This invention relates to medical diagnostic imaging systems and, in particular, to fluid filled probes for three dimensional imaging with ultrasonic diagnostic imaging systems.
Ultrasonic probes always benefit from a good acoustic path between the transducer and the patient's body. With static probes in which the transducer does not move, the transducer is covered with an acoustic window that is highly transmissive of ultrasound, generally a rubber-like material. The acoustic path to the body is completed with an acoustic gel between the acoustic window and the skin of the patient, which prevents air and air bubbles from obstructing or interfering with the passage of acoustic energy into and out of the body.
Mechanical probes have been in use for many years in which the transducer is oscillated to sweep a beam of ultrasound through the body for imaging. The first mechanical probes employed single element transducers which were oscillated in a fluid bath that provided acoustic coupling from the transducer element to the acoustic window. The acoustic gel would then couple the ultrasound energy between the acoustic window and the body. This same approach was used with oscillating annular array transducers, in which the annular array was electronically focused as its beam was swept through an image plane of the patient.
In recent years the use of mechanical single element probes has declined as static array probes with electronic beam steering have replaced them. But with the recent advent of live three dimensional ultrasonic imaging, mechanical probes have begun to appear which oscillate an array of transducers that electronically steers beams as the array is moved. The array transducer can be electronically controlled to steer beams in the azimuth direction as the mechanical oscillation moves the array in the elevation direction, thereby transmitting and acquiring ultrasonic energy from a volumetric region rather than just a plane. In essence, the array electronically scans a plane and the plane is mechanically swept through the volume being imaged. These three dimensional mechanical array probes, like their single element ancestors, need a means for coupling ultrasonic energy from the moving transducer to the acoustic window. Like their predecessors, a fluid bath is the most common and convenient coupling means.
With fluid bath probes come the familiar challenges which fluids present. Fluid-filled probes can leak and can develop air bubbles in the fluid which can interfere with the acoustic path. These problems are often the result of expansion and contraction of the fluid and probe components with changes in temperature and pressure. A probe shipped by airplane can be subject to low pressures and temperatures during transit, whereas a probe left in an automobile in the sun can be subject to high temperatures. Accordingly it is desirable to provide a means for accommodating such expansion and contraction within the probe itself so as to prevent the development of fluid leakage and air bubbles in the acoustic path of the probe.
In accordance with the principles of the present invention a fluid-filled ultrasound probe with a moving transducer is described which includes a volume compensation system that accommodates changes in fluid volume with changes in temperature or pressure. The volume compensation system includes a thin-walled balloon made of a high performance thermoplastic material which is coupled to the fluid chamber for the transducer. At nominal temperatures and pressures the balloon is partially collapsed so that it will respond to increases or decreases in pressure or temperature by collapsing or expanding without reaching its expansion limit. The thermoplastic material enables the balloon wall to be very thin which offers little resistance to the filling and emptying of the balloon with changes in fluid temperature or pressure. In accordance with a further aspect of the present invention, the thin-walled balloon has a low permeability to the fluid which fills the fluid chamber of the probe. A desirable balloon material should also exhibit high integrity in terms of tear strength, burst strength, and puncture strength.
In the drawings:
Referring first to
Referring now to
Because ultrasonic energy does not efficiently pass through air, the array transducer 46 is surrounded by a liquid which is transmissive of ultrasound and closely matches the acoustic impedance of the body which is approximately that of water. The liquid is contained within a fluid chamber 42 inside the transducer mount assembly 40 which also contains the array transducer 46. Water-based, oil-based, and synthetic polymeric liquids may be used. In a constructed embodiment silicone oil is used as the acoustic coupling fluid in the transducer fluid chamber. Further details of the fluid chamber of the embodiment of
In accordance with the principles of the present invention a thin-walled balloon 44 of high performance thermoplastic is coupled to be in fluidic communication with the transducer fluid chamber 42. The balloon is shown in perspective in
In the constructed embodiment in which the balloon 44 was made of a high performance thermoplastic, the balloon wall thickness was approximately 0.6 mil. Wall thicknesses of 5-50 microns (0.0002-0.002 inches) have been found to exhibit the desired compliance (ΔP/ΔV) while at the same time exhibiting the desired low permeability (p) to fluids. Partially filled balloons of the above wall thickness have been found to offer little resistance to filling and emptying, with compliances of less than 2 psi per ml. The permeability of the high performance thermoplastic material has been found to be an order of magnitude better than typical elastomers, with the permeation rate p per unit thickness generally less than 1.0 and usually in the range of 0.1 to 0.2 cm3·mm/m2·day·atm (for gas permeability). It should be noted that silicone fluids are known to aggressively permeate many plastics which are impermeable to most other fluids.
The volume compensation balloon should also exhibit good thermal stability. High performance thermoplastics such as PET are used below the glass transition temperature Tg of the material. Elastomeric balloons have a low glass transition temperature, usually in the range of −10° C. to −80° C. and consequently are always highly elastic and rubber-like at their intended operating temperatures, e.g., room temperature. Elastomeric balloons also undesirably have a higher ΔP vs ΔV and will become stiff and resist collapsing in colder temperatures. A volume compensation balloon of the present invention should be thermally stable so that the compliance does not deteriorate throughout the intended temperature range of use.
Elastomers, being highly elastic by design, will stretch considerably and can usually be stretched in excess of 100% and still return to the original size and shape. A high performance thermoplastic balloon of the present invention is rigid and can generally only be stretched about 2%-7% before yielding. When excessively stretched the balloon will not return to its original size and shape but will experience plastic deformation. A constructed embodiment exhibited a coefficient of expansion of only ˜0.00013/° C.
A high performance thermoplastic balloon should also exhibit a high burst strength. A constructed embodiment exhibited a burst strength pressure resistance of 18 atmospheres.
In summary, the factors that should be considered when choosing a balloon material are the ability to fabricate a thin-walled balloon since stiffness of the balloon is principally controlled by wall thickness. The balloon should be rigid yet compliant due to the thin wall design. The material should also exhibit low permeability to the acoustic fluid; good thermal stability so that the balloon will not become stiff at low temperatures; and good integrity in terms of burst strength, tear strength, and puncture resistance.
The fluid chamber 42 in the transducer mount assembly 40 of the embodiment of
This application claims the benefit of U.S. provisional application Ser. No. 60/559,379 filed Apr. 02, 2004, which is incorporated herein.
Filing Document | Filing Date | Country | Kind | 371c Date |
---|---|---|---|---|
PCT/IB2005/050985 | 3/22/2005 | WO | 00 | 9/25/2006 |
Publishing Document | Publishing Date | Country | Kind |
---|---|---|---|
WO2005/094690 | 10/13/2005 | WO | A |
Number | Name | Date | Kind |
---|---|---|---|
4517840 | Thompson et al. | May 1985 | A |
4784148 | Dow et al. | Nov 1988 | A |
5117831 | Jang et al. | Jun 1992 | A |
5178150 | Silverstein et al. | Jan 1993 | A |
5226422 | McKeighen et al. | Jul 1993 | A |
5738901 | Wang et al. | Apr 1998 | A |
5830182 | Wang et al. | Nov 1998 | A |
5882302 | Driscoll et al. | Mar 1999 | A |
6592526 | Lenker | Jul 2003 | B1 |
7081113 | Sutton | Jul 2006 | B2 |
7479128 | Lenz | Jan 2009 | B1 |
20030083653 | Maguire et al. | May 2003 | A1 |
Number | Date | Country |
---|---|---|
0089131 | Sep 1983 | EP |
20050141428 | May 2004 | JP |
Number | Date | Country | |
---|---|---|---|
20070293761 A1 | Dec 2007 | US |
Number | Date | Country | |
---|---|---|---|
60559379 | Apr 2004 | US | |
60559321 | Apr 2004 | US | |
60559390 | Apr 2004 | US |