FIELD OF THE INVENTION
This invention relates to implantable medical devices and, in particular, to energy transfer devices, systems and methods for implantable medical devices.
BACKGROUND OF THE INVENTION
Implantable medical devices for producing a therapeutic result in a patient are well known. Examples of such implantable medical devices include implantable drug infusion pumps, implantable neurostimulators, implantable cardioverters, implantable cardiac pacemakers, implantable defibrillators and cochlear implants. Of course, it is recognized that other implantable medical devices are envisioned which utilize energy delivered or transferred from an external device.
A common element in all of these implantable medical devices is the need for electrical power in the implanted medical device. The implanted medical device requires electrical power to perform its therapeutic function whether it be driving an electrical infusion pump, providing an electrical neurostimulation pulse or providing an electrical cardiac stimulation pulse. This electrical power is derived from a power source.
Typically, a power source for an implantable medical device can take one of two forms. The first form utilizes an external power source that transcutaneously delivers energy via wires or radio frequency energy. Having electrical wires which perforate the skin is disadvantageous due, in part, to the risk of infection. Further, continuously coupling patients to an external power for therapy is, at least, a large inconvenience. The second form utilizes single cell batteries as the source of energy of the implantable medical device. This can be effective for low power applications, such as pacing devices. However, such single cell batteries usually do not supply the lasting power required to perform new therapies in newer implantable medical devices. In some cases, such as an implantable artificial heart, a single cell battery might last the patient only a few hours. In other, less extreme cases, a single cell unit might expel all or nearly all of its energy in less than a year. This is not desirable due to the need to explant and re-implant the implantable medical device or a portion of the device. One solution is for electrical power to be transcutaneously transferred through the use of inductive coupling. Such electrical power or energy can optionally be stored in a rechargeable battery. In this form, an internal power source, such as a battery, can be used for direct electrical power to the implanted medical device. When the battery has expended, or nearly expended, its capacity, the battery can be recharged transcutaneously, via inductive coupling from an external power source temporarily positioned on the surface of the skin.
Transcutaneous energy transfer through the use of inductive coupling involves the placement of two coils positioned in close proximity to each other on opposite sides of the cutaneous boundary. The internal coil, or secondary coil, is part of or otherwise electrically associated with the implanted medical device. The external coil, or primary coil, is associated with the external power source or external charger, or recharger. The primary coil is driven with an alternating current. A current is induced in the secondary coil through inductive coupling. This current can then be used to power the implanted medical device or to charge, or recharge, an internal power source, or a combination of the two.
Since it may be necessary for the patient to operate the external power source and since the patient may not be medically skilled nor electronically skilled, it is desirable that the external power source have a user interface that is both intuitive and simple to use and operate.
SUMMARY OF THE INVENTION
In an embodiment, the present invention provides an implantable medical device system. An implantable medical device provides a therapeutic output to a patient. An external patient control unit can be operatively coupled via telemetry with the implantable medical device allowing the patient to control the therapeutic output of the implantable medical device. An external energy transfer unit can be operatively coupled to the implantable medical device to transcutaneously transfer energy to the implantable medical device. The external patient control unit has a user interface operable by the patient having patient controls and display icons indicative of a status of operation of the patient control unit. The energy transfer unit has a user interface operable by the patient having patient controls and display icons indicative of a status of operation of the energy transfer unit. At least some of the patient controls and the display icons of the energy transfer unit are common with at least some of the patient controls and the display icons of the patient control unit.
In another embodiment, the present invention provides an external energy transfer unit adapted to transcutaneously transfer energy to an implantable medical device for providing a therapeutic output to a patient. The implantable medical device is capable of communication with an external patient control unit allowing the patient to control the therapeutic output of the implantable medical device. The external patient control unit has a user interface operable by the patient having patient controls and display icons indicative of a status of operation of the patient control unit. A user interface operable by the patient has patient controls and display icons indicative of a status of operation of the energy transfer unit. At least some of the patient controls and the display icons of the energy transfer unit are common with at least some of the patient controls and the display icons of the patient control unit.
In another embodiment, the present invention provides a method of providing therapeutic output to a patient using an implantable medical device. An external patient control unit is operatively coupled via telemetry with the implantable medical device allowing the patient to control the therapeutic output of the implantable medical device. An external energy transfer unit is operatively coupled to the implantable medical device to transcutaneously transfer energy to the implantable medical device. The external patient control unit has a user interface operable by the patient having patient controls and display icons indicative of a status of operation of the patient control unit. The energy transfer unit has a user interface operable by the patient and has patient controls and display icons indicative of a status of operation of the energy transfer unit. At least some of the patient controls and the display icons of the energy transfer unit are common with at least some of the patient controls and the display icons of the patient control unit.
In a preferred embodiment, at least one of the display icons is an icon representative of a state of charging of an energy storage device in the implantable medical device.
In a preferred embodiment, at least one of the patient controls is a control for adjusting the therapeutic output of the implantable medical device.
In a preferred embodiment, the external energy transfer unit inductively couples energy to the implantable medical device.
In a preferred embodiment, the implantable medical device has a rechargeable energy storage device and wherein the energy transfer unit charges the rechargeable energy storage device.
In another embodiment, the present invention provides an implantable medical device system. An implantable medical device provides a therapeutic output to a patient. An external energy transfer unit can be operatively coupled to transcutaneously transfer energy to the implantable medical device. The energy transfer unit is operable by the patient with less than three operative controls to control energy transfer from the external energy transfer unit to the implantable medical device.
In another embodiment, the present invention provides an external energy transfer unit adapted to transcutaneously transfer energy to an implantable medical device for providing a therapeutic output to a patient. Energy transfer circuitry transcutaneously transfers energy to the implantable medical device. Less than three operative controls control energy transfer from the external energy transfer unit to the implantable medical device.
In another embodiment, the present invention provides a method providing a therapeutic output to a patient with an implantable medical device. An external energy transfer unit is operatively coupled with the implantable medical device to transcutaneously transfer energy to the implantable medical device. The energy transfer unit is controlled with less than three operative controls to control energy transfer from the external energy transfer unit to the implantable medical device.
In a preferred embodiment, the energy transfer unit is operable with two operative controls to control energy transfer from the external energy transfer unit to the implantable medical device.
In a preferred embodiment, the two operative controls comprise a start control and a stop control.
In a preferred embodiment, the start control is a start button and wherein the stop control is a stop button.
In a preferred embodiment, the energy transfer unit also has operative controls to start and stop the therapeutic output of the implantable medical device.
In a preferred embodiment, the energy transfer unit also has a control to enable and silence audio feedback from the energy transfer unit.
In a preferred embodiment, the external energy transfer unit inductively couples energy to the implantable medical device.
In a preferred embodiment, the implantable medical device has a rechargeable energy storage device and wherein the energy transfer unit charges the rechargeable energy storage device.
In another embodiment, the present invention provides an implantable medical device system. An implantable medical device provides a therapeutic output to a patient. The implantable medical device is capable of receiving inductively coupled energy with a secondary coil and is capable of being transcutaneously controlled using an internal telemetry coil. An external antenna has a primary coil adapted to inductively transfer energy to the secondary coil of the implantable medical device when the external antenna is externally placed in proximity of the secondary coil. An external energy transfer unit is adapted to be operatively coupled by a cord to the external antenna driving the primary coil to inductively transfer energy to the implantable medical device. The energy transfer unit has an external telemetry coil allowing the energy transfer unit to communicate with the implantable medical device through the internal telemetry coil in order to at least partially control the therapeutic output of the implantable medical device.
In another embodiment, the present invention provides an external energy transfer system adapted to transcutaneously transfer energy to an implantable medical device for providing a therapeutic output to a patient. The implantable medical device is capable of receiving inductively coupled energy with a secondary coil and is capable of being transcutaneously controlled using an internal telemetry coil. An external antenna has a primary coil adapted to inductively transfer energy to the secondary coil of the implantable medical device when the external antenna is externally placed in proximity of the secondary coil. An external energy transfer unit is adapted to be operatively coupled by a cord to the external antenna driving the primary coil to inductively transfer energy to the implantable medical device. The energy transfer unit has an external telemetry coil allowing the energy transfer unit to communicate with the implantable medical device through the internal telemetry coil in order to at least partially control the therapeutic output of the implantable medical device.
In another embodiment, the present invention provides a method of controlling an implantable medical device for providing a therapeutic output. The implantable medical device is capable of receiving inductively coupled energy with a secondary coil and is capable of being transcutaneously controlled using an internal telemetry coil. Energy is inductively transferred to the secondary coil with an external antenna having a primary coil when the external antenna is externally placed in proximity of the secondary coil. The primary coil is driven with an external energy transfer unit operatively coupled by a cord to the external antenna to inductively transfer energy to the implantable medical device. An external telemetry coil in the energy transfer unit communicates with the implantable medical device through the internal telemetry coil in order to at least partially control the therapeutic output of the implantable medical device.
In a preferred embodiment, the inductively transferring step and the communicating step occur simultaneously.
In a preferred embodiment, the energy transfer unit can control the implantable medical device to turn the therapeutic output on and off.
In a preferred embodiment, the energy transfer unit can control the implantable medical device using the external telemetry coil while transferring energy to the implantable medical using the primary coil of the antenna.
In a preferred embodiment, the implantable medical device has a rechargeable energy storage device and wherein the energy transfer unit charges the rechargeable energy storage device.
BRIEF DESCRIPTION OF THE DRAWINGS
FIG. 1 is a block diagram of an external programmer and an external charging device for an implantable medical device such as can be utilized in the present invention;
FIG. 2 is a detailed block diagram of an implantable medical device implanted sub-cutaneously and an associated external charging device in accordance with an embodiment of the present invention;
FIG. 3 illustrates the start-up flow of an external programmer usable in conjunction with an implantable medical device;
FIG. 4 illustrates operation flow of an external programmer usable in conjunction with an implantable medical device;
FIG. 5 illustrates a therapy status screen of an external programmer;
FIG. 6 illustrates access to menu selection of an external programmer;
FIG. 7 illustrate group selections of an external programmer;
FIG. 8 illustrate parameter selections of an external programmer;
FIG. 9A illustrate screens of an external programmer;
FIG. 9B illustrate messages of an external programmer;
FIG. 9C illustrate warnings of an external programmer;
FIG. 10 illustrate menu selections of an external programmer;
FIG. 11 illustrates audio status of an external programmer;
FIG. 12 illustrates contrast of an external programmer;
FIG. 13 illustrates INS/ENS time of an external programmer;
FIG. 14 illustrate time/number formats of an external programmer;
FIG. 15 illustrate patient programmer information;
FIG. 16 illustrates INS/ENS information;
FIG. 16A illustrates Telemetry N information;
FIG. 17 illustrates the navigation flow of an external charger for recharging the INS;
FIG. 18 illustrates the navigation flow of an external charger for recharging the external power source;
FIG. 19 illustrates screen shots of charge status of an external charger;
FIG. 20 illustrates status messages displayed during charging of the INS;
FIG. 21 illustrates external power source charge status screens;
FIG. 22 illustrates status messages displayed during recharging of the external power source; and
FIGS. 23A and 23B illustrate screens, information messages and warnings.
DETAILED DESCRIPTION OF THE INVENTION
In FIG. 1, external programming unit 2 is attachable via cord 4 to an external telemetry coil 6. External programming unit 2 can program operations implantable medical device 16 (FIG. 2) in a conventional manner using external telemetry coil 6. External antenna 52 is attachable via cord 56 to external charging device 48 is inductively transfer power to implantable medical device 16 when external antenna 52 is placed in proximity of a secondary coil 24 associated with implantable medical device 16. External charging device 48 receives AC power from cord 8 through transformer 10. External antenna 52 can be held in position on patient 18 with belt 12.
FIG. 2 illustrates an embodiment of implantable medical device 16 situated under cutaneous boundary 38. Charging regulation and therapy control is conventional. Implantable medical device 16 also has internal telemetry coil 44 configured in conventional manner to communicate through external telemetry coil 46 to an external programming device 2, charging unit 50 or other device in a conventional manner in order to both program and control implantable medical device and to externally obtain information from implantable medical device 16 once implantable medical device has been implanted. Internal telemetry coil 44, rectangular in shape with dimensions of 1.85 inches (4.7 centimeters) by 1.89 inches (4.8 centimeters) constructed from 150 turns of 43 AWG wire, is sized to be larger than the diameter of secondary charging coil 34. Secondary coil 34 is constructed with 182 turns of 30 AWG wire with an inside diameter of 0.72 inches (1.83 centimeters) and an outside diameter of 1.43 inches (3.63 centimeters) with a height of 0.075 inches (0.19 centimeters). Magnetic shield 36 is positioned between secondary charging coil 34 and housing 32 and sized to cover the footprint of secondary charging coil 34.
Internal telemetry coil 44, having a larger diameter than secondary coil 34, is not completely covered by magnetic shield 36 allowing implantable medical device 16 to communicate with the external programming device with internal telemetry coil 44 in spite of the presence of magnetic shield 36.
Rechargeable power source 24 can be charged while implantable medical device 16 is in place in a patient through the use of external charging device 48. In a preferred embodiment, external charging device 48 consists of charging unit 50 and external antenna 52. Charging unit 50 contains the electronics necessary to drive primary coil 54 with an oscillating current in order to induce current in secondary coil 34 when primary coil 54 is placed in the proximity of secondary coil 34. Charging unit 50 is operatively coupled to primary coil by cable 56. In an alternative embodiment, charging unit 50 and antenna 52 may be combined into a single unit. Antenna 52 also contains external telemetry coil 46 which may be operatively coupled to charging unit 50 if it is desired to communicate to or from implantable medical device 16 with external charging device 48. Alternatively, antenna 52 may contain external telemetry coil 46 which can be operatively coupled to external programming device 2, either individually or together with external charging unit 48.
The user interface of external programming unit 2 is illustrated by reference to FIG. 3 through FIG. 16. The following terms are used throughout the description of the user interface of external programming unit 2.
As noted below in the explanation of acronyms utilized in the following paragraphs and table, DTC refers to desktop charger which is used to denote the AC power supply (transformer 10) and power cord (cord 8). The DTC is “docked” when external charger 48 is connected to an source of AC power. The DTC is “undocked” when external charger 48 is disconnected from a source of AC power.
- DTC Desktop Charger
- ENS External Neurostimulator
- INS Implanted Neurostimulator
- INSR Another name for the RX1 Patient Recharger. Acronym stands for Implanted Neurostimulator Recharger
- LCD Liquid Crystal Display
Refer to FIG. 3 and FIG. 4 for start-up flow and navigation flow, respectively.
User interface components in the keypad are shown below.
The LCD is backlit and consists of 96 (Horizontal)×58 (Vertical) dots with asymmetrical dots 0.35 (H)×0.4 (V). The active area on the LCD is 33.3 mm (H)×23.2 mm (V). Four colors of display available. These are named below (from lightest to darkest):
- White
- Light Gray
- Dark Gray
- Black
Icons are shown in Tables 1-6 below.
Therapy status screen display layout is divided vertically into three lines of information: status line, groups line and parameters line as shown in FIG. 5. When a successful initial interrogation has been performed, software will display the Therapy Status Screen of the active set with the following highlighting priority: parameters line if displayed; groups line if displayed; and status line.
Scroll up and scroll down keys will be used to access the status line, groups line and parameters line on the display. A box highlights user selection. Scroll left and right will be used to cycle through available choices for a selected line. Software will display a “real-time” INS charge level on the status line. The charge level will be updated and displayed whenever a parameter adjustment is made. Software will display programmer battery status on the status line.
Access to Menu Selections is from the status line are shown in FIG. 6. Initial interrogation will include data on all groups currently available to the patient (“Baseline” setting).
Software will display labels for all groups the user can activate, including those containing no adjustable parameters. Refer to FIG. 7 for Group Selections. All unnecessary text will be removed from the display: e.g., if there is only 1 group, software will not display the groups line. Software will not display parameters that the user cannot adjust: i.e., the clinician programmer disallows access or upper and lower limits are the same. The user will access all available groups by scrolling through a linear list: A, B, C, . . . , Z. When the user scrolls to a different group, software will display the first adjustable parameter in the group. The user will scroll left/right through available groups to access an inactive group.
Software will display all adjustable parameters for inactive groups. The user will scroll down to highlight parameters and scroll left or right to view them. If a user presses Therapy Off or Therapy On when the screen displays an inactive group, software will turn the INS off, and display the Therapy Status Screen with the active group highlighted. A check in its check box will be used to indicate an active group (check_on); a check box without a check will be used to indicate inactive groups (check_off). The user will press the Sync key to activate a group at the displayed parameter value(s).
The neurostimulator does not have to be on to activate a new group. The user can activate a group when therapy is off. The user can activate a new group when the groups line is highlighted or when the parameters line is highlighted. The check box will be updated whenever a new group is activated.
Software will display labels for all parameters the user can adjust as shown in FIG. 8, Parameter Selections. Software will display the first adjustable parameter of the active group when a Sync has been performed for an initial interrogation. The user will access all adjustable parameters by scrolling through a linear list. The order of scrolling through adjustable parameters in a group is Amp1, Amp2, Amp3, Amp4, PW1, PW2, PW3, PW4, Rate, Amp1, . . . Refer to FIG. 8, Parameter Selections. Increase/decrease keys will be inactive until a session has been initiated—i.e., the INS has been interrogated. When the user presses and holds the increase/decrease keys, software will send a change to the neurostimulator every 12 second. When the user changes a parameter, software will display the updated parameter value.
Therapy control events are described in screens, information messages and warnings that are displayed on the programmer as shown in FIGS. 9A, 9B and 9C. Operational requirements are described in tables that specify conditions for entry and expected changes in control.
|
|
User Control/System EventSystem Response: Table 10 reference
|
|
Power Down Screen (blank display)
Screen Entry: User Control/System Event
System timeout or Power Button keypress
Power buttonStart Screen, 14, 15
Therapy On1, 2, 3, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 20, 23, 24
Therapy Off1, 2, 3, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 20, 23, 24
Sync1, 2, 3, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 20, 23, 24
Scroll up/down/Scroll left/rightIgnore
Increase/DecreaseIgnore
Replace programmer batteriesSplash Screen
Start Screen
Screen Entry: User Control/System Event
Power button
Therapy On1, 2, 3, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 20, 23, 24
Therapy Off1, 2, 3, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 20, 23, 24
Sync1, 2, 3, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 20, 23, 24
Power button/System timeoutPower Down Screen
Scroll up/down/Scroll left/rightIgnore
Increase/DecreaseIgnore
Increase + Decrease and hold10
for 2-3 seconds
Therapy Status Screen
Screen Entry: User Control/System Event
Telemetry success
Therapy On1, 2, 3, 4, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 20, 23, 24
Therapy Off1, 2, 3, 4, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 20, 23, 24
Sync1, 2, 3, 4, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 20, 23, 24
Increase1, 2, 3, 4, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 17, 18, 19,
20, 23, 24
Decrease1, 2, 3, 4, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 16, 19, 20,
23, 24
Scroll left/right from status lineAudio Screen, Time/Number Formats
Power button/System timeoutPower Down Screen
|
The i symbol will be displayed on all information messages as shown in FIG. 9B. The user can clear all messages by pressing the 4-way button (Scroll up/down, Scroll left/right), Therapy Off to turn stimulation off or the Power button to turn the display off. If the system times out, the display will turn off (Power Down Screen). Refer to Table 7.
TABLE 7
|
|
Processing Requirements for Information Messages
MessageUser Control/System Response: Feedback/Next Screen
|
INS Battery Low4-way button: Therapy Status
Stim Off: Table 10-1, 2, 3, 6, 11, 12, 13, 14, 15, 23, 24
Stim On: ignore
Sync: ignore
Increase/Decrease: ignore
System timeout or Power button: Power Down
ENS Battery Low4-way button: Therapy Status
Stim Off: Table 10-1, 2, 3, 8, 9, 12, 13, 14, 15, 20, 23, 24
Stim On: ignore
Sync: ignore
Increase/Decrease: ignore
System timeout or Power button: Power Down
Programmer Battery4-way button: Previous screen
LowStim Off: Table 10-1, 2, 3, 5, 6, 7, 8, 9, 11, 12, 13, 15, 20, 23, 24
Stim On: ignore
Sync: ignore
Increase/Decrease: ignore
System timeout or Power button: Power Down
Amp at Upper Limit4-way button: Therapy Status
Stim Off: Table 10-1, 2, 3, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 20, 23,
24
Stim On: ignore
Sync: ignore
Increase/Decrease: ignore
System timeout or Power button: Power Down
PW at Upper Limit4-way button: Therapy Status
Stim Off: Table 10-1, 2, 3, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 20, 23,
24
Stim On: ignore
Sync: ignore
Increase/Decrease: ignore
System timeout or Power button: Power Down
Rate at Upper Limit4-way button: Therapy Status
Stim Off: Table 10-1, 2, 3, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 20, 23,
24
Stim On: ignore
Sync: ignore
Increase/Decrease: ignore
System timeout or Power button: Power Down
Amp at Lower Limit4-way button: Therapy Status
Stim Off: Table 10-1, 2, 3, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 20, 23,
24
Stim On: ignore
Sync: ignore
Increase/Decrease: ignore
System timeout or Power button: Power Down
PW at Lower Limit4-way button: Therapy Status
Stim Off: Table 10-1, 2, 3, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 20, 23,
24
Stim On: ignore
Sync: ignore
Increase/Decrease: ignore
System timeout or Power button: Power Down
Rate at Lower Limit4-way button: Therapy Status
Stim Off: Table 10-1, 2, 3, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 20, 23,
24
Stim On: ignore
Sync: ignore
Increase/Decrease: ignore
System timeout or Power button: Power Down
Poor Communication -Allow a repeat of the last keypress for Stim On, Sync, Increase and
INSDecrease
4-way button: Sync Up Required
Stim Off: Table 10-1, 2, 3, 5, 6, 11, 12, 13, 14, 15
Stim On: Table 10-1, 2, 3, 5, 6, 11, 12, 13, 14, 15
Sync: Table 10-1, 2, 3, 5, 6, 11, 12, 13, 14, 15
Increase/Decrease: Table 10-1, 2, 3, 5, 6, 11, 12, 13, 14, 15, 16,
17, 18, 19
System timeout or Power button: Power Down
Poor Communication -Allow a repeat of the last keypress for Stim On, Sync, Increase and
ENSDecrease
4-way button: Sync Up Required
Stim Off: Table 10-2, 7, 8, 9, 12, 13, 14, 15, 20, 23, 24
Stim On: Table 10-2, 7, 8, 9, 12, 13, 14, 15, 20, 23, 24
Sync: Table 10-2, 7, 8, 9, 12, 13, 14, 15, 20, 23, 24
Increase/Decrease: Table 10-2, 7, 8, 9, 12, 13, 14, 15, 16, 17,
18, 19, 20, 23, 24
System timeout or Power button: Power Down
Sync Up4-way button: Therapy Status
Stim Off: Table 10-1, 2, 3, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 20, 23,
24
Stim On: Table 10-1, 2, 3, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 20, 23,
24
Sync: Table 10-1, 2, 3, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 20
Increase/Decrease: ignore
System timeout or Power button: Power Down
Turn Stimulation On4-way button: Therapy Status
Stim Off: ignore
Stim On: Table 10-1, 2, 3, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 20, 23,
24
Sync: ignore
Increase/Decrease: ignore
System timeout or Power button: Power Down
Connect ENS Cable4-way button: Sync Up Required
Stim Off: Table 10-2, 7, 8, 9, 12, 13, 14, 15, 20, 23, 24
Stim On: Table 10-2, 7, 8, 9, 12, 13, 14, 15, 20, 23, 24
Sync: ignore
Increase/Decrease: ignore
System timeout or Power button: Power Down
|
The exclamation mark symbol will be displayed on all warnings as shown in FIG. 9C. All warnings will remain on the display until the user takes appropriate action. Refer to Table 8.
TABLE 8
|
|
Processing Requirements for Warning Messages
User Action/System Response:
MessageFeedback/NextScreen
|
Recharge StimulatorSystem timeout or Power button: Power Down
Replace ENS BatterySystem timeout or Power button: Power Down
Replace ProgrammerReplace programmer battery: Start Screen
BatterySystem timeout or Power button: Power Down
Sync Up RequiredSync button or Therapy On or Therapy Off:
Therapy Status
System timeout or Power button: Power Down
Call Your DoctorTherapy Off and telemetry success with
audio feedback: Call Your Doctor
Therapy Off and telemetry failure with
audio feedback: Call Your Doctor
System timeout or Power button: Power Down
Call Your Doctor - EOSSystem timeout or Power button: Power Down
Call Your Doctor - PORSystem timeout or Power button: Power Down
|
“In the box” messages are described in Table 9.
“In the Box” Messages
- Software will notify the user with 3 beeps when “In the Box” messages are displayed
All “In the Box” messages will remain on the display until the system times out or the user presses the Power button to turn off the display. Refer to Table 9.
TABLE 9
|
|
Processing Requirements for “In the Box” Messages
User Action/System Response: Feedback/Next
MessageScreen
|
INS in the BoxSystem timeout or Power button: Power Down
ENS in the BoxSystem timeout or Power button: Power Down
|
Therapy control events are described in Table 10.
TABLE 10
|
|
Processing Requirements for Therapy Control Events and Device Communication
User Control/System EventSystem Response
|
1.Communicating with the INSDisplay Communicating - INS Message
2.Successful telemetryDisplay Therapy Status Screen
3.Telemetry failure - INSDisplay Poor Communication - INS
Message
4.Software is unable to refresh/update theDisplay Sync Up Required Warning
Therapy Status Screen after a telemetry
failure
5.INS battery lowDisplay INS Battery Low Message
6.INS battery depletedDisplay Recharge Stimulator Warning
7.ENS battery low or batteryDisplay ENS Battery Low Message
disconnected
8.ENS battery depletedDisplay Replace ENS Battery Warning
9.ENS screening cable disconnectedDisplay Connect ENS Cable Message
10.User presses Increase + Decrease andDisplay Patient Information Screen
holds for 2-3 seconds(Error! Reference source not found.)
11.INS error - EOSDisplay Call Your Doctor - EOS Warning
12.INS error - battery voltage too high;/Display Call Your Doctor Warning
ENS error - POR, battery voltage too
high
13.System Error and WarningDisplay Call Your Doctor Warning
14.Patient programmer battery lowDisplay Programmer Battery Low
Message
15.Patient programmer battery depletedDisplay Replace Programmer Battery
Warning
16.Parameter at lower limitDisplay At Lower Limit Message
17.Parameter at upper limit or out of regDisplay At Upper Limit Message
18.Increase with therapy offDisplay Turn Stimulation On Message
19.Increase/decrease parameter of inactiveDisplay Sync Up Message
group
20.ENS error - POR, battery voltage tooDisplay Communicating - ENS,
lowfollowed by INS/ENS Time with 12:00
displayed
21.Communicating via IRDisplay IR Communication Screen
22.Replace programmer batteriesDisplay Splash Screen
23.Communicating with the ENSDisplay Communicating - ENS
Message
24.Telemetry failure - ENSDisplay Poor Communication - ENS
Message
25.INS error - locked normalDisplay INS in the Box Message
26.ENS error - ENS has no programs orDisplay ENS in the Box Message
groups
27.INS error - PORDisplay Call Your Doctor - POR
Warning
|
Access to menu selections is from the status line on the Therapy Screen as shown in FIG. 10.
The display is divided vertically into two lines of information: menu option line and value line. Scroll up and down keys will be used to access the menu option line and the value line on the display. A box highlights user selection. Scroll right and left and increase/decrease keys will be used to change values for Audio, Contrast, Time/Number Format and the Information options. In addition, Increase/decrease keys will be used to change values.
Software will display screens for the menu options in Table 11 Menu Selections.
TABLE 111
|
|
Menu Selections and Navigation Flow
Menu ItemReferences
|
AudioFIG. 11, Table 12-1, 3, 9
ContrastFIG. 12, Table 12-1, 3, 4, 6, 7, 9
INS/ENS TimeFIG. 13, Table 12-1, 2, 4, 7, 8, 9
Time/Number FormatFIG. 14, Table 12-1, 3, 9
Patient Programmer InformationFIG. 15, Table 12-1, 2, 9, 10
INS/ENS InformationFIG. 16, Table 12-1, 2, 9, 10
Telemetry N InformationFIG. 16A, Table 12-1, 2, 9, 10
|
The user interface of the external charger is described below. A description of the user interface for an external recharger for an implantable medical device is described. The appearance of the screens and how users interact with these screens are defined.
The following terms are used throughout the description of the user interface for the external charger.
- DTC Desktop Charger
- ENS External Neurostimulator
- INS Implanted Neurostimulator
- INSR Another name for the RX1 Patient Recharger. Acronym stands for Implanted
- Neurostimulator Recharger
- LCD Liquid Crystal Display
Refer to FIG. 17 and FIG. 18 for navigation flow.
FIG. 17 illustrates a process for charging rechargeable power source 24 in implantable medical device 16 by using external charging device 48.
In order to charge rechargeable power source 24, a patient may use belt 12 to attach external antenna 52 to the patient's body at approximately the location of implanted medical device 16. It is desirable that primary coil 54 of external antenna 52 be aligned as closely as possible with secondary coil 34 of implantable medical device 16. Techniques for attaching external antenna 52 to the patient's body and positioning primary coil 54 in close proximity to secondary coil 34 can be found in co-pending U.S. patent application Ser. No. 10/837,506, filed Apr. 30, 2004, the entire contents of which are hereby incorporated by reference. This application claims priority from U.S. Provisional Patent Application Ser. No. 60/508,174, filed Oct. 2, 2003.
After affixing external antenna 52 to the approximate location of implantable medical device 16, the patient (or other user such as a medical professional or family member assisting the patient) may then press (step 110) the “Start Charge” button 90 (FIG. 1) on charging unit 50.
Charging unit 50 then determines (step 112) if telemetry with implantable medical device 16 is successful. If telemetry is unsuccessful (step 114), an error message is generated such as a telemetry timeout (Reposition Antenna—Press Start Charge), system error (Antenna Too Hot) or external charging device 48 error due to INSR battery low or INSR battery depleted and external charging device displays the appropriate message. If telemetry is successful, then charging unit 50 waits (step 116) and the appropriate charge status may be indicated (step 118) in the display of charging unit 50.
If charging of rechargeable power source 24 is not yet complete (step 120), charging unit 50 returns through step 118. The surface temperature of antenna 52 is monitored as well as the charging current and battery voltage levels. An error message is generated and charging is suspended if measurements are not within limits.
Once charging of rechargeable power 24 is complete (step 122), charging unit 50 indicates in the display that charging is complete and the user may press the “Stop Charge” button 92 on charging unit 50 to terminate the charging process. Of course, it is to recognized and understood that the “Stop Charge” button 92 may be pressed at any time during the charging process in order to terminate the charging process, i.e., the user need not wait until the complete charging has been indicated.
Charging unit 50 may be operated to charge rechargeable power 24 in implantable medical device 16 by the use of only two buttons, namely the “Start Charge” button 90 and the “Stop Charge” button 92. These are the only two buttons which need be used. While a preferred embodiment of charging unit 50 does contain a third button, namely “Silence” button 94, this third button is not necessary. “Silence” button 94 is used only to disable, or to re-enable, audio output of charging unit 50. Since audio feedback is not required for the charging process, this button may be used or not at the convenience of the user.
In an alternative embodiment, charging unit 50 may itself be powered by a rechargeable power source, such as a battery. Since the charging process could take several hours, it would be desirable.
FIG. 18 illustrates the process used to charge the external charging unit 48. The external charging unit 48 is connected (step 124) to a suitable power source, such as a readily available AC power outlet. A check is made (step 126) to determine whether a “recharge event”, i.e., an alert or alarm condition, has occurred and, if so, issue the appropriate message (step 128). If a “recharge event” has not occurred, the process updates and displays (step 130) the charge status of external charging device 48. If recharge of external charging is not complete (step 132), the process again updates and displays the charging status in step 130. Once charging of external charging device 48 is complete (step 134), charging unit 50 displays a message indicative of a complete charging process.
The LCD is backlit and consists of 96 (Horizontal)×58 (Vertical) dots with asymmetrical dots 0.35 (H)×0.4 (V). The active area on the LCD is 23.2 mm (V)×33.3 mm (H).
Four colors of display are available. These are named below (from lightest to darkest):
- White
- Light Gray
- Dark Gray
- Black
The keypad has the following buttons.
The INS charge status screen display is divided vertically into three lines of information: status line, recharge progress line and recharge efficiency line. Refer to FIG. 19 and FIG. 20.
Therapy On, Therapy Off, Start Charge and Stop Charge are dedicated function keys. When the display is blank and the user presses Therapy On, Therapy Off or Start Charge, software will perform the function and turn the display on. System timeout is <30 seconds.
The INS battery status display has 5 settings: 0%, 25%, 50%, 75% and 100%. During recharge, software will display the appropriate quartiles in solid fill and the next higher quartile in a flashing on/off manner every second. The exception is 100%. Even if the INS battery voltage indicates 100%, software will flash the last quartile until the recharge session is complete.
The INSR charge status screen display is divided vertically into two lines of information: status line and recharge progress line. Refer to FIG. 21, INSR Charge Status Screen, and FIG. 22, Recharging the INSR.
The INSR battery status has 5 settings: 0%, 25%, 50%, 75% and 100%. During recharge, software will display the appropriate quartiles in solid fill and the next higher quartile in a flashing on/off manner every second. The exception is 100%. Even if the INSR battery voltage indicates 100%, software will flash the last quartile until the recharge session is complete.
Recharging events are described in screens, information messages and warnings that are displayed on the recharger as shown in FIG. 23A and FIG. 23B. Operational requirements are described in tables that specify conditions for entry and expected changes in control. Refer to the operational requirements for each screen in this section.
The i symbol will be displayed on all information messages. With the exception of INSR System Information screen, the user can exit the messages by pressing Audio or wait for the system to timeout. Therapy On and Off will always be active when the messages are displayed. When the user presses Therapy On or Off, software will turn stimulation on/off and clear the information message.
The exclamation mark symbol will be displayed on all warnings. All warnings will remain on the display until the user takes appropriate action.
|
|
1-Power Down Screen (blank display)
Screen Entry: User Control/System Event
System timeout from Start Screen - Not Docked
User Control/
System EventSystem Response: Feedback/Next Screen (requirement)
|
Therapy On/OffTelemetry success: Start Screen - Not Docked with Therapy
On/Off icon on status line (Table 17-22)
Telemetry event: refer to Table 18
Start ChargeAntenna temperature OK: Wait Screen (Table 17-1)
Antenna too hot: Antenna Too Hot Message (Table 17-30)
Antenna failure: Call Your Doctor Warning (Table 17-18)
Antenna disconnected: Connect INSR Antenna Message (Table
17-14)
INSR battery low: INSR Battery Low Message (Table 17-8)
INSR battery depleted: Recharge INSR Warning (Table 17-9)
Stop ChargeNo change
AudioStart Screen - Not Docked (Table 17-24, 25)
INSR battery low: INSR Battery Low Message (Table 17-8)
INSR battery depleted: Recharge INSR Warning (Table 17-9)
Attach DTCINSR recharge complete: INSR Recharge Complete Message
(Table 17-11)
INSR recharge incomplete: INSR Charge Status Screen (Table
17-7)
Audio + Stop ChargeINSR System Information Message (Table 17-17)
Audio + Start ChargeAntenna temperature OK: Physician Recharge Screen (Table 17-
and hold for >1021)
secondsAntenna too hot: Physician Recharge - Antenna Too Hot (Table
17-31)
|
2-Start Screen - Not Docked
Screen Entry: User Control/System Event
Therapy On/Off keys when the recharger is powered down
Audio key when the recharger is powered down
Any key from Splash Screen
Detach DTC
System timeout or message clear when DTC is detached
Stop Charge from Physician Recharge Screen or Physician Recharge - Antenna
Too Hot Screen
Host terminates the session, the Reset button is activated
or the user presses Audio from IR Communication Screen
User Control/
System EventSystem Response: Feedback/Next Screen (requirement)
|
Therapy On/OffTelemetry success: Therapy On/Off icon on status line (Table
17-22)
Telemetry event: refer to Table 18
Start ChargeAntenna temperature OK: Wait Screen (Table 17-1)
Antenna too hot: Antenna Too Hot Message (Table 17-30)
Antenna failure: Call Your Doctor Warning (Table 17-18)
Antenna disconnected: Connect INSR Antenna Message (Table
17-14)
INSR battery low: INSR Battery Low Message (Table 17-8)
INSR battery depleted: Recharge INSR Warning (Table 17-9)
Stop ChargeNo change
AudioAudio icon (Table 17-24)
INSR battery lowINSR Battery Low Message (Table 17-8)
INSR battery depletedRecharge INSR Warning (Table 17-9)
System errorCall Your Doctor Warning (Table 17-18)
Attach DTCINSR recharge complete: INSR Recharge Complete Message
(Table 17-11)
INSR recharge incomplete: INSR Charge Status Screen (Table
17-7)
Audio + Stop ChargeINSR System Information Message (Table 17-17)
Audio + Start ChargeAntenna temperature OK: Physician Recharge Screen (Table
and hold for >1017-21)
secondsAntenna too hot: Physician Recharge - Antenna Too Hot
(Table 17-31)
System timeoutPower Down Screen
|
3-Start Screen - Docked
Screen Entry: User Control/System Event
System timeout or message clear
Any key from Splash Screen
Host terminates the session, the Reset button is activated or the user presses Audio
from the Splash Screen
Stop Charge from Physician Recharge Screen or Physician Recharge - Antenna Too Hot
Screen
User Control/
System EventSystem Response: Feedback/Next Screen (requirement)
|
Therapy On/OffTelemetry success: Therapy On/Off icon on status line (Table
17-22)
Telemetry event: refer to Table 18
Start ChargeAntenna temperature OK: Wait Screen (Table 17-1)
Antenna too hot: Antenna Too Hot Message (Table 17-30)
Antenna failure: Call Your Doctor Warning (Table 17-18)
Antenna disconnected: Connect INSR Antenna (Table 17-14)
Stop ChargeNo change
AudioAudio icon (Table 17-24)
System errorCall Your Doctor Warning (Table 17-18)
Detach DTCStart Screen - Not Docked
Audio + Stop ChargeINSR System Information Message (Table 17-17)
Audio + Start ChargeAntenna temperature OK: Physician Recharge Screen (Table
and hold for >1017-21)
secondsAntenna too hot: Physician Recharge - Antenna Too Hot
(Table 17-31)
|
4-Wait Screen
Screen Entry: User Control/System Event
Start Charge
Antenna temperature OK
System timeout or message clear
User Control/
System EventSystem Response: Feedback/Next Screen (requirement)
|
Therapy On/OffNo change
Telemetry successINS Charge Status Screen or INS Charge Status - 90 to 100%
Screen (Table 17-2)
Telemetry failure orReposition Antenna - Press Start Charge Warning (Table 17-12)
poor charging condition
INS error: ERICall Your Doctor - ERI Message (Table 17-19)
INS error: PORCall Your Doctor - POR Message (Table 17-27)
INS error: EOS, batteryCall Your Doctor Warning (Table 17-18)
voltage too high
INSR battery lowINSR Battery Low Message (Table 17-8)
INSR battery depletedRecharge INSR Warning (Table 17-9)
Antenna too hotINS Charge Status - Antenna Too Hot (Table 17-13)
Antenna failureCall Your Doctor Warning (Table 17-18)
Antenna disconnectedConnect INSR Antenna Message (Table 17-14)
System errorCall Your Doctor Warning (Table 17-18)
Start ChargeNo change
Stop Charge (notStart Screen - Not Docked
docked)
Stop Charge (docked)INSR recharge incomplete: INSR Charge Status Screen (Table
17-7)
INSR recharge complete: INSR Recharge Complete Message
(Table 17-11)
AudioNo change
Attach DTCNo change
Detach DTCNo change
Audio + Stop ChargeNo change
Audio + Start ChargeNo change
and hold for >10
seconds
|
5-INS Charge Status Screen
Screen Entry: User Control/System Event
Telemetry success
System timeout or message clear
Antenna temperature OK
User Control/
System EventSystem Response: Feedback/Next Screen (requirement)
|
Therapy On/OffTelemetry success: Therapy On/Off icon on status line (Table
17-22)
Telemetry event: refer to Table 18
Start ChargeAntenna temperature OK: refresh recharge efficiency line
(Table 17-26)
Antenna too hot: INS Charge Status - Antenna Too Hot (Table
17-13)
Stop ChargeCharge complete: INS Recharge Complete Message (Table 17-
6)
Charge incomplete and not INS battery low: INS Recharge
Interrupted Message (Table 17-5)
INS battery low: INS Battery Low Message (Table 17-3)
INS battery depleted: Recharge Stimulator Warning (Table 17-
4)
AudioToggle Audio icon (Table 17-24)
Attach DTCPower source symbol on status line (Table 17-23)
Detach DTCINSR battery low: INSR Battery Low Message (Table 17-8)
INSR battery depleted: Recharge INSR Warning (Table 17-9)
Status line without power source symbol (Table 17-23)
INS recharge completeINS Recharge Complete Message (Table 17-6)
INS error: ERICall Your Doctor ERI Message (Table 17-19)
INS error: PORCall Your Doctor - POR Message (Table 17-27)
INS error: EOS, batteryCall Your Doctor Warning (Table 17-18)
voltage too high
Antenna too hotINS Charge Status - Antenna Too Hot (Table 17-13)
Antenna failureCall Your Doctor Warning (Table 17-18)
Antenna disconnectedConnect INSR Antenna Message (Table 17-14)
INSR battery lowINSR Battery Low Message (Table 17-8)
INSR battery depletedRecharge INSR Warning (Table 17-9)
INSR rechargeINSR Batt1 on status line (Table 17-25)
complete
System errorCall Your Doctor Warning (Table 17-18)
Audio + Stop ChargeNo change
Audio + Start ChargeNo change
and hold for >10
seconds
|
6-INSR Charge Status Screen
Screen Entry: User Control/System Event
Attach DTC from Power Down Screen, Start Screen - Not Docked or Recharge
INSR Warning
Attach DTC from INSR Battery Low Message when not charging INS
System timeout or message clear
User Control/
System EventSystem Response: Feedback/Next Screen (requirement)
|
Therapy On/OffTelemetry success: Therapy On/Off icon on status line (Table
17-22)
Telemetry event: refer to Table 18
Start ChargeAntenna temperature OK: Wait Screen (Table 17-1)
Antenna too hot: Antenna Too Hot Message (Table 17-30)
AudioToggle Audio icon (Table 17-24)
Detach DTCCharge incomplete and not INSR battery low: INSR Recharge
Interrupted Message (Table 17-10)
INSR battery low: INSR Battery Low Message (Table 17-8)
INSR battery depleted: Recharge INSR Warning (Table 17-9)
INSR recharge complete: INSR Recharge Complete Message
(Table 17-11)
INSR rechargeINSR Recharge Complete Message (Table 17-11)
complete
Audio + Start ChargeAntenna temperature OK: Physician Recharge Screen (Table
and hold for >1017-21)
secondsAntenna too hot: Physician Recharge - Antenna Too Hot
(Table 17-31)
System errorCall Your Doctor Warning (Table 17-18)
|
7-Physician Recharge Screen
Screen Entry: User Control/System Event
Audio + Start Charge and hold for >10 seconds
Antenna temperature OK
System timeout or message clear
User Control/
System Event*System Response: Feedback/Next Screen (requirement)
|
SuccessfulINS Charge Status Screen (Table 17-2)
communication with
INS
Timer goes to zero andStart Screen - Not Docked without Therapy On/Off icon on
system times out (notstatus line (Table 17-22)
docked)
Timer goes to zero andINSR recharge incomplete: INSR Charge Status Screen
system times out(Table 17-7)
(docked)INSR recharge complete: INSR Recharge Complete Message
(Table 17-11)
Stop Charge (notStart Screen - Not Docked
docked)
Stop Charge (docked)Start Screen - Docked
Detach DTCINSR battery low: INSR Battery Low Message (Table 17-8)
INSR battery depleted: Recharge INSR Warning (Table 17-9)
Antenna too hotPhysician Recharge - Antenna Too Hot (Table 17-31)
Antenna failureCall Your Doctor Warning (Table 17-18)
Antenna disconnectedConnect INSR Antenna Message (Table 17-14)
System errorCall Your Doctor Warning (Table 17-18)
INSR battery lowINSR Battery Low Message (Table 17-8)
INSR battery depletedRecharge INSR Warning (Table 17-9)
INSR rechargeNo change
complete
|
8-IR Communication Screen
Screen Entry: User Control/System Event
Reset button from any screen when DTC is not docked and IR command is received
|
Processing Requirements
All keys inactive
Software will display the screen until the host terminates the session, the Reset button is
activated or the Audio key is pressed. (Table 17-28)
When the host terminates the session or the Reset button is activated, software will display
Start Screen - Not Docked.
|
9-INS Battery Low Message
Screen Entry: User Control/System Event
Therapy On/Off
Entry ScreenUser Control/System Event: Next Screen
|
Power DownAudio key or system timeout: Start Screen - Not Docked with
Start Screen - NotTherapy On/Off icon on status line (Table 17-22)
DockedTherapy On/Off and telemetry success: Start Screen - Not
Docked with Therapy On/Off icon on status line (Table 17-
22)
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen
(Table 17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge: no change
Attach DTC: no change
Start Screen - DockedAudio key or system timeout: Start Screen - Docked with
Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off and telemetry success: Start Screen - Docked
with Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen
(Table 17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge: no change
Detach DTC: no change
INSR Charge StatusAudio key or system timeout: INSR Charge Status Screen
with Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off and telemetry success: INSR Charge Status
with Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen/INS
Charge Status (Table 17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge: no change
Detach DTC: no change
|
INS Battery Low Message - continued
Screen Entry: User Control/System Event
Stop Charge
Entry ScreenUser Control/System Event: Next Screen
|
INS Charge StatusAudio key or system timeout: Start Screen - Not Docked
(not docked)Therapy On/Off and telemetry success: Start Screen - Not
Docked with Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge: no change
Attach DTC: no change
INS Charge StatusAudio key and INSR recharge incomplete: INSR Charge Status
(docked)with Therapy On/Off icon on status line (Table 17-22)
Audio key and INSR recharge complete: INSR Recharge
Complete Message (Table 17-11)
Therapy On/Off and telemetry success and INSR recharge
incomplete: INSR Charge Status with Therapy On/Off icon on
status line (Table 17-22)
Therapy On/Off and telemetry success and INSR recharge
complete: Start Screen - Docked with Therapy On/Off icon on
status line (Table 17-22)
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge: no change
Detach DTC: no change
|
10-INS Recharge Interrupted Message
Screen Entry: User Control/System Event
Stop Charge
Entry ScreenUser Control/System Event: Next Screen
|
INS Charge StatusAudio key or system timeout: Start Screen - Not Docked
(not docked)Therapy On/Off and telemetry success: Start Screen - Not
Docked with Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge: no change
Attach DTC: no change
INS Charge StatusAudio key or system timeout and INSR recharge incomplete:
(docked)INSR Charge Status with Therapy On/Off icon on status line
(Table 17-22)
Audio key or system timeout and INSR recharge complete:
INSR Recharge Complete Message (Table 17-11)
Therapy On/Off and telemetry success and INSR recharge
incomplete: INSR Charge Status with Therapy On/Off icon on
status line (Table 17-22)
Therapy On/Off and telemetry success and INSR recharge
complete: Start Screen - Docked with Therapy On/Off icon on
status line (Table 17-22)
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge: no change
Detach DTC: no change
|
11-INS Recharge Complete Message
Screen Entry: User Control/System Event
Stop Charge
INS recharge complete
Entry ScreenUser Control/System Event: Next Screen
|
INS Charge StatusAudio key or system timeout: Start Screen - Not Docked
(not docked)Therapy On/Off and telemetry success: Start Screen - Not
INS Charge Status -Docked with Therapy On/Off icon on status line (Table 17-22)
90 to 100% (notTherapy On/Off and telemetry event: Refer to Table 18
docked)Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge: no change
Attach DTC: no change
INS Charge StatusAudio key or system timeout and INSR recharge incomplete:
(docked)INSR Charge Status
INS Charge Status -Audio key or system timeout and INSR recharge complete:
90 to 100% (docked)INSR Recharge Complete Message (Table 17-11)
Therapy On/Off and telemetry success and INSR recharge
incomplete: INSR Charge Status with Therapy On/Off icon on
status line (Table 17-22)
Therapy On/Off and telemetry success and INSR recharge
complete: Start Screen - Docked with Therapy On/Off icon on
status line (Table 17-22)
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge: no change
Detach DTC: no change
|
12-INSR Battery Low Message
Screen Entry: User Control/System Event
Therapy On/Off
Entry ScreenUser Control/System Event: Next Screen
|
Power DownAudio key or system timeout: Start Screen - Not Docked with
Start Screen - NotTherapy On/Off icon on status line (Table 17-22)
DockedTherapy On/Off and telemetry success: Start Screen - Not
Docked with Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge: no change
Attach DTC: INSR Charge Status with Therapy On/Off icon on
status line (Table 17-23)
INS Charge StatusAudio key or system timeout: previous screen with Therapy
INS Charge Status -On/Off icon on status line (Table 17-22)
90 to 100%Therapy On/Off and telemetry success: Wait Screen
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: INS Charge Status - Antenna
Too Hot Screen (Table 17-13)
Stop Charge: Start Screen - Not Docked
Attach DTC: previous screen with Therapy On/Off icon and
Power Source Symbol on status line (Table 17-22, 23)
|
Screen Entry: User Control/System Event
Start Charge
Entry ScreenUser Control/System Event: Next Screen
|
Power DownAudio key or system timeout: Wait Screen/INS Charge Status/
Start Screen - NotINS Charge Status - 90 to 100%
DockedTherapy On/Off and telemetry success: Wait Screen/INS
Charge/INS Charge Status - 90 to 100% Status with Therapy
On/Off icon on status line (Table 17-22)
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge: no change
Attach DTC: Wait Screen/INS Charge Status/INS Charge
Status - 90 to 100%
|
INSR Battery Low Message - continued
Screen Entry: User Control/System Event
Audio
Entry ScreenUser Control/System Event: Next Screen
|
Power DownAudio key or system timeout: Start Screen - Not Docked
Therapy On/Off and telemetry success: Start Screen - Not
Docked with Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off and telemetry event: Refer to Table 18
Attach DTC: INSR Charge Status
|
Screen Entry: User Control/System Event
INSR battery low condition
Entry ScreenUser Control/System Event: Next Screen
|
Wait ScreenAudio key or system timeout or Attach DTC: Wait Screen/INS
Charge Status/INS Charge Status - 90 to 100%
Physician RechargeAudio key or system timeout or Attach DTC: Physician Recharge
Physician Recharge -Audio key or system timeout or Attach DTC: Physician Recharge -
Antenna Too HotAntenna Too Hot
INS Charge StatusAudio key or system timeout or Attach DTC: INS Charge Status
INS Charge Status 90 toAudio key or system timeout or Attach DTC: INS Charge Status - 90
100%to 100%
INS Charge Status -Audio key or system timeout or Attach DTC: INS Charge Status -
Antenna Too HotAntenna Too Hot
Start Screen - NotAudio key or system timeout: Start Screen - Not Docked
DockedAttach DTC: INSR Charge Status
|
INSR Battery Low Message - continued
Screen Entry: User Control/System Event
Detach DTC
Entry ScreenUser Control/System Event: Next Screen
|
INS Charge StatusAudio key or system timeout: previous screen
INS Charge Status -Therapy On/Off and telemetry success: previous screen with
90 to 100%Therapy On/Off icon on status line (Table 17-22)
INS Charge Status -Therapy On/Off and telemetry event: Refer to Table 18
Antenna Too HotStart Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: INS Charge Status - Antenna
Too Hot Screen (Table 17-13)
Stop Charge: Start Screen - Not Docked
Attach DTC: previous screen with power source symbol on status
line (Table 17-23)
INSR Charge StatusAudio key or system timeout: Start Screen - Not Docked
Therapy On/Off and telemetry success: Start Screen - Not
Docked with Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge: no change
Attach DTC: INSR Charge Status
Physician RechargeAudio key or system timeout: previous screen
ScreenTherapy On/Off: no change
Physician Recharge -Start Charge: no change
Antenna Too HotStop Charge: Start Screen - Not Docked
Attach DTC: previous screen
|
13-INSR Recharge Interrupted Message
Screen Entry: User Control/System Event
Detach DTC
Entry ScreenUser Control/System Event: Next Screen
|
INSR ChargeAudio key or system timeout: Start Screen - Not Docked
StatusTherapy On/Off and telemetry success: Start Screen - Not Docked
with Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table 17-
1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge: no change
Attach DTC: INSR Charge Status
|
14-INSR Recharge Complete Message
Screen Entry: User Control/System Event
INSR recharge complete
Entry ScreenUser Control/System Event: Next Screen
|
INSR Charge StatusAudio key: Start Screen - Docked
Therapy On/Off and telemetry success: Start Screen - Docked
with Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge: no change
Detach DTC: Start Screen - Not Docked
|
Screen Entry: User Control/System Event
Attach DTC
Entry ScreenUser Control/System Event: Next Screen
|
Power DownAudio key: Start Screen - Docked
Start Screen - NotTherapy On/Off and telemetry success: Start Screen - Docked
Dockedwith Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge: no change
Detach DTC: Start Screen - Not Docked
|
15-INSR System Information Message
Screen Entry: User Control/System Event
Audio + Stop Charge
Entry ScreenUser Control/System Event: Next Screen
|
Power DownAudio key: Cycle through information
Start Screen - NotStop Charge or System timeout and not docked: Start Screen -
DockedNot Docked
Start Screen -Stop Charge or System timeout and docked: Start Screen -
DockedDocked
Therapy On/Off and telemetry success: Start Screen - Not
Docked with Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Attach/Detach DTC: no change
|
16-Call Your Doctor - ERI Message
Screen Entry: User Control/System Event
Therapy On/Off
Entry ScreenUser Control/System Event: Next Screen
|
Power DownAudio key or system timeout: Start Screen - Not Docked with
Start Screen - NotTherapy On/Off icon on status line (Table 17-22)
DockedTherapy On/Off and telemetry success: Start Screen - Not
Docked with Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge/Attach DTC: no change
INS Charge StatusAudio key or system timeout: previous screen with Therapy
INS Charge Status -On/Off icon on status line (Table 17-22)
90 to 100%Therapy On/Off and telemetry success: previous screen with
INS Charge Status -Therapy On/Off icon on status line (Table 17-22)
Antenna Too HotTherapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: INS Charge Status - Antenna
Too Hot Screen (Table 17-13)
Stop Charge and not docked: Start Screen - Not Docked
Stop Charge and docked: Start Screen - Docked
Attach DTC/Detach DTC: no change
INSR Charge StatusAudio key or system timeout: INSR Charge Status Screen with
Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off and telemetry success: INSR Charge Status
with Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge/Detach DTC: no change
Start Screen - DockedAudio key or system timeout: Start Screen - Docked with
Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off and telemetry success: Start Screen - Docked
with Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge/Detach DTC: no change
|
Call Your Doctor - ERI continued
Screen Entry: User Control/System Event
INS error: ERI
Entry ScreenUser Control/System Event: Next Screen
|
Wait ScreenAudio key or system timeout: Wait Screen
Therapy On/Off and telemetry success: Wait Screen
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: INS Charge Status - Antenna
Too Hot Screen (Table 17-13)
Stop Charge/Attach DTC/Detach DTC: no change
INS Charge StatusAudio key or system timeout: previous screen
INS Charge Status -Therapy On/Off and telemetry success: previous screen with
90 to 100%Therapy On/Off icon on status line (Table 17-22)
INS Charge Status -Therapy On/Off and telemetry event: Refer to Table 18
Antenna Too HotStart Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: INS Charge Status - Antenna
Too Hot Screen (Table 17-13)
Stop Charge and not docked: Start Screen - Not Docked
Stop Charge and docked: Start Screen - Docked
Attach DTC/Detach DTC: no change
|
17-Call Your Doctor - POR Message
Screen Entry: User Control/System Event
Therapy On/Off
Entry ScreenUser Control/System Event: Next Screen
|
Power DownAudio key or system timeout: Start Screen - Not Docked
Start Screen - Notwith Therapy On/Off icon on status line (Table 17-22)
DockedTherapy On/Off: no change
Start Charge/Stop Charge/Attach DTC/Detach DTC: no
change
INS Charge StatusAudio key or system timeout: previous screen with Therapy
INS Charge Status -On/Off icon on status line (Table 17-22)
90 to 100%Therapy On/Off: no change
INS Charge Status -Stop Charge and not docked: Start Screen - Not Docked
Antenna Too HotStop Charge and docked: Start Screen - Docked
Start Charge/Attach DTC/Detach DTC: no change
INSR Charge StatusAudio key or system timeout: INSR Charge Status Screen
with Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off: no change
Start Charge/Stop Charge/Attach DTC/Detach DTC: no
change
Start Screen - DockedAudio key or system timeout: Start Screen - Docked with
Therapy On/Off icon on status line (Table 17-22)
Therapy On/Off: no change
Start Charge/Stop Charge/Attach DTC/Detach DTC: no
change
|
Screen Entry: User Control/System Event
INS error: POR
Entry ScreenUser Control/System Event: Next Screen
|
Wait ScreenAudio key or system timeout: Wait Screen
Therapy On/Off and telemetry success: Wait Screen
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge/Attach DTC/Detach DTC: no change
INS Charge StatusAudio key or system timeout: previous screen
INS Charge Status -Therapy On/Off and telemetry success: previous screen with
90 to 100% INSTherapy On/Off icon on status line (Table 17-22)
Charge Status -Therapy On/Off and telemetry event: Refer to Table 18
Antenna Too HotStop Charge and not docked: Start Screen - Not Docked
Stop Charge and docked: Start Screen - Docked
Start Charge/Attach DTC/Detach DTC: no change
|
18-Antenna Too Hot Message
Screen Entry: User Control/System Event
Start Charge
Entry ScreenUser Control/System Event: Next Screen
|
Power DownAudio key or system timeout and not docked and antenna too
Start Screen -hot: Start Screen - Not Docked
DockedAudio key or system timeout and docked and antenna too hot:
Start Screen - NotStart Screen - Docked
DockedAudio key or system timeout and docked/not docked and
INSR Chargeantenna cools: INS Charge Status/INS Charge Status - 90 to
Status100%
INS Battery LowTherapy On/Off and telemetry success and not docked: Start
INS RechargeScreen - Not Docked with Therapy On/Off icon on status line
Interrupted(Table 17-22)
INS RechargeTherapy On/Off and telemetry success and docked: Start
CompleteScreen - Docked with Therapy On/Off icon on status line
INSR Battery Low(Table 17-22)
INSR RechargeTherapy On/Off and telemetry event: Refer to Table 18
InterruptedStart Charge: no change
INSR RechargeStop Charge and charge complete: INS Recharge Complete
CompleteMessage (Table 17-6)
Call Your Doctor -Stop Charge and charge incomplete and not INS battery low:
ERIINS Recharge Interrupted Message (Table 17-5)
Call Your Doctor -Stop Charge and INS battery low: INS Battery Low Message
POR Reposition(Table 17-3)
Antenna - PressStop Charge and INS battery depleted: Recharge Stimulator
Start ChargeWarning (Table 17-4)
WarningAttach/detach DTC: no change
Connect INSR
Antenna
|
19-Warning Messages
WarningUser Control/System Event: Next Screen
|
Recharge StimulatorStart Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
System timeout: Power Down
Recharge INSRAttach DTC: INSR Charge Status Screen with Power Source
Symbol
System timeout: Power Down
Reposition Antenna -Start Charge and antenna temperature OK: Wait Screen (Table
Press Start Charge17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
System timeout: Power Down
Reposition Antenna -Therapy Off and telemetry success: Start Screen - Docked or
Press Therapy OffStart Screen - Not Docked with Therapy Off icon on status
line (Table 17-22)
Therapy Off and telemetry event: refer to Table 18
System timeout: Power Down
Reposition Antenna -Therapy On and telemetry success: Start Screen - Docked or
Press Therapy OnStart Screen - Not Docked with Therapy On icon on status
line (Table 17-22)
Therapy On and telemetry event: refer to Table 18
System timeout: Power Down
Call Your DoctorSystem timeout: Power Down
Call Your Doctor -System timeout: Power Down
EOS
|
20-Splash Screen
Screen Entry: User Control/System Event
Application is loading
Reset button is activated and application is reloading
|
Processing Requirements
All keys active
Software will display the screen until the user presses any key or the system times out.
When the user presses any key or the system times out, software will display the Start
Screen - Docked or Start Screen - Not Docked (Table 17-29)
|
21-Connect INSR Antenna Message
Screen Entry: User Control/System Event
Antenna disconnected
User Control/System Event: Next Screen (assuming antenna
Entry Screenand recharger are making full contact)
|
Wait ScreenAudio key or system timeout and not docked: Start Screen -
INS Charge StatusNot Docked
ScreenAudio key or system timeout and docked: Start Screen -
INS Charge Status -Docked
90 to 100%Therapy On/Off and telemetry success and not docked: Start
Screen - Not Docked with Therapy On/Off icon on status line
(Table 17-22)
Therapy On/Off and telemetry success and docked: Start
Screen - Docked with Therapy On/Off icon on status line
(Table 17-22)
Therapy On/Off and telemetry event: Refer to Table 18
Start Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: INS Charge Status -
Antenna Too Hot Screen (Table 17-13)
Stop Charge: no change
Attach/detach DTC: no change
Physician RechargeAudio key or system timeout and not docked: Start Screen -
ScreenNot Docked
Physician RechargeAudio key or system timeout and docked: Start Screen -
Screen - AntennaDocked
Too HotTherapy On/Off: no change
Start Charge: no change
Stop Charge: no change
Attach/detach DTC: no change
|
Screen Entry: User Control/System Event
Therapy On/Off
Start Charge
User Control/System Event: Next Screen (assuming antenna
Entry Screenand recharger are making full contact)
|
All screens andAudio key or system timeout and not docked: Start Screen -
messages with theNot Docked
exception ofAudio key or system timeout and docked: Start Screen -
Physician RechargeDocked
Screen, PhysicianTherapy On/Off and telemetry success and not docked: Start
Recharge -Screen - Not Docked with Therapy On/Off icon on status line
Antenna Too Hot(Table 17-22)
Screen, IRTherapy On/Off and telemetry success and docked: Start
CommunicationScreen - Docked with Therapy On/Off icon on status line
Screen, Splash(Table 17-22)
Screen and allTherapy On/Off and telemetry event: Refer to Table 18
reposition antennaStart Charge and antenna temperature OK: Wait Screen (Table
warnings17-1)
Start Charge and antenna too hot: Antenna Too Hot Message
(Table 17-30)
Stop Charge: no change
Attach/detach DTC: no change
|
22-INS Charge Status - 90 to 100% Screen
Screen Entry: User Control/System Event
System timeout or message clear
Antenna temperature OK
Battery voltage indicates 90%
User Control/
System EventSystem Response: Feedback/Next Screen (requirement)
|
Therapy On/OffTelemetry success: Therapy On/Off icon on status line (Table
17-22)
Telemetry event: refer to Table 18
Start ChargeStart Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: INS Charge Status -
Antenna Too Hot Screen (Table 17-13)
Stop ChargeINS Recharge Complete Message (Table 17-6)
AudioToggle Audio icon (Table 17-24)
Attach DTCPower source symbol on status line (Table 17-23)
Detach DTCINSR battery low: INSR Battery Low Message (Table 17-8)
INSR battery depleted: Recharge INSR Warning (Table 17-9)
Status line without power source symbol (Table 17-23)
INS recharge completeINS Recharge Complete Message (Table 17-6)
INS error: ERICall Your Doctor ERI Message (Table 17-19)
User Control/
System EventSystem Response: Feedback/Next Screen (requirement)
INS error: PORCall Your Doctor - POR Message (Table 17-27)
INS error: EOS, batteryCall Your Doctor Warning (Table 17-18)
voltage too high
Antenna too hotINS Charge Status - Antenna Too Hot (Table 17-13)
Antenna failureCall Your Doctor Warning (Table 17-18)
Antenna disconnectedConnect INSR Antenna Message (Table 17-14)
INSR battery lowINSR Battery Low Message (Table 17-8)
INSR battery depletedRecharge INSR Warning (Table 17-9)
INSR rechargeINSR Batt1 on status line (Table 17-25)
complete
System errorCall Your Doctor Warning (Table 17-18)
Audio + Stop ChargeNo change
Audio + Start ChargeNo change
and hold for >10
seconds
|
23-INS Charge Status - Antenna Too Hot Screen
Screen Entry: User Control/System Event
Antenna too hot
System timeout or message clear
User Control/
System EventSystem Response: Feedback/Next Screen (requirement)
|
Therapy On/OffTelemetry success: Therapy On/Off icon on status line (Table
17-22)
Telemetry event: refer to Table 18
Start ChargeStart Charge and antenna temperature OK: Wait Screen (Table
17-1)
Start Charge and antenna too hot: INS Charge Status -
Antenna Too Hot Screen (Table 17-13)
Stop ChargeNot docked: Start Screen - Not Docked
Docked and INSR recharge complete: INSR Recharge
Complete Message (Table 17-11)
Docked and INSR recharge incomplete: INSR Charge Status
Screen (Table 17-7)
AudioToggle Audio icon (Table 17-24)
Attach DTCPower source symbol on status line (Table 17-23)
Detach DTCINSR battery low: INSR Battery Low Message (Table 17-8)
INSR battery depleted: Recharge INSR Warning (Table 17-9)
Status line without power source symbol (Table 17-23)
INS error: ERICall Your Doctor ERI Message (Table 17-19)
INS error: PORCall Your Doctor - POR Message (Table 17-27)
INS error: EOS, batteryCall Your Doctor Warning (Table 17-18)
voltage too high
Antenna too hotINS Charge Status - Antenna Too Hot (Table 17-13)
Antenna cooledWait Screen/INS Charge Status (Table 17-1)
Antenna failureCall Your Doctor Warning (Table 17-18)
Antenna disconnectedConnect INSR Antenna Message (Table 17-14)
INSR battery lowINSR Battery Low Message (Table 17-8)
INSR battery depletedRecharge INSR Warning (Table 17-9)
INSR rechargeINSR Batt1 on status line (Table 17-25)
complete
System errorCall Your Doctor Warning (Table 17-18)
Audio + Stop ChargeNo change
Audio + Start ChargeNo change
and hold for >10
seconds
|
24-Physician Recharge - Antenna Too Hot Screen
Screen Entry: User Control/System Event
Antenna too hot
User Control/
System Event*System Response: Feedback/Next Screen (requirement)
|
SuccessfulINS Charge Status Screen (Table 17-2)
communication with
INS
Timer goes to zero andStart Screen - Not Docked without Therapy On/Off icon on
system times out (notstatus line (Table 17-22)
docked)
Timer goes to zero andINSR recharge incomplete: INSR Charge Status Screen
system times out(Table 17-7)
(docked)INSR recharge complete: INSR Recharge Complete Message
(Table 17-11)
Stop Charge (notStart Screen - Not Docked
docked)
Stop Charge (docked)Start Screen - Docked
Detach DTCINSR battery low: INSR Battery Low Message (Table 17-8)
INSR battery depleted: Recharge INSR Warning (Table 17-9)
Antenna too hotPhysician Mode - Antenna Too Hot (Table 17-31)
Antenna failureCall Your Doctor Warning (Table 17-18)
Antenna disconnectedConnect INSR Antenna Message (Table 17-14)
System errorCall Your Doctor Warning (Table 17-18)
INSR battery lowINSR Battery Low Message (Table 17-8)
INSR battery depletedRecharge INSR Warning (Table 17-9)
INSR rechargeNo change
complete
|
*Note:
|
all keys inactive except Stop Charge
|
TABLE 18
|
|
|
1
When the user presses start charge or the antenna has cooled,
|
software will display the Wait Screen until successful
|
communication with the INS is established..
|
2
Software will display the INS Charge Status Screen as
|
soon as successful communication with the INS is established.
|
3
Display INS Battery Low Message
|
4
Display Recharge Stimulator Warning
|
5
Display INS Recharge Interrupted Message
|
6
Display INS Recharge Complete Message
|
7
Display INSR Charge Status Screen
|
8
Display INSR Battery Low Message
|
9
Display Recharge INSR Warning
|
10
Display INSR Recharge Interrupted Message
|
11
Display INSR Recharge Complete Message
|
12
Display Reposition Antenna Press Start Charge Warning
|
13
Display INS Charge Status - Antenna Too Hot Screen
|
14
Display Connect INSR Antenna Message
|
15
Display Reposition Antenna Press Therapy On Warning
|
16
Display Reposition Antenna Press Therapy Off Warning
|
17
Display INSR System Infomiation Message
|
18
Display Call Your Doctor Warning
|
19
Display Call Your Doctor - ERI Message
|
20
Entry intentionally left blank to preserve numbering
|
21
Display Physician Recharge Screen
|
22
Software will display an indicator for stimulation status
|
(On/Off) on the status line.
|
23
Software will display the power source symbol when the
|
desktop charger is docked.
|
24
Software will display an indicator for audio control
|
on the status line.
|
25
Software will display the INSR charge level on the status line.
|
26
Software will update the recharge efficiency line each time
|
user presses Start Charge
|
27
Display Call Your Doctor - POR Message
|
28
Display IR Communication Screen
|
29
Display Splash Screen
|
30
Display Antenna Too Hot Message
|
31
Display Physician Recharge - Antenna Too Hot Screen
|
|
TABLE 19
|
|
|
Therapy On/Off Events
|
System Event
Next Screen (requirement)
|
|
1.
Telemetry failure
Reposition Antenna - Press Therapy On
|
for Therapy On
Warning (Table 17-15)
|
2
Telemetry failure
Reposition Antenna - Press Therapy Off
|
for Therapy Off
Warning (Table 17-16)
|
3.
INS battery low
INS Battery Low Message (Table 17-3)
|
4.
INS battery depleted
Recharge Stimulator Warning (Table 17-4)
|
or in sleep mode
|
5.
INS error - ERI
Call Your Doctor - ERI Message
|
(Table 17-19)
|
6.
INS error - POR
Call Your Doctor - POR Message
|
(Table 17-27)
|
7
INS error - EOS or
Call Your Doctor Warning (Table 17-18)
|
battery voltage too high
|
8.
INSR battery low
INSR Battery Low Message (Table 17-8)
|
9.
INSR battery depleted
Recharge INSR Warning (Table 17-9)
|
10.
Antenna failure
Call Your Doctor Warning (Table 17-18)
|
11.
Antenna disconnected
Connect INSR Antenna Message
|
(Table 17-14)
|
12.
System error
Call Your Doctor Warning (Table 17-18)
|
13.
Antenna too hot
Display INS Charge Status - Antenna Too
|
Hot Screen (Table 17-13) or Antenna Too
|
Hot Message (Table 17-30)
|
|
User familiarity with both external programming unit 2 and charging unit 50 is enhanced through the use of common patient controls and/or display icons across both devices. The user is presented with common controls and displays across both platforms.
The primary purpose of external programming unit 2 is to manipulate therapeutic programs such as starting therapy and stopping therapy. The primary purpose of charging unit 50 is to charge or recharge rechargeable power source 24. However, patients may want to control their therapy while charging unit 50 is charging the rechargeable power source 24 of implantable medical device 16. Thus, charging unit 50 is also equipped with controls to start and stop therapy as well as controls to start and stop charging.
External programming unit 2 and charging unit 50 both have similar “Therapy On” controls as illustrated in Table 1 and Table 11, respectively. Similarly, both external programming unit 2 and charging unit 50 also have similar “Therapy Off” controls as illustrated in Table 1 and Table 11, respectively.
To aid in familiarity by the user, both external programming unit 2 and charging unit 50 also have similar or identical icons to display the same or similar messages to the user. Battery charging level icons “INSR Batt 1”, “INSR Batt 2”, “INSR Batt 3”, “INSR Batt 4” and “INSR Batt 5” icons are similar or identical between external programming unit 2 and charging unit 50 as illustrated in Table 1 and Table 11, respectively.
Similarly, the “Call Doctor” icon illustrated in Table 5 with respect to external programming unit 2 is the same as, or very nearly the same as, the “Call Doctor” icon illustrated in Table 15 with respect to charging unit 50.
The “Warnings” icon illustrated in Table 5 with respect to external programming unit 2 is the same as, or very nearly the same as, the “Warnings” icon illustrated in Table 16 with respect to charging unit 50.
“Audio On” and “Audio Off” icons illustrated in Table 6 with respect to external programming unit 2 are the same as, or very nearly the same as, the “Audio On” and “Audio Off” icons illustrated in Table 11 with respect to charging unit 50.
These are examples of controls and icons that are the same, or very similar, between both platforms, the external programming unit 2 and the charging unit 50 to aid in user familiarity.
As noted with respect to FIG. 17 describing the process of charging rechargeable power source 24 using external charging device 48, charging unit 50 receives information via telemetry from implantable medical device 16 during the charging process. External antenna 52 contains both primary coil 54 and external telemetry coil 46 and implantable medical device 16 contains both secondary coil 34 and internal telemetry coil 44. When external antenna 52 is positioned properly at or near the site of implantable medical device 16, primary coil 54 can be inductively coupled to secondary coil 34 to facilitate the charging process. Similarly, external telemetry coil 46 can be coupled with internal telemetry coil 44 to communicate information across cutaneous boundary 38 during the charging process.
Also as noted above, magnetic shield 36 prevent electromagnetic radiation stimulating secondary coil 34 from reaching the interior of housing 32 of implantable medical device. However, external telemetry coil 46 and internal telemetry coil 44 are sized to be larger than magnetic shield 36 allowing electromagnetic communication between external telemetry coil 46 and internal telemetry coil 44.
At step 112 of the charging process illustrated in FIG. 17, a check is made to determine if telemetry between charging unit 50 and implantable medical device 16. Further, the status of charging rechargeable power source 24 is communicated from implantable medical device 16 and displayed by charging unit 50 (step 118). The charging status displayed by charging unit 50 can be the actual known charging status of rechargeable power source 24 as communicated from implantable medical device 16 to charging unit 50. This prevents essentially guessing the internal charging status of rechargeable power source 24 by some external means such as simply by the passage of time. Rather, the actual charging status of rechargeable power source 24, e.g., the battery voltage, can be measured and transmitted to charging unit 50 for display to the user.
Typically, inductive charging and telemetry operations are time division multiplexed during the charging operation. That is, a portion of the time during which charging occurs is dedicated to telemetry. During this dedicated telemetry time, the inductive charging by primary coil 54 to secondary coil 34 is momentarily stopped and telemetry allowed to occur. As soon as telemetry has occurred and the required information has been communicated, inductive coupling can be resumed. Typically, only a relatively small portion of the time period during charging is required to be devoted to telemetry. Inductive charging can be stopped only for a short, e.g., a few seconds, to allow for telemetry and then inductive charging can be resumed. Although the time division multiplexing occurs over the entire charging period, only a small portion of the charging time is lost to telemetry. Time division multiplexing of inductive coupling charging and telemetry prevents the possibility of inductive charging from interfering with telemetry signals.
Thus, embodiments of the user interface for external charger for implantable medical device are disclosed. One skilled in the art will appreciate that the present invention can be practiced with embodiments other than those disclosed. The disclosed embodiments are presented for purposes of illustration and not limitation, and the present invention is limited only by the claims that follow.