The present invention relates to an equipment for continuously cleaning a strip in a tank with ultrasound emitting means. Such an invention eases the global management of said cleaning tank.
In the metallurgical field, producing strip having a high surface quality is of a major importance. During the rolling step, iron, metallic particles, dirt and grease adhere to the metal strip. Such adhesions engender a degradation of the strip surface quality post-coating because they will be entrapped under the coating and thus the surface will not be smooth. In order to avoid such drawbacks, the strip is cleaned before the coating step. Generally, it occurs after the rolling operation and before the annealing or the coating. To do so, most of the cleaning lines uses an electrolytic process among their cleaning operations. However, such a technique presents a high safety risk due to the H2 accumulation leading to safety hazards such as fire. Consequently, cleaning lines using ultrasound have been developed to replace the electrolytic process. Naturally, new problems have arisen, especially concerning the management of the ultrasound emitting means. Usually, transducers converting oscillating electrical energy into mechanical energy are used, creating the ultrasound. Despite those emerging problems, such lines are interesting because they are safer, create [[less]] fewer by-products and have a lower electrical consumption, being thus more eco-friendly.
Ultrasound cleaning works thanks to the propagation of an ultrasound wave (or more generally an acoustic wave) through an aqueous solution which induces local variations of the aqueous solution pressure. When the negative pressure is low enough (lower than the aqueous solution vapour pressure), the aqueous solution cohesive forces break down, and gas bubbles (also called cavitation bubbles) are formed. These bubbles are then submitted to pressure variations (due to acoustic wave propagation), which cause them to expand and contract successively until they collapse. Ultrasonic waves induce a thermal effect, but also a mechanical effect due to cavitation. Indeed, two phenomena occur when cavitation bubbles break down:
Patent KR 2005 006 3145 discloses an apparatus cleaning a steel sheet. Said steel sheet is passed through a tank filled with an alkaline solution in which ultrasound emitting means are arranged inside boxes placed on each side of the passing sheet.
However, by using the above method and its equipment, two major drawbacks appear. Firstly, the intensity of the ultrasounds emitted by ultrasound emitting means placed inside the housing is lowered when the ultrasounds pass through the housing wall. Secondly, the maintenance is time-consuming because the replacement of a box containing the ultrasound emitting means requires the removal of the several pieces, the use of a scaffolding and also raises safety concerns.
It is an object of the present invention to provide a solution solving the aforementioned problems.
The present invention provides an equipment (1) for the continuous cleaning of a moving strip (S) comprising:
Another object of the present invention is to provide a method for cleaning of a moving metal strip using the equipment, and to provide a method for replacing an ultrasound emitter in the equipment.
Other characteristics and advantages of the invention will become apparent from the following detailed description of the invention.
To illustrate the invention, various embodiments and trials of non-limiting examples will be described, particularly with reference to the following figure:
The invention relates to an equipment 1 for the continuous cleaning of a moving strip S comprising:
at least an impermeable closable opening 11 on at least a lateral side of said tank 2 through which the at least one ultrasound emitting mean 5 can be introduced,
a wire W connecting said means for controlling the power 10 of the at least one ultrasound emitting mean 5 and the at least one ultrasound emitting mean 5.
As illustrated in
The at least one roller 4 is preferentially at the bottom of the tank but above the emptying means 7, such an arrangement increases the distance travelled by the strip S through the aqueous solution 3 and the cleaning time thus improving the cleaning.
The aqueous solution 3 is introduced into the tank by the feeding means 6 such as pipes and valves, preferentially connected to another tank filled with the solution.
As illustrated in
The means for estimating 8 the aqueous solution level can be a differential pressure captor or any means used in a hydrostatic method. The means for estimating 8 the aqueous solution level can also be composed of several aqueous solution level indicators, disposed along the bath height indicating the presence or not of an aqueous solution permitting to estimate the aqueous solution level between two indicators. Such level indicators can be vibrating level switches.
The at least one ultrasound emitting mean 5 is placed inside said tank 2 preferably under the feeding means 6 and preferably above the roller 4.
In the prior art, it seems that it is not possible to easily and quickly remove an ultrasound emitting mean from the tank. The equipment according to the invention enables a quicker and easier removal of an ultrasound emitting mean for several reasons. Firstly, no scaffolding is required which reduces the replacement time and makes it safer for the workers. Secondly, the tank is not totally emptied but the bath level is set under the ultrasound emitting mean to be replaced, so the filling time after the operation is shorter.
Preferably, said at least an ultrasound emitting mean 5 is immerged in the aqueous solution 3. It permits to enhance the efficiency of the cleaning.
Preferably, as illustrated in
Several tests have been done to demonstrate the improved efficiency of [[the]] a cleaning tank equipped with transducers, such as push-pull transducers, compared to one equipped with submersible boxes. In those tests, the cleanliness of a strip sample has been measured before and after a cleaning step. In those experiments, a strip is immersed during 24 sec in a box containing a cleaning bath, having 10 g·L−1 of NaOH, at 65° C. and either a set of two push-pull piezo transducers having a power of 2 kW or a submersible box having a power of 2 kW. It is assumed that an immersion time of 24 seconds in the experiment conditions corresponds to a direct exposition time of about 6 seconds because a strip portion is faced by an ultrasound emitter means only during a quarter of the experiment time due to its displacement through the aqueous solution.
The cleaning efficiency, as noted in the following table, is: “the estimated cleanliness before the cleaning step” divided by “the estimated cleanliness after the cleaning step”. To estimate the cleanliness, a 3M 595 Scotch™ adhesive is pressed on a strip surface in order to stick the iron fines and the oil onto the adhesive. Then the reflectance of the scotch is measured by a reflectometer. This reflectance is linked to the density of iron fines per square meter. The more iron fines have adhered to the adhesive, the lower will be its reflectance. Consequently, the higher is the adhesive reflectance, the cleaner is the strip. The following table contains the main parameters of the experiment. In
Preferably, as illustrated in
Preferably, as it can be seen in
Preferably, as illustrated in
For example, if six ultrasound emitters working at a frequency of 25 kHz are used in an environment comparable to water, the wave speed, which depends on numerous factors (e.g.: the temperature and the pressure) is approximately of 1500 m·s−1. The wavelength is equal to the wave speed divided by the wave frequency so in this case, 1 500/25 000=0.06, the wavelength is approximately of 6 cm. In the case where the ultrasound emitting means produce ultrasound with a wavelength of 6 cm, they should be laterally shifted of, (1/6)×6=1 cm one from another.
As it can be seen in
Preferably, said resonator rod 15 and the strip S are spaced by a distance comprised between 40 mm and 250 mm. Such spacing enables to efficiently use the ultrasound emitting mean. Such spacing distance improves the installation 1 because if the spacing if less than 40 mm, the ultrasound emitting mean will eventually be broken by the strip S due for example strip bending or strip flatness irregularities. But if the spacing is bigger than 200 mm then the efficiency of the ultrasound emitting mean cleaning power seems to be severely reduced.
Preferably, said at least closable opening 11 is separable from the tank 2 and attached to the ultrasound emitting mean 5. Such an arrangement eases the removal of the ultrasound emitting mean.
Preferably, the strip S to be cleaned has two opposite surfaces and the equipment according to the invention comprises preferably at least one ultrasound emitting mean 5 facing each of said surface. Even though an ultrasound emitting mean placed on one side of a strip cleans both surfaces, having ultrasound emitting means facing each surface increases the cleaning quality.
Consequently, in an installation similar to the one represented in
Consequently, in an installation similar to the one represented in
Preferably, said equipment has a power density between 5 Watt per litre and 25 Watt per litre. Even more preferentially, the power per litre is between 10 and 20 W·l−1. Using a power density in this range seems to be the best compromise between a sufficient cleaning and energy saving, it allows a good and sufficient cleaning of the strip and avoid energy waste.
The equipment accord to the invention can be used for cleaning any strip that is compatible with it. Preferably, said strip is a metal strip. More preferably, said metal strip is a steel strip.
The invention also relates to a method for the continuous cleaning of a moving strip S using the equipment according to the present invention, wherein said strip is a metal strip.
Preferably, said aqueous solution contains between 10 grams per litre and 40 grams per litre of alkali product. Apparently, an alkali product concentration in this range improves the cleaning and efficiently uses the alkali product. Other solutions such as acidic or neutral solutions can be used, the solution selection depends on the substrates and the pollutants.
Preferably, said aqueous solution is kept at a temperature between 30° C. and 80° C. Apparently, higher is the cleaning solution temperature, better is the cleaning efficiency of the process but shorter is the ultrasound emitting mean lifespan. This range seems to be the best compromise between cleaning efficiency and the ultrasound emitting mean lifespan.
The invention also relates to a method for replacing an ultrasound emitting mean 5 of an equipment according to the present invention, comprising:
The process management system estimates for impermeable closable opening 11 its distance to the aqueous solution level. This estimation is made by the means for calculating 9 based on the impermeable closable opening position, i.e.: at what height they are positioned, and the estimation of the aqueous solution level thanks to the means for estimating the aqueous solution level 8. The means for calculating 9 calculates the distance between an impermeable closable opening position and the estimated level of the aqueous solution.
For safety reasons, the aqueous solution level should be set below an impermeable closable opening of the ultrasound emitting means to be removed because when the closable opening is opened, it prevents the solution 3 to flow out of the bath 2. Consequently, a determined value defines the minimal distance between an impermeable closable opening of an ultrasonic emitting mean to be replaced and the aqueous solution level required to safely replace said ultrasound emitting mean.
In the case where means for estimating the aqueous solution level are composed of at least several aqueous solution indicators, such as vibrating level switches, the aqueous solution level indicators are preferentially placed below each impermeable closable opening 11 at the predetermined distance.
When using an equipment according to the prior art, the procedure to replace an ultrasound emitting means is the following:
Whereas when the equipment according to the invention is used, the procedure is much shorter and simpler:
As it can be observed, no scaffolding is used with the present invention so the replacement duration is reduced by 8 hours, as it takes 1 hour instead of 9 hours.
The present invention is applicable to every process in which a strip is cleaned by passing it through an aqueous solution filled tank comprising ultrasound emitting means.
The invention has been described above as to the embodiment which is supposed to be practical as well as preferable at present. However, it should be understood that the invention is not limited to the embodiment disclosed in the specification and can be appropriately modified within the range that does not depart from the gist or spirit of the invention, which can be read from the appended claims and the overall specification, and a manufacturing method of a hot-rolled steel sheet and a manufacturing apparatus of a hot-rolled steel sheet with such modifications are also encompassed within the technical range of the invention.
Number | Date | Country | Kind |
---|---|---|---|
PCT/IB2018/058711 | Nov 2018 | IB | international |
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/IB2019/059493 | 11/5/2019 | WO | 00 |