Often, when a person loses most of their kidney function, dialysis is required to perform some of the functions of the kidneys. Namely, dialysis removes waste, salt, and excess water from the blood to prevent a toxic build-up in the body. Dialysis also helps to maintain a safe level of chemicals in the blood (e.g., potassium, sodium, and biocarbonate) as well as help control the person's blood pressure. Approximately half a million Americans are on dialysis alone.
The most popular form of dialysis is hemodialysis. In hemodialysis, blood is removed from a patient's blood vessel, ran through a dialysis machine which functions as an artificial kidney, and returned back to the patient's blood vessel. Often, in hemodialysis, an arteriovenous (AV) fistula is created to provide suitable blood pressure and flow in the patient's blood vessel. The AV fistula is a connection between an artery and a vessel, usually created in the arm with a surgical procedure. Surgically created AV fistulas are a preferred mode of vascular access for long-term vascular dialysis treatment and is favored by Medicare over other options. The AV fistula results in increased flow and pressure within a vessel which becomes the access point for dialysis needles. Indeed, the closer the access point in the vessel is to the AV fistula itself, the better the blood flow. After surgery, there is an average 60 day waiting period for the fistula to mature before it can be used.
In approximately 40% of cases, the AV fistula fails to become useable and subsequent surgeries are required. Approximately 33% of cases require transposition surgery, which is needed when the vessel is too deep within the arm to access. After a useable AV fistula is created, the patient and caregivers still endure the challenges of getting a “good stick” each time there is a dialysis treatment, performed at least three times per week. The access vessel must be accurately targeted, sometimes requiring sticks with two 15-gauge needles. Bad punctures are common and, in the most severe cases, may result in significant blood loss and collapse of the fistula (approximately 5.2% patients have significant infiltrations which are a result of a bad puncture and frequently lead to loss of the AV fistula). Furthermore, most AV fistulas only last about three years, largely failing because of excessive and bad punctures. The failed AV fistula site requires additional surgical or radiological intervention to be recovered. If recovery is unsuccessful, a new AV fistula needs to be created in a different location.
In some situations, the increased blood flow through the vessel due to the AV fistula can enlarge the vessel over time, further increasing blood flow through the vessel. Too much blood flow through the AV fistula and into the vessel can leave the outer extremities, which are served by the artery, without enough blood, causing pain and discomfort to the patient. Too much blood flow through the AV fistula and into the vessel can also send too much blood directly to the heart, causing heart issues up to and including heart failure.
Vascular access devices with arteriovenous fistula support are provided. Arteriovenous fistulas are often needed to supply enough blood flow and pressure to support hemodialysis of a patient's blood. The vascular access device provides a place to draw blood from a vessel that is close to the artery (and therefore has adequate blood flow and pressure to support hemodialysis of a patient's blood) and provides structural support to the vessel, artery, and arteriovenous fistula to prevent damage. Advantageously, the described vascular access device can protect an AV fistula and therefore prevent or at least minimize likelihood of invasive and redundant surgeries, as well as prevent too much blood flow through the AV fistula that causes heart issues and/or not enough blood to get to a patient's outer extremities.
A vascular access device with arteriovenous fistula support can include a top portion and a bottom portion that couple together. The top portion has a partial artery channel for receiving a top half of an artery, a partial vessel channel for receiving a top half of a vessel, and a partial arteriovenous joint between the partial artery channel and the partial vessel channel for receiving a top half of an arteriovenous fistula. The partial vessel channel of the top portion includes a vascular access aperture for exposing the vessel so that blood may be drawn from the vessel. The bottom portion has a partial artery channel for receiving a bottom half of the artery, a partial vessel channel for receiving a bottom half of the vessel, and a partial arteriovenous joint between the partial artery channel and the partial vessel channel for receiving a bottom half of the arteriovenous fistula. When the top portion is coupled to the bottom portion, the partial artery channel of the top portion and the partial artery channel of the bottom portion couple to one another to form an artery channel, the partial vessel channel of the top portion and the partial vessel channel of the bottom portion couple to one another to form a vessel channel, and the partial arteriovenous joint of the top portion and the partial arteriovenous joint of the bottom portion couple to form an arteriovenous joint for an arteriovenous fistula.
In some cases, the partial artery channel of the top portion and the partial artery channel of the bottom portion are semi-cylindrical, and when the top portion is coupled to the bottom portion, the semi-cylindrical artery channel of the top portion and the semi-cylindrical artery channel of the bottom portion couple to one another to form a cylindrical artery channel. In some cases, the partial vessel channel of the top portion and the partial vessel channel of the bottom portion are semi-cylindrical, and when the top portion is coupled to the bottom portion, the semi-cylindrical vessel channel of the top portion and the semi-cylindrical vessel channel of the bottom portion couple to one another to form a cylindrical vessel channel. In some cases, all of the partial artery channels and the partial vessel channels are semi-cylindrical such that both the artery channel and the vessel channel are cylindrical.
In some cases, the partial vessel channel of the bottom portion is elongated relative to the partial vessel channel of the top portion so that as a dialysis needle is inserted into the vessel at an angle relative to a surface of the skin of a patient, the needle does not rupture or damage the back wall of the vessel (i.e., does not go completely through the vessel and out the other side). In some cases, the vascular access aperture is elongated parallel to the vessel of the patient so that as a dialysis needle is inserted into the vessel at an angle relative to a surface of the skin of a patient, the needle does not rupture or damage the back wall of the vessel (i.e., does not go completely through the vessel and out the other side).
In some cases, the surfaces of the top portion and the bottom portion are porous to allow for collagen to form around the top portion and the bottom portion after implantation into a patient. In some cases, a guide lip surrounding the vascular access aperture of the partial vessel channel of the top portion is included to help guide the dialysis needle into the correct position needed for dialysis treatment.
In some implementations, the vessel channel of the vascular access device curves in to the arteriovenous joint while the artery channel of the vascular access device is straight. In some implementations, the artery channel of the vascular access device curves in to the arteriovenous joint while the vessel channel of the vascular access device is straight. In some implementations, both the artery channel and the vessel channel curve in to the arteriovenous joint. In some implementations, both the artery channel and the vessel channel are straight.
In some implementations, an angle between a vertical axis perpendicular to the vascular access aperture of the device and an axis running through center points of the vessel channel and the artery channel is greater than 45 degrees. In some implementations, an angle between a vertical axis perpendicular to the vascular access aperture of the device and an axis running through center points of the vessel channel and the artery channel is between 30 degrees and 45 degrees. In some implementations, an angle between a vertical axis perpendicular to the vascular access aperture of the device and an axis running through center points of the vessel channel and the artery channel is less than 30 degrees.
In some implementations in which an end-to-side AV fistula is created, the vessel channel is perpendicular to the artery channel and the arteriovenous joint is disposed in a position where the artery channel abuts the vessel channel.
A method of using the vascular access device can include creating an incision to insert the vascular access device, positioning the partial artery channel of the bottom portion of the vascular access device around the bottom half of an artery and the partial vessel channel of the bottom portion of the vascular access device around the bottom half of a vessel, positioning the partial artery channel of the top portion of the vascular access device around the top half of the artery and the partial vessel channel of the top portion of the vascular access device around the top half of a vessel to couple the top portion to the bottom portion, closing the incision used to insert the vascular access device and allowing the tissue surrounding the vascular access device to heal, and after the incision is healed, creating an AV fistula between the artery and the vessel without uncoupling the top portion of the vascular access device from the bottom portion of the vascular access device.
This Summary is provided to introduce a selection of concepts in a simplified form that are further described below in the Detailed Description. This Summary is not intended to identify key features or essential features of the claimed subject matter, nor is it intended to be used to limit the scope of the claimed subject matter.
Vascular access devices with arteriovenous fistula support are provided. Arteriovenous fistulas are often needed to supply enough blood flow and pressure to support hemodialysis of a patient's blood. The vascular access device provides a place to draw blood from the vessel that is close to the artery (and therefore has adequate blood flow and pressure to support hemodialysis of a patient's blood) and provides structural support to the vessel, artery, and arteriovenous fistula to prevent damage. Advantageously, the described vascular access device can protect an AV fistula and therefore prevent or at least minimize likelihood of invasive and redundant surgeries, as well as prevent too much blood flow through the AV fistula that causes heart issues and/or not enough blood to get to a patient's outer extremities.
As used herein, “successful cannulation” or a “good puncture” refers to when a needle/cannula is placed in a vessel to provide vascular access and does not damage any tissue any more than necessary to be placed in the vessel.
As used herein, “unsuccessful cannulation” or a “bad puncture” refers to when a needle/cannula damages more tissue than necessary to be placed in the vessel, whether the needle/cannula is actually placed in the vessel (or not) and/or enters the AV fistula and/or artery. Examples of a bad puncture include when a needle/cannula goes into the vessel and out the back wall of the vessel, or when the vessel is missed altogether, both of which can damage the vessel and/or the tissue surrounding the vessel and contribute to the collapse of the vessel and/or AV fistula, as well as cause blood loss into the surrounding tissue and formation of a hematoma. Another example of a bad puncture includes when a needle/cannula enters the artery, which can be dangerous if any air (e.g., an air bubble) from the needle/cannula enters the artery, which can lead to an air embolism.
As used herein, referring to a portion of the device as “curved” or having “curvature” means that portion of the device has a substantially uniform deviation from straight with respect to a plane perpendicular to the bottom surface of the device.
As used herein, “semi-cylindrical” refers to an object that has a shape of a cylindric section (of a cylindrical surface) where the intersecting plane of the cylinder forms a general shape of a horizontal cylindrical segment. However, the semi-cylindrical shape does not require the cylindric section of the cylindrical surface to appear as being formed from a single plane. For example, the perpendicular-to-the-axis cross-section of the semi-cylindrical shape can appear as a pie shape/sector.
As used herein, “joint” refers to a connection and/or absence of material between the vessel channel and the artery channel that may or may not include structure around the connection and/or absence of material.
As used herein, “side-to-side” refers to a side of a vessel and a side of an artery (or another vessel) that abut and/or come close to one another. As used herein, “end-to-side” refers to an end of a vessel abutting a side of an artery (or another vessel).
As used herein, “artery” may refer to an actual artery or an artery graft.
The vessel channel 110 provides structural support and external protection for a vessel 115 of a patient. By providing structural support and external protection for the vessel 115 of a patient, the vessel channel 110 prevents a bad puncture (e.g., through a back wall of the vessel 115), which also decreases the amount of trauma incurred to the vessel 115 of the patient and diminishes subsequent scarring and stenosis of the vessel 115. The vessel channel 110 also includes a vascular access aperture 112 that is elongated parallel to the vessel of the patient. The vascular access aperture 112 provides easy access for a dialysis needle to deliver a good puncture to the vessel of the patient and limits the risk of infection by utilizing the skin as a natural barrier to pathogens. Indeed, the vascular access aperture 112 provides a greater surface area for a good puncture than other devices or no devices at all. This enables good punctures in different locations along the surface of the patient's skin, which helps prevent skin breakdown and gives the patient's skin a chance to heal.
The artery channel 120 provides structural support and external protection for an artery 125 of a patient. By providing structural support and external protection for the artery 125 of the patient, the artery channel 120 prevents the artery from being punctured or damaged (e.g., through a bad puncture of the vessel).
An AV fistula 130 can be formed between the vessel 115 and artery 125 and encased in an AV joint 135. The vessel channel 110 and the artery channel 120 can be any shape suitable to house a vessel and artery and therefore, although the channels 110, 120 are illustrated as cylindrical, the cross-section perpendicular to the axis of the cylinder is not required to be a perfect circle or ellipse. Indeed, in other cases, the cross-sectional shape of the channels 110, 120 may be any shape and/or size to fully enclose portions of the vessel 115 and the artery 125 near the AV fistula 130.
In the illustrated example, the vessel channel 110 curves towards the artery channel 120. However, as illustrated in
In some cases, the surfaces of the device 100 are porous. The pores in the device 100 allow for collagen (i.e., scar tissue) to form around the top portion 102 and the bottom portion after implantation of the device 100 into a patient. The porosity of the device 100 allows ingrowth of fibrovascular tissue, which adheres to the device 100. The fibrovascular tissue allows the body to mount an immune response to any bacteria which are instilled during needle cannulation as well as prevents implantation of bacteria on the surface of the device 100. The porosity of the channels 110, 120 allow collagen to integrate into the vessel and arterial walls, supporting them biologically as well as mechanically. The collagen that forms around the vessel and into the porous channels 110, 120 acts as a scaffold that helps keep the vessel and artery open. The collagen also aids in prevention of collapse of the vessel from repeated sticks and weakening of the vessel wall as well as create a biological seal across the vascular access aperture 112. The porosity of the device 100 may include spaces/holes that can be one nanometer in size up to almost 1 millimeter in size. In some implementations, the pores can be a range of sizes or a specific size, the range or specific size being 100, 150, 200, 250, 300, 350, 400, 450, 500, 550, 600, 650, 700, 750, and/or 800 microns in size. The porosity of the surfaces of the device 100 provides radial support for the vessel of the patient and helps to hold it open during the needle's entry into the vessel. It should be noted that the pores are small enough to prevent the passage of the needle through the surface of the device 100. The surface of the device 100 (having the pores) may be formed by sintering metallic beads (e.g., titanium beads) or powders onto the surface, machining, sandblasting, laser etching, injection molding, and/or 3D printing. It should also be noted that, in some cases, all of the surfaces of the device 100 that are exposed to human tissue are porous.
In some cases, some or all of the device 100 is made of biocompatible plastic, metal, or ceramic, such as titanium, Poly Ether Ether Ketone (PEEK), aluminum oxide, stainless steel, polyvinylchloride and the like. In some cases, some or all of the device 100 is made of radiolucent and/or radiopaque materials. This can help a surgeon if any subsequent procedures are needed (e.g., creation of the AV fistula subsequent to the implantation of the device 100). For example, radiolucent and/or radiopaque markers may be fixed to the device 100 so that surgeons and/or interventionalists who are implanting/revising/accessing the device 100 can see the device under various imaging modalities. The radiolucent/radiopaque markers can also be utilized to identify appropriate sites for withdrawal and/or injection of fluid.
In some cases, the partial vessel channel 312 of the top portion 310 and the partial vessel channel 322 of the bottom portion 320 are semi-cylindrical. Therefore, when the top portion 310 is coupled to the bottom portion 320, the partial vessel channel 312 of the top portion 310 and the partial vessel channel 322 of the bottom portion 320 couple to one another to form a cylindrical vessel channel. In some cases, the partial artery channel 314 of the top portion 310 and the partial artery channel 324 of the bottom portion 320 are semi-cylindrical. Therefore, when the top portion 310 is coupled to the bottom portion 320, the partial artery channel 314 of the top portion 310 and partial artery channel 324 of the bottom portion 320 couple to one another to form a cylindrical artery channel. In some cases, all of the partial vessel channels 312, 322 and partial artery channels 314, 324 are semi-cylindrical such that both the artery channel and the vessel channel are cylindrical.
As illustrated in
Furthermore, because the punctures (via the vascular access aperture) are so close to the AV fistula, a very high blood flow (e.g., approximately 400 ml/min) and pressure (compared to standard hemodialysis through a vessel relatively far away from an AV fistula) is provided. This allows patients and/or technicians performing the hemodialysis to reduce the capacity needed of a pump or possibly eliminate the pump that is required for normal hemodialysis. That is because the blood is naturally pressurized and pumped by the heart of the patient in an artery more than a vessel that is not connected to an artery via an AV fistula. In other words, the blood flow and pressure of a vessel that is directly connected to an AV fistula and is relatively close to that AV fistula will have greater blood flow and pressure than a vessel without an AV fistula or even a vessel with an AV fistula that is not relatively close to that AV fistula.
With reference to both
Advantageously, when the top portion 310 is coupled to the bottom portion 320, the structure of the top portion 310 and the bottom portion 320 around the vessel prevents or at least minimizes the chance of a bad puncture by preventing the needle from going somewhere unintended in the patient. In addition, the porous surface of the device 300 enables the device 300 to integrate into the surrounding tissues which allows the body to mount an immune response to any bacteria which are instilled during needle cannulation. Having the device integrate into the surrounding tissues removes bare surfaces for bacteria to grow on and develop a biofilm.
In some patients, depending on the part of the body that the AV fistula 530 is created and the physiology of the individual patient, the vessel 510 may be proximal to the surface of the patient's skin and the artery 520 may be distal from the surface of the patient's skin (i.e., the artery 520 is beneath the vessel 510 in relation to the surface of the patient's skin). Therefore, as explained below with respect to
As can be seen in
In
It should be understood that the angles 620, 640 illustrated in
Furthermore, because the needles 702, 704 withdraw blood (via the vascular access aperture) so close to the AV fistula, a very high blood flow and pressure (compared to standard hemodialysis through a vessel relatively far away from an AV fistula) is provided. This allows patients and/or technicians performing the dialysis to reduce the capacity needed of a pump or possibly eliminate the pump that is required for normal hemodialysis treatment. That is because the blood is naturally pressurized and pumped by the heart of the patient in an artery 710 more than a normal vessel; the vessel 708 is directly connected to the artery 710 through an AV fistula, thus creating pressure and flow in the vessel 708 that is similar to that of the artery 710.
The offset design may be useful in cases where damage to the AV fistula is of great concern to the physician and/or patient. Indeed, because the vascular access aperture 902 is offset from the AV fistula/arteriovenous joint 904, the likelihood of damage to the AV fistula from a bad puncture is further reduced. Furthermore, because of the reduced likelihood of damage to the AV fistula from a bad puncture, patients may have more confidence in administering in-home dialysis treatment themselves. It should also be understood that, except for that which is inconsistent with the apparatus and techniques described with respect to
In any case, the AV fistula 1040 is created and may be completely surrounded by the device 1010 so that a dialysis needle (or any other type of needle) cannot penetrate the AV fistula 1040. This allows a needle to provide a good puncture to the vessel 1020 of the patient through the vascular access aperture 1070 of the device 1010 without worry that the AV fistula 1040 will be damaged.
During implantation of the device 1010, the top portion 1012 and the bottom portion 1014 of the device 1010 are coupled together to form a vessel channel 1050 that can encase the vessel 1020 and an artery channel 1060 that can encase the artery 1030. The vessel channel 1050 provides structural support and external protection for a vessel 1020 of a patient. By providing structural support and external protection for the vessel 1020 of the patient, the vessel channel 1050 prevents a bad puncture, which also decreases the amount of trauma to the vessel 1020 of the patient and diminishes subsequent scarring and stenosis of the vessel 1020. The vessel channel 1050 also includes a vascular access aperture 1070 that is elongated parallel to the vessel 1020 of the patient. The vascular access aperture 1070 provides easy access for a dialysis needle to deliver a good puncture to the vessel 1020 of the patient and limits the risk of infection by utilizing the skin as a natural barrier to pathogens. Indeed, the vascular access aperture 1070 provides a greater surface area for a good puncture than other devices or no devices at all. This enables good punctures in different locations along the surface of the patient's skin, which helps prevent skin breakdown and gives the skin a chance to heal.
The artery channel 1060 provides structural support and external protection for the artery 1030 of a patient. By providing structural support and external protection for the artery 1030 of the patient, the artery channel 1060 prevents the artery 1030 from being punctured or damaged (e.g., through a bad puncture of the vessel 1020).
An AV fistula 1040 can be formed through a connection of the end of the vessel 1020 and a side of the artery 1030. The AV fistula 1040 is fully encased by the device 1010. The vessel channel 1050 and the artery channel 1060 can be any shape suitable to house a vessel 1020 and an artery 1030. Therefore, although the channels 1050, 1060 are illustrated as cylindrical, the cross-section perpendicular to the axis of the cylinder is not required to be a perfect circle or ellipse. Indeed, in other cases, the cross-sectional shape of the channels 1050, 1060 may be any shape and/or size to fully enclose portions of the vessel 1020 and the artery 1030 near the AV fistula 1040.
In some cases, the surfaces of the device 1010 are porous. As explained above with respect to
It should be understood that, except for that which is inconsistent with the apparatus and techniques described with respect to
Certain aspects of the invention provide the following non-limiting embodiments:
Example 1. A vascular access device comprising a top portion having a partial artery channel for receiving a top half of an artery, a partial vessel channel for receiving a top half of a vessel, and a partial arteriovenous joint between the partial artery channel and the partial vessel channel, the partial vessel channel having a vascular access aperture for exposing the vessel; and a bottom portion having a partial artery channel for receiving a bottom half of the artery, a partial vessel channel for receiving a bottom half of the vessel, and a partial arteriovenous joint between the partial artery channel and the partial vessel channel; wherein when the top portion is coupled to the bottom portion: the partial artery channel of the top portion and the partial artery channel of the bottom portion couple to one another to form an artery channel; the partial vessel channel of the top portion and the partial vessel channel of the bottom portion couple to one another to form a vessel channel; and the partial arteriovenous joint of the top portion and the partial arteriovenous joint of the bottom portion couple to form an arteriovenous joint for an arteriovenous fistula.
Example 2. The vascular access device of Example 1, wherein the partial artery channel of the top portion and the partial artery channel of the bottom portion are semi-cylindrical; wherein when the top portion is coupled to the bottom portion, the semi-cylindrical artery channel of the top portion and the semi-cylindrical artery channel of the bottom portion couple to one another to form a cylindrical artery channel.
Example 3. The vascular access device of Examples 1 or 2, wherein the partial vessel channel of the top portion and the partial vessel channel of the bottom portion are semi-cylindrical; wherein when the top portion is coupled to the bottom portion, the semi-cylindrical vessel channel of the top portion and the semi-cylindrical vessel channel of the bottom portion couple to one another to form a cylindrical vessel channel.
Example 4. The vascular access device of any of the preceding Examples, wherein the partial vessel channel of the bottom portion is elongated relative to the partial vessel channel of the top portion.
Example 5. The vascular access device of any of the preceding Examples, wherein the vascular access aperture is elongated parallel to a length of the vessel of a patient.
Example 6. The vascular access device of any of the preceding Examples, wherein surfaces of the top portion and the bottom portion are porous to allow for collagen to form around the top portion and the bottom portion after implantation.
Example 7. The vascular access device of any of the preceding Examples, wherein the vessel channel is curved towards the arteriovenous joint.
Example 8. The vascular access device of any of the preceding Examples, wherein the artery channel is curved towards the arteriovenous joint.
Example 9. The vascular access device of any of Examples 1-6, wherein the vessel channel extends perpendicular to the artery channel; and wherein the arteriovenous joint is disposed in a position where the artery channel abuts the vessel channel.
Example 10. The vascular access device of any of the preceding Examples, wherein an angle between a vertical axis perpendicular to the vascular access aperture and an axis running through center points of the vessel channel and the artery channel is greater than 45 degrees.
Example 11. The vascular access device of any of Examples 1-9, wherein an angle between a vertical axis perpendicular to the vascular access aperture and an axis running through center points of the vessel channel and the artery channel is between 30 degrees and 45 degrees.
Example 12. The vascular access device of any of Examples 1-9, wherein an angle between a vertical axis perpendicular to the vascular access aperture and an axis running through center points of the vessel channel and the artery channel is less than 30 degrees.
Example 13. The vascular access device of any of the preceding Examples, wherein the arteriovenous joint is offset from the vascular access aperture.
Example 14. A method of using the vascular access device of any of the preceding Examples includes creating an incision to insert the vascular access device, positioning the partial artery channel of the bottom portion of the vascular access device around the bottom half of the artery and the partial vessel channel of the bottom portion of the vascular access device around the bottom half of the vessel, positioning the partial artery channel of the top portion of the vascular access device around the top half of the artery and the partial vessel channel of the top portion of the vascular access device around the top half of the vessel to couple the top portion to the bottom portion, closing the incision used to insert the vascular access device and allowing the tissue surrounding the vascular access device to heal, and after the incision is healed, creating an AV fistula between the artery and the vessel without uncoupling the top portion of the vascular access device from the bottom portion of the vascular access device.
Although the subject matter has been described in language specific to structural features and/or acts, it is to be understood that the subject matter defined in the appended claims is not necessarily limited to the specific features or acts described above. Rather, the specific features and acts described above are disclosed as examples of implementing the claims and other equivalent features and acts are intended to be within the scope of the claims.
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/US2020/054169 | 10/3/2020 | WO |
Number | Date | Country | |
---|---|---|---|
62910984 | Oct 2019 | US |