Referring first to
In addition, the vacuum control system 20 includes a controller 28 for continuously comparing the preselected vacuum level with the actual vacuum level. It still further includes a proportional valve, generally designated 30, for adjusting the actual vacuum level so that it will correspond to the preselected vacuum level in response to an electronic signal from the controller 28. Accordingly, the vacuum control system 20 will be understood to comprise a closed loop system which is capable of providing a precise vacuum during operation.
Still referring to
The motor 22b turns the rotary-to-linear crank mechanism 22e through the speed reducing gearbox 22d in one direction only. The output of the speed reducing gearbox 22d provides the required torque to drive the piston 22a within the vacuum cylinder 22c. The piston 22a is driven within the vacuum cylinder 22c to achieve the desired operating vacuum profile.
Referring to
As for the sensor 26, it preferably comprises a differential vacuum transducer for monitoring the actual vacuum level being produced. The differential vacuum transducer 26 monitors the actual vacuum level through an airway or tube 34 in communication with an airway or tube 36 extending from the vacuum cylinder 22c to a milk collection kit, i.e., it monitors the actual vacuum level between the cylinder 22c and the milk collection kit. In addition, the differential vacuum transducer 26 provides feedback to the controller 28 through an electrical line 38.
Referring to
If the comparison shows a difference more than a preprogrammed amount, the proportional valve 30 will cause the actual vacuum level to be adjusted in response to a signal from the controller 28. The proportional valve 30 comprises an air flow valve (see
As will be appreciated from
With the foregoing, it will be appreciated that the vacuum control system 20 is a closed loop system capable of maintaining a preselected vacuum level for a breast pump independent of varying ambient barometric conditions.
Referring to
As for other functionality, the breast pump 42 includes another set of up and down membrane switches 44a and 44b for setting the cycle speed. There is also a cycle speed LCD 44c above the membrane switches 44a and 44b. Further, the front panel 42a of the breast pump 42 also includes a session timer LCD 46 and a session timer reset membrane button 48.
Still additionally, the breast pump 42 may advantageously include a membrane button 50 for initiating start up of the pump and placing the pump in a standby mode (“off”) after it has been used.
When the membrane button 50 is pushed, the breast pump 42 will automatically start at a minimum preprogrammed vacuum level of 30 mmHg and a maximum preprogrammed pumping cycle speed of 80 cycles/min. Subsequently, the user can adjust the cycle speed and/or adjust the vacuum level using the membrane switches 44a, 44b and 32a/32b, respectively. Preferably, the breast pump 42 is programmed to have a vacuum range of 30 mmHg-250 mmHg developed by the vacuum source 22. The vacuum source 22 may suitably include a 24 V brushless DC (BLDC) motor 22b and a 20:1 speed reducing gearbox 22d. Referring to
Still referring to
More specifically, the piston shaft link 22g is coupled to the rotary-to-linear crank mechanism 22e off-center to the rotation of the motor shaft. This serves to impart linear reciprocating motion to the piston shaft 22f and the piston 22a. Further, the vacuum cylinder 22c is coupled to the base of the breast pump 42 through a mounting plate 52 that supports a shaft 54 in generally spaced relation as shown in
As will be appreciated, the pivoting movement of the vacuum cylinder permitted by the collar 56 mounted on the shaft 54 maintains linear alignment with the piston shaft 22f and the piston 22a for each revolution of the rotary-to-linear crank mechanism 22e. Under hardware/software control, this rotary-to-linear actuator 22e moves the piston 22a with a fixed stroke length and adjustable cycle speed toward the motor/gearbox assembly 58 comprised of the motor 22b and the speed reducing gearbox 22d to increase the vacuum and away from the motor/gearbox assembly 58 to reduce the vacuum. In this connection, the vacuum produced by movement of the piston 22a within the vacuum cylinder 22c will be available to the milk collection kit through a suitable airway or tubing 36 leading from the vacuum cylinder 22c (see
In the present disclosure, the sensor 26 and proportional valve 30 are disposed within a closed-loop system that maintains consistent pump performance in both single and double pumping modes under varying ambient barometric conditions. The ability to maintain closed-loop control of the vacuum amplitude is implemented as a result of the proportional valve 30 and the differential vacuum transducer 26 and which are in electrical communication with the controller 28. As previously noted, the proportional valve 30 comprises a stepper motor 30a, a valve housing 30e with a defined orifice pathway in the form of the port 30d which cooperates with the cone-shaped pin 30c, and an opening to atmosphere 30f.
With this arrangement, the cone-shaped pin 30c attached to the stepper motor 30a by means of the shaft 30b is used to precisely regulate the amount of ambient air flowing into the vacuum system. Also, the electronic differential vacuum transducer 26 is used to monitor gauge vacuum amplitude relative to ambient barometric pressure and provide feedback to the controller 28. The controller 28 compares the gauge vacuum reading to the desired vacuum and adjusts the proportional valve 30 so that the desired peak vacuum level is achieved.
The vacuum system initial air volume at ambient pressure or zero vacuum is defined by the position of the piston 22a nearest to the output of the vacuum cylinder 22c, the status of the proportional valve 30, the airway or tube 34 from the proportional valve 30 to the vacuum cylinder 22c, the airway or tube 34a from the output of the vacuum cylinder 22c to the vacuum transducer 26 and the airway or tube 36 from the vacuum cylinder output to the pump vacuum port.
In single pumping mode, one milk collection kit is connected to the breast pump 42 and the initial air volume is less than that for double pumping utilizing two milk collection kits. There is automatic compensation for variation in milk collection kit volume as a result of measurement of system vacuum and subsequent regulation of vacuum via of the proportional valve 30 in a closed-loop control manner by the controller 28. In a piston pump, change in vacuum is directly proportional to atmospheric pressure multiplied by the ratio of vacuum system initial volume to final volume so it follows there is at least the potential for varying output peak vacuum for a fixed piston stroke length as atmospheric pressure varies. In the present disclosure, the compensation is achieved over a barometric range of 0.82 atm/83 kPa to 1.05 atm/106 kPa.
Another advantage of the present disclosure and its system vacuum measurement scheme is the ability to alert the nursing mother if there is a leak in the milk collection kits connected to the breast pump 42. A “Check Kit” text message is visible in the vacuum LCD 32c if the measured system output vacuum cannot reach the desired vacuum setting and, thus, conditions such as a milk collection kit being disconnected from the vacuum port, an open milk collection kit tubing adapter, or a missing diaphragm in a milk collection kit will elicit display of this kit error message. As a comfort measure, the user cannot change the cycle speed or vacuum settings at any time while this kit error message is being displayed on the vacuum LCD 32c.
If the “Check Kit” message appears under normal pumping conditions for the breast pump 42, the user can remedy the problem within several seconds to maintain current vacuum settings. However, if the problem is not remedied within several seconds, the pump control system will return the vacuum output to its default setting of 30 mmHg as a comfort measure.
Referring to
The motor voltage drive software algorithm provides the gradual acceleration/deceleration phases necessary under dynamic loading of the motor/gearbox assembly 58 to achieve an “intermittent” vacuum waveform including a “rest” phase at ambient barometric pressure or zero gauge vacuum, vacuum rise time to peak vacuum amplitude, and vacuum fall time back to ambient barometric pressure for each vacuum cycle. The rest phase of the vacuum cycle at ambient barometric pressure is achieved with a 0.1 psi mechanical one-way check valve 62 mounted on the vacuum cylinder 22c that exhausts the vacuum cylinder to ambient air only after peak vacuum is reached and the piston 22a moves toward the vacuum cylinder output, i.e., toward the end of the vacuum cylinder 22c opposite the piston shaft 22f (compare
For this purpose, a 5.5 psi mechanical relief valve 64 may be mounted on the vacuum cylinder 22c at the vacuum output port in order to prevent vacuum cylinder output from exceeding 300 mmHg.
The vacuum front panel controls, i.e., membrane switches 32a and 32b, are independent of the cycle speed front panel controls, i.e., membrane switches 44a and 44b. Step changes over the peak vacuum range of 30-250 mmHg are made possible using the up and down membrane switches 32a and 32b wherein a single push-and-release results in a target 2 mmHg change and a push-and-hold results in one or more target 10 mmHg changes. The vacuum LCD 32c comprises a graphic liquid crystal display positioned above the vacuum level controls 32a and 32b providing the user with information in two different formats. The graphical liquid crystal display presents a bar graph representing 0-100% of the available 30-250 mmHg peak vacuum range changes in response to use of the vacuum level controls 32a and 32b whereas a numerical display of the percentage facilitates observing the vacuum setting. As for the numerical display, the setting can be used for future pumping sessions and/or communication to a lactation consultant.
The cycle speed front panel controls 44a and 44b are independent of the vacuum level controls 32a and 32b. Step changes in cycles/min over the range of 30-80 cycles/min are made possible using the up and down membrane switches 44a and 44b wherein a single push-and-release results in a target 1 cycle/min change and a push-and-hold results in one or more target 10 cycles/min changes. The cycle speed LCD 44c also comprises a graphic liquid crystal display positioned above the cycle speed controls 44a and 44b providing the user with information in two different formats. The graphical liquid crystal display presents a bar graph representing 0-100% of the available 30-80 cycles/min range changes in response to use of the cycle speed controls 44a and 44b whereas a numerical display of actual cycle speed facilitates observing the cycle speed setting. As for the numerical display, the setting can be used for future pumping sessions and/or communication to a lactation consultant.
As shown in
The nursing mother has the option of using the session timer reset membrane button 48 to reset the session timer to 00:00. The elapsed session timer LCD 46 keeps track of the total pumping time; however, an automatic shut-down mode of operation is enabled in the event that the elapsed session timer reaches 60 minutes of continuous run time. This automatic shut-down feature is intended to help guard against excessive pumping and/or unattended operation of the breast pump 42.
After the session timer reaches 60:00, the pump vacuum and cycle speed are reduced to 0 mmHg and 0 cycles/min, respectively, the pump returns to a standby mode and turns off.
The breast pump 42 can be electrically powered over a range of line power voltages (100-240 VAC, 50/60 Hz) and socket configurations encountered throughout the world. It will, of course, be understood that country-specific, detachable power cord sets for the breast pump 42 can be provided as needed. When the breast pump 42 is connected to an electrical outlet, an AC power plug icon is visible in the session timer LCD 46 to indicate the connection to the user.
If desired, the breast pump 42 can be battery-operated with a rechargeable Li-Ion battery pack as an optional accessory. With this optional accessory, the battery capacity and charging status are indicated with icons visible in the session timer LCD 46.
Referring to
As shown in
While in the foregoing there has been set forth a detailed description of the present disclosure, it will be appreciated by those skilled in the art that the details herein given may be varied without departing from the true spirit and scope of the appended claims.
Number | Date | Country | |
---|---|---|---|
60806886 | Jul 2006 | US |