europium ion and 4,7-bis(sulfonatophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid or a derivative of 4,7-bis(sulfonatophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid, and an ammonium acetate buffer solution.
Embodiments of the methods can also include any of the other features described in this application, including combinations of features described in connection with different embodiments, unless expressly stated otherwise.
In another aspect, this application features systems that include: a purification unit configured to receive a groundwater sample that includes multiple organic tracer species from a petroleum containing reservoir and featuring a solid phase extraction material configured to isolate an aqueous fraction of the groundwater sample; a first chromatographic separation device that includes at least one chromatographic column configured to receive the aqueous fraction and to separate the aqueous fraction into a plurality of components, where each component corresponds to a different one of the organic tracer species; a fluid source coupled to an output of the first chromatographic separation device and configured to combine at least one lanthanide element with each of the separated components; a second chromatographic separation device that includes at least one chromatographic column configured to receive each of the separated components and to separate an excess amount of the at least one lanthanide element from the organic tracer species of each separated component; and an analyzer coupled to an output of the second chromatographic separation device and configured to measure fluorescence emission from complexes of each of the organic tracer species and the at least one of the lanthanide elements, and to determine a relative amount of each organic tracer species in the groundwater sample based on the measured fluorescence emission, where the fluid source is configured to supply the at least one lanthanide element to each of the component solutions so that a ratio of the at least one lanthanide element to the separated component in each component solution is 10:1 or greater.
Embodiments of the systems can include any one or more of the following features.
The multiple organic tracer species can include at least twenty organic tracer species (for example, at least thirty organic tracer species, at least fifty organic tracer species). The multiple organic tracer species can include 2,6-pyridine dicarboxylic acid. The multiple organic tracer species can include at least one derivative of 2,6-pyridine dicarboxylic acid. The multiple organic tracer species can include 4,7-bis(sulfonatophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid. The multiple organic tracer species can include at least one derivative of 4,7-bis(sulfonatophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid.
The purification unit can include a flow through solid phase extraction bed. The purification unit can include an ion-exchange material configured to remove one or more ionic species from the aqueous fraction. The solid phase extraction material can be configured to remove an amount of one or more chromogenic species from the aqueous fraction. A concentration of at least one of the multiple organic tracer species in the aqueous fraction can be increased relative to a concentration of the at least one species in the groundwater sample.
The first chromatographic separation device can include a first chromatography column and a second chromatography column in fluid communication with the first chromatography column, the aqueous fraction can pass through the first chromatography column to generate a first plurality of chromatographic fractions, and at least some members of the first plurality of chromatographic fractions can pass through the second chromatography column to generate the plurality of components. A chemical selectivity of the first chromatography column can be different from a chemical selectivity of the second chromatography column. At least one of the first and second chromatography columns can be a reversed-phase chromatography column. At least one of the first and second chromatography columns can be an ion-exchange chromatography column. At least one of the first and second chromatography columns can be a hydrophilic interaction chromatography column.
The at least one lanthanide element can include two or more lanthanide elements (for example, five or more lanthanide elements).
The fluid source can be configured to supply the at least one lanthanide element so that the ratio of the at least one lanthanide element to the separated component in each component solution can be 20:1 or greater (for example, 100:1 or greater).
The analyzer can include a radiation source, a detector, and a controller connected to the radiation source and to the detector, and configured to analyze each separated component by directing the radiation source to expose the separated component to illumination radiation and terminating the exposure to the illumination radiation at an initial time t0, and measuring fluorescence emission from the separated component at a time tm>t0, where (tm−t0) is 2 microseconds or more. The controller can be configured to determine relative amounts of each organic tracer species in the groundwater sample based on the measured fluorescence emission from the separated components.
Embodiments of the systems can also include any of the other features described in this application, including combinations of features described in connection with different embodiments, except as expressly stated otherwise.
As used in this document, a “derivative” of a chemical substance or moiety is a structurally modified version of that substance or moiety, which contains that substance or moiety as a structural unit. For example, derivatives of a chemical substance or moiety A can include, but are not limited to, substances that include the chemical structure of substance or moiety A functionalized at one or more atoms to include an additional chemical structure (for example, an additional group of one or more atoms). Derivatives of the chemical substance or moiety A can also include chemical substances corresponding to the structure of substance or moiety A with at least one atom removed.
In some embodiments, due to the structural similarity between chemical substance or moiety A and its derivative, both the chemical substance or moiety A and its derivative participate in similar chemical reactions.
Unless otherwise defined, all technical and scientific terms used have the same meaning as commonly understood by one of ordinary skill in the art to which this disclosure belongs. Although methods and materials similar or equivalent to those described can be used in the practice or testing of the present subject matter, suitable methods and materials are described below. All publications, patent applications, patents, and other references mentioned are incorporated by reference in their entirety. In case of conflict, the present specification, including definitions, will control. In addition, the materials, methods, and examples are illustrative only and not intended to be limiting.
The details of one or more embodiments are set forth in the accompanying drawings and the description. Other features and advantages will be apparent from the description, drawings, and claims.
Like symbols in the drawings indicate like elements.
Accurate quantitative detection of the relative amounts of large numbers of organic molecular tracers from groundwater samples taken from a petroleum containing reservoir, particularly over a relatively wide range of possible concentrations, depends in part on the linearity of the measurement signal with respect to the tracer concentration in the groundwater samples. When fluorescence emission from lanthanide-complexed tracers is used as the detection modality, if at least one property of the fluorescence emission—such as instantaneous intensity or time-integrated intensity—scales approximately linearly with tracer concentration, then measurements of fluorescence emission can be used to deduce relative concentrations of organic molecular tracers in groundwater samples. Because such tracers can be reliably distinguished, quantitation of large numbers of different tracers (for example, from many different injector wells) is possible, which allows detailed flow field maps to be constructed for a particular reservoir.
In aqueous solution, certain organic molecular tracers can form a variety of adducts with lanthanide elements. In particular, depending upon factors such as the relative concentrations of the tracers, the lanthanide elements, the presence of interfering ions that may bind to tracer cleft sites, and the solution pH, tracers can form—in addition to 1:1 complexes with lanthanide elements—various multi-coordination complexes in which the lanthanide:tracer ratio is 1:2, 1:3, 1:4, for example. Such multi-coordination complexes affect the linear scaling of fluorescence emission from lanthanide-complexed tracers, thereby making accurate quantitation of the tracers more difficult.
Other factors can also impede quantitation of molecular tracers. For example, free lanthanide ions in solution exhibit detectable fluorescence emission. Further, certain matrix components in groundwater samples (such as polycyclic aromatic hydrocarbons) also exhibit fluorescence emission. These anomalous sources of fluorescence emission can affect measurements of fluorescence emission from complexed molecular tracers, introducing error into quantitative measurements. Still further, certain matrix components of groundwater samples can act either as fluorescence quenchers or enhancers, such that measurements of fluorescence emission even from 1:1 complexed tracers may not scale linearly with tracer concentration.
The methods and systems described in this disclosure include various steps that can be used to reduce or eliminate sources of non-linearity that may affect fluorescence emission measurements from complexed tracer species, and which may therefore interfere with the accurate determination of tracer concentrations in groundwater samples.
Flow field analysis of a petroleum reservoir can be initiated by the injection of multiple organic tracer species at different corresponding injection sites or wells within the reservoir. The organic tracer species propagate through the reservoir according to complex flow patterns. Groundwater samples withdrawn at a producing site or well will typically include concentrations of multiple different organic tracer species corresponding to multiple injection sites. The concentrations of the various tracer species represent information about flow field patterns linking the injection and production sites.
In some embodiments, for example, a groundwater sample can include at least 10 different organic tracer species (such as at least 15 different species, at least 20 different species, at least 25 different species, at least 30 different species, at least 40 different species, at least 50 different species). The methods and systems described can be used to accurately quantify each of these different tracer species, providing a complete multi-dimensional picture of reservoir flow patterns.
Examples of methods for injecting and recovering organic tracer species to and from petroleum reservoirs, and examples of suitable organic tracer species, are described in U.S. Patent Application Publication No. US 2019/0226326, the entire contents of which are incorporated in this disclosure by reference.
Purification unit 200 includes a solvent reservoir 202, a sample reservoir 204, a multi-way valve 206, a SPE bed 208 (implemented as a SPE cartridge or disk, for example), an outlet 210, and a waste container 212. Valve 206 is connected to an electronic controller 150 which regulates the operation of the valve.
During operation, with valve 206 in the first position as shown in
After the sample has been loaded onto SPE bed 208, controller 150 switches valve 206 to the second position as shown in
Although purification unit 200 includes only a single solvent reservoir 202 in
Purification unit 200 can also include reservoirs for additional solvents such as wash solvents that can be used during purification of the groundwater sample. As described above, such reservoirs can be connected to SPE bed 208 through valves, conduits, and other fluid directing components under the control of controller 150.
In certain embodiments, purification unit 200 can include multiple SPE beds 208 that include different solid phase extraction media. For example, depending upon the nature of the groundwater sample, it can be advantageous in some circumstances to flow the sample through a sequence of SPE beds, each of which is designed to adsorb and retain specific components from the sample. Different elution conditions (for example, different elution solvents, solvent volumes, and times) can be used for each SPE bed under the control of controller 150. Further, because SPE beds 208 may need to be periodically replaced, by using multiple SPE beds, the beds can be individually replaced at different times as they become saturated.
The solid phase extraction medium is a chemically active stationary phase that functions to retain substances as the groundwater sample is passed over it. Different solid phase extraction media can be used for specific tracer species to yield desired retention times, depending upon the interactions between the tracer molecules and the SPE material, and the elution solvents being used.
For example, for tracer species corresponding to 2,6-pyridine dicarboxylic acid (DPA) or a derivative of DPA (examples of which are shown in
In general, a variety of different derivatives of DPA can be used as tracer species. In addition to the examples shown in
The pKa values of DPA are −1.05, 2.22 and 5.22, and the related molecular species are shown in
For tracer species based on 4,7-bis(sulfonatophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid (BSPPDA)—that is, BSPPDA and derivatives of BSPPDA—the BSPPDA molecule contains negatively charged sulphonic groups in its structure, which offers good solubility in wide pH range in water but little retention in the non-charged ISOLUTE® ENV+cartridge.
In general, a variety of different derivatives of BSPPDA can be used as tracer species. Suitable derivatives of BSPPDA that can be used as tracers include, but are not limited to, derivatives that include one, two, three, or four substituents on the phenanthroline ring system. Examples of substituents that can be present include, but are not limited to, linear and branched alkyl groups that include from 1-20 carbon atoms, linear and branched alkenyl groups that include from 1-20 carbon atoms, linear and branched alkynyl groups that include from 1-20 carbon, hydroxyl groups, amino groups, nitro groups, cyclic aliphatic groups that include from 2-20 carbon atoms and can optionally include one or more heteroatoms such as N, O, S, and P, one or more aromatic groups that include from 2-20 carbon atoms and can optionally include one or more heteroatoms such as N, O, S, and P, aliphatic carboxylic acid groups, aromatic carboxylic acid groups, aliphatic amide groups, aromatic amide groups, aliphatic aldehyde groups, aromatic aldehyde groups, aliphatic ketone groups, aromatic ketone groups, aliphatic ether groups, aromatic ether groups, and combinations of any of the foregoing groups (for example, one of the foregoing groups with one or more of the foregoing groups as substituents thereon).
In addition, suitable derivatives of BSPPDA include species in which one or both of the sulfonatophenyl moieties includes any one or more of the above groups, in addition to, or as an alternative to, groups present on the phenanthroline ring system. As discussed above in connection with the phenanthroline ring system, combinations of any of the groups discussed above can be present on one or both of the sulfonatophenyl moieties. If substituted, the sulfonatophenyl moieties can be substituted in the same manner or can be substituted differently.
As an example, for BSPPDA-based tracer species, an anionic exchange type resin, such as CHROMABOND® HR-XAW cartridge or CHROMABOND® HR-XA cartridge (both available from Macherey-Nagel, Dueren, Germany) can be used for purification of the groundwater sample. The CHROMABOND® HR-XAW cartridge contains a solid phase extraction material formed from a polystyrene-divinylbenzene-based copolymer with secondary and tertiary amine groups. The CHROMABOND® HR-XA cartridge contains a solid phase extraction material formed from a polystyrene-divinylbenzene-based copolymer with quaternary ammonium modification. At a pH<6, the quaternary ammonium, secondary and tertiary amine groups of these solid phase extraction resins are protonated to form a positively charged structure, which has strong electrostatic interactions with negatively changed BSPPDA molecules in solution. In addition, the phenanthroline ring in the molecule offers additional interactions with the polystyrene ring structure in the SPE bed through partial hydrophobic and π-π interactions. At neutral or basic conditions, the secondary and tertiary amine groups are deprotonated, weakening the interaction with BSPPDA molecules.
Accordingly, purification of the groundwater sample can be performed under acidic conditions in diluted HCl or H3PO4 solution at pH<5, preferably pH<2, and then the tracer species can be eluted from the SPE bed with a solvent such as methanol-water at basic pH. An extraction recovery of approximately 85%, starting from an initial solution of BSPPDA in a groundwater sample, at a molar concentration of 1 ppb, has been achieved.
Purifying the groundwater sample by flowing the sample over or through SPE bed 208 performs a number of advantageous functions. As described above, one such function is to separate certain sample matrix components from the organic tracer species of interest. Solid matrix components make flow-through handling of the sample difficult, and elimination of these components reduces contamination of fluidic components.
In addition, certain matrix components may be chromophoric and absorb incident radiation, thereby interfering with fluorescence emission measurements of tracer species. By reducing or eliminating such components from the aqueous fraction, a source of error in the fluorescence emission measurements can be reduced.
In some embodiments, a groundwater sample can include dissolved ions that bind specifically to cleft sites of the tracer species. If binding sites of the tracer species are occupied by these interfering ions, the binding kinetics of complex formation between the tracer species and lanthanide elements can be significantly impaired, and complex formation may even be prevented. By reducing or eliminating these interfering ions from the aqueous fraction, complex formation between tracer species and lanthanide elements (in a subsequent analysis step) is promoted, which yields more accurate quantitative determination of tracer species concentrations.
In certain embodiments, by suitable choice of elution solvent, volume, and time, the concentrations of individual tracer species in the aqueous fraction can be increased relative to their concentrations in the groundwater sample. By increasing the concentrations of tracer species, the species are detected more easily—and with higher accuracy—in fluorescence emission measurements, relative to detection without increasing the concentrations. Because it is the relative concentrations of various tracer species in the groundwater sample that provides information about fluid flow patterns within the reservoir, increasing the concentrations of tracer species in the aqueous fraction by means of solid phase extraction does not impede reservoir management operations.
In some embodiments, other methods can be used in addition to, or as alternatives to, removing matrix substances from groundwater samples using solid phase extraction materials. Such methods can include, for example, sedimentation and selective liquid-liquid extraction.
Returning to
The separation step 104 effectively accomplishes two important functions. First, separation step 104 separates individual tracer species from one another, allowing the individual species to be identified by retention time. Second, separation step 104 separates other light-absorbing matrix components of the groundwater sample from the tracer species. In general, groundwater samples include a mixture of substances, which includes tracer species and “background” substances. It has been observed that a single chromatographic separation run may not always provide a complete isolation of tracer species from background substances, especially if the matrix contains multiple different dissolved organic matter components (such as, for example, a mixture of different hydrocarbon components). Typically, chromatographic separation of tracer species is implemented by collecting eluted fractions from a chromatographic column during particular time windows. If a background substance co-elutes during a time window associate with a particular tracer species, then the background substance will be present along with the tracer species in the corresponding component solution.
Accordingly, for a particular component solution that nominally corresponds to a single tracer species, it has been observed that fluorescence emission from the solution includes a continuous background signal composed of many overlapping peaks, representing unknown background substances co-eluting along with the tracer species of interest. Examples of such background substances include polycyclic aromatic hydrocarbons and other chromophores. As a result of interference from such background substances, the measured fluorescence emission signal from a particular component solution can be enhanced or quenched by background substances.
Nonetheless, for a particular tracer species, it is generally observed that most background substances in a raw groundwater sample either can be removed from the matrix by solid phase extraction. Even if some of the background substances do adsorb on the SPE materials, they do not co-elute with the tracer species and can be separated by chromatography. As such, the effect on the fluorescence (or other) signal arising from interference by background substances following separation step 104 is generally a few orders of magnitude less than the effect that would be observed on the fluorescence signal for the raw groundwater sample.
Each of the unique organic tracer species in the aqueous fraction that is separated into components in step 104 has a different elution time (or retention time) on the chromatography column. In certain embodiments, these elution times can be measured, and used to uniquely identify the different tracer species corresponding to each component of the aqueous fraction. For example, controller 150 (which can be connected to the first chromatographic separation device, as will be discussed later) can measure the elution times of the components, and correlate or match the elution times with stored reference information to identify the tracer species associated with each component.
In some embodiments, separation step 104 can be performed using a first chromatographic separation device that includes multiple chromatography columns. For example, a two-dimensional (2D) liquid chromatography procedure can be implemented in a first chromatographic separation device that includes two chromatography columns fluidly connected in series so that some components (selective, or “heart cutting” 2D liquid chromatography) or even all components (comprehensive 2D liquid chromatography) of the aqueous fraction that are initially separated by elution from the first chromatography column are then further separated by loading onto, and elution from, the second chromatography column, as they emerge from the first chromatography column. To implement selective 2D liquid chromatography, individual components selected for loading and elution on the second chromatography column are stored in fluid loops and then individually loaded onto the second chromatography column for elution.
More generally, separation step 104 can be performed on a multi-dimensional first chromatographic separation device that includes multiple (for example, two or more, three or more, four or more, five or more, or even more) chromatography columns connected in fluidic series. In some embodiments, each of the multiple chromatography columns has the same chemical selectivity. In certain embodiments, however, at least some of the multiple chromatography columns have different chemical selectivities. Multi-dimensional chromatographic separation procedures are typically implemented to further enhance the separation of organic tracer species from one another, and from background substances in the aqueous fraction.
For example, when the first chromatographic separation device is implemented as a 2D liquid chromatography system with two chromatography columns connected in fluidic series, the chemical selectivities of the two columns can differ. Examples of suitable chemical selectivities for any one or more of the chromatography columns in a multi-dimensional first chromatographic separation device include, but are not limited to, reversed phase, ion-exchange, and hydrophilic interaction.
To demonstrate the separation of a variety of different DPA-based organic tracers using liquid chromatography procedures, a mixture of DPA and eight DPA-based derivatives of DPA was prepared and injected onto a Gemini C18 reversed-phase HPLC column (obtained from Phenomenex, Torrance, Calif.). Separation of the mixture was performed in gradient mode, where the mobile phase was composed of (A) water with 0.3% by volume phosphoric acid, and (B) methanol. The organic component (the percentage of methanol in the mobile phase) was ramped from 10% to 30% by volume over 5.5 minutes, and then stepped up to 40% by volume until 8 minutes.
In some embodiments, the separated components (the tracer species) can be identified based on their retention times in step 104. For example, controller 150 can measure the component retention times (or receive information about the retention times, for example, from a storage device or medium, or from a user interface) and, by comparing the measured retention times to reference information that includes retention times for known tracer species, identify some or all of the separated components from the groundwater sample.
Returning to
When multiple tracer molecules bind to a single lanthanide ion, the fluorescence response can be greatly enhanced. However, in general, the largest enhancement occurs generally under lanthanide-starved conditions. For example, significant enhancement for pyridine dicarboxylic acid-based tracer species occurs when the concentration of the tracer species is at least three times in excess of the lanthanide ion concentration. In practice, maintaining this balance of conditions can be challenging. Certain tracer molecules will not contribute to measured fluorescence emission under lanthanide-starved conditions as they will not be complexed to a lanthanide ion. Further, certain lanthanide:tracer complexes will include only a 1:2 or even a 1:1 ratio of lanthanide:tracer. As a result, the measured fluorescence emission signal will include a complex mixture of contributions from lanthanide:tracer adducts of different ratios, and will not include contributions from free tracer molecules. Typically, this measured signal will not scale linearly with tracer species concentrations in the groundwater sample, and may scale substantially non-linearly with tracer species concentration. This unpredictable relationship between the measured fluorescence emission properties and the tracer species concentration makes accurate quantitative determination of the tracer species concentrations difficult.
Another factor that makes maintaining of high tracer:lanthanide concentration ratios challenging is the concentrations of the tracer species in the groundwater sample. In many field measurements, tracer species are present in groundwater samples at concentrations that can be difficult to measure accurately. As a result, maintaining high tracer:lanthanide concentration ratios is impractical for many groundwater samples.
To address the foregoing challenges in step 106, an excess amount of one or more lanthanide elements is added to each component. By using excess lanthanide elements, each lanthanide ion will bind to either 1 or 0 tracer molecules. Furthermore, the lanthanide:tracer binding kinetics will be favored, such that all (or nearly all) of the tracer molecules will bind to a lanthanide ion. Accordingly, even though the overall fluorescence emission intensity is lower relative to the situation described above in which the tracer:lanthanide molar ratio is greater, the fluorescence emission intensity for each component scales approximately linearly with the amount of the corresponding tracer species in the groundwater sample, so that relative amounts (or concentrations) of each tracer species can be accurately determined.
In some embodiments, for example, one or more lanthanide elements are combined with each of the separated components from step 104 so that in each component solution, a molar ratio of the one or more lanthanide elements to the component (organic tracer species) is 5:1 or more (for example, 10:1 or more, 20:1 or more, 30:1 or more, 50:1 or more, 70:1 or more, 100:1 or more, 150:1 or more, 200:1 or more, 300:1 or more).
The lanthanide elements that are combined with the separated components from step 104 can generally include any of the elements in the lanthanide series. Examples of such elements include terbium, europium, cerium, praseodymium, neodymium, promethium, samarium, gadolinium, dysprosium, holmium, erbium, thulium, ytterbium, and lutetium.
In certain embodiments, a single type of lanthanide element is combined with each of the separated components. In some embodiments, however, more than one type of lanthanide element is combined with the separated components. A first lanthanide element can be combined with some of the separated components, while a second lanthanide element can be combined with other separated components. In general, any number of lanthanide elements (for example, two or more, three or more, four or more, five or more, six or more, eight or more) can be combined with the components separated in step 104.
After lanthanide:tracer complexes have been formed in step 106, excess unbound lanthanide elements are removed in step 108. In general, step 108 is performed by a second chromatographic separation device, which can include one or more chromatography columns as will be described further below.
Free lanthanide ions generate measurable fluorescence emission, and although the emission is significantly weaker than the fluorescence emission from lanthanide:tracer complexes, spectral interference from free lanthanide emission can nonetheless affect the accurate quantitative measurement of tracer species in groundwater samples. In effect, fluorescence emission from free lanthanide ions sets up a noise floor for the limit of detection (LOD) for tracer species at trace and ultra-trace concentrations.
Improving the LOD involves reducing the noise floor (contributions to the measured tracer signals arising from free lanthanide ions in the component solutions). In turn, reducing the noise floor due to free lanthanide ions can be accomplished by separating free lanthanide ions from the components (organic tracer species) in the component solutions by performing a liquid chromatographic separation using the second chromatographic separation device.
As an example, separation of free terbium ions from Tb:DPA was performed using hydrophilic interaction liquid chromatography (HILIC). To maintain the stability of Tb:DPA complexes, a buffered ammonium acetate solution (200 millimoles per Liter (mM), pH 5.6) was used as the mobile phase, employing acetonitrile as the organic modifier. Separations were performed using 55% acetonitrile at 30° C. and 1 mL/min flow rate.
Three types of stationary phase functionalities were evaluated and compared for their separation performance: (a) bare silica, (b) poly-hydroxy fructan, and (c) zwitterionic. All of these columns were obtained from Agilent Technologies, as the Poroshell 120 HILIC, Poroshell HILIC-OH5, and Poroshell 120 HILIC-Z columns, respectively.
To detect DPA:Tb complexes, an inline fluorimeter was coupled to the column output. The fluorimeter provided illumination radiation at a wavelength of 270 nanometers (nm), and measured fluorescence emission at 540 nm. Free Tb ions were detected by measuring UV absorption at an illumination wavelength of 230 nm.
As discussed above in connection with step 104, the separation operation implemented in step 108 can be a one-dimensional or multi-dimensional chromatographic separation, and the second chromatographic separation device can include one or more chromatographic columns. Combinations of any one or more of the different types of columns and chemical sensitivities can be used in the second chromatographic separation device. Where the device includes multiple columns, the multiple columns can be connected in fluidic series, and some or all of the fractions that elute from an earlier column in the series can be loaded onto and eluted onto later columns in the series. In particular, for workflows in which multiple different types of lanthanide ions are used to complex different tracer species, multi-dimensional chromatographic separations may be advantageous (but are not required) to separate all of the different types of free lanthanide ions from the complexed tracer species.
Returning to
A variety of different methods can be used to analyze the complexed tracer species. In some embodiments, for example, fluorescence emission (measured by an analyzer that receives each of the component solutions in a flow-through manner from the second chromatographic separation device) is used for detection. Alternatively, in certain embodiments, absorption measurements (for example, ultraviolet (UV) absorption, measured in a flow-through manner) is used for detection. As another alternative, in some embodiments, eluted components are identified by mass spectrometry.
In some embodiments, non-gated fluorescence or other measurements are performed in which each component solution is illuminated with incident radiation and fluorescence emission, absorption, or another quantity is measured (either a peak measurement or an integrated measurement) over a relatively wide temporal window that includes the time at which illumination occurs.
In certain embodiments, time-gated fluorescence measurements are performed. Aspects of such measurements are described in detail in U.S. Patent Application Publication No. US 2019/0226326. Following flash excitation to illuminate a component solution with incident radiation—a process that ends at a time to—the lanthanide:tracer complexes has a relatively long excited state lifetime. For example, in some embodiments, a complex's luminescence occurs for a time period greater than 5 microseconds (for example, greater than 8 microseconds, greater than 10 microseconds, greater than 15 microseconds, greater than 20 microseconds, greater than 30 microseconds, greater than 40 microseconds, greater than 50 microseconds, greater than 100 microseconds, greater than 200 microseconds, greater than 500 microseconds, greater than 1 millisecond (ms), greater than 2 ms, greater than 5 ms, greater than 10 ms) after excitation is complete.
Without wishing to be bound by theory, it is believed that the long excited state lifetime occurs due to parity-forbidden transitions between lanthanide ion excited and ground states, which occur more slowly than parity-allowed state-to-state transitions during excitation and further, more slowly than background fluorescence arising from parity allowed excited-to-ground state transitions among other substances that may be present in a component solution.
When a lanthanide ion is in close proximity to a sensitizing chromophore (such as a tracer molecule) that either has a strong dipole moment or is anionic such that it can associate ionically with the lanthanide ion, significant enhancement of the lanthanide ion luminescence occurs because the sensitization process circumvents the LaPorte selection rules that normally forbid f-f transitions.
Accordingly, the difference in time scales separates fluorescence or luminescence signals from the lanthanide:tracer complexes (which occur at later times) from fluorescence signals due to other substances that may be present in the component solutions (“background” fluorescence). The difference in excited state lifetimes and concomitant difference in the temporal evolution of emission from the lanthanide:tracer complexes and background substances allows gating of the optical detector such that the detector can be used to measure luminescence from the complex alone, after fluorescence from background substances in the component solution has decayed.
Specifically, by exposing each component solution to incident radiation and terminating the exposure at an initial time t0, and then measuring fluorescence emission beginning at a later time tm, the measured fluorescence emission signal—when (t0-tm) is chosen appropriately—can correspond nearly entirely to fluorescence emission from lanthanide:tracer complexes.
In general, it has been found that by choosing (t0-tm) to be 2 microseconds or more (2.5 microseconds or more, 3 microseconds or more, 3.5 microseconds or more, 4 microseconds or more, 4.5 microseconds or more, 5 microseconds or more), the measured fluorescence emission signal contains contributions from only the lanthanide:tracer complexes in component solutions, without measurable interference from lanthanide luminescence.
After fluorescence emission has been measured in step 110, a controller (for example, controller 150) determines relative amounts or concentrations of each of the tracer species in the groundwater samples based on the measurements. For example, calibration information (such as a calibration curve) or stored reference information can be used to convert measurements of gated, time-integrated fluorescence intensity for each tracer species to relative quantities or concentration values for the tracer species. The relative quantities or concentration values can then optionally be used by controller 150 (or another processing unit) to execute a variety of reservoir reporting and management operations.
Relative quantities or concentration values can be determined from fluorescence or other measurements in various ways. In some embodiments, peak areas (that is, the area under each peak) are used to calculated relative quantities or concentrations of each tracer species. To convert these relative quantities or concentrations to absolute values, reference information (in the form of previously measured concentration values or a calibration curve) is used. For example, a calibration curve can be determined by measuring fluorescence emission or other signals for solutions with known concentrations of tracer species, so that peak area is determined as a function of known concentration for each tracer species of interest. By establishing a functional relationship between peak area and concentration (for example, by regression analysis) for each tracer species, a calibration curve is generated. Subsequently, measured peak areas for solutions of unknown tracer species concentration can be used to determine absolute tracer species concentration values using the calibration curve.
The procedure shown in
(i) Solvent and Buffer
It has been determined that the nature of the solvent (including the solvent's buffered pH) may play an important role in ensuring that fluorescence emission measurements of the component solutions can be used to accurately determine relative amounts or concentrations of tracer species in a groundwater sample. To investigate the role of the solvent, fluorescence emission measurements were performed in 1.0 mol/L (M) bioluminescence grade sodium acetate buffer (pH adjusted with 1.0 M acetic acid to pH˜5.6), and in 1.0 M HPLC-grade ammonium acetate (pH adjusted with 1.0 M acetic acid to pH˜5.7) to evaluate the background fluorescence intensity against fluorescence emission from a trace amount of a molecular tracer species (molar concentration of 100 parts per trillion (ppt) of BSPPDA, and molar concentration of 1 ppb of DPA). Lanthanide ion concentrations were varied between 5×10−6 M and 5×10−8 M for europium ions in the analysis with BSPPDA, and between 5×10−6 M and 1×10−7 M for terbium in the analysis with DPA.
Similarly,
Various literature sources (see, for example, Hindle et al., “Dipicolinic acid (DPA) assay revisited and appraised for spore detection,” Analyst 124: 1599-1604 (1999)), have reported that sodium acetate buffer is preferred for stabilizing the fluorescence of lanthanide ions bound at a 1:1 ratio to sensitizing ligands (such as the tracer species), because they effectively exclude water from complexing the lanthanide ions, thereby ensuring that the ions remain complexed by the tracer species.
However, the measurements shown in
The effect of the solvent's buffered pH on the linearity of the fluorescence emission response was also investigated with tracer molecules at trace and ultra-trace concentration levels. Lanthanide ion concentrations were maintained constant (1×10−7 M for europium and 1×10−6 M for terbium), and fluorescence emission measurements were performed in 1.0 M ammonium acetate and 1.0 M sodium acetate solutions.
In
The compatibility of the solvent buffer solution used to introduce lanthanide species with the chromatography mobile phase can also affect the measured fluorescence emission. The mobile phase in reversed-phase chromatography system typically consists of a mixture of aqueous and organic solvents, of which acetonitrile is a common choice, and is acidified to approximately pH 2. As discussed above, to realize improved fluorescence emission from tracer:lanthanide complexes, the solvent mixture generally includes a relatively high concentration of acetate ions, which displace water molecules from the lanthanide ion so that the lanthanide luminescence is not quenched by O—H vibration of the coordinated water molecules. In addition, the solvent mixture typically has a pH of between approximately 5 and 6.
Static experiments showed that a 1 M sodium acetate buffer is miscible with up to 30% (by volume) acetonitrile. Using a crystal violet indicator to determine miscibility (which has high solubility in acetonitrile and moderate solubility in sodium acetate buffer), a phase separation was observed once the acetonitrile content exceeded 30% by volume, although no precipitation was visible.
In some embodiments, the mobile chromatography phases used in the methods described herein can include acetate ions at concentrations of between 0.5 M and 3.0 M (for example, between 0.5 M and 2.8 M, between 0.5 M and 2.6 M, between 0.5 M and 2.4 M, 0.5 M and 2.2 M, between 0.5 M and 2.0 M, between 0.5 M and 1.8 M, between 0.5 M and 1.6 M, between 0.5 M and 1.5 M, between 0.7 M and 3.0 M, between 0.7 M and 2.8 M, between 0.7 M and 2.6 M, between 0.7 M and 2.4 M, between 0.7 M and 2.2 M, between 0.7 M and 2.0 M, between 0.9 M and 3.0 M, between 0.9 M and 2.8 M, between 0.9 M and 2.6 M, between 0.9 M and 2.4 M, between 0.9 M and 2.4 M, between 0.9 M and 2.2 M, between 0.9 M and 2.0 M, or any other range between 0.5 M and 3.0 M).
In certain embodiments, in any of the mobile chromatography phases described above, acetonitrile can be present at a concentration of between 0% and 30% (for example, between 2% and 30%, between 4% and 30%, between 6% and 30%, between 8% and 30%, between 10% and 30%, between 12% and 30%, between 14% and 30%, between 16% and 30%, between 18% and 30%, between 20% and 30%, between 25% and 30%, or any other range between 0% and 30%).
In addition to acetonitrile, the chromatography methods used for quantifying lanthanide tracers typically also include phosphoric acid to acidify the aqueous component, which protonates the carboxylic acid groups in the tracer ligands and promotes retention on the chromatography column.
The compatibility of phosphoric acid with the in-line mixed lanthanide solution was investigated in connection with the methods described in this disclosure. The strong binding of the phosphate anion to a lanthanide ion is known to quench its fluorescence intensity. This would have a detrimental effect on inline detection, since the mobile phase from the chromatography column(s) is typically mixed directly with the lanthanide solution prior to detection.
To assess the effect of phosphoric acid on fluorescence emission measurements, a static fluorescence experiment was carried out using dipicolinic acid (DPA) as a model tracer to investigate the effects of both acetonitrile and phosphoric acid on the fluorescence intensity.
As an alternative to phosphoric acid, other acids can also be used to acidify the mobile chromatography phase. For example, acids commonly used in reversed-phase chromatography are acetic acid, formic acid, and trifluoroacetic acid (TFA). Acetic acid and formic acid were previously determined to be too weak to achieve a mobile phase pH of 2-3, so TFA was selected as an alternative. The same static experiment described above was carried out using 0.1% and 0.5% (by volume) TFA in a sodium acetate buffer containing 10−6 M TbCl3.
In general, in any of the solution mixtures described herein that are used to measure fluorescence intensity of tracer complexes, the volumentric concentration of TFA can be between 0.1% and 3.0% (for example, between 0.1% and 2.8%, between 0.1% and 2.6%, between 0.1% and 2.4%, between 0.1% and 2.2%, between 0.1% and 2.0%, between 0.1% and 1.8%, between 0.1% and 1.6%, between 0.1% and 1.4%, between 0.1% and 1.2%, between 0.1% and 1.0%, between 0.1% and 0.8%, between 0.1% and 0.6%, between 0.1% and 0.5%, between 0.2% and 3.0%, between 0.2% and 2.5%, between 0.2% and 2.0%, between 0.2% and 1.5%, between 0.2% and 1.0%, between 0.2% and 0.5%, between 0.3% and 3.0%, between 0.3% and 2.5%, between 0.3% and 2.0%, between 0.3% and 1.5%, between 0.3% and 1.0%, between 0.3% and 0.5%, between 0.4% and 3.0%, between 0.4% and 2.5%, between 0.4% and 2.0%, between 0.4% and 1.5%, between 0.4% and 1.0%, between 0.4% and 0.5%, or any other range between 0.1% and 3.0%).
Buffer composition was also investigated to determine its effect on fluorescence measurement sensitivity. A series of static experiments was carried out comparing the sodium acetate buffer at pH 5.6 with (4-(2-hydroxyethyl)-1-piperazineethanesulfonic acid) (HEPES) buffers at pH 6.5 and 7.3. For each buffer solution, fluorescence emission intensity (integrated area under fluorescence emission peak) was determined as a function of DPA concentration for a Tb-DPA complex.
The measurements shown in
In some embodiments, the solution mixture used to measure fluorescence emission from the tracer complexes described in this disclosure can have a pH of between 5.0 and 8.0 (for example, between 5.0 and 7.5, between 5.0 and 7.0, between 5.5 and 8.0, between 5.5 and 7.5, between 5.5 and 7.0, between 6.0 and 8.0, between 6.0 and 7.5, between 6.5 and 8.0, between 6.5 and 7.5, or any other range between 5.0 and 8.0).
As discussed above, in certain embodiments, the solution mixture used to measure fluorescence emission from the tracer complexes can include acetate ions. For example, the solution mixture can include acetate ions as part of a buffered composition. Acetate ions can be present as a sodium acetate buffered solution, an ammonium acetate buffered solution, or as part of a buffer composition in combination with one or more other cations such as, but not limited to, calcium, potassium, lithium, and magnesium.
(ii) Lanthanide Ion Concentration
Further investigations were performed to determine—with ammonium acetate used as the solvent for conducting fluorescence emission measurements—appropriate lanthanide ion concentrations to maintain a desired signal-to-noise ratio in the fluorescence emission measurements. Fluorescence measurements were performed in ammonium acetate solutions at various DPA:Tb concentration ratios (with 1:100 as the lowest ratio), with the Tb ion concentration held constant in each solution.
In some embodiments, to achieve suitable signal-to-noise ratios for measurement of fluorescence emission from tracer complexes, the Tb ion concentration is maintained between 1×10−4 M and 1×10−6 M (for example, between 9.5×10−5 M and 1.5×10−6 M, between 9.0×10−5 M and 2.0×10−6 M, between 8.5×10−5 M and 2.5×10−6 M, between 8.0×10−5 M and 3.0×10−6 M, between 7.5×10−5 M and 3.5×10−6 M, between 7.0×10−5 M and 4.0×10−6 M, between 6.5×10−5 M and 4.5×10−6 M, between 6.0×10−5 M and 5.0×10−6 M, between 5.5×10−5 M and 5.5×10−6 M, between 5.0×10−5 M and 6.0×10−6 M, between 4.5×10−5 M and 6.5×10−6 M, between 4.0×10−5 M and 7.0×10−6 M, between 3.5×10−5 M and 7.5×10−6 M, between 3.0×10−5 M and 8.0×10−6 M, between 2.5×10−5 M and 8.5×10−6 M, between 2.0×10−5 M and 9.0×10−6 M, or any other range between 1×10−4 M and 1×10−6 M).
Fluorescence measurements were also performed in ammonium acetate solutions at various BSPPDA:Eu concentration ratios (with 1:10 as the lowest ratio), with the Eu ion concentration held constant in each solution.
In certain embodiments, to achieve suitable signal-to-noise ratios for measurement of fluorescence emission from tracer complexes, the Tb ion concentration is maintained between 1 30×10−5 M and 1×10−7 M (for example, between 9.5×10−6 M and 1.5×10−7 M, between 9.0×10−6 M and 2.0×10−7 M, between 8.5×10−6 M and 2.5×10−7 M, between 8.0×10−6 M and 3.0×10−7 M, between 7.5×10−6 M and 3.5×10−7 M, between 7.0×10−6 M and 4.0×10−7 M, between 6.5×10−6 M and 4.5×10−7 M, between 6.0×10−6 M and 5.0×10−7 M, between 5.5×10−6 M and 5.5×10−7 M, between 5.0×10−6 M and 6.0×10−7 M, between 4.5×10−6 M and 6.5×10−7 M, between 4.0×10−6 M and 7.0×10−7 M, between 3.5×10−6 M and 7.5×10−7 M, between 3.0×10−6 M and 8.0×10−7 M, between 2.5×10−6 M and 8.5×10−7 M, between 2.0×10−6 M and 9.0×10−7 M, or any other range between 1×10−5 M and 1×10−7 M).
In some embodiments, to achieve suitable signal-to-noise ratios for measurement of fluorescence emission from tracer complexes, the lanthanide ion concentration is maintained between any of the foregoing combinations of range limiting values discussed above. For example, the lanthanide ion concentration can be maintained between 1×10−4 M and 1×10−7 M (such as between 0.5×10−4 M and 2.0×10−7 M, between 0.5×10−4 M and 3.0×10−7 M, between 0.5×10−4 M and 4.0×10−7 M, between 0.5×10−4 M and 5.0×10−7 M, between 0.5×10−4 M and 6.0×10−7 M, between 0.5×10−4 M and 7.0×10−7 M, between 0.5×10−4 M and 8.0×10−7 M, between 0.5×10−4 M and 9.0×10−7 M, between 0.5×10−4 M and 1.0×10−6 M, between 0.5×10−4 M and 2.0×10−6 M, between 0.5×10−4 M and 3.0×10−6 M, between 0.5×10−4 M and 4.0×10−6 M, between 0.5×10−4 M and 5.0×10−6 M, between 0.5×10−4 M and 6.0×10−6 M, between 0.5×10−4 M and 7.0×10−6 M, between 0.5×10−4 M and 8.0×10−6 M, between 0.3×10−4 M and 2.0×10−7 M, between 0.3×10−4 M and 3.0×10−7 M, between 0.3×10−4 M and 4.0×10−7 M, between 0.3×10−4 M and 5.0×10−7 M, between 0.3×10−4 M and 6.0×10−7 M, between 0.3×10−4 M and 7.0×10−7 M, between 0.3×10−4 M and 8.0×10−7 M, between 0.3×10−4 M and 9.0×10−7 M, between 0.3×10−4 M and 1.0×10−6 M, between 0.3×10−4 M and 2.0×10−6 M, between 0.3×10−4 M and 3.0×10−6 M, between 0.3×10−4 M and 4.0×10−6 M, between 0.3×10−4 M and 5.0×10−6 M, between 0.3×10−4 M and 6.0×10−6 M, between 0.3×10−4 M and 7.0×10−6 M, between 0.3×10−4 M and 8.0×10−6 M, between 0.7×10−4 M and 2.0×10−7 M, between 0.7×10−4 M and 3.0×10−7 M, between 0.7×10−4 M and 4.0×10−7 M, between 0.7×10−4 M and 5.0×10−7 M, between 0.7×10−4 M and 6.0×10−7 M, between 0.7×10−4 M and 7.0×10−7 M, between 0.7×10−4 M and 8.0×10−7 M, between 0.7×10−4 M and 9.0×10−7 M, between 0.7×10−4 M and 1.0×10−6 M, between 0.7×10−4 M and 2.0×10−6 M, between 0.7×10−4 M and 3.0×10−6 M, between 0.7×10−4 M and 4.0×10−6 M, between 0.7×10−4 M and 5.0×10−6 M, between 0.7×10−4 M and 6.0×10−6 M, between 0.7×10−4 M and 7.0×10−6 M, between 0.7×10−4 M and 8.0×10−6 M, and any other range between 1×10−4 M and 1×10−7 M).
(iii) Lanthanide Ions and Tracer Species
In the preceding description, multiple different types of lanthanide elements can be added in step 106 to the separated components from step 104. In some embodiments, however, steps 106 and 108 can be combined, and a separation step can be performed after each type of lanthanide element is added to the separated components. For example, a first type of lanthanide element can be combined with the separated components as described above, and then each of the component solutions can be purified to remove excess quantities of the first type of lanthanide element using the second chromatographic separation device as described above. Next, a second type of lanthanide element can be added to each of the purified, separated components as described above, and then each of the component solutions can then be purified again to remove excess quantities of the second type of lanthanide element using the second chromatographic separation device. A similar procedure can be repeated for third, fourth, fifth, and additional different types of lanthanide elements that are added to the separated components.
This type of work flow in which different types of lanthanide elements are added sequentially can have certain advantages. Certain tracer species will bind to, but not sensitize, certain types of lanthanide ions. For example, BSPPDA generally will bind but not sensitize terbium ions. Different tracer species can then be distinguished based on their selective sensitization of different types of lanthanide ions.
During operation of the system, purification unit 200 removes matrix and other substances from the groundwater sample. As described above, in some embodiments, purification unit 200 includes a solid phase extraction material onto which matrix substances are adsorbed. An aqueous fraction of the groundwater sample is first adsorbed onto, and then eluted from, the solid phase extraction material. The aqueous fraction is then conveyed from an outlet of purification unit 200 to an inlet of first chromatographic separation device 1302 via conduit 1320.
First chromatographic separation device 1302 can include one or more chromatographic columns, as described above. If multiple columns are present, they can be connected in series so that some or all of the components eluted from an earlier column in the series can be selectively loaded onto and eluted from one or more later columns in the series.
Fluid source 1304 includes one or more fluid reservoirs 1305, and during operation of system 1300, delivers fluids that include lanthanide elements to the separated components generated by first chromatographic separation device 1302.
During operation of fluid source 1304, controller 150 adjusts flow regulator 1506 so that a discrete volume of fluid that includes one or more lanthanide elements is delivered from fluid reservoir 1305 into conduit 1503. The lanthanide element(s) form complexes with tracer species in the component solutions flowing through conduit 1503 as described above, yielding component solutions with tracer:lanthanide complexes. These solutions emerge from outlet 1504 and are transported to second chromatographic separation device 1306 through conduit 1324.
In some embodiments, system 1300 includes a fluid mixer 1520. Fluid mixer 1520 can be implemented as part of fluid source 1304, as shown in
The manner in which the fluids are mixed in fluid mixer 1520 to achieve chelation of tracer ligands to lanthanide species can be an important aspect of system 1300. In particular, higher intensity fluorescence signals are generally measured as the extent of chelation of the lanthanide species by the tracer ligands increases. However, if mixing significantly increases temporal dispersion of analytes, then subsequent chromatographic sensitivity (peak height) and/or resolution can be compromised. If chromatographic separations of tracer complexes occur at lower resolution, fluorescence signal intensities can be reduced.
A variety of different fluid mixers can be used in system 1300. In some embodiments, in-line chelation can be accomplished using a post-chromatographic column reactor, which dampens pulsations from a fluid pump driving fluid flow. In addition, the reactor includes a mixing chamber that allows for sufficient residence time to complete chelation.
During operation, the tracer-containing effluent solution and the lanthanide-containing solutions are mixed in mixing chamber 1910. After mixing, the combined solution is delivered to outlet 1914 via conduit 1912, from which the combined solution is transported to second chromatographic separation device 1306.
Certain commercially available reactors can be used in system 1300 to mix tracer-containing and lanthanide-containing solutions. An example of such a reactor is the Metrohm post column reactor 6.2836.000 (available from Metrohm AG, Germany).
In certain embodiments, a fluidic junction can be used as a fluid mixer. Examples of such junctions include, but are not limited to, T-shaped junctions, Y-shaped junctions, and fluid-loop junctions.
In some embodiments, the fluid mixer can be implemented as a mixing chip. A variety of mixing chips with different configurations can be used. As one example, in certain embodiments, fluid mixer 1520 can be implemented as a microfluidic chip with a teardrop-shaped mixing channel.
To evaluate the feasibility of using different fluid mixers, three different fluid mixing devices were investigated: a mixing T-junction, similar to junction 1920 in
For the mixing T-junction, a pulsation dampener was constructed by connecting a short piece of 0.5 mm capillary, capped at one end, to the outlet of a pump driving transport of the lanthanide-containing solution, using a tee. The other side of the tee was connected to the T-junction fluid mixer with a 0.12 mm capillary to increase the backpressure on the pump and reduce ripple. The outlet of the T-junction fluid mixer was connected to a detector flow cell, and then to waste.
For the post-chromatography column reactor, to reduce peak dispersion, the reactor was installed in such a way to minimize the length of capillary between the column and reactor, and between the reactor and the detector. A wide capillary (0.5 mm diameter) was used at the reactor outlet to reduce the backpressure on the reactor.
Based on static experiments, suitable conditions for fluorescence measurement of DPA concentration were 1 M acetate buffer, with 10−6 M Tb3+ and up to 30% acetonitrile. Therefore, the final solution after mixing was targeted to have these characteristics. The final solution is prepared by mixing a lanthanide-containing solution and the effluent solution (the chromatographic mobile phase) from the first chromatographic separation device, according to a mixing ratio between two solutions. In principle, the optimum mixing ratio maximizes chelation of the lanthanide species by the tracer ligands, while ensuring that dilution of the chromatographic mobile phase is minimized. Three different mixing ratio conditions, shown in Table 1, were investigated, while ensuring that the final lanthanide and sodium acetate concentrations remained constant.
It was observed that mixing ratios did not measurably affect the peak intensity of the measured fluorescence emission. However, at a mixing ratio of 2:1, the baseline fluctuation was reduced in comparison to the 1:1 and 1:4 mixing ratios. This mixing ratio was selected for measurement of fluorescence emission intensity in certain experiments, as described below.
As discussed previously, it was determined that up to 0.5% TFA and up to 30% acetonitrile can be added to the buffer solution without significantly impacting the fluorescence sensitivity. Therefore, to assess the limit of detection (LOD) of the system, the aqueous phase (A) selected for chromatographic separation was water with TFA, and the organic phase (B) was acetonitrile. Additionally, it was determined that a TFA concentration of only 0.1% would be adequate for the chromatographic separation operation, giving a pH of less than 2.2 which is sufficient to protonate both carboxylic acid groups of DPA.
An Agilent Poroshell EC-C18 column, 2.1×150 mm, maintained at a temperature of 40° C., was used to chromatographically separate tracer complexes for purposes of assessing LOD. The high performance liquid chromatography (HPLC) separation was run in gradient mode, starting from 10% B, and increasing to 30% B over a time of 10 minutes. The column was then flushed with 100% B for 5 minutes. Due to the immiscibility of high acetonitrile concentrations with the acetate buffer, the buffer pump was programmed to stop during the flushing part of the run, and resume during the post-run equilibration for 2 minutes.
To assess the LOD when using the microfluidic mixing chip shown in
To improve the sensitivity of the system, the mixing chip was replaced by the Metrohm post column reactor 6.2836.000 discussed above. This increased post-mixing residence time from about 0.03 min to about 0.5 min. Also, a degassing membrane filter (obtained from Whatman, Maidstone, UK) was added to remove dissolved gas bubbles from the buffer solution. The acetate buffer solution was replaced with a 1 M sodium acetate solution (pH approximately 9), and the Tb3+ concentration was adjusted to 5 μM. The flow rate of the acetate solution was lowered to 0.5 mL/min.
Returning to
As described above, in some embodiments, analyzer 1308 is configured to measure fluorescence emission from the separated components of the sample and can include a radiation source 1310 and a detector 1312, each of which is connected directly or indirectly to controller 150. During operation of analyzer 1308, controller 150 directs radiation source 1310 to expose each separated component to illumination radiation and directs detector 1312 to measure fluorescence emission from each separated component as described above. The raw fluorescence emission measurement information is transmitted to controller 150, which analyzes the information to determine relative amounts or concentrations of the tracer species in the groundwater sample.
Suitable radiation sources include, but are not limited to, incandescent sources, fluorescent sources, LED-based sources, laser diodes, lasers including, but not limited to, gas-based lasers, solid-state lasers, and dye-based lasers, metal-vapor and metal halide lamps, flash lamps, and other common radiation sources. Suitable detectors include, but are not limited to, photodiodes, charge-coupled devices (CCDs), CMOS-based detectors, photomultipliers, and arrays of any of the foregoing devices.
To assess the effectiveness of in-line fluid mixers independent of the chromatographic separation devices, samples of the mobile phase stream from the first chromatographic separation device 1302 (with no tracer species present) were injected with 50 mL aliquots of DPA solutions of differing DPA concentration. The mobile phase stream was a 90:10 mixture of (A) 0.1% TFA in water, and (B) acetonitrile. The mobile phase stream flowed into the post column reactor at a rate of 0.5 mL/min. A lanthanide-containing solution of a 1 M sodium acetate buffer (at a pH of approximately 9) and containing TbCl3 at a concentration of 5 μM was introduced into the post column reactor at a flow rate of 0.5 mL/min. Following mixing of the solutions, fluorescence emission measurements were performed on each sample as described above.
Any of the sample processing steps and control functions described above can be executed by controller 150, or by another system controller, connected to any of the system components discussed herein, and to other components such as pumps, valves, optical components, and fluid handling components. In general, controller 150 can include a single electronic processor, multiple electronic processors, one or more integrated circuits (such as application specific integrated circuits), and any combination of the foregoing elements. Software instructions and hardware-based instructions are executed by controller 150 to perform the steps and functions discussed above. As shown in
While this specification contains many specific implementation details, these should not be construed as limitations on the scope of the disclosure or on the scope of what may be claimed, but rather as descriptions of features that may be specific to particular embodiments. Features that are described in this specification in the context of separate embodiments can also be implemented in combination in a single embodiment unless expressly stated otherwise. Conversely, various features that are described in the context of a single embodiment can also be implemented in multiple embodiments separately or in any suitable subcombination. Moreover, although features may be described above as acting in certain combinations and even initially be claimed as such, one or more features from a claimed combination can be excised from the combination, and the claimed combination may be directed to a subcombination or variation of a subcombination.
Similarly, while operations may be depicted in the drawings in a particular order, this should not be understood as requiring that such operations be performed in the particular order shown or in sequential order, or that all illustrated operations be performed, to achieve desirable results. Moreover, the separation of various system modules and components in the embodiments described above should not be understood as requiring such separation in all embodiments. While particular embodiments have been described above, other embodiments are within the scope of the following claims.
This application claims priority to U.S. Provisional Patent Application No. 62/878,281, filed on Jul. 24, 2019, the entire contents of which are incorporated herein by reference.
Number | Date | Country | |
---|---|---|---|
62878281 | Jul 2019 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 16938801 | Jul 2020 | US |
Child | 18079644 | US |