The described embodiments relate generally to tunable light source substance detection. More particularly, the described embodiments relate to systems, methods and apparatuses for tunable light source cavity detection using a plurality of axial-plus-transverse modes.
There are many applications and markets today where much effort is being invested to develop sensor technologies that have greater performance in terms of sensitivity (parts per million or billion) and specificity (unambiguously detecting identified molecules) while offering the economics needed to deploy in a large scale to provide both qualitative and quantitative data to improve safety, their infrastructure and to better the environment. One example of such a need is around natural gas sensing (methane), a common energy source around the world. In the United States there are over 68 million homes and over 6 million buildings and thousands of factories that rely on natural gas. This gas is delivered via over 2.4 million miles of main gas pipelines having distributing pipes extending to homes/buildings and factories. This is a common means of distributing energy in major cities around the world. As with any infrastructure, the pipes continue to age and are prone to hazardous leaks. Natural gas is a highly combustible odorless and colorless hydrocarbon gas largely composed of methane. Gas leaks present a serious safety and environment hazard and much effort has been invested in developing higher sensitivity and specificity sensors. The need to deploy sensors extending to residential homes and buildings has placed price/performance criteria on sensors that are not currently satisfied by existing technologies. The primary performance concerns with existing commercial sensors are that they do not have specificity and hence generate false positives to common household agents; level of detection is not sufficient and the range of sensing is limited.
The Oil and Gas industry also utilize a wide variety of gases that present safety and environmental issues if not widely monitored for leaks. In the United State, Leak Detection And Repair (LDAR) programs are required as part of the Federal Regulations. A LDAR program is a facility's system of procedures used to locate and repair leaking components (e.g., valves, pumps, connectors, compressors, and agitators) to minimize fugitive Volatile Organic Compounds (VOC) and Hazardous Air Pollutants (HAP) emissions including but not limited to methane.
It is desirable to have methods apparatuses, and systems for tunable light source cavity detection using a plurality of axial-plus-transverse modes.
An embodiment includes a system for detecting a substance. The system includes a tunable light source, an optical cavity, a cavity detector, and a processor. The tunable light source generates a beam of electro-magnetic radiation, wherein a wavelength of the beam of electro-magnetic radiation is tuned to operate at a plurality of wavelengths. The optical cavity receives the beam of electro-magnetic radiation, wherein physical characteristics of the optical cavity define a plurality of allowed axial-plus-transverse electro-magnetic radiation modes, wherein only a subset of the plurality of allowed axial-plus-transverse electro-magnetic radiation modes are excited when the optical cavity receives the beam of electro-magnetic radiation. The cavity detector senses electro-magnetic radiation emanating from the optical cavity. The processor operates to receive information relating to the sensed electro-magnetic radiation, and detect the substance within the optical cavity based on at least one of an intensity, an amplitude, a phase, or an amplitude and phase of the sensed electro-magnetic radiation emanating from the optical cavity at one or more of the plurality of wavelengths while the optical cavity receives the beam of electro-magnetic radiation.
Another embodiment includes a method of detecting a substance. The method includes generating, by a tunable light source, a beam of electro-magnetic radiation, wherein a wavelength of the beam of electro-magnetic radiation is tuned to operate at a plurality of wavelengths, receiving, by an optical cavity, the beam of electro-magnetic radiation, wherein physical characteristics of the optical cavity define a plurality of allowed axial-plus-transverse electro-magnetic radiation modes, wherein only a subset of the plurality of allowed axial-plus-transverse electro-magnetic radiation modes are excited when the optical cavity receives the beam of electro-magnetic radiation, sensing, by a cavity detector, electro-magnetic radiation emanating from the optical cavity, receiving, by a processor, information relating to the sensed electro-magnetic radiation, and detecting the substance within the optical cavity based on at least one of an intensity, an amplitude, a phase, or an amplitude and phase of the sensed electro-magnetic radiation emanating from the optical cavity at one or more of the plurality of wavelengths while the optical cavity receives the beam of electro-magnetic radiation.
Other aspects and advantages of the described embodiments will become apparent from the following detailed description, taken in conjunction with the accompanying drawings, illustrating by way of example the principles of the described embodiments.
The embodiments described include methods, apparatuses, and systems for tunable light source cavity detection using a plurality of allowed axial-plus-transverse modes.
For at least some embodiments, the subset of the plurality of the allowed axial-plus-transverse electro-magnetic radiation modes is excited by mode mismatching a waist size and/or a waist location of the incoming beam of electro-magnetic radiation to the allowed axial-plus-transverse modes that are defined by the optical cavity. For an embodiment, the incoming beam of electro-magnetic radiation comprises more than one transverse mode. For at least some embodiments, the beam of electro-magnetic radiation in the cavity impacts one location on each mirror that defines the cavity.
A first implementation could include cavity-based implementations that utilize the amplitude of a single axial-plus-transverse electro-magnetic radiation mode of the cavity which is enhanced using a single wavelength laser. Further, the absorption of one axial-plus-transverse electro-magnetic radiation mode is quantified, when the single wavelength laser is no longer enhancing (the laser diode is turned off and/or not transmitting electro-magnetic radiation) the axial-plus-transverse electro-magnetic radiation mode of the cavity. A second implementation could utilize a single wavelength laser so that the interaction between the laser and the cavity always excites the entire plurality of allowed axial-plus-transverse electro-magnetic radiation mode defined by the cavity. However, both of these implementations require very precise cavity alignment and very expensive optical components. In contrast, the described embodiments do not require precise cavity alignment and can be configured with low-cost optical components.
As shown, the embodiment of
For at least some embodiments, the fiber isolator 255 serves two purposes, (1) to help suppress all but the desired mode or modes (TEMOO Transverse Electro-magnetic Mode, and (2) to isolate the laser source from back reflections that might be generated from optical components.
For at least some embodiments, the free-space beam of electro-magnetic radiation emitted from the second fiber collimator (FiberPort) 257 then impacts lens #2 (252) and lens #3 (253) and passes through beam steering mirrors 254.
As shown in
The incoming beam of electro-magnetic radiation is considered aligned with the cavity when the incoming beam is steered into the optical cavity 220 such that the incoming beam of electro-magnetic radiation is congruent with the optical cavity axis. When the incoming beam of electro-magnetic radiation differs (not congruent or misaligned) with the cavity axis, a subset of the plurality of allowed axial-plus-transverse electromagnetic radiation modes in the cavity which is greater than one can be excited.
For an embodiment, the incoming beam of electro-magnetic radiation includes one or more transverse modes. When the incoming beam of electro-magnetic radiation includes more than one transverse mode, a subset of the plurality of allowed axial-plus-transverse electromagnetic radiation modes in the cavity which is greater than one can be excited.
Referring back to
In an embodiment, the incoming beam of electro-magnetic radiation includes one transverse mode. The beam is aligned with the optical axis of the cavity. A waist size and/or waist location of the incoming beam of electro-magnetic radiation is selected to be a different (not mode matched) from that defined by the physical design of the optical cavity 220 thereby exciting a subset of the plurality of allowed axial-plus-transverse electro-magnetic radiation modes in the cavity which is greater than one.
In an embodiment, the incoming beam of electro-magnetic radiation is described as one transverse electro-magnetic mode, a waist size and/or waist location of the incoming beam of electromagnetic radiation is selected to be the same as that defined by the physical design of the optical cavity 220. The beam is steered into the optical cavity 220 so that the incoming beam of electro-magnetic radiation is misaligned with the optical cavity axis thereby exciting a subset, which is greater than one, of the plurality of the allowed axial-plus-transverse electro-magnetic radiation modes. In an embodiment, this misalignment is implemented using two beam steering mirrors 254 to steer the electro-magnetic radiation into the optical cavity 220 such that a misalignment is created by introducing a displacement between the incoming beam of electro-magnetic radiation and the optical cavity axis of the optical cavity 220 thereby resulting in a beam of electro-magnetic radiation within the optical cavity 220 that excites only a subset of the plurality of allowed axial-plus-transverse electro-magnetic radiation modes which is greater than one. One method to displace the incoming beam of electro-magnetic radiation is to use the two steering mirrors 254 to tilt the incoming beam of electro-magnetic radiation so that it is misaligned with the optical cavity axis thereby exciting a subset, which is greater than one, of the plurality of the allowed axial-plus-transverse electro-magnetic radiation modes.
In another embodiment, the incoming beam of electro-magnetic radiation includes more than one transverse mode(s), thereby exciting a subset of the plurality of allowed axial-plus-transverse electro-magnetic radiation modes in the cavity which is greater than one.
In another embodiment, the incoming beam of electro-magnetic radiation includes more than one transverse mode(s). The incoming beam of electro-magnetic radiation is aligned with the cavity axis. A waist size and/or waist location of the incoming beam of electro-magnetic radiation is selected to be a different size than that defined by the physical design of the optical cavity 220 thereby exciting a subset of the plurality of allowed axial-plus-transverse electro-magnetic radiation modes in the cavity which is greater than one.
In another embodiment, the incoming beam of electro-magnetic radiation comprises more than one transverse mode. A waist size and/or waist location of the incoming beam of electromagnetic radiation is selected to be the same as that defined by the physical design of the optical cavity 220. The incoming beam of electro-magnetic radiation is misaligned with the optical axis of the cavity, thereby exciting a subset of the plurality of allowed axial-plus-transverse electro-magnetic radiation modes in the cavity which is greater than one
In addition to the previously mentioned embodiments utilizing a single tunable light source to generate more than one transverse mode in the cavity, there are at least two other methods to generate more than one transverse modes in the cavity. One embodiment includes a light source that generates multiple transverse modes, and another embodiment includes using multiple light sources.
For at least some embodiments, the cavity detector 230 is placed such that the cavity detector 230 senses electro-magnetic radiation emanating from the optical cavity 220. The cavity detector 230 receives electro-magnetic radiation exiting the optical cavity 220 and produces a voltage or current that is proportional to the electro-magnetic radiation sensed by the cavity detector 230. The electro-magnetic radiation sensed by the cavity detector 230 is representative of the electro-magnetic radiation intensity inside the optical cavity 220, and for at least some embodiments, it is this electro-magnetic radiation that represents the electro-magnetic intensity, amplitude, a phase, or an amplitude and phase at a plurality of wavelengths.
For at least some embodiment, the cavity detector 230 generates data (sensed) in an analog form. For at least some embodiments, for input into and processing by the processor 240, the analog data is converted into digital form by and ADC (analog to digital converter) 291. The processor 240 records the received output at the plurality of wavelengths and detects the substance. For an embodiment the substance is detected based on the intensity, amplitude, a phase, or an amplitude and phase of the sensed electro-magnetic radiation emanating from the optical cavity 220 at one or more of the plurality of wavelengths while the optical cavity 220 receives the beam of electro-magnetic radiation.
For at least some embodiments, the incoming beam of electro-magnetic radiation can be described or represented by a linear combination of transverse modes. The transverse mode of electro-magnetic radiation is defined as a particular (transverse) electro-magnetic field pattern of radiation measured in a plane perpendicular (i.e., transverse) to the propagation direction of the electro-magnetic radiation. If a transverse wave is moving in the z direction, its oscillations lie in the x-y plane, and the Electric field oscillates in one plane, for instance the x plane, while the Magnetic field oscillates in the perpendicular plane, for instance the y plane.
The term ũ0(x,z)ũo(y,z) is referred to as the lowest-order Gaussian mode (TEM 00) and its transverse component can be described as a round “slab” or disk. The function ũ1(x,z)ũ0(y,z) is called the first-order Gaussian mode in x and the lowest-order mode in y and as can be seen in
A beam of Electro-magnetic radiation in an optical cavity (such as, optical cavity 220) can also be described using axial—plus-transverse modes that are allowed. Allowed cavity modes (allowed axial-plus-transverse electro-magnetic radiation modes) can be understood using a framework in the wavelength domain.
When resonator theory is applied to an arrangement of 2 or more mirrors forming an optical cavity, the steady-state resonance condition becomes,
where the coefficients r1, r2, r3, . . . rn are the wave-amplitude reflection coefficients for mirror 1, mirror 2, mirror 3, ... mirror n which defines the cavity; ω is the angular frequency of the optical wave, λ is the wavelength of the optical wave, c is the speed of electro-magnetic radiation,
is the phase shift for one full trip around the optical cavity with a round—trip distance designated as p, and ψnm is the “Guoy phase shift” phase angle. The resonance condition equation expresses the round-trip phase shift conditions under which the system oscillates with greater electro-magnetic amplitudes than when the resonance condition is not met. The greater amplitudes produced while meeting this resonance condition is called allowed axial-plus-transverse modes.
Allowed axial-plus-transverse modes can occur at a plurality of wavelengths. Axial-plus-transverse modes in an optical cavity can be described mathematically in the following manor. The electro-magnetic amplitude in an optical cavity can be described in Cartesian coordinates using the paraxial wave equation:
where {tilde over (E)}(x, y, z)=0 is the electro-magnetic amplitude in the cavity.
The “Guoy phase shift” is associated with the fact that different axial-plus-transverse modes can accumulate phase at different rates. If one identifies the axial-plus-transverse modes of a given optical cavity, one can also find the exact angular frequencies (wavelengths) at which the resonance condition is met. The resonance angular frequencies, including the Guoy phase shift, of the Hermite-Gaussian axial-plus-transverse modes for a two mirror cavity made up of mirrors M1 and M2 are given by a plurality of wavelengths described by:
where p is the round-trip path length in the optical cavity, c is the speed of electro-magnetic radiation, and where the parameters “g” are given by;
where the parameters R1 and R2 are the radius of curvatures of mirror M1 and M2, respectively. The resonance angular frequencies are degenerate with respect to the indices n and m through the (n+m+1) term
In cylindrical coordinates, the resonance angular frequencies of the Laguerre-Gaussian axial-plus-transverse modes in the optical cavity are then given by a plurality of wavelengths described by,
In this case, the resonance angular frequencies are degenerate with respect to the indices l and p through the (|l|+2p+1) term.
In an embodiment, electro-magnetic amplitudes at a plurality of wavelengths can be excited using two beam steering mirrors to create a mode mismatch between the incoming beam of electro-magnetic radiation and the optical cavity. The number of allowed axial-plus-transverse electro-magnetic radiation modes excited is related to the amount of misalignment between the incoming beam of electro-magnetic radiation and the optical cavity axis. The electro-magnetic amplitude of the axial-plus-transverse modes is related to the amount of misalignment between the incoming beam of electro-magnetic radiation and the optical cavity axis.
In mathematical terms, in the case where the optical cavity has losses, the angular frequencies, which make up optical cavity modes cannot be completely described as delta functions. With optical loss, the electro-magnetic radiation in the optical cavity can be described as a superposition of a large number of waves, separated by equal phase shifts but with amplitudes that are geometrically reduced by loss in the optical cavity. Because of this geometric term, the higher the optical cavity loss translates to a wider axial-plus-transverse cavity mode in the frequency domain. When considering a plurality of axial-plus-transverse modes, overlapping frequencies of two or more modes can cause transverse mode beating which will introduce large intensity variations, as a function of time, in each undistinguished axial-plus-transverse mode. This inter-modulation distortion will result in a noisier measurement thereby degrading system signal-to-noise ratio and sensitivity. As an example, suppose an optical cavity is oscillating simultaneously in two indistinguishable modes in the frequency domain. The indices of these modes is q1n1m1 and q2n2m2. The total output power from the cavity associated with these two indistinguishable modes can be written as;
i(t)=I01+I02+I12 cos[(ω2−ω1)t+ϕ12]
where I01 and I02 are the dc intensities due to each separate mode, plus a cross product or beat frequency term I12 cos[(ω2−ω1)t+ϕ12] between the two signals. If both modes have the exact same frequency, as in the degenerative case, then (ω2−ω1)=0 and there is no time dependence on i(t). However, if (ω2−ω1)≠0, then there will be a time dependent variation of the total intensity, i (t) with a period defined by the frequency separation of the two modes. This time dependent variation would introduce unwanted noise on the intensity measurement thereby impacting performance. The axial-plus-transverse modes excited is a subset of the allowed axial-plus-transverse optical cavity modes that, except in the degenerative case, do not overlap in the frequency domain and/or do not produce mode beating (mixing products) and/or minimize mode beating in the frequency domain.
Laser Tuning
As previously described, for an embodiment, the laser generates a beam of electro-magnetic radiation, wherein a wavelength of the beam of electro-magnetic radiation is tuned to operate at a plurality of wavelengths. For at least some embodiments, each axial-plus-transverse optical cavity mode is excited when electro-magnetic radiation entering the optical cavity 220 meets the resonance conditions of the optical cavity, including excitation wavelength.
At least some embodiments include continuous Laser Current and/or Temperature Tuning. Specifically, in an embodiment, the laser current and/or laser temperature is varied over a range of currents and/or temperatures thus causing the laser to radiate electro-magnetic energy over a range of different wavelengths (frequencies). For an embodiment, these variations in current and/or temperature are periodic and can be described using linear (such as a triangle wave in
At least some embodiments include discrete laser current tuning. Specifically, in an embodiment, the laser current is varied using a sequence of current steps. These discrete steps in current result in discrete steps in the laser wavelength. By stepping the laser wavelength, a plurality of axial-plus-transverse electro-magnetic radiation modes is activated.
For at least some embodiments, detecting the substance is based on a variation in a shape of an intensity, an amplitude, a phase, or an amplitude and phase of axial-plus-transverse electro-magnetic modes of the sensed electro-magnetic radiation. When optical loss is introduced, axial-plus-transverse electro-magnetic radiation modes no longer resonate at a single frequency.
Hermite-Gaussian modes can be used to describe the electro-magnetic amplitudes in the cavity.
In an embodiment, the laser wavelength is varied to activate a plurality of allowed axial-plus-transverse electro-magnetic radiation modes and at least one of an intensity, an amplitude, a phase, or an amplitude and phase of the sensed electro-magnetic radiation emanating from the optical cavity are detected. The repetitive pattern can be used to uniquely identify each mode and associate the measured electro-magnetic radiation to each axial-plus-transverse mode.
In another embodiment, the laser wavelength is varied to activate a plurality of allowed axial-plus-transverse electro-magnetic radiation modes and at least one of an intensity, an amplitude, a phase, or an amplitude and phase of the sensed electro-magnetic radiation emanating from the optical cavity are detected. The amplitude of the laser tuning voltage can be used to uniquely identify each mode and associate the measured electro-magnetic radiation to each axial-plus-transverse optical cavity mode.
For at least some embodiments, detecting the substance includes summing at least one of an intensity, an amplitude, a phase, or an amplitude and phase of allowed axial-plus-transverse electro-magnetic modes of the sensed electro-magnetic radiation. Because the electro-magnetic radiation in an optical cavity is governed by optical losses and wherein the cavity detector senses electro-magnetic radiation emanating from the optical cavity, the cavity detector can provide a measurement of optical loss in the cavity. at least one of an intensity, an amplitude, a phase, or an amplitude and phase of each excited axial-plus-transverse electro-magnetic radiation mode can be measured. For example, by summing the peaks intensities of identified axial-plus-transverse optical cavity modes over a plurality of wavelengths, information about optical cavity loss can be obtained.
At least some embodiments further include a second normalization detector (such as, normalization detector 261) sensing electro-magnetic radiation reflected off an entrance to the optical cavity, wherein the processor further operates to monitor the sensed electro-magnetic radiation reflected off the entrance to the optical cavity. In order to represent the electro-magnetic radiation intensity entering the optical cavity, the beam reflected from the entrance to the optical cavity can be monitored.
For at least some embodiments, the processor 240 further operates to account for effects of electro-magnetic radiation changes emanating from the optical cavity 220 caused by changes in an intensity, amplitude, phase, or amplitude and phase of the electro-magnetic radiation entering the optical cavity 220 including measuring the intensity, amplitude, phase, or amplitude and phase of the sensed electro-magnetic radiation reflected off the entrance to the optical cavity 220 and correcting the electro-magnetic radiation emanating from the cavity based on the intensity, amplitude, phase, or amplitude and phase measured by the normalization detector 261.
In an embodiment, by comparing the output of the cavity detector with the output of the normalization detector 261, corrections for electro-magnetic radiation changes emanating from the optical cavity 220 caused by changes in the intensity of the electro-magnetic radiation entering the optical cavity 220 can be made.
For at least some embodiments, the processor further operates to tune the wavelength of the beam of electro-magnetic radiation based on monitoring electro-magnetic radiation exiting from a reference substance, wherein at least a portion of the electro-magnetic radiation reflected off the entrance to the optical cavity is passed through the reference substance, and wherein the reference substance is selected to insure that a frequency of the electro-magnetic radiation corresponds to a frequency at which an absorption of the substance is measured.
In an embodiment, the beam of electro-magnetic radiation from the laser enters the optical cavity 220 through the partially reflecting optical cavity entrance mirror 223. A portion of that electro-magnetic radiation passes through the partially reflecting optical cavity entrance mirror 223 while another portion is reflected off the mirror 223. The electro-magnetic radiation reflected off the cavity entrance mirror 223 is transported to the partially reflecting beam splitter 260. At this beam splitter 260 a portion of the electro-magnetic radiation is reflected off the beam splitter 260 and is directed through the reference substance 262 and sensed by the reference substance detector 263 as shown in
For an embodiment, the reference substance has enough reference species of interest to make an absorption measurement possible, and must have sufficient path length to make an absorption measurement.
The processor 240 receives the information of the sensed electro-magnetic radiation from the Reference Substance detector 263 and further operates to tune the wavelength of the beam of electro-magnetic radiation by adjusting the laser ramp offset voltage such that the average laser ramp voltage corresponds to a wavelength consistence with the absorbance wavelength of the substance of interest. This ensures that the laser remains centered at the wavelength of the absorbance feature and as such centered on the absorbance of the substance of interest.
For at least some embodiments, as in the case of a normalization determination, a valid reference substance determination requires that the measurement be made when the optical cavity 220 is not in resonance. For an embodiment, this requirement is, again, met by selecting laser frequencies where the optical cavity is not in resonances. For an embodiment, this is accomplished by monitoring the signal on the cavity detector 230 and selecting laser frequencies where no optical cavity resonances occur.
For at least some embodiments, the processor further operates to tune the beam of electro-magnetic radiation over the plurality of wavelengths one or multiple times, and record the information related to the sensed electro-magnetic radiation at the plurality of wavelengths. For at least some embodiments, the processor further operates to sum the recorded information related to the sensed electro-magnetic radiation for each of the plurality of wavelengths over the one or multiple times. For at least some embodiments, the processor further operates to identify repetitive patterns in the recorded information related to the sensed electro-magnetic radiation for each of the plurality of wavelengths over the one or multiple times. For at least some embodiments, the processor further operates to detect an amount of the substance based on the identified repetitive patterns in the summed recorded information related to the sensed electro-magnetic radiation for each of the plurality of wavelengths over the one or multiple times.
More information regarding the optical loss can be obtained by tuning the laser to the same plurality of wavelengths multiple times.
For an embodiment, the cavity includes an optical cavity 220 wherein the optical cavity 220 comprises 2 or more mirrors forming the optical cavity, wherein the optical cavity receives the beam of electro-magnetic radiation; wherein successive reflections of the electro-magnetic radiation at each of the 2 or more mirrors occurs at one common physical location on each mirror.
For an embodiment, the beam of electro-magnetic radiation is selected to ensure that one axial-plus-transverse electro-magnetic mode is dominate at any one of the plurality of wavelengths.
For at least some embodiments, the processor receives information from one or more detectors relating to a portion of the beam of electro-magnetic radiation not entering the cavity, wherein the detectors (261, 263) are arranged in a physical configuration such that the information they provide in conjunction with the sensed (sensed by cavity detector 260) electro-magnetic radiation emanating from the optical cavity increases the measurement range of the substance detectable, wherein the sensed signal in conjunction with the cavity detector increases the measurement range of the substance detectable. Specifically, information received from the cavity detector and information received from one or more detectors is used to detect the sub stance.
For an embodiment, the amount of the substance detected is limited by the cavity detector and processor's capability to measure excited allowed axial-plus-transverse electro-magnetic radiation modes when the majority of electro-magnetic radiation in the cavity is absorbed by the substance. Information from one or more detectors sensing electro-magnetic radiation not entering the cavity may be used to detect the substance when a substance absorbs the majority of electro-magnetic radiation in the cavity. Combining the information from the cavity detector and the information from one or more detectors sensing electro-magnetic radiation not entering the cavity may increase the measurement range of the substance detectable.
Referring back to
The substance absorbs electro-magnetic radiation outside the cavity in the optical paths of a beam splitter 260 and a reference Substance 262. As shown, for an embodiment, the detectors are arranged in a physical configuration such that the propagation length of the path from the beam splitter 260 to the normalization detector 261 and the path from the Beam splitter 260 to the reference substance detector 263 is different; therefore the information from the normalization detector 261 and reference substance detector 263 can be used to determine the amount of the substance detected by measuring losses external to the cavity.
The cavity 220 improves measurement sensitivity by adding gain to the system to measure the substance at very low levels of concentration. In the absence of the cavity 220 the substance may be detected at higher levels of concentration. By combining the information derived from the cavity 220 and cavity detector 230 with information derived outside the cavity through the normalization detector 261 and reference substance detector 263 the concentration range of substance detected can be extended.
As previously described, for at least some embodiments, detecting the substance is based on a variation in a shape of an intensity, an amplitude, a phase, or an amplitude and phase of axial-plus-transverse electromagnetic modes of the sensed electro-magnetic radiation. As previously described, for at least some embodiments, detecting the substance comprises summing at least one of an intensity, an amplitude, a phase, or an amplitude and phase of axial-plus-transverse electromagnetic modes of the sensed electro-magnetic radiation.
As previously described, at least some embodiments include sensing, by a normalization detector, electro-magnetic radiation reflected off an entrance to the optical cavity, wherein the processor further operates to monitor the sensed electro-magnetic radiation reflected off the entrance to the optical cavity. As previously described, at least some embodiments include accounting for effects of electro-magnetic radiation changes emanating from the cavity caused by changes in an intensity, amplitude, phase, or amplitude and phase of the electro-magnetic radiation entering the cavity comprising measuring the intensity, amplitude, phase, or amplitude and phase of the sensed electro-magnetic radiation reflected off the entrance to the optical cavity and correcting the electro-magnetic radiation emanating from the cavity based on the measured intensity.
As previously described, at least some embodiments include tuning the wavelength of the beam of electro-magnetic radiation based on monitoring of a reference electromagnetic radiation exiting from a reference substance, wherein at least a portion of the electro-magnetic radiation reflected off the entrance to the optical cavity is passed through the reference substance, and wherein the reference substance is selected to insure that a frequency of the electro-magnetic radiation corresponds to a frequency at which an absorption of the substance is measured.
As previously described, at least some embodiments include tuning the beam of electro-magnetic radiation over the plurality of wavelengths one or multiple times, and recording the sensed electro-magnetic radiation at the plurality of wavelengths. As previously described, at least some embodiments include summing the recorded sensed electro-magnetic radiation for each of the plurality of wavelengths over the one or multiple times. As previously described, at least some embodiments include identifying repetitive patterns in the recorded sensed electro-magnetic radiation for each of the plurality of wavelengths over the one or multiple times. As previously described, at least some embodiments include detecting an amount of the substance based on the identified repetitive patterns in the summed recorded sensed electro-magnetic radiation for each of the plurality of wavelengths over the one or multiple times.
As previously described, at least some embodiments include receiving, by the processor, information from one or more detectors relating to a portion of the beam of electro-magnetic radiation not entering the cavity, wherein information received from the cavity detector and information received from one or more detectors are used to detect the substance.
Although specific embodiments have been described and illustrated, the embodiments are not to be limited to the specific forms or arrangements of parts so described and illustrated. The described embodiments are to only be limited by the claims.
This patent application is a continuation of U.S. patent application Ser. No. 15/909,909, filed Mar. 1, 2018, and claims priority to U.S. Provisional Patent Application Ser. No. 62/560,235 filed Sep. 19, 2017, all of which are herein incorporated by reference.
Number | Date | Country | |
---|---|---|---|
62560235 | Sep 2017 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 15909909 | Mar 2018 | US |
Child | 16907491 | US |