1. Field of the Invention
The present invention relates generally to the field of orthopedic surgery and particularly to an orthopedic plate and screw system using bushings.
2. Description of the Related Art
How to best promote rapid and robust healing of fractured bone without residual deformity continues to be the primary goal of orthopedic trauma surgeons. Plate and screw constructs have long been and continue to be a staple in the surgeon's toolbox. Classically, these plates work by a tightened screw pulling and holding the plate tightly against the bone. If the screw loosens, the construct will generally fail resulting in either a healed, deformed bone and/or a non-healed, unstable bone.
One of the more recent advances in the standard plate and screw system was the advent of “locking” screws. These screws were designed for the screw head to become rigidly fixed to the plate. This improved pull-out strength, and allowed complex, fragmented fractures to be held firmly in place. An added bonus is that the locking screws don't come loose from the plate.
The primary problem with locking plates is that these constructs are so rigid the fractures take much longer to heal. The same effect is observed when external fixators are used. If there is no mechanical load allowed across the fracture site, the chemical, electrical, and mechanical signals generated by loading the bone aren't generated. The result is greatly delayed, or, in some cases, the fracture healing is completely arrested. The key to healing is to put compression across the fracture site. Classically, this has been done by placing the screws in the bone in such a way that once tightened the plate acts like a tension band while the screws push against the plate to force the fractured ends together at high pressure. While this does get the bone to heal more rapidly, local bone necrosis at the fracture site has long been recognized. Allowing for intermittent, controlled loading/compression without loss of fracture reduction would be the ideal.
A lesser problem with current locking plates is that once the screw head comes in contact with the plate, the screw locks to the plate. There is no further tightening possible. The plate remains suspended above the bone. Because of this mechanical similarity with external fixators which also do not come in contact with the bone, these locking plates have been promoted as “internal external fixators”. If the surgeon wants the plate to be in firm contact with the bone, he must use specially designed screws to pull the plate to the bone, then put in the locking screws.
A third problem with the current locking plates is cold welding of the screw head to the plate which can make screw removal surgically challenging.
The problems and the associated solutions presented in this section could be or could have been pursued, but they are not necessarily approaches that have been previously conceived or pursued. Therefore, unless otherwise indicated, it should not be assumed that any of the approaches presented in this section qualify as prior art merely by virtue of their presence in this section of the application.
In one embodiment, a specially designed bushing, or similar means, is used between one or more screws and an orthopedic plate, for locking the plate to a bone in a non-rigid, micro-motion-enabling and controlling manner. The direction and amount of flexibility that the screw is allowed is carefully engineered. Unlike any of the prior art, the screw is held to the bushing by friction (not screw threading) between the bushing and the screw. Because the bushing holding the screw is flexible, the screw has some flexibility. Again, the direction and amount of flexibility that the screw is allowed is carefully engineered. Some adjustments to that flexibility can be made by choices allowed to the surgeon in the operating room.
A flexible locking is achieved in the sense that friction and flexibility relieving stress on the bushing is not allowing the screw to back out of the plate. When the screw does “flex,” it springs back to its original position. Thus, an advantage is the controlled micro-motion, which enhances the healing of said bone and/or its fracture. Another advantage is allowing the plate to be firmly pulled onto the bone if the surgeon so chooses. Yet, another advantage is that when (and if) the screw needs to be removed, there is no cold weld, so the screw head should not strip on removal as is fairly common with the current locking plates.
In the prior art, when the screw is allowed to move it stays moved in its new position. The proposed bushing is flexible. Thus the screw has a tendency to spring back to its original position after loading. Not only is the fracture loaded with patient movement in the first place, but the loading is relieved when the patient rests. This means that the compression of the fracture will be intermittent, which is a much more beneficial physiologic situation for healing than the current technologies. Because only specifically controlled micromotion is allowed, there is no loss of fracture reduction.
The friction fit between the bushing and the screw allows the surgeon to stop tightening the screw at any degree of pressure desired between the plate and the bone. The friction between the bushing “locks” the screw, preventing it from backing out at any point the surgeon quits tightening the screw. Cold weld between the bushing and screw, with the resultant stripping of screw head, should not occur, making removal of the hardware surgically more predictable.
Because it is still a locking plate, the advantages for dealing with complex fractures are still realized.
In another embodiment, the screws may, at a pressure level set during installation, be allowed to slide after the installation, and thus, allowing the fracture to be compressed while still controlling its alignment and rotation. After any sliding there will still be controlled intermittent loading to promote healing. The post-installation controlled compression of the fracture is also very advantageous to healing.
In another embodiment, the micro-motion enabling bushing may be shaped to have the screw inserted into the plate in a pre-engineered fixed angle, or so that the screw may be inserted at a variable angle.
The above embodiments and advantages as well as other embodiments and advantages will become apparent from the ensuing description and accompanying drawings.
For exemplification purposes, and not for limitation purposes, embodiments of the invention are illustrated in the figures of the accompanying drawings, in which:
a illustrates a perspective view of a portion of an orthopedic plate system according to one embodiment.
b illustrates a longitudinal-sectional view of the orthopedic plate system from
a illustrates a perspective view of a portion of an orthopedic plate system according to another embodiment.
b illustrates a top view of an orthopedic plate system as in
c illustrates a partial longitudinal-sectional view of an orthopedic plate system as in
d illustrates a perspective view of an orthopedic plate system as in
a illustrates a top view of a portion of an orthopedic plate system according to another embodiment.
b illustrates a perspective view of an orthopedic plate system as in
c illustrates a partial longitudinal-sectional view of an orthopedic plate system as in
a illustrates a top view of a portion of an orthopedic plate system according to another embodiment.
b illustrates a perspective view of an orthopedic plate system as in
c illustrates a longitudinal-sectional view of an orthopedic plate system as in
a illustrates a perspective view of a portion of an orthopedic plate system according to another embodiment.
b illustrates a perspective view of a portion of an orthopedic plate system according to another embodiment.
c illustrates a cross-sectional view taken along lines 6c-6c of
d illustrates a cross-sectional view taken along lines 6d-6d of
e illustrates a partial longitudinal-sectional view of an orthopedic plate system as in
a illustrates a perspective view of a portion of an orthopedic plate system according to another embodiment.
b illustrates a cross-sectional view of an orthopedic plate system as in
a-c illustrate perspective views of an orthopedic plate system according to one embodiment being positioned at various distances from a fractured bone.
a illustrates a perspective view of a portion of an orthopedic plate system according to another embodiment.
b illustrates a cross-sectional view taken along lines 10b-10b of
a illustrates a partial top view of a plate-bushings subsystem, wherein the holes 1113 in the bushings have different diameters, and a side view of the corresponding screws according to several embodiments.
b illustrates partial cross-sectional views taken through the holes 1113 in
a-b illustrate a perspective and a side view, respectively, of an orthopedic plate system, using a “bushing” system comprising two flexible members 1301, to allow micromotion along one axis only, in accordance to another embodiment.
What follows is a detailed description of the preferred embodiments of the invention in which the invention may be practiced. Reference will be made to the attached drawings, and the information included in the drawings is part of this detailed description. The specific preferred embodiments of the invention, which will be described herein, are presented for exemplification purposes, and not for limitation purposes. It should be understood that structural and/or logical modifications could be made by someone of ordinary skills in the art without departing from the scope of the present invention. Therefore, the scope of the present invention is defined only by the accompanying claims and their equivalents.
The terms micro-motion and motion are used herein interchangeably.
a and
The flexible friction lock, as opposed to a rigid friction lock is beneficial to the healing of the bone. The amount, the type and/or direction of the motion the bone or bones will be allowed to experience after the installation of the orthopedic plate system can be pre-determined as will be described in detail below. Thus, this orthopedic plate system gives the surgeon the ability to control what type and how much motion the bone(s) will experience after the surgery according to the specifics of the surgical intervention (i.e., type of fracture, type of bones being secured, etc). The flexible friction lock allows the screw to be fixed in a non-rigid fashion. This promotes carefully controlled and limited micromotion which is beneficial to the healing of the bone.
The flexible friction lock is the result of the screw 102 pressing onto the bushing 101, which is made of an elastic material, and which, in turn, presses onto the surface of the opening 108. The material used for the bushing may be any material acceptable for medical use and having desired elastic properties, such as rubber, silicon rubber, high density polypropylene or others.
One way of controlling how much pressure the screw neck 207 exercises onto the bushing 201, and therefore, how much post-installation motion is allowed, is by pre-determining the diameter of the screw neck 207. For example, when all other factors are equal (e.g., shape and size of the bushing, density and elastic properties of the bushing material, diameter of the bushing hole 290, etc), a greater diameter of the screw neck 207 translates into lower flexibility of the friction lock, and therefore, into less post-installation bone motion allowed. This is because less room remains between the screw neck 207 and the plate 203, and because, all else constant, the more compressed the bushing 201 is, the less elastic it becomes. Hence, for a given bushing 201, should the surgeon determine in a particular case that less post-surgery motion should be experienced by the bone(s) in order to achieve the best healing results, he/she may choose a screw 202 with a neck 207 having a correspondingly greater diameter.
There are obviously many other ways to control the amount of motion the bone(s) will be allowed to experience after surgery. For example, all else constant, the greater the elasticity of the material used for the bushing, the greater the amount of motion allowed. As another example, the greater the thickness of the bushing wall, meaning the distance between the interior surface of the bushing 201, created by the bushing hole 290, and the exterior-opposite surface of the bushing 201, the greater the amount of motion allowed is. As another example, all else constant, the amount of motion allowed can be controlled by varying the degree of tightening of a screw 202 having a tapered neck 1107t as depicted in
Furthermore, the maximum amount of motion the screw 202, and consequently the bone, is allowed may be firmly controlled by pre-determining the area of the exit end of opening 208. As illustrated in
In addition, for example when the bushing 201 has a cylindrical shape, and therefore, naturally allowing equal range of motion radially, the shape of the exit end 205 of opening 208 may be so predetermined and sized as to allow a range of motion that is greater in one plane, and smaller in another, for example, a perpendicular plane. Such a shape may be an oval, oblong or obround shape. Opting for a smaller exit end 205 has a secondary benefit. It creates the shelf 204 around the exit end 205, thus preventing the bushing-screw assembly from pulling out of the plate 203.
In
As depicted in
a and
A similar bone motion controlling effect may be achieved using other oblong bushing such as the rectangular bushing 401r in
c illustrates a partial longitudinal-sectional view of an orthopedic plate system as in
Fifteen degrees range of allowed motion is illustrated in
a illustrates a perspective view of a portion of an orthopedic plate system according to another embodiment. In this embodiment, the assembly created by the bushing 601o′ and the screw 602 is allowed to slide after installation a predetermined distance at a predetermined pressure and in a predetermined direction (longitudinally in
The sliding direction may be predetermined by selecting the location of the size difference between bushing 601o′ and opening 608′ (i.e., lengthwise as in
Also, as suggested in
And, as explained above, the surgeon may also control the direction and the maximum amount of sliding by choosing the appropriate plate design (e.g., as the one in
It should be apparent from
It should also be apparent that one or more static screws 602a may initially be used. After the surgery, a few weeks (or months) into the healing, the static screw(s) 602a may be removed and the plate becomes dynamic. Instead of immediately compressing after static screw removal, the fracture and/or the bone will only compress to the degree that the forces exceed the pre-determined force/pressure necessary to slide the bushing(s) 601o′ along the sliding opening 608′ in the plate 603. Once compressed, the fracture is again exposed to intermittent compression as the flexible screw/bushing construct is exposed to force, and then the screw 602 springs back when the pressure is removed. This intermittent stress has been shown to enhance bone healing.
b illustrates a perspective view of a portion of an orthopedic plate system according to another embodiment. This embodiment is very similar with the one described above and depicted in
c illustrates a cross-sectional view taken along lines 6c-6c of
d illustrates a cross-sectional view taken along lines 6d-6d of
e illustrates a partial longitudinal-sectional view of an orthopedic plate system as in
The surgeon may also skip the use of static crews, whether without or with bushing, and use only dynamic screws 602 and bushings 601v. Hence, depending on the particular fracture and/or bones, the surgeon may plan for a one-phase, two-phase, three-phase, etc, process for securing the fracture/bone(s) and promoting healing. For example, the surgeon could start with true static screws (no bushing at all) and dynamic screws, continue in phase two by replacing the true static screws with less “static” screws (having a small bushing) and ending with phase three by removing the less “static” screws so that the fracture/bone(s) may experience increased degree of motion as predetermined to be permitted by the dynamic screws. Thus, in this example, the fracture and/or the bone(s) will experience a progressive degree of motion: from zero (phase one), to some degree of motion (phase two), and finally, to an increased, yet controlled degree of motion (phase three).
It should be understood that the direction and degree of flexibility for the less “static” screws 602a (having a bushing 601a) may be also predetermined, the same way as for the dynamic screw/bushing assembly 602/601v, as described earlier under
a and 7b illustrate, respectively, a perspective and a cross-sectional view of a portion of an orthopedic plate system according to another embodiment. The advantage of this embodiment is that, besides achieving flexible/dynamic friction lock, the surgeon has a wide range of options as to what angle to insert screws 702 at into the bone. This is facilitated by the spherical bushings 701s.
a-c illustrate perspective views of an orthopedic plate system 800 according to one embodiment being positioned at various distances from a fractured bone 810. The advantage of the flexible friction lock plate is that the surgeon may choose to lock the plate at any of the positions depicted in the three figures or at any other intermediate positions. The position depicted in
a illustrates a perspective view of a portion of an orthopedic plate system according to another embodiment. It may be desirable to preload the bushing on the plate. It can be mounted with adhesive and/or mechanically. An optional mechanical mounting of a sample cylindrical bushing is depicted here, which is better understood by turning to
b illustrates a cross-sectional view taken along lines 10b-10b of
a illustrates a partial top view of a plate-bushings subsystem formed by the plate 1103 and bushings 1101, wherein the holes 1113 in the bushings 1101 have different diameters, and a side view of the corresponding screws 1102 having different diameters at their neck section 1107, according to several embodiments. For exemplification purposes, the three bushings are oval in shape, made of the same material and of equal exterior size. However, starting from the left, their holes 1113 are increasingly smaller in diameter. This means that after the installation of the corresponding screws 1102 (as seen in
By analogy, a higher degree of motion will be allowed by the assembly formed by the bushing 1101″ and the screw 1102″ and the highest degree of motion will be allowed by the assembly formed by the bushing 1101′ and the screw 1102′. Thus, for the same size and shape of opening 1108, the surgeon may choose screw-bushing assemblies which allow the most desirable degree of motion in order to enhance the healing of the particular fracture and/or bone(s).
Many other configuration possibilities exist, which may afford the surgeon the option of choosing a screw-bushing assembly which will allow the desired amount and direction of motion for the particular case. For example, bushings of the same shape and size may be made available to the surgeon, which are made of different elastic materials or of the same elastic material but of different densities, and therefore, each bushing being capable of enabling a different degree of motion. For easy use, the bushings may be labeled to indicate the degree of motion allowed (e.g., one degree, two degrees, etc). In another example, a plurality of screws having various diameters at their neck section, may accompany a particular bushing. Thus, as described before under
It is to be understood that the various bushing configurations described in this description, may be pre-mounted on the plate, on the screw, or, it may be used as a washer that's added at the time of insertion.
As illustrated in
Other flexible/elastic structures may be used between the screw and the plate, as long as they are suited for achieving controlled post-surgery motion of the screws, and therefore, of the bone to enhance healing. For example, spring-based structures may be used. Such a structure may comprise, for example, an interior cylinder with an opening for the passage of the screw, an exterior cylinder larger in diameter than the interior cylinder and configured to sit in the plate opening, and a plurality of springs positioned between the two cylinders in, for example, a perpendicular (to both cylinders) arrangement. The interior cylinder and the exterior cylinder may need to have a side cut (or two side cuts, hence, converting in half-cylinders) so that they expand during the insertion of the screw. Such a structure would the functional equivalent of a cylindrical bushing described earlier in this description. Similarly, the functional equivalent of an oval, obround, spherical, etc bushing may be constructed.
Although specific embodiments have been illustrated and described herein for the purpose of disclosing the preferred embodiments, someone of ordinary skills in the art will easily detect alternate embodiments and/or equivalent variations, which may be capable of achieving the same results, and which may be substituted for the specific embodiments illustrated and described herein without departing from the scope of the present invention. Therefore, the scope of this application is intended to cover alternate embodiments and/or equivalent variations of the specific embodiments illustrated and/or described herein. Hence, the scope of the present invention is defined only by the accompanying claims and their equivalents.